left ventricular assist device (lvad) heartware hvad t . schlöglhofer
DESCRIPTION
Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer Center for Medical Physics and Biomedical Engineering, Department of Cardiac Surgery, Medical University of Vienna, Austria; Ludwig-Boltzmann-Cluster for Cardiovascular Research, Vienna, Austria. HeartWare HVAD. - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/1.jpg)
Schulung HVAD Oktober 2012
Left Ventricular Assist Device (LVAD)HeartWare HVAD
T. Schlöglhofer
Center for Medical Physics and Biomedical Engineering,
Department of Cardiac Surgery, Medical University of Vienna, Austria; Ludwig-Boltzmann-Cluster for Cardiovascular Research, Vienna, Austria
![Page 2: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/2.jpg)
Schulung HVAD Oktober 2012
HeartWare HVAD
![Page 3: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/3.jpg)
Schulung HVAD Oktober 2012
• Implantierte Komponenten:– HVAD Pumpe– Ausstrom Graft– Transcutane Driveline
• Weitere Komponenten:– Controller– System Monitor– Energiequellen
• Akkus• AC Adapter• 12V Adapter
– Zubehör
Systemkomponenten-HeartWare HVAD
![Page 4: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/4.jpg)
Schulung HVAD Oktober 2012
HeartWare HVAD Pumpe
• Zentrifugal Pumpe mit kontinuierlichem Fluss
• Hydrodynamische Lagerung des Impellers
![Page 5: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/5.jpg)
Schulung HVAD Oktober 2012
• Einstrom: aus der linken Herzkammer
• Pumpe an der Herzspitze
• Rückfluss: in die aufsteigende Aorta
• Speed Bereich: 1800 - 4000 rpm
• Flow Bereich: bis 10 L/min
Pumpe/Systemlage
![Page 6: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/6.jpg)
Schulung HVAD Oktober 2012
Controller (Steuereinheit)
![Page 7: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/7.jpg)
Schulung HVAD Oktober 2012
Stromversorgung• Akku (ca. 4-6h Betriebszeit) • AC oder DC Adapter für Netzbetrieb
![Page 8: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/8.jpg)
Schulung HVAD Oktober 2012
• Die Leuchten bei jeder Bat. bedeuten:– Grün bei Ready: Bat. voll– Bei Status:
• Rot: Bat zu kalt oder heiß• Blinkend rot: Bat. Defekt• Gelb: Bat wird geladen
Ladegerät
![Page 9: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/9.jpg)
Schulung HVAD Oktober 2012
Zubehör• Umhängetasche/Bauchgurt
• Duschtasche
![Page 10: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/10.jpg)
Schulung HVAD Oktober 2012
Monitor
• Anzeige aller Pumpenparameter
• Drehzahländerung
• Anzeige von Alarmen bzw. Energiestatus des Controllers
![Page 11: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/11.jpg)
Schulung HVAD Oktober 2012
• Flow beeinflußt von Pre- und Afterload
• Pulsatilität: erzeugt vom nativen Herzen
• Pulsatilität: Änderung des Flow zwischen Systole/Diastole
Waveform/Pulsatilität
![Page 12: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/12.jpg)
Schulung HVAD Oktober 2012
Waveform/Pulsatilität
![Page 13: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/13.jpg)
Schulung HVAD Oktober 2012
Abhängigkeit Drehzahl vs. Druck
![Page 14: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/14.jpg)
Schulung HVAD Oktober 2012
Übersicht Hinweis-Alarme
![Page 15: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/15.jpg)
Schulung HVAD Oktober 2012
Übersicht Notfall-Alarme
![Page 16: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/16.jpg)
Schulung HVAD Oktober 2012
• Low Flow• High Watts• VAD gestoppt - Pumpe diskonnektiert• VAD gestoppt – beide Stromquellen
diskonnektiert
Wichtigste Alarme
![Page 17: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/17.jpg)
Schulung HVAD Oktober 2012
• Hypovolämie• Rechtsherzdekompensation• Hypertonie• Orthostase Probleme• Arrhythmie
Low Flow Alarm – mögliche Ursachen
![Page 18: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/18.jpg)
Schulung HVAD Oktober 2012
• Pumpen Thrombus – bei Hämolyse (fHb) über 10mg/dl Kontakt mit dem Kunstherzteam aufnehmen!
• Pumpe defekt• Controller defekt• Kabelbruch
High Watts Alarm– mögliche Ursachen
![Page 19: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/19.jpg)
Schulung HVAD Oktober 2012
Waveforms
![Page 20: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/20.jpg)
Schulung HVAD Oktober 2012
Waveforms
![Page 21: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/21.jpg)
Schulung HVAD Oktober 2012
Waveforms
![Page 22: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/22.jpg)
Schulung HVAD Oktober 2012
Waveforms
![Page 23: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/23.jpg)
Schulung HVAD Oktober 2012
Waveforms
![Page 24: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/24.jpg)
Schulung HVAD Oktober 2012
MARCOUMAR: • Therapeutischer Bereich INR 2.0 - 3.0 (entspricht etwa einem
Thrombotest von 10 - 14%)• Achtung: bei Lupus-antikoagulans-positiven Patienten ist ein
höherer INR notwendig!• Wird bei Kontrolle der therapeutische Bereich unterschritten, ist
Lovenoxgabe meist notwendig - nach Rücksprache mit dem Kunstherzteam!
• Macoumar-Diät (Achtung bei Speisen mit Vitamin K!), die Patienten sind darauf hinzuweisen.
Thrombo-ASS : • Weiter nach Verordnung (typisch: 1x100mg) bzw. individuell
Gerinnungsmanagement
![Page 25: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/25.jpg)
Schulung HVAD Oktober 2012
VerbandswechselVorbereitungBeim Verbandwechsel der Kabelaustrittstelle bei Pumpenpatienten sind immer ein Mundschutz und Gummihandschuhe zu verwenden! Mundschutz für Schwester/Pfleger und Patient!
Schritt 1: Reinigen der Kabelaustrittstelle• Gegebenfalls Haare abrasieren, danach Kabelaustrittstelle mit Octenisept reinigen/desinfizieren (Keine alkoholischen Reinigungsmittel!)• Bei infizierten Kabelaustrittstellen Verwendung von Silbernitratlösung (ca. 10 Minuten mit befeuchtetem Tupfer einwirken lassen)
![Page 26: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/26.jpg)
Schulung HVAD Oktober 2012
Verbandswechsel
Schritt 2: Abdecken der Kabelaustrittstelle mit Verband• Die gereinigte Kabelaustrittstelle wird mit einem geschlitzten kleinen Tupfer bedeckt• Durchsichtige Katheterabdeckfolie (Smith&Nephew IV3000 1-Hand) wie abgebildet auf die Tupfer kleben
![Page 27: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/27.jpg)
Schulung HVAD Oktober 2012
Verbandswechsel
Schritt 3: Befestigen des Pumpenkabels• Das Pumpenkabel muss so nahe als moglich an der Austrittstelle mit der Schlauchbefestigung fixiert werden!
• Es muss sichergestellt sein, dass sich das Kabel an der Austrittstelle so wenig wie moglich bewegen kann, um eine mechanische Reizung zu verhindern!
![Page 28: Left Ventricular Assist Device (LVAD) HeartWare HVAD T . Schlöglhofer](https://reader033.vdocument.in/reader033/viewer/2022061322/56816578550346895dd80dfb/html5/thumbnails/28.jpg)
Schulung HVAD Oktober 2012
Kontaktdaten AKH Wien - Kunstherz
Verantwortliche Ärzte:Doz. Dr. Daniel Zimpfer 0664/243 72 22 oder 01/40400 6835Dr. Stephane Mahr 0676/377 96 19Dr. Thomas Haberl 0664/207 79 46Dr. Julia Riebandt 0650/523 83 77
Technische Betreuung:24h-Bereitschaftsnummer 01/40400 2200Prof. Heinrich Schima 0664/454 60 33 oder 01/40400 2001KT Zeno Hartner 0676/342 31 01 oder 01/40400 2001KT Johann Horvat 0664/504 56 26 oder 01/40400 2001BSc. Thomas Schlöglhofer 0650/711 91 29 oder 01/40400 2001