High End Anesthesia Units

TRANSCRIPT

Page 1: Leon Plus Engl 1.2
Page 2: Leon Plus Engl 1.2

Production Leon, Leon plus, etc.

Page 3: Leon Plus Engl 1.2

• (S) IMV (Synchronized) Intermittend Mandatory Ventilation

• (S) PCV (Synchronized) Pressure Controlled Ventilation

• PSV Pressure Supported Ventilation

• Man. Manual Ventilation

• Spont. Spontaneous Ventilation

• HLM – Mode (optional) Heart – lung machine

Ventilation Modes

Page 4: Leon Plus Engl 1.2

Description Patient module

Patient module

• Fresh gas decoupled• Tempered: preventing condensation warming respiratory gas • APL decoupled during controlled ventilation• Visual observation of the „Hanging Bellow“

Page 5: Leon Plus Engl 1.2

Patient module

• Inspiratory and expiratory flow sensors• Absorber can be exchanged

during operation• Can be fully sterilized• Compliance compensated

Patient module

Page 6: Leon Plus Engl 1.2

CO2 - Absorber

• New Absorber• Volume complete 2000 ml• Fill limit 1750 ml• Longer operating time• Exchange during operation• Easy in handling

Page 7: Leon Plus Engl 1.2

Features

• 15“ TFT color display• Operation via touchscreen and / or control

knob• Electronic fresh gas blender• One - hour rechargeable battery buffering• Two Interface connections RS 232• Ethernet interface• Gas monitor with automatic detection of

anesthetic gas and anesthetic gas mixture

Page 8: Leon Plus Engl 1.2

Features

• Receptacle for two anesthetic vaporizers• Undercarriage with 4 antistatic rollers, all with brakes• 4 Drawers• 1 cupboard reserve bottle connections for O² and N²O• Pull out writing surface, DIN A4• Rails on right and left to fit support arms and other accessories

Page 9: Leon Plus Engl 1.2

O² Safety Flow System

• Mechanic Valve (independent from Voltage)

• Settings 0 – 15 l/min.• Use for quick start

(emergency)• Use for device is turned

off• Use for fail of fresh gas

blender

Page 10: Leon Plus Engl 1.2

Gas Management

• Sidestream Monitor with real time measurement and real time graphs

• Paramagnetical O² measurement

• Different water trap for exactly measures in patient category „child“

Page 11: Leon Plus Engl 1.2

Configuration of the options panel

Solution for half open breathingsystem

KuhnJackson ReesMagillBain

Page 12: Leon Plus Engl 1.2

Configuration of the options panel

• Solution „external O² outlet“

• O² – Flowmeter adjustable between 0 – 15 l / min.

Page 13: Leon Plus Engl 1.2

Pressure gauges and control elements

O² flush

Turn switch for vacuum dosing

Vacuumdosing

Vacuum pressure

O² bottle pressure

N²O bottle pressure

Page 14: Leon Plus Engl 1.2

New Ventilation Mode

PSV Pressure Supported Ventilation

• Psv provides pressure support during insufficient spontaneous breathing.

• If the Leon plus is not triggered by the patient after the adjustable apnoe time, the Leon plus initiates inspiration on ist own.

• In addition, you can use a button to start a manual breath, not activated by the patient.

Page 15: Leon Plus Engl 1.2

New Ventilation Mode

HLM – Mode (Heart – Lung machine mode)

• Mode is working on a CPAP Level• Measurement of CPAP• Without any patient alarm• Stopwatch for time of bypass

Page 16: Leon Plus Engl 1.2

Real time graphs and complete Measurement

Page 17: Leon Plus Engl 1.2

Fresh gas control

• Air or N2O as carrier gas

• Set percent oxygen to total fresh gas flow

• Set total fresh gas flow• Setting range 200 ml / min. – 18 l / min.

• Ensure O2 concentration in O2 / N2O mixture of 25 % (Ratio)

• Ensure O2 concentration in „Minimal Flow“ (up from 200 ml/min.) is with 100%, adapted to the flow

Page 18: Leon Plus Engl 1.2

Fresh gas control

• Backup Emergency bottles (2 – 3 litres)Anlog pressure gauges in front

• Reservegas bottles (10 litres) with digital input (optional)Digital bottle pressures in the window „thresholds“

Page 19: Leon Plus Engl 1.2

Systemtest

New Systemtest• Traffic light system

(green, yellow, red)• Clear optical

information• Repetition of the

tests one by one• comment window• Self explanable

Page 20: Leon Plus Engl 1.2

Real time graphs

Values as real time graphs

• Airway pressure

• Flow

• Volume

• CO² (%, hPa or mmHg)

• O²

• N²O

• Halothane

• Enflurane

• Isoflurane

• Sevoflurane

• Desflurane

Page 21: Leon Plus Engl 1.2

Values displayed in numerical form

Real time measurement:

Minute volume, tidal volume, Peak / Plateau / Mean pressure, PEEP, Frequency

Calculated values: MAC, Compliance dyn., Compliance stat., C20/C, Resistance

Adapted values:CPAP (only in HLM – Mode)

Page 22: Leon Plus Engl 1.2

Loops

One Loop at a time can be displayed in full screen

Page 23: Leon Plus Engl 1.2

Loops

Loop window

Three loops can be displayed in the Loops window– Volume over pressure– Flow over pressure– Flow over volume

Page 24: Leon Plus Engl 1.2

• Alarm limits for monitoring

(adjustable)

• Function of „Autoset“

• Alarms with three different

priorities (high, medium, informal)

• Complete history of alarms in the

Log

• Complete history of events in the

Event Log (only in Standby)

• Adjustable loudness of alarm tone

• Gas supply pressure

Thresholds