Lowering the Dose: mA and kVp

Marilyn J. Siegel, M.D

Mallinckrodt Institute of Radiology

Washington University Medical Center

St. Louis, MO

Page 2: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Disclosure of Commercial Interest

I have a financial relationship with a

commercial organization that may have a direct

or indirect interest in the content as follows:

• Siemens Medical Solutions: consultant,

speakers bureau

Page 3: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Objective

• Describe role of mA and kVp in

lowering CT in an adult population

Page 4: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Major CT Parameters Affecting Dose

• Tube current (mA)

• Tube voltage (kV)

• Pitch

• Collimation

Page 5: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Effect of CT Adaptations on Dose

• 50% decrease in mA = 50% dose decrease

• 80 kVp = 30% to 50% dose decrease

• 50% increase in pitch = 50% dose decrease

• Thicker slices = dose decrease

Page 6: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

mA and kV

• Tube milliamperage (mA)

– rate of x-ray production

• Kilovoltage (kV)

–number of x-rays produced

Page 7: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

mA

Tube Current Reduction

Page 8: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Best Approach for mA Reduction

• Automated technology introduced in 1990s

• Superior to technique charts (no guess work)

• Exceptions: screening chest CT

• Other exception is head CT where beam

attenuation comes from bone formation and

this process is age dependent so mA is age

dependent-use fixed tube current

Graser AJR 2006; 659

Herzog et al. AJR 2008, 1232

Mulkens Radiology 2005; 213

Page 9: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Automated mA Reduction

• mA is adapted to body parts, not patient weight,

thinner parts need less radiation

• mA based on attenuation data from scout image

• Dose modulation done in x, y, z planes

Tube

current

Z-position

Page 10: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Ex: shoulders to pelvis Ref value 150 mA

Page 11: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Automated mA Modulation Mean Dose Reduction

• Stone disease 64%

– Mulkens AJR 2007; 188: 553

• Cardiac CT 60%

– Herzog, AJR 2008; 190:1232

• Colonography 35%

– Graser, AJR 2006; 187:695

Page 12: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Lung Cancer Screening

• Fixed tube current

• mA of 20 to 40 rather than standard 150

• Low attenuation background of air-filled

lungs offers a high contrast setting which

can be used favorably to assess lung

abnormalities

Page 13: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Older woman breast cancer low dose vs. standard CT

Similar image quality

150 mA 20 mA

Page 14: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

National Lung Cancer Screening Trial

• 26,724 participants

• Scanned with 20-40 mA

• Average effective dose

–2 mSv for low-dose CT vs.

–7-8 mSv for standard chest CT

• Dose >> females than males (2.4/1.6)

Larke et al. AJR Nov 2011; 197:1165-1169

Page 15: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

A Tale of Two Techniques

kV Modulation

Page 16: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Major Untapped Potential has been kV

?

Distribution of scans in clinical practice

Page 17: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Why use low kV---Rationale

• Lowers radiation dose

• Improves contrast

• Greatest benefit in contrast CT exams

Siegel MJ Radiology 2004; 233:515

Funama, et al., Radiology 2005

Nakayama, et al., Radiology 2005

Huda, et al., Med Phys 2004

Nakayama, et al. AJR 2006 80kV 120kV

Page 18: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Low kV-Effect on Dose: Phantom Study

Siegel MJ: Radiology 2004; 233: 515

0

5

10

15

20

25

30

35

40

45

0 4 8 12 16 20 24 28 32 36

Phantom diameter, cm

Do

se,

mG

y

140 kV

120 kV

100 kV

80 kV

•Reducing kV reduces dose

Page 19: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

0

50

100

150

200

250

300

350

400

0 4 8 12 16 20 24 28 32 36

Phantom diameter, cm

Iod

ine

co

ntr

as

t(H

U)

80 kV

100 kV

120 kV

140 kV

Low kVp: Effect on Contrast

•Reducing kV increases Iodine contrast (HU)

HU

Siegel MJ: Radiology 2004; 233: 515

Page 20: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Low kVp

• Decreases dose & increases contrast

• Rationale: K-edge of iodine 32 keV

• Mean photon energy

– 80 kVp 44 keV

–100 kVp 52 keV

–120 kVp 57 keV

–140 kVp 62 keV

Huda W, et al. Radiology 2000; 217:430

Page 21: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Low kVp: Effect on Image Noise

0

10

20

30

40

50

60

0 4 8 12 16 20 24 28 32 36

Phantom diameter, cm

No

ise

(HU

) 80 kV

100 kV

120 kV

140 kV

•Reducing kV increases image noise in large phantoms

Noise

Page 22: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

kV Modulation-Key Point

• The use of lower-kV is highly dependent

on patient size and diagnostic task

• Increases in contrast are seen ONLY

with high atomic number substances

–Iodine

–NOT water, soft tissue, calcium

• Benefit greatest in children and small

sized adults and contrast examinations

Page 23: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Approaches for kV Reduction?

• Conventionally done using technique charts

• Recently automatic selection tools have

become available (2011)

• Most experience with kV reduction is in PE

and cardiac CTA

Page 24: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Low kV Coronary CTA

• 100 patients (≤ 85 kg)

–Dual Source 64 CT

–Retrospective gating

–120 kVp / 330 mAs: 12 mSv

–100 kVp / 330 mAs: < 8 mSv

• 39% decrease in radiation exposure

Pflederer T, et al. AJR 2009; 192:1045–1050

Page 25: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Low kVp and High Pitch CT

• Dose < 1 mSv

• Total scan time 0.25-0.27s

100 kV and 0.8 mSv

Achenbach et al Eur Heart J. 2010;31:340-6

Leschka et al. Eur Radiol. 2009;19:2896-903

Page 26: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Low kV Pulmonary Embolism CTA

• Park et al. (Korean J Radiol 2009; 10:235)

–dose reduction of 24%

–10.1 to 7.8 mSv (120 to 100 kV)

–Smaller patients (BMI < 25)

• Matsuoka et al. (AJR 2009; 1651-1656)

–40 to 50% dose reduction (120 to 100 kV)

Page 27: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Low kV—Pulmonary Embolism CTA

• Lower dose, better contrast

• Mean enhancement > with 100 kVp (492 vs. 345 HU)

• Sensitivity 100% for detecting PE

120 kV 80 kV

Szucs-Farkas AJR 2011: 197:852

Page 28: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Automated kV Selection Technology

• A tool which automatically adjusts kV for body size

determined from topogram and the exam type

• Up to 60% dose reduction

Page 29: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Auto kV User Interface

• With auto kV off: choose exam type as you would routinely do, screen shows mA, kV & dose -

• Turn auto kV on: take scout CT-program selects best kV to lower dose and maintain image quality

Small adult

Contrast

chest CT

Page 30: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Auto kV

• Used in conjunction with auto mA

• kV decreases AND mA increases

• But overall dose is lower than with fixed kV

Page 31: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Auto kV contrast abdomen, pelvis

• 15 year old boy

• REF: 120kV / 75mA, CTDIvol = 5.77 mGy

• 80 kV / 177mA CTDIvol = 3.2 mGy

Page 32: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Adults--Automated kV 2011

Automated kV selection based on the attenuation

profile of the topogram is feasible, provides a

diagnostic image quality for body CTA, and reduces

overall radiation dose by 25% as compared with a

standard protocol with 120 kV

Investigative Radiology 2011

Page 33: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Auto kV vs. Image Quality

Investigative Radiology 2011

Page 34: Lowering the Dose: mA and kVp - scbtmr.org · Major CT Parameters Affecting Dose

Summary

• Beautiful pictures do NOT need to come

at the cost of higher radiation dose

• Adjusting mA and kV can dial down the

dose and drive up imaging quality-ALARA