Page 1: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Mass Spectrometry

Page 2: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Mass Spectrometry (MS)

• An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase.– Mass spectrometry is our most valuable

analytical tool for determining accurate molecular masses.

– Also can give information about structure.– Proteins can now be sequenced by MS.

Page 3: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

A Mass Spectrometer

• A mass spectrometer is designed to do three things:– Convert neutral atoms or molecules into a

beam of positive (or rarely negative) ions.– Separate the ions on the basis of their

mass-to-charge (m/z) ratio.– Measure the relative abundance of each

ion.

Page 4: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Mass Spectrometry (MS)

• Schematic of an electron ionization mass spectrometer (EI-MS).

Page 5: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

A Mass Spectrometer

• Electron Ionization MS– In the ionization chamber, the sample is

bombarded with a beam of high-energy electrons.

– Collisions between these electrons and the sample result in loss of electrons from sample molecules and formation of positive ions.

++

+

C HH

H

H

CH

H

He 2eH

Molecular ion(a radical cation)

Page 6: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Molecular Ion

• Molecular ion (M):Molecular ion (M): A radical cation formed by removal of a single electron from a parent molecule in a mass spectrometer.

• For our purposes, it does not matter which electron is lost; radical cation character is delocalized throughout the molecule; therefore, we write the molecular formula of the parent molecule in brackets with– a plus sign to show that it is a cation.– a dot to show that it has an odd number of

electrons.

Page 7: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Molecular Ion

At times, however, we find it useful to depict the radical cation at a certain position in order to better understand its reactions.

CH3CH2OCH(CH3)2CH3CH2OCH(CH3)2.

Page 8: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Mass Spectrum

• Mass spectrum:Mass spectrum: A plot of the relative abundance of ions versus their mass-to-charge ratio.

• Base peak:Base peak: The most abundant peak. Assigned an arbitrary intensity of 100.

• The relative abundance of all other ions is reported as a % of abundance of the base peak.

Page 9: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

MS of dopamine

A partial mass spectrum of dopamine showing all peaks with intensity equal to or greater than 0.5% of the base peak.

Page 10: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Other MS techniques

• What we have described is called electron ionization mass spectrometry (EI-MS).

• Other mass spectrometry techniques include– fast atom bombardment (FAB).– matrix-assisted laser desorption ionization

(MALDI).– chemical ionization (CI).– electrospray ionization (ESI).

Page 11: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Resolution

• Resolution:Resolution: A measure of how well a mass spectrometer separates ions of different masses.– Low Resolution MS (LR-MS):Low Resolution MS (LR-MS): Refers to

instruments capable of separating only ions that differ in nominal mass; that is ions that differ by at least 1 or more atomic mass units (amu).

– High resolution MS (HR-MS):High resolution MS (HR-MS): Refers to instruments capable of separating ions that differ in mass by as little as 0.0001 amu.

Page 12: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Resolution

– C3H6O and C3H8O have nominal masses of 58 and 60, and can be distinguished by low-resolution MS.

– C3H8O and C2H4O2 both have nominal masses of 60 which occurs due to isotopes of the same elements.

– distinguish between them by high-resolution MS.

C2H4O2

C3H8O

60.02112

60.05754

60

60

MolecularFormula

Nominal Mass

PreciseMass

Page 13: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Isotopes

Virtually all elements common to organic compounds are mixtures of isotopes with different relative abundances.

0.3815.000110014.00311.1113.0034

10012.00000.016

1002.014101.00783

Mass(amu)

14.007

12.011

1.0079

nitrogen

carbon

hydrogen

Relative AbundanceIsotopeElement

Atomicweight

98.080.916310078.918332.536.9659

10034.96894.4033.9679

10031.97210.2017.9992

10015.9949

79.904

35.453

32.066

15.999

bromine

chlorine

sulfur

oxygen

15N14N

12C13C

2H

1H

35Cl

81Br

37Cl79Br

32S34S

16O18O

Page 14: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Isotopes

– Carbon, for example, in nature is 98.90% 12C and 1.10% 113C.

– There are 1.11 atoms of carbon-13 in nature for every 100 atoms of carbon-12. So low intensity M+1 peak is very likely to appear along with M+ peak.

1.10 13C x 10098.90 12C

= 1.11 atoms 13C per 100 atoms 12Cx 100

Page 15: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

M+2 and M+1 Peaks

• The most common elements giving rise to significant M + 2 peaks are chlorine and bromine.– Chlorine in nature is 75.77% 35Cl and 24.23% 37Cl.– A ratio of M to M + 2 of approximately 3:1 indicates the

presence of a single chlorine in a compound, as seen in the MS of chloroethane.

Page 16: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

M+2 and M+1 Peaks

– Bromine in nature is 50.7% 79Br and 49.3% 81Br.– A ratio of M to M + 2 of approximately 1:1

indicates the presence of a single bromine atom in a compound, as seen in the MS of

1-bromopropane.

Page 17: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

M+2 and M+1 Peaks

• Sulfur is the only other element common to organic compounds that gives a significant M + 2 peak

32S = 95.02% and 34S = 4.21%• Because M + 1 peaks are relatively low in

intensity compared to the molecular ion and often difficult to measure with any precision, they are generally not useful for accurate determinations of molecular weight.

Page 18: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry
Page 19: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Interpreting a mass spectrum

• The only elements to give significant M + 2 peaks are Cl and Br.– If no large M + 2 peak is present, these elements

are absent.• Is the mass of the molecular ion odd or even?• Nitrogen Rule:Nitrogen Rule: If a compound has

– zero or an even number of nitrogen atoms, its molecular ion will have an even m/z value.

– an odd number of nitrogen atoms, its molecular ion will have an odd m/z value.

Page 20: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Fragmentation of the Molecular Ion

• To attain high efficiency of molecular ion formation and give reproducible mass spectra, it is common to use electrons with energies of approximately 70 eV [6750 kJ (1600 kcal)/mol].– This energy is sufficient not only to dislodge one or

more electrons from a molecule, but also to cause extensive fragmentation.

– These fragments may be unstable as well and, in turn, break apart to even smaller fragments.

Page 21: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Fragmentation of M

• Fragmentation of a molecular ion, M, produces a radical and a cation.– Only the cation is detected by MS.

A-B

A

A

B

B

•+

Molecular ion(a radical cation)

+•

•+

+

+

Cation Radical

Radical Cation

Page 22: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Fragmentation of M

• A great deal of the chemistry of ion fragmentation can be understood in terms of the formation and relative stabilities of carbocations in solution.– Where fragmentation occurs to form new cations,

the mode that gives the most stable cation is favored.

– The probability of fragmentation to form new carbocations increases in the order.

CH3+ 1°

2°1° allylic

1° benzylic< < <

3°2° allylic

2° benzylic< 3° allylic

3° benzylic

Page 23: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Interpreting a mass spectrum

• The only elements to give significant M + 2 peaks are Cl and Br.– If no large M + 2 peak is present, these elements

are absent.• Is the mass of the molecular ion odd or even?• Nitrogen Rule:Nitrogen Rule: If a compound has

– zero or an even number of nitrogen atoms, its molecular ion will have an even m/z value.

– an odd number of nitrogen atoms, its molecular ion will have an odd m/z value.

Page 24: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alkanes

• Fragmentation tends to occur in the middle of unbranched chains rather than at the ends.

• The difference in energy among allylic, benzylic, 3°, 2°, 1°, and methyl cations is much greater than the difference among comparable radicals.– Where alternative modes of fragmentation are

possible, the more stable carbocation tends to form in preference to the more stable radical.

Page 25: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alkanes

– Mass spectrum of octane.

Page 26: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alkanes

– Mass spectrum of 2,2,4-trimethylpentane.

Page 27: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alkanes

– Mass spectrum of methylcyclopentane.

Page 28: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alkenes• Alkenes characteristically

– show a strong molecular ion peak.– cleave to form resonance-stabilized allylic cations.

[CH2=CHCH2CH2CH3] CH2=CHCH2+ + •CH2CH3

+•

Page 29: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Cyclohexenes

– Cyclohexenes give a 1,3-diene and an alkene, a process that is the reverse of a Diels-Alder reaction.

A radical cation (m/z 68)

+•

•+

+

CH3

CH3C CH2

CCH2H3C

CH3

Limonene(m/z 136)

A neutral diene(m/z 68)

Page 30: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alkynes• Alkynes typically

– show a strong molecular ion peak.– cleave readily to form the resonance-stabilized

propargyl cation or substituted propargyl cations.

HC C=CH2HC C-CH2+ +3-Propynyl cation

(Propargyl cation)

Page 31: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alcohols

• One of the most common fragmentation patterns of alcohols is loss of H2O to give a peak which corresponds to M-18.

• Another common pattern is loss of an alkyl group from the carbon bearing the OH to give a resonance-stabilized oxonium ion and an alkyl radical.

Molecular ion(a radical cation)

A radical

••

•++O HC R

••

••

+R'-C O H

A resonance-stabilized oxonium ion

RR"

R'

R"R'-C=O-H

R"

+

Page 32: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Alcohols

– Mass spectrum of 1-butanol.

Page 33: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Aldehydes and Ketones

• Characteristic fragmentation patterns are

-cleavage– McLafferty rearrangement

m/z 128

+•-cleavage

m/z 43+

+

+

m/z 113CH3

O

O

O

+

•+

m/z 58

McLaffertyrearrangement

Molecular ionm/z 114

•+O

HO

H+

Page 34: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Aldehydes and Ketones

– Mass spectrum of 2-octanone.

Page 35: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Carboxylic Acids

• Characteristic fragmentation patterns are -cleavage to give the ion [CO2H]+ with m/z 45.

– McLafferty rearrangement.

-cleavage• +

m/z 45

O=C-O-HOH

O

Molecular ionm/z 88

+

•+•+

+

McLaffertyrearrangement

m/z 60

OH

OH

OH

OHMolecular ion

m/z 88

Page 36: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Carboxylic Acids

– Mass spectrum of butanoic acid.

Page 37: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Esters

-cleavage and McLafferty rearrangement

+

+

•+•+ m/z 71

m/z 59

Molecular ionm/z 102

-cleavage

OCH3

OCH3

O

O

OCH3

O+

m/z 74

McLaffertyrearrangement +

+•+•

O

OCH3

H OH

OCH3

Molecular ionm/z 102

Page 38: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Esters

– Mass spectrum of methyl butanoate.

Page 39: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Aromatic Hydrocarbons

– Most show an intense molecular ion peak.– Most alkylbenzenes show a fragment ion of m/z 91.

+•

+ Tropylium cation (m/z 91)

Toluene radical cation

CH3 H

HH

H

HH

H

-H•

Page 40: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Amines

• The most characteristic fragmentation pattern of 1°, 2°, and 3° aliphatic amines is -cleavage.

m/z 30

++•

-cleavageCH3 CH3

CH3-CH-CH2-CH2-NH2 CH3-CH-CH2 CH2=NH2

Page 41: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Mass spectrometry by chemical ionization

• In chemical ionization (C.I.) mass spectrometry the electron beam is used to ionize a simple molecule (reagent gas) such as methane which in turn ionizes our molecule by collision and transfer of a proton.

Under electron bombardment, methane loses a bonding electron to give CH4

+• which reacts with an unionized methane molecule to give CH3

• and CH5+.

CH5+ does have a carbon atom with five bonds. But it has only

eight electrons! These are distributed between five bonds (hence the + charge) and the structure is thought to be trigonal bipyramidal.

This structure has not been determined as it is too unstable. It is merely proposed from theoretical calculations.

Page 42: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Chemical Ionization (CI)

Page 43: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

• This unstable compound is a powerful acid, and can

protonate just about any other molecule.

• This unstable compound is a powerful acid, and can

protonate just about any other molecule.

When it protonates our sample, a proton has been added rather

than an electron removed, so the resulting particles are simple

cations, not radical cations, and are generally more stable than the

radical cations produced by direct electron impact.

So the molecular ion has a better chance of lasting the necessary

20 ms to reach the detector.

Note :We observe [M + H]+ (i.e. one more than the molecular mass)

rather than M+ by this method.

Page 44: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Fast Atom Bombardment (FAB)

• Polar molecules, such as peptides, with molecular weights up to 10,000 Da can be analyzed by a "soft" ionization technique called fast atom bombardment (FAB).

• The bombarding beam consists of xenon (or argon) atoms of high translational energy (Xe). This beam is produced by first ionizing xenon atoms with electrons to give xenon radical cations:

Page 45: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

• The radical cations are accelerated to 6-10 keV to give radical cations of high translational energy (Xe) +∙ , which are then passed through xenon. During this passage, the charged high-energy xenon obtains electrons from the xenon atoms to become high-energy atoms (Xe) or fast atoms, and the (Xe) +∙ ions are removed by an electric field.

• The compound of interest is dissolved in a high-boiling viscous

solvent such as glycerol; a drop is placed on a thin metal sheet, and the compound is ionized by the high-energy beam of xenon atoms (Xe). Ionization by translational energy minimizes the amount of vibrational excitation, and this results in less destruction of the molecule.

Page 46: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

Matrix Assisted Laser Desorption Ionization (MALDI)

h

Laser

1. Sample (A) is mixed with excess matrix (M) and dried on a MALDI plate.

2. Laser flash ionizes matrix molecules.

3. Sample molecules are ionized by proton transfer from matrix:

MH+ + A M + AH+.

+20 kVVariable Ground Grid Grid

AH+

Sample plate

Page 47: Mass Spectrometry. Mass Spectrometry (MS) An analytical technique for measuring the mass-to-charge ratio (m/z) of ions in the gas phase. –Mass spectrometry

• In the MALDI procedure - used mainly for large biomolecules - the sample in a matrix is dispersed on a surface, and is desorbed and ionized by the energy of a laser beam.

• The MALDI procedure has been used recently in several variations to determine the molecular weight of large protein molecules - up to several hundred kDa.

• Peptide sequencing is another application.

• Matrix selection is critical and depends on the wavelength of the laser beam and on the nature of the sample.

• ESI (Electrospray ionization) and MALDI are the preferred procedures for large biopolymers.

Matrix Assisted Laser Desorption lionization (MALDI)