maintenance of capillary gc columns - young incoating efficiency: retention index: peak height...
TRANSCRIPT
1
Young In Scientific Co., Ltd.
46 - 1
Lee, Do-Woo
Maintenance of Capillary GC ColumnsMaintenance of Capillary GC Columns
or...."It's not what your column can do for you, but what you can do for your column"
Young In Scientific Co., Ltd.
46 - 2
Objectives
• To know correct column installation procedure
• To know why column performance decreases
2
Young In Scientific Co., Ltd.
46 - 3
Contents
1. Column Installation Procedure
2. To know correct column installation procedure
3. Column Storage
4. Summary
Young In Scientific Co., Ltd.
46 - 4
Column Installation Procedure
1. Install the column
2. Leak and installation check
3. Column conditioning
4. Setting linear velocity or flow rate
5. Bleed profile
6. Test mix
"Getting off to a good start"
3
Young In Scientific Co., Ltd.
46 - 5
Cutting The Column
Gently scribe through the polyimide coating.Do not attempt to cut the glass.
Recommended tools:Diamond or carbide tipped pencil; or sapphire cleaving tool, ceramic wafer Ocular
Do not use:Scissors, file, etc.
Install The Column
Young In Scientific Co., Ltd.
46 - 6
Examples of Column Cuts
Good
Bad
4
Young In Scientific Co., Ltd.
46 - 7
Column Installation
White out Septa
Measuring the right distance
Young In Scientific Co., Ltd.
46 - 8
Column Installation
How tight is tight?
5
Young In Scientific Co., Ltd.
46 - 9
Overtightened Ferrule
Young In Scientific Co., Ltd.
46 - 10
2. Leak and Installation Check
Electronic leak detector
IPA/Water
Inject a non-retained peak
Leak Check
DO NOT USE SNOOP
6
Young In Scientific Co., Ltd.
46 - 11
Leak and Installation Check
The peak should be sharp and symmetrical
Detector Compound
FID Methane or Butane
ECD MeCl2 (headspace or diluted)NPD CH3CN (headspace or diluted)TCD AirMS Air or Butane
Inject a non-retained compound vs DB-1
Young In Scientific Co., Ltd.
46 - 12
Non-Retained Peak Shapes
Good Installation Improper Installation orInjector Leak
Check for: Injector or septum leakToo low of a split ratioLiner problem (broken, leaking, misplaced) Column position in injector and detector
7
Young In Scientific Co., Ltd.
46 - 13
3. Column Conditioning
System must be leak free before conditioning column
Heat the column to the lower of:Isothermal maximum temperatureOR20° to 30°C above highest operation temperatureTemperature programming is not necessary
Stop conditioning when the stable baseline is obtained:1 to 2 hours in most cases
Young In Scientific Co., Ltd.
46 - 14
Calculating Linear VelocityInject a non-retained compound and obtain the retention time:
µ is dependent on column temperature
µ=toL µ = Average linear velocity (cm/sec)
L = Column length (cm)to = Retention time (sec)
He 35-40 cm/secH2 45-60 cm/sec
4. Setting Linear Velocity or Flow Rate
8
Young In Scientific Co., Ltd.
46 - 15
F is dependent on column temperatureMeasuring flow with a flow meter is often inaccurate
F = Flow rate (mL/min)r = Column radius (cm)L = Column length (cm)to= Retention time (sec)
=F r2Lto
Inject a non-retained compound and obtain the retention time:
4. Setting Linear Velocity or Flow Rate
Calculating Flow Rate
Young In Scientific Co., Ltd.
46 - 16
Generating a Bleed Profile
*DB-1 30m x .32mm x .25µmTemperature program
40°C, hold 1 min20°/min to 320°Chold 10 min.
Temperature program the column without an injection*
Time (min.)0 5 10 15 20 25
6000
7000
8000
9000
1.0e4
1.1e4
1.2e4
1.3e4
5. Bleed Profile
9
Young In Scientific Co., Ltd.
46 - 17
6. Test Mixes
Used to determine how "good" the column is
#1
Young In Scientific Co., Ltd.
46 - 18
Column Performance Summary
THEORETICAL PLATES/METER: MIN SPEC ACTUAL
COATING EFFICIENCY:
RETENTION INDEX:
PEAK HEIGHT RATIO:
MIN SPEC MAX SPEC ACTUAL
PENTADECANE
PENTADECANE
1-UNDECANOL
ACENAPHTHYLENE
4-CHLOROPHENOL/METHYL NONANOATE
4-PROPYLANILINE/METHYL NONANOATE
0.83
1.14
1371.04 1372.04 1371.43
1459.34 1460.34 1459.53
3900 4389
90.0 95.5
COMPOUNDIDENTIFICATION
RETENTION TIME(Tr )
PARTITION RATIO
(k)
PEAK WIDTH(W 1/2)
1,6-HEXANEDIOL
4-CHLOROPHENOL
METHYL NONANOATE
4-PROPYLANILINE
TRIDECANE
1-UNDECANOL
ACENAPHTHYLENE
PENTADECANE
2.51
2.95
3.21
3.81
4.20
5.52
8.00
9.58
0.9
1.3
1.5
1.9
2.2
3.3
5.2
6.4
0.019
0.022
0.022
0.026
0.027
0.036
0.053
0.062
PART NO:COLUMN I.D. NO.:LIQUID PHASE:FILM THICKNESS:COLUMN DIMENSIONS:
m X mmTEMPERATURE LIMITS:
C TO C
12250323303121
DB-50.25 µm
30 0.252
-60° 325° ( C PROGRAM)350°
10
Young In Scientific Co., Ltd.
46 - 19
Chromatographic Performance
RETENTION TIME (MIN)
Max.Bleed10.0 pA325
9.0 pA
135
0 5
TEST TEMPERATURE:CARRIER GAS: (H2) 38.7cm/sec, 1.2ml/minINJECTION:ANALYST ERIK
135 C
SPLIT
Young In Scientific Co., Ltd.
46 - 20
Test Mixture Components
Compounds Purpose
Hydrocarbons EfficiencyRetention
Alcohols ActivityFAME’s, PAH’s RetentionAcids Acidic CharacterBases Basic Character
11
Young In Scientific Co., Ltd.
46 - 21
Own Test Mixture
• More specific
• Selective detectors
• Actual concentrations
• No conditions or instrument changes
Young In Scientific Co., Ltd.
46 - 22
An Ounce of Prevention......
12
Young In Scientific Co., Ltd.
46 - 23
Common Causes of Column Performance Degradation
1. Physical damage to the polyimide coating
2. Thermal damage
3. Oxidation (O2 damage)
4. Chemical damage by samples
5. Contamination
Young In Scientific Co., Ltd.
46 - 24
1. Physical Damage to The Polyimide Coating
• Bending radius decreases with increasing tubing diameter
• Avoid scratches and abrasions
• Immediate breakage does not always occur upon physical damage
13
Young In Scientific Co., Ltd.
46 - 25
2. Thermal Damage
Degradation of the stationary phase is increased at higher temperatures. Breakage along the polymer backbone.
Dimethylpolysiloxane
CH3 CH3CH3
CH3 CH3CH3
O
Si SiSi
OOO
Young In Scientific Co., Ltd.
46 - 26
Thermal Damage
Disconnect column from detector
“Bake out” overnight at isothermal limit
Remove 10-15 cm from column end
What To Do If It Happens
14
Young In Scientific Co., Ltd.
46 - 27
Thermal Damage
Rapid degradation of the stationary phase caused by excessively high temperatures
Isothermal limit = Indefinite time
Programmed limit = 5-10 minutes
Temporary "column failure" below lower temperature limit
Young In Scientific Co., Ltd.
46 - 28
3. Oxidation (O2 Damage)
Oxygen in the carrier gas rapidly degrades the stationary phase.The damage is accelerated at higher temperatures. Damage along the polymer backbone is irreversible.
O2
CH3 CH3CH3
CH3 CH3CH3
O
Si SiSi
OOO
Dimethylpolysiloxane
15
Young In Scientific Co., Ltd.
46 - 29
Oxygen Damage
• Rapid damage to the column
• Usually results in irreversible column damage
What To Do If It Happens
Young In Scientific Co., Ltd.
46 - 30
How to Prevent Column Damage by Oxygen
• High quality carrier gas (4 nine's or greater)
• Leak free injector and carrier linesChange septaMaintain gas regulator fittings
• Appropriate impurity traps
16
Young In Scientific Co., Ltd.
46 - 31
Configurations for Carrier Gas Purifiers
OVEN TEMP 100
IndicatingMoisture Trap
High CapacityOxygen Trap
Indicating Oxygen Trap
GC
Make sure the traps are in the upright position!
Young In Scientific Co., Ltd.
46 - 32
4. Chemical Damage by Sample
Bonded and cross-linked columns have excellent chemical resistance except for inorganic acids and bases
Chemical damage will be evident by excessive bleed, lack of inertness or loss of resolution/retention lack of inertness or loss of resolution/retention
HCl NH3 KOH NaOH
H2SO4 H3PO4 HF etc.
17
Young In Scientific Co., Ltd.
46 - 33
Chemical Damage
• Remove 1/2 - 1 meter from the front of the columns
• Severe cases may require removal of up to 5 meters
What To Do If It Happens
Young In Scientific Co., Ltd.
46 - 34
5. Contamination
Fouling of GC and column by contaminants
Mimics nearly every chromatographic problems
18
Young In Scientific Co., Ltd.
46 - 35
Symptoms of Contamination
• Poor peak shape
• Loss of separation (resolution)
• Changes in retention
• Reduced peak size
• Baseline disturbances (semi-volatiles only)
Young In Scientific Co., Ltd.
46 - 36
Typical Samples Containing Large Amount of Residues
Biological (Blood, Urine, Tissue, Plants)
Soils Foods
Waste Water Sludge
All samples contain residues!!
19
Young In Scientific Co., Ltd.
46 - 37
Other Sources of Contamination
Septum and ferrule particles
Gas and trap impurities
Unknown sources (vials, syringes,etc.)
Young In Scientific Co., Ltd.
46 - 38
Non-Volatile Residues
Any portion of the sample that does not elute from the column or remains in the injector.
Semi-Volatile Residues
Any portion of the sample that elutes from the column after the current chromatographic run.
Other Sources of Contamination
20
Young In Scientific Co., Ltd.
46 - 39
Methods to Minimize Non-Volatile Residue Problems
Sample cleanup
Packed injection port liners
Guard columns
Young In Scientific Co., Ltd.
46 - 40
Guard Column
The guard column is 0.5 - 10 meters of deactivated fused silica tubing with the same diameter as the analytical column. It is connected with a zero dead volume union.
INJECTORDETECTOR
21
Young In Scientific Co., Ltd.
46 - 41
Non-Volatile Contamination
Do not “bake out” the column
Front End Maintenanceclean or change the injector linerclean the injectorcut off 1/2 -1 meter of the front of the column
Turn the column around
Solvent rinse the column
Cut the column in half
What To Do If It Happens
Young In Scientific Co., Ltd.
46 - 42
Rinse Kit
1/16" flexible teflon line to regulated pressure source
Beaker for solvent collection
Capillary column Special connectorand ferrule
Flexible teflontubing
Special adapter
CapVial
Capillary column
22
Young In Scientific Co., Ltd.
46 - 43
Semi-Volatile Contamination
“bake out” the columnLimit to 1-2 hoursMay polymerize some contaminationReduces column life
Solvent rinse the column
What To Do If It Happens
Young In Scientific Co., Ltd.
46 - 44
Column Storage
• Place septa over the ends
• Return to column box
23
Young In Scientific Co., Ltd.
46 - 45
Always Remember to:Always Remember to:
Start with a good installation
Maintain an oxygen free system
Avoid physical, thermal, and chemical damage
Take steps to prevent contamination
Summary
Young In Scientific Co., Ltd.
46 - 46
Thank you very much for your attention !