1. SYSTEM DESCRIPTION
MARSIC200 + MARSIC300
MARSIC200 and MARSIC300 are especially developed to measure emissions
aboard of ships according to all relevant maritime regulations like
MARPOL Annex VI, MEPC.259(68)1, MVR, etc.
Designed, that maintenance and service is simple, even for a board engineer
Developed with less use of movable parts to extent maintenance intervals and keep service & TCO down
Cost optimized systems due to multiple sampling points with only one analyzer
Complete Type Approved by DNV GL and ABS (CCS applied) according to MARPOL Annex VI, related
guidelines and international classification standards IACS E101
No need of calibration gas due to certified internal calibration filters (MARSIC300)
Service concept included spare part hubs all over the world and 24/7 maritime Helpline
Extended documentation incl. easy understandable short descriptions, installation plans, etc.
1) International Association of Classification Societies
Center line content
3 Hinrich Brumm| Confidential 2016-03-10
2. TARGET APPLICATIONS
MARSIC200 + MARSIC300
Emission monitoring acc.to MARPOL Annex VI and MEPC.259(68)1 for:
Monitoring of DeSOx and DeNOx exhaust gas cleaning systems
Process Measurements for:
Monitoring of O2, CO, CO2, NOx for optimization of engine performance
Monitoring of CH4 / Methane-Slip Measurement on LNG / Dual Fuel Engines
Scrubber and SCR Efficiency (measurement in raw gas and after EGCS)
1) Former: MEPC.184(59)
Center line content
4 Hinrich Brumm | Confidential 2016-03-10
3. MARSIC200 - TECHNOLOGY MARSIC200 + MARSIC300
Sampling probe
Conditioning unit
Distribution unit
Analyzer
Pending
Center line content
5 Hinrich Brumm | Confidential 2016-03-10
3. MARSIC200 – SHORT BRIEF
MARSIC200 + MARSIC300
Measurement Principle Cold-dry extractive system
Short heated sampling gas line to conditioning unit (cooler), long unheated line to analyzer
General Technical Arrangement Modular design: Sampling conditioning unit, distributor and analyzer
Each sampling point has its own sampling conditioning unit
Measuring Components Up to 5 components: CO2 with NDIR, NO, NO2, SO2 with NDUV, O2 with ECS
Calibration Concept With calibration gas every three months, zero setting automatically with instrument air
Sampling Points 1-4 Sampling points can be measured with one analyzer in 4:45 min1
Applications DeNOx (prior and aft SCR), DeSOx (after Scrubber)
Process (engine optimization CO2, O2, NOx) 1) As required by MEPC.259(68)
Center line content
6 Hinrich Brumm | Confidential 2016-03-10
3. MARSIC200 - ARRANGEMENT
MARSIC200 + MARSIC300
1 2
3 4
Long distance unheated sample gas line (PTFE)
One sampling probe each sampling point
Only short self-regulated (120°C) heated sample gas line for each sampling point
One conditioning unit incl. gas cooler each sampling point
One distribution unit for all sampling points
One analyzer for all sampling points
Up to 4 Sampling Points with only one analyzer
Max. measuring distance more than 35 m Show animation
Center line content
7 Hinrich Brumm | Confidential 2016-03-10
3. MARSIC200 - ARRANGEMENT
MARSIC200 + MARSIC300
1
For measuring of one sampling point, only one sample gas pump in the distribution unit is required
Distribution unit
Sample pump
Center line content
8 Hinrich Brumm | Confidential 2016-03-10
3. MARSIC200 - ARRANGEMENT
MARSIC200 + MARSIC300
1
For measuring of one sampling point, only one sample gas pump in the distribution unit is required
Measuring of 2-4 sampling points with two pumps (sample- and bypass pump) in the distributor unit
With this technology, up to 4 sampling points can be
measured within the 4:45min acc. to MEPC.259(68)1
2
Outlet / back to exhaust pipe
Distribution unit
Bypass pump
Sample pump
Klick to activate Sampling Point 2
Klick to activate Sampling Point 1
1 Former MEPC.184(59)
Center line content
9 Hinrich Brumm | Confidential 2016-03-10
3. MARSIC200 – TECHNICAL DETAILS
MARSIC200 + MARSIC300
For a technical excursion
click into the modules.
To skip, click here
Center line content
10 Hinrich Brumm | Confidential 2016-03-10
3. MARSIC200 - SUMMARY
MARSIC200 + MARSIC300
Feature 1: Up to four sampling points with only one analyzer
Your Benefit: Less amount of invest compared to in-situ solution
Feature 2: Modular design (3 Boxes) and only short heated lines
Your Benefit: Easily placeable, high flexibility for installation
Feature 3 SO2-measurement with UV instead of NDIR
Your Benefit: High reliability and accuracy, because no water cross sensitivity
Feature 4: Maintenance optimized technology (e.g. 4 year maintenance interval of UV Lamp)
Your Benefit: Low maintenance effort & TCO costs
Feature 5: Extended calibration interval of three months, zero setting automatically
Your Benefit: Low calibration effort & costs
Center line content
11 Hinrich Brumm | Confidential 2016-03-10
4. MARSIC300 - TECHNOLOGY MARSIC200 + MARSIC300
Filter Sampling Unit
Analyzer
Bundle of pipes
Heated Sampling Line
Sampling Probe heated/unheated
Pending Pending
Center line content
12 Hinrich Brumm | Confidential 2016-03-10
4. MARSIC300 – SHORT BRIEF
MARSIC200 + MARSIC300
Measurement Principle Hot-wet extractive system
Sample gas is complete heated up to 200°C from sample probe to analyzer means no condensation of
H2O, H2SO4, HNO3, and no condensate handling
General Technical Arrangement Compact design: Only one analyzer, all signal- power and PTFE-tubes in a combined „bundle of pipe“
Measuring components
Up to 9 Components: NO, NO2, CO2, SO2, O2, CO, NH3, CH4 and H2O
Calibration concept Calibration with internal, certified filters. calibration with cal.- gas optional
Sampling Points 1-2 Sampling points can be measured with one analyzer in 4:45 min1
Applications
DeNOx (prior and aft SCR), DeSOx (prior and after scrubber)
Process: Engine optimization (CO2, O2, NOx) , Methane slip measurement
1) As required by MEPC.259(68)
Center line content
13 Hinrich Brumm | Confidential 2016-03-10
4. MARSIC300 - CABINET MARSIC200 + MARSIC300
Bundle of pipes
Valve unit control of the
sampling filter and test gas
Analyzer with internal calibration
Heated sampling line (200°C)
External Instrument Air conditioning
unit included
Instrument Air pressure regulation unit
Photometer
Click here for more
technical details…
Sampling point 1
Sampling point 2
Center line content
14 Hinrich Brumm | Confidential 2016-03-10
Sampling Filter Unit SFU
Fine filter cartridge: glass fibre 0,1 µm
Back purge function
Main valve for sample point switching + standby
Flange (NW 65 / ND 6) others optional
Stainless steel heated probe (0.5 or 0.8 m length) or
unheated probe (0.2 ; 0.5 or 0.8 m) length
Glass fibre filter 0,1µm
IP54 Wether Protection (incl. Heating for temperature -25°C ) available
Bundle of Pipes
For connection of analyzer with sampling filter unit SFU
All electrical lines (power, signal) and pneumatics in one bundle
Very easy installation
Halogen reduced
4. MARSIC300– SAMPLING FILTER UNIT SFU MARSIC200 + MARSIC300
Center line content
15 Hinrich Brumm | Confidential 2016-03-10
4. MARSIC300 – INTERNAL CALIBRATION
MARSIC200 + MARSIC300
0. MARSIC300 in measurement mode
1. flushing the cuvette with instrument air
2. Setting signal zero (internal zero adjustment)
3. Swing in of specific calibration filter
4. Setting span point calibration
5. Swing of specific calibration filter
6. Switch back to measurement
Calibration curve for SO2
SO
2 C
once
ntra
tion
(pp
m)
SO2 raw signal
Light Source
Filter wheel (component specific wavelength)
Cuvette
Detector
Filter wheel (internal calibration filter)
SO2
SO2 SO2
SO2
SO2
SO2
SO2
SO2
SO2
SO2
SO2
SO2
SO2
Measuring value
Center line content
16 Hinrich Brumm | Confidential 2016-03-10
4. MARSIC300– INTERNAL CALIBRATION
MARSIC200 + MARSIC300
Advantage of certified calibration filters on MARSIC300:
The optical path during calibration is exact the complete light path during the "normal" measurement
mode. Aging effects of all optical components are taken into account and are corrected accordingly
Checking with calibration gas only necessary after optical parts have been replaced!
No hazzle with expired calibration gas bottles (usually expired after 12months)
No costs for calibration gas (10l bottle of calibration gas (CO2/SO2) costs around 600€)
The operator of the vessel is always in the position to show compliance with the applied regulations
against flag state, class or PSC (port state control) without additional measures
The MARSIC300 is the only device with this feature available on the market!
Center line content
17 Hinrich Brumm | Confidential 2016-03-10
4. MARSIC300– SHORT BRIEF
MARSIC200 + MARSIC300
Feature 1: Certified Internal calibration filters
Your Benefit: Always compliant, no need for expensive gases, less calibration effort
Feature 2: Hot extractive principle, no condensation of H2O, H2SO4 HNO3, etc.
Your Benefit: No condensate handling, less maintenance effort & costs
Feature 3: Up to nine components with high and low ranges
Your Benefit: Covers every maritime application
Feature 4: Integrated power transformer and network balancing for nearly every situation aboard
Your Benefit: No additional transformer needed, less invest and installation costs
Feature 5: Compact design
Your Benefit: Easy installation (bundle of pipes)
Center line content
18 Hinrich Brumm | Confidential 2016-03-10
5. DIFFERENTIATION MARSIC200 AND MARSIC300 MARSIC200 + MARSIC300
MARSIC200 MARSIC300
Center line content
MARSIC200 + MARSIC300 5. DIFFERENTIATION MARSIC200 AND MARSIC300
19 2016-03-10 Hinrich Brumm| Confidential
…But different
Different technologies: MARSIC200: cold/dry extractive MARSIC300: hot/wet extractive
Different design: MARSIC200: 3 Boxes, modular MARSIC300: 1 Box, compact
Different nos. of standard sampling points: MARSIC200: 4; MARSIC300: 2 Sampling Points
Different options (components and measuring ranges)
Same basic application maritime exhaust gas measurement (DeSOx and DeNOx)
Same requirements due to the given MARPOL guidelines
Same basic design concept adopted to special situation aboard
Same service concept
Same same….
Center line content
20 Hinrich Brumm | Confidential 2016-03-10
5. DIFFERENTIATION MARSIC200 AND MARSIC300
MARSIC200 + MARSIC300
Component MARSIC200 MARSIC300
Measurement Principle Cold-dry Hot-wet
Design 3 Boxes, Modular 1 Box, Compact
No of standard sampling points 4 2
Fully Type Approved Yes Yes
NOX Before SCR Yes Yes
NOx After SCR Yes Yes
SOX Before Scrubber No (SO2 max =500ppm)
Yes
SOX After Scrubber Yes Yes
CO2, SO2, O2, NO, NO2 Yes Yes
CH4, CO, NH3 and H2O No Yes
Integrated calibration filters No (calibration interval= 3 months)
Yes
Integrated power transformer No Yes
I-air conditioning inclusive No Yes
Center line content
21 Hinrich Brumm | Confidential 2016-03-10
5. DIFFERENTIATION MARSIC200 AND MARSIC300
MARSIC200 + MARSIC300
Component MARSIC200 MARSIC300
SO2
0 … 100 ppm 0 … 30 ppm
0 … 500 ppm 0 … 2.000 ppm
CO2 0 … 25 Vol.% 0 … 25 Vol.%
O2 0 ... 21 Vol.-% 0 ... 21 Vol.-%
NO
0 ... 300 ppm 0 ... 300 ppm
0 ... 1.500 ppm 0 ... 2000 ppm
NO2 0 ... 300 ppm 0 ... 300 ppm
0 ... 1.500 ppm 0 ... 2000 ppm
CO 0 ... 200 ppm 0 ... 2000 ppm
CH4 0 ... 500 ppm 0 ... 10 000 ppm
NH3 0 … 50 ppm 0 … 500 ppm
H2O 0 … 40 Vol.%
Center line content
MARSIC200 + MARSIC300 5. DIFFERENTIATION MARSIC200 AND MARSIC3001
22 2015-09-29 Hinrich Brumm| Confidential
Spare Parts and Consumables2) MARSIC200 MARSIC300
Maintenance / Year 1.366 € 1.062€
Manpower / year 16 hours 8 hours
Calibration Costs MARSIC200 MARSIC300
Manpower / year 4 hours 0 hours
Cost / Year 500 € - €
1) Standard System with one Sampling Point, 10m heated measureing line and unheated probe 2) Replacing UV-Lamp in MARSIC200 incl. Maintenance every 4 Years / IR-Source for MARSIC300 every two years not included 3) Calculated as 360 days x 24h x 50% nominal power
Total Costs of Ownership MARSIC200 MARSIC300
Manpower / year 20 hours 8 hours
Consumption / year 6.264 kWh 10.152 kWh
Spares, Consumables, Cal.-gas / Year 1.866 € 1.062 €
Power Consumption MARSIC200 MARSIC300
Consumption / year3
6.264 kWh 10.152 kWh
Center line content
23 Hinrich Brumm | Confidential 2016-03-10
6. USER INTERFACE CONCEPT AND CONNECTIVITY MARSIC200 + MARSIC300
SOPAS-ET
(SICK Engineering Tool)
I/O signals (digital* & analog*), Modbus, OPC
Routine Operation via Control Unit All required routine functions via instrument display
Multilingual: English, German, French, Spanish
Logbook function for status and alarms
Password protected operating levels
Ethernet Connection; Direct or via network
Remote diagnosis
Remote maintenance
Remote operation
Connectivity to customers devices Control via SICK solution MEAC
Control of the MARSIC via customers SPS
Data Recording on external devices
Prepared to embed the MARSICin customers
overall solution
GSM or
Satellite
SICK Meeting Point Router
*MARSIC200 only
Center line content
24 Hinrich Brumm | Confidential 2016-03-10
Complete new service concept for MARSIC
MARSIC is designed that the service can be done even by board engineer
In case of failures, only pre-calibrated modules are to be exchanged
Loan modules are also available for fast replacement
Easy to understand description within operating instructions
Short instructions for commissioning, calibration, etc.
SICK Maritime Hotline
Worldwide accessibility for customers of MARSIC on ships
Acceptance for parts orders or technical support – 24/7 worldwide availability to take the call
SICK as a Partner
50 years of experience in CEMS from SICK as a market leader (2015: ~60 Board installations)
Maritime tradition for more than 70 years
More than 7.000 employees in more then 88 countries
7. MARITIME SERVICE CONCEPT & SICK MARITIME HOTLINE MARSIC200 + MARSIC300
Center line content
25 Hinrich Brumm | Confidential 2016-03-10
8. CONCLUSION
MARSIC200 + MARSIC300
Conclusion
Center line content
MARSIC200 + MARSIC300 8. CONCLUSION
26 2016-03-10 Hinrich Brumm| Confidential
Reliability and Acceptance
Developed especially for maritime application → Insensitive to vibration, power fluctuation, etc.
Good accessibility of components, modular design → easy to maintain & servicing
Fully Type Approved (MARPOL AND Class) by major societies → No acceptance problems, nowhere
Cost optimized system
Up to four sampling points covered by only one analyzer → Less invest and service effort
Minimum of movable parts inside and extended maintenance intervals → Keeps service effort and TCO down
Internal calibration filters → Safe huge amount of money over the lifetime of the analyzer
Ready for the future
Low measurement ranges covering also tougher future limits in required accuracy → No need for further invests
Well-thought-out Service Package
Remote Service, Maritime Helpline 24/7 and a global Network → Help and Support whenever and wherever needed
Center line content
MARSIC200 + MARSIC300
27 2016-03-10 Hinrich Brumm| Confidential
SAFELY ON THE RIGHT COURSE