mamografo philips

Upload: william-m-camacho

Post on 03-Jun-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/12/2019 Mamografo Philips

    1/12

    Womens healthdeserves the best

    Philips MammoDiagnost DR Specifi cation

    Superb image quality as well as efficient, convenient

    workflow characterize this system. The intuitive,

    streamlined workflow at the Eleva workspot and

    the systems ergonomic design make working with it

    comfortable and efficient for the radiographer, and

    ensures a quick and easy examination for the patient.

    The excellent image quality resulting from our UNIQUE

    image processing software and the high image resolution

    of the MammoDiagnost DR provide a safe diagnosis

    for both radiologist and the patient.

  • 8/12/2019 Mamografo Philips

    2/12

    2

  • 8/12/2019 Mamografo Philips

    3/12

    3

    Content

    04 Console

    06 Gantry

    08 Detector

    09 Tube

    10 Generator

    10 Ambient conditions

    10 Conformity

    11 Dimensions

  • 8/12/2019 Mamografo Philips

    4/12

    44

    Eleva workspot

    Dimensions Height: 2000 mm / 78.7 "

    Width: 750 mm / 29.5 "

    Depth: 550 mm / 21.7 "

    Glass shield 2000 mm x 750 mm / 78.7 " x 29.5 ", meets IEC Protective Shielding requirements

    Monitor 19 inch colour touch screen monitor, adjustable in height and angle

    Software handling Simultaneous handling of acquisition, processing and export of images

    Internal disk space 200 GB, sufficient to store at least 700 examinations

    Automatic deletion Automatic deletion of the oldest images after a configurable t ime

    Worklist Support of DICOM worklist management, worklist filter enables quick

    organization

    Exposure An audible and visual signal on the Eleva workspot indicates exposures

    Image acquisit ion After exposure it takes < 16 seconds until the processed image is displayed

    on the Eleva workspot monitor

    Image processing UNIQUE image processing for excellent image quality and with optimized

    visibility of all clinical relevant details

    Archiving Image is automatically sent to PACS and/or printer as configured. Images

    for presentation and / or images for processing can be archived as configured.

    Alternatively the user can manually send the images to PACS and/or printer

    Export status for archiving

    and printing

    Displayed for each image

    IHE Support of the IHE-Mammography Image Integration Profile

    HIPAA HIPAA security profile conformance including audit trail and time

    synchronization

    Login Individual login for each user visible after login, fast change of the user

    possible

    Security Update of the virus pattern signature files by license

    DICOM grayscale standard Support for export of images

    Eleva workspot

  • 8/12/2019 Mamografo Philips

    5/12

  • 8/12/2019 Mamografo Philips

    6/12

    6

    Gantry

    Dimensions Height: 2030 mm / 79.9 "

    Width: 1570 mm / 61.8 "

    Depth: 1175 mm / 46.3 "

    Weight 400 kg

    SID 65 cm / 25.6 "

    Movements All movements are motor driven

    Speed of vertical movement Configurable between 40 80 mm per/s / 1.6 " 3.2 " per/s

    Rotation of C-arm 180 degree to + 180 degree, isocentric

    Speed of rotation Appr. 20 degree / second

    Height adjustment 650 1500 mm / 25.6 " 59.1 " above the floor to the object

    Wheelchair accessibility Yes

    Location of control panels On both sides of the tube head and on both sides of the compression unit

    Foot switches Mirrowed foot switches for elevation and compression

    Optimized compression Display when the most sensible compression is applied

    Grid Motorized linear grid, ratio 5 : 1, 31 lines/cm, 79 lines/inch

    Magnification unit Magnification factor of 1.5

    Automatic retraction of the grid and

    automatic selection of small focus

    Collimation Automatic collimation according to the paddle size

    Paddles

    standard:

    optional:

    18 x 24, 24 x 30

    18 x 24 flexible, 18 x 24 high edge, 24 x 30 high edge, 24 x 30 flexible,

    axillary paddle, magnification spot paddle (all dimensions in cm)

    Automatic compression Motorized with max. compression force of 200 N

    Manual compression Max. compression force of 200 N

    Max. distance breast support to paddle 320 mm / 12.6 "

    Decompression Automatic decompression after each exposure to 150 mm / 5.9 "

    above object table

    Display Display at the foot of the gantry indicating compression force,

    thickness, angulation and patient name

    Gantry

  • 8/12/2019 Mamografo Philips

    7/12

    7

  • 8/12/2019 Mamografo Philips

    8/12

    8

    Detector

    Detector material Amorphous Selenium

    Field of view 23,9 x 30,5 cm / 0.9 " x 1.2 "

    Pixel size 85 m

    Pixel matrix 3584 x 2816

    Image depth 13 bit

    Refresh time between exposures < 30 seconds

    Patient throughput 15 patients / hour

    Spatial Resolution > 7 lp / mm

    DQE

    DQE @ 1 lp / mm > 0.5

    DQE @ 5.8 lp / mm > 0.15

    MTF > 40% at 5.8 lp / mm

    Ghost image factor < 0.3

    Detector

  • 8/12/2019 Mamografo Philips

    9/12

    9

    Tube

    Anode target / filter combination Mo / Mo, Mo / Rh, W / Rh

    Focus 0,3 x 0,3 and 0,1 x 0,1

    Max. tube current:

    Molybdenum focal spot

    Tungsten focal spot

    large focal spot: 150 mA at 25 kV

    small focal spot: 28 mA at 25 kV

    large focal spot: 188 mA at 25 kV

    small focal spot: 34 mA at 25 kV

    Total heat storage capacity 1.8 MJ

    Anode heat storage capacity 120 kJ

    Anode speed 8800 rpm

    Filter 30 m Molybdenum

    25 m Rhodium

    50 m Rhodium

    Tube

  • 8/12/2019 Mamografo Philips

    10/12

    10

    Generator

    Generator Integrated into the gantry

    mAs range 2 600 mAs

    kV range 23 35 kV in 1 kV increments

    Mains voltage: 1-phase 208 V, 230 V, 240 V, 277 V 10%

    Mains voltage: 2-phase 208 V, 230 V, 240 V, 277 V, 440 V 10%

    Exposure technique AEC with pre-pulse technique or manual

    Dose optimization 4 pre-defined programs available, according to

    compression automatic selection of target / filter

    combination and kV value

    Exposure times 10 ms to 4.0 s for large focus

    60 ms to 6.0 s for small focus

    (tolerance 10%)

    Conformity

    IHE, DICOM, QA EUREF, MQSA

    Ambient conditions

    TemperatureOperation

    Storage

    +18 C to +30 C

    10 C to +55 C

    Rel. humidity

    Operation

    Storage

    20% to 75%

    10% to 95%

    Air pressure

    Operation & Storage 700 hPa to 1060 hPa

    Generator

  • 8/12/2019 Mamografo Philips

    11/12

    11

    836,4 mm / 32.9"

    1220,4mm/

    48"

    1

    500mm/

    59.1

    "

    2310mm/

    90.9

    "

    680 mm / 26.8"

    Dimensions

    1247,7 mm / 49.1"

    1220,4mm/

    48"

    836,4 mm / 32.9"49,5 mm / 1.9" 361,8 mm / 14.2"

    2031mm/

    80"

  • 8/12/2019 Mamografo Philips

    12/12

    Philips Medical Systems is part ofRoyal Philips Electronics

    Interested?

    Would you like to know more about our imaginative

    products? Please do not hesitate to contact us.

    We would be glad to hear from you.

    On the web

    www.medical.philips.com

    Via [email protected]

    By fax

    +31 40 27 64 887

    By mail

    Philips Medical Systems

    Global Information Center

    P.O. Box 1286

    5602 BG Eindhoven

    The Netherlands

    By phone

    Asia

    Tel: +852 2821 5888

    Europe, Middle East, Africa

    Tel: +49 7031 463 2254

    Latin America

    Tel: +55 11 2125 0764

    North America

    Tel: +1 800 229 6417

    2007 Koninklijke Philips Electronics N.V.

    All rights are reserved.

    Philips Medical Systems DMC GmbH reserves the right to make changes in specifications and/or to discontinue any product at any time without notice

    or obligation and will not be liable for any consequences resulting from the use of this publication.

    Printed in The Netherlands.

    4522 962 27161/714 * NOV 2007