The Linatron ® -M is a modular system. The control console, modulator and RF unit are common to all model configurations. Only the X-ray head changes to match the application. The Linatron-M is designed to fit mobile, gantry, and fixed installations. 1.0 Standard Equipment and Services 1.1 Control Console The standard control console is a touch screen display system. An optional desktop PC control console is available (see section 4.6). 1.2 X-ray Head Low Leakage (0.1%) 1.3 Modulator/Power Distribution Cabinet External signal interface 1.4 Temperature Control Unit (TCU) The TCU is used to keep the system components at a nominal 30˚C (86˚F). It is available in high voltage and low voltage configurations for environments ranging from -40/+55˚C (-40/131˚F), condensing. 1.5 Standard Spare Parts Kit The standard spare parts kit includes over 40 items such as PC boards and individual components. 1.6 Interconnecting Cables (X-ray Head to Modulator. Modulator to Console) and Hoses (TCU to X-ray Head) Included. Lengths up to 100 meters. 1.7 Manuals and Data Books Two sets of operator and maintenance manuals and data books are included in English. 1.8 Installation Supervision and Start-up Assistance 1.9 Varian’s Standard Warranty 2.0 Performance 2.1 X-ray Beam Quality The X-ray beam quality is specified using Half Value Layer (HVL) steel. This corresponds to the nominal X-ray energy shown in Table 1. These HVL numbers are derived from a compilation of broad beam data measurements. 2.2 X-ray Beam Dose Rate* (10 cm x 10 cm field) The maximum continuous dose rate at 1 meter is listed in Table 1. (Without flattening filter.) * Dose rate is reduced with flattening filter. 2.3 X-ray Field Size A 30˚ cone or 22.5˚ square defines the field. Also see section 4.1. 2.4 X-ray Beam Focal Spot Size The focal spot size does not exceed 2.0 mm in diameter. 2.5 X-ray Beam Symmetry The beam asymmetry does not exceed 5% at ±7.5˚ off the central axis for all energies. 2.6 Radiographic Quality The Linatron-M system will demonstrate at least ASTM E 94 1-2T, or equivalent, sensitivity over the ranges given in Table 2 using film detection. Linatron-M6 & M6A Modular high-energy X-ray source Touchscreen Control Console X-ray Head and RF Unit Table 1 Model Nominal HVL Flatness Max. Dose Rate Energy (MeV) (in) (% @ ±7.5˚) (Gy/min) M6 3.5 0.96 71.0 2.5 5.0 1.06 65.5 5.0 6.0 1.10 62.0 8.0

Upload: anonymous-ipqcbb

Post on 01-Feb-2016

220 views

Category:

Documents


3 download

DESCRIPTION

Linatron M6X ray solutionacurate x ray test

TRANSCRIPT

Page 1: LinatronM6

The Linatron®-M™ is a modular system. The controlconsole, modulator and RF unit are common to allmodel configurations. Only the X-ray head changes tomatch the application. The Linatron-M is designed tofit mobile, gantry, and fixed installations.

1.0 Standard Equipment and Services1.1 Control Console

The standard control console is a touch screendisplay system. An optional desktop PC controlconsole is available (see section 4.6).

1.2 X-ray Head Low Leakage (0.1%)

1.3 Modulator/Power Distribution CabinetExternal signal interface

1.4 Temperature Control Unit (TCU)The TCU is used to keep the system components ata nominal 30˚C (86˚F). It is available in high voltageand low voltage configurations for environmentsranging from -40/+55˚C (-40/131˚F), condensing.

1.5 Standard Spare Parts KitThe standard spare parts kit includes over 40 itemssuch as PC boards and individual components.

1.6 Interconnecting Cables (X-ray Head to Modulator.Modulator to Console) and Hoses (TCU to X-rayHead) Included. Lengths up to 100 meters.

1.7 Manuals and Data BooksTwo sets of operator and maintenance manuals anddata books are included in English.

1.8 Installation Supervision and Start-up Assistance

1.9 Varian’s Standard Warranty

2.0 Performance2.1 X-ray Beam Quality

The X-ray beamquality is specifiedusing Half ValueLayer (HVL) steel.This corresponds tothe nominal X-rayenergy shown inTable 1. These HVL numbers are derived from a compilation of broad beam data measurements.

2.2 X-ray Beam Dose Rate* (10 cm x 10 cm field)The maximum continuous dose rate at 1 meter islisted in Table 1. (Without flattening filter.)* Dose rate is reduced with flattening filter.

2.3 X-ray Field SizeA 30˚ cone or 22.5˚ square defines the field. Alsosee section 4.1.

2.4 X-ray Beam Focal Spot SizeThe focal spot size does not exceed 2.0 mm indiameter.

2.5 X-ray Beam SymmetryThe beam asymmetry does not exceed 5% at ±7.5˚off the central axis for all energies.

2.6 Radiographic QualityThe Linatron-M system will demonstrate at leastASTM E 94 1-2T, or equivalent, sensitivity over theranges given in Table 2 using film detection.

Linatron-M6 & M6A Modular high-energy X-ray source

Touchscreen Control Console

X-ray Head and RF Unit

Table 1Model Nominal HVL Flatness Max. Dose Rate

Energy (MeV) (in) (% @ ±7.5˚) (Gy/min)M6 3.5 0.96 ≥71.0 2.5

5.0 1.06 ≥65.5 5.06.0 1.10 ≥62.0 8.0

Page 2: LinatronM6

2.7 Standard Leakage RadiationThe leakage radiation is specified along thehorizontal axis at 1 meter from the beamcenterline at angles 60˚ and greater, outside theprimary beam. The values in Table 3 are afraction of the primary beam central axis doserate measured with a 10 cm x 10 cm collimator.Leakage is taken with the primary beamcompletely blocked. See section 4.2 for lowerleakage options.

3.0 Customer Facility Requirements3.1 Electrical Requirements

3.1.1 The Linatron-M operates from a single 15-kVA 50/60 Hz power source. Two voltageranges are available.3.1.1.1 Low Voltage Option

208 VAC, 3 phase, 3 or 4 wire plusground, 60 Amp minimum surgeper leg. ±10% voltage regulation isrequired.

3.1.1.2 High Voltage Option400 VAC, 3 phase, 4 wire plusground, 40 Amp minimum surgeper leg. ±10% voltage regulation isrequired.

3.1.2 The TCU is connected to a separate 13-kVApower source. Models are available that canoperate on a line voltage of 220 VAC and400 VAC, at 50 Hz; or 220 VAC and 480VAC, at 60 Hz. A separate 10-kVA powersource may be required for the in-line heaterpackage.

3.2 Operating Requirement

3.2.1 Indoor RequirementThe operatingenvironment forcontrol console andmodulator must bebetween 4˚C (39˚F)and 35˚C (95˚F),with 90% maximumrelative humidity(non-condensing).

3.2.2 Outdoor RequirementThe available temperature range for X-rayhead/RF unit is dependent on the TCU and thermal insulation blanket. The rangecan be absorbed as -40/+55˚C (-40/131˚F),condensing.

3.2.3 VentilationThe approximate heat given to room airfrom each component with system operatingat full power is given below:X-ray Head/RF Unit: 1.0 kWModulator Cabinet: 2.0 kWTemperature Control Unit: 6.0-12.0 kWTouchscreen Control Console: Negligible

4.0 Optional Equipment4.1 Custom Beam Collimation

Nonstandard field sizes are available per customer’srequirements. A motorized collimator is alsoavailable to quickly change the beam collimation.

4.2 Lower Leakage Options are listed in Table 4.

4.3 Voltage RegulatorRecommended for installations where line powershort-term fluctuations are greater than ±5%. Astep-up or step-down transformer can also beordered to adapt a non-standard voltage source foruse with the Linatron or TCU. The regulator isCE and UL approved.

Linatron-M6 & M6A Modular high-energy X-ray source

Table 3Model Leakage

(fraction)M6 1x10-3

Table 4Model Leakage RF Unit/Head Wt.

(fraction) (lbs)Super Low Ultra Low Super Low Ultra Low

M6 2 x 10-5 2.5 x 10-6 2,100 5,100

Modulator

Table 2Model Nominal Range

Energy (MeV) (mm)M6 3.5 38-203

5.0 45-2306.0 51-254

Page 3: LinatronM6

4.4 Beam FlattenerThis option provides a more uniform beam intensityover the exposed region at 3.5, 5, and 6 MeV. Use ofa flattening filter will reduce dose. See table below.Not available with ULLP leakage option.

4.5 Dual EnergyThe dual energy specifications are given in Table 6.Select two operating energies:

4.6 Desktop PC Control ConsoleThe desktop PC control console provides the samesystem control as the touch screen console but has alarger viewing screen plus data storage capability.Heat given to room air is 0.5 kW.

4.7 Laser Alignment SystemAn internally mounted single spot laser is availableto align the X-ray beam to an object beingradiographed. Not available with ULLP leakageoption.

4.8 Variable External CollimatorThe dependent jaw variable external collimatormounts to thefront of the X-ray head. Thefield size variesbetween 1˚and 24˚. Arotatingversion isavailable thatrotates over arange from -50˚ to +50˚.

4.9 Remote Customer InterfaceA 37-pin Amphenol socket is provided on themodulator for interface to customers equipment.Signals include:

For a complete description of these signals requestdocument # 100015302.

CE Marking

All Linatron-M models are designed and manufactured inaccordance with the Electromagnetic CompatibilityDirective 89/336/EEC and Low Voltage Directive73/23/EEC.

ETL Marking

All Linatron-M models conform to UL STD 61010A-1and are certified to CSA 1010.1.

Quality Standard

Varian Security & Inspection Products, Las Vegas Facility,Quality Management Systems is registered to ISO9001:2008.

External Collimator with Rotation

Table 6Model Nominal HVL Flatness Max. Dose

Energy (in) (% @ ±7.5˚) Rate(MeV) (Gy/min)

M6A 3.5 0.96 ≥71.0 2.505.0 1.06 ≥65.5 5.006.0 1.10 ≥62.0 8.00

Table 5Energy Flatness Dose rate Coverage steel (MV) (Gy/min-m) thickness range

t (mm)3.5 better than 80% 2 or more 50-2505 better than 80% 3 or more 50-2506 better than 80% 3.5 or more 50-250

Remarks: Effective field size of flatness specified is 400 mmin diameter at SFD=1.5 m. Flatness is measured by densityof film exposed.

Flatness Specification

Flatness Specification

• External Trigger• Emergency Off• Remote Interlock• Warning Lights

• Warning Alarm• X-ray on Request• Warm Up and Power On Status• Fault Information and Reset

Page 4: LinatronM6

5.0 Physical Description

4 2 . 0 2 4 . 0

4 0 . 0

4 2 . 03 2 . 5

2 6 . 0

7 . 0

1 9 . 0

1 2 . 3

3 6 . 4

2 4 . 04 2 . 0

3 0 . 0

3 6 . 0

5 5 . 0

M6 X-ray Head(1,700 lbs)

Super Low Leakage(2,100 lbs)

Control Console(9 lbs)

Modulator(735 lbs)

M6 X-ray HeadUltra-low Leakage Low Profile(5,000 lbs)

* Dimensions are in inches.

Linatron-M6 & M6A Modular high-energy X-ray source