Lab Safety Design Standard
Preface This document has been assembled by the Vancouver Island University, Department of Health and
Safety Services. This document specifically pertains to the built environment of laboratories and not
operational practices. This document should be read and understood by all parties involved in lab
development. Please refer to the Roles section to learn more.
Scope These are design and construction standards to be applied to new construction and renovation of
facilities containing teaching and research laboratories and clinics at Vancouver Island University. For
renovations of existing laboratories or new laboratories in existing buildings not all standards will be
practical to implement, such as changing existing building ventilation systems. For this and other
provisions H&SS will provide exemptions on a case by case basis. H&SS must be consulted in the
commissioning of new labs and all renovations to existing labs.
Roles
Health and Safety Services Establishes standards to ensure compliance with legislation governing laboratory activities
within all of the following: the applicable municipality, the province of British Columbia,
and Canada. H&SS ensures that compliance is met when commissioning of laboratories is
executed.
Principal Investigators (PI) and faculty Determines the research activities to be carried out in the lab. Principal Investigators must
ensure that the design of lab space complies with the requirements of university, provincial
and federal safety legislation applicable to the research activities. Refer to the Definitions
section to find some of the applicable standards.
Space Designers (Project Manager, Project Planners and/or Architects) Coordinates the requirements of this standard, other applicable standards and/or guidelines
with H&SS, the Faculty or Principle Investigators and the contracted constructor, designer or
other provider. Space Designers are to facilitate the requirements indicated above, the needs
of the Faculty or PI and costs in execution of the laboratory construction.
2 | P a g e
Regulations and Standards Human Pathogens and Toxins Act Canadian Biosafety Standard, 2nd Edition:
http://canadianbiosafetystandards.collaboration.gc.ca/cbs-ncb/index-eng.php
Canadian Biosafety Handbook, 2nd Edition:
http://canadianbiosafetystandards.collaboration.gc.ca/cbh-gcb/index-eng.php
Nuclear Safety and Control Act Regulations: http://nuclearsafety.gc.ca/eng/acts-and-regulations/regulations/index.cfm
GD-52: https://cnsc-ccsn.gc.ca/eng/acts-and-regulations/regulatory-
documents/published/html/gd52/
BC Worker’s Compensation Act and Regulations Act: https://www.worksafebc.com/en/law-policy/occupational-health-safety/searchable-ohs-
regulation/workers-compensation-act/workers-compensation-act
Regulations: https://www.worksafebc.com/en/law-policy/occupational-health-
safety/searchable-ohs-regulation/ohs-regulation
BC Fire Code
BC Building Code
Definitions and Abbreviations
ANSI American National Standards Institute
ASHRAE American Society of Heating, Refrigerating and Air-Conditioning Engineers
BCBC BC Building Code
CNSC Canadian Nuclear Safety Commission
CCAC Canadian Council on Animal Care
CL Containment Level
Minimum physical containment and operational practice requirements for
handling infectious material or toxins safely in laboratory and animal work
environments. There are four containment levels ranging from a basic
laboratory (CL1) to the highest level of containment (CL4).
CIFA Canadian Food Inspection Agency
CSA Canadian Standards Association
HEPA High Efficiency Particulate Air
HVAC Heating, Ventilation and Air Conditioning
PHAC Public Health Agency of Canada
WCA Worker’s Compensation Act (BC)
3 | P a g e
Table of Contents
Contents Preface........................................................................................................................................................... 1
Scope ............................................................................................................................................................. 1 Roles ............................................................................................................................................................. 1
Health and Safety Services ....................................................................................................................... 1 Principal Investigators (PI) and faculty .................................................................................................... 1 Space Designers (Project Manager, Project Planners and/or Architects) ................................................. 1
Regulations and Standards ............................................................................................................................ 2 Human Pathogens and Toxins Act ............................................................................................................ 2
Nuclear Safety and Control Act ................................................................................................................ 2 BC Worker’s Compensation Act and Regulations ................................................................................... 2 BC Fire Code ............................................................................................................................................ 2 BC Building Code ..................................................................................................................................... 2
Definitions and Abbreviations ...................................................................................................................... 2 Standards ....................................................................................................................................................... 5
1.0 Containment Level 1: Architectural.................................................................................................... 5 General Laboratory Design .................................................................................................................. 5 Finishes ................................................................................................................................................. 6
Furnishings ........................................................................................................................................... 6 Storage and Equipment ......................................................................................................................... 6
2.0 Containment Level 1: Mechanical Requirements ............................................................................... 8
HVAC .................................................................................................................................................... 8
Plumbing ............................................................................................................................................... 9 2.1.2 Sinks ................................................................................................................................................ 9
2.2.3 Eye wash ............................................................................................................................................ 10 3.0 Containment Level 1: Electrical ........................................................................................................ 10
Light fixtures ....................................................................................................................................... 10
Power .................................................................................................................................................. 10 4.0 Containment Level 2: Architectural.................................................................................................. 10
General Laboratory Design ................................................................................................................ 10 Finishes ............................................................................................................................................... 11
Furnishings ......................................................................................................................................... 12 Storage and Equipment ....................................................................................................................... 12
5.0 Containment Level 2: Mechanical Requirements ............................................................................. 13
HVAC .................................................................................................................................................. 13 Plumbing ............................................................................................................................................. 14
Lab Type Specific Requirements ............................................................................................................ 16 High hazard Chemical ........................................................................................................................ 16
Radiation ............................................................................................................................................. 16 Engineering ......................................................................................................................................... 16 Laser ................................................................................................................................................... 17
X-Ray................................................................................................................................................... 17 Power Supply and Servers .................................................................................................................. 17
4 | P a g e
Sterilization Rooms ......................................................................................... 17
8.0 Offices: Seating Requirements ......................................................................................................... 17 Chair requirements ............................................................................................................................. 18 Intensive computer use/sitting (5 or more hours a day) ..................................................................... 18 Moderate computer use/sitting (3-5 hours a day)............................................................................... 18 Comfort ............................................................................................................................................... 18
Ease of adjustment .............................................................................................................................. 18 Warranty ............................................................................................................................................. 19
5 | P a g e
Standards
1.0 Containment Level 1: Architectural General Laboratory Design
1. Laboratories shall have separations from public areas by a secured door with a non-
duplicable key (ensures authorized entry only)
2. Laboratories shall have door openings with a clear width of 36” (915mm) or more
where possible and a minimum clear width of 32”(813mm) to allow passage of all
anticipated equipment and for persons with a disability.
3. Doors shall be equipped with door closers.
4. Doors shall have lever type handles, preferably J-Shaped
5. Doors should have vision panels to prevent collisions and accommodate for persons
with disabilities. The vision panel should include a decal or strip, at eye level, and
provide contrast for the surrounding area to increase safety and visibility.
6. Doors shall not be frameless glass
7. Locations must be provided for mounting emergency posters and other warning signs
immediately outside the lab. Preferred locations are on door between 30”(762mm)
42”(1067mm) high or located beside the door on the handle side between
30”(762mm) 42”(1067mm) high.
8. Laboratories shall have walls that extend to the ceiling from public/common areas or
at containment barrier location.
9. Laboratories should provide reasonable protection against vermin and insects. This
includes sealing or screening breaks in the containment barrier such as windows,
wiring, pipes and conduit penetrations.
10. Office areas to be located outside of containment laboratory. Paperwork stations for
data collection can be within containment laboratory provided they are located away
from laboratory work areas ideally this work should be separated from the bench
space used for experimentation to assist in preventing contamination of office areas.
11. Clothing and PPE donning and doffing areas should be accommodated at the exit
from the laboratory.
12. Supply and waste removal activities should be located at the lab perimeter.
6 | P a g e
Finishes
1. Laboratory flooring should be continuous, non-porous and impervious surface.
Preferred Homogenous floor with a static coefficient of friction of 0.6 greater to
ASTM C1028-07 - Static Coefficient of Friction Testing.
2. Baseboards shall be provided between floor and wall and should be sealed and
continuous to prevent spills from penetrating under the flooring.
3. Carpeting must not be used.
4. All surfaces shall be washable, non-absorptive and easy to decontaminate in the
event of a spill. Exposed wood grains should be avoided.
5. Walls must be finished with a smooth and washable surface
6. T-Bar Ceilings, if provided, must be flush profile to reduce possible contamination
locations, also tiles are to have acoustic qualities.
Furnishings
1. Work Bench surfaces should be scratch, stain, moisture, and heat resistant in
accordance with laboratory function. Examples of surfaces: Stainless steel (316
grade), 2 part chemical resistant epoxy paints, Chemical and abrasive resistant
laminates, Solid Epoxy.
2. Surfaces should provide impact resistance in accordance with laboratory function.
Surfaces should be compatible with adjacent and overlapping materials in order to
maintain a continuous surface and perimeter.
3. The counter top should include a lip to prevent run-off onto the floor. If the
countertop abuts a wall, it will either be coved or have a back-splash against the wall
that is adequately sealed at the joint between the countertop and the wall.
4. Adequate clearance in aisles 48”-60” (1200mm to 1500m) between benches and other
work stations, to accommodate persons with a disability; more distance may be required.
5. Normal working height for benches is 900 mm. Height adjustable benches preferred to
accommodate wheelchairs and other ergonomic considerations. For writing purposes 760
mm is the normal height. Average counter height variations may occur due to unleveled
floors and different casework manufactures.
6. Impervious, chemically resistant furniture in the working area. Finishing for chairs
to be easy to decontaminate if a spill is to occur. No porous fabrics to be used.
Leather, vinyl, stainless steel, or hard plastic only for chair seats.
7. Adjustable chairs for ergonomic considerations. Accessibility should be addressed in
general laboratory design.
8. Adequate/ample storage on shelves should be provided to prevent improper storage,
which can lead to trip and egress hazards.
9. Consideration should be given to restricting the height of shelving should ideally to
96”(2440mm) from the ground
Storage and Equipment
Flammables
1. All flammable storage shall comply with the BC Fire Code.
2. The maximum amount of flammable and combustible liquids that can be stored in a
room outside of a flammable/combustible storage cabinet is:
7 | P a g e
• 10 litres, of which 5 litres can be a
flammable liquid (Class I liquid) per room.
• 250 L, including not more than 60L of a combustible liquid
(Class II), or 10L of a Class I liquid, in a single fire compartment
having at least 45 min fire separation.
3. In the automotive shops or industrial arts areas, storage up to 75L of flammable liquids and
combustible liquids, including not more than 25L of a Class I liquid, shall be permitted
outside of a cabinet or designated Class I or Class II storage room.
4. Labs must be equipped with an adequate numbers of flammable cabinets. Cabinets may be
vented or sealed cabinets.
5. Special consideration is needed for flammable compressed gases. In general, one should
limit what can be stored in the lab. Additional storage spaces for compressed gases should
be provided within the facility.
6. Explosion proof refrigerators are required for cold storage of flammables.
7. Note that there may be considerable additional requirements if large quantities of
flammable gasses are to be used (e.g. hydrogen)
Gas Cylinders
1. Facilities must include provision for safe storage of unused gas cylinders.
2. Adequate physical support is required for gas cylinders in use
3. Additional ventilation may be required for toxic compressed or liquefied gasses.
4. Shipping areas must include a holding area for receipt of new cylinders.
5. Systems for securing gas cylinders while in use should be available
General Chemical Storage
1. Chemicals should be stored in accordance with the appropriate separation of chemical
classes.
2. Adequate provision for separation of incompatible chemicals.
3. Ventilation is deemed adequate as per appropriate codes of practice
4. Separate storage for acids includes corrosion resistant shelving and secondary
containment. Special acid storage cabinets should be provided
Waste Disposal and Storage
1. Unneeded chemical, biological and radioactive materials must be disposed of in
accordance with the all federal, provincial and municipal regulations and bylaws. Design
considerations to aid compliance include:
• Separate hazardous waste storage room located near load bay/shipping-
receiving area
• Designed with specialized ventilation as per the BCFC and BCBC
• Adequate floor space and shelving units for waste containers that are
constructed from materials compatible with the materials being held for
disposal.
• Space allocation is to be determined by the facility users and will be
based on the activities in the facility. Take into consideration the size of
the bins required to handle the volume of waste generated between waste
pick up cycles.
8 | P a g e
2.0 Containment Level 1: Mechanical Requirements HVAC
1.1 All air exhausted from the lab will be exhausted to the outside of the building
without recirculation.
1.2 Fume hoods are required for hazardous materials that can produce a hazardous
atmosphere of vapour, gas, particles, fume, mist or combustion products.
1.3 Functioning alarms (approved by EHS) such as Ventalert alarms, are required
with all installations of constant volume hoods. Note that the factory installed
alarms on the fume hoods are not acceptable except on a case-by case basis, and
where approved in written by EHS.
1.4 Variable air volume systems must have alarms on each hood and must be
approved by EHS.
1.5 Fume hood face velocity must be adjusted to be within the range of 80 – 130 feet
per minute before the space is used by the intended occupants. ASHRAE 110
validation of fume hood performance should be included as part of the project
cost.
1.6 Low velocity fume hoods may be considered on a case by case basis
1.7 Recirculating fume hoods are not acceptable in the place of a chemical fume
hood.
1.8 Location of fume hoods should be away from doors, high velocity diffusers, and
high traffic areas.
1.9 Ventilation shall have:
• 100% exhaust to the outdoors;
• 10 air exchanges per hour for biohazard labs;
• 4-12 air exchanges per hour for chemical labs (per ASHRAE);
• 6 air exchanges per hour for radiation labs;
• Negative pressure with respect to the surrounding areas
such as offices and hallways;
• an alarmed vacuum gauge or equivalent to warn of prolonged
failure of negative pressure in the lab;
• air flow from the area of low hazard activity to the areas of
higher hazard activity;
• fume hoods located away from the doors;
• thermostatically controlled HVAC systems located in each lab
room.
1.10 Negative pressure is required in newly built labs; and renovated biological
containment level 2, chemical containment 2 or higher labs. Exceptions will be
considered on a case-by-case basis with approval from EHS.
1.11 In special circumstances higher rates of air exchange may be needed. Examples
include bench top use of higher hazard materials, or where the potential for high
hazard spills exists.
1.12 HEPA filtration on local exhaust ventilation systems are recommended for
nanotech labs.
1.13 For Biosafety containment level 2 and nanotech labs 100% fresh air is
recommended.
1.14 Climate controlled rooms (e.g. cold rooms) should have separate temperature
9 | P a g e
controls and adequate air exchange rates.
Plumbing
2.1.1 Showers 1.1 All shower installations must include piped in emergency showers in labs
1.2 Drainage for emergency showers is required by the building code.
1.3 Emergency showers should be located near the door and within a 10 second
walking distance from the hazardous activity.
1.4 Emergency shower placement in corridors, to serve a number of labs is
discouraged, and only approved by EHS in exceptional circumstances.
1.5 Special consideration provided; rooms with electrical equipment may be
exempted from shower installation
1.6 Drench hoses are only accepted in a grandparented situation and should not be
used in new or renovated labs
2.1.2 Sinks 1.1 A separate hand washing sink should be placed within 15 feet of the door to allow
hand washing before exiting the lab.
1.2 A separate utility/glassware washing sink should be provided.
1.3 Sinks must be stainless steel. or an epoxy chemical resistant coating
1.4 Pipes and fixtures must be chemically resistant. Glass pipe traps should be
specified in the design stage. The height of the trap must be sufficient to prevent
the negative pressure in the lab from allowing gasses to escape from the drain
line (off-gassing).
1.5 Appropriate back-flow prevention should be included in the design.
1.6 Each sink will have an overflow outlet.
10 | P a g e
2.2.3 Eye wash 1.1 All eyewash installations must follow the VIU “Emergency Eyewash and
Shower Standard”: - see Appendix A.
1.2 A plumbed tepid eyewash is required for new installations. Tepid installations
require hot and cold water with a mixing valve to allow for a full 15 minutes of
flushing.
1.3 An emergency eyewash station will be provided in the room or in close proximity
to the room.
1.4 Tap fixture eye wash will not be accepted; separate plumbing is required.
3.0 Containment Level 1: Electrical Light fixtures
1.1 Adequate lighting as per international lighting standards (IEE).
1.2 Provisions to optimize natural lighting should be considered at design stage
1.3 Emergency lighting must be provided within the room.
1.4 Within radiation laboratories the lighting fixtures must be washable
Power 1.1 Sufficient outlets adequately spaced to prevent the use of power cords.
1.2 GFI circuit interrupters required as per code and within 36”(915mm) of sinks.
1.3 Power outlets required at the end of the bench closest to the corridor to provide cleaning staff easy
access to power.
• Emergency power outlets should be provided in each lab and properly spaced to
run equipment that requires continuous power (refrigerators, freezers, incubators,
etc.)
• Provision for internet and/or server access (preferably wireless) should be
considered in the wet lab and office spaces
4.0 Containment Level 2: Architectural All Containment Level 2 standards here are to be on top of Canadian Biosafety Standards.
General Laboratory Design
1.1 Windows positioned on containment barrier to be sealed in place; window glazing
material to provide required level of security
1.2 Laboratories shall have separations from public areas by a secured door with a
non-duplicable key (ie Authorized users)
1.3 Laboratories shall have door openings with a clear width of 36” (915mm) or more
where possible and a minimum clear width of 32”(813mm) to allow passage of
all anticipated equipment and for persons with a disability.
1.4 Doors shall be equipped with door closers.
1.5 Doors shall have lever type handles, preferably J-Shaped
1.6 Doors shall not exceed 22N of force required to open.
11 | P a g e
1.7 Doors should have vision panels to prevent collisions and
accommodate for persons with disabilities. The vision panel should include a
decal or strip, at eye level, and provide contrast for the surrounding area to
increase safety and visibility.
1.8 Doors shall not be frameless glass
1.9 Locations must be provided for mounting emergency posters and other warning
signs immediately outside the lab. Preferred locations are on door between
30”(762mm) 42”(1067mm) high or located beside the door on the handle side
between 30”(762mm) 42”(1067mm) high. Signage must meet all standards from
the governing bodies and should comply with the
1.10 Laboratories shall have walls that extend to the ceiling from public/common areas
or at containment barrier location.
1.11 Office areas to be located outside of containment laboratory. Paperwork stations
for data collection can be within containment laboratory provided they are
located away from laboratory work areas ideally this work should be separated
from the bench space used for experimentation to assist in preventing
contamination of office areas.
1.12 Clothing and PPE donning and doffing areas should be accommodated at the exit
from the laboratory.
1.13 Supply and waste removal activities should be located at the lab perimeter.
Finishes
1.1 Doors, frames, casework and bench tops to be non-absorptive (i.e., the use of
organic materials should be avoided).
1.2 Windows positioned on containment barrier to be sealed in place; window glazing
material to provide required level of security
1.3 Doors, frames, casework and bench tops to be non-absorptive (i.e., the use of
organic materials should be avoided).
1.4 Working surfaces of bench tops to be non-absorptive
1.5 Surfaces to provide impact resistance in accordance with laboratory function
1.6 Floors to be slip-resistant and washable
1.7 Floors to be slip-resistant and washable
1.8 Surfaces to be scratch, stain, moisture, chemical and heat resistant in accordance
with laboratory function.
1.9 Benches, doors, drawers, door handles, etc. to have rounded rims and corners.
1.10 Surfaces to be scratch, stain, moisture, chemical and heat resistant in accordance
with laboratory function.
12 | P a g e
Furnishings
1.1 Surfaces to provide impact resistance in accordance with laboratory function
1.2 Interior coatings to be gas and chemical resistant in accordance with
laboratory function (e.g., will withstand chemical disinfection, fumigation
1.3 Bench tops to have no open seams
1.4 Interior coatings to be gas and chemical resistant in accordance with
laboratory function (e.g., will withstand chemical disinfection, fumigation
1.5 Bench tops to have no open seams
1.6 Bench tops to contain spills of materials (e.g., with marine edges and drip stops).
1.7 Backsplashes, if installed tight to wall, to be sealed at wall-bench junction
1.8 Reagent shelving to be equipped with lip edges
1.9 Drawers to be equipped with catches, i.e., to prevent the drawer from being
pulled out of the cabinet
1.10 Cabinet doors not to be self-closing
1.11 No fabric or absorptive material to be used on chairs used in wet labs.
Chairs should be constructed with impervious chemically resistant material.
Storage and Equipment
Autoclave
1.1 Autoclave or other acceptable means of waste treatment/disposal to be provided.
1.2 Separate exhaust must be provided from the autoclave
Flammables
1.1 All flammable storage shall comply with the EHS flammable storage
standards found in Appendix A.
1.2 Open flammables in the lab cannot exceed 50 litres. Space must be allocated
within the lab for flammable cabinets. Cabinets may be vented or sealed
cabinets.
1.3 Special consideration is needed for flammable compressed gases. In general,
one should limit what can be stored in the lab. Additional storage spaces for
compressed gases should be provided somewhere else within the building
1.4 Explosion proof refrigerators are required for cold storage of flammables.
1.5 Note that there may be considerable additional requirements if large quantities
of flammable gasses are to be used (eg hydrogen)
Cryogenic Materials
1.1 All installations using cryogenic materials must comply with the VIU “Control
Program for Liquid Cryogenic Transfer Facilities” and the “Standard for Inert
Cryogenic Liquid Usage in the Laboratory” as well as special precautions for
flammable, toxic or reactive cryogens such as hydrogen, carbon monoxide, and
oxygen. Provisions in the standards include the following.
1.2 There must be appropriate ventilation in areas where cryogens are
being dispensed.
1.3 Oxygen sensors and panic buttons linked into the ventilation system may
13 | P a g e
be required, as well as specialized signage.
Gas Cylinders
1.1 Labs must include provision for safe storage of unused gas cylinders.
1.2 Adequate physical support is required for gas cylinders in use
1.3 Additional ventilation may be required for toxic compressed or liquefied gasses.
1.4 Shipping areas must include a holding area for receipt of new cylinders.
1.5 Systems for securing gas cylinders while in use should be available
1.6 Special consideration is required for LPG cylinders.
General Chemical Storage
1.1 Chemicals should be stored in accordance with the separation classes specified
in the VIU “Laboratory Safety Manual”.
1.2 Adequate provision for separation of incompatible chemicals.
1.3 Ventilation is deemed adequate as per appropriate codes of practice
1.4 Separate storage for acids includes corrosion resistant shelving and
secondary containment. Special acid storage cabinets can be purchased
1.5 Waste disposal
1.6 Unneeded chemical, biological and radioactive materials must be disposed of in
accordance with the VIU “Laboratory Hazardous Waste Management and
Disposal Manual”. Design considerations to aid compliance include:
1.7 Adequate floor space for waste containers.
1.8 Waste generated from the activities of the lab will require a large enough space
for waste. Space required will be determined by the waste bins required to handle
the volume of waste generated between waste pick up cycles.
1.9 If large amounts of chemical waste are to be generated, provision for special
ventilation in the short term holding area.
1.10 An area to store radioactive waste that is separate from the working area.
Biosafety Cabinet
1.1 BSC should be located away from high traffic areas, doors and air supply/exhaust
grilles that may interrupt airflow patterns. A minimum unobstructed distance of
40 cm should be provided between the exhaust outlet on top of the cabinet and
any overhead obstructions. Whenever possible, a 30 cm clearance should be
provided on each side of the cabinet to allow for maintenance access. For ducted
cabinets, blowers
1.2 Biosafety cabinets should be located 1.5 m away from any doors and air vents and
away from high traffic area
5.0 Containment Level 2: Mechanical Requirements
HVAC
1.1 The HVAC system (heating, ventilation, air conditioning) should not
14 | P a g e
i
n
t
e
r
f
ere with the airflows of biological safety cabinets
1.2 The exhaust system should be located at the terminal end of the ductwork; failure
of exhaust flow should signal an alarm to the user. To prevent pressurization of
the cabinet, an interlock system should be installed to prevent the cabinet blower
from operating whenever the exhaust flow is insufficient; an anti-backflow
device to prevent reverse airflow through the HEPA filter may be required.
Plumbing
Showers
1.1 All shower installations must follow the VIU “Emergency Eyewash and
Shower Standard” – see Appendix A – and must include piped in emergency
showers in labs where a hazard to the skin exists. :
15 | P a g e
1.2 Drainage for emergency showers is required by the building code.
1.3 Emergency showers should be located near the door and within a 10 second
walking distance from the hazardous activity.
1.4 Emergency shower placement in corridors, to serve a number of labs is
discouraged, and only approved by EHS in exceptional circumstances.
1.5 Special consideration provided; rooms with electrical equipment may be
exempted from shower installation
1.6 Drench hoses are only accepted in a grandparented situation and should not be
used in new or renovated labs
Sinks 1.1 A separate hand washing sink should be placed within 15 feet of the door to allow
hand washing before exiting the lab.
1.2 A separate utility/glassware washing sink should be provided.
1.3 Sinks must be stainless steel. or an epoxy chemical resistant coating
1.4 Pipes and fixtures must be chemically resistant. Glass pipe traps should be
specified in the design stage. The height of the trap must be sufficient to prevent
the negative pressure in the lab from allowing gasses to escape from the drain
line (off-gassing).
1.5 Appropriate back-flow prevention should be included in the design.
1.6 Each sink will have an overflow outlet.
16 | P a g e
5.2.3 Eye wash
1.1 All eyewash installations must follow the VIU “Emergency Eyewash
and Shower Standard”: - see Appendix A.
1.2 A plumbed tepid eyewash is required for new installations. Tepid
installations require hot and cold water with a mixing valve to allow for a
full 15 minutes of flushing.
1.3 An emergency eyewash station will be provided in the room or in close
proximity to the room.
1.4 Tap fixture eye wash will not be accepted; separate plumbing is required.
Lab Type Specific Requirements High hazard Chemical
6.1.1 Labs which boil large quantities of acids, or use particularly hazardous
materials like hydrogen fluoride, perchloric acid or other explosive or
dangerous materials, may have additional requirements for fume hoods or
other ventilation of the work area. (Codes specific to chemicals need to be
followed.)
6.1.2 For highly toxic chemicals, special provisions may be needed to ensure that
materials remain within a certain area and do not migrate to less well
controlled areas. An example of a design consideration is the provision of
an anteroom or a wash up room as the exit to the lab.
6.1.3 The flooring material must be solvent and acid resistant.
Radiation
6.1.4 The design of all radiation labs must be approved by the federal regulator
(CNSC) GD-52
6.1.5 Intermediate and high level radiation laboratories must satisfy all
design conditions for wet labs.
6.1.6 Since access to these rooms is restricted to persons with radiation training,
persons wearing radiation badges and protective equipment, controlled
access is required.
6.1.7 Bench tops should have no open seams, and will have a seamless one-piece
design.
6.1.8 All windows must be covered with security film
6.1.9 An area to store radioactive waste that is separate from the working area,
preferable in a waste storage room that is located near the loading bay/dock.
Engineering
Physical design of spaces for engineering labs must consider diverse issues including the following.
a) Working at heights.
b) Confined spaces.
17 | P a g e
c) Design for minimal ladder use.
d) Lock-out and tag-out of equipment.
e) Engineered noise control at source.
f) Lifting devices, cranes, hoists etc.
g) Safety interlocks for equipment not otherwise protected.
h) Special wiring for specialized equipment.
i) Heat and cold stress.
j) Provision of mechanical aids to prevent unneeded manual material handling.
k) Ergonomic design.
l) Guarding and ventilation to minimize PPE (respirators, face shields etc.)
requirements for hazards like the following.
• Guarding of pinch points, cutting tools and rotating shafts etc.
• Welding and hot work.
• Grinders, woodworking equipment.
• Battery charging stations.
• Any other equipment that generates a hazardous atmosphere or
condition.
Laser
Laboratories containing any lasers are required to follow the Laser Safety Program which includes
the following design requirements.
1.1 Visible sign at entrance.
1.2 Light at entrance.
1.3 Interlocks on doors.
1.4 The arrangement for the lit sign which signals that a laser is operational must
be clearly defined and included in the design phase
1.5 Curtain around the laser inside the door.
X-Ray 1.1 Sign at the entrance.
1.2 Special construction of walls to address shielding considerations.
1.3 Interlock if the x-ray machine is open and not contained in a cabinet.
1.4 Light at entrance if machine is open.
1.5 Self-locking mechanism on door.
Power Supply and Servers
1.1 Shielding may be required if servers are not able to meet the EMF emissions
according to Vancouver Island University standards
1.2 Control access and signage for workers
Sterilization Rooms
1.1 Mange Humidity, with controls and ventilation.
8.0 Offices: Seating Requirements
18 | P a g e
The following recommendations are for minimum requirements for an office chair
at Vancouver Island University. The recommendations are classified by time spent sitting as the
major risk factor.
The two classifications are intensive and moderate. Intensive being more than 5 hours a day and
moderate being 3-5 hours a day.
Chair requirements Intensive computer use/sitting (5 or more hours a day)
□ Height adjustable
□ Adjustable seat pan depth
□ Adjustable lumbar support
□ Tension adjustable recline (active back)
□ Adjustable armrests (must adjust on the horizontal and vertical plane (swivel in and out and
slide forward and back)
□ Comfort (section 3)
□ Ease of adjustment (section 4)
Examples: Steelcase Amia, Leap
Moderate computer use/sitting (3-5 hours a day)
□ Height adjustable
□ Adjustable seat pan depth
□ Adjustable backrest angle
□ Adjustable armrests (must adjust on both the horizontal and vertical plane (swivel in and out
and slide forward and back)
□ Comfort (section 3)
□ Ease of adjustment (section 4)
Examples: Steelcase Reply, Crew
Comfort Chairs that have all the features described in this document may still vary in comfort. A chair test or
demo, ideally under actual working conditions, is helpful in determining whether a chair will truly be
comfortable for an employee and should be explored if possible.
Ease of adjustment If more than one person will be using a chair or to obtain an optimum fit for an individual, a chair
should be able to be adjusted easily.
Chair adjustment controls should:
• Be operable while sitting in the chair
• Be logical in placement and direction of activation
• Have control markings that are visible from a seated position or be distinguishable by some other
means (e.g. different shapes)
19 | P a g e
• Not require undue force for activation
• Be designed to prevent unintentional activation
Warranty
Chairs should include a lifetime warranty. Repairs should be timely within a 1-2 week window after
the initial request is made and the request should be simple and efficient with one contact person.