Logical In-Vial Dilution with Ultrafiltration for Anion
Analysis
IC Application Work AW US6-0149-012011
Presented by Carlos Bazán
![Page 2: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/2.jpg)
2
Overview
• Dosino technology provides a highly efficient and accurate method for performing dilutions in-line; saving customers time, money, and reducing human error
• Previous configurations required external mixing with added components and added $$
• The In-Vial Dilution configuration can perform dilutions in an empty sample vial and utilizes the Dosino to mix the diluted sample
• Intelligent re-dilution of samples above the calibration range is possible
• Next-Vial Dilution allows the next sample in the series to be diluted while the current sample is still running, maintaining high sample through-put
![Page 3: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/3.jpg)
3
Components needed• 881, 882, or 850 IC
• 858 Sample Processor with peristaltic pump (2.858.0020)
• 800 Dosino with 10 mL Dosing Unit
• IC Equipment for Ultrafiltration (6.5330.010)
• T-Connector 3 X UNF 10-32 (6.2744.290; diverts flow for dilution or filtration)
• In-line check valve (6.2744.190; prevents Dosino from pulling from ultrafiltration cell waste)
• 10 mL Transfer Capillary (6.1562.130)
• Wash station for IC (6.2841.100)
![Page 4: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/4.jpg)
4
Basic function• Sample is aspirated into the transfer capillary by the Dosino. The in-
line check valve prevents UF cell waste from being aspirated.
![Page 5: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/5.jpg)
5
Basic function• The aspirated sample and a quantity of UHP H2O to complete the
dilution are dispensed into an empty sample vial.
![Page 6: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/6.jpg)
6
Basic function• The diluted sample is aspirated and dosed several times to mix it
completely.
![Page 7: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/7.jpg)
7
Basic function• The peristaltic pump moves the diluted sample through the
ultrafiltration cell and to the IC.
![Page 8: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/8.jpg)
8
MagIC Net Sample Table
Injections= 1 for normal analysis, 2 when an intelligent re-dilution check should be performed
Dilution= the Manual Dilution Factor for hand-dilutions made
Value 1= the Dosino Dilution Factor; the dilution the instrument will perform
Value 2= the Dilution Vial; vial position the dilution is performed in
Value 3= the Intelligent Dilution Vial; vial position for intelligent re-dilution when needed
![Page 9: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/9.jpg)
9
Features: Calibration• Calibration from one stock standard; levels made by variable
dilution• Makes standard prep easier, saves time, reduces human error
Dilution Factor (Value 1)
200 100 50 20 10
![Page 10: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/10.jpg)
10
Features: Next-Vial Dilution• Prepares the dilution of the next sample in the series while the
previous sample is still running• Reduces the overall dilution time needed for a sample series;
very valuable for labs needing high through-put• Can be used with intelligent re-dilution; during the re-dilution run
the next sample is prepared• Uses the SD.NEXT values in the program to access the POSITION,
VALUE1 and VALUE2 of the next line in the sample table
![Page 11: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/11.jpg)
11
Features: Multi-component Intelligent Re-dilution• Software can evaluate which out of the entire list of
component ions is most above the calibration range • Re-dilutes samples accordingly
If this is not an intelligent dilution run (Injection=1),or the Area of Fluoride =0,then Autodilution Factor for Fluoride=0.
If it is an intelligent dilution run (Injection=2) then Autodilution Factor for Fluoride = (Concentration of Fluoride in current run / (Concentration of Fluoride in the highest standard X the Manual Dilution Factor )) X the Dosino Dilution Factor
![Page 12: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/12.jpg)
12
Features: Multi-component Intelligent Re-dilution
![Page 13: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/13.jpg)
13
Performance: Calibration Linearity
• Low calibration RSD
• Linear across a broad calibration range
![Page 14: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/14.jpg)
14
Performance: Dilution Accuracy
• 99%+ Accuracy compared to a 1:10 hand-dilution• 96%+ Accuracy compared to a 1:100 hand-dilution
(1:100 is maximum advisable Dosino-prepared dilution with 10 mL Dosing Unit)
![Page 15: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/15.jpg)
15
Performance: Reproducibility
• Approximately 1.5% RSD on 6 replicate injections of 1:100 dilution (with lower RSD at lesser dilution factors).
![Page 16: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/16.jpg)
16
Performance: Intelligent Re-dilution• System can re-dilute samples based on which ion is most above the
calibration range; can compare all analytes in component table
Highest standard
Samples
Fluoride Cl N02 as N Br NO2 as N P04 as P SO4
Std 5 3.972 ppm 100.463 ppm 1.507 ppm 1.977 ppm 10.072 ppm 9.699 ppm 98.899 ppm
![Page 17: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/17.jpg)
17
Market Potential
• A large percentage of environmental and QC laboratories must routinely perform sample dilutions
• User proficiency is going down, so reliable in-line, automated methods for dilution and standard preparation are highly valuable
• Automatic re-dilution of samples above the calibration range saves time, especially on samples with 48 hr. hold-time
• LOW COST OF THE SYSTEM MEANS WE SHOULD SELL A LOT OF THESE TO A LOT OF USERS!
![Page 18: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/18.jpg)
18
Acknowledgements
• Jay Kim for invaluable contributions to programming of the intelligent dilution and next-vial dilution components of the method
• Dr. Shibu Paul for the initial development request and continuing drive to perfect the method
• Dr. Hari Narayanan for advice and guidance
• Larry Tucker for suggestions on sample flow path plumbing and wisdom on Dosino capabilities
• Dr. Stuart Procter for method testing during technical training sessions
![Page 19: Logical In-Vial Dilution with Ultrafiltration for Anion Analysis IC Application Work AW US6-0149-012011 Presented by Carlos Bazán 1](https://reader035.vdocument.in/reader035/viewer/2022062716/56649dff5503460f94ae7ca4/html5/thumbnails/19.jpg)
19
Questions?
?