manual -ipc 7th semester

Upload: nayyer-shah

Post on 03-Apr-2018

230 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 Manual -IPC 7th Semester

    1/36

    INSTRUMENTATION & PROCESS CONTROL

    LAB MANUAL (7th Semester)

    LAB INCHARGE:

    Prof. Dr. Arshad Chughtai

    FACULTY TEAM:

    Ms. Rabya Aslam

    Institute of Chemical Engineering & Technology

    1

  • 7/28/2019 Manual -IPC 7th Semester

    2/36

    Experiments List:

    A. Calibration of Instruments:

    1. Calibrate the given Bourdon Gauge using the Mercury Filled Manometer. Also find

    out the span, error and accuracy of the bourdon gauge.

    2. Calibrate and report the accuracy of the given Bourdon Gauge using Dead Weight

    Tester.

    3. Calibrate the Given Resistance Temperature Detector using the mercury filled

    Thermometer.

    4. Calibrate the given Thermocouple using Thermometer.

    B. Process Analysis:

    5. Find out the time constant of the given Mercury Filled Thermometer and also find

    the response y(t) of the system when t =, t = 2, t = 3

    6. Find out the time constant of the Liquid Level System

    C. Control Loops:

    7. Report the response and variations in the process variable PV( flow of water)output

    in

    i. Proportional Mode (P- mode) by giving the values of gain as 1.00, 0.6 and

    1.6 while the output of controller set on 30 % i.e. set point = 30 %.

    ii. Proportional Integral mode (PI mode) by giving the values of Reset time, as

    0.1, 0.15 and 0.5 while the output of controller set on 30 %.

    iii. Proportional Integral Derivative mode (PID mode) by giving the values of

    rate minute as 1, 2, 3 while the output of controller set on to 30 %.

    8. Study the response of the process variable (temperature) in the on-off algorithm.

    Also plot the graph between time and temperature during heating and cooling.

    2

  • 7/28/2019 Manual -IPC 7th Semester

    3/36

    Table of Contents:

    Contents Page no.

    1. Calibration of Bourdon gauge using Manometer 4

    2. Calibration of Bourdon gauge using dead weight tester 8

    3. Calibration of RTD using thermometer 12

    4. Calibration of thermocouple using thermometer 15

    5. Time constant of thermometer 18

    6. Time constant of liquid level system 22

    7. Flow control loop 26

    8. Temperature control loop 32

    3

  • 7/28/2019 Manual -IPC 7th Semester

    4/36

    Experiment # 1: Calibration of Bourdon Gauge using

    Manometer

    Objective:To calibrate the given Bourdon gauge using the Mercury Filled Manometer also report

    the accuracy of the Bourdon gauge.

    Apparatus:

    Bourdon gauge

    Mercury filled thermometer

    Air compressor

    Procedure:

    Start the air compressor before performing the experiment.

    Mercury level in the U-shaped manometer is checked.

    The Manometer shows the value of the change in pressure by inches of mercury and Bourdon

    gauge report the value of same pressure in psi or bar.

    The readings of both the manometer i.e., inches of mercury and Bourdon gauge i.e., psi or

    bar are noted after the alteration in the pressure by means of controlling valve.

    The readings of manometer i.e., in inches of mercury are converted into psi by multiplying

    with an appropriate factor.

    The graph between manometer readings at X-axis and the Bourdon gauge reading at Y-axis

    is plotted. A straight line at 45 form origin is drawn. The maximum difference between the

    actual plotted line and the 45 line is the Span error.

    Formulae Used:

    1. Pressure Reading using Manometer

    4

  • 7/28/2019 Manual -IPC 7th Semester

    5/36

    =

    c

    airHg

    g

    hgP

    )(

    Where

    Hg is density of Mercury

    airis density of Air

    h = difference in height between two limbs of manometer

    2. Average Error

    ==

    n

    E

    AvgError

    n

    i 1

    Where

    E = Error for ith observation

    n = number of observations

    3. Accuracy:

    n

    ValuedardS

    ValueMeasuredValuedardS

    ErrorX

    n

    i

    =

    ==1

    100tan

    tan

    %

    Accuracy = X %

    Bourdon Gauge is X % inaccurate

    5

  • 7/28/2019 Manual -IPC 7th Semester

    6/36

    Observations and calculations:

    Density of Air = air= 1.2 kg/m3

    Density of Mercury = Hg = 13,650 kg/m3

    No. of

    obs.

    Gauge

    Pressure

    (kg/cm2)

    Gauge Pressure

    (X1)

    (kPa)

    Manometer

    Reading, H

    (inHg)

    Differential

    pressure

    measured by

    manometer

    X2

    (kPa)

    Error

    (X1-X2)

    (kPa)

    6

  • 7/28/2019 Manual -IPC 7th Semester

    7/36

    RESULT:

    Results

    AccuracyRange

    Average Error

    Experiment # 2: Calibration of Bourdon Gauge Using Dead

    Weight Tester

    7

  • 7/28/2019 Manual -IPC 7th Semester

    8/36

    Objective:

    To calibrate and report the accuracy of the given Bourdon gauge, using dead weight tester.

    Apparatus:

    Bourdon gauge

    Dead weight tester

    Weights.

    Working Principle:

    The working principle of the above depicted Dead weight tester is based on Pascal

    Law. This law states that if pressure is applied on a fluid at rest, the pressure is equally

    distributed to all directions, i.e. one to the piston of the dead weight tester and the other to the

    Bourdon gauge.

    Procedure:

    1. Initially the gauge which is to be calibrated is connected with the dead weight tester.

    2. The lever is moved outward completely

    3. After putting a pressure plate of suitable weight upon dead weight tester, the lever is

    gradually moved outwards.

    4. During closure of lever, Bourdon gauge showed increase in pressure. The piston is moved

    inwards until the scale of Bourdon gauge stops with the jerk.

    5. At that point Bourdon gauge reading is noted.

    6. Again the same procedure is repeated right from the beginning by placing another

    pressure plate on the top of already placed pressure plate.

    7. The graph between dead weight tester reading on X-axis and Bourdon gauge reading on

    Y-axis is plotted.

    Formulae used:

    8

  • 7/28/2019 Manual -IPC 7th Semester

    9/36

    1. Accuracy:

    n

    ValuedardS

    ValueMeasuredValuedardS

    ErrorX

    n

    i

    =

    ==1

    100tan

    tan

    %

    Accuracy = X % inaccurate

    2. Average Error

    ==

    n

    E

    AvgError

    n

    i 1

    Where

    E = Error for i

    th

    observationn = number of observations

    Observations & Calculations:

    Sr. No.

    Pressure applied

    by weights,

    P1

    Bourdon gauge

    Reading

    P2

    Error,

    E=P1-P2

    kg/cm2 kg/cm2 kg/cm2

    Calibration Curve:

    9

  • 7/28/2019 Manual -IPC 7th Semester

    10/36

    Result:

    Results

    Accuracy

    Range

    Average Error

    10

  • 7/28/2019 Manual -IPC 7th Semester

    11/36

    SUMMARYOF BOURDON GAUGE

    Category Pressure Measuring Device

    Working Principle Mechanical displacement due to pressure

    on fluid.

    Material of construction

    Berylium

    Copper

    steel

    chrome alloy steel

    stainless steel

    Accuracy 1-5% of full span

    Limits of application Up to 100 MPa.

    Advantages

    Low cost with reasonable accuracy.

    wide limits of application

    can be used in harsh environment

    Disadvantages

    affected by shock and vibrations

    have slow response time as compared to

    bellow

    Experiment # 3: Calibration of RTD

    Objective:

    11

  • 7/28/2019 Manual -IPC 7th Semester

    12/36

    To calibrate the given Resistance Temperature Detector using the Mercury Filled

    Thermometer.

    Apparatus:

    Resistance Thermometer (Platinum)

    Beaker

    Oil bath

    Thermometer

    Avometer

    Procedure:

    Initially the Avometer is standardized by joining the two ends of the wires of RTD to

    Avometer and then wires are short circuited in order to set the pointer at zero.

    The resistance thermometer is inserted in the oil beaker which already had the mercury

    filled thermometer.

    The oil in the beaker is heated and different sets of readings are taken for resistance and

    the temperature for every five degree centigrade temperature rise.

    Finally a graph between temperature and the resistance is plotted. The straight line drawn

    showed the fitness of the resistance thermometer under consideration for the required

    purpose.

    Observations:

    12

  • 7/28/2019 Manual -IPC 7th Semester

    13/36

    Sr. No.Temperature

    (C)

    Resistance

    (ohm)

    Calibration Curve:

    Summary Of RTD

    Category Temperaturesensor

    13

  • 7/28/2019 Manual -IPC 7th Semester

    14/36

    Working Principle

    Resistance of metal increase with increase

    in temperature.

    Material of construction Platinum

    Nickel

    Sensitivity 0.004/C to 0.005/C

    Limits of application

    Up to 650C for Platinum

    Up to 300C for Nickel

    Advantages

    High accuracy.

    Wide Range of application

    Good reproducibility

    Higher signal to noise ratio.

    Can be used in radiation environment

    Disadvantages

    Slower response time

    Expensive

    Experiment # 4: Calibration of Thermocouple usingThermometer

    Objective:

    To calibrate the given Thermocouple with the help of Thermometer

    14

  • 7/28/2019 Manual -IPC 7th Semester

    15/36

    Apparatus:

    Potentiometer

    Thermocouple

    Oil bath

    Mercury filled Thermometer

    Procedure:

    1. The standardization of potentiometer is done, when the thermocouple is not connected to

    it.

    2. After connecting the thermocouple with potentiometer, the thermocouple and a mercury

    filled thermometer are inserted into oil beaker.

    3. The oil in beaker is heated up to 180 C then for every 5 C drop in temperature a

    corresponding change in E.M.F. is noted via potentiometer.

    4. The change in E.M.F. along with the change in temperature of the system is plotted on a

    graph.

    Observations & calculations:

    No. of Obs.Temperature,

    TE.M.F.

    C mV

    Calibration Curve:

    15

  • 7/28/2019 Manual -IPC 7th Semester

    16/36

    Summary of thermocouple

    16

  • 7/28/2019 Manual -IPC 7th Semester

    17/36

    Category Temperature Sensor

    Working Principle SEEBECK effect

    Material of construction

    Chromel-Alumel

    Copper-Constantan

    Platinum,Rhodium-Platinum

    Limits of application -100 to 1100C

    Advantages

    Low cost with reasonable accuracy.

    Wide Range of application

    Good reproducibility

    Good accuracy

    Disadvantages

    Cannot be used for radiation environment

    Low value of emf is corrupted with noise

    Temperature is not exactly linear with emf

    Thermocouple types

    Type E- Type E (chromelconstantan)

    Type K- Type K (chromel{90 percent nickel and 10

    percent chromium}alumel)

    Type J- Type J (ironconstantan)

    Type N- Type N (NicrosilNisil)

    Type R- Type R thermocouples use a platinumrhodium

    alloy containing 13% rhodium

    Type S- Type S thermocouples are of 90% Platinum and

    10% Rhodium

    17

  • 7/28/2019 Manual -IPC 7th Semester

    18/36

    Experiment # 5: Time Constant of Mercury Filled

    Thermometer

    Objective: Determine the time constant of the given mercury filled thermometer dipped in an oil

    bath.

    Draw graph between Y(t)/A and t/ and report response Y(t) of the system

    when t=, t=2, t=3.

    Apparatus:

    Two mercury filled thermometer

    Oil bath

    Heating device

    Stop watch.

    Procedure:

    1. Initially the room temperature is noted.

    2. The mercury filled thermometer is dipped in the oil bath which is placed on the heating

    arrangement.

    3. The whole arrangement is heated till the temperature of the bath is reached to 220C. The

    attained temperature is maintained.

    4. Another thermometer, whose time constant is to determine is dipped in the same oil bath

    and the rise in temperature after every 5 seconds is noted along with time.

    5. The required parameters are calculated and a graph between above mentioned quantities

    is plotted.

    18

  • 7/28/2019 Manual -IPC 7th Semester

    19/36

    Observations and Calculations:

    Room temperature= y(s) = CMaximum temperature for heater oil bath = y() = 220C

    Amplitude = A = y() - y(s)

    Sr. No

    Time

    t

    Temperature

    of

    Thermometer

    y(t),

    Y(t)=

    y(t)-

    y(s)

    Y(t)/

    At/ = - ln(1- (Y(t)/A))

    (sec) C C

    19

  • 7/28/2019 Manual -IPC 7th Semester

    20/36

    Response of system at when t=, t=2, t=3.

    At t/ = 1,

    Y(t)/A = 0.63 (From graph)

    Y(t) = 0.63 x A C

    Results:

    Results

    Time Constant (sec)

    Response of System

    Y(t) when t/ =

    1C

    Y(t) when t/ =

    2C

    Y(t) when t/ =

    3C

    20

  • 7/28/2019 Manual -IPC 7th Semester

    21/36

    Summary of thermometer

    Category Temperature Measuring Device

    Working Principle Expansion of fluid with increase in

    temperature.

    Material of construction

    Ordinary Soda Lime

    Fused Quartz

    Advantages of mercury

    Hg is opaque.

    Hg is good conductor.

    It does not wet the glass surface

    Limits of application

    Up to 350C for mercury

    Less than 120C for alcohol

    Advantages

    Low cost

    Can be used easily.

    Used as STANDARD equipment for

    calibration of temperature sensors

    Good accuracy

    Disadvantages Cannot be used in industry for automatic

    control.

    21

  • 7/28/2019 Manual -IPC 7th Semester

    22/36

    Experiment # 6: Liquid Level System

    Objective:

    Find the Time Constant of the given Liquid Level System, also report the effect of dia and

    length of tube on time constant of system

    Apparatus:

    Storage tank,

    Tubes (of certain lengths),

    Beaker,

    Scale

    Stop watch.

    Procedure:

    1. A tube of certain length and diameter was fitted at the bottom of tank. Storage tank was

    filled up to a certain level (say h1). A finger was put at the end of the tube so that no

    water can flow.

    2. Then a beaker was placed under the tube and the finger was removed from the lower end

    of the tube. The water began to flow and at the same time, a stopwatch was operated and

    the time for which the level of water fell to a certain height in the storage tank (say h2)

    was noted. The volume of the water that fell into the beaker was measured.

    3. The mean of height or level of water was also noted.

    4. Diameter of Storage tank was also measured.

    5. By drawing a graph of mean level of water (along X axis) vs. the flow rate(along Y-axis),

    the resistance was noted.6. The time constant of the system can be noted by the formula.

    Time Constant = Resistance x Storage Capacitance

    7. The experiment was repeated by taking tubes of different length and diameter.

    22

  • 7/28/2019 Manual -IPC 7th Semester

    23/36

    Formulae Used:

    1. Area of tank

    A = cm2

    Whered = diameter of tank

    2. Time Constant:

    = Rx A

    Where

    A = Area of Tank

    R= Resistance to flow

    3. Resistance (from graph):

    Observations & Calculations:

    Diameter of the storage tank = d cm

    Area of the storage tank = A cm2

    1. Length of tube = L1 cm

    Serial #

    Initial

    level of

    water(h1)

    Final level

    of water

    (h2)

    Mean Level

    (H)

    Volume

    Time

    of

    flow

    Flow rate

    Q

    1 in in in cm (ml) (sec) (ml/sec)

    2 7.5 6.5 7

    3 6.5 5.5 6

    4 5.5 4.5 5

    23

  • 7/28/2019 Manual -IPC 7th Semester

    24/36

    5 4.5 3.5 4

    6 3.5 2.5 3

    7 2.5 1.5 2

    From graph:

    R=Resistance to flow =

    Time constant:

    = R A

    Result were repeated for different tube lengths and diameters.

    24

    0

    2

    4

    6

    8

    10

    12

    2 4 6 8 10 12

    Flowrate,Q(ml/s)

    Mean Level (cm)

    Time Constant of Liquid Level System with tube oflength L1

  • 7/28/2019 Manual -IPC 7th Semester

    25/36

    Results:

    Results

    Dia of tube = constant = d cm

    Time Constant for L1 = 1

    Time Constant for L2 = 2

    Time Constant for L3 = 3

    Length of tube = constant = L cm

    Time Constant for d1 = 1

    Time Constant for d2 = 2

    Time Constant for d3 = 3

    25

  • 7/28/2019 Manual -IPC 7th Semester

    26/36

  • 7/28/2019 Manual -IPC 7th Semester

    27/36

    7. The pump is still not started. When the pump is started immediately one person will

    record the readings of magnetic flow transmitter and other person simultaneously

    record the reading (maxima and minima)of the output displayed with reference to the

    time i.e. readings can be taken after every 5 seconds or any other time which may be

    suitable. Also observes how quickly the system stabilizes.

    8. When set point is achieved and approximately oscillations are stopped then the pump

    is to be stopped from the control panel switch. Same procedure is repeated for

    different gain values (0.6, 1.6) and keeping other parameters constant.

    Procedure:

    Methodology adopted would be the same and the variables will have the following values: Gain = 0.6 for all the three inputs

    Rate minute = 0.00 for all the three inputs

    Reset min = 0.1, 0.15 and 0.5

    Same procedure will be followed to develop the graph.

    3) PID Mode

    Procedure:

    Methodology adopted would be the same and the variables will have the following values,

    Gain = 0.6 for all the three inputs

    Rate minute = 1, 2, 3

    Reset min = 0.1 for all the three inputs

    Same procedure will be followed to develop the graph..

    27

  • 7/28/2019 Manual -IPC 7th Semester

    28/36

    Block Diagram:

    28

  • 7/28/2019 Manual -IPC 7th Semester

    29/36

    Flow Diagram:

    1. Storage Tank

    2. Pump

    3. Control Valve

    4. Restriction Valve

    5. Bottom valve

    FT-1: Flow Transmitter

    FC-1: Flow Controller

    LC-2: Positioner

    PI-1: Pressure gauge

    29

  • 7/28/2019 Manual -IPC 7th Semester

    30/36

    FT-2: Flow Transmitter (Magnetic)

    2) PI mode

    Proportional Controller

    Set point=30%Plot Graph for all gains by taking time against x-axis and Controller Output across Y axis.

    30

    Tuning

    ParametersGain = 0.6 Gain =1.0 Gain =1.6

    Time

    (s)

    ControlledVariable

    (%TO)

    ControlledVariable

    (%TO)

    ControlledVariable

    (%TO)

    0

    5

    10

    15

    20

  • 7/28/2019 Manual -IPC 7th Semester

    31/36

    Proportional Integral Derivative Controller:

    Set point=30%

    31

    Tuning

    Parameters

    Gain = 0.6

    Integral time=0.1

    Gain =0.6

    Integral time=0.15

    Gain =0.6

    Integral time=0.5

    Time

    (s)

    ControlledVariable

    (%TO)

    ControlledVariable

    (%TO)

    ControlledVariable

    (%TO)

    0

    5

    10

    15

    20

  • 7/28/2019 Manual -IPC 7th Semester

    32/36

    32

    Tuning

    Parameters

    Gain = 0.6Integral time=0.1

    Rate min=1

    Gain =0.6Integral time=0.1

    Rate min=2

    Gain =0.6Integral time=0.1

    Rate min=3

    Time

    (s)

    Controlled

    Variable

    (%TO)

    Controlled

    Variable

    (%TO)

    Controlled

    Variable

    (%TO)

    0

    5

    10

    1520

    25

  • 7/28/2019 Manual -IPC 7th Semester

    33/36

    Experiment # 8: Temperature Control Loop

    Objectives:

    To study the response of the process variable (temperature) in the on-off

    algorithm. Also plot the graph between time and temperature during heating and cooling.

    Procedure:

    1. Initially the water supply tube from the cooling element is connected to the main water

    supply.

    2. The outlet tube from the solenoid valve is connected to the drainage.

    3. The process container is filled with water so as to cover the cooling element.

    4. The electric power supply is connected.

    5. The controller is set according to the following procedure.

    6. The set point Select key is pressed until Algorithm is displayed.

    7. The set point value is set to the required value (according to heating or cooling system)

    8. For the above set point, note down reading of temperature against time.

    9. Plot a graph b/w temperature (along y-axis and time (along x-axis) both for cooling and

    heating process.

    33

  • 7/28/2019 Manual -IPC 7th Semester

    34/36

    During heating:-

    Set Point=55oC

    During Cooling:

    Set Point=46.7 oC

    34

    Sr no. Time (sec) Temperature (oC)

    1. 5

    2. 10

    3. 15

    4. 20

    5. 25

    6. 30

    7. 35

    8. 409. 45

    10. 50

  • 7/28/2019 Manual -IPC 7th Semester

    35/36

    Reference:

    1. Coughanowr, D. R.:Process Systems Analysis and Control, 2nd ed., McGraw-Hill.

    2. Fribance, A. E.: Industrial InstrumentationFundamentals,McGraw-Hill, 1962.

    3. Luyben, W. L.: Process Modeling, simulation, and Control for Chemical Engineers,

    3rd ed,McGraw-Hill, Inc., 1997.

    4. Seborg, D. E. et al.:Process Dynamics and Control, 2nd ed, John Wiley & Sons, Inc.,

    1989.

    35

    Sr no. Time (sec) Temperature (oC)

    1.2.

    3.

    4.

    5.

    6.

    7.

    8.

    9.

  • 7/28/2019 Manual -IPC 7th Semester

    36/36