lee transact[ons on instrumentation and …

5
lEE" TRANSACT[ONS ON INSTRUMENTATION AND MEASUREMENT. VOL. 42. NO.2. APRIL [993 [3[ Automatic Inductive Voltage Divider Bridge for Operation from 10 Hz to 100 kHz ;' that is in phase with the Channel A signal and provides the reference signal for the detector. The detector is iso- lated from the measurement point (the tap of the Test IVD) using transformer To. This allows both the test signal V and the detector input Vo to be grounded. When the quadrature error component is much larger than the in-phase component, an appropriate quadrature voltage VQ is injected through transformer TQ to reduce the quadrature error to a level that can be measured by the detector without compromising the in-phase measurement accuracy. The amplitude of VQ' which is a function of the ratio of resistors R. and R2' is used to compute the quad- rature error of the Test IVD. For complete automation, bridge components either span the frequency range or are remotely switched for low- and high-frequency ranges. For example, two sets of iso- lation and injection transformers (TI> TQ' and To) are needed to cover the frequency range of 10 Hz-IOO kHz. These transformers can be selected either manually or re- motely. Also, the voltage divider is adjustable in three ranges using different resistors for R[ and R2. [ I I I I Svetlana Avramov, Nile M. Oldham, Dean G. Jarrett, and Bryan C. Waltrip Abstract-A bridge to calibrate programmable and manual inductive voltage dividers is described. The bridge is based on a programmable 30 b binary inductive voltage divider with ter- minallinearity of :to.1 ppm in phase and :t2 ppm quadrature at 400 Hz. Measurements of programmable test dividers can be automated using software developed to align the bridge com- ponents and perform an automatic balance. I. INTRODUCTION I NDUCTIVE voltage dividers (IVD's) and bridges to compare them have been extensively described in the literature. Most of the work has concentrated on manually operated, decade IVD's for use in the low audio fre- quency range [1]-[4], with some mention of operation at higher frequencies [5]. In the late 1960's, Hoer described a fixed ratio binary inductive voltage divider (BlVD) de- signed to operate at 1 MHz [6]. Consequently, several programmable BlVD's were developed for use in auto- matic measurement systems [7]-[9]. Further development was motivated by the need for a system to support new commercial programmable IVD's. The BlVD-based bridge described in this paper was originally developed to calibrate specific ratio ranges of a programmable IVD used for temperature control in the Zeno project (a fluid dynamics experiment scheduled to fly on the Space Shuttle in 1993 [10]). II. INDUCTIVEVOLTAGEDIVIDERBRIDGE The primary objective of the new bridge is to automate IVD calibrations rather than to improve accuracy. With this in mind, we adopted an approach that made use of commercial instrumentation where possible. A programmable, digitally synthesized, sine-wave gen- erator was selected as the voltage source for the IVD bridge, which is shown schematically in Fig. 1. This source produces two amplitude- and phase-adjustable sine waves from 10 Hz to 100 kHz. Channel A of the voltage source supplies the sinusoidal test voltage V that is iso- lated through transformer TJ. Channel B supplies a second sinusoidal voltage that is in quadrature with V and pro- grammable in amplitude. The Ref output is a TTL signal Manuscript received June 12, 1992; revised September 21, 1992. This work was supported in part by NASA under Contract NAS 3-25370. The authors are with the Electricity Division, National Institute of Stan- dards and Technology. U.S. Department of Commerce, Technology Administration, Gaithersburg, MD 20899. IEEE Log Number 9206513. L III. ALIGNMENTPROCEDURE A commercial lock-in amplifier serves as a detector for the bridge. Lock-in amplifiers are able to resolve small in-phase signals in the presence of large quadrature com- ponents. However, because of alignment complexities, they are seldom used to their full potential. An alignment procedure (described below) allows the lock-in amplifier to provide direct reading of in-phase and quadrature error components without a precise null balance. An automatic alignment procedure is performed by phase shifting the reference signal in the detector to find the in-phase and quadrature projections of the voltage dif- ference between the Standard and Test IVD's, Vs - V" on the Standard IVD output Vs. This procedure consists of the two steps described below. A. Detector Reference Alignment (See Fig. 2) The Standard and Test IVD's are set to the nominal test ratio, and the voltage difference between their taps VOl is measured. The Standard divider is then set to 1.001 times the nominal test ratio, and the corresponding difference VD2 is measured. We assume that linear step D is in phase with the output of the Standard IVD, and the magnitude of this step can be calculated from measurement data. The ~ U.S. Government work not protected by U.S. copyright

Upload: others

Post on 04-Dec-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

lEE" TRANSACT[ONS ON INSTRUMENTATION AND MEASUREMENT. VOL. 42. NO.2. APRIL [993 [3[

Automatic Inductive Voltage Divider Bridge forOperation from 10 Hz to 100 kHz

;'

that is in phase with the Channel A signal and providesthe reference signal for the detector. The detector is iso-lated from the measurement point (the tap of the Test IVD)using transformer To. This allows both the test signal Vand the detector input Vo to be grounded.

When the quadrature error component is much largerthan the in-phase component, an appropriate quadraturevoltage VQ is injected through transformer TQ to reducethe quadrature error to a level that can be measured by thedetector without compromising the in-phase measurementaccuracy. The amplitude of VQ' which is a function of theratio of resistors R. and R2' is used to compute the quad-rature error of the Test IVD.

For complete automation, bridge components eitherspan the frequency range or are remotely switched for low-and high-frequency ranges. For example, two sets of iso-lation and injection transformers (TI> TQ' and To) areneeded to cover the frequency range of 10 Hz-IOO kHz.These transformers can be selected either manually or re-motely. Also, the voltage divider is adjustable in threeranges using different resistors for R[ and R2.

[

I

III

Svetlana Avramov, Nile M. Oldham, Dean G. Jarrett, and Bryan C. Waltrip

Abstract-A bridge to calibrate programmable and manualinductive voltage dividers is described. The bridge is based ona programmable 30 b binary inductive voltage divider with ter-minallinearity of :to.1 ppm in phase and :t2 ppm quadratureat 400 Hz. Measurements of programmable test dividers can beautomated using software developed to align the bridge com-ponents and perform an automatic balance.

I. INTRODUCTION

INDUCTIVE voltage dividers (IVD's) and bridges tocompare them have been extensively described in the

literature. Most of the work has concentrated on manuallyoperated, decade IVD's for use in the low audio fre-quency range [1]-[4], with some mention of operation athigher frequencies [5]. In the late 1960's, Hoer describeda fixed ratio binary inductive voltage divider (BlVD) de-signed to operate at 1 MHz [6]. Consequently, severalprogrammable BlVD's were developed for use in auto-matic measurement systems [7]-[9]. Further developmentwas motivated by the need for a system to support newcommercial programmable IVD's.

The BlVD-based bridge described in this paper wasoriginally developed to calibrate specific ratio ranges of aprogrammable IVD used for temperature control in theZeno project (a fluid dynamics experiment scheduled tofly on the Space Shuttle in 1993 [10]).

II. INDUCTIVEVOLTAGEDIVIDERBRIDGE

The primary objective of the new bridge is to automateIVD calibrations rather than to improve accuracy. Withthis in mind, we adopted an approach that made use ofcommercial instrumentation where possible.

A programmable, digitally synthesized, sine-wave gen-erator was selected as the voltage source for the IVDbridge, which is shown schematically in Fig. 1. Thissource produces two amplitude- and phase-adjustable sinewaves from 10 Hz to 100 kHz. Channel A of the voltagesource supplies the sinusoidal test voltage V that is iso-lated through transformer TJ. Channel B supplies a secondsinusoidal voltage that is in quadrature with V and pro-grammable in amplitude. The Ref output is a TTL signal

Manuscript received June 12, 1992; revised September 21, 1992. Thiswork was supported in part by NASA under Contract NAS 3-25370.

The authors are with the Electricity Division, National Institute of Stan-dards and Technology. U.S. Department of Commerce, TechnologyAdministration, Gaithersburg, MD 20899.

IEEE Log Number 9206513.

L

III. ALIGNMENTPROCEDURE

A commercial lock-in amplifier serves as a detector forthe bridge. Lock-in amplifiers are able to resolve smallin-phase signals in the presence of large quadrature com-ponents. However, because of alignment complexities,they are seldom used to their full potential. An alignmentprocedure (described below) allows the lock-in amplifierto provide direct reading of in-phase and quadrature errorcomponents without a precise null balance.

An automatic alignment procedure is performed byphase shifting the reference signal in the detector to findthe in-phase and quadrature projections of the voltage dif-ference between the Standard and Test IVD's, Vs - V"

on the Standard IVD output Vs. This procedure consistsof the two steps described below.

A. Detector Reference Alignment (See Fig. 2)The Standard and Test IVD's are set to the nominal test

ratio, and the voltage difference between their taps VOl ismeasured. The Standard divider is then set to 1.001 timesthe nominal test ratio, and the corresponding differenceVD2is measured. We assume that linear step D is in phasewith the output of the Standard IVD, and the magnitudeof this step can be calculated from measurement data. The

~U.S. Government work not protected by U.S. copyright

111f!

,...

~

132 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. VOL. 42. NO.2. APRIL 1993

-

TTL REFERENCE SIGNAL

Fig. I. Bridge for calibration of both manual and programmable IVD's.

'::'

Fig. 2. Vector representation of the detector reference alignment proce-dure.

lock-in amplifier records the magnitude and phase of themeasured signals, and by vector algebra, it is possible tocalculate the in-phase P and quadrature Q errors:

d = Vbl + V~ - 2 VOlVm cos cf> (1)

a = cos-I (Vb I + d - V~)/2VOID

P = VOlcos a, and Q = VOl sin a. (3)

By performing these two measurements, the detector ref-erence signal can be aligned with the linear step, allowingP and Q to be read directly from the detector display.

B. Quadrature Alignment (See Fig. 3)The Standard and Test IVD's are set to the nominal test

ratio. The phase of the quadrature injection signal VQ isset to 90° at the appropriate magnitude. The voltage dif-ference VD3is measured. A second measurement VD4isperformed after phase shifting the quadrature injectionsignal by l1cf>s. These two measurements provide enoughinformation to compute the correction ~ for the phase an-gle of the quadrature injection signal.

l1P = VD4 cos 'Y - VD3 cos {j (4)

l1Q :;:: VD4 sin'Y - VD3sin {j (5)

~ = arc tan «l1Q/ l1P» - l1cf>s/2. (6)

10.

INJEC110N IN QUADRAJIJRE

\WH DEIECIOR REfERENj:E

DEIECIOR

Fig. 3. Vector representation of the quadrature injection alignment pro-cedure.

When properly aligned, the detector is able to resolve anin-phase signal that is 40 dB below the quadrature signal,allowing the bridge to be balanced with a single iteration.

(2) IV. BINARY INDUCTIVE VOLTAGE DIVIDER

A 30 b BlVD had been developed to measure the dif-ferentiallinearity of a decade IVD used in a temperaturebridge for the Zeno project [10]. For this application, aresolution of one part in 109 was required, and it was feltthat a binary divider was a good choice because it offereda structure with a different error pattern than the decadedivider. Several BlVD designs were considered, includ-ing a technique similar to that described in [8], whichcombines inductive and resistive dividers to achieve therequired resolution. However, we thought that electronicnoise from the resistive divider (a multiplying digital-to-analog converter) might cause problems at the low signallevels, so it was decided to construct a relay-switched 30b BlVD similar to the 20 b BlVD's described briefly in[6] and [8]. The BlVD consists of n center-tapped trans-formers that can be connected to form 2n different ratios

(230 is approximately 109). We later decided to use this30 b BlVD as a standard in the bridge described abovefor calibrating programmable IVD's.

Shown schematically in Fig. 4, the BlVD consists offour relay-switched, binary transformers that are con-

STANDARD TEST

VOLTAGESOURCE IIVD IVD

REF CHANNELCHANNELI I TO TD r J :ARDB A

AVRAMOV el al.: 1VD BRIDGE FOR OPERATION FROM 10 Hz TO 100 kHz~

Section 1(7 Bita)

Magnetizing

Input

RaUoOUtput*

133

Section 2(8 Bits)

Section 3(8 Bits)

Section 4(7 Bits)

Fig. 4. 30 b BlVD.

trolled via the General Purpose Interface Bus (GPIB). Thefirst section is a two-stage 7 b binary inductive voltagedivider designed to operate at low frequencies. It consistsof a magnetizing winding and seven separate ratio wind-ings in a binary sequence. The second section consists of8 b coupled to the first by two-stage techniques. The thirdand fourth sections are small single-stage binary trans-formers with 8 and 7 b respectively.

One of the advantages of the binary divider over theconventional decade divider is that it takes only about onethird as many switches to achieve the same resolution. Adisadvantage is that all of the relays are in series in thesignal path, increasing the effective winding resistance.To minimize this influence, relays with less than 0.01 ncontact resistance and 0.001 n repeatability were se-lected.

Each transformer consists of a twisted pair connectedto form a center tap. The winding technique provides ex-cellent symmetry, resulting in a well-defined center tapover a wide frequency range. The disadvantage of thetwisted pair is that it produces a large interwinding ca-pacitance. The BlVD is designed to have optimum per-formance in the low audio frequency range when all foursections are used. As the operating frequency increases,the most significant sections are switched out to maintainmaximum accuracy. The sections used (and subsequentresolution) at various frequencies are given in Table I.

A. BlVD Performance Tests

The simplest IVD is a I b BlVD, which is simply acenter-tapped autotransformer. It is relatively easy to testthe symmetry of the center tap by measuring the tap volt-age with the input leads in direct and reversed positions[see Fig. 5(a»). If a second transformer with half the turnsof the first is wound on the same core, ratios of 1/ 4 or3 /4 may be obtained by connecting this winding, asshown in Fig. 5(b). Ideally, this connection should not

-

TABLE IBlVD RESOLUTION ON DIFFERENT FREQUENCY RANGES

load the first transformer since the voltage developedacross the second transformer is approximately the sameas the voltage across 1/2 of the first transformer. Theextent to which this second transformer degrades the cen-ter tap of the first can be measured using the reversingtechnique, with and without the second transformer con-nected. A 3 b divider was constructed and characterizedusing this bootstrap approach.

Uncertainties En for the center taps of the first 3 b onthe standard divider are given by

whereEnis the difference between the direct and reversed po-

sitions (VDn - VRn)/ V,n is the bit number (the errors are measured with lower

bits connected),VDnis the voltage difference between the taps of the

BlVD and the standard divider n in the direct position,VRn is the voltage difference between the taps of the

BlVD and the standard divider n in the reversed position,and

V is the voltage applied to both dividers.

"

I'j~.I.!..~'

Frequency Sections Total Number Resolution

Range Used of Bits (ppm)

10 Hz-2 kHz 1,2,3,4 30 0.0012 kHz-20 kHz 2,3,4 23 0.120 kHz-IOO kHz 3,4 15 30

E. = 1/2 E] (7)

E2 = E. + 1/4 E2 (8)

E3 = E2 + 1/8 E3 (9)

f1

...*

134 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 42. NO.2. APRIL 1993

TABLE IITHE BlVD BIT ERRORS AND BIT TRANSITION ERRORS FOR THE FIRST 3 BAT

400 Hz

(a)

0

.0.875

1/2

0.75

.

0.25

:H0-1 0-2 0-3

(b)

Fig. 5. (a) Reversing technique. (b) Standard 3 blVD.

The 3 b standard divider was then used as the standardto calibrate the errors of the first 3 b of the programmable30 b BlVD. The BlVD bit error bn is given by

bn = (VOn - Vo)/V (10)

whereVOnis the voltage difference between the taps of the

standard divider and the BlVD with both set to ratio 2 -n,Vo is the voltage difference between the taps of the

standard divider and the BlVD with both set to ratio O.Because of the binary structure of this instrument, sig-

nificant errors are likely to occur at the bit transitions. Forexample, the binary code for ratio 0.5 is 10000000 . . . ,where the most significant bit (MSB) is on and the loweror least significant bits (LSB' s) are off. In this 'condition,the output is connected directly to the center tap of thefirst transformer. Reducing the ratio by one LSB producesthe binary code 0111 1111 .. . . In this condition, theoutput is connected through the center taps of all 30 trans-

° These ratios were obtained by setting a single bit and reversing trans-fonner.

TABLE IIITHE BlVD BIT ERRORS AND BIT TRANSITION ERRORS FOR THE MSB

formers. We call the difference between these errors thebit transition error en and define it as

en = (VOn - VLSB - VLn)/V (11)

whereVLSBis the ideal voltage of the least significant bit,VLn is the voltage difference between the taps of the

standard divider set to ratio 2 -n and the BlVD set to ratio2-n - 1 LSB.

Early measurements [11] indicated that the bit transi-tion errors below the third bit contribute no more than onepart in 108. Both the bit errors and the bit transition errorsof the 30 b BlVD were measured using the 3 b standarddivider described above. The total uncertainties Un rep-resent a combination of the center-tap uncertainties of thestandard BlVD En and the random uncertainties of thebridge measurements. Results of measurements at 400 Hz,

.as well as the estimated uncertainties, are summarized inTable II.

Similar test results at other frequencies are given in Ta-ble III. For simplicity, only figures for the MSB (nominalratio 0.5) are provided.

The results in Tables II and III were obtained by com-paring the 30 b BlVD to a characterized 3 b standard di-vider using the bridge described above. The uncertaintiesare the same for both bit and bit transition errors. TheBlVD was also tested at decade ratios. The comparisonbetween binary and decade structures is quite powerful inthe way that it exercises many of the BlVD bits, ratherthan just the most significant bits. Results of the tests madeusing the standard decade IVD calibration system [12], at

Bit Errors (bn) and Bit Transition ErrorsUncertainties (Un) in ppm (en) in ppm

NominalRatios In-Phase Quadrature In-Phase Quadrature

"'v) I I 1/2 DETECTOR0.125 0.04 :t 0.11 0.07 :t 1.00 0.01 0.030.25 0.03 :t 0.06 0.30 :t 1.00 0.03 0.200.5 0.05 :t 0.04 0.70 :t 0.85 0.02 1.500.75° 0.09 :t 0.06 0.60 :t 1.00 0.02 0.300.875° 0.07 :t 0.11 0.60 :t 1.00 0.02 0.60

Bit Errors (b,) and Bit Transition ErrorsUncertainties (Ut) in ppm (e,) in ppm

Frequency(kHz) In-Phase Quadrature In-Phase Quadrature

0.010 1.00 :t 0.40 4.50 :t 0.50 0.80 7.000.100 0.10 :t 0.06 0.05 :t 0.10 0.05 0.15I 0.02 :t 0.04 I.O :t 0.50 0.30 3.80

10 1.00 :t 0.11 2.50 :t 1.50 1.30 1.80100 2.50 :t 0.90 43.0 :t 6.00 8.00 12.3

AVRAMOV el al.: IVD BRIDGE FOR OPERATION FROM 10 Hz TO 100 kHz

TABLE IVTHE BlVD ERRORS BASED ON THE STANDARD DECADE IVD

100 Hz and 1 kHz, are given in Table IV. Measurementuncertainties were :1::0.5ppm for the in-phase and :1::5ppmfor the quadrature components. The results of measure-ments made using these two systems agree to within theircombined uncertainties.

V. CONCLUSIONS

A bridge for calibrating programmable IVD's has beendescribed. A procedure to align the bridge detector wasdeveloped to simplify balancing by providing direct read-ing of the in-phase and quadrature error components. A30 b BlVD, which operates from 10 Hz to 100 kHz, servesas the standard for the new bridge. Once the alignmentprocedures have been completed, automatic measure-ments can normally be performed in less than 1 min/testpoint. In general, the accuracy of measurements made us-ing this bridge is comparable to those made using the bestmanual IVD standards.

Since complete characterization of the BlVD errors islaborious, future work will include an analysis of the bitand bit transition errors to reduce complexity and possiblyenhance accuracy. Also, further study is needed to char-acterize the bridge at higher frequencies where ratio un-certainties are expected to be on the order :l::0.3Fppm outto 100 kHz, where F is the test frequency in kilohertz.

--

135

ACKNOWLEDGMENT

In addition to providing financial and staff support forthe Zeno project, Prof. R. W. Gammon of the Universityof Maryland has promoted the continuation of this workto develop new wide-band IVD standards at NIST. Theauthors also wish to thank R. H. Palm and A. D. Koffmanfor their assistance in developing the BlVD.

REFERENCES

11] T. L. Zapf, C. H. Chinburg, and H. K. Wolf, "Inductive voltagedividers with calculable relative corrections," IEEE Trans. Instrum.Meas., vol. IM-12, pp. 80-85, Sept. 1963.

(2] J. J. Hill and T. A. Deacon, "Theory, design and measurement ofinductive voltage dividers," Proc. Inst. Elec. Eng., vol. 1I5, pp.727-735, May 1968. .

(3) -, "Voltage-ratio measurement with a precision of parts in 109andperformance of inductive voltage dividers," IEEE Trans. Instrum.Meas., vol. 1M-I?, pp. 269-278, Dec. 1968.

(4] w. C. Sze, "An injection method for self-calibration of inductivevoltage dividers," J. Res.. Nat. Bur. Stand., vol. nc, pp. 49-59,Jan.-Mar. 1968.

(5) K. Grohmann and T. L. Zapf, "An international comparison of in-ductive voltage divider calibration methods between 10 kHz and 100kHz," Metrologia, vol. 15, pp. 69-75, 1979.

(6) C. A. Hoer and W. L. Smith, "A I-MHz binary inductive voltagedivider with ratios of2n to lor 6n dB," IEEE Trans. Instrum. Meas.,vol. IM-17, pp. 278-284, Dec. 1968.

(7] N. M. Oldham, "A 50 ppm ac reference standard which spans 1 Hzto 50 kHz," IEEE Trans. Instrum. Meas., vol. IM-32, pp. 176-179,Mar. 1983.

(8) G. Ramm, R. Vollmert, and H. Bachmair, "Microprocessor-con-trolled binary inductive voltage divider," IEEE Trans. Instrum.Meas., vol. IM-34, pp. 335-337, June 1985.

[9] N. M. Oldham, O. Petersons, and B. C. Waltrip, "Audio-frequencycurrent comparator power bridge: Development and design consid-erations," IEEE Trans. Instrum. Meas., vol. 38, pp. 390-394, Apr.1989.

[10) R. w. Gammon, "Critical fluid light scattering," in The Nation'sFuture Material Needs. Proc. 19th Int. Tech. Conf. Soc. Advance-ment Mater. Process Eng., Oct. 1987.

[II] S. Avramov, "Initial NIST test results of Zeno EDM ratio trans-former linearity against binary inductive voltage divider," InterimRep., July 1991 (available from Institute for Physical Sciences andTechnology, University of Maryland, College Park).

[12) W. C. Sze, "Instruction manual for NBS reference inductive di-vider." NBS/CCG Project No. 70-40, Mar. 1973 (available from theElectricity Division, NIST).

!

"

Frequency: 100 Hz BlVD Frequency: I kHz BlVDErrors in ppm Errors in ppm

NominalRatio In-Phase Quadrature In-Phase Quadrature

0.1 0.01 0.09 0.05 0.720.2 0.01 0.34 0.09 2.060.3 0.02 0.79 0.15 3.700.4 0.00 0.28 0.19 5.250.5 0.02 0.30 0.03 2.580.6 0.03 0.29 0.02 3.870.7 0.01 0.76 0.04 5.880.8 0.04 0.46 0.23 2.720.9 0.02 0.29 0.18 4.90