NuAire,Inc.|2100FernbrookLane|Plymouth,MN55447|U.S.A.|763.553.1270|WWW.NUAIRE.COM
LABGARD NU-629 CLASS II TYPE A2
ANIMAL HANDLING BIOLOGICAL SAFETY CABINET
Safety
Greatervisibilityandreachthroughalarge12(305mm)access
opening along with Class II Type A2 bio-hazard containment
througha105FPMairbarrier.
Uni-directional laminar airflow minimizes air turbulence and
crosscontaminationbetweenanimals.
Allcontaminatedducts,plenums,andgasketsareundernegative
pressureorsurroundedbynegativepressuretoeliminatethe
possibility of cabinet or gasket leaks and to guard against
HEPAfailure.
Ergonomic Auto-Lift
Electronically adjustable work surface height from 23.75
(603 mm) to 38.75 (984 mm) or 42.75 (1086 mm) with
optional lift system, accommodates the ergonomic needs of
multipleuserswhomaybesittingorstanding.
Asabenchstylecabinetwiththemotor/blowerlocatedinthe
top,theNU-629isidealforseatedanimalprocedurework.The
NU-629 is designed for ergonomics and allows a chair to fit
completelyunderneaththeworksurface.
Thenarrowprofileoftheliftsystemdoesnotaddtotheoverall
widthofthecabinet.
HEPA Filtration
High Efficiency Particulate Air (HEPA) filtration is
99.99% efficient on removal of particulates for
supply(re-circulated)andexhaustedair.
EnergyefficientDCstyleECMmotoroffersanenergy
savingsofmorethan50%overconventionalmotor
technology.TheECMmotoralsocreatesanincrease
in total filter load capacity of 250% to extend the
HEPAfilterlifeexpectancyfrom710years.
Adjustableinternaldamperandmotorspeedcontrol
furtherextendHEPAfilterlife.
Cleanablility
Pre-filterscapturebedding,hairanddanderandare
easytoaccessforcleaning.[C]
NU-629
Theworksurfacecanbeliftedandproppedorremovedquicklywithouttoolstoallowforroutinedisinfectionorcleaningliquidspills.[A] Theheightoftheworksurfaceisadjustablewithina15(381mm)rangefromminimumof23.75(603mm)toamaximumof38.75(984mm).Thisflexibilityallowsausertositorstand,aswellasaccommodatingawiderangeofuserheights.TheNU-629featuresa redesigned front panel to allowtheusertostandclosertotheunit,providingabettersightlinetotheworksurface.[B]
[B]
[A]
[C]
The LabGard NU-629 Class II Type A2 Biological Safety Cabinet/Cage Changing Station provides personnel, product, and environmental protection for sensitive animal handling and cage changing procedures. Built to withstand the rigors of use in an animal facility, the NU-629 is both mobile and height adjustable with the following key features:
Airflow
ContaminatedAir
Pre-FilteredAir
CleanHEPAFilteredAir
AirInsideHEPEX
RoomAir
Exhaust
Experience Ergonomics
2
Allmodelsaredesignedto
meetUL,CUL,andCEcriteria
forsafetyandperformance.
Standard Features
EnergySaverDC/ECMMotorUpgrade
HEPEXZeroLeakAirflow
EssentialsControlSystem
ExperienceErgonomicSystem
105fpm(0.50m/s)AirBarrier PersonnelProtection
BackWallPre-filter
FlatPropUpWorkTray
ADACompliantDesign
12(305mm)AccessOpening
SinglePieceMonolithicConsole StyleConstruction
15(381mm)AutoLiftBaseStand
TwoDuplexOutlets(BackWall)
HEPAFilters99.999%@0.3Microns
MotorSpeedController
InternalExhaustDampener
InteriorMountedCoolWhiteLighting
5(127mm)LockableCastors
RemovableControlCenter
Optional Features
GermicidalUVLight
GroundFaultInterrupter
AdditionalServiceValves
RemoteServiceValves
AdditionalOutlet
AlarmSystems
ExhaustTransitions
Sinksw/Hot/ColdorDIWaterFaucets
StoragePull-OutTrays
LayinSorbentExhaustFilter
DecorativeSidePanels (HidesPlumbingFixtureConnections)
18(457mm)HeightRangeAutoLift BaseStand
FixedHeightBaseStand
PermanentPlenumw/Quick ReleaseSupplyFilter
MetalFramedHEPAFilters
ULPAFilters
StainlessSteelArmrest
HeatedWorkSurfaceWith Micro-ProcessorTemperature ControlSystem
1
Effective Work Zone Area for
More User-Friendly Space
NuAires cabinets provide the
largest effective work zone
which helps reduce arm/neck/
shoulder strain. Extend reach
up to 12 with 100% stainless
steel turntable (ball bearing
construction,easytoclean,and
canbeautoclaved).
3
Turntable
Ability to Sit or Stand at a Range of Heights
The cabinet design maximizes knee/thigh
clearancetoimprovepostureandtheadjust-
able base stand
allows optimiza-
tion for leg and
forearmsupport.
Expanded Vision Zone Reduces Awkward
Postures and Proper Lighting Reduces Glare
Cabinets incorporate coolwhite lighting and
framelesspolishededgewindowstoallowfor
greater visibility
and better sight
lines.
LabGard NU-629BiologicalSafetyCabinet
Specifications
* Cratedshippingweightdoesnotincludeweightforaccessoriesoroptions.
Baseoncabinetwithnewfiltersrunningat115VAC.
Catalog Number NU-629-400Nominal4Foot(1.2m)
NU-629-500Nominal5Foot(1.5m)
NU-629-600Nominal6Foot(1.8m)
PerformanceSpecifications:PersonalandProductProtection NSF/ANSI49 NSF/ANSI49 NSF/ANSI49
NSFStd.No.49Class ClassII,TypeA2 ClassII,TypeA2 ClassII,TypeA2
StyleofCabinet BenchTop/Consolew/BaseStand
BenchTop/Consolew/BaseStand
BenchTop/Consolew/BaseStand
CabinetConstruction(PressureTightDesign)
AllWeldedStainlessSteel16GA,Type304
AllWeldedStainlessSteel16GA,Type304
AllWeldedStainlessSteel16GA,Type304
DiffuserforAirSupply(Metal) Non-flammable Non-flammable Non-flammable
HEPAFilterSealType:SupplyFilter-99.99%Eff.on0.3ExhaustFilter-99.99%Eff.on0.3
HEPEXSealNeoprene,Springloaded
HEPEXSealNeoprene,Springloaded
HEPEXSealNeoprene,Springloaded
FumigationperNIH/NSFProcedure Yes Yes Yes
StandardServices:ServiceCoupling(3/8inchNPT)DuplexOutlet
Two,RightSideWallTwo,BackWall
Two,RightSideWallTwo,BackWall
Two,RightSideWallTwo,BackWall
OptionalServices:GasCocks(3/8inchNPT)UltravioletLight
UptoThreeEachSideWallOne,BackWall
UptoThreeEach,SideWallOne,BackWall
UptoThreeEach,SideWallOne,BackWall
CabinetSize,Inches(mm):Height(Min/Max)WidthDepth(withControlCenter)
821/2(2096)/951/2(2426)535/8(1362)327/8(835)
821/2(2096)/951/2(2426)655/8(1667)327/8(835)
821/2(2096)/951/2(2426)775/8(1972)327/8(835)
WorkAccessOpening:StandardOpeningHeight,Inches(mm)StandardInflowVelocity(DIM)WorkSurfaceHeight(Min/Max)
12(305)105FPM(.53m/s)26(660)/39(991)
12(305)105FPM(.53m/s)26(660)/39(991)
12(305)105FPM(.53m/s)26(660)/39(991)
WorkZone,Inches(mm):HeightWidthDepth
281/2(724)463/8(1178)231/2(597)
281/2(724)583/8(1483)231/2(597)
281/2(724)703/8(1788)231/2(597)
ViewingWindow,Inches(mm)StandardisTemperedSlidingGlass
FullyClosedto191/2(495)Open FullyClosedto191/2(495)Open FullyClosedto191/2(495)Open
RequiredExhaustCFM(CMH)Standard:Canopy(NU-918/917)Canopy(NU-916)
506(860)530(818)
627(1065)657(1117)
750(1274)772(1314)
PlantDuctStaticPressure,Inches(mm) 0.05-0.1(1.27-2.54) 0.05-0.1(1.27-2.54) 0.05-0.1(1.27-2.54)
HeatRejected,BTU/Hour(Non-vented/Vented) 2551/157 2669/178 2787/198
Electrical:Volts,AC(60Hz)Amps:Blower/Lights/Autolift
Amps:DuplexAmps:Total12ftPowerCord(One)
U.L./U.L.-CListed1156.531514GA-3Wire,20A
U.L./U.L.-CListed1156.831614GA-3Wire,20A
U.L./U.L.-CListed1157.131614GA-3Wire,20A
NetWeightCratedShippingWeight*
750lbs(340kg)800lbs(362kg)
815lbs(369kg)865lbs(392kg)
880lbs(398kg)930lbs(421kg)
LabGard NU-629BiologicalSafetyCabinet
Process#20-1091DRev.25/11
Formoreinformationpleasevisitwww.nuaire.comorcall1.800.328.3352
http://www.facebook.com/nuairehttp://www.linkedin.com/company/nuairehttp://www.twitter.com/nuairehttp://www.youtube.com/nuaire