lab 2 final (compiled)

Upload: snomasnayr

Post on 18-Oct-2015

21 views

Category:

Documents


0 download

DESCRIPTION

Physics 242 Lab for UK

TRANSCRIPT

  • 5/28/2018 Lab 2 Final (Compiled)

    1/21

    Page 1

    Lab: 2

    Section #: 004

    Group #: 5

    Experiment #: 2

    Date : 3/6/2014

    Oscilloscope

    Your signature indicates that you have completely read the entire report and agree witheverything here in. Failure to sign will result in a zero for your personal grade unless a formal

    exception is filed with your TA.

    Please Print and Sign Full Name

    Principal investigator: Ryan Samons_________________________________________

    Skeptic: Morgan Culver_________________________________________

    Researcher Logan Murphy

    _________________________________________

    TA Justin Woods__________________________________________

    Lab Report Score:________________

    Role I DC AD RC Q1 Q2 PI PG

    Legend:

    I introductionDC data and calculationAD analysis and discussion

    RC results and conclusion

    Q1/Q2 quiz/prelabPI principal investigator points

    PG personal grade (average of individual score and Lab Report Score)

  • 5/28/2018 Lab 2 Final (Compiled)

    2/21

    Page 2

    Introduction:

    After familiarizing ourselves with equipment during lab one, lab two focuses on learning

    how to use and becoming familiar with the oscilloscope. We learned how to look at the data

    displayed on the oscilloscopes screen and interpret that data to find other measurements. From

    this experiment will verify ohms law, and confirm that there are different ways to calculatevoltages but ohms law works for all the methods and the values calculated are not significantlydifferent between the methods used.

    Equipment:

    1 Digital Multimeter 91910044 1 Oscilloscope C050155 1 DC power supply C1010444 2 AC power supply (function generator) FG24243,FG24241 1 10-k resistor 5 Banana clip wires 2 BNC cables

    Post Lab Meetings:

    Monday February 24th

    there was a meeting in the lab to finish and finalize the data from

    the lab. Morgan and Logan were in attendance. Ryan had a class at the time of the office hours

    were the lab was completed therefore he was unfortunately unable to attend. The meeting lasted

    from approximately 11:00 am to 11:40 am.

    Thursday February 27th

    there was a meeting during the lab time. We discussed how we

    were going to proceed on the lab and discussed the results and comments given by the TA. This

    meeting lasted approximately 30 minutes.

    References:

    Ellis, Steven L. Laboratory Manual Spring 2013-2014 Appendix Fa - Tektronix_PWS4305 Appendix Fb - Tektronix_PWS4305

    Appendix H- Tenma function generator 72-7210 Appendix K - Digital Multimeter Appendix S - Tenma Function Generator72 7210 Manual Appendix JError propagation Appendix CTDS 100B Oscilloscopes users manual Standard Cover Sheet

  • 5/28/2018 Lab 2 Final (Compiled)

    3/21

    Page 3

    Procedure:

    The research team began the experiment by reviewing the lab manual and trying to learnthe general procedures of the lab before moving any further.

    Next, Ryan Samons and Morgan Culver retrieved the required equipment from the filingcabinet and Logan Murphy recorded the serial numbers of the equipment.

    Table 1

    The PI Ryan Samons referenced the lab manual for the proper setup of the equipment forthis part of the experiment. By placing cables between the oscilloscope and the DC

    power supply we were able to complete the circuit.

    These banana clips were ran through the resistor. The Digital Multimeter was then placed where it would read the voltage running through

    the system. At this point the DMM Voltage heading of Table 1 was completed by reading

    the DMM.

    Next the oscilloscope was used to record the scale and divisions of the data given in orderto find the voltage reading on that device.

    These steps were repeated for .5V,1V,2V,4V, and 8V. The percent difference were then calculated by Skeptic Morgan Culver and recorded in

    Table 1.

    The diagram below shows the setup for this part of the experiment.

    Table 2

    For this part of the experiment the DC power supply was replaced with the functiongenerator in order to find the same measurements using an AC power supply. All group

    members helped in switching these devices.

    The connection between the oscilloscope and function generator was made using BNC tobanana clip cables.

    The TA Mr. Wood was asked by Logan Murphy to check the setup before continuing.

  • 5/28/2018 Lab 2 Final (Compiled)

    4/21

    Page 4

    The DMM readings were taken in the same manner as before, with the DMM beingplaced on top of the banana wires in the experiment in order to accurately record the data.

    Next the oscilloscope was used to find the divisions and scale of the data given in orderto calculate the voltage readings for this part of the experiment.

    The percent difference was then calculated by the Skeptic Morgan Culver and recordedalong with the other data

    The table below shows the setup for this part of the experiment.

    Table 3

    The next part of the experiment focused on using the function generator and seeing theeffect the shape of the wave and the frequency strength had on the voltage given off.

    The same setup from table 2 was used in this part of the experiment also. Mr. Wood was asked by Morgan Culver to check the setup before preceding. First the data was recorded for all of the sine, triangle, and square waves at the 400Hz

    values, then the 4000Hz value, and the 40000Hz value.

    The divisions and scale factor data were gathered by Ryan Samons and Logan Murphyfrom the screen of the Oscilloscope itself. The divisions were the number of lines in

    between the two points needed.

    All the data was then recorded by the Skeptic Morgan Culver. The table below shows the setup for this part of the experiment.

    Table 4

  • 5/28/2018 Lab 2 Final (Compiled)

    5/21

    Page 5

    The last part of the experiment involved 2 function generators. The two function generators were connected to the oscilloscope. One in channel one and

    the other in channel two.

    Mr. Wood was asked by Ryan Samons to check the setup before preceding.

    Both were turned on and the frequency values set to about equal, one double the other,and one triple the other, with the data being recorded along the way.

    The data recorded were the number of loops along the Lissajous pattern. After findingthe display button we were able to find the proper display to show the data in this form.

    The number of loops were counted in the horizontal and vertical directions. The diagram below shows the setup for this part of the experiment.

    Data and Calculations:

    Table 1

    The DC power supply was used in this part of the experiment changing the voltage asdesired anywhere from .5V to 8V. The device was connected directly to the Oscilloscope and

    the reading came out to be in Volts as well. Below is the data collected.

    Power

    Supply

    DMM

    Voltage Oscilloscope

    %

    Difference

    Divisions Scale DC Voltage

    .5V .5V 2.5 200mV .5V 0

    1.00V 1V 2 500mV 1V 0

    2.00V 2V 2 1V 2V 0

    4.00V 4V 2 2V 4V 0

    8.00V 8V 1.6 5V 8V 0

    Although the percent difference equation does not apply here because the data was

    exactly the same the equation for percent difference will be used throughout the experiment and

    thus will be referenced in this section.

  • 5/28/2018 Lab 2 Final (Compiled)

    6/21

    Page 6

    Equation 1

    Below is a sample equation for Equation 1. Since the values from the table above all give

    the % to be zero, the sample calculation will be if the values were 4V and 4.2V.

    This next equation applies to all the numbers underneath the oscilloscope heading in the

    charts throughout the experiment. In order to find the DC voltage you must multiply the scale by

    the divisions.

    Equation 2 Sample equation using the row of the .5V.

    Table 2

    This part of the lab used the function generator to produce current using a sine wave and

    was measured both with the DMM and the oscilloscope. Before comparing the data the DMMvoltage reading had to be converted to the peak voltage readings and the divisions and Scale had

    to be found on the oscilloscope. The data for this part of the experiment is shown below.

    Function

    Generator

    Oscilloscope

    Digital Multimeter V(rms) Digital Multimeter V(pp) %differeDivisions Scale V(pp)

    min 0 db 5 .1V .5V .1738V .491V 1.81

    mid 0 db 6.4 1V 6.4V 2.21V 6.25V 2.37

    max 0 db 4.6 5V 23V 7.68V 21.73V 5.68

    mid -20 db 6.4 .1V .64V .22V .62V 3.17

    max -20 db 4.6 .5V 2.3V .795V 2.25V 2.2

    There a few equations needed for the completion of the table above. The first is the conversion

    of the Digital Multimeter V(rms) into the V(pp). This equation is listed below.

    Equation 3

  • 5/28/2018 Lab 2 Final (Compiled)

    7/21

    Page 7

    Sample equation for equation 2 will be used at the mid 0 db row of calculations above.

    The percent difference for this part of the experiment was found using Equation 1from

    earlier in the experiment. The V(pp) underneath the oscilloscope was calculated by simply

    multiplying the divisions and the scale together. Also under the oscilloscope in the table there is

    a divisions and scale column. Using Equation 2you find the values for the next column in the

    table.

    Equation 4

    Table 3In this part of the experiment the function generator was used to find the changes sine,

    triangular, and square waves affected the frequency readings. The table below shows the data

    gathered from this section of the lab.

    Table 3Wave

    form

    type

    Function

    generato

    r

    frequenc

    y

    Amplitude Peak-to-Peak

    Amplitude

    Period Frequ

    ency

    F

    (HZ)

    % di

    in fr

    Div Scale

    factor

    r (V)

    Volt

    s

    Div Scale

    factor

    r (V)

    Volts Div Scale

    factor

    r (ms)

    Sec.

    Sine 400 Hz 3.2 .1 .32 6.4 .1 .64 2.46 1 .00246 406.5 1.61

    Sine 4000 Hz 3.3 .1 .33 6.6 .1 .66 2.6 .1 .00026 3846.15 3.921

    Sine 40000 Hz 3.1 .1 .31 6.2 .1 .62 2.5 .01 .000025 40000 0

    Triangl

    e

    400 Hz 3.7 .1 .37 7.4 .1 .74 2.48 1 .00248 403.22 .8

    Triangl

    e

    4000 Hz 3.65 .1 .365 7.5 .1 .75 2.6 .1 .00026 3846.15 3.92

    Triangl

    e

    40000 Hz 3.75 .1 .375 7.5 .1 .75 2.6 .01 .000026 38461.5 3.92

    Square 400 Hz 2.75 .2 .55 5.5 .2 1.1 2.46 1 .00246 406.5 1.611Square 4000 Hz 2.7 .2 .54 5.4 .2 1.08 2.5 .1 .00025 4000 0

    Square 40000 Hz 2.25 .2 .45 5.5 .2 1.1 2.5 .01 .000025 40000 0

    The percent difference is found in this table by once again using Equation 1. The two

    volts categories under amplitude and peak to peak were calculated by taking the divisions and

    multiplying them by the scale factor. Another equation used in this chart is Equation 2. This is

  • 5/28/2018 Lab 2 Final (Compiled)

    8/21

    Page 8

    used by multiplying the divisions and scales in order to get the values. This just leaves the

    calculated frequency. Calculating frequency has the following equation.

    Equation 5

    The sample calculation for this equation will be using the information in the square

    40000Hz row. Table 4

    The last part of the experiment involved using two function generators changing the

    frequency of one of them and creating Lissajous Pattern. We had to set up the oscilloscope in

    the XY mode and looked at the ratio of vertical loops to horizontal loops. The chart below was

    created after looking at different ratios of frequencies given off by the function generator.

    Frequency ratio Number of Vertical

    Loops

    Number of Horizontal

    Loops

    Ratio

    1/1 1 1 1/1

    1 2

    1/3 1 3 1/3

    The picture below shows the oscilloscope at the time when both function generators were

    set at 400Hz. As you can see there is one vertical and one horizontal loop.

  • 5/28/2018 Lab 2 Final (Compiled)

    9/21

    Page 9

    The next picture shows the oscilloscope when one of the function generators is at double

    the frequency of the other. As you can see there is two horizontal loops and one vertical loop.

    The last ratio tested was when one function generator was 3 times the other. As you can

    see there is three horizontal loops and one vertical loop.

  • 5/28/2018 Lab 2 Final (Compiled)

    10/21

    Page 10

  • 5/28/2018 Lab 2 Final (Compiled)

    11/21

    Page 11

    Discussion and Analysis

    Throughout this entire experiment there were errors in the measurements taken. Some of

    these were measurable while others were more random and unmeasurable. Each piece of data

    collected will have a certain measurable or unmeasurable amount of error. This error is a result

    from the error in the equipment used during the experiment as well as random human error.These errors can be analyzed by breaking them down into four different sections, ranking them,

    and then further discussing systematic versus random error. The first sources of error occurred in

    part two of this experiment.

    Part 2: Measurement of DC Voltage with the Oscilloscope

    The error in this section of the lab will affect the data values collected in Data Table 1.

    Each piece of equipment used in this section will have a certain amount of error. These pieces of

    equipment include the digital multi-meter, power supply, and the oscilloscope. The oscilloscope

    will have sub errors that must be propagated such as the error in divisions, scale, and DC voltage.

    The systematic error for the power supply can be found using the following equation, where: Vpsis the error in the power supply output measured in volts

    : Vpsis the voltage setting on the power supply

    : [2] is the number added to the smallest significant digit of the voltage of the power

    supply.

    Below is a sample equation of the error for the Power Supply using the voltage at 0V.

    By using this equation, the error regarding the power supply can be found for each individual

    measurement taken. These errors can be found in Errors in Table 1.

    The systematic error in the digital multi-meter can also be found using a similar equation show

    below,

    where: V is the digital multi-meter voltage measurement.

    :is the total error in the power supply: [1] is the number added to the smallest significant digit

  • 5/28/2018 Lab 2 Final (Compiled)

    12/21

    Page 12

    Because the digital multi-meter is measuring the power supply voltage, the error in the

    digital multi-meter will also include the error in the power supply as seen above. Below is a

    sample equation of the error for the DMM device using the voltage at 0V.

    The error in the oscilloscope will also affect the measurements found in Data Table 1.

    The error regarding this piece of equipment is more complex than the previous two. The total

    error in the DC voltage measurements taken has three different components. The first two are the

    error in the number of divisions and the error in the scale factor. The error in the divisions was

    given as0.1 because it is designated as half of the smallest possible measurement. The error inthe scale factor will be as 0.3% of the scale factor. In order to find the total error in DCmeasurement, these two values are multiplied together. Because of this, the total error must be

    propagated using the following equation:

    This equation can be simplified, however, to an even simpler equation show below. The

    partial derivative of the voltage with respect to the divisions is just S. The partial derivative of

    the voltage with respect to the scale will be D. These can be substituted into the equation and

    then it be further simplified,

    where:V is the error in the voltage

    : S is the scale factor

    : D is the error in the divisions: D is the number of divisions

    : S is the error in the scale

    Similar to the digital multi-meter, the error in the oscilloscope will also inherit the error

    in the power supply. Because of this, the total error in the voltage measured by the oscilloscope

    can be found using the equation below. Below is a sample calculation where S is .2V with an uncertainty of .006, a division of 2.5 and

    an uncertainty of .1, and the error in the power supply is .045V.

  • 5/28/2018 Lab 2 Final (Compiled)

    13/21

    Page 13

    Error in Table 1

    Power Supply

    Setting

    Error inPower

    Supply

    Voltage

    Error inDMM

    Voltage

    Reading

    Oscilloscope

    Error in

    Divisions

    Error in

    Scale

    Error in DC

    Voltage

    0.50V .045 .07 .1 .0006 .065

    1.00V .07 .11 .1 .0015 .12

    2.00V .12 .19 .1 .003 .22

    4.00V .22 .35 .1 .006 .42

    8.00V .42 .67 .1 .015 .92

    Part 3: Comparison of AC voltages measured with either the oscilloscope or DMM

    The errors regarding the section of the lab will be similar to those seen in part 2 of the

    lab. The oscilloscope and digital multi-meter were used and will have the same measurable error.

    The error for measurements taken using the digital multi-meter will be slightly different,

    however, because now it is being used to measure the root mean square voltage and peak to peak

    voltage of AC voltages. A new piece of equipment was introduced in this section of the lab as

    well. For this part the function generator was used as well and it will have a measurable amount

    of error. All systematic error that will affect the measurements taken can be found in Errors in

    Table 2.

    The error in the digital multi-meter will have two different measurable errors due to the two

    different ways it was used in this section of the lab. The first error that can be calculated in the

    error in root mean square measurements. This error can be found using the simple equation, where: Vrmsis the total error in the measurement

  • 5/28/2018 Lab 2 Final (Compiled)

    14/21

    Page 14

    : Vrmsis the original measurement

    : [2] is the number added to the smallest significant figure.

    Below is a sample calculation where the peak to peak voltage is .1738V.

    The other measurable error for the digital multi-meter is the error in measuring peak to peak

    voltage. This error can be found using a simple process. Because the peak to peak voltage is

    found by multiplying the root mean square measurement by 22, the error in peak to peakvoltage is simply:

    =* 2Below is a sample calculation with Vrmsis .=.0037* 2=.0105The measurements taken using the oscilloscope will also have a certain amount of measurable

    error. These can be found by solving the following equation for peak to peak voltage.

    In this equation the variables represent the same values that they did for the error in the

    oscilloscope measurement in the previous section. A sample calculation using this equation can

    be seen below where the function generator is set to min 0 db, the scale is .1V, the uncertainty in

    divisions is .1, the number of divisions is 5 and the uncertainty in the scale is .0003V.

  • 5/28/2018 Lab 2 Final (Compiled)

    15/21

    Page 15

    Error in Table 2

    Function

    Generator

    Oscilloscope Error in

    DMM Vrms

    Error in

    DMM VppError in

    Divisions

    Error in

    Scale

    Error in Vpp

    Min 0 db .1 .0003 .01 .01Mid 0 db .1 .003 .1 .0642 .18Max 0 db .1 .015 .5 .1736 .491

    Mid -20 db

    .1 .0003 .01 .0244 .069

    Max -20 db .1 .0015 .05 .0179 .051Part 4: Measurement AC voltage for Different Waveforms

    This section of the lab will have similar errors as in the previous section along with new

    errors regarding the error in time and frequency. The error in the divisions and scale factor will

    be the same as in part 3 on the lab. These can be found in Error in Table 2. The error in seconds

    can be found using a similar equation used to find the error in DC voltage in part 2 of the lab.

    The main difference in the equation below is that this equation is tailored to seconds instead of

    volts,

    where: S is the seconds

    : D is the error in the divisions

    : D is the number of divisions

    : S is the error in the seconds

    A sample calculation can be seen below where the scale factor is .001s, the error in the divisionsis .1, the number of divisions is 2.46, and

  • 5/28/2018 Lab 2 Final (Compiled)

    16/21

    Page 16

    The error in frequency can also be calculated using a similar equation. The main difference isnow the derivative of frequency is taken with respect to the divisions and the scale.

    Because seconds is equal to the divisions multiplied by the scale, this equation can be

    rewritten by replacing the two partial derivatives. The equation for error in frequency is

    now:

    A sample calculation using this equation is shown below where the number of divisions in 2.46,

    the scale is .001, the error in divisions in .1, and the error in the scale is 3.0E-6.

    There will also be a measurable error in volts in amplitude for this section of the lab. This error

    can be calculated using the equation below.

  • 5/28/2018 Lab 2 Final (Compiled)

    17/21

    Page 17

    This equation is the same one used to calculate the total error in voltage in part 2 of the lab. A

    sample equation can be seen below.

    The error in peak to peak amplitude can be found using the same equation seen in part 3 of the

    lab. =* 2A sample calculation regarding this section of the lab can be seen below.

    =* 2=.01* 2Vpp=.028

    Error in Table 3Waveform

    type

    Function

    generator

    frequency

    Amplitude Peak-to-Peak

    Amplitude

    Period Error

    Frequ

    y

    F (HZError

    in Div

    Error

    in

    Scalefactor

    r (V)

    Error

    in

    Volts

    Error

    in Div

    Error

    in

    Scalefactor

    r (V)

    Error

    in

    Volts

    Error

    in Div

    Error

    in

    Scalefactor

    r (ms)

    Error in

    Sec.

    Sine 400 Hz .1 .0003 .01 .1 .0003 .028 .1 .003 .0001 16.77Sine 4000 Hz .1 .0003 .01 .1 .0003 .028 .1 .0003 .00001 167.7Sine 40000 Hz .1 .0003 .01 .1 .0003 .028 .1 .0000

    3

    .000001 1677

    Triangle 400 Hz .1 .0003 .01 .1 .0003 .028 .1 .003 .0001 16.77

  • 5/28/2018 Lab 2 Final (Compiled)

    18/21

    Page 18

    Triangle 4000 Hz .1 .0003 .01 .1 .0003 .028 .1 .0003 .00001 167.7Triangle 40000 Hz .1 .0003 .01 .1 .0003 .028 .1 .0000

    3

    .000001 1677

    Square 400 Hz .1 .0006 .02 .1 .0006 .057 .1 .003 .0001 16.77Square 4000 Hz

    .1 .0006 .02

    .1 .0006 .057

    .1 .0003 .00001 167.7

    Square 40000 Hz .1 .0006 .02 .1 .0006 .057 .1 .00003 .000001 1677Random vs. Systematic Errors

    The systematic errors in this experiment are the ones that can be calculated and

    mathematically accounted for. The random errors in this experiment are those that cant such ashuman errors. One example of a random human error seen in this experiment that the function

    generator would never stay at a set value for frequency. Because it would randomly change, it

    cannot be accounted for mathematically. There are also unknown human errors in this lab that

    will affect the results. The majority of the error seen in this lab can be broken down into the twocategories.

    Error Types

    1. Random errors. These are errors that you cannot account for exactly but you know theyexist throughout the experiment.

    The function generator fluctuated in its Hz given off. The function generator didnt have a knob to stop at exactly the MID. The research team had faulty cables that may have affected the results. Other unknown sources of error not calculable.

    2. Systamatic errors. These are errors that we know occur and have specific equations inorder to find the exact value for the error.

    The error in the power supply including that for root mean square and peak topeak.

    The error in the digital multi-meter. The error in the oscilloscope readings including the scale and divisions. The error in frequency. The error in time.

  • 5/28/2018 Lab 2 Final (Compiled)

    19/21

    Page 19

    Results and Conclusions

    The purpose of this lab was to become familiar with using the oscilloscope. One

    advantage that the oscilloscope has when compared to the DMM we learned to use in Lab 1, is

    that the oscilloscope displays a graph of the voltage as a function of time. This allows us to view

    how the voltage changes over a longer period of time vs. the DMM which shows the voltage

    using the root-mean-squared method. One advantage the DMM has is that it can show a more

    precise measurement of the voltage at the instant it is measuring.

    During the first part of the lab, we measured the DC voltage. The error has been

    calculated and organized in the Error in Table 1 table. Below is a sample comparison of the

    DMM vs. Oscilloscope measurements for the power supply setting of 2 volts.

    Number Line 1:

    We can see from the number line that the DMM and Oscilloscope voltage overlap (they

    were stacked to aid viewing). This shows that these instruments can be trusted to produce an

    accurate value for the voltage.

    Now the AC voltage is measured. The AC voltage is provided by the function generator.

    Below is a number line that compares the oscilloscope, and the DMM measurements for the

    voltage. The number line is based on the 400 Hz and 0 db setting for the function generator. The

    errors are taken from the Error in Table 2 table from the skeptic.

    Number Line 2:

    0 0.5 1 1.5 2 2.5

    DMM vs Oscilloscope (2 V)

    DMM 2 0.19V

    Oscilloscope 2 0.22V

    0.475 0.48 0.485 0.49 0.495 0.5 0.505 0.51 0.515

    DMM vs Oscilloscope (Min 0 db)

    DMM 0.491 0.01V

    Oscilloscope 0.5 0.01

  • 5/28/2018 Lab 2 Final (Compiled)

    20/21

    Page 20

    Here we can see that the DMM and the Oscilloscope measurements for voltage overlap.

    The percent difference between the two measurements was 1.81%. This, plus the overlapping

    error bars, shows that the oscilloscope and the DMM can be used to accurately determine the

    voltage. We can conclude because these values overlap, as well as the overlap from Number

    Line 1, that the adjusting of the DC power supply and the Function generator using only the front

    control knobs can yield a precise output from either unit.

    This next section compares the different frequencies of the function generator and that of

    the oscilloscope. The uncertainty in the frequency could be calculated by defining the uncertainty

    to be that of the error of the inverse of the period. The first number line in this section will be

    that of the sin at 400 Hz. The frequency measured by the oscilloscope is the one to trust in this

    situation because setting the frequency on the function generator is difficult because of the

    sensitivity of the knob on the front of the unit. It is also important to note that one should try to

    fit the entire period onto the screen of the oscilloscope to allow for a more accurate

    measurement. If the entire period was not on the screen, it would be difficult to determine where

    the period began and ended. The errors came from the Error in Table 3 table in from the skeptic.

    Below is the comparison for the sin wave at 400 Hz vs. the function generator at 400 Hz.

    Number Line 3:

    We can see that the error bounds on the oscilloscope's measurements overlapped the

    frequency of the function generator. This shows that the oscilloscope can produce an accurate

    measurement for the frequency from the function generator; however, it may not be a precise

    measurement.

    Below is the comparison for the sin wave at 4000 Hz vs. the function generator at 4000 Hz.

    380 390 400 410 420 430

    Function Generator vs Oscilloscope (sin 400 Hz)

    Function Generator 400Hz

    Oscilloscope 406.5 16.77Hz

  • 5/28/2018 Lab 2 Final (Compiled)

    21/21

    Page 21

    Number Line 4:

    Because the oscilloscope's measurement and errors overlapped, we can draw the same

    conclusion as we did from number line 4, that is, that the oscilloscope can return an accurate

    measurement of the frequency. One thing to notice is that the measured frequency of 3846.15 Hz

    is lower than the frequency of 4000 Hz we were supposed to compare the oscilloscope to. This is

    probably due to a random error or a mistake in setting the function generator's frequency. The

    random errors and systematic errors were discussed in the Skeptic's portion of this paper.

    Below is the comparison for the sin wave at 40000 Hz vs. the function generator at 40000 Hz.

    Number Line 5:

    We can see again that the oscilloscope's measurement and errors overlap the frequency of

    the function generator. Based on this, we can again conclude that the oscilloscope will produce

    an accurate measurement of the frequency.

    Conclusion:

    In this lab, we learned how to use the oscilloscope and accompanying software, althoughnot required, to determine voltage and frequency. We learned about using the oscilloscope to

    measure AC and DC voltages. We learned how to measure the period and frequency of a signal

    using the oscilloscope, and we learned how to make a Lissajous figure using two AC voltage

    input signals. This lab also introduced us to the oscilloscope, which allows us to make a more in-

    depth analysis of electrical signals for future labs.

    3600 3700 3800 3900 4000 4100

    Function Generator vs Oscilloscope (sin 4000 Hz)

    Function Generator 4000Hz

    Oscilloscope 3846.15

    167.7Hz

    38000 39000 40000 41000 42000

    Function Generator vs Oscilloscope (sin 40000 Hz)

    Function Generator 40000Hz

    Oscilloscope 40000 16.77

    Hz