laryngeal and tracheal stents
TRANSCRIPT
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 1/15
Laryngeal and Tracheal Stents Treatment & Management
Background
Laryngeal and tracheal stents are solid or hollow absorbable or nonabsorbable tubes of various shapes, sizes, and
materials. Stents are used as primary treatment for lumen collapse or to stabilize a reconstructive effort of the larynx or
trachea to prevent collapse. Stents can be used for the larynx and the trachea individually, or they can be used
interchangeably or concomitantly. This article reviews (1) various laryngeal, tracheal, and combination stents; (2)
disease processes for which stents are used; and (3) new stent designs.
An image depicting laryngeal and tracheal stents can be seen below.
Side view of a Montgomery laryngeal stent.
For excellent patient education resources, see eMedicineHealth's patient education article Bronchoscopy.
History of the Procedure
The word stent is derived from Charles B. Stent, a British dentist who practiced in the late 19th century. Stent
developed material that was used for dental impressions and, later, to support skin grafts. In current practice, the word
stent refers to material that supports some form of tissue against collapse. Most often, stent describes devices that
maintain the lumens of tubular organs.
In 1965, Montgomery described the use of silicone prostheses for tracheal stenting. Surgeons in other subspecialties,
such as gastroesophageal, genitourinary, and vascular surgery, developed stents that remain useful in expanding
lumens in these areas. This evolution increased the interest of otolaryngologists to develop stents for the larynx and
tracheobronchial tree.
Problem
See the Indications section.
Indications
Laryngeal stents are often used to keep the airway expanded after surgical reconstruction or trauma. Occasionally,
laryngeal stents can be used for expansion of a disease process. Stents can be used for similar reasons in the trachea,
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 2/15
especially to maintain lumen patency to prevent death. Stenting in the tracheobronchial tree is usually used as a last
resort for severe conditions such as recurrent carcinoma and severe tracheal collapse that results in periods of
prolonged apnea.
The most common indication for laryngeal stenting follows reconstruction of laryngotracheal stenosis (LTS; see the
Medscape Reference articles Glottic Stenosis and Subglottic Stenosis in Children). Laryngeal stents can be used to keep
the laryngeal lumen open and the reconstruction supported and stable. Occasionally, laryngeal stents are usedfollowing trauma to the larynx resulting in laryngeal fracture or injury. Stenting may help maintain lumen patency and
prevent mucosal lacerations from scarring.
Laryngotracheal stenosis
The first reason to use stents in cases of LTS is to support the larynx, typically with some form of autologous cartilage,
after a reconstructive technique has been performed. A stent can be used to stabilize the cricoid plate once it has been
divided anteriorly or posteriorly, with or without cartilage placement, to keep the complex in an expanded formation
during healing. Stenting to help stabilize the laryngeal structure normally lasts for 2-6 weeks. Zalzal has showed that
when a cartilage graft is inserted, the minimum time necessary for a posterior cricoid split to heal in a distracted
manner is 2 weeks. If stenting is performed for a shorter duration, the graft may prolapse into the lumen.
For anterior graft placement only, stenting commonly lasts for a week or less, which usually occurs while a patient is
endotracheally intubated in an intensive care unit (ICU). If the cartilage graft is made into a boat formation with flanges
over the edge of the intraluminal portion of the graft, thus providing stabilization and preventing prolapse, see the
images below, then ventilation with an endotracheal tube is not necessarily used for support but for protection of the
airway while the edema resolves and the air leak around the reconstructive effort seals.
Note the diamond-shaped internal intraluminal component. The shape of the rib is seen on the reverse
side of the carved-out diamond-shaped wound.
Another anterior graft with a diamond shape. Note that it is approximately 1.7 mm long. Again, the intraluminal site is
seen facing up.
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 3/15
This helps prevent pneumomediastinum, pneumothorax, or both in concert. Often, a child can be extubated within a
few days following surgery. With this graft design, a stent is not always required because the reconstructive area is
already stabilized, provided no air leak has developed.
The second reason for stent placement in LTS repair is to counteract scar contraction. Theoretically, this requires
stenting for a 6-month period. In general, guidelines to determine the duration of stenting depend on the consistency of
the stenosis, the anatomical distortion of the disease process, and the stability of the reconstructive areas.
Laryngeal web or atresia
For a child with laryngeal web or atresia, a keel stent, shown below, is often placed to prevent the laryngeal web from
reforming. Frequently, this stent is left in place for several weeks.
Laryngeal keel.
Tracheal lesions
The purpose of stenting for tracheal lesions varies from palliation, to cure, to stabilization while a reconstructive effort
heals. In adults, primary cancer of the tracheobronchial tree or cancer from the head, neck, or chest that extends into
the tracheobronchial tree frequently causes lumen compromise and airway obstruction.
The intraluminal component can be removed with laser treatment, mechanical debulking, electrocautery,
brachytherapy, photodynamic therapy, or cryotherapy. A stent can then be placed to maintain the airway lumen
following debridement to counteract collapse or edema. Alternatively, stents can be placed that help compress any
lesion extending into the trachea or bronchi, without the need for debulking. Stents have been used successfully to
palliate patients with inoperable bronchogenic cancer, primary tracheal tumors, and metastatic malignancies. Placing a
stent in a patient with a terminal illness allows that patient to breathe comfortably and prevents death from
asphyxiation.
Tracheomalacia and bronchomalacia
Tracheomalacia and bronchomalacia (sometimes called tracheobronchomalacia) may be primary or secondary in
nature. These conditions usually occur in children, but they can be observed in adults. Tracheomalacia andbronchomalacia are termed primary if they arise from primary cartilage abnormalities of the trachea, such as immature
formation of the cartilaginous rings in neonates or relapsing polychondritis in children and adults. Primary
tracheomalacia is often observed with a tracheoesophageal fistula.
Secondary tracheomalacia or bronchomalacia is caused by extrinsic compression from a structure in the mediastinum.
These structures may be normal vascular structures (eg, aortic artery, heart) or abnormal vascular structures (eg,
aortic arch ring, pulmonary artery sling, other congenital vascular abnormality). Additionally, secondary
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 4/15
tracheomalacia or bronchomalacia can be caused by mediastinal cysts or neoplasms, including teratomas, lymphomas,
and neuroblastomas.
Stenting of the trachea following surgical resection or reconstruction
If tracheomalacia or stenosis occurs following the resection of lesions, stents can be placed in the trachea to prevent
scarring or to provide support for the operated segment while it heals.
In most situations, surgical correction as primary treatment for a disease process is preferred to stenting alone because
complications are decreased. However, situations arise in which patients cannot undergo formal corrective surgery,
and stent placement is the only way to prevent a severe apneic event or death. Tracheal stents have been placed
following unsuccessful tracheal repairs, postanastomotic stricture after sleeve resection for lung transplantation, and
following malacia from infections such as tuberculosis.
Following lesion excision
Any time the airway is opened to excise a lesion of the larynx, a stent can be considered for stabilization, scar
prevention, or airway protection while edema resolves and the area heals. Frequently, this requires several days of
intubation in the ICU, with the endotracheal tube acting as a temporary stent.
Types of stents
Different stents have different indications. Types include primarily laryngeal stents, primarily tracheal stents,
combination tracheal and laryngeal stents, and stents that can be used either in the larynx or trachea.
Laryngeal stents
If stenosis is confined to the larynx (ie, glottis, subglottis), stenting can be short- or long-term. Short-term stenting is
defined as stenting for less than 6 weeks. Long-term stenting is defined as stenting for more than 6 weeks.
Place a short stent for 6 weeks or less because granulation tissue forms at the lower end of the stent above the
tracheotomy, potentially leading to tracheal stenosis or collapse above the tracheotomy site. Use short-term stenting
for stabilization of cartilage grafts following laryngotracheal reconstruction (LTR) and/or for separation of mucosal
surfaces during healing following laryngeal trauma, repair of web formation or atresia, or excision of a laryngeal lesion.
Stents for these indications include Aboulker stents, silicone stents, Montgomery laryngeal stents, endotracheal tubes,
and laryngeal keels. See the images below.
Representative (noninclusive) sample demonstrating various sizes of Aboulker stents, ranging from 15 mm in diameter
on the left to 3 mm in diameter on the right. These stents are hollow and coated with Teflon.
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 5/15
An end view of an Aboulker stent, showing the central opening. These stents are hollow and coated with Teflon.
Side view of a Montgomery laryngeal stent.
Laryngeal keel.
Occasionally, long-term stenting is required when the trachea above the tracheotomy tube requires stenting for either
collapse or stenosis following reconstruction. In this case, a long stent wired to the trachea, shown in the images below,
or a tracheal tube (T-tube), such as the Montgomery T-tube, shown below, can be used.
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 6/15
Radiographic lateral neck view of a long stent connected to a metal Jackson
tracheotomy tube at the bright inferior portion of the picture. The stent is seen in the airway as an oblong translucent
area, with a rim of opacification around it that extends up through the larynx. A thin wire is seen connecting the stent
and the tracheotomy tube.
A long Aboulker stent wired to a metal Jackson tracheotomy tube.
A Jackson tracheotomy tube wired to a long Aboulker stent.
Montgomery T-tube (7 mm) stent with caps.
Relevant Anatomy
Larynx
When a laryngeal stent is necessary, place it through and above the true and false vocal cords or a significant distance
below the true vocal cords. If the stent is placed directly under the true vocal cords, granulation tissue can form and
scarring of the glottis can occur; if at all possible, avoid this problem because repair and correction of glottic stenosis is
difficult.
Trachea
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 7/15
In infants, the trachea is short. Thus, a tracheotomy tube can often stent the airway open itself, occasionally almost
down to the carina. If this does not suffice, then other corrective measures must be performed for tracheomalacia in an
infant.
Contraindications
Larynx
Absolute contraindications for use of a laryngeal stent include patients who are medically unstable and not able to
undergo general anesthesia or individuals who may have an allergic reaction to the material of a particular stent.
Relative contraindications include risks of particular stents in various patients that could result in a poor outcome or
complications. Historically, concern has been voiced regarding use of the Montgomery T-tube in young children (see
Montgomery stent).
Trachea
Absolute and relative contraindications for tracheal stents are similar to contraindications listed for laryngeal stents;
however, tracheal stents are often used as a last resort to prevent death by asphyxiation when other measures havefailed or are not feasible. Another relative contraindication is the placement of a nonabsorbable stent when other
techniques are available.
Imaging Studies
• The criterion standard for evaluating the larynx or the trachea is direct laryngoscopy and bronchoscopy.
• Fluoroscopy can be used to help define a narrowing of the trachea or abnormal radiographic appearance of the
larynx. The diagnostic information yielded from fluoroscopy is inferior to that of direct laryngoscopy and
bronchoscopy in the operating room (OR); however, radiopaque stents can be placed under fluoroscopic
guidance in adults and children.
• Occasionally, an imaging study such as MRI, CT scan, or plain x-ray film helps in diagnosing laryngeal lesions
that might require stenting or in defining the length of stenosis or malacia in the trachea.
Diagnostic Procedures
The criterion standard for evaluating the larynx and trachea is direct laryngoscopy and bronchoscopy. Using these
tools, the segment that requires stenting can be defined.
Histologic Findings
Graft vascularization
Most histological studies have been performed in animals. Concern has been raised that a stent could cause impaired
lumen healing, increasing the complication rate; however, a study on 36 white rabbits that had tracheoplasty with
autologous cartilage grafting with and without stents showed the rate of vascularization of the graft was equal amongboth groups and possibly slightly superior in the stented group at 10 days following insertion.
Mucociliary transport
The laryngotracheal mucociliary transport in guinea pigs with and without stenting was evaluated in studies conducted
in 1997 and 2000 by S.Y. Lee of the Department of Otolaryngology, Taiwan University, Taiwan. In these studies, stenting
appeared to preserve or increase the mucociliary function of laryngotracheal mucosa in the acute phase of stenting
compared to mucosa without stenting.[1, 2]
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 8/15
Tracheal stents
Histological evaluation has been accomplished with metal, silicone, and bioabsorbable stents.
For metal stents, the most common (eg, Palmaz stent) are placed intraluminally and are expanded with a balloon. One
study reported that the stent was overgrown with tracheal epithelium; granulation tissue formation was the only
complication cited. In experimental studies in cats, histological analysis revealed a mild inflammatory reaction withgranulation tissue present in all animals in which the stent was not manipulated. When the stent was overexpanded,
the reaction was more severe; epithelial ulceration that had sealed the lumen in most animals was noted.
In the same study, researchers opened and closed cats' tracheas with or without stent placement. Increased
inflammatory reaction, granulation tissue, and epithelial damage were observed in animals with stents as opposed to
control subjects without stents. Based on that study, the Palmaz stent, when used in a normal trachea on which an
operation has not been performed, provokes only a very mild and clinically insignificant inflammatory reaction when
appropriately expanded. However, once surgery has been performed on the trachea, the inciting inflammatory reaction
can cause formation of granulation tissue.
A study from Japan examined expandable metallic stents after tracheal patch reconstruction with omentoplasty.
Polypoid granulation tissue developed at both ends of the metallic stent. Histologically, epithelium was regenerated in
the patched area 4 weeks postoperatively; the patched area was covered with respiratory epithelium 12 weeks
postoperatively.[3]
Another study investigated the histocompatibility of 4 different types of material used to cover expandable Gianturco
metallic stents in dogs: polypropylene mesh, silicone-covered mesh, polyester mesh, and expanded
polytetrafluoroethylene. The polypropylene mesh appeared to be the most biocompatible of all the coverings and was
histocompatible to the airway.[4]
For bioabsorbable and silicone stents, a comparison of bioabsorbable airway stents, shown in the images below, with a
silicone airway stent and metallic stent was performed in Finland on rabbits.
Intraoperative picture showing a solid dissolvable airway stent next to a trachea in a 3-kg New Zealand white rabbit.
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 9/15
Solid spiral dissolvable stent produced for the trachea of a rabbit. Scale is in centimeters.
The bioabsorbable spiral stent was manufactured with self-reinforced poly-L-lactide material, and tracheomalacia was
created by extramucosal resection of cartilaginous arches of the cervical trachea. Hyperplastic polyps occurred at both
ends of the silicone stent, and the internal diameter of the stent became encrusted. The bioabsorbable stent and the
metallic stents were tolerated well.
Other experimental studies have been performed to evaluate histological events that occur when a bioabsorbable stent
is placed, as compared with silicone stents. Silicone stents in one animal study again showed a tendency to become
stenosed with encrusted material and to develop a hyperplastic bulge at both ends. Bioabsorbable stents of poly-L-
lactide material showed no foreign body reaction and had a tendency to penetrate into the tracheal wall. These stents
had been reabsorbed at 10-month follow-up.
Surgical Therapy
The placement of stents in the larynx or trachea is usually performed under anesthesia. Transcervical or transthoracic
approaches provide exposure to place the stents under direct vision. Stents can be sutured in place or affixed in other
ways.
Depending on the clinical situation, laryngeal and tracheal stents can be placed transluminally under endoscopic view
in the OR by a surgeon or they can be placed under fluoroscopic control in the radiology suite by a radiologist.
Laryngeal stents
• Aboulker stent
o The most common stent used for stability following LTR in children is the Aboulker stent. In the early
1960s, Aboulker introduced this cigar-shaped prosthesis, which is available in several different
outside diameters at a length of 120 mm, as in the image below. Aboulker originally used his stents in
adults, after cricoid plate reconstructions. In the late 1960s, Aboulker used a stent to decannulate 3 of
5 children after airway reconstructions. In the early 1970s, Grahne, Cotton, and Crysdale also used this
stent following LTR in children. Representative (noninclusive)
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 10/15
sample demonstrating various sizes of Aboulker stents, ranging from 15 mm in diameter on the left to
3 mm in diameter on the right. These stents are hollow and coated with Teflon.
o Zalzal noted that one of the benefits of the Aboulker stent is that the stent is made of highly polished
Teflon, which minimizes irritation and granulation tissue formation compared with other stents.[5]
However, the stent has been shown to cause some mucosal irritation, with granulation tissue
formation occurring at the superior or inferior end. Not uncommonly, the base of the epiglottis is
affected, and granulation tissue often forms in this area.o The Aboulker stent has been shown effective for counteracting scar contracture and for keeping
reconstructed areas stable during healing. Zalzal, Cotton, and others have had good success
decannulating children with severe LTS using the Aboulker stent in association with LTR.
o The Aboulker stent can be used in both a short and long form. In the short form, the stent is used for 6
weeks or less and is placed in the larynx and upper trachea. As with all laryngeal stents, place it
between the true and false vocal cords superiorly and suture it in place with a large Prolene stitch tied
externally to the strap muscles. Place a portion of an angiocatheter or small feeding tube over this
suture to decrease the chance of erosion through the strap muscles. Throw multiple knots for
identification during later stent removal.
o The Aboulker stent can also be used in the long form. When the long form is used, suture the stent to a
metal Holinger tracheotomy tube (Holinger T-tube); see the images below. Long stents are used when
the suprastomal area needs support or when long-term stenting (ie, > 6 wk) is necessary. Long stents
help prevent granulation tissue and secondary stenosis in the suprastomal area.
Radiographic lateral neck view of a long stent connected to a metal
Jackson tracheotomy tube at the bright inferior portion of the picture. The stent is seen in the airway
as an oblong translucent area, with a rim of opacification around it that extends up through the larynx.
A thin wire is seen connecting the stent and the tracheotomy tube.
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 11/15
o A long Aboulker stent wired to a metal Jackson tracheotomy tube.
o A Jackson tracheotomy tube wired to a long Aboulker stent.
• Montgomery stent
o The Montgomery T-tube, shown in the image below, is a silicone stent with a long center lumen and a
smaller lumen projecting from the side of the stent at either a 90° or 75° angle. Ensure that the upperend extends through the true and false vocal cords. The lower end can extend all the way to the carina,
depending on the length of the trachea and stent. The side lumen extends through the tracheostoma. If
long-term stenting is required, this tube can be an alternative to a wired stent-tracheotomy complex.
Montgomery T-tube (7 mm) stent with caps.
o The Montgomery T-tube has been used successfully to stent the adult larynx and trachea following
reconstruction and to stent areas of malacia and stenosis. Concern exists regarding the safety of the
stent in children because the internal diameter of the stent is small and the tube could become plugged
with dried secretions, which could result in obstruction of the stent and could end in airway
obstructive symptoms and death if not promptly removed. With an indwelling Montgomery T-tube,
plug the side lumen as much as possible and intermittently perform proper suctioning to prevent
plugging.
o Stern and Cotton evaluated the Montgomery T-tube in children, reporting on 26 children whounderwent laryngeal procedures including LTR and cricotracheal resection.[6] Most of the stents used
had lumens wider than 8 mm in diameter; only 1 had a diameter of 6 mm. Stern et al used the stents
from 2 weeks to almost 2 years and did not observe increased granulation tissue formation in children
who had stents for longer periods of time. The researchers concluded that the Montgomery T-tube is
safe to use in children. In France, Froehlich et al used Montgomery T-tubes in 12 children for an
average of 6 months and noted that complications include removal of the T-tube by the child, forward
migration resulting in tube expulsion, formation of granulation tissue, and clogging.[7] Neither set of
researchers reported deaths from stent malfunction.
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 12/15
• Silastic sheet (Swiss roll)
o This stent, shown in the image below, was reported by Evans in 1977 as a form of stenting for
laryngotracheoplasty.[8] The silastic sheet was initially one of the main stents used following reconstructive
efforts in children. The silastic sheet is rolled up and inserted into the larynx and upper trachea, where it is
fixed in place with a suture. The roll has a constant tendency to unroll, producing general pressure on the
mucosa. This pressure allows obliteration of any dead space and allows mucosal regeneration to occur.
Because of its propensity to form granulation tissue, the silastic sheet was replaced by the Teflon Aboulkerstent
.
Roll of silastic sheeting.
o Foam and a finger cot were used initially in the larynx, especially following acute laryngeal traumatic
injuries. Most of these stents have been replaced by Aboulker stents, Montgomery T-tubes, or
Montgomery laryngeal stents.
• Endotracheal tubes: Originally reported by Brick in 1970, the use of endotracheal tubes made from polyvinylchloride was reported again from a series in India in 1995. These tubes were used to stent the laryngotracheal
area following LTR. Fibrosis, inflammation, and granulation tissue occurred after 4 weeks. Brick believed these
stents should only be used for 4 weeks or less to prevent permanent laryngeal damage.
• Silicone stents: Silicone stents used in the larynx have mainly been reported in adults. These stents are placed
endoscopically after dilating the larynx up to 18 mm. If dilation does not enlarge the larynx enough to place the
stent, carbon dioxide laser excision of the scar is occasionally performed to enlarge the subglottic stenosis.
Suspend the patient in the OR, and place the stent through the vocal cords into the subglottis. Place an external
stitch by passing it through the skin into the stent and then out to the skin again, and secure it in position. In
adults, place stents 5-30 mm below the vocal folds. Complications and concerns of this stent include transient
vocal fold edema, stent migration, and recurrence of stenosis following stent removal, which has occurred even
after leaving stents for 11 months.
• Inflatable stents: Over the last few years, inflatable stents composed of a small balloon attached to a port to
allow expansion with air have been evaluated at the Cleveland Clinic. These stents have been used only inexperimental canine models.[9] Inflatable stents are inflated slowly and expanded over 3-46 days.
Complications included small superficial ulceration of the true vocal cords, minor polypoid changes, and
granulomas. Of these experimental dogs, 20-30% exhibited inflammation extending into the underlying
cartilages. Based on this information, the stent was associated with minimum local tissue reaction.
Tracheobronchial stents
Tracheobronchial stents can be broadly divided into metal or silicone stents. General differences exist between these 2
categories.
Silicone stents were the first developed in the mid 1960s. Most studies report that the tracheal mucosa tolerates long-
term use of these stents relatively well. Silicone stents can be left in place for several years and, generally, are easily
removable. Because this stent has the propensity to slide well over the mucosa, stent migration is the most commonproblem. Metal stents are usually composed of a mesh and become incorporated into the mucosa, making them
extremely difficult to remove in some cases.
This same propensity for mucosa to grow around the metallic stents allows for better mucociliary clearance and
function. Mucociliary flow is hindered by the silicone stents, which are mostly solid. Silicone stents are thicker than
metallic stents and, thus, cause more restrictive airflow than the thin metal stents that can self-expand or be expanded
to adhere closely to the tracheal wall. Additionally, metal stents have the potential to be expanded intermittently as the
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 13/15
airway enlarges or changes. Metal stents are more distensible and conform better to the airway than silicone stents,
especially if the airway is somewhat tortuous.
Generally, more reaction and granulation tissue is seen with metallic stents than with silicone stents, although reports
vary, some significantly. Metallic stents can be coated with various materials that seem to reduce the rate of granulation
tissue formation.
Metal stents are easier to use in the distal trachea and bronchi because they are mesh and do not obstruct primary or
secondary bronchi that come into contact with the side of the stent. A bronchus could become obstructed with a
silicone stent, unless a port has been fashioned into the side of the stent. The Dumon stent, shown in the image below,
comes with some stents open to the side.
The Dumon stent with its opening for the right mainstem bronchus.
Generally, both types of stents are structurally strong, and they resist forces that cause stent collapse. Metal stents are
usually used in the radiological suite because they are radiopaque; silicone stents are used in the OR because they are
radiolucent, do not expand, and generally require some type of affixation.
Specific stents are discussed below. Metal stents include the categories of balloon- and self-expandable stents.
Bioabsorbable stents are also discussed and compared.
• Balloon-expandable metallic stents
o Palmaz stent
This stent (Johnson & Johnson; New Brunswick, NJ and Interventional Systems; Warren, NJ)
was initially developed as a vascular prosthesis. The Palmaz stent reportedly is the most
common stent used in children, in part because of its small size.
The stent consists of 150-µm slotted stainless steel in a tubular mesh configuration with
lengths varying from 10-40 mm. A balloon of 6- to 10-mm diameter fits inside the stent for
manual expansion of as much as 6-12 mm. An appropriate size for expansion of the trachea
and bronchi in children is 8 mm and 6 mm in diameter, respectively. After balloon expansion,
the stent does not exert a continual outward pressure on the airway wall.
The stent has been used in primary tracheomalacia or bronchomalacia, external compression
of the trachea or bronchi, or collapse of the trachea or bronchi from previous surgery. Fraga et al from the Hospital for Sick Children in Toronto, Ontario, Canada, reported results in
16 patients who required 30 stents. Half of these patients had primary tracheomalacia, and all
did well, with relief of symptoms following stent placement. The stents were easily removable,
and granulation tissue was seen in approximately 20% of the group. Additionally, the children
who required stenting following stenosis at the site of a previous surgical repair did well and
underwent stent removal without complications, except for some granulation tissue. In the 4
children who had stents placed for external compression, 2 had stent migration and 1 had
bronchial erosion. Three patients died of underlying cardiac disease, but stenting provided
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 14/15
good palliation of respiratory symptoms. Sommer and Forte believe use of this stent is the
preferred management for significant tracheobronchomalacia in children.
o Strecker stent
This stent is made of a tantalum filament that is structured into a cylindrical wire mesh. The
stent is flexible, whether compressed or expanded. When expanded, the stent does not change
length.
The Strecker stent is 2-4 cm long and can be expanded from 8 to 11 mm.
This stent is being used successfully in tracheobronchial obstructions.
• Self-expanding metallic stents: These stents have "memory" that allows a return to normal shape after
compression for placement in the airway. The device self-expands but occasionally requires balloon inflation
for complete expansion. Self-expanding metallic stents include the Gianturco-Z (William Cook, Bjaeverskov,
Denmark), the Wallstent (Boston Scientific; Natick, Mass), and the Nitinol.
o Gianturco-Z stent
Developed by Cesare Gianco in the 1980s, this stent was initially designed for stenting
obstructions in the vascular system; it has been the most widely used of the expandable
metallic stents in adults. The Gianturco-Z is composed of 460-µm stainless steel filaments
arranged in a zigzag configuration. The stent has been modified several times since initial
development. One modification added small hooks to the outer perimeter to prevent
migration, but they contributed to difficulty with stent removal.
The diameter of the stent when expanded is 15-40 mm. The stent is available in 2- and 2.5-cmlengths. Two stents can be combined to double the length.
The Gianturco-Z stent has been used to expand benign disease and with posterior anastomotic
strictures, tracheal stenosis, and tracheobronchomalacia. The stent exerts good radial force
and does not shorten when deployed, but it has a tendency to spring forward if released too
quickly.
Complications are sometimes reported and include breakdown or unraveling of the stent and
fatal hemoptysis after erosion into the pulmonary artery 10 days after insertion.
o Wallstent
The Wallstent (Schneider; Minneapolis, Minn) is a stainless steel device composed of
approximately 15-20 braided (100-µm diameter) filaments. The filaments are arranged in a
criss-cross fashion to form a cylindrical mesh. The individual alloy filaments make a precise
number of turns to maintain the mesh pattern and are good in tortuous airways because they
maintain shape. Stent diameters range from 6-25 mm; lengths range from 2-7 cm. Ensure that the diameter of
the stent is at least 2 mm wider than the diameter measured at the proximal region of the
normal airway. The stent exerts good radial force and good flexibility but shortens to 20-40%
upon deployment. The Wallstent can be delivered with a rigid bronchoscope or flexible
telescope.
Initially, this stent had difficulty with tumor in-growth, and a silicone body covering was
developed. An important advantage of this stent is the ability to cut small openings into the
mesh of a stent that lies across bronchial openings.
o Nitinol/InStent/Ultraflex stent
This stent is thermally triggered and changes shape in response to temperature changes. This
response is known as a Marmen effect and occurs in certain metals or alloys that distort at low
temperatures (martensitic state) then revert to the original shape when reheated (austenitic
state). The Nitinol wire is heated and made into a helical shape and is then cooled for deployment.
With release into the target site, high body temperature causes the Nitinol stent to coil back to
its original helical shape. Alternatively, a current of 1.5-3 amperes or 3-5 volts can be applied
to this stent for 1-2 seconds to heat it to 40°C, thus converting it to the fully expanded state.
A group from Japan noted that Teflon coating of this stent decreased tissue reactivity. The
stent was found to have poor visualization under fluoroscopy. Haeck compared this stent to
the Strecker stent and found it preferable because of ease of use, greater flexibility, and faster
deployment.
7/27/2019 Laryngeal and Tracheal Stents
http://slidepdf.com/reader/full/laryngeal-and-tracheal-stents 15/15
• Silicone stents: Available stents include the Montgomery T-tube, Dumon (Novatech; Aubagne, France), Reynder
(Reynder's Medical Supply; Lennik, Belgium), dynamic (Rusch AG; Duluth, Ga), Polyflex (Rusch AG; Duluth, Ga),
and Nova (Novadis; Saint-Victoret, France). The Montgomery T-tube and laryngeal stents were discussed
earlier. A form of the Montgomery T-tube can be extended to include the trachea and upper bronchi.
o Dumon stent
In the late 1980s, Dumon introduced a tracheobronchial stent that could be inserted through a
bronchoscope. This stent is a cylindrical silicone stent with external studs placed at regularintervals to prevent stent migration and to reduce mucosal ischemia by limiting contact with
the airway wall. The Dumon is the most widely used silicone stent. The stent is widely
available in varying widths and lengths from 10-35 mm. Once placed in the airway, the stent
can be adjusted with a forceps and bronchoscope.
These stents have been used more commonly in the adult population. Dumon's series of 574
stents placed between 1987 and 1994 indicated the stents were tolerated for as long as years.
The Dumon stent has a thick wall, but a thin-walled stent in pediatric sizes has been
developed.
o Reynder stent: This stent is a cylindrical silicone prosthesis that is more rigid than regular silicone
tubes but requires a special introducer and a bronchoscope for placement.
o Dynamic stent: The dynamic stent is a silicone Y-stent with anterior and lateral walls reinforced to
simulate the tracheal wall. Special forceps are available for insertion within the rigid laryngoscope.
o Polyflex: This device is a self-expandable stent made of polyester wire mesh within layers of silicone.
o Novastent: The Novastent is a thin silicone sheet containing a small metallic hoop of Nitinol alloy. The
silicone bands on the ends are designed to prevent migration.
• Bioabsorbable tracheal stents: Corpela from Finland described absorbable stents in animal models. These
stents have been efficacious in preventing airway collapse in rabbit models and in created tracheomalacia,
although no current publications report studies on the use of bioabsorbable stents in humans.