47
ORIGINAL ARTICLEMalaysian Journal of Medical Sciences, Vol. 7, No. 1, January 2000 (47-53)
ABSORBED DOSE TO WATER DETERMINATION USING IAEA,HPA, NACP, AAPM, NCRP AND ICRU PROTOCOLS FOR 1.25 MEVGAMMA RAY 6 MV AND 10 MV X-RAYS: AN INTERCOMPARISON
OF RESULTS WHEN IAEA WAS TAKEN AS A STANDARDPROTOCOL
M T Dolah, S B Samat and T Kadni*
School of Applied Physics, Universiti Kebangsaan Malaysia43600 UKM BANGI, Selangor, Malaysia
*Secondary Standard Dosimetry Laboratory MalaysiaMalaysian Institute for Nuclear Technology Research
43000 KAJANG, Selangor, Malaysia
Absorbed dose to water was measured with ionisation chambers NE 2561 (#267),NE 2581 (#334), NE 2571 (#1028), using the IAEA standard water phantom. Theionisation chamber was inserted in the water phantom at a reference depthdependent on the type of the radiation quality used. Three radiation qualities wereused namely 1.25 MeV gamma ray, 6 MV x-rays and 10 MV x-rays. The values ofthe absorbed dose to water were determined by the NK- and NX- based methods, i.ewith the use of IAEA, HPA, NACP, AAPM, NCRP and ICRU protocols. The aimof this study was to make an intercomparison of the results, by taking the IAEAprotocol as a standard. The largest deviation contributed by any of these protocolswas recorded for each quality. It was found that AAPM, NCRP and ICRU protocolscontributed 0.94% for 1.25 MeV gamma ray, NACP contributed 2.12% for the 6MV x-rays, and NACP contributed 2.35% for 10 MV x-rays. Since the acceptablelimit of deviation set by the IAEA for this absorbed dose work is ± 3%, it is clearthat the overall deviations obtained were all satisfactory.
Key words : gestational Diabetes mellitus, diagnosis, management
Introduction
At present there are many protocols that arebeing used in variaus countries to determine theabsorbed dose to water. In 1987, InternationalAtomic Energy Agency (IAEA) recommended aprotocol for absorbed dose to water determinationfor high energy photon dosimetry (1). Prior to thisrecommendation, several protocols have beenrecommended, for example; Hospital Physicists’Association in 1983 with its HPA protocol (2),American Association of Physicists in Medicine in1983 with AAPM protocol (3), National Council on
Radiation Protection and Measurement in 1981 withits NCRP protocol (4), Nordic Association ofClinical Physics in 1980 with its NACP protocol(5) and International Commission on RadiationUnits and Measurement in 1973 with its ICRUprotocol (6). Table 1 summarises these six protocolstogether with their respective formulae forcalculating the absorbed dose to water. The,meanings of symbols that are used in the formulaeare given in the final part of this paper.
For the purpose of dosimetry accuracy inradiotherapy, these six protocols should in practiceyield a single value in the absorbed dose to water,
48
Table 1. Absorbed dose to water DW formulae according to variaus protocols
Table 2. The radiotherapy machices that provided the three radiation qualities
Table 3. Calibration factor of the three ionisation chambers
Ionisation chamber Calibration faktor
Model Serial number NK NX
NE 2561 267 9.353 mGy/sd(a 1.064 R/sd(u
NE 2581 334 52.94 mGy/sd(b 6.022 R/nC(d
NE 2571 1028 41.34 mGy/sd(c 4.703 R/nC(d
uCertificate values as reference (8).bCertificate values as reference (9).cCertificate values as reference (10).dDerived from the NK Value.
for a given radiation quality and reference condition.The study acms to confirm whether the differentprotocols yield almost identical absorbed dose towater result, for a given radiation quality andexperiment set-up.
The goalo of this study are: (1) to determine
the absorbed dose to water using IAEA, HPA,NACP, AAPM, NCRP and ICRU protocols for 1.25MeV gamma ray, 6 MV x-rays and 10 MV x-rays;and (2) to compare the results obtained by the HPA,NACP, AAPM, NCRP and ICRU with the mostcommonly used IAEA protocol (7).
M T Dolah et al
Protocols Equation no. this work ReferencesNameIAEA
Dw formulasMuNk (1-g) Katt Km (Sw air) Pu Pdis
HPA 1.139 M NK Cλ
NACP MuNK (1-g) Katt Km (Sw air) Pu
AAPM M NX Kq (a (Li p)med.gas Pton Prepl Pwall
NCRP M Nx Cλ
ICRU
123
456
[1][2][5]
[3][4][6]M Nx Cλ
akq =ion wall wall
α × L/pwall
airμen/p
wall
airμen/p
air
cap+ (1-α)× L/p
cap
air
Radiation quality MachineNo. Energy Name Located atModel
1
2
3
1.25 MeV
6 MV x-ray
10 MV x-ray
Co-60 teletherapy unit
Medical linear accelerator (linac)
Linac
SSDL Malaysia
Radiotherapy and Oncology Unit, HUKM
Radiotherapy and Oncology Unit, HUKM
Eldorado-8 (#104)
Mevatron KD2
Mevatron KD2
× × × ×
××
49
Material and Methods
2.1 The dependence of the three elements(calibration factors, dosimeter readings and theinteraction coefficients) on radiation quality andionisation chamber in the protocol formulae.
The six formulae (given by the six protocols)for the determination of the absorbed dose to waterin Table 1, mainly comprise of three elementsnamely:
Table 4. Dosimeter reading of the three ionisation chambers, in three radiation qualities
Table 5. Values of interaction coefficients as a function of radiation quality, ionisation chamber,and protocol.
(i) the ionisation chamber calibration factors (NKor NX), which depend on the type of chamberused,
(ii) the dosimeter reading (M or M u), whichdepends on both the chamber and the radiationquality, and
(iii) the interaction coefficients, which depend onthe ionisation chamber, the radiation quality andthe protocol used.
The present work investigated thedetermination of the absorbed dose to water in three
ABSORBED DOSE TO WATER DETERMINATION USING IAEA, HPA, NACP, AAPM, NCRP AND ICRU PROTOCOLS FOR 1.25 MEV GAMMA RAY 6 MV AND 10 MV X-RAYS: AN INTERCOMPARISON OF RESULTS WHEN IAEA WAS TAKEN AS A STANDARD PROTOCOL
Dosimeter models Dosimeter average reading M(a or Mu(b
Ionisation chanber Electrometer 1.25 MeV gamma ray 6 MV x-ray 10 MV x-rayNE 2561 (#267)
NE 2581 (#334)
NE 2571 (#1028)
NE 2560 (n151)
PTW-Unidos 10005 (n 50013)
PTW-Unidos 10005 (n 50013)
28.979 ± 0.040sd/min
4.957 ± 0.002nC/min
6.109 ± 0.004nC/min
85.933 ± 0.047sd/100mu
35.336 ± 0.042nC/230mu
15.701 ± 0.000nC/80mu
201.347 ± 0.12sd/300mu
-
-
(a For HPA, AAPM, NCRP and ICRU protocols(b For IAEA and NACP protocols
1.25 MeV gamma rayNE 2561 NE 2581 NE 2571 NE 2571
IAEAIAEA
HPA
NACP
AAPM
NCRP
ICRU
(1-g)Katt
Km
(Sw,air)Pu
Pdis
Cλ
(1-g)Katt
Km
Pu
(Sw,air)u
(1-g)(L/p)med,gas
Pion
Prepl
Pwall
Cλ
F
0.9970.9950.9841.1330.9930.985
0.951
0.9970.990.9910.971.15
0.0085(a
1.1341.0000.99530.9965
0.95md/R
0.95md/R
0.9970.9950.991.1331.00750.987
0.951
0.9970.990.9630.991.15
0.00851.1341.0000.99581.0099
0.95md/R
0.95md/R
0.9970.9940.9911.1330.9930.987
0.951
0.9970.990.9910.971.15
0.0086(b
1.1341.0000.99580.9970
0.95md/R
0.95md/R
0.9970.9950.9841.1190.99460.985
0.95
0.9970.990.9910.981.14
0.0085(a
1.1271.0000.99550.9965
0.94md/R
0.94md/R
NE 2581
0.9970.9750.991.1191.0060.987
0.95
0.9970.990.9630.991.14
0.0085(a
1.1271.0000.99611.0050
0.94md/R
0.94md/R
NE 2571
0.9970.9740.9911.1190.99460.987
0.95
0.9970.990.9610.981.14
0.0086(b
1.1271.0000.99610.9960
0.94md/R
0.94md/R
NE 2551
0.9970.9950.9841.1050.99660.985
0.943
0.9970.990.9610.9851.125
0.0085(a
1.1171.0000.99250.9976
0.93md/R
0.93md/R
NE 2581
0.9970.9950.991.1051.24070.987
0.943
0.9970.990.9630.991.125
0.00851.1171.0000.99362.0047
0.93md/R
0.93md/R
NE 2571
0.9970.9940.9911.1050.99660.987
0.943
0.9970.990.9910.9851.125
0.0086(a
1.1171.0000.99360.9976
0.93md/R
0.93md/R
Reference (12)
50
radiation qualities using three different ionisationchambers.
Table 2 shows types of machines that providedthese radiation qualities and the theirlocation.
Table 3 shows the three chambers used togetherwith their calibration factors.
Table 4 shows the dosimeter readings obtainedfrom the three chambers in the threeradiation qualities.
The types of electrometers used with theseionisation chambers to yield the readings (charge)are given in this table. To obtain these readings, forexample in the 1.25 MeV Co-60 beam, experimentalset-up as shown in Fig 1 was used. Similar set upwas used for the x-rays beam. Source to chamberdistance (SCD) was set at 110 cm for 10 MV x-ray,and 105 cm for other radiations.
The final part of the absorbed dose to waterformula is the interaction coefficients. Table 5 showsthese coefficients with their values (1, 2, 3, 4, 5, 6,
11, 12). It can be seen that these values vary withthe type of radiation quality, the type of ionisationchamber used for measurement and the protocolused.
2.2 Numerical examples for calculating theabsorbed dose to water and the percentagedeviation in absorbed dose to water
2.2.1 The absorbed dose to water
Example 1. Suppose that the absorbed doseto water in the 1.25 MeV gamma ray beam is to bedetermined using NE 2561 ionisation chamber basedon the IAEA protocol. By looking at Table 1, eqn. 1is the formula that should be used. For the presentwork, the values for eqn. 1 are: NK = 9.353 mGy/sd(Table 3); Mu = 28.979 ± 0.040 sd/min (Table 4);(1-g) = 0.997, Katt = 0.995, Km = 0.984, Sw,air = 1.133,Pu = 0.933 and Pdis = 0.985 (Table 5). Uponcalculating, we obtained DW = 293.24 mGy/min.
Example 2. Suppose that the absorbed doseto water in the 6 MV x-rays beam is to be determined
Table 6. DW and D values for three ionisation chambers, calculated using six protocols, and threeradiation qualities.
Table 7. Comparison of absorbed dose to water let the present study and previous studentsTwo protocolthat are beingcompared
______________HPA, incomparisonwith AIEA
AAPM, incomparisonwith AIEA
Study__________________
Condition
__________________Radiation qualityCo-60, 6 MV and10 MV using NE2561 chamberRadiation qualityCo-60, 6 MV and10 MV using NE2561, NE 2581andNE 2571 chambers
__________________Results ofdeviation (%)
__________________0.11% to 1.42%
0.10% to 1.14%
_______________________Condition
_______________________Same as present work
Different from presentstudy: This other studyuses radiation qualityCo-60, 4 MV and 25 MVusing PTW, Capintec andFarmer chambers
Previous student_________________
Results ofdeviation %
_________________-1.29% to -0.22%
-0.40% to-1.20%
___________Reference
___________13
14
Comments on theresults of presentstudy
_________________It is in a goodagreement
One to onecomparisoncannot bedone asconditions aredifferent
M T Dolah et al
Radiationquality
1.25MeV
gammaray
6 MVx-ray
10 MVx-ray
Ionisationchamber
NE2561
NE 2581NE 25 71
NE 2561NE 2581NE 2571
NE 2561
Units
mGy/min
mGy/minmGy/min
Gy/100muGy/230muGy/80mu
Gy/30mu
293.24*
281.65275.53
0.8602.0010.701
1.994
IAFA
293.57
284.25273.56
0.8702.0240.702
2.002
HPA
295.74
283.98275.56
0.8782.0070.712
2.041
NACP
296.00
286.57275.81
0.8692.0210.709
2.014
AAFM
292.92
283.59272.94
0.8592.0000.694
1.992
ICRU
Dw(Eqns. (1) to (6))
&(Eqn. (7))
292.92
283.59272.94
0.8592.0000.694*
1.992
NCRP
0.11*
-0.14-0.72
1.121.160.26
1.42
HPA
0.85
-0.230.01
2.120.291.66
2.35
NACP
0.94
0.670.10
1.051.001.14
1.00
AAPM
-0.11
-0.37-0.94
-0.070.04-0.92
-0.10
ICRU
-0.11
-0.37-0.94
-0.070.04-0.92
-0.10
NCRP
51
with NE 2571 ionisation chamber using the NCRPprotocol. By looking at Table 1, eqn. 5 is the formulathat should be used. For the present work, the valuesfor eqn. 5 are: NX = 4.703 R/nC (Table 3); Mu =15.701 ± 0.000 nC/80mu (Table 4); Cl = 0.94 rad/R(Table 5). Upon calculating, we obtained DW = 0.694Gy/80mu
Similar methods as shown in examples 1 and2 were used to calculate DW for other radiationqualities using the three chambers for the sixprotocols. The results are shown in Table 6.
2.2.2 The percentage deviation in absorbed doseto water
The percent age deviation in the absorbeddose when we compare with the IAEA protocol is
DW (Other protocol) - DW (IAEA protocol) x100%
DW (IAEA protocol)
Example 3. Suppose the deviation of theabsorbed dose to water in the HPA protocol results(obtained in the determination of absorbed dose towater in 1.25 MeV gamma-ray beam using an NE2561 chamber) is to be calculated. By looking at
Table 6, we have = 100% x (293.57 – 293.24)/293.24 = 0.11 %.
Similar method as shown in example 3 wasused for the other protocols to calculate other forthe three radiation qualities using the three chambers.The results are shown in Table 6.
Result and Discussion
Column 4 of Table 6 shows the values of theabsorbed dose to water determined by the IAEAprotocol for three radiation qualities using the threetypes of ionisation chambers. Columns 5, 6, 7, 8and 9 show the absorbed dose to water calculatedby the HPA, NACP, AAPM, NCRP and ICRUprotocols respectively. Columns 10, 11, 12, 13 and14 show how the other protocols deviate from theIAEA results.
In this study the NCRP and ICRU protocolsyielded results which are in good agreement withthe IAEA protocols, followed by the HPA, AAPMand NACP protocols. From Table 6, it can be seenthat all the deviation values given by these two(NCRP and ICRU) protocols are lower than ±1%.
Table 7 shows the comparison between thepresent study and present studies. For the HPA
Figure 1. Experimental set-up for the absorbed dose to water determination in a 60 Co beam atthe SSDL.
ABSORBED DOSE TO WATER DETERMINATION USING IAEA, HPA, NACP, AAPM, NCRP AND ICRU PROTOCOLS FOR 1.25 MEV GAMMA RAY 6 MV AND 10 MV X-RAYS: AN INTERCOMPARISON OF RESULTS WHEN IAEA WAS TAKEN AS A STANDARD PROTOCOL
52
protocol, a study that was newformed by Kadni [13]shawed a good agreement with the present study.The other study done by Huq and Nath [14] howevercan not be compared to this study as differentconditions were used.
The largest deviation contributed by any ofthese protocols was recorded for each quality. It wasfound that AAPM, NCRP and ICRU contributed0.94% for 1.25 MeV gamma ray, NACP contributed2.12% for the 6 MV x-rays, and NACP contributed2.35% for 10 MV x-rays. Since the acceptable limitof deviations set by the IAEA for this absorbed dosework is ± 3% [15], it is clear that the overalldeviations obtained were all satisfactory.
Conclusions
HPA, NACP, AAPM, NCRP and ICRUprotocols have not yielded significant differencesin absorbed dose to water value when compared withthe recent IAEA protocol. The differences in datafor interaction coefficients have minor influenceson the final results. It can be concluded that, despitethe many differences in the values of these protocols,
the final results were almost identical. Therefore,not suprisingly, the five protocols are still being usedin the western countries. In Malaysia, a preliminarysurvey [16] showed that four out of thirteen healthinstitutions use the HPA protocol while two healthinstitutions use the ICRU and AAPM protocols.
Acknowledgement
We are grateful to the Malaysian Governmentfor the IRPA Grant (Project code: 09-02-02-0040)and to the Universiti Kebangsaan Malaysia for aResearch Grant (Project code: S/11/97).
Correspondence :
Dr. M T DolahSchool of Applied PhysicsUniversiti Kebangsaan Malaysia43600 UKM BANGI, Selangor, Malaysia
References
1. International Atomic Energy Agency. A b s o r b e ddose determination in photon and electron beams - aninternational code of practice. TRS No: 277. Vienna:IAEA, 1987.
2. Hospital Physicists’ Association. Revised code ofpractice for the dosimetry of 2 to 35 MV x-ray, andcaesium-137 and cobalt-60 gamma ray beams. Phys.Med. Biol. 1983; 28(10): 1097-1104.
3. American Association of Physicists in Medicine. AProtocol for the Determination of Absorbed Dose fromHigh Energy Photon and Electron Beams, Task Group21. Med. Phys. 1983; 10(6): 741-771.
4. National Council on Radiation Protection andMeasurement. Dosimetry of x-ray and gamma raybeams for radiation therapy in the energy range 10 keVto 50 MeV. NCRP Report No. 69. Washington, 1981:62-66.
5. Nordic Association of Clinical Physics. Procedures inexternal radiation dosimetry with electron and photonbeams with maximum energies between 1 and 50 MeV.Acta Radiol. Oncology. 1980; 19(55):145-167.
6. International Commission on Radiation Units andMeasurement. Measurement of absorbed dose in aphantom irradiated by a single beam of x or gammarays. ICRU Report 23. Washington, 1973.
7. Metcalfe, P., Kron, T. and Hoban, P. The physics ofradiotherapy x-rays from linear accelerators. Madison:Medical Physics Publishing, 1997:138.
8. National Physical Laboratory. Certificate of calibrationan x-ray exposure meter. Ref. No. D 1910, 14 May1985. Reading: NPL, 1985.
Figure 2. List of symbols and meaning.List of symbols
(l-g) Value for fragtion energy of secondary charges particles thatis lost to bremsstrahlungFraction of ionization due to electron from buildup cap
ion Ion collection efficiencywall Wall corection factor
wall Absorbed dose/collision fraction of kermaC Exposure to absorbed dose conversion factorD Absorbed dose to waterF Coefficient relating the exposure in roentgens to absorbed
dose in water expressed in cGyk Charge per unit mass of air per unit exposureK Attenuation factorK Non-airk Chamber correction factor(L/ ) Stopping power ratio-medium to gasM or M Electrometer readingN Air kerma calibration faktorN Exposure calibration factorPdis Correction for displacement correctionPrepl Replacement correctionPu Perturbation factor
att
w
m
q
u
K
X
M T Dolah et al
53
9. International Atomic Energy Agency. IAEA/WHOnetwork of secondary standard dosimetry laboratoriesintercomparison of calibration factors. Ref. No. MAL/INTER/97. Vienna: IAEA, 1997.
10. International Atomic Energy Agency. IAEA/WHOnetwork of secondary standard dosimetry laboratoriesintercomparison of calibration factors. Ref. No. MAL/9801. Vienna: IAEA, 1999.
11. International Atomic Energy Agency. The use ofparallel plate ionization chamber in high energyelectron and photon beams. TRS No: 381. Vienna:IAEA, 1997.
12. Dolah, M.T. Calculation of interaction coefficient kqfor NE 2561 and NE 2571 for radiation qualities of1.25 MeV gamma ray, 6 MV x-ray and 10 MV x-ray.Unpublished report, 2000.
13. Kadni, T. Determination of absorbed dose to water inhigh energy photon beam by application of two
different dosimetry codes of practice. Presented atRegional Seminar on Radiotherapy Dosimetry:Radiaton dose in radiotherapy from prescription todelivery. Thailand, 1995.
14. Huq, M.S. and Nath, R. Comparison of IAEA 1987and AAPM 1983 protocols for dosimetry calibrationof radiotherapy beams. Med. Phys. 1991; 18(1): 26-35.
15. International Atomic Energy Agency. Radiotion dosein radiotherapy from prescription to delivery, IAEA -TECDOC - 896, Dosimetry and Medical RadiotionPhysics Section, and Radiation Safety Section) Vienna:IAEA, 1996: 281
16. Dolah, M.T. A survey of protocols used in thedetermination of absorbed dose in high energy photonbeams in Malaysian health institutions. Unpublishedreport, 2000.
ABSORBED DOSE TO WATER DETERMINATION USING IAEA, HPA, NACP, AAPM, NCRP AND ICRU PROTOCOLS FOR 1.25 MEV GAMMA RAY 6 MV AND 10 MV X-RAYS: AN INTERCOMPARISON OF RESULTS WHEN IAEA WAS TAKEN AS A STANDARD PROTOCOL