Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation. Daniel Mihalcea. Northern Illinois University Department of Physics. Outline:. Fermilab/NICADD overview Michelson interferometer Bunch shape determination Experimental results Conclusions. - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/1.jpg)
Fermilab, Jan. 16, 2007
Longitudinal Diagnostics of Electron Bunches Using
Coherent Transition Radiation
Daniel Mihalcea
Northern Illinois University
Department of Physics
![Page 2: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/2.jpg)
•Fermilab/NICADD overview
• Michelson interferometer
• Bunch shape determination
• Experimental results
• Conclusions
Outline:
![Page 3: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/3.jpg)
Fermilab NICADD Photo-injector Laboratory
FNPL is a collaborative effort of several institutes and universities to operate a high-brightness photo-injector dedicated to accelerator and beam physics research.
Collaborators include:
U. of Chicago, U. of Rochester, UCLA, U. of Indiana, U. of Michigan, LBNL, NIU, U. of Georgia, Jlab, Cornell University
DESY, INFN-Milano, IPN-Orsay, CEA-Saclay
![Page 4: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/4.jpg)
FNPL layout
Michelson interferometer for longitudinal diagnostics
![Page 5: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/5.jpg)
Michelson Interferometer
(University of Georgia & NIU)
Autocorrelation = I1/I2 Detectors: - Molectron pyro-electric
- Golay cell (opto-acoustic)
![Page 6: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/6.jpg)
Data Flow
Interferometer
ICT Detectors Stepping motor
Scope
Controller
Get Q Get I1 and I2
LabView code: Advances stepping motor between x1 and x2 with adjustable step size (50 m)
At each position there are N readings (5)
A reading is valid if bunch charge is within some narrow window (Ex: 1nC 0.1 nC)
Position, average values of I1 and I2 and their ’s are recorded.
Autocorrelation function is displayed.
![Page 7: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/7.jpg)
Basic Principle (1)
222
22
03
22
)cos1(
sin
4
c
e
dd
UdIe
Ginsburg-Franck:
)(fIe
Detector aperture 1 cm
Backward transition radiation
![Page 8: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/8.jpg)
Basic Principle (2)
2)()1()( fINNI eIntensity of Optical Transition Radiation:
Coherent part N2
Form factor )(f related to longitudinal charge distribution: dzczizf )/exp()()(
To determine (z) need to know I() and the phase of f()
dttEctEI2
1 )()/(
dttE
dtctEtE
I
IS 2
2
1
)(
)/()(Re)(
dE
deES
ci
2
/2
)(
)(Re)( deSEI ci
/2)()()(
Kramers-Kröning technique
![Page 9: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/9.jpg)
Coherence condition
z mm3max
Due to detector sensitivity:
Acceptable resolution: mmz 6.0 Need bunch compression !
![Page 10: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/10.jpg)
Bunch Compression RF field in booster cavity
Electron bunch before compression
Energy-Position correlation
Head
Tail
After compression
mm1
mm15.0
2566 c
Tt f
![Page 11: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/11.jpg)
Measurement Steps
Ideal apparatus
mmz 19.0
2/
FT
K-K
mmz 18.0
Kramers-Kröning method:
dczIc
z
x
IxIdx
/)(cos)(1
)(
)(/)(ln)(
0
022
![Page 12: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/12.jpg)
Experimental results (1)
Path difference (mm)
Frequency (THz)c
ndE
22 sin||
Molectron pyro-electric detectors
Interference effect Missing frequencies
![Page 13: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/13.jpg)
Experimental results (2)
Golay detectors: no problem with interference !
Still need to account for:
low detector sensitivity at low frequencies
diffraction at low frequencies
absorption at large frequencies
![Page 14: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/14.jpg)
Experimental results (3)
Low detector sensitivity
DiffractionAbsorption
Interference
Apparatus response function:
)(
)()(
simulation
measured
I
IR
Beam conditions:
Q = 0.5 nC
maximum compression
![Page 15: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/15.jpg)
Experimental results (4)
Auto-correlation function:
Q = 3nC
9-cell phase was 3 degrees from maximum compression
Power spectrum:
Asymptotic behavior
low frequencies:
high frequencies:
Least square fit.
21)( aI
)(I
Molectron pyroelectric detectors
![Page 16: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/16.jpg)
Experimental results (5)
• Molectron pyroelectric detectors
• Kramers-Kroning method
Head
Tail
Parmela simulation
Head-Tail ambiguity
![Page 17: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/17.jpg)
Experimental results (6)
FT
Spectrum correction with R()
Spectrum completion for:
andzc /0 0
Beam conditions:
Q = 3.0 nC
moderate compression
K-K
Golay cell
Start point
z 1ps
![Page 18: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/18.jpg)
Complicated bunch shapes
Stack 4 laser pulses
Select 1st and 4th pulses (t 15ps)
Before compression
After compression
(Parmela simulations)
![Page 19: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/19.jpg)
Experimental results (7)
Double-peaked bunch shapes
Beam conditions:
Q = 0.5 nC each pulse
15 ps initial separation between the two pulses
both pulses moderately compressed
K-K method may not be accurate for complicate bunch shapes !
![Page 20: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/20.jpg)
K-K method accuracy
R. Lai and A. J. Sievers, Physical Review E, 52, 4576, (1995)
Generated
Reconstructed
K-K method accurate if:
• Simple bunch structure
• Stronger component comes first
Calculated widths are still correct !
![Page 21: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/21.jpg)
Other approaches
Major problem: the response function is not flat.
1. Complete I() based on some assumptions at low and high frequencies.
R. Lai, et al. Physical Review E, 50, R4294, (1994).
S. Zhang, et al. JLAB-TN-04-024, (2004).
2. Avoid K-K method by assuming that bunches have a predefined shape and make some assumptions about I() at low frequencies.
A. Murokh, et al. NIM A410, 452-460, (1998).
M. Geitz, et al. Proceedings PAC99, p2172, (1999).
This work:
D. Mihalcea, C. L. Bohn, U. Happek and P. Piot, Phys. Rev. ST Accel. Beams 9, 082801 (2006).
![Page 22: Longitudinal Diagnostics of Electron Bunches Using Coherent Transition Radiation](https://reader036.vdocument.in/reader036/viewer/2022062517/56813815550346895d9fc86d/html5/thumbnails/22.jpg)
Conclusions:
Longitudinal profiles with bunch lengths less than 0.6 mm can be measured.
Systematic uncertainties dominated by approximate knowledge of response function and completion procedure.
Golay cells are better because the response function is more uniform.
Some complicate shapes (like double-peaked bunches) can be measured.