BA 209C/07/en/07.02 No. 51517599 Software version 2.30 or later Supplement to BA 193C/07/en BA 194C/07/en BA 199C/07/en BA 200C/07/en BA 214C/07/en PROFIBUS-PA/-DP Field Communication for Liquisys M CXM 223/253 Operating Instructions Endress Hauser The Power of Know How Quality made by Endress+Hauser ISO 9001 ENDRESS+HAUSER LIQUISYS M ENDRESS+HAUSER LIQUISYS M

Upload: api-3727411

Post on 18-Nov-2014

437 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Liquisys (CCM253)_Chlorine_analyser

BA 209C/07/en/07.02No. 51517599Software version 2.30 or later

Supplement toBA 193C/07/enBA 194C/07/enBA 199C/07/enBA 200C/07/enBA 214C/07/en

PROFIBUS-PA/-DPField Communication forLiquisys M CXM 223/253Operating Instructions

Endress HauserThe Power of Know How

Quality made byEndress+Hauser

ISO 9001

ENDRESS+HAUSERLIQUISYS M

ENDRESS+HAUSERLIQUISYS M

Page 2: Liquisys (CCM253)_Chlorine_analyser

You need information on the instrument?Please read the following chapters:

SafetyGeneralinformation

You wish to install and start up the instrument.The required steps are described in these chapters:

Installation

You wish to operate or reconfigure the instrument:

Start-up

Communication

Technical data

Index

Accessories % Ω°C% Ω°C

Page 3: Liquisys (CCM253)_Chlorine_analyser

XM53PE0.CHP

TABLE OF CONTENTS1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43.1 System equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43.2 Electrical connection PA instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53.3 Electrical connection DP instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

4 Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94.1 Setting of device address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94.2 Device data and type files. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

5 Communication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125.1 Cyclical data exchange (Data_Exchange) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125.2 Acyclical data exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145.3 Operation via Commuwin II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

6 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

7 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

8 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Liquisys M PROFIBUS-PA/-DP

Endress+Hauser 1

Page 4: Liquisys (CCM253)_Chlorine_analyser

1 General information

These operating instructions were designedspecifically for use with transmitters of thefamily Liquisys M CXM 223/253. They containspecific information for instruments with thePROFIBUS-PA interface (Process Field Bus –Process Automation) or the PROFIBUS-DPinterface (Process Fieldbus DecentralizedPeriphery).

Please refer to the corresponding standardoperating instructions for information oninstallation and general transmitter operation:BA 193C/07/en (CLM 223/253)BA 194C/07/en (CPM 223/253)BA 199C/07/en (COM 223/253)BA 200C/07/en (CUM 223/253)BA 214C/07/en (CCM 223/253).

1.1 Symbols used

Warning:

This symbol alerts to hazards whichmay cause serious injuries as well asdamage to equipment if ignored.

Note:

This symbol indicates importantitems of information.Ignoring this information may resultin malfunction.

General information Liquisys M PROFIBUS-PA/-DP

2 Endress+Hauser

Page 5: Liquisys (CCM253)_Chlorine_analyser

2 Safety

2.1 Intended use

Operation via PROFIBUS-PA interface

PROFIBUS is an open field bus standardaccording to DIN 19 245 / EN 50 170. It hasbeen created specifically for process controlpurposes and permits connection of severalmeasuring instruments to a bus line.

The transmission method according toIEC 1158-2 guarantees safe signaltransmission. The PROFIBUS-PA interfacepermits operation of the instrument from thePC via the Commuwin II operating program.

2.2 General safety instructions

Warning:

• The notes and warnings in theseoperating instructions must bestrictly adhered to.

• The notes and warnings containedin the standard operatinginstructions for the instrument inquestion (193C, 194C, 199C,200C, 214C) must be strictlyadhered to.

XM53PE2.CHP

Liquisys M PROFIBUS-PA/-DP Safety

Endress+Hauser 3

Page 6: Liquisys (CCM253)_Chlorine_analyser

3 Installation

3.1 System equipment

A complete system equipment comprises thefollowing components:

• Transmitter Liquisys M CXM 223/253-xx3• Segment coupler (only with PA)• Programmable logic controller (PLC) or PC

with PROFIBUS interface and Commuwin IIoperating program

• PROFIBUS-PA terminating resistor• Cabling incl. bus distributor.

The maximum number of transmitters in a bussegment is determined by their powerconsumption, the bus coupler output and therequired bus length.

Note:

For detailed information on functionand connection of a PROFIBUSsystem, please refer to TechnicalInformation TI 260F/00/en, which isavailable via the Endress+Hausersales agencies (see back cover ofthese operating instructions).

PROFIBUS-PA

PROFIBUS-DP

ENDRESS+HAUSERLIQUISYS SENDRESS+HAUSERLIQUISYS S

ENDRESS+HAUSERLIQUISYS

CAL REL

REL1

REL1

ALARMALARMREL2

REL2E

ENDRESS+HAUSERLIQUISYS

CAL REL

REL1

REL1

ALARMREL2

REL2E

Segmentcoupler

PC with Commuwin IIoperating program

PLC

XM253-02.CDRFig. 3.1Measuring systems withPROFIBUS interface

e.g. Liquisys M CPM 223with CPA 140

e.g. Liquisys M COM 253with COS 4

Installation Liquisys M PROFIBUS-PA/-DP

4 Endress+Hauser

Page 7: Liquisys (CCM253)_Chlorine_analyser

3.2 Electrical connection PA instrument

Bus cable connection to field instrument

The bus cable connection to the fieldinstrument can be performed with or withoutM12 plug. Connect the bus cable as follows:

• Unscrew the four recessed-head screwsand remove the instrument cover.

• Thread the cable through the openedcable entry into the connectioncompartment.

• Connect the wires of the bus cable to theterminal block according to the figurebelow. Polarity reversal of the connectionsPA+ and PA– does not affect operation.

• Tighten the cable entry.

Connection of several field instruments

565554 51 52 53 5957 58 47 48 49 41 42 43

3231 33 34 9385 86 94 81 82

565554 51 52 53 5957 58 47 48 49 41 42 43

3231 33 34 9385 86 94 81 82

XM253AN.CDR

Pg couplingM12 plug

Connection terminal block

bk bn blPA+ PA–

Fig. 3.2

Bus cable connectionto field instrumentLiquisys M-PACPM 253 / CLM 253inductive / COM 253 /CUM 253 / CCM 253

left:Connection withM12 plug

right:Connection withoutM12 plug

565554 51 52 53 5957 58 47 48 49 41 42 43

3231 33 34 9385 86 94 81 82

565554 51 52 53 5957 58 47 48 49 41 42 43

3231 33 34 9385 86 94 81 82

Pg couplingM12 plug

PA+ PA–

bn bl bk Connection terminal block

PA+ PA–

XM253AN.CDR

screen

Fig. 3.3

Bus cable connectionto field instrumentLiquisys M-PACLM 253 conductive

left:Connection withM12 plug

right:Connection withoutM12 plug

Supplyinstrument /segmentcoupler

Connection box Connection box

Condensatorsmax. 10 nF/250 V AC

Field instrument 1 Field instrument 2

XM253A5.CDR Fig. 3.4Connection of severalfield instruments

Screen PA+ PA–

XM53PE3.CHP

Liquisys M PROFIBUS-PA/-DP Installation

Endress+Hauser 5

Page 8: Liquisys (CCM253)_Chlorine_analyser

Bus cable connection to panel-mounted instrument

The bus cable connection to the panel-mounted instrument is performed directly at the terminalblock on the rear of the instrument.

Connection of several panel-mounted instruments

For the connection of several panel-mounted instruments to the bus cable, the connection linesmust be bridged according to the figure below.

82819493868594 78959899

XM253A2.CDR

PA+ PA– Screen

Connection terminal block

Fig. 3.5

Bus cable connection topanel-mounted instrumentLiquisys M-PACPM 223 / CLM 223inductive / COM 223 /CUM 223 / CCM 223

82819493868594 78959899Connection terminal block

PA+ PA– Screen

XM253A4.CDR

bn bl

Fig. 3.6

Bus cable connection topanel-mounted instrumentLiquisys M-PACLM 223 conductive

XM253-03.CDR

Panel-mountedinstrument 1

Panel-mountedinstrument 2

Connect and ground shields

Fig. 3.7

Connection of severalpanel-mountedinstruments

bn bl

Installation Liquisys M PROFIBUS-PA/-DP

6 Endress+Hauser

Page 9: Liquisys (CCM253)_Chlorine_analyser

3.3 Electrical connection DP instrument

Bus cable connection to field instrument

Bus termination

Install a bus termination for each instrumentconnected.

565554 51 52 53 5957 58 47 48 49 41 42 43

3231 33 34 9385 86 94 81 82

Connection terminal block

S B A

XM253A6.CDR Fig. 3.8

Bus cable connectionto field instrumentLiquisys M-DPCPM 253 / CLM 253inductive / COM 253 /CUM 253 / CCM 253

565554 51 52 53 5957 58 47 48 49 41 42 43

3231 33 34 9385 86 94 81 82

Connection terminal block

B A S

XM253A7.CDR Fig. 3.10

Bus cable connectionto field instrumentLiquisys M-DPCLM 253 conductive

31(99)

32(98)

33(67)

390 Ω 220 Ω 390 Ω

34(68)

XM253A9.CDR Fig. 3.9 Bus terminating resistor

XM53PE3.CHP

Liquisys M PROFIBUS-PA/-DP Installation

Endress+Hauser 7

Page 10: Liquisys (CCM253)_Chlorine_analyser

Bus cable connection panel-mounted instrument

The bus cable connection to the panel-mounted instrument is performed via a plug-in blockintegrated in the cover frame.

82819493868568 69679899Connection terminal block

gn ye br gr wt

Plug-in block

XM253A8.CDRFig. 3.11

Bus cable connection topanel-mounted instrumentLiquisys M-DPCPM 253 / CLM 253ind. / cond. / COM 253 /CUM 253 / CCM 253

Installation Liquisys M PROFIBUS-PA/-DP

8 Endress+Hauser

Page 11: Liquisys (CCM253)_Chlorine_analyser

4 Start-up

4.1 Setting of device address

The addresses of all PROFIBUS-PA devicesmust be set. An incorrect address setting willprevent detection of the measuring instrumentby the control system.

All instruments are set to address 126 at thefactory. This address can be used to test thefunction of the instrument and for connectionto a PROFIBUS-PA network. Subsequently,this address must be changed to connnectadditional devices. The device address canbe set via

• field operation• PROFIBUS communication• DIL switch at the instrument.

Note:

• Valid device address range:0 ... 126.

• Each address may only beassigned once in a PROFIBUS-PAnetwork.

• The double arrow in the displayindicates communication viaPROFIBUS.

Setting the device address via field operation (software setting)

The device address is set via keys inthe menu field Ι1 of the function groupINTERFACE.

Note:

The address can only be set viasoftware if DIL switch 8 is set tosoftware position. DIL switch 8 is setto software position at the factory.The opposite figure shows thedefault setting of the switch.

Setting the device address via PROFIBUS communication

The device address is set via the Set Slave Adr service.

Setting the device address via DIL switch (hardware setting)

The DIL switch is situated at the electronics module above the display. To set the deviceaddress, proceed as follows:

• Field instrument:– Unscrew the four recessed-head

screws and remove the instrumentcover. The electronics module withDIL switch is situated in the instrumentcover.

• Panel-mounted instrument:– The DIL switch is accessible through

a break in the instrument above thedisplay.

• Set the device address (0 ... 126) atswitches 1 to 7, e.g. 18 = 2 + 16.

• For setting the address via DIL switch→ set switch 8 to OFF.

• For field instrument, close the instrumentcover again.

126 I1

Address

1 2 3 4 5 6 7

ON

OFF

8

ON = software address

XM253-06.CDR

1

1

2

2

3

4

4

8

5

16

6

32

7

64

ON

OFF

8

OFF = hardware address

XM253-04.CDR

XM53PE4.CHP

Liquisys M PROFIBUS-PA/-DP Start-up

Endress+Hauser 9

Page 12: Liquisys (CCM253)_Chlorine_analyser

4.2 Device data and type files

The device data file (GSD) is required toconfigure a PROFIBUS-DP network. The GSD(a simple text file) describes e.g., which datatransfer rate is supported by the device orwhich digital information in which format thePLC receives from the device.

Each device is assigned an ID number by thePROFIBUS user organization (PNO). Thename of the device data file (GSD) is derivedfrom this ID number.

Note:

• For Endress+Hauser devices,the ID number always starts with“15XX”.

The GSD files for all Endress+Hauser devicescan be obtained from the following sources:

– via the Internet:E+H: http://www.endress.com(Product Avenue → DownloadStreet → Field CommunicationStreet)PNO: http://www.profibus.com(GSD library)

– on floppy disk from E+H:order no. 943157-0000

Contents of download file from the Internetand of floppy disk:

• All Endress+Hauser GSD files• All Endress+Hauser bitmap files• Useful information on the devices

Name of device ID no. GSD Type file Bitmaps

Liquisys MCLM 223/253-xx3 1515 hex EH_ _1515.gsd EH_1515x.200

EH1515_d.bmpEH1515_n.bmpEH1515_s.bmp

Liquisys MCPM 223/253-xx3 1516 hex EH_ _1516.gsd EH_1516x.200

EH1516_d.bmpEH1516_n.bmpEH1516_s.bmp

Liquisys MCUM 223/253-xx3 1517 hex EH_ _1517.gsd EH_1517x.200

EH1517_d.bmpEH1517_n.bmpEH1517_s.bmp

Liquisys MCOM 223/253-xx3 1518 hex EH_ _1518.gsd EH_1518x.200

EH1518_d.bmpEH1518_n.bmpEH1518_s.bmp

Liquisys MCCM 223/253-xx3 1519 hex EH_ _1519.gsd EH_1519x.200

EH1519_d.bmpEH1519_n.bmpEH1519_s.bmp

Liquisys MCLM 223/253-xx4 1521 hex EH_ _1521.gsd EH_1521x.200

EH1521_d.bmpEH1521_n.bmpEH1521_s.bmp

Liquisys MCPM 223/253-xx4 1520 hex EH_ _1520.gsd EH_1520x.200

EH1520_d.bmpEH1520_n.bmpEH1520_s.bmp

Liquisys MCUM 223/253-xx4 151f hex EH_ _151f.gsd EH_151fx.200

EH151f_d.bmpEH151f_n.bmpEH151f_s.bmp

Liquisys MCOM 223/253-xx4 151e hex EH_ _151e.gsd EH_151ex.200

EH151e_d.bmpEH151e_n.bmpEH151e_s.bmp

Liquisys MCCM 223/253-xx4 151d hex EH_ _151d.gsd EH_151dx.200

EH151d_d.bmpEH151d_n.bmpEH151d_s.bmp

Start-up Liquisys M PROFIBUS-PA/-DP

10 Endress+Hauser

Page 13: Liquisys (CCM253)_Chlorine_analyser

Working with the GSD / type files

The GSD files must be copied to a specificsubdirectory of the PROFIBUS-DPconfiguration software of your PLC.

Note:

• The subdirectory to be used forthe Siemens STEP7 configurationsoftware of the SiemensS7-300/400 PLC is\siemens\step7\s7data\gsd.

• If you are using a differentconfiguration software, then contactthe manufacturer of your PLC forthe correct subdirectory to use.

• If you have a Siemens S5 PLC andif the PROFIBUS-PA network isconfigured with the COM ET200configuration software, then youwill need type files (x.200 files).

• The GSD directory contains asubdirectory which holds the GSDfiles with the standard identifier(0x94). For example, these GSDfiles are intended for use with anAllen-Bradley PLC5.

XM53PE4.CHP

Liquisys M PROFIBUS-PA/-DP Start-up

Endress+Hauser 11

Page 14: Liquisys (CCM253)_Chlorine_analyser

5 Communication

The Liquisys M CXM 223/253 transmitter supplies the measured values using the cyclicalservice according to the PROFIBUS-DP protocol. Other parameters are made available usingthe acyclical service.

5.1 Cyclical data exchange (Data_Exchange)

The Data_Exchange service enables a PLC to read input data from the Liquisys M CXM223/253 transmitter. The cyclical data message for the maximum configuration of the LiquisysM CXM 223/253 is as follows:

Maximum configuration Liquisys M CXM 223/25

IEEE 754 floating point number

The Liquisys M CXM 223/253 transmitter uses a subset of the status codes defined by thePROFIBUS user organization (PNO). (See also: PNO specification “PROFIBUS-PA Profile forProcess Control Devices; General Requirements V2.0”.)

Input dataindex Data Access Data format /

remarksUnit

(factory setting)

0, 1, 2, 3 Main parameter read 32-bit floating pointnumber (IEEE 754)

CPM 223/253: pH; mV; %CLM 223/253: mS/cm; %; MΩCUM 223/253: FNU; ppm; g/l;%COM 223/253: mg/l; % SATCCM 223/253: mg/l

4 Main parameterstatus read See status codes –

5, 6, 7, 8 Temperature read 32-bit floating pointnumber (IEEE 754) °C

9 Temperaturestatus read See status codes –

10,11,12,13 pH / ORP read 32-bit floating pointnumber (IEEE 754)

pH or mV(dep. on operating mode)

14 pH / ORP status read See status codes –Tab. 5.1 Maximum configuration

D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0

S Exponent (E) Fraction (F)

27 26 25 24 23 22 21 20 2–1 2–2 2–3 2–4 2–5 2–6 2–7

Fraction (F)

2–8 2–9 2–10 2–11 2–12 2–13 2–14 2–15 2–16 2–17 2–18 2–19 2–20 2–21 2–22 2–23Tab. 5.2

IEEE 754 floating pointnumber

Communication Liquisys M PROFIBUS-PA/-DP

12 Endress+Hauser

Page 15: Liquisys (CCM253)_Chlorine_analyser

Status codes for main parameter (pH, conductivity, turbidity, oxygen) and temperature

If you do not wish to use all cyclical data provided by the Liquisys M CXM 223/253, then youcan eliminate cyclical data blocks from the cyclical message with the aid of the deviceconfiguration data string (Chk_Cfg) using the PLC software. You should only activate those datablocks which are actually processed by the system. This will improve the data throughput of aPROFIBUS network.

On the Liquisys M CXM 223/253 transmitter, two data blocks can be activated for cyclical dataexchange. The following table shows the configuration data needed for this (h means that thenumber is a hexadecimal number). The shortened identifier 94h can also be used. Theextended format is to be preferred, however, because it includes information for interpretationof the data.

Liquisys M CXM 223/253 configuration data

The following configurations are useful for the data blocks available with the Liquisys M CXM223/253:

Data blocks that are not activated must be identified by a zero as a placeholder in theconfiguration data string if other configuration data follow. Zeros at the end of the configurationdata may be omitted.

Status code Meaning Device status

hex. (dec.)

08h (8)Not connected(no meas. valueavailable)

BAD

0Ch (12) Device failure BAD

80h (128) O.K. GOOD Tab. 5.5

Status codes formain parameterand temperature

Cyclical data block Sequencein Chk_Cfg

Extended format Short formatData block active active Data block inactive

Main parameter 1 42h, 84h, 08h, 05h 94h 00h

Temperature 2 42h, 84h, 08h, 05h 94h 00h

pH / ORP withCCM 223/253 3 42h, 84h, 08h, 05h 94h 00h

Tab. 5.3 Configuration data

Activated data blocks Configuration data string(Chk_Cfg) Length (Chk_cfg)

Main parameter with status+ temperature with status

42h, 84h, 08h, 05h,42h, 84h, 08h, 05halternatively: 94h, 94h

8 bytes(2 bytes)

Main parameter with status 42h, 84h, 08h, 05h, (00h)alternatively: 94h (00h)

4 ... 5 bytes(1 ... 2 bytes)

Temperature with status 00h, 42h, 84h, 08h, 05halternatively: 00h, 94h

5 bytes(2 bytes)

pH/ORP with CCM 223/253 42h, 84h, 08h, 05h 4 BytesTab. 5.4 Useful configurations

XM53PE5.CHP

Liquisys M PROFIBUS-PA/-DP Communication

Endress+Hauser 13

Page 16: Liquisys (CCM253)_Chlorine_analyser

5.2 Acyclical data exchange

Acyclic data transmission is used to transfer parameters during commissioning, duringmaintenance or to display other measured variables that are not contained in the useful datatraffic.

The software of the Liquisys M CXM 223/253 transmitter includes the so-called physical blockcontaining instrument-specific information, such as measuring point designation (tag), softwareversion, etc., and a block with manufacturer-specific data used for instrument operation.

Generally, a distinction is made between Class 1 and Class 2 master connections.The Liquisys M transmitter only supports Class 2 master connections.

• Two Class 2 masters are permitted with Liquisys M. This means that two Class 2 masterscan access Liquisys M at the same time. However, you must make certain that they do notboth attempt to write to the same data. Otherwise the data consistency can lo longer beguaranteed.

• When a Class 2 master reads parameters, it sends a request telegram to the transmitterspecifiying the device address, the slot/index and the expected record length.The transmitter answers with the requested record if the record exists and is the correctlength (byte).

• When a Class 2 master writes parameters, it transmits the address of the transmitter, theslot/index, length information (byte) and the record. The transmitter acknowledges this writejob after completion. A Class 2 master can access the blocks mentioned above.

The device management and physical block are identical for all parameters measured.Please refer to the following tables for the index (all parameters are situated in slot 1).

Communication Liquisys M PROFIBUS-PA/-DP

14 Endress+Hauser

Page 17: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for device management and physical block

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

Device Management

Directory Object Header 1 / 0 12 UNSIGNED16 R

Composite List Directory Entries 1 / 1 24 UNSIGNED16 R

Physical Block Parameter

PB Block Objekt 1 / 14 20 DS-32 R

PB Static Revision 1 / 15 2 UNSIGNED16 R

PB Tag Description VAH0 1 / 16 32 OSTRING R / W

PB Strategy 1 / 17 2 UNSIGNED16 R / W

PB Alert Key 1 / 18 1 UNSIGNED8 R / W

PB Target Mode 1 / 19 1 UNSIGNED8 R / W

PB Mode Block 1 / 20 3 DS-37 R

PB Alarm Summary 1 / 21 8 DS-42 R

PB Software Revision 1 / 22 16 OSTRING R

PB Hardware Revision 1 / 23 16 OSTRING R

PB Device Manufacturer ID 1 / 24 2 UNSIGNED16 R

PB Device ID 1 / 25 16 OSTRING R

PB Device Serial Number 1 / 26 16 OSTRING R

PB Diagnosis 1 / 27 4 OSTRING R

PB Diagnosis Extension 1 / 28 6 OSTRING R

PB Diagnosis Mask 1 / 29 4 OSTRING R

PB Diagnosis Mask Extension 1 / 30 6 OSTRING R

PB Device Certification 1 / 31 16 OSTRING R / W

PB Security Locking 1 / 32 2 UNSIGNED16 R / W

PB Factory Reset 1 / 33 2 UNSIGNED16 R / W

PB Descriptor 1 / 44 32 OSTRING R / W

PB Device Message 1 / 45 32 OSTRING R / W

PB Device Installation Date 1 / 46 8 OSTRING R / W

PB Actual Errorcode V8H0 1 / 62 2 UNSIGNED16 R

PB UpDown Features Supported 1 / 64 1 OSTRING R

PB UpDown Control Parameter 1 / 65 1 UNSIGNED8 W

PB UpDown Parameter 1 / 66 20 UPDOWNDATA R / W

PB Device Bus Address VAH1 1 / 67 1 SIGNED8 R

PB Device & Software Number 1 / 68 2 UNSIGNED16 R

PB View_1 1 / 74 17 OSTRING RTab. 5.6

Slot / index list fordevice managementand physical block

XM53PE5.CHP

Liquisys M PROFIBUS-PA/-DP Communication

Endress+Hauser 15

Page 18: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for pH / redox

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

Manufacturer specific Block

Block Object 1 / 100 20 DS-32 R

Main measured value V0H0 1 / 108 4 FLOAT R

Measured temperature value V0H1 1 / 109 4 FLOAT R

Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W

Connection type V0H3 1 / 111 1 UNSIGNED8 R / W

Electrode type V0H5 1 / 112 1 UNSIGNED8 R / W

Measured value damping V0H6 1 / 113 1 UNSIGNED8 R / W

Measured value in mV V1H7 1 / 114 4 FLOAT R

Current measured / offset V0H8 1 / 115 4 FLOAT R / W

Calculated offset V0H9 1 / 116 4 FLOAT R

Temperature compensation type V1H0 1 / 117 1 UNSIGNED8 R / W

Entry of MTC temperature V1H1 1 / 118 4 FLOAT R / W

Temp. compensation for calibration V1H2 1 / 119 1 UNSIGNED8 R / W

Entry of process temperature V1H8 1 / 120 4 FLOAT R / W

Entry of temperature offset V1H9 1 / 121 4 FLOAT R / W

Beg. of meas. (bar graph) V2H3 1 / 122 4 FLOAT R

End of meas. (bar graph) V2H4 1 / 123 4 FLOAT R

Contact selection V3H0 1 / 124 1 UNSIGNED8 R / W

Contact configuration V3H1 1 / 125 1 UNSIGNED8 R / W

Switch contact on or off V3H2 1 / 126 1 UNSIGNED8 R / W

Switch-on point (LC) V4H0 1 / 127 4 FLOAT R / W

Switch-off point (LC) V4H1 1 / 128 4 FLOAT R / W

Pickup delay (LC) V4H2 1 / 129 2 UNSIGNED16 R / W

Dropout delay (LC) V4H3 1 / 130 2 UNSIGNED16 R / W

Alarm threshold (LC) V4H4 1 / 131 4 FLOAT R / W

Setpoint (PID / neutra) V4H0 1 / 132 4 FLOAT R / W

Control gain Kp (PID / neutra) V4H1 1 / 133 4 FLOAT R / W

Integral action time Tn (PID / neutra) V4H2 1 / 134 4 FLOAT R / W

Derivative action time Tv (PID / neutra) V4H3 1 / 135 4 FLOAT R / W

Control characteristic (PID) V4H4 1 / 136 1 UNSIGNED8 R / W

Operating mode of controller (PID / neutra) V4H5 1 / 137 1 UNSIGNED8 R / W

Pulse period (PID / neutra) V4H6 1 / 138 4 FLOAT R / W

Pulse frequency (PID / neutra) V4H7 1 / 139 2 UNSIGNED16 R / W

Minimum ON time (PID / neutra) V4H8 1 / 140 4 FLOAT R / W

Rinse time (timer) V4H0 1 / 141 2 UNSIGNED16 R / W

Pause time (timer) V4H1 1 / 142 2 UNSIGNED16 R / W

Minimum pause time (timer) V4H2 1 / 143 2 UNSIGNED16 R / W

Start pulse (cleaning) V4H0 1 / 144 1 UNSIGNED8 R / W

Pre-rinse time (cleaning) V4H1 1 / 145 2 UNSIGNED16 R / W

Cleaning time (cleaning) V4H2 1 / 146 2 UNSIGNED16 R / W

Post-rinse time (cleaning) V4H3 1 / 147 2 UNSIGNED16 R / W

Repeat cycles (cleaning) V4H4 1 / 148 1 UNSIGNED8 R / W

Pause time (cleaning) V4H5 1 / 149 2 UNSIGNED16 R / W

Minimum pause time (cleaning) V4H6 1 / 150 2 UNSIGNED16 R / W

Economy (cleaning) V4H7 1 / 151 1 UNSIGNED8 R / WTab. 5.7Slot / index list forpH / redox

Communication Liquisys M PROFIBUS-PA/-DP

16 Endress+Hauser

Page 19: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for pH / redox (continued)

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

Temperatursensor

V6H0

V0H4

1 / 160

1 / 152

1

1

UNSIGNED8

UNSIGNED8

R / W

R / W

Messfrequenz

V6H1

V0H7

1 / 161

1 / 153

1

1

UNSIGNED8

UNSIGNED8

UNSIGNED8

R / W

R / W

Transmitter type

V6H2 1 / 162

1 / 154

4

4

FLOAT R / W

Limit contact status

V6H3

V4H5

1 / 163

1 / 155

1

1

UNSIGNED8 R / W

Serial number

V8H1

VAH3

1 / 164

1 / 188

1

1

UNSIGNED8

OSTRING

R / W

Basic load

V8H2

V4H9

1 / 165

1 / 189

1

1

UNSIGNED8

UNSIGNED8

R / W

R / W

Controller type

V8H3

V5H8/V4H8

1 / 166

1 / 198

2

2

UNSIGNED16

UNSIGNED8

UNSIGNED8

R / W

R / W

Set current input active

Current input off delay

Current input on delay

Current input switch-off direction

Current input feedforward control

Current input gain

Current input available

Current input switch-off point

V8H4

V5H0

1 / 167

1 / 199

1

1

UNSIGNED8 R / W

R / W

R / W

V8H5

V5H1

1 / 168

1 / 200

1

1

UNSIGNED8 R

V8H6

V5H2

1 / 169

1 / 201

1

1

UNSIGNED8 R / W

R / W

V8H8

V5H3

1 / 170

1 / 202

1

1

UNSIGNED8

UNSIGNED8

UNSIGNED8

R / W

R / W

V8H9

V5H4

1 / 171

1 / 203

2

2

UNSIGNED16

UNSIGNED16

UNSIGNED16

R / W

R / W

V9H0

V5H5

1 / 172

1 / 204

1

1

UNSIGNED8

UNSIGNED8

UNSIGNED8

R / W

R / W

V9H1

V5H6

1 / 173

1 / 205

2

2

UNSIGNED16 R / W

R / W

V9H2 1 / 174

1 / 206

1

1

UNSIGNED8

UNSIGNED8

R

R

Alarm threshold monitoring

Process monitoring

Alarm delay

Max. period lower limit exceeded

Max. period upper limit exceeded

Lower alarmthreshold

Upper alarmthreshold

Limit of process monitoring

V9H3

V7H1

V7H5

1 / 175

1 / 207

1 / 211

1

1

1

UNSIGNED8

UNSIGNED8

UNSIGNED8

R

V9H4

V7H2

V7H6

1 / 176

1 / 208

1 / 212

1

1

1

UNSIGNED8 R / W

R / W

R / W

R / W

R / W

R / W

R / W

R / W

R / W

VAH2

V7H3

V7H4

V7H8

V7H7

1 / 177

1 / 209

1 / 213

2

2

2

UNSIGNED16

FLOAT

FLOAT

FLOAT

UNSIGNED16

UNSIGNED16

UNSIGNED16

R

R

R

R

1 / 178

1 / 210

1 / 214

1

1

1

UNSIGNED8 R

SCS alarm measuring electrode

SCS alarm reference electrode

SCS alarm threshold

PCS alarm

Alarm contact type

Unit alarm delay

Alarm delay

Error number selection

Error status

Alarm contact effective

Cleaning trigger

Release code

Hold configuration

Hold dwell period

Plus package released

Chemoclean released

Reset of instrument

Software version

Number of relays

Tab. 5.8Slot index list forpH / redox (continued)

XM53PE5.CHP

Liquisys M PROFIBUS-PA/-DP Communication

Endress+Hauser 17

Page 20: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for conductivity

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

Manufacturer specific Block

Block Object 1 / 100 20 DS-32 R

Main measured value V0H0 1 / 108 4 FLOAT R

Measured temperature value V0H1 1 / 109 4 FLOAT R

Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W

Display unit conc. V0H3 1 / 111 1 UNSIGNED8 R / W

Display format V0H4 1 / 112 1 UNSIGNED8 R / W

Display unit cd. / res. V0H5 1 / 113 1 UNSIGNED8 R / W

Measured value damping V0H6 1 / 114 1 UNSIGNED8 R / W

Uncompensated meas. value V0H7 1 / 115 4 FLOAT R

Temperature sensor V1H0 1 / 116 1 UNSIGNED8 R / W

Temperature compensation type V1H1 1 / 117 1 UNSIGNED8 R / W

Temperature coefficient alpha V1H2 1 / 118 4 FLOAT R / W

Process temperature V1H3 1 / 119 4 FLOAT R / W

Cell constant V1H4 1 / 120 4 FLOAT R / W

Cable resistance V1H5 1 / 121 4 FLOAT R / W

Entry of process temperature V1H8 1 / 122 4 FLOAT R / W

Entry of temperature offset V1H9 1 / 123 4 FLOAT R / W

Beg. of meas. (bar graph) –

1 / 124 4 FLOAT R

End of meas. (bar graph) 1 / 125 4 FLOAT R

Contact selection V3H0 1 / 126 1 UNSIGNED8 R / W

Contact configuration V3H1 1 / 127 1 UNSIGNED8 R / W

Switch contact on or off V3H2 1 / 128 1 UNSIGNED8 R / W

Switch-on point (LC) V4H0 1 / 129 4 FLOAT R / W

Switch-off point (LC) V4H1 1 / 130 4 FLOAT R / W

Pickup delay (LC) V4H2 1 / 131 2 UNSIGNED16 R / W

Dropout delay (LC) V4H3 1 / 132 2 UNSIGNED16 R / W

Alarm threshold (LC) V4H4 1 / 133 4 FLOAT R / W

Setpoint (PID / neutra) V4H0 1 / 134 4 FLOAT R / W

Control gain Kp (PID / neutra) V4H1 1 / 135 4 FLOAT R / W

Integral action time Tn (PID / neutra) V4H2 1 / 136 4 FLOAT R / W

Derivative action time Tv (PID / neutra) V4H3 1 / 137 4 FLOAT R / W

Control characteristic (PID) V4H4 1 / 138 1 UNSIGNED8 R / W

Operating mode of controller (PID / neutra) V4H5 1 / 139 1 UNSIGNED8 R / W

Pulse period (PID / neutra) V4H6 1 / 140 4 FLOAT R / W

Pulse frequency (PID / neutra) V4H7 1 / 141 2 UNSIGNED16 R / W

Minimum ON time (PID / neutra) V4H8 1 / 142 4 FLOAT R / W

Rinse time (timer) V4H0 1 / 143 2 UNSIGNED16 R / W

Pause time (timer) V4H1 1 / 144 2 UNSIGNED16 R / W

Minimum pause time (timer) V4H2 1 / 145 2 UNSIGNED16 R / W

Start pulse (cleaning) V4H0 1 / 146 1 UNSIGNED8 R / W

Pre-rinse time (cleaning) V4H1 1 / 147 2 UNSIGNED16 R / W

Cleaning time (cleaning) V4H2 1 / 148 2 UNSIGNED16 R / W

Post-rinse time (cleaning) V4H3 1 / 149 2 UNSIGNED16 R / W

Repeat cycles (cleaning) V4H4 1 / 150 1 UNSIGNED8 R / W

Pause time (cleaning) V4H5 1 / 151 2 UNSIGNED16 R / W

Minimum pause time (cleaning) V4H6 1 / 152 2 UNSIGNED16 R / W

Economy (cleaning) V4H7 1 / 153 1 UNSIGNED8 R / WTab. 5.9Slot / index list forconductivity

Communication Liquisys M PROFIBUS-PA/-DP

18 Endress+Hauser

Page 21: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for conductivity (continued)

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

V5H0 1 / 154 1 UNSIGNED8 R / W

V5H1 1 / 155 1 UNSIGNED8 R / W

V5H2 1 / 156 1 UNSIGNED8 R / W

V5H3 1 / 157 1 UNSIGNED8 R / W

V5H4 1 / 158 1 UNSIGNED8 R / W

V5H5 1 / 159 4 FLOAT R / W

V5H6 1 / 160 4 FLOAT R / W

V5H7 1 / 161 4 FLOAT R / W

V5H8 1 / 162 1 UNSIGNED8 R / W

V6H0 1 / 163 1 UNSIGNED8 R / W

V6H1 1 / 164 1 UNSIGNED8 R / W

V6H2 1 / 165 1 UNSIGNED8 R / W

V6H3 1 / 166 4 FLOAT R / W

V6H4 1 / 167 4 FLOAT R / W

V6H5 1 / 168 1 UNSIGNED8 R / W

V7H0 1 / 169 1 UNSIGNED8 R / W

V4H5 1 / 170 1 UNSIGNED8 R

V8H1 1 / 171 1 UNSIGNED8 R / W

V8H2 1 / 172 1 UNSIGNED8 R / W

V8H3 1 / 173 2 UNSIGNED16 R / W

V8H4 1 / 174 1 UNSIGNED8 R / W

V8H5 1 / 175 1 UNSIGNED8 R

V8H6 1 / 176 1 UNSIGNED8 R / W

V8H8 1 / 177 1 UNSIGNED8 R / W

V8H9 1 / 178 2 UNSIGNED16 R / W

VAH5 1 / 179 1 UNSIGNED8 R / W

VAH6 1 / 180 2 UNSIGNED16 R / W

VAH7 1 / 181 1 UNSIGNED8 R

V9H7 1 / 182 1 UNSIGNED8 R

V9H8 1 / 183 1 UNSIGNED8 R / W

V9H3 1 / 184 2 UNSIGNED16 R

1 / 185 1 UNSIGNED8 R

V1H6 1 / 186

4

4

4

4

4 FLOAT R

Reference temperature V1H3 1 / 187 R / W

R / WVAH3

V4H9

1 / 188 11 OSTRING

UNSIGNED81 / 189

2

Serial number

Basic load

Controller type

UNSIGNED8V9H6 1 / 206

1

1

R / W

R1 / 199 11

V9H0 1 / 200 1 R / W

V9H1 1 / 201

UNSIGNED8

UNSIGNED8

R / W

V9H2 1 / 202 2

2 UNSIGNED16

UNSIGNED16 R / W

V9H3 1 / 203 1

1

UNSIGNED8

UNSIGNED8

R / W

V9H4

V9H5

1 / 204 R / WR / W1 / 205 1 UNSIGNED8

Alarm threshold monitoring

Process monitoring

Alarm delay

Max. period lower limit exceeded

Max. period upper limit exceeded

Lower alarm threshold

Upper alarm threshold

Limit of process monitoring

V7H1

V7H5

1 / 190

1 / 194

1

1

UNSIGNED8

UNSIGNED8

UNSIGNED8

V7H2

V7H6

1 / 191

1 / 195

R / W

R / W

R / W

R / W

R / W

R / WR / W

R / W

R / W

V7H3

V7H4

V7H8

V7H7

1 / 192

1 / 196

22

FLOAT

FLOAT

FLOAT

FLOAT

UNSIGNED16

UNSIGNED16

UNSIGNED16

R

1 / 193

1 / 1971 / 198

Active concentration curve

Edit table (conc)

Table option (conc)

Number of value pairs (conc)

Selection of value pair (conc)

Uncomp. cd. value (conc)

Concentration value (conc)

Temperature value (conc)

Table status (conc)

Table option (alpha)

Number of table value pairs (alpha)

Selection of table value pair (alpha)

Temperature value (alpha)

Temperature coefficient (alpha)

Table status (alpha)

Polarisation detection (cond)

Limit contact status

Alarm contact type

Unit of alarm delay

Alarm delay

Error number selection

Error status

Alarm contact effective

Cleaning trigger

Release code

Hold configuration

Hold dwell period

Plus package released

Chemoclean released

Reset of instrument

Software version

Number of relays

Installation factor

Current input off delay

Current input on delay

Current input switch-off direction

Current input feedforward control

Current input gain

Current input available

Current input switch-off point

Set current input active

Tab. 5.10Slot / index list forconductivity (continued)

XM53PE5.CHP

Liquisys M PROFIBUS-PA/-DP Communication

Endress+Hauser 19

Page 22: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for turbidity

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

Manufacturer specific Block

Block Object 1 / 100 20 DS-32 R

Main measured value V0H0 1 / 108 4 FLOAT R

Measured temperature value V0H1 1 / 109 4 FLOAT R

Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W

Display unit V0H3 1 / 111 1 UNSIGNED8 R / W

Display format V0H4 1 / 112 1 UNSIGNED8 R / W

Sensor type V0H5 1 / 113 1 UNSIGNED8 R / W

Measured value damping V0H6 1 / 114 1 UNSIGNED8 R / W

Measured value in FNU V0H7 1 / 115 4 FLOAT R

Current measured value / offset V0H8 1 / 116 4 FLOAT R / W

Calculated offset V0H9 1 / 117 4 FLOAT R

Wiper on/off V1H0 1 / 118 1 UNSIGNED8 R / W

Wiping time V1H1 1 / 119 2 UNSIGNED16 R / W

Pause time V1H2 1 / 120 2 UNSIGNED16 R / W

Calibration data set V1H3 1 / 121 1 UNSIGNED8 R / W

Copy calibration data set V1H4 1 / 122 1 UNSIGNED8 R / W

Installation adaption V1H5 1 / 123 1 UNSIGNED8 R / W

Entry of process temperature V1H8 1 / 124 4 FLOAT R / W

Entry of temperature offset V1H9 1 / 125 4 FLOAT R / W

Beg. of meas. (bar graph) – 1 / 126 4 FLOAT R

End of meas. (bar graph) – 1 / 127 4 FLOAT R

Contact selection V3H0 1 / 128 1 UNSIGNED8 R / W

Contact configuration V3H1 1 / 129 1 UNSIGNED8 R / W

Switch contact on or off V3H2 1 / 130 1 UNSIGNED8 R / W

Switch-on point (LC) V4H0 1 / 131 4 FLOAT R / W

Switch-off point (LC) V4H1 1 / 132 4 FLOAT R / W

Pickup delay (LC) V4H2 1 / 133 2 UNSIGNED16 R / W

Dropout delay (LC) V4H3 1 / 134 2 UNSIGNED16 R / W

Alarm threshold (LC) V4H4 1 / 135 4 FLOAT R / W

Setpoint (PID / neutra) V4H0 1 / 136 4 FLOAT R / W

Control gain Kp (PID / neutra) V4H1 1 / 137 4 FLOAT R / W

Integral action time Tn (PID / neutra) V4H2 1 / 138 4 FLOAT R / W

Derivative action time Tv (PID / neutra) V4H3 1 / 139 4 FLOAT R / W

Control characteristic (PID) V4H4 1 / 140 1 UNSIGNED8 R / W

Operating mode of controller (PID / neutra) V4H5 1 / 141 1 UNSIGNED8 R / W

Pulse period (PID / neutra) V4H6 1 / 142 4 FLOAT R / W

Pulse frequency (PID / neutra) V4H7 1 / 143 2 UNSIGNED16 R / W

Minimum ON time (PID / neutra) V4H8 1 / 144 4 FLOAT R / W

Rinse time (timer) V4H0 1 / 145 2 UNSIGNED16 R / W

Pause time (timer) V4H1 1 / 146 2 UNSIGNED16 R / W

Minimum pause time (timer) V4H2 1 / 147 2 UNSIGNED16 R / WTab. 5.11 Slot / index list for turbidity

Communication Liquisys M PROFIBUS-PA/-DP

20 Endress+Hauser

Page 23: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for turbidity (continued)

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

V4H0 1 / 148 1 UNSIGNED8 R / W

V4H1 1 / 149 2 UNSIGNED16 R / W

V4H2 1 / 150 2 UNSIGNED16 R / W

V4H3 1 / 151 2 UNSIGNED16 R / W

V4H4 1 / 152 1 UNSIGNED8 R / W

V4H5 1 / 153 2 UNSIGNED16 R / W

V4H6 1 / 154 2 UNSIGNED16 R / W

V4H7 1 / 155 1 UNSIGNED8 R / W

V5H0 1 / 156 1 UNSIGNED8 R / W

V5H1 1 / 157 1 UNSIGNED8 R / W

V5H2 1 / 158 1 UNSIGNED8 R / W

V5H3 1 / 159 1 UNSIGNED8 R / W

V5H4 1 / 160 1 UNSIGNED8 R / W

V5H5 1 / 161 4 FLOAT R / W

V5H6 1 / 162 4 FLOAT R / W

V5H7 1 / 163 1 UNSIGNED8 R / W

V1H6 1 / 164 2 UNSIGNED8 R / W

V8H1 1 / 165 1 UNSIGNED8 R / W

V8H2 1 / 166 1 UNSIGNED8 R / W

V8H3 1 / 167 2 UNSIGNED16 R / W

V8H4 1 / 168 1 UNSIGNED8 R / W

V8H5 1 / 169 1 UNSIGNED8 R

V8H6 1 / 170 1 UNSIGNED8 R / W

V8H8 1 / 171 1 UNSIGNED8 R / W

V8H9 1 / 172 2 UNSIGNED16 R / W

V9H0 1 / 173 1 UNSIGNED8 R / W

V9H1 1 / 174 2 UNSIGNED16 R / W

V9H2 1 / 175 1 UNSIGNED8 R

V9H3 1 / 176 1 UNSIGNED8 R

V9H4 1 / 177 1 UNSIGNED8 R / W

VAH2 1 / 178 2 UNSIGNED16 R

1 / 179 1 UNSIGNED8

UNSIGNED8

R

R1

4

4

4

Limit contact status V1H3 1 / 180

R / WVAH3

V4H9

1 / 188 11 OSTRING

UNSIGNED81 / 189

2

UNSIGNED8V9H6 1 / 215

1

1

R / W

R1 / 208 1

1

V9H0 1 / 209 1 R / W

V9H1 1 / 210

UNSIGNED8

UNSIGNED8

R / W

V9H2 1 / 211 2

2 UNSIGNED16

UNSIGNED16 R / W

V9H3 1 / 212 1

1

UNSIGNED8

UNSIGNED8

R / W

V9H4

V9H5

1 / 213 R / WR / W1 / 214 1 UNSIGNED8

V7H1

V7H5

1 / 199

1 / 203

1

1

UNSIGNED8

UNSIGNED8

UNSIGNED8

V7H2

V7H6

1 / 200

1 / 204

R / W

R / W

R / W

R / W

R / W

R / W

R / W

R / W

R / W

V7H3

V7H4

V7H8

V7H7

1 / 201

1 / 205

22

FLOAT

FLOAT

FLOAT

UNSIGNED16

UNSIGNED16

UNSIGNED16

R

1 / 202

1 / 206

1 / 198

Serial number

Basic load

Controller type

Alarm threshold monitoring

Process monitoring

Alarm delay

Max. period lower limit exceeded

Max. period upper limit exceeded

Lower alarm threshold

Upper alarm threshold

Limit of process monitoring

Current input off delay

Current input on delay

Current input switch-off direction

Current input feedforward control

Current input gain

Current input available

Current input switch-off point

Set current input active

Start pulse (cleaning)

Pre-rinse time (cleaning)

Cleaning time (cleaning)

Post-rinse time (cleaning)

Repeat cycles (cleaning)

Pause time (cleaning)

Minimum pause time (cleaning)

Economy (cleaning)

Active concentration curve

Edit table (conc)

Table option (conc)

Number of value pairs (conc)

Selection of value pair (conc)

Entry of turbidity value (conc)

Concentration value (conc)

Table status (conc)

Gas bubble barrier

Alarm contact type

Unit of alarm delay

Alarm delay

Error number selection

Error status

Alarm contact effective

Cleaning trigger

Release code

Hold configuration

Hold dwell period

Plus package released

Chemoclean released

Reset of instrument

Software version

Number of relays

Tab. 5.12Slot / index list for turbidity(continued)

XM53PE5.CHP

Liquisys M PROFIBUS-PA/-DP Communication

Endress+Hauser 21

Page 24: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for oxygen

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

Manufacturer specific Block

Block Object 1 / 100 20 DS-32 R

Main measured value V0H0 1 / 108 4 FLOAT R

Measured temperature value V0H1 1 / 109 4 FLOAT R

Operating mode V0H2 1 / 110 1 UNSIGNED8 R / W

Pressure compensation V0H4 1 / 111 1 UNSIGNED8 R / W

Altitude V0H5 1 / 112 2 UNSIGNED16 R / W

Measured value damping V0H7 1 / 113 1 UNSIGNED8 R / W

Sensor current in nA V0H8 1 / 114 4 FLOAT R

Salinity V1H0 1 / 115 4 FLOAT R / W

Entry of process temperature V1H8 1 / 116 4 FLOAT R / W

Entry of temperature offset V1H9 1 / 117 4 FLOAT R / W

Beg. of meas. (bar graph) – 1 / 118 4 FLOAT R

End of meas. (bar graph) – 1 / 119 4 FLOAT R

Contact selection V3H0 1 / 120 1 UNSIGNED8 R / W

Contact configuration V3H1 1 / 121 1 UNSIGNED8 R / W

Switch contact on or off V3H2 1 / 122 1 UNSIGNED8 R / W

Switch-on point (LC) V4H0 1 / 123 4 FLOAT R / W

Switch-off (LC) V4H1 1 / 124 4 FLOAT R / W

Pickup delay (LC) V4H2 1 / 125 2 UNSIGNED16 R / W

dropout delay (LC) V4H3 1 / 126 2 UNSIGNED16 R / W

Alarm threshold (LC) V4H4 1 / 127 4 FLOAT R / W

Setpoint (PID / neutra) V4H0 1 / 128 4 FLOAT R / W

Control gain Kp (PID / neutra) V4H1 1 / 129 4 FLOAT R / W

Integral action time Tn (PID / neutra) V4H2 1 / 130 4 FLOAT R / W

Derivative action time Tv (PID / neutra) V4H3 1 / 131 4 FLOAT R / W

Control characteristic (PID) V4H4 1 / 132 1 UNSIGNED8 R / W

Operating mode of controller (PID / neutra) V4H5 1 / 133 1 UNSIGNED8 R / W

Pulse period (PID / neutra) V4H6 1 / 134 4 FLOAT R / W

Pulse frequency (PID / neutra) V4H7 1 / 135 2 UNSIGNED16 R / W

Minimum ON time (PID / neutra) V4H8 1 / 136 4 FLOAT R / W

Rinse time (timer) V4H0 1 / 137 2 UNSIGNED16 R / W

Pause time (timer) V4H1 1 / 138 2 UNSIGNED16 R / W

Minimum pause time (timer) V4H2 1 / 139 2 UNSIGNED16 R / W

Start pulse (cleaning) V4H0 1 / 140 1 UNSIGNED8 R / W

Pre-rinse time (cleaning) V4H1 1 / 141 2 UNSIGNED16 R / W

Cleaning time (cleaning) V4H2 1 / 142 2 UNSIGNED16 R / W

Post-rinse time (cleaning) V4H3 1 / 143 2 UNSIGNED16 R / W

Repeat cycles (cleaning) V4H4 1 / 144 1 UNSIGNED8 R / W

Pause time (cleaning) V4H5 1 / 145 2 UNSIGNED16 R / W

Minimum pause time (cleaning) V4H6 1 / 146 2 UNSIGNED16 R / W

Economy (cleaning) V4H7 1 / 147 1 UNSIGNED8 R / WTab. 5.13 Slot / index list for oxygen

Communication Liquisys M PROFIBUS-PA/-DP

22 Endress+Hauser

Page 25: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for oxygen (continued)

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

V7H0 1 / 148 1 UNSIGNED8 R / W

V7H1 1 / 149 2 UNSIGNED16 R / W

V7H2 1 / 150 4 FLOAT R / W

V7H3 1 / 151 4 FLOAT R / W

V7H4 1 / 152 1 UNSIGNED8 R / W

V7H5 1 / 153 2 UNSIGNED16 R / W

V7H6 1 / 154 2 UNSIGNED16 R / W

V7H7 1 / 155 4 FLOAT R / W

V8H1 1 / 156 1 UNSIGNED8 R / W

V8H2 1 / 157 1 UNSIGNED8 R / W

V8H3 1 / 158 2 UNSIGNED16 R / W

V8H4 1 / 159 1 UNSIGNED8 R / W

V8H5 1 / 160 1 UNSIGNED8 R

V8H6 1 / 161 1 UNSIGNED8 R / W

V8H8 1 / 162 1 UNSIGNED8 R / W

V8H9 1 / 163 2 UNSIGNED16 R / W

V9H0 1 / 164 1 UNSIGNED8 R / W

V9H1 1 / 165 2 UNSIGNED16 R / W

V9H2 1 / 166 1 UNSIGNED8 R

V9H3 1 / 167 1 UNSIGNED8 R

V9H4 1 / 168 1 UNSIGNED8 R / W

V9H3

1 / 169 2 UNSIGNED16 R

1 / 170 1 UNSIGNED8

UNSIGNED8

R

Range switching V0H6 1 / 171 1

Pressure measurement available 1 / 172 1 UNSIGNED8

UNSIGNED8

R

R1Limit contact status V4H5 1 / 173

R / WVAH3

V4H9

V0H3

1 / 188 11 OSTRING

UNSIGNED81 / 189

Serial number

Basic load

Controller type

Unit of measured value

UNSIGNED8

UNSIGNED8

V9H6 1 / 205

1 / 205

1

R / W

R

R1 / 199 1

1

1

V9H0 1 R / W

R / W

V9H1 1 / 200UNSIGNED8

R / W

V9H2 1 / 201 2

2 UNSIGNED16

UNSIGNED16 R / W

V9H3 1 / 20 1

1

UNSIGNED8

UNSIGNED8

R / W

V9H4

V9H5

1 / 203 R / WR / W1 / 204 1 UNSIGNED8

1

1

UNSIGNED8

UNSIGNED8

R / W

R / W

R

1 / 198

1 / 190

Alarm threshold monitoring

Alarm delay

Lower alarm threshold

Upper alarm threshold

Process monitoring

Limit value of process monitoring

Alarm contact type

Unit of alarm delay

Alarm delay

Error number selection

Error status

Alarm contact effective

Cleaning trigger

Release code

Hold configuration

Hold dwell period

Plus package released

Chemoclean released

Reset of instrument

Software version

Number of relays

Current input off delay

Current input on delay

Current input switch-off direction

Current input feedforward control

Current input gain

Current input available

Current input switch-off point

Set current input active

Max. period lower limit exceeded

Max. period upper limit exceeded

Tab. 5.14Slot / index list for oxygen(continued)

XM53PE5.CHP

Liquisys M PROFIBUS-PA/-DP Communication

Endress+Hauser 23

Page 26: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for chlorine

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

1 / 100 20 DS-32 R

V0H0 1 / 108 4 FLOAT R

Measured value pH / ORPV0H1

V0H2

V0H3

1 / 109

1 / 110

4

4

FLOAT

FLOAT

R

R

Controller switch-on delay

Controller switch-off delay

Controller switch-off

Type of sensor

V0H5 1 / 112

1 / 111

1

1

UNSIGNED8

UNSIGNED8

R / W

R / W

V0H6 1 / 113 2 UNSIGNED16 R / W

V0H7 1 / 114 2 UNSIGNED16 R / W

Selection of digital input 1 V0H8 1 / 115 1 UNSIGNED8

UNSIGNED8

UNSIGNED8

UNSIGNED8

R / W

Messwertdämpfung V0H9 1 / 116 1 R

Operating mode pH / ORP V1H0 1 / 117 1 R / W

pH compensation

Entry of process temperature

V1H8

V1H8

1 / 118

1 / 120

1

4 FLOAT

R / W

R / W

Manual pH compensation

Entry of temperature offset

V1H2

V1H9

1 / 119

1 / 121

4

4

FLOAT

FLOAT

R / W

R / W

V3H0 1 / 122 1 UNSIGNED8 R / W

V3H1 1 / 123 1 UNSIGNED8 R / W

V3H2 1 / 124 1 UNSIGNED8 R / W

V4H0 1 / 125 4 FLOAT R / W

V4H1 1 / 126 4 FLOAT R / W

V4H2 1 / 127 2 UNSIGNED16 R / W

V4H3 1 / 128 2 UNSIGNED16 R / W

V4H4 1 / 129 4 FLOAT R / W

V4H0 1 / 130 4 FLOAT R / W

V4H1 1 / 131 4 FLOAT R / W

V4H2 1 / 132 4 FLOAT R / W

V4H3 1 / 133 4 FLOAT R / W

V4H4 1 / 134 1 UNSIGNED8 R / W

V4H5 1 / 135 1 UNSIGNED8 R / W

V4H6 1 / 136 4 FLOAT R / W

V4H7 1 / 137 2 UNSIGNED16 R / W

V4H8 1 / 138 4 FLOAT R / W

V4H0 1 / 139 2 UNSIGNED16 R / W

V4H1 1 / 140 2 UNSIGNED16 R / W

V4H2 1 / 141 2 UNSIGNED16 R / W

V4H0 1 / 142 1 UNSIGNED8 R / W

V4H1 1 / 143 2 UNSIGNED16 R / W

V4H2 1 / 144 2 UNSIGNED16 R / W

V4H3 1 / 145 2 UNSIGNED16 R / W

V4H4 1 / 146 1 UNSIGNED8 R / W

V4H5 1 / 147 2 UNSIGNED16 R / W

V4H6 1 / 148 2 UNSIGNED16 R / W

V4H7 1 / 149 1 UNSIGNED8 R / W

Switch-on point (LC)

Switch-off (LC)

Pickup delay (LC)

Dropout delay (LC)

Alarm threshold (LC)

Setpoint (PID / neutra)

Control gain Kp (PID / neutra)

Integral action time Tn (PID / neutra)

Derivative action time Tv (PID / neutra)

Control characteristic (PID)

Operating mode of controller (PID / neutra)

Pulse period (PID / neutra)

Pulse frequency (PID / neutra)

Minimum ON time (PID / neutra)

Rinse time (timer)

Pause time (timer)

Minimum pause time (timer)

Start pulse (cleaning)

Pre-rinse time (cleaning)

Post-rinse time (cleaning)

Repeat cycles (cleaning)

Pause time (cleaning)

Minimum pause time (cleaning)

Economy (cleaning)

Cleaning time (cleaning)

Contact selection

Contact configuration

Switch contact on or off

Manufacturer specific Block

Block Object

Main measured value

Measured temperature value

Tab. 5.15 Slot / index list for chlorine

Communication Liquisys M PROFIBUS-PA/-DP

24 Endress+Hauser

Page 27: Liquisys (CCM253)_Chlorine_analyser

Slot / index list for chlorine (continued)

Parameter Matrix(Standard)

Slot /Index

Sizein byte(record)

Type read/write

Selection of monitoring parameters

V7H1

V7H0

1 / 151

1 / 150

1

1

UNSIGNED8

UNSIGNED8

R / W

R / W

V7H2 1 / 152 2 UNSIGNED16 R / W

V7H3 1 / 153 4 FLOAT R / W

V7H4 1 / 154 4 FLOAT R / W

V7H5 1 / 155 1 UNSIGNED8 R / W

V7H6 1 / 156 2 UNSIGNED16 R / W

V7H7 1 / 157 2 UNSIGNED16 R / W

V7H8 1 / 158 4 FLOAT R / W

V8H1 1 / 159 1 UNSIGNED8 R / W

V8H2 1 / 160 1 UNSIGNED8 R / W

V8H3 1 / 161 2 UNSIGNED16 R / W

V8H4 1 / 162 1 UNSIGNED8 R / W

V8H5 1 / 163 1 UNSIGNED8 R

V8H6 1 / 164 1 UNSIGNED8 R / W

V8H8 1 / 165 1 UNSIGNED8 R / W

V8H9 1 / 166 2 UNSIGNED16 R / W

V9H0 1 / 167 1 UNSIGNED8 R / W

V9H1 1 / 168 2 UNSIGNED16 R / W

V9H2 1 / 169 1 UNSIGNED8 R

V9H3 1 / 170 1 UNSIGNED8 R

V9H4 1 / 171 1 UNSIGNED8 R / W

Hardware version

VAH2

1 / 172

1 / 173

2

2

2

UNSIGNED16 R

R

1 / 174 1

1

UNSIGNED8

UNSIGNED8

R

Motor run time (3-pt. step controller)

Neutral zone (3-pt. step controller)

V4H5

V4H6

1 / 176

1 / 177

UNSIGNED8

R

V4H5

R / W

VAH3

V4H9

V0H4V5H9/V4H8

OSTRING

UNSIGNED8

Display unit

UNSIGNED8V5H6 1 / 184

1 / 185

1 / 188

1 / 189

1 / 190

1 / 198

R / W

1 / 178

1

1

V5H0 1 R / W

R / W

R / W

V5H1 1 / 179UNSIGNED8

R / WR / W

R / W

R / W

R / W

V5H2 1 / 180 2

2 UNSIGNED16

UNSIGNED16

UNSIGNED16UNSIGNED16

V5H3 1 / 181 1 UNSIGNED8

UNSIGNED8

R / W

V5H4

V5H5

1 / 182R / W1 / 183 1 UNSIGNED8

1

11

1

1

14

UNSIGNED8

UNSIGNED8

UNSIGNED8

R / W

R

1 / 175

Limit contact status

Serial number

Basic load

Controller type

Alarm threshold monitoring

Alarm delay

Lower alarm threshold

Upper alarm threshold

Process monitoring

Limit value of process monitoring

Alarm contact type

Unit of alarm delay

Alarm delay

Error number selection

Error status

Alarm contact effective

Cleaning trigger

Release code

Hold configuration

Hold dwell period

Plus package released

Chemoclean released

Reset of instrumentSoftware version

Number of relays

Current input off delay

Current input on delay

Current input switch-off direction

Current input feedforward control

Current input gain

Current input available

Current input switch-off point

Set current input active

Max. period lower limit exceeded

Max. period upper limit exceeded

Tab. 5.16Slot / index list for chlorine(continued)

XM53PE5.CHP

Liquisys M PROFIBUS-PA/-DP Communication

Endress+Hauser 25

Page 28: Liquisys (CCM253)_Chlorine_analyser

5.3 Operation via Commuwin II

PROFIBUS devices can be operated usingthe Commuwin II operating program. Theconnection to Commuwin II is established viathe PROFIBUS server. The device list showsall devices connected to the selected bussegments. The profile parameters areconfigured either via the operating matrix orvia the graphical user interface.

The operating structures of the variousmeasuring parameters are shown on thenext pages.

Note:

• Operation via Commuwin II isdescribed in the operatinginstructions BA 124F/00/en.

• If the off-line parameterisationdoes not correspond with theversion of the device(e.g., no release code for pluspackage), then error 03 will be setafter completion of the download,and the transmitter will not switchto the normal operating status.In this case, please repeat thedownload with correct data, orperform a device reset (VH 94).

Example of a Commuwin II matrix (e.g. CLM 2X3)

Communication Liquisys M PROFIBUS-PA/-DP

26 Endress+Hauser

Page 29: Liquisys (CCM253)_Chlorine_analyser

6 Accessories

Fieldbus connection box for PAFieldbus connection box for direct mountingto the instrument. Aluminium housing, ingressprotection IP 67. With 4-pole connector anda bus termination via a jumper. 2 cableglands Pg 9.Order no.: 017 481-0130

Fieldbus connection box for PAFieldbus connection box for direct mountingto the instrument. Aluminium housing, ingressprotection IP 67. With 4-pole connector anda bus termination via a jumper. 2 cableglands Pg 9. Internal ground condensator.Order no.: 017 481-0110

Fieldbus M12 instrument plug for PA4-pole metal plug for mounting to theinstrument for connection box or cable socketconnection. Cable length: 150 mm.Order no.: 51502184

Fieldbus Y adapterMetal Y adapter with 2 cable glands Pg 13.5.Order no.: 51502183

E+H

PROFIBUS-PA

PROFIBUS-DP

ENDRESS+HAUSERLIQUISYS SENDRESS+HAUSERLIQUISYS S

ENDRESS+HAUSERLIQUISYS

CAL REL

REL1

REL1

ALARMREL2

REL2E

ENDRESS+HAUSERLIQUISYS

CAL REL

REL1

REL1

ALARMREL2

REL2E

ENDRESS+HAUSERLIQUISYS

CAL REL

REL1

REL1

ALARMREL2

REL2E

PLCconfiguration

tool Device data file

PLC

Segment coupler

CXM 223separatemounting

CXM 253separatemountingCXM 253

directmounting

Fieldbus connection box

M12 plug

Bus termination box

XM253-07.CDR

CXM 253separatemounting

Fig. 6.1Instrumentation overviewLiquisys M CXM 2X3

n.c.

A (PA-)

B (PA+)

S

C2

58

64

18.5

17.5

18.5

37

n.c. A (PA-)

SB (PA+)

A (PA-)SB (PA+)

45 150/300 7

62

Pg 13.5

65

Pg 13.5

XM253D07.CHP

Liquisys M PROFIBUS-PA/-DP Accessories

Endress+Hauser 27

Page 30: Liquisys (CCM253)_Chlorine_analyser

Device data file• via the Internet:

E+H: http://www.endress.comProduct Avenue → Download Street →Field Communication Street)PNO: http://www.profibus.com(GSD library)

• Floppy disk by E+H:Order no.: 943157-0000

Commuwin IIGraphical operation software for allPROFIBUS instruments. Connection via theDDE server. Supported systems are Windows3.1/3.11, WIN 95 and WIN NT 4.0.Order no.: FXS 113-A7B

• 16 bit ISA plug-in board for PCOrder no.: 016 570-5300

• PCMCIA plug-in board for notebooksOrder no.: 016 570-5200

Cable for PATerminated cable with M12 plug andM12 coupling made of polyurethane andnickel-plated brass joints. Ingress protectionIP 67. Screen is connected to the threadedjoint. PVC sheath. Drilled and screenedconductor pair, 2/18 AWG.Temperature range – 40... +70°C.Lengths 1, 2, 5, 10 meter.

Cable type Order no.Cable terminated, yellow, 1 m 52001025

Cable terminated, yellow, 2 m 52001040

Cable terminated, yellow, 5 m 52001041

Cable terminated, yellow, 10 m 52001042

For further accessories, please refer to theSpecial Documentation SD 096F/00/en,available from the Endress+Hauser salesagencies (see back cover of these operatinginstructions).

Accessories Liquisys M PROFIBUS-PA/-DP

28 Endress+Hauser

Page 31: Liquisys (CCM253)_Chlorine_analyser

7 Technical data

General specifications Manufacturer Endress+Hauser

Product designation Liquisys M CXM 223/253-xx3

Output PROFIBUS-PA Output signal Digital communication signal,PROFIBUS-PA

PA function Slave

Transfer rate 31.25 kbit/s

Signal coding Manchester II

Response time of slave approx. 20 ms

Failure signal Signal: status bit is set, last valid measuredvalue is maintainedDisplay: error code

Physical layer IEC 1158-2

Bus voltage 9 ... 32 V

Current consumption of bus 10 mA ± 1 mA

Output PROFIBUS-DP Output signal PROFIBUS-DP acc. to EN 50 170 Part 3

DP function Slave

Transfer rate 9.6 kBit/s, 19.2 kBit/s, 93.75 kBit/s,187.5 kBit/s, 500 kBit/s, 1.5 MBit/s,3 MBit/s, 6 MBit/s, 12 MBit/s

Signal coding NRZ code

Failure signal Signal: status bit is set, last valid measuredvalue is maintainedDisplay: error code

Physical layer RS 485

Display and user interface Field operation via +, –, E, CAL, REL, AUTO keys

PC operation via PROFIBUS with Commuwin IIoperating program

Bus address Setting via DIL switch orvia operating menu orvia Set Slave Adr service

Communication interface PROFIBUS-PA/-DP

Standards and guidelines PROFIBUS EN 50 170, part 2; DIN 19 245, parts 1-3

PROFIBUS-DP EN 50 170, part 3; DIN 19 245, part 3PNO guidelines for PROFIBUS-DP

PROFIBUS-PA EN 50 170, part 4; DIN 19 245, part 4PNO guidelines for PROFIBUS-PA

Physical layer EN 61 158-2; IEC 1158-2, RS 485

Subject to modifications.

% Ω°C% Ω°C

XM53PE8.CHP

Liquisys M PROFIBUS-PA/-DP Technical data

Endress+Hauser 29

Page 32: Liquisys (CCM253)_Chlorine_analyser

8 Index

AAccessories . . . . . . . . . . . . . . . . . . . . . . 27-28Acyclical data exchange . . . . . . . . . . . . . . . 14Application . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

BBus cable connection to field instrument . . . . 5Bus cable connection to panel-mountedinstrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

CCommunication. . . . . . . . . . . . . . . . . . . . 12-26Commuwin II . . . . . . . . . . . . . . . . . . . . . . . . . 26Configuration. . . . . . . . . . . . . . . . . . . . . . . . . 13Configuration software . . . . . . . . . . . . . . . . . 11Cyclical data exchange . . . . . . . . . . . . . . . . 12

DData block. . . . . . . . . . . . . . . . . . . . . . . . . . . 13Data_Exchange . . . . . . . . . . . . . . . . . . . . . . 12Device address. . . . . . . . . . . . . . . . . . . . . . . . 9Device data file . . . . . . . . . . . . . . . . . . . . 10-11

EElectrical connection . . . . . . . . . . . . . . . . . 5-8

FFloating point number. . . . . . . . . . . . . . . . . . 12

GGeneral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2General safety instructions . . . . . . . . . . . . . . . 3GSD. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11

HHazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

IInstallation . . . . . . . . . . . . . . . . . . . . . . . . . 4-8Intended use . . . . . . . . . . . . . . . . . . . . . . . . . 3

MMaximum configuration . . . . . . . . . . . . . . . . 12

PPhysical block . . . . . . . . . . . . . . . . . . . . . . . 14Physical layer . . . . . . . . . . . . . . . . . . . . . . . . 29PLC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11PNO . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 12PROFIBUS user organization. . . . . . . . . . . . 10Programmable logic controller . . . . . 4, 12-13

SSafety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Safety instructions . . . . . . . . . . . . . . . . . . . . . 3Segment coupler . . . . . . . . . . . . . . . . . . . . . . 4Slot / index list . . . . . . . . . . . . . . . . . . . . 15-25Start-up. . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11Status codes . . . . . . . . . . . . . . . . . . . . . . . . 13Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Symbols used. . . . . . . . . . . . . . . . . . . . . . . . . 2System equipment . . . . . . . . . . . . . . . . . . . . . 4

TTechnical data . . . . . . . . . . . . . . . . . . . . . . . 29Terminating resistor . . . . . . . . . . . . . . . . . . . . 4Transmission method. . . . . . . . . . . . . . . . . . . 3Type file . . . . . . . . . . . . . . . . . . . . . . . . . 10-11

Index Liquisys M PROFIBUS-PA/-DP

30 Endress+Hauser

Page 33: Liquisys (CCM253)_Chlorine_analyser
Page 34: Liquisys (CCM253)_Chlorine_analyser
Page 35: Liquisys (CCM253)_Chlorine_analyser
Page 36: Liquisys (CCM253)_Chlorine_analyser

Europe

Austria Endress+Hauser Ges.m.b.H.WienTel. ++43 (1) 88056-0, Fax (1) 8 8056-35

BelarusBelorgsintezMinskTel. ++3 75 (1 72) 26 3166, Fax (172) 2631 11

Belgium / Luxembourg Endress+Hauser S.A./N.V.BrusselsTel. ++32 (2) 24806 00, Fax (2) 2 4805 53

BulgariaINTERTECH-AUTOMATIONSofiaTel. ++3 59 (2) 6648 69, Fax (2) 9 6313 89

Croatia Endress+Hauser GmbH+Co.ZagrebTel. ++3 85 (1) 6 63 7785, Fax (1) 663 7823

CyprusI+G Electrical Services Co. Ltd.NicosiaTel. ++3 57 (2) 4847 88, Fax (2) 48 4690

Czech Republic Endress+Hauser GmbH+Co.PrahaTel. ++4 20 (26) 678 4200, Fax (26) 67841 79

Denmark Endress+Hauser A/SSøborgTel. ++45 (70) 13 1132, Fax (70) 13 2133

EstoniaElvi-AquaTartuTel. ++3 72 (7) 4227 26, Fax (7) 42 2727

Finland Endress+Hauser OyEspooTel. ++3 58 (9) 8 59 6155, Fax (9) 859 6055

France Endress+HauserHuningueTel. ++33 (3) 89 6967 68, Fax (3) 89 69 4802

Germany Endress+Hauser Meßtechnik GmbH+Co.Weil am RheinTel. ++49 (7621) 975 01, Fax (76 21) 9 755 55

Great Britain Endress+Hauser Ltd.ManchesterTel. ++44 (161) 286 5000,Fax (161) 998 18 41

GreeceI & G Building Services Automation S.A.AthensTel. ++30 (1) 92415 00, Fax (1) 9 2217 14

HungaryMile Ipari-ElektroBudapestTel. ++36 (1) 26155 35, Fax (1) 2 6155 35

IcelandVatnshreinsun HFReykjavikTel. ++3 54 (5) 6196 16, Fax (5) 61 9617

IrelandFlomeaco Company Ltd.KildareTel. ++3 53 (45) 8686 15, Fax (45) 8681 82

Italy Endress+Hauser Italia S.p.A.Cernusco s/N MilanoTel. ++39 (02) 92 106421,Fax (02) 92 10 7153

LatviaRaita Ltd.RigaTel. ++3 71 (7) 3128 97, Fax (7) 31 2894

LithuaniaAgava Ltd.KaunasTel. ++3 70 (7) 2024 10, Fax (7) 20 7414

Netherlands Endress+Hauser B.V.NaardenTel. ++31 (35) 6 9586 11, Fax (35) 6 95 8825

Norway Endress+Hauser A/STranbyTel. ++47 (32) 85 98 50, Fax (32) 85 9851

Poland Endress+Hauser Polska Sp. z o.o.WarszawyTel. ++48 (22) 7 201090, Fax (22) 720 1085

PortugalTecnisis - Tecnica de Sistemas IndustriaisLinda-a-VelhaTel. ++351 (1) 417 2637, Fax (1) 418 52 78

RomaniaRomconseng SRLBucharestTel. ++40 (1) 4 1016 34, Fax (1) 4 101634

Russia Endress+Hauser Moscow OfficeMoscowTel.++7 09 (5) 1 5875 71, Fax (5) 1 589864

Slovak RepublicTranscom Technik s.r.o.BratislavaTel. ++421 (74) 488 86 84, Fax (74) 4 8871 12

Slovenia Endress+Hauser D.O.O.LjubljanaTel. ++386 (61) 159 22 17, Fax (61) 1 5922 98

Spain Endress+Hauser S.A.BarcelonaTel. ++34 (93) 4 803366, Fax (93) 473 3839

Sweden Endress+Hauser ABSollentunaTel. ++46 (8) 55 511600, Fax (8) 5551 1600

Switzerland Endress+Hauser AGReinach/BL 1Tel. ++41 (61) 7 157575, Fax (61) 711 1650

TurkeyIntek Endüstriyel Ölcü ve Kontrol SistemleriIstanbulTel. ++90 (212) 275 13 55,Fax (2 12) 26627 75

UkraineIndustria UkraïnaKievTel. ++380 (44) 268 81, Fax (44) 2 69 08

YugoslaviaMeris d.o.o.BeogradTel. ++381 (11) 444 61 64, Fax (11) 4 4419 66

Africa

EgyptAnasiaHeliopolis/CairoTel. ++20 (2) 4 179 00, Fax (2) 4 179 00

MoroccoOussama S.A.CasablancaTel. ++212 (2) 2413 38, Fax (2) 40 26 57

NigeriaJ F Technical Invest. Nig. Ltd.LagosTel. ++234 (1) 6223 45 46, Fax (1) 62 2345 48

South Africa Endress+Hauser Pty. Ltd.SandtonTel. ++27 (11) 4 441386, Fax (11) 444 1977

TunisiaControle, Maintenance et RegulationTunisTel. ++216 (1) 7930 77, Fax (1) 78 85 95

America

Argentina Endress+Hauser Argentina S.A.Buenos AiresTel. ++54 (1) 1 4522 79 70,Fax (1) 1 452279 09

BoliviaTritec S.R.L.CochabambaTel. ++591 (42) 569 93, Fax (42) 5 09 81

Brazil Samson Endress+Hauser Ltda.Sao PauloTel. ++55 (11) 5 03134 55,Fax (11) 50 3130 67

Canada Endress+Hauser Ltd.Burlington, OntarioTel. ++1 (905) 681 92 92,Fax (9 05) 6 8194 44

ChileDIN Instrumentos Ltda.SantiagoTel. ++56 (2) 2 050100, Fax (2) 2 25 8139

ColombiaColsein Ltd.Bogota D.C.Tel. ++57 (1) 2 367659, Fax (1) 6 10 7868

Costa RicaEURO-TEC S.A.San JoseTel. ++506 (2) 961542, Fax (2 ) 961542

EcuadorInsetec Cia. Ltda.QuitoTel. ++593 (2) 269148, Fax (02) 4618 33

GuatemalaACISA Automatizacion Y Control Industrial S.A.Ciudad de Guatemala, C.A.Tel. ++502 (3) 345985, Fax (2) 3274 31

Mexico Endress+Hauser I.I.Mexico CityTel. ++52 (5) 56 89 65, Fax (5) 56 8418

ParaguayIncoel S.R.L.AsuncionTel. ++595 (21) 21 3989, Fax (21) 226583

UruguayCircular S.A.MontevideoTel. ++598 (2) 925785, Fax (2) 9291 51

USA Endress+Hauser Inc.Greenwood, IndianaTel. ++1 (317) 535 71 38,Fax (3 17) 5 3584 89

VenezuelaH. Z. Instrumentos C.A.CaracasTel. ++58 (2) 9 440966, Fax (2) 9 44 4554

Asia

China Endress+Hauser Shanghai

Instrumentation Co. Ltd.ShanghaiTel. ++86 (21) 5490 2300,Fax (21) 5490 23 03

Endress+Hauser Beijing OfficeBeijingTel. ++86 (10) 6834 4058,Fax (10) 6834 40 68

Endress+Hauser (H.K.) Ltd.Hong KongTel. ++852 (2) 528 31 20, Fax (2) 86541 71

India Endress+Hauser India Branch OfficeMumbaiTel. ++91 (22) 8 52 1458, Fax (22) 852 19 27

IndonesiaPT Grama BazitaJakartaTel. ++62 (21) 7 97 5083, Fax (21) 797 50 89

Japan Sakura Endress Co., Ltd.TokyoTel. ++81 (4 22) 54 0611, Fax (4 22) 55 02 75

Malaysia Endress+Hauser (M) Sdn. Bhd.Petaling Jaya, Selangor Darul EhsanTel. ++60 (3) 7 334848, Fax (3) 7 33 8800

PakistanSpeedy AutomationKarachiTel. ++92 (21) 7 72 2953, Fax (21) 773 68 84

Papua New GuineaSBS Electrical Pty LimitedPort MoresbyTel. ++675 (3) 25 1188, Fax (3) 2595 56

PhilippinesBrenton Industries Inc.Makati Metro ManilaTel. ++63 (2) 6 38 8041, Fax (2) 638 8042

Singapore Endress+Hauser (S.E.A.) Pte., Ltd.SingaporeTel.++65 (5) 66 8222, Fax (2) 6668 48

South Korea Endress+Hauser (Korea) Co., Ltd.SeoulTel. ++82 (2) 6 58 7200, Fax (2) 659 2838

TaiwanKingjarl CorporationTaipei R.O.C.Tel. ++886 (2) 2718 39 38, Fax (2) 27 1341 90

Thailand Endress+Hauser Ltd.BangkokTel. ++66 (2) 9 96 7811 20, Fax (2) 9 967810

VietnamTan Viet Bao Co. Ltd.Ho Chi Minh CityTel. ++84 (8) 8 33 5225, Fax (8) 833 5227

IranTelephone Technical Services Co. Ltd.TehranTel. ++98 (21) 874 6750, Fax (21) 87372 95

IsraelInstrumetrics Industrial Control Ltd.Tel-AvivTel. ++972 (3) 64802 05, Fax (3) 6 4719 92

JordanA.P. Parpas Engineering S.A.AmmanTel. ++962 (6) 46432 46, Fax (6) 4 6457 07

Kingdom of Saudi ArabiaAnasiaJeddahTel. ++966 (2) 67100 14, Fax (2) 6 7259 29

KuwaitKuwait Maritime & Mercantile Co. K.S.C.SafatTel.++9 65 (2) 4414 81, Fax (2) 44 1486

LebanonNabil IbrahimJbeilTel. ++961 (3) 25 4052, Fax (9) 5480 38

Sultanate of OmanMustafa & Jawad Science & Industry Co.L.L.C.RuwiTel. ++968 (60) 20 09, Fax (60) 70 66

United Arab EmiratesDescon Trading EST.DubaiTel. ++971 (4) 65 3651, Fax (4) 6532 64

YemenYemen Company for Ghee and Soap IndustryTaizTel. ++976 (4) 23 0664, Fax (4) 2123 38

Australia + New Zealand

AustraliaALSTOM Australia Ltd.SydneyTel. ++61 (2) 9722 4777, Fax (2) 972248 88

New ZealandEMC Industrial InstrumentationAucklandTel. ++64 (9) 4 15 5110, Fax (9) 415 5115

All other countries

Endress+Hauser GmbH+Co.Instruments International

D-Weil am RheinGermanyTel. ++49 (76 21) 9 7502, Fax (76 21) 97 5345

BA 209C/07/en/07.02Printed in Germany / CV5 / DT