Light element sensitivity:
EDS detectors are NOT created equal Keith Thompson Aug 2014
![Page 2: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/2.jpg)
2 Proprietary & Confidential
EDS detectors have come a long way over the past decade
Most people will tell you that all detectors are basically the same
123 eV
• Faster • Better resolution • Better sensitivity
![Page 3: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/3.jpg)
3 Proprietary & Confidential
In the “middle – high” energy range: detector performance is mostly the same between detectors.
123 eV 129 eV
183 eV
133 eV 138 eV
![Page 4: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/4.jpg)
4 Proprietary & Confidential
EDS map at 145 eV
In the “middle – high” energy range: detector performance is mostly the same between detectors.
![Page 5: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/5.jpg)
5 Proprietary & Confidential
Standard detector
Extreme detector
In the “middle – high” energy range: detector performance is mostly the same between detectors.
![Page 6: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/6.jpg)
6 Proprietary & Confidential
In the light element world: some EDS detectors just don’t hold up.
Standard detector
Extreme detector
![Page 7: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/7.jpg)
7 Proprietary & Confidential
Why are some detectors better than others at low energies?
• Impact of the SDD Module
• Impact of the architecture
• Impact of the window
![Page 8: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/8.jpg)
8 Proprietary & Confidential
Three main factors in SDD light element performance
Traditional – circular / symmetric vs.
Modern – tear drop, small FET
Low energy x-ray
SDD
mod
ule
Light elementwindow
Low energy x-ray
SDD
mod
ule
Light elementwindow
Traditional: N2 backed window vs.
Modern: evacuated window
Traditional: wire-bonded FET vs.
Modern: integrated, “on-chip” FET
![Page 9: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/9.jpg)
9 Proprietary & Confidential
Light element sensitivity: “Sensitive to B”
Pure B metal
![Page 10: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/10.jpg)
10 Proprietary & Confidential
Light element sensitivity: “Sensitive to Be”
Extreme EDS
![Page 11: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/11.jpg)
11 Proprietary & Confidential
Impact of the module
N2 back-filled window Traditional round geometry
Evacuated window Traditional round geometry
Evacuated window Tear-drop geometry
Normalized intensity scales Stronger B peak
Smaller zero width
Better signal to noise
![Page 12: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/12.jpg)
12 Proprietary & Confidential
Some Examples - Trace B in steel
![Page 13: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/13.jpg)
13 Proprietary & Confidential
Mapping 2% B in Fe-Cr: Traditional detector
0
1000
2000
3000
4000
5000
6000
0200400600800
100012001400160018002000
80 180 280 380 480
EDS
coun
ts
WDS
cou
nts
Energy eV)
B - WDSB - EDS
Trace B (2% B in Fe-Cr) is harder
B
C
0
2000
4000
6000
8000
10000
0
20000
40000
60000
80000
100000
120000
80 180 280 380 480
EDS
coun
ts
WDS
cou
nts
Energy eV)
B - WDSB - EDS
B metal is easy for EDS/WDS
B
![Page 14: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/14.jpg)
14 Proprietary & Confidential
B raw counts
B Net counts (processed)
B WDS
Mapping 2% B in Fe-Cr
1. With an evacuated tear-drop detector trace B mapping is possible.
2. “Processing” the maps to remove background helps. 3. WDS still provides the best answer.
![Page 15: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/15.jpg)
15 Proprietary & Confidential
Some Examples - Sn: Ni-Cu intermetallic
![Page 16: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/16.jpg)
16 Proprietary & Confidential
20 kV 5 kV
EDS Ni-Cu intermetallic
The goal is to examine potential interdiffusion through a thin barrier layer.
Low kV analysis to avoid an interaction volume that may pollute the data
Cu only
1 µm
![Page 17: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/17.jpg)
17 Proprietary & Confidential
Spectrum analysis: EDS Ni-Cu intermetallic
Cu only
1 µm
Serious overlap
![Page 18: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/18.jpg)
18 Proprietary & Confidential
Standard module Advanced module
![Page 19: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/19.jpg)
19 Proprietary & Confidential
Spectrum analysis: WDS vs. EDS Ni-Cu intermetallic
1
2
3
Cu only
Ni only
Cu & Ni
1 µm
![Page 20: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/20.jpg)
20 Proprietary & Confidential
EDS vs. WDS maps – Ni, Cu, Sn
EDS element maps are confounded
WDS element maps provide complete confidence
Copper Ni Sn
![Page 21: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/21.jpg)
21 Proprietary & Confidential
EDS vs. WDS maps – Ni, Cu, Sn – Processed!
Copper Ni Sn
WDS element maps provide complete confidence
The PROCESSED EDS element maps provide a better look
![Page 22: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/22.jpg)
22 Proprietary & Confidential
EDS vs. WDS maps – Ni, Cu, Sn – Processed!
Raw EDS element maps
PROCESSED – Net Counts – EDS element maps
![Page 23: Light element sensitivity: EDS detectors are NOT created equaltools.thermofisher.com/.../brochures/WS52655-Light...• The light element performance is very different based on •](https://reader034.vdocument.in/reader034/viewer/2022050112/5f497c474b82c212277c7d1c/html5/thumbnails/23.jpg)
23 Proprietary & Confidential
Summary • EDS detectors are not all the same.
• The light element performance is very different based on
• SDD module type • Detector window • Overall detector architecture
• Know your application
• Mid – high energy applications: Most EDS detectors are fine • Low energy applications: Need the best possible EDS detector.
• Often post-processing algorithms can extract the correct answer even when the raw data is confounding.