Density, Contrast, Distortion & Detail
Lab Test #3
![Page 2: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/2.jpg)
Contrast and the 15% Rule
Objective: To demonstrate the influence of the 15% rule on contrast
1st exposure at recommended technique
Used as control
![Page 3: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/3.jpg)
Contrast and the 15% rule
Apply the 15% rule twice and ↑ kVp ↓ mAs accordingly
All other factors stay the same
The math doesn’t quite add up on this one, so I think the images are from 2 different groups
![Page 4: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/4.jpg)
Contrast and the 15% rule
Again, this image doesn’t match my lab sheet, I believe we lowered the kVp 15% and shot another on just for S&G
At this point, we should compare the 2 images to determine which has the longest scale of contrast
See next slide
![Page 5: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/5.jpg)
Contrast and the 15% rule
Low kVp=High Contrast= Short scale High kVp=Low Contrast= Long scale
More B&W More Shades of Gray
![Page 6: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/6.jpg)
Density- Control Image
Objective: To demonstrate the effect of select factors on density
Control image of knee phantom taken @ 5mAs, 60kVp, & 40” SID.
![Page 7: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/7.jpg)
Density- Filter
With no change in tech factors, a copper sheet is taped to the face of the collimator
Filters have an inverse relationship on density thus:
↑ Filtration ↓ Density
![Page 8: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/8.jpg)
Density- Grid
Copper filter removed No change in technical
factors Grid is introduced to IR Result: Adding a grid
decreases density this is an inverse relationship
(+) ↑ Grid ↓Density
![Page 9: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/9.jpg)
Density- SID
Grid removed No change in technical
factors SID increased from 40”
to 60”
![Page 10: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/10.jpg)
Density-Tissue Thickness
Grid is removed and hand phantom is substituted for knee
No change in technical factors
Overall, the hand the tissue thickness of the hand is much less than that of the knee
↑ tissue thickness ↓density
![Page 11: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/11.jpg)
Density- Collimation-Control
Torso phantom is substituted for hand
14x17 collimation Use technique
recommended by console + 20 mAs
This image will be compared with the highly collimated image.
![Page 12: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/12.jpg)
Density- Collimation
Collimation is increased to a 4x4 square.
Repeat exposure of torso with no change in technical factors
↑collimation ↓density
![Page 13: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/13.jpg)
Density- Anode Heel Effect
Use a foot phantom and a 14x17 cassette
2 mAs, 60 kVp, 40” SID Leave collimation open
lengthwise to 17” Orient toes over the
anode side
![Page 14: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/14.jpg)
Density- Anode Heel Effect
All technical factors remain the same
Foot is moved to opposite end of cassette to place toes on cathode side
In practice, the thickest tissue should be placed at the cathode end of tube
In theory, placed thicker tissue at anode end would effect density
A visible change in density would only be appreciated with film/screen
![Page 15: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/15.jpg)
Size Distortion: Control
Objective: To demonstrate the various types of distortion
Control image of hand shot at 40” SID
Technique: Pre-programmed
![Page 16: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/16.jpg)
Size Distortion: 6” OID
2nd image 6” OID is created with
sponges All other factors remain
the same Compare with control
image to note distortion (magnification)
![Page 17: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/17.jpg)
Size Distortion: 12” OID
Image #3 OID is ↑ to 12” Collimation is exactly the
same No other factors have
been changed Note magnified
appearance of hand Such distortion can mask
pathology
![Page 18: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/18.jpg)
Distortion: OID + SID
OID is ↓ to 6” SID is ↓ to 20” All remaining tech factors
are unchanged Result: Size distortion
(magnification in this case) can be caused by OID, SID or both
![Page 19: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/19.jpg)
Shape Distortion: Control
Control exposure of knee phantom w/recommended tech factors
This image will be used for comparison with others
![Page 20: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/20.jpg)
Shape Distortion: Angled Tube
2nd image of knee phantom
Tube is angled All tech factors remain
the same
![Page 21: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/21.jpg)
Shape Distortion: Angled IR
3rd exposure of knee phantom
IR is angled All tech factors are
unchanged from control
Note closed joint spaces
![Page 22: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/22.jpg)
Shape Distortion: Angled Anatomy
4th exposure of knee phantom
All tech factors remain constant except…
Phantom is angled Result: Shape distortion
can be caused by the angulation of tube, IR or anatomy being imaged.
![Page 23: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/23.jpg)
Density & mAs
Objective is to demonstrate the effects of overexposure & underexposure on CR images
First image is made @ 60kVp, 10mAs w/40” SID
![Page 24: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/24.jpg)
Density & mAs 2
mAs is ↑ 50 All other technical factors
remain the same LGM#’s of all images will be
compared with control LGM represents the # of
photons reaching IR to form the latent image
LGM is proportional to mAs
![Page 25: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/25.jpg)
Density & mAs 3
↑ to 100 mAs No other technical factors
are changed At this stage, LGM#’s
appear to be ↑ as mAs is ↑ (a direct relationship).
![Page 26: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/26.jpg)
Density & mAs 4
↑ to 200 mAs Remaining tech factors
unchanged LGM#’s increase with
each increase in mAs mAs is THE controlling
factor of density
![Page 27: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/27.jpg)
Detail and Distortion: Motion
Objective: To demonstrate the effect of motion on detail
Technical factors: The programmed technique for a foot but ↓ mA
Decrease in mA is necessary to ↑ exposure time (applicable in next image)
1st image of top taken on 10x12 w/no motion
![Page 28: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/28.jpg)
Detail and Distortion: Motion
2nd image- top is spinning while exposure is taken
Technical factors are unchanged
The lower mA ↑ exposure time, allowing the motion to be caught on film
![Page 29: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/29.jpg)
Detail and Distortion: Motion
Additional image of this experiment
Compared to the 1st image (stationary), the lead letter attached to top is blurred.
Result: Increased motion decreases detail
Motion↑↓ Detail Although motion is generally a
detriment to good films, it can be used to the RT’s advantage
Ex: Using breathing technique to blur ribs
![Page 30: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/30.jpg)
Quantum Mottle
Definition: a lack of sufficient incoming data to process an image; AKA quantum noise
No idea what this image has to do with anything, but that’s how it was labeled
That would be a cool name for a band though
![Page 31: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/31.jpg)
Contrast- Control Image
Objective: To demonstrate how selected factors effect contrast
66kVp, 10 mAs, 40” SID on table top
![Page 32: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/32.jpg)
Contrast 66kv
Skull phantom is replaced w/step wedge
Exposure taken with no changes in technical factors
Compare this image w/second exposure using a higher technique
![Page 33: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/33.jpg)
Contrast: ↑kVp ↓mAs
↑ to 86kVp Compensate by ↓mAs to
¼ of its original value All other factors remain
the same
![Page 34: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/34.jpg)
Contrast- 86kv
Again, step wedge takes the place of skull phantom
No change in technical factors
Compare shades of gray in adjacent densities
![Page 35: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/35.jpg)
Comparison of Contrast
Low kVp High kVp
![Page 36: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/36.jpg)
Contrast: +Grid
3rd image of skull Grid is introduced mAs ↑ 4x to
compensate All remaining factors
unchanged
![Page 37: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/37.jpg)
Contrast: Scatter
4th image of skull Collimation open wide
to expose IR to scatter All remaining factors
unchanged Compare this series of
images to note differences in adjacent densities
![Page 38: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/38.jpg)
Contrast Collimation
Skull is now positioned laterally
This exposure will use bucky instead of table top
100 kVp @ 20mAs
![Page 39: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/39.jpg)
Contrast: Collimation
Repeat exposure of lateral skull
Collimated to 3”x3” area
All remaining factors are unchanged
![Page 40: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/40.jpg)
Short vs Long Scale Contrast: kVp
Objective: To demonstrate short & long scale contrast
Elbow phantom @ 46kVp & 5mAs
![Page 41: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/41.jpg)
Stepwedge 2 Contrast ????
I think this is supposed to be part of a contrast lab but I couldn’t match it up with anything
Also, it was posted twice in the images file…so there’s nothing to compare it to as far as I can tell
![Page 42: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/42.jpg)
Short vs Long Scale Contrast; kVp
2nd image of elbow phantom
↑ to 70kVp ↓to 3mAs
![Page 43: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/43.jpg)
Density Chart
Variables Effects Density Relationship
Filtration ↑↓ ↓↑ Inverse
Grid ↑↓ ↓↑ Inverse
Tissue Thickness ↑↓ ↓↑ Inverse
SID ↑↓ ↓↑ Inverse
Collimation ↑↓ ↓↑ Inverse
Anode Heel Effect Cathode: ↑Anode: ↓
*In theory the anode heel effect does cause a change in density but this is not evident in digital imaging
N/A unless film screen technology is involved
![Page 44: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/44.jpg)
Contrast ChartVariable Effect Contrast Relationship
kVp ↑↓ ↓↑ inverse
mAs Ø Ø none
SID Ø Ø none
OID ↑↓ ↑↓ direct
Filtration ↑↓ ↓↑ inverse
Collimation ↑↓ ↑↓ direct
Tissue Thickness ↑↓ ↓↑ inverse
Contrast Media + ↑ direct
Grid ratio ↑↓ ↑↓ direct
Focal spot size Ø Ø none
Film Processing *+/- developing time and temp beyond optimal
↓ Always ↓ contrast
![Page 45: Lab Test 3](https://reader033.vdocument.in/reader033/viewer/2022061214/549a50a0b47959043f8b45be/html5/thumbnails/45.jpg)
Detail Chart
Variable Effect Detail Relationship
SID ↑↓ ↑↓ direct
OID ↑↓ ↓↑ inverse
Tissue Thickness ↑↓ ↓↑ inverse
Focal Spot Size ↑↓ ↓↑ inverse
Motion ↑↓ ↓↑ inverse