Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013 A reality check?

Upload: phamdan

Post on 30-Jul-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Laboratory instrumentation the last 10 years

Energy, Mining and Environment

Pamela Whitfield

Accuracy in Powder Diffraction IV, NIST, April 2013

A reality check?

Page 2: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Overview

•Where we’ve come from

•The goniometer – heart of the system

•Kinematic-type mounts

•PSDs

•Multi-layer optics

•Selected example application

Page 3: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Diffraction in the lab

• Lab diffractometers often in multi-user environments

• No single setup has been able to satisfy everyone • However CuKα Bragg-Brentano with graphite monochromator

came closest and was the de-facto standard for years…

• Simple

• Built like tanks

• Minimal electronics • The Scintag XDS2000 has 1 safety

circuit that can be defeated with a magnet!

Page 4: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

The march of analogue to digital….

• Systems have become much more complex • Systems designed to be flexible and modular

– The electronics themselves sometimes less so…..

• More sophisticated microprocessor electronics and safety systems • Architectures familiar to anyone who`s looked inside a PC

Old New

Page 5: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

The Goniometer heart of the diffractometer

• Basic concept unchanged • Axes attached to very large bearings… usually ball races

• Drive gears do eventually wear

• Goniometers still largely mechanical but… • Optical encoders now widespread to improve positioning accuracy • Goniometer arms → `optical benches` • Realignments reduced or even eliminated

Heidenhain RON905 optical encoder accuracy ±0.4”

Page 6: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

The Goniometer • When pushing the limits of accuracy every little matters…

• Flexing of goniometer arm limits accuracy in vertical setup • → detector/source weight & counterbalancing become issues

• θ−2θ often preferable in terms of pure positional accuracy

Jim Cline being fussy! Stress analysis on Jim’s custom D5000 detector arm

Page 7: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

System alignment • Manual system alignment can be…

• boring • frustrating • source of much rude language! • impediment to switching configurations

– increases my ‘activation energy’

• Don’t have someone with the ‘Midas touch’?

• 2 main alternative approaches • Fully motorized and automated alignment • Kinematic-type mounts for reproducibility

Page 8: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Automated alignment

• Alignment automated at the touch of a button • Everything motorized and software controlled • Easy for the user

• However, system never exactly the same twice after realigning • No different from manual realignment in this regard

• System reproducibility shouldn’t change over time • Mechanical wear excepted

Page 9: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Kinematic mounts • Kinetic mounts used routinely for mounting optical mirrors

• 3-leg stool (tripod) concept

• Dirty environments? • Forces should force dirt out from between surfaces – no effect • However, turbidity from environment/surfaces can prevent

equilibrium being reached – Chocking effect (www.precisionball.com/kinematic_encyclopedia.php)

3-point kinematic mirror mount

Page 10: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

• Precision engineering – expensive!

• Hardened surfaces desirable (e.g. WC) • Reduce contact distortion + low friction • High shear strength to resist lateral forces (chocking)

• Mounts need to be robust • Delicate stuff tends to not fare well in a university environment!

• Factory-aligned • Portability between multiple systems?

Page 11: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

From the old to the new….. • Bragg-Brentano + graphite mono/scintillation detector

• gives excellent data but low counts

• Fear not! • Many new toys to spend your money on!

• Seems every component has a possible upgrade • X-ray source

– Microfocus, rotating anode, ‘liquid metal jet’

• Optics – Mirrors, monocaps, polycaps, monochromators (primary and secondary)

• Detectors – PSDs, 2D detectors, energy-dispersive

Page 12: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

X-ray sources

• Standard X-ray tube still the most common source • Ceramic insulation used increasingly • Ceramic is a better heat-conductor • Better tube lifetime claimed

• Rotating anode • Still quite a niche product for powder diffraction • Large increase in tube power (↑ 18kW) • Generally very heavy and bulky

– Not normally used for θ−θ systems due to issues mentioned previously but….

Rigaku TTRAX

GE X-ray tubes

Page 13: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Spot-focus X-ray sources • Microfocus

• Widely available • Standard anode materials • High brilliance

– Comparable to older synchrotron beamlines

• Liquid-metal jet • Don`t have to worry about melting the

anode as it`s already liquid • Wavelength GaKα (other metals as

dopants in liquid) – GaKα similar to CuKα but users often

uncomfortable changing wavelengths..

Page 14: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Position Sensitive Detectors rule? • Historically proportional detectors using a counting gas

e.g. mBraun, Inel

• Calibration required to relate ‘channel number’ to angle

• Defocussing in linear PSDs

• Inel known for their curved PSDs • No defocussing but calibration functions more complex

mBraun PSD

Inel 120° PSD

Page 15: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

PSDs rule! • PSDs largely moved away from gas-based technologies

• PSDs now a common sight in XRD labs • What effect has this had on aspects of accuracy?

•GOOD • Count rates… • Excellent signal-to-noise

•BAD • Count rates! • Artefacts in data now visible

Page 16: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Newer PSD technologies

2 (degrees CoK)5 10 15 20 25 30 35

Inte

nsity

(cou

nts)

0

1000

2000

3000

4000

35kV, 40mA30kV, 45mA

Quartz stoneCo tube, Vantec, Fe filter

Xe escape peaks

K

K

• More advanced gas-based counting (e.g. Vantec-1) • 33keV Xe escape peaks can be problematic

– not all detectors the same…. one of ours less sensitive then the other • Works up to MoKα (Vantec AgKα )

Effect of generator voltage on appearance of 33keV Xe escape peaks without primary optic

Page 17: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Solid state strip detectors • Current mainstream PSD technology

• Cost effective

• PSD energy discrimination improving

• Higher X-ray energies require thicker Si films to maintain efficiency

• Image-plate based curved PSD - Stoe

Page 18: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

2 (degrees - CuK)20 30 40 50 60 70 80 90

Arbi

trary

inte

nsity

20 30 40 50

Sqrt(

inte

nsity

)

* * * *

PSDs rule! – not so good? • More stuff you`d rather ignore

• Absorption-edges of Kβ-filters – Background fitting errors → inaccurate peak intensities

• Tungsten Lα lines & residual Kβ intensity becomes very obvious

LaB6 with CuKα and a nickel Kβ filter

Page 19: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Lin

(Cps

)

0

10

20

30

40

50

2-Theta - Scale20 30 40 50 6

Which dataset do you prefer?

PSDs • Normally poor energy discrimination

• In absence of monochromator fluorescence a major problem • Poor peak-to-background impacts accuracy (CPD Rietveld RR) • Tweaking detector electronic discriminators can help…

LiMn1.5Ni0.5O4 CoKα with varying detector discriminator settings

Page 20: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Discriminators…

Lin

(Cps

)

0

1

2

3

4

5

6

7

8

9

10

2-Theta - Scale21.2 22 23 24

PHA0.1 0.2 0.3 0.4 0.5 0.6

LL = 0.36WW = 0.06 P/B = 13.4

PHA0.1 0.2 0.3 0.4 0.5 0.6

LL = 0.28WW = 0.34

P/B = 4.5

PHA0.1 0.2 0.3 0.4 0.5 0.6

LL = 0.1WW = 0.5

P/B = 4.2

Rescaled to normalize background

Sacrifice intensity to improve P/B ratio

P/B still way off 50

• Recent developments in electronics have greatly improved possible energy discrimination of PSDs

Page 21: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Alternative low tech approach… • High-purity Al foil filters can reduce fluorescent background

• costs intensity • but once again less can be more!

Mn3O4 CuKα primary mirror mBraun PSD Al foil mounted on Ni filter

2 (degrees CuK)20 30 40 50 60

Inte

nsity

(cou

nts)

0

1000

2000

3000

4000

5000no Al5 Al layers/Ni10 Al layers/Ni

2 (degrees CuK)30 32 34 36 38 40

Inte

nsity

(cou

nts)

0

50

100

150

200

250no Al 15 Al layers 10 Al layers

Normalized to background

P/B = 20

Page 22: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

2 (degrees CoK)0 20 40 60 80 100 120 140

Sqrt(

Inte

nsity

)

0

20

40

60

80

100

120

140

160 Rwp = 10.2%

15 20 25 30 35 40

0

10

20

30

40

50

60

Residual K

PSDs and monochromators… Data from Ian Madsen

Y2O3 data from an APD system Bragg-Brentano proportional 0D detector graphite monochromator

2 (degrees CoK)0 20 40 60 80 100 120 140

Sqrt(

Inte

nsity

)

0

50

100

150

200

250Rwp = 9.2%

15 20 25 30 35 40

0

10

20

30

40

50

60

Residual K

• Out-of-plane monochromators fitted to some PSDs

• Gets rid of fluorescence but can cause other issues

Y2O3 data from an MPD system Bragg-Brentano PSD strip detector with monochromator

Page 23: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

2 (degrees AgK)10 15 20 25

Inte

nsity

(cou

nts)

-2000

0

2000

4000

6000

8000

10000

AgKα Pd filter

PSDs – flexibility? • Most lab-based PSDs optimized for CuKα

• MoKα – reduced efficiency even if it works…. • AgKα – good luck!

• Detector electronics may need better shielding

• Absorption edge from β filter difficult to miss with MoKα & AgKα!

2 (degrees AgK)10 15 20 25 30 35 40 45

Inte

nsity

(cou

nts)

0

5000

10000

15000

20000

SRM1976 500µm Lynxeye, Pd β filter

2 (degrees AgK)10 15 20 25 30 35 40 45

Inte

nsity

(cou

nts)

0

5000

10000

15000

20000Fit when K-edge effects modelled

Page 24: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

However - all is not lost….. • E.g. PSD not

optimised for MoKα …

• Calcium silicate hydrate (CSH) carbonation

• Fixed snaphots - 8°

• Damp sample vs dry

• Damp sample 3s time resolution

6.3 MPa (915psi) CO2 dry sample 24 hours 30s snapshots

2θ (MoKα) 10 12 14 16 18

Flowing 1atm CO2 damp sample 20 mins 3s snapshots

10 12 14 16 18

Page 25: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

2D detectors • Common at synchrotrons

• Increasingly in lab as well…

• Positional accuracy more problematic as more sources of error in calibration • Complex aberration functions • With lab wavelengths integrated data from multiple frames may

have to be merged…

• We never used ours much – too many Mn-rich samples • Fluorescence a real problem with either CuKα or CoKα • Al foil filter also works with 2D detectors but not ideal… • Currently sat in a box!

Page 26: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Multi-layer Mirrors…. • Technologies for multilayers developed over last 10 years

• The layer materials in use have evolved • Mirror corrosion issues solved

• Originally only CuKα widely available

• Now available in all wavelengths in divergent/parallel/ convergent

• The right mirror for the right application can be great

Fraunhofer 40mm CuKα W-based mirror (circa 2002)

Incoatec sealed 60mm CuKα Ni-based mirror (circa 2006)

Page 27: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Multi-layer mirror optics mirrors versus monochromators – pros & cons • Pros

• Choose from convergent, parallel or divergent • Very high intensities possible • Excellent Kβ suppression • Good resolution can be achieved in focussing geometry

• Cons • Wavelength specific • Kα1/Kα2 doublet • Past history of surface corrosion • Temperature sensitive

• Despite cons I have an unsealed CuKα focussing mirror

Page 28: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Not for those of a weak disposition…! Horror pic from 2005

• Eventually explained by ozone attack

• Sealed units or inert flowing gas solved problem… • or reduce Bremmstralung energy by reducing voltage

A nice new ‘mirror-like’ mirror After intensities started to drop…

Page 29: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

2 (degrees CuK)10 20 30 40 50 60

Inte

nsity

(cou

nts)

0

10000

20000

30000

CuKα focussing mirror – Pro (intensity) • Fluoroethylene carbonate (focussing mirror + 10° PSD)

• Structure indexed using a 4 minute dataset • Solved & refined with 3.5hr pseudo-VCT dataset → 140° 2θ

Page 30: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

CuKα focussing mirror – Con (odd profile) • Can be tricky to get Kα2 at 50% during manual alignment

• The change easily modelled but not ideal

• Alignment for max intensity often doesn’t give best result….

2Th Degrees9796959493929190898887

Cou

nts

4,500

4,000

3,500

3,000

2,500

2,000

1,500

1,000

500

0

-500

Si 0.00 %

SRM640c Kα2 = 63% (simple 2 line emission profile)

Page 31: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Multi-layer mirrors • Make sure you have the right one for the job…

• Parallel-beam mirror should have been fine for capillary transmission work

Two theta (degrees)26 28 30 32 34

Arbi

trary

inte

nsity

focussingpowderpeak positions

• Divergent ‘powder mirrors’ can be problematic • strange peak shapes,

poor resolution

Comparison of capillary data from jadarite with a ‘powder mirror’ and a focussing mirror

Focussing for transmission

Page 32: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Twin mirrors • Second mirror focuses onto scintillation detector

• decent peak resolution • excellent backgrounds • the data is still student-proof!

• Quite rare….and not cheap….

• Particle statistics aren’t great

2 Theta (degrees)34.0 34.5 35.0 35.5 36.0

Inte

nsity

(cou

nts)

0

200

400

600

800

1000

1200

1400 -0.6 mm -0.5 mm -0.4 mm -0.3 mm -0.2 mm-0.1 mm0.0 mm0.1 mm 0.2 mm0.3 mm0.4 mm0.5 mm0.6 mm

Al2O3 104 reflection with displacement for contoured sample

Oddity? 60mm Ni-based sealed primary with 40mm W-based unsealed secondary….

Page 33: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

2 (degrees CuK)80 85 90 95 100 105 110

Inte

nsity

(cou

nts)

-20

0

20

40

60

80

100

120

140 5º grazing incidence (0.2+1+0.2mm slits)symmetrical

micronized yttria

Grazing incidence with twin mirrors… • Secondary mirror → no loss of peak resolution

• Complex intensity/angle relationship but can be modelled • Dependence gets quite ‘interesting’ with sample displacement…

2 (degrees CuK)0 20 40 60 80 100

I asym

m/I sy

mm

0.0

0.5

1.0

1.5

2.0

Toraya equation 1 for 5º incident angleToraya equation 3 for 0.2mm beam height and 1mm detector slit

Calculated intensity/angle curve for the 5º grazing setup

2 (degrees CuK)20 40 60 80 100

Inte

nsity

(cou

nts)

0

500

1000

1500

2000

2500

3000

Toraya corrections (no displacement)

Incident angle refined to 5.5ºMirror recieving slit refined to 1.3mm

Rwp = 19.8%, GOF = 1.1

Page 34: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Increasingly Popular Application Areas • Combinatorial screening

• 2D detectors with well plates

• Pharmaceuticals • Polymorphs and structure solution

• Tomography

• Residual stress • 2D detectors

• Pair Distribution Function (PDF) • High energy

Page 35: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Pair Distribution Function

• Synchrotron beam-time a finite resource • Lab studies viable for some systems where resolution not as vital

• Was always possible but painfully slow with lab instrumentation

• Need very low noise to high scattering angles • PSDs desirable but efficiency ↓ as energy ↑

Apologies for the pun!

Page 36: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

PDF – instrument considerations • Want highest-energy to maximize Q-range, but…..

• MoKα – Qmax ~18 Å-1, 3kW LFF tubes • AgKα – Qmax ~22 Å-1, 1.5kW LFF tubes (50kV, 28mA)

• Capillary geometry desirable • Monochromator versus mirror?

• Trade-off between • Higher Q Ag Mo • Tube power Ag Mo • PSD efficiency Ag Mo

Page 37: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

2 (degrees - MoK)20 40 60 80 100 120 140

Inte

nsity

(cou

nts)

0

10000

20000

30000

40000

50000

60000

2 (degrees - MoK)

100 110 120 130 140 150

Inte

nsity

(cou

nts)

0

200

400

600

800

1000

Example - LiMn1/3Co1/3Ni1/3O2 • MoKα with Vantec PSD (10° window)

• Efficiency ~50% in use versus CuKα (N.B. higher tube power)

3 hour scan time (reflection) Radial Soller slits to reduce background 40kV 55mA to minimize Xe escape peaks without primary optic (the ‘good’ Vantec)

Page 38: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

LiMn1/3Co1/3Ni1/3O2

• PDFFIT just as happy with lab data

• Despite low resolution distortion away from average structure (red line) visible

Mn4+-O (1.91Å)

Ni2+-O (2.06Å)

Fit using PDFFIT of the average R-3m structure versus G(r )

Blowup shows the misfit in the TM-O environment

Page 39: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Conclusion • Faster is not always better…..

• Counts aren’t everything • less really can be more!

• Use best setup for the experiment, not the most expensive!

• In multi-user environments the single perfect instrument still doesn’t exist

• Competing demands of speed, flexibility and accuracy lead to trade-offs

Page 40: Laboratory instrumentation - NIST · Laboratory instrumentation the last 10 years Energy, Mining and Environment Pamela Whitfield Accuracy in Powder Diffraction IV, NIST, April 2013

Gazing into the future? • Better energy discrimination?

• Hopefully…

• Better angular resolution? • Probably not….

• Is the future in the lab 2D? • For some but not all

• High energy • As above….

• Do the instrument vendors have more rabbits to pull from the hat?