Low-temperature primary thermometry development at NRC. Dr. Patrick M.C. Rourke Measurement Science and Standards (MSS) National Research Council Canada (NRC) CAP Congress, Sudbury, 19 June 2014. Thermometry. Primary thermometer Directly measure “real” thermodynamic temperature T - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Low-temperature primary thermometry development at NRC

Low-temperature primary thermometry development at NRC

Dr. Patrick M.C. RourkeMeasurement Science and Standards (MSS)National Research Council Canada (NRC)

CAP Congress, Sudbury, 19 June 2014

Page 2: Low-temperature primary thermometry development at NRC

Thermometry

• Primary thermometer• Directly measure “real” thermodynamic temperature T• Complicated, large, difficult to use not many in existence

• Secondary thermometer• Needs calibration in order to set scale• Almost all thermometers are secondary

• International Temperature Scale of 1990 (ITS-90)• Used for secondary thermometer calibrations worldwide between

0.65 K and 1357.77 K• Based on best thermodynamic data from primary thermometers

available up to 1990• Newer measurements suggest the scale should be improved

2

Page 3: Low-temperature primary thermometry development at NRC

ITS-90 scale deviates from thermodynamic temperature

0 50 100 150 200 250 300

-12

-10

-8

-6

-4

-2

0

2

4

Gas Thermometers

Constant Volume(CVGT)

Kemp 1986 Steur 1986 Astrov rev.

1995/96

Acoustic(AGT)

Moldover 1999 Ewing 2000 Benedetto 2004 Pitre 2006

Dielectric Constant(DCGT)

Gaiser 2008 Gaiser 2010

T - T

90 (m

K)

Temperature (K)

3

Adapted from CCT-WG4 report (2008), Fischer et al., Int. J. Thermophys. 32, 12 (2011),Astrov et al., Metrologia 32, 393 (1995/96) and Gaiser et al., Int. J. Thermophys. 31, 1428 (2010)

Page 4: Low-temperature primary thermometry development at NRC

• Microwave resonances in a gas-filled conducting cavity• Fixed temperature & gas pressure

• Resonance frequency f gas refractive index n• c0: speed of light in vacuum• ξ: electromagnetic eigenvalue for microwave

resonance• a: radius of spherical cavity• Thermal expansion coefficient αL and isothermal

compressibility κT important

• Calculate thermodynamic temperature T from n using virial equations• Helium gas: quantum mechanics

• Similarities to other techniques• Acoustic gas thermometry (AGT)• Dielectric constant gas thermometry (DCGT)• Resolve differences between them?

Refractive index gas thermometry (RIGT) in principal

4

Page 5: Low-temperature primary thermometry development at NRC

RIGT in practice

2.612 2.613 2.614 2.615 2.616 2.617 2.618 2.6190

5

10

15

20

25

30

35

40

45

2.621 2.622 2.623 2.624 2.625 2.626 2.627 2.628

0

5

10

15

20

25

30

35

40

45

Frequency (GHz) at T = 5 K

TM11 mode at 297 K and 5 K, in vacuum

g3

g2

g1

f3

f2

peak 3 ("z")

peak 2 ("x")

106 |S 21

|

Frequency (GHz) at T = 297 K

peak 1 ("y")f1

• Quasi-spherical resonator• Controllably lift resonance

degeneracy

• Finite electrical conductivity• microwaves penetrate into skin

layer• resonances broadened &

shifted

• Eigenvalue corrections• Shape effects• Disturbances due to waveguides

5

Page 6: Low-temperature primary thermometry development at NRC

Experimental details

• Motivation: RIGT to measure T - T90: 5 K – 300 K• Initially, characterize resonator in vacuum• Microwave resonances resonator size, shape,

conductivity

• Prototype copper resonator

• Copper pressure vessel• Resistive thermometers (ITS-90) on copper coupling rod

• Two-stage pulse-tube cryocooler• Home-made thermal control system

6

Page 7: Low-temperature primary thermometry development at NRC

Microwave fitting

2.612 2.613 2.614 2.615 2.616 2.617 2.618 2.619

-10

-8

-6

-4

-2

0

2

4

6

8

10

10

6 [Re

(S21

) or I

m(S

21)]

Frequency (GHz)

TM11 mode at 297 K, in vacuum

2.621 2.622 2.623 2.624 2.625 2.626 2.627 2.628

-15

-10

-5

0

5

10

15

20

25

30

35

40

106 [

Re(S

21) o

r Im

(S21

)]

Frequency (GHz)

TM11 mode at 5 K, in vacuum

• Measure microwave resonances using 2-port Portable Network Analyzer

• Complex 3-Lozentzian + polynomial background fitting routine• Peak frequencies and half-widths

• Several microwave modes measured• Optimized spectral fitting background

terms, 1st- & 2nd-order shape corrections, and waveguide corrections

• Room temperature results agree with those done at NIST May et al., Rev. Sci. Instrum. 75, 3307 (2004)

7

Page 8: Low-temperature primary thermometry development at NRC

Electrical conductivity

0 50 100 150 200 250 3000

1x108

2x108

3x108

4x108

5x108

6x108

7x108

8x108

9x108

1x109

Present study

OFHC Cu from Simon et al. 1992 / Hust & Lankford 1984

+/- 15% of Simon et al. 1992 / Hust & Lankford 1984 curve

r,Cu

Cu (S

·m-1)

Temperature (K)

Copper conductivity, TM11 peak 1 half-width • Temperature dependence of resonator conductivity (from peak width)

• Stable, fixed temperatures over entire temperature range

• Agrees with literature within literature curve’s 15% uncertainty Simon et al., NIST Monograph 177, 1992

• Free parameter σ(T = 0) ≡ 1/ρ0 set to present experimental data at 5 K

8

Page 9: Low-temperature primary thermometry development at NRC

Thermal expansion coefficient αL

0 50 100 150 200 250 300

0.0

2.0x10-6

4.0x10-6

6.0x10-6

8.0x10-6

1.0x10-5

1.2x10-5

1.4x10-5

1.6x10-5

Present study, TM11 mode Present study, TE11 mode Present study, TM12 mode

OFHC Cu from Simon et al. 1992 / NIST CMPD 2010 +/- 1.4 × 10-7 K-1 standard deviation of Simon et al. 1992

L (K

-1)

Temperature (K)

Copper thermal expansion coefficient• Experimental data

from 3 microwave modes• Good consistency

• Literature curve – no free parameters!• Simon et al., NIST

Monograph 177, 1992• NIST Cryogenic

Materials Properties Database (2010 revision)

• Excellent agreement with literature values over entire temperature range

9

Page 10: Low-temperature primary thermometry development at NRC

Thermal expansion coefficient αL

0 50 100 150 200 250 300

-2.0x10-7

-1.0x10-7

0.0

1.0x10-7

2.0x10-7

Present study, TM11 mode Present study, TE11 mode Present study, TM12 mode +/- 1.4 × 10-7 K-1 standard deviation of Simon et al. 1992

L, p

rese

nt s

tudy

- L,

lite

ratu

re(K

-1)

Temperature (K)

Copper thermal expansion coefficientwith Simon et al. 1992 / NIST CMPD 2010 curve subtracted

• Present data is within 1 st. dev. of literature curve at all temperatures measured

10

Page 11: Low-temperature primary thermometry development at NRC

Conclusions & future directions

Conclusions

• International Temperature Scale of 1990 deviates from thermodynamic temperature• More measurements needed to resolve issues before replacement scale created• NRC developing microwave RIGT for Canadian thermodynamic temperature measurement capability

• Microwave resonances measured in quasi-spherical copper resonator• Vacuum, 5 K – 300 K

• Comparison to literature properties of copper measured with other methods• Excellent agreement over wide temperature range• Increased confidence in our microwave implementation

Next steps

• Measure triaxial ellipsoid resonator• Better shape, reduced background effects

• Gas in resonator• Refractive Index Gas Thermometry

11

Page 12: Low-temperature primary thermometry development at NRC

We’re looking for a few good physicists: do you have what it takes?

THE PROJECT• Electrical resistivity and Seebeck voltage of platinum-group metals

(and other metals and alloys) – considerable interest to thermometry• Solid-state theory and experimental measurements to understand the

temperature dependencies of these properties• Electronic band structure, electron-phonon scattering, electron-electron (s-d)

scattering, oxidation, recrystallization, and scattering from vacancies and dislocations

• Suitability of various phase transformations as reference temperatures• Typically liquid/solid and solid/liquid transformations of pure elements or

eutectics• Various metal-carbon eutectics and peritectics are of current interest at high

temperatures

KEY SPECIFICATIONS• Ph.D. in Physics (experimental solid state / condensed matter physics preferred)

• Ability to design, construct, and operate experimental equipment with a minimum of technical assistance

• Innovative “hands on” approach towards the solution and attainment of high accuracy in a variety of measurement problems

• Attention to detail commensurate with the operation of a primary standards facility

• Ability to work effectively within a small group devoted to the research, development, and dissemination of temperature standards

Get in touch for more information: [email protected]

Page 13: Low-temperature primary thermometry development at NRC

Thank youDr. Patrick RourkeMeasurement Science and Standards