Maxum PD PA AP Maxum II Official Product Specifications Brochure A5E03012338001 Rev03 ECO# DP5A00199469 12/2016 A5E03012338001 Rev03 Specifications Overview 1 Maxum II Common Specifications 2 Airless or Airbath Oven 3 Modular Oven 4 Ovens 5 Detectors 6 Valve Specifications 7 Air Treater Specifications 8 Network Access Unit 9

Upload: others

Post on 21-Oct-2019

19 views

Category:

Documents


0 download

TRANSCRIPT

Maxum

PD PA APMaxum II Official Product Specifications

Brochure

A5E03012338001 Rev03ECO# DP5A00199469

12/2016A5E03012338001 Rev03

Specifications Overview 1Maxum II Common Specifications 2

Airless or Airbath Oven 3

Modular Oven 4

Ovens 5

Detectors 6

Valve Specifications 7

Air Treater Specifications 8

Network Access Unit 9

Legal informationWarning notice system

This manual contains notices you have to observe in order to ensure your personal safety, as well as to prevent damage to property. The notices referring to your personal safety are highlighted in the manual by a safety alert symbol, notices referring only to property damage have no safety alert symbol. These notices shown below are graded according to the degree of danger.

DANGERindicates that death or severe personal injury will result if proper precautions are not taken.

WARNINGindicates that death or severe personal injury may result if proper precautions are not taken.

CAUTIONindicates that minor personal injury can result if proper precautions are not taken.

NOTICEindicates that property damage can result if proper precautions are not taken.If more than one degree of danger is present, the warning notice representing the highest degree of danger will be used. A notice warning of injury to persons with a safety alert symbol may also include a warning relating to property damage.

Qualified PersonnelThe product/system described in this documentation may be operated only by personnel qualified for the specific task in accordance with the relevant documentation, in particular its warning notices and safety instructions. Qualified personnel are those who, based on their training and experience, are capable of identifying risks and avoiding potential hazards when working with these products/systems.

Proper use of Siemens productsNote the following:

WARNINGSiemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems. The permissible ambient conditions must be complied with. The information in the relevant documentation must be observed.

TrademarksAll names identified by ® are registered trademarks of Siemens AG. The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner.

Disclaimer of LiabilityWe have reviewed the contents of this publication to ensure consistency with the hardware and software described. Since variance cannot be precluded entirely, we cannot guarantee full consistency. However, the information in this publication is reviewed regularly and any necessary corrections are included in subsequent editions.

Siemens AGDivision Process Industries and DrivesPostfach 48 4890026 NÜRNBERGGERMANY

Document order number: A5E03012338001Ⓟ 12/2016 Subject to change

Copyright © Siemens AG 2016.All rights reserved

Table of contents

1 Specifications Overview...............................................................................................................................5

2 Maxum II Common Specifications................................................................................................................7

3 Airless or Airbath Oven...............................................................................................................................11

4 Modular Oven.............................................................................................................................................15

5 Ovens.........................................................................................................................................................19

6 Detectors....................................................................................................................................................21

7 Valve Specifications...................................................................................................................................25

7.1 Model 50 Valve Specifications...............................................................................................25

7.2 Model 20 Valve Specifications...............................................................................................26

7.3 Model 20 High Temperature Valve Specifications.................................................................26

7.4 Model 11 Valve Specifications...............................................................................................27

7.5 Model 11 Low Dead Volume Valve (LDV) Specifications......................................................27

7.6 Liquid Injection Valve.............................................................................................................28

8 Air Treater Specifications...........................................................................................................................29

9 Network Access Unit..................................................................................................................................31

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 3

Table of contents

Maxum II Official Product Specifications4 Brochure, 12/2016, A5E03012338001 Rev03

Specifications Overview 1The specifications in this document are provided to assist in product comparison and selection. See the product documentation for more details.

NoteSpecifications Are Application-Dependent

The specifications given in this document reflect the performance of the component when used in a Maxum II Gas Chromatograph as a typical system. Performance may vary between individual installations.

NoteSee Individual Documentation Package

Each analyzer is shipped with an Individual Documentation Package that includes drawings, specifications and certifications for the individual unit that supercede the information in this document.

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 5

Specifications Overview

Maxum II Official Product Specifications6 Brochure, 12/2016, A5E03012338001 Rev03

Maxum II Common Specifications 2Table 2-1 General

Temperature control ±0.02°C (±0.05°F)Calibration Comparison measurement with external standard● Type Manual or automatic● Zero value Automatic baseline correction● Span Standard sample cylinder (single or multipoint calibration possible)

Table 2-2 Measuring Response

Sensitivity (depending on applica‐tion)

±0.5% of span

Linearity (depending on application) ±2% of spanEffects of vibrations NegligibleRepeatability in % of full scale(Depending on application)

Range Value2 and 100% ±0.5%0.05 and 2% ±1%; 50 and 500 ppm ±2%5 and 50 ppm ±3%0.5 and 5 ppm ±5%

Table 2-3 Climatic conditions

Ambient temperature and humidity -18 to 50°C (0 to 122°F) application-dependentSpecific installations may require additional environmental constraints. See the custom documentation package.. Note: If the Maxum II is exposed to high condensing humidity with the electronics open or without dry purge air applied, then it must be allowed to re-stabilize at the above stated conditions for at least 8 hours before electrical power is applied.** Depending on application characteristics such as number of detectors, oven tem‐perature, and electronic loading, the acceptable ambient temperature range may be reduced. For application-specific detail, contact Siemens through the Siemens Industry Online Support (SIOS) website:https://support.industry.siemens.com.

Altitude Up to 2000m (6561') for analyzers using 230VAC Supply Up to 3000m (9842') for analyzers using 115VAC Supply

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 7

Table 2-4 Gas supply

Instrument air ● At least 350 kPa (50 psig) for units with Model 11 or Valco valves● At least 825 kPa (120 psig) for units with Model 50 valves● At least 175 kPa (25 psig) for airbath ovens; 85 l/min per oven● Instrument air quality requirement: ISO-8573-1:2010 [3:2:3]

[ 3 : - : - ] Dust filtering down to 1 micron [ - : 2 : - ] Dew point of -40°C [ - : - : 3 ] Total hydrocarbon below 1 mg per cubic meter (appr. 0.8 ppmw)

Carrier gas ● Hydrogen, Nitrogen, Helium, Argon, or Synthetic Air in compressed gas cylinder, purity 99.999%, or hydrogen with a purity of 99.999% (depending on application).

● Typical consumption quantity: 5 to 100 l/month per detector moduleCorrosion protection ● Purging with dry air to protect the electronics

● Airbath oven with stainless steel lining● Airless oven made of aluminum● Steel lining painted on outside (epoxy powder coating)

These are general parameters that vary depending on the individual application and configuration.

Table 2-5 Typical Gas Requirements

Type Quality Pressure Typical ConsumptionCarrier gas Hydrogen

NitrogenHeliumArgonSynthetic Air

5.0 (≥ 99,999%) 500 to 1000 kPa (72.5 to 145 psig)

50 to 150 ml/min/train

Combustion gas Hydrogen 5.0 (≥ 99,999%) 500 to 1000 kPa (72.5 to 145 psig)

30 to 50 ml/min/FIDApprox. 100 ml/min/FPD

Combustion air Synthetic air Hydrocarbon free 500 to 1000 kPa (72.5 to 145 psig)

Approx. 400 ml/min/FIDApprox. 100 ml/min/FPD

Actuation air Instrument air Purge air(py purge)

Instrument air or Nitrogen

Instrument air 400 to 1655 kPa (58 to 240 psig)

Initial purge: Approx. 65l/minOperating purge: Approx. 3.5 l/min

Purge air(px purge)

Instrument air or Nitrogen

Instrument air 400 to 600 kPa (58 to 87 psig)

Initial purge: 60 to 80 l/minOperating purge: 12 to 18 l/min

Valve gas(for Model 50)

Instrument air 500 to 1000 kPa (72.5 to 145 psig)

Oven heating (for Airbath oven)

Instrument air 400 to 600 kPa (58 to 87 psig)

85 l/min/oven

Vortex cooler Instrument air 300 to 600 kPa (43.5 to 87 psig)

130 to 170 l/min

PTGC Instrument air 400 to 600 kPa (58 to 87 psig)

210 l/min

Maxum II Common Specifications

Maxum II Official Product Specifications8 Brochure, 12/2016, A5E03012338001 Rev03

Type Quality Pressure Typical ConsumptionPTGC + Vortex Instrument air 400 to 600 kPa (58 to 87

psig)Approx. 575 l/min

Strip gas Nitrogen 5.0 (≥ 99,999%) 400 to 600 kPa (58 to 87 psig)

6 to 8 l/h

Table 2-6 Influencing variables

Effects of ambient temperature Negligable with electronic pressure control. Different effects with mechanical pressure control (depending on application)

Table 2-7 Mounting Clearance

Mounting clearance ● Clearance on left: 460 mm (18") from walls and other devices ● Clearance on right: 460 mm (18") in all cases ● Clearance at front: 654 mm (26") in all cases ● Center-to-center: 1,120 mm (44") in all cases

Maxum II Common Specifications

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 9

Maxum II Common Specifications

Maxum II Official Product Specifications10 Brochure, 12/2016, A5E03012338001 Rev03

Airless or Airbath Oven 3Table 3-1 Dimensions

Weight 77 kg (170 lbs.) application dependentOverall Dimensions(including mounting hardware)

Height: 1053.6 mm (41.5”)Width: 752 mm (26.6”)Depth: 417.4 mm (16.4”)

Figure 3-1 Maxum II Airless/Airbath Dimensions

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 11

Table 3-2 Configuration

Oven options ● Single isothermal airbath oven or divided airbath oven with two independent isothermal zones

● Single oven or two independent, airless ovens in separate oven areas.● Programmable-temperature (PTGC) oven

Detector options ● Thermal conductivity● Flame ionization● Flame photometry● Pulse discharge detector in Helium ionization mode● Pulse discharge detector in Photo-ionization mode● Pulse discharge detector in Electron capture mode

Number of detectors ● 1, 2 or 3 in any combination of detector module types for airbath ovens (max. 2 FPDs)

● 1, or 2 in any combination of detector module types for airless ovens, up to 3 in special configurations

Sampling and column valves ● Diaphragm valves● Diaphragm piston valves● Sliding vane rotary valves● Slider valves● Liquid injection valve

Valveless option Live switchingColumn options Packed, micropacked or capillary columnsRegulation of gas supply Up to 8 electronic pressure regulator channels and up to 6 mechanical pressure reg‐

ulators

Table 3-3 Enclosure Protection

Ingress protection IP54, Category 2Explosion protection ● Standard configurations:

● Certified according to ATEX with air or nitrogen purging for Zone 1 (II2G Ex ... IIB + H2 ... Gb)

● Suitable for use in non-hazardous areas and with non-dangerous conditions● Certified according to CSA C/US for use in Class 1, Div. 1, Groups B, C, D with air

or nitrogen purging ● Certified according to CSA C/US for use in Class 1, Div. 2, Groups B, C, D. Important note! Purging of the electronics area with instrument air or nitrogen is rec‐ommended to maintain optimum measurement performance. The PDD is not certified for hazardous areas.

EMI/RFI design ● CE-compatible; in conformity with 2014/30/EU (EMC directive) ● CE-compatible; in conformity with 2014/35/EU (low-voltage directive)● Tested according to EN 61010-1 / IEC 61010-1

Airless or Airbath Oven

Maxum II Official Product Specifications12 Brochure, 12/2016, A5E03012338001 Rev03

Table 3-4 Electrical characteristics

Power supply ● Single-phase AC, 100 to 130 V or 195 to 260 V (selectable), 47 to 63 Hz● Single oven: max. 1800 VA● Dual oven: 2 circuits, max. 1800 VA each

Table 3-5 Electrical inputs and outputs

Standard input and output ● 2 analog outputs● 4 digital outputs (1 for output of system faults, 3 are user-configurable)● 4 digital inputs;

Card slots for optional inputs and outputs via internal I2C bus

2

Input and output cards ● I2C_AIO: 8 analog outputs, 8 analog inputs, 2 digital inputs● I2C_DIO: 6 digital inputs and 8 digital outputs ● I2C_ADIO: 4 digital inputs and 4 digital outputs, 4 analog inputs and 4 analog

outputsDigital inputs ● Optically coupled with a common for all inputs.

● Self powered floating-contact input, or configurable for sinking or sourcing current. ● Sourcing current mode: 24 V internal isolated supply, with positive terminal of

supply at common. ● Sinking current mode: 5 V internal isolated supply, with negative terminal of supply

at common.Digital outputs Floating double-throw contacts, maximum contact load rating 1 A at 30 V (AC or DC).

External diode shunt suppression should be used for inductive DC loads, preferably at the load

Analog inputs -20 to 20 mA into 50 Ω or-10 to 10 V Rin = 0.1 MΩWith alternate insulation up to 10 V

Analog outputs 0/4 to 20 mA into max. 750 Ω, common negative pole, electrically isolated from ground; freelyconnectable to ground

Termination SYSCON-based I/O: Terminal strip for braided or solid cable with maximum section of 1.5 mm2 (16AWG) Expansion board-based I/O: Screw terminal for shielded or solid cable with a maximum area of 0.82 mm2 (18AWG)

Table 3-6 Communication

Serial ports 4 RS232/RS485 ports, e.g. ModbusEthernet ● Standard 10/100 BaseT Ethernet with 4 RJ45 connectors e.g. Modbus TCP IP or

OPC● Optional ESBF board Fiber-optic 100Base FX multimode with ST connection 3 x

RJ45 and 1 x optical

Table 3-7 Gas sample inlet conditions

Sample flow 5 to 100 ml/min (depending on application)Sample filter size 0.1 to 5 μm with gaseous samples (depending on type of valve)

Airless or Airbath Oven

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 13

Minimum sample pressure 35 kPa (5 psig) standardMaximum sample pressure 200 kPa (29 psig) standard, higher pressure as optionMaximum sample temperature 121°C (250°F) standard; higher temperature as optionMaterials in contact with sample Stainless steel and Teflon; other materials as option

Table 3-8 Combustion gas supply

Combustion gas ● Hydrogen with a purity of 99.999%● Typical consumption quantity: 2 000 l/month per detector module

Combustion air ● Reference air (< 1 ppm THC, O2 concentration 20 to 21%). Synthetic air, or supply through instrument air with catalytic purification (optional).

● Typical consumption quantity: 26 000 l/month

Airless or Airbath Oven

Maxum II Official Product Specifications14 Brochure, 12/2016, A5E03012338001 Rev03

Modular Oven 4Table 4-1 Dimensions

Weight 60 kg (132 lbs.) application dependentOverall Dimensions(including mounting hardware)

Height: 729.9 mm (28.75")]Width: 752 mm (26.6”)Depth: 415.9 mm (16.4")

Figure 4-1 Maxum II Modular Oven Dimensions

Table 4-2 Configuration (application dependent)

Oven options ● Single airless oven or two independent, airless ovens. ● Optionally small oven for one small analytical module, large oven for two small

analytical modules or one large analytical module. ● Two small ovens, two large ovens or any combination of 2 ovens is possible. ● Dual-oven version has two independent ovens with separate doors.

Detector module type Thermal conductivityNumber of detector modules ● One 4-cell TCD for small analytical modules

● One or two 4-cell TCD for large analytical modulesSampling and column valves ● One Model 50 diaphragm valve in small analytical module

● One, two, or three Model 50 diaphram valves in large analytical module

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 15

Columns Packed, micropacked or metal capillary columnsRegulation of gas supply Up to 6 electronic pressure regulator channels and up to 4 mechanical pressure reg‐

ulators

Table 4-3 Enclosure Protection

Ingress protection IP54, Category 1Explosion protection ● Standard configurations:

● Certified according to ATEX with air or nitrogen purging for Zone 1 (II2G Ex ... IIC ... Gb)

● Suitable for use in non-hazardous areas and with non-dangerous conditions● Certified according to CSA C/US for use in Class 1, Div. 1, Groups A, B, C, D with

air or nitrogen purging ● Certified according to CSA C/US for use in Class 1, Div. 2, Groups A, B, C, D. Important note! Purging of the electronics area with instrument air or nitrogen is rec‐ommended to maintain optimum measurement performance.

EMI/RFI design ● CE-compatible; in conformity with 2014/30/EU (EMC directive) ● CE-compatible; in conformity with 2014/35/EU (low-voltage directive)● Tested according to EN 61010-1 / IEC 61010-1

Table 4-4 Electrical characteristics

Power supply ● Single-phase AC, 85 to 264 V, 47 to 63 Hz ● Max. 655 VA, nominal 280 VA ● Optional: External 24 V DC ±10% 10 A with 32 V voltage limiting Max. 100 mV

residual ripple and interferences minimum to maximum at 20 MHz ● Fusing at max. 20 A ● External 24 V supply must accept minus to ground

Table 4-5 Electrical inputs and outputs

CIM-based IO ● 2 digital outputsCard slots for optional inputs and outputs via internal I2C bus

● 4 slots, maximum 2 IO cards (in IO tray)

Input and output cards ● I2C_AIO: 8 analog outputs, 8 analog inputs, 2 digital inputs● I2C_DIO: 6 digital inputs and 8 digital outputs ● I2C_ADIO: 4 digital inputs and 4 digital outputs, 4 analog inputs and 4 analog

outputsDigital inputs ● Optically coupled with a common for all inputs.

● Self powered floating-contact input, or configurable for sinking or sourcing current. ● Sourcing current mode: 24 V internal isolated supply, with positive terminal of

supply at common. ● Sinking current mode: 5 V internal isolated supply, with negative terminal of supply

at common.

Modular Oven

Maxum II Official Product Specifications16 Brochure, 12/2016, A5E03012338001 Rev03

Digital outputs Floating double-throw contacts, maximum contact load rating 1 A at 30 V (AC or DC). External diode shunt suppression should be used for inductive DC loads, preferably at the load

Analog inputs -20 to 20 mA into 50 Ω or-10 to 10 V Rin = 0.1 MΩWith alternate insulation up to 10 V

Analog outputs 0/4 to 20 mA into max. 750 Ω, common negative pole, electrically isolated from ground; freelyconnectable to ground

Termination SYSCON-based I/O: Terminal strip for braided or solid cable with maximum section of 1.5 mm2 (16AWG) Expansion board-based I/O: Spring contacts for shielded or solid cable with a maxi‐mum area of 0.82 mm2 (18AWG)

Table 4-6 Communication

Serial ports 2● Serial Port 1: RS232/RS485 selectable● Serial Port 2: RS485 mode only

Ethernet ● Standard 10/100 BaseT Ethernet with 4 RJ45 connectors e.g. Modbus TCP IP or OPC

● Optional ESBF board Fiber-optic 100Base FX multimode with ST connection 3 x RJ45 and 1 x optical or

Table 4-7 Gas sample inlet conditions

Sample flow 5 to 100 ml/min (depending on application)Sample filter size 0.1 μm with gaseous samplesMinimum sample pressure 35 kPa (5 psig) standardMaximum sample pressure 200 kPa (29 psig) standardMaximum sample temperature 80°C (176°F) standard; higher temperature as optionMaterials in contact with sample Stainless steel, aluminum, Viton, polyimide and Teflon

Modular Oven

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 17

Modular Oven

Maxum II Official Product Specifications18 Brochure, 12/2016, A5E03012338001 Rev03

Ovens 5Table 5-1 Oven Characteristics

Oven Temperature Range(application dependent)

Oven Type Temperature Range (dependent on T-rating)Airless* 50 to 260°C (122 to 535°F)Airbath* 5 to 225°C (40 to 440°F) Airbath PTCG* 5 to 330ºC (40 to 625°F)Modular 60 to 80°C (140 to 176°F)

Temperature Control ± 0.02°C (± 0.05°F)

* Note: 10°C (18°F) differential above ambient is required for control in all cases unless equiped with a vortex tube.

Table 5-2 Maximum power consumption (application and ambient temperature dependent):

Oven configuration WAirless single oven 500Airless dual oven 1000Airbath single oven 650 or 1400Airbath dual oven 2800Airbath PTGC 2800Modular oven system (24 VDC) Small oven (total) 60

Large Oven (total) 100

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 19

Ovens

Maxum II Official Product Specifications20 Brochure, 12/2016, A5E03012338001 Rev03

Detectors 6The detector specifications given in this section depend on the specific application in which they are installed, and reflect the performance of the component when used in a Maxum II Gas Chromatograph as a system.

Table 6-1 Thermal Conductivity Detector

Parameter ValueTarget Components AllMeasurement Type ConcentrationSmallest Measuring Range (at 60°C oven temperature) 0 to 20 ppm ±5%Typical Measuring Range (at 60°C oven temperature) 0 to 500 ppm ±2% FS

up to 0 to 100% ±0.5% FS

Dynamic Range 105

Linear Range 104

Selectivity NoneTemperature Range (Thermistor bead detector inside oven compartment) 5 to 120°C Temperature Range (Filament detector inside oven compartment) 5 to 180°C

Table 6-2 Flame Ionization Detector (FID)

Parameter ValueTarget Components CarbonMeasurement Type MassSmallest Measuring Range (at 60°C oven temperature) 0 to 1 ppm ±3%Typical Measuring Range (at 60°C oven temperature) 0 to 100 ppm ±1% FS

up to 0 to 100% ±0.5% FS

Dynamic Range 107

Linear Range 106

Selectivity Combustible ComponentsTemperature Range 105 to 260°C

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 21

Table 6-3 Flame Photometric Detector (FPD)

Parameter ValueTarget Components SulfurMeasurement Type MassSmallest Measuring Range (at 60°C oven temperature) 0 to 1 ppm ±5%Typical Measuring Range (at 60°C oven temperature) 0 to 10 ppm ±3% FS

up to 0 to 500 ppm ±2% FS

Dynamic Range 102

Linear Range 70 (70:1) linearizedSelectivity 104 to 1

Sulfur to CarbonTemperature Range 105 to 150°C

Table 6-4 Pulse Discharge Detector, Helium Ionization Mode

Parameter ValueTarget Components AllMeasurement Type MassSmallest Measuring Range (at 60°C oven temperature) 0 to 2 ppm ±5%Typical Measuring Range (at 60°C oven temperature) 0 to 5 ppm ±5% FSDynamic Range 103 at fixed attenuationLinear Range 103 at fixed attenuationSelectivity None without He and NeTemperature Range 70 - 110°C

(higher temperature ranges on request; system configuration dependent)

Table 6-5 Pulse Discharge Detector, Electron Capture Mode

Parameter ValueTarget Components Components with electron affinityMeasurement Type MassSmallest Measuring Range (at 60°C oven temperature) 0 to 2 ppm ±5%Typical Measuring Range (at 60°C oven temperature) 0 to 5 ppm ±5% FSDynamic Range 103 at fixed attenuationLinear Range 103 at fixed attenuationSelectivity Components with electron affinity to carbon

103 to 1Temperature Range 70 - 110°C

(higher temperature ranges on request; system configuration dependent)

Detectors

Maxum II Official Product Specifications22 Brochure, 12/2016, A5E03012338001 Rev03

Table 6-6 Pulse Discharge Detector, Photoionization Mode

Parameter ValueTarget Components Photon sensitive componentsMeasurement Type MassSmallest Measuring Range (at 60°C oven temperature) 0 to 2 ppm ±5%Typical Measuring Range (at 60°C oven temperature) 0 to 5 ppm ±5% FSDynamic Range 103 at fixed attenuationLinear Range 103 at fixed attenuationSelectivity Photon sensitive components

(example: benzene to hexane 103 to 1)Temperature Range 70 - 110°C

(higher temperature ranges on request; system configuration dependent)

Detectors

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 23

Detectors

Maxum II Official Product Specifications24 Brochure, 12/2016, A5E03012338001 Rev03

Valve Specifications 77.1 Model 50 Valve Specifications

Table 7-1 Configuration

Valve Type Double-acting, pneumatically operatedNumber of Ports 10Flow Paths On state: flow between ports 1 & 10, 2 & 3, 4 & 5, 8 & 9

Off state: flow between ports 1 & 2, 3 & 4, 5 & 6, 7 & 8, 9 & 10 Note: flow between ports 6 & 7 is always blocked

Table 7-2 Performance and Requirements

Sample Types Vapor samples only, free of particulatesActuation Gas Inert gas, particulate freeSwitching Time 20 ms (Typical in Maxum gas chromatographs.) In other uses, actual switching time

is dependent on the type of solenoids used to deliver the pneumatic signal to the valve and the size of the pneumatic tubing to the valve.

Sample Filtration 5 micronMaximum Actuation Pressure 690 kPa (100 psig); must be 140-175 kPa (20-25 psig) higher than maximum carrier

gas or sample gas pressureMaximum Port Pressure (Carrier or Sample)

515 kPa (75 psig); contact factory for higher pressure options

Minimum Port Pressure (Carrier or Sample)

35 kPa (5 psi)

Pressure Drop < 35 kPa (5 psi) for 200 scc/min (air) Pressure drop was measured in air at room temperature with flow from port to port. Actual pressure drop will vary with temperature and gas type. Pressure drop through the sample loop is dependent on the inside di‐ameter and length of the sample loop.

Maximum Temperature 120°C (248°F); contact factory for higher temperature optionsCarrier Gas Flow 0.3 ml/min to 40 ml/min; contact factory for higher flow optionsMean Time Between Failures >107 cycles at 60°C (140°F); >4×106 cycles at 150°C (300°F). Valve failure is typically

caused by particulates preventing the diaphragms from sealing against the center plate or long term wear resulting in deformation of the diaphragms.

Table 7-3 Dimensions

Mounting 2 clearance holes for mounting screws (mounting brackets available as separate spare part if desired)

Actuation & Sample Connections 1.6 mm (1/16") OD tubing with Valco® or Swagelok® fittingsDimensions Height: 33 mm (15/16")

Body Diameter: 38 mm (11/2")Outside Diameter: 50 mm (2")

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 25

Weight 0.25 kg (0.5 lb)Materials of Construction Body: 316 stainless steel

Diaphragms: Teflon® coated stainless steel

7.2 Model 20 Valve Specifications

Internal loop sample size: 2 μl standard. (Other sizes are available on special request.)External loop sample size: 170 μl standard. (Other sizes are available on special request.)Sample filter: 5 μm recommended (not furnished with valve).Maximum operating temperature: 122°C (250°F) standard Maximum sample pressure: 1380 kPa (200 psi) gage instrument air standard.

10350 kPa (1500 psi) gage with 2 80 550 kPa (80psi) gage air signals. Higher operating pressure available on special request.

Minimum sample pressure: 34.5 kPa (5 psig)Maximum leak rate between ports: Less than 1 μl per minute between ports.Material in contact with sample: 316 stainless steel and Teflon, standard. Other materials are available on special

request.Connection size and type: 1.6 mm (1/16") O. D. tubing, with Swagelok® fittingsActuating air: 345 kPa to 550 kPa (50 to 80 psig) instrument air standard. 550 kPa (80 psig)

maximum. Helium or nitrogen may be substituted for air. Total switching time: 150 ms (total time from initiation of command signal to completion of change in

valve state)Body size: 70 mm (23/4") diameter by 70 mm (23/4") heightMounting: Clamp typeOther: A carrier purge (helium, hydrogen, or nitrogen) may be used (an external low-

pressure gas supply is required).

7.3 Model 20 High Temperature Valve Specifications

Internal loop sample size: 2 μl standard. (Other sizes are available on special request.)External loop sample size: 170 μl standard. (Other sizes are available on special request.)Sample filter: 5 μm recommended (not furnished with valve).Maximum operating temperature: 150°C (300°F) at 3450 kPa (500 psig) maximum standard

234°C (450°F) at 690 kPa (100 psig) maximum for special applications Maximum leak rate between ports: Less than 1 μl per minute between ports.Material in contact with sample: 316 stainless steel. Other materials are available on special request.Connection size and type: 1.6 mm (1/16") O. D. tubing, with Swagelok® fittingsActuating air: 345 kPa (50 psig) instrument air standard. 550 kPa (80 psig) required for oper‐

ating pressure above 1380 kPa (200 psig). Helium or nitrogen may be substituted for air.

Total switching time: 150 ms (total time from initiation of command signal to completion of change in valve state)

Body size: 70 mm (23/4") diameter by 70 mm (23/4") height

Valve Specifications7.3 Model 20 High Temperature Valve Specifications

Maxum II Official Product Specifications26 Brochure, 12/2016, A5E03012338001 Rev03

Mounting: Clamp typeOther: A helium, hydrogen, or nitrogen purge may be used (an external supply of low

pressure helium is required).

7.4 Model 11 Valve Specifications

Internal loop sample size: 0.5, 10, 20 μl standard. Other sizes on special request.External loop sample size: 15 μl and larger. 170 μl is standard.Sample filter: 5 μm recommended (not furnished with valve).Cap screw (33) Torque Setting: 3.95±0.11 Nm (35±1 inch-pounds)Base Screw (12) Torque Setting: 0.73±0.11 Nm (6.5±1 inch-pounds)Maximum operating temperature: 122°C (250°F) standard

66°C (150°F) 2000627-014Maximum sample pressure: Up to 2068 kPa (300 psig). Up to 414 kPa (60 psig) 2000627-014Minimum sample pressure before it is neces‐sary to apply vacuum assistance to lower chamber:

34.5 kPa (5 psig)

Maximum leak rate between ports: Less than 1 μl per minuteMaterial in contact with sample: 316 stainless steel and TFE, standard. Monel and Teflon, or Hastelloy and

Teflon available on request. PFA available on request.Connection size and type: 1.6 mm (1/16") O. D. tubing, with Swagelok® fittingsActuating air: (Lower port on manifold) Max. 345 kPa (50 psig) instrument air standard. Helium or nitrogen may be

substituted for air. 20 ml (at standard temperature and pressure) of actuating air required for each valve operation.

Lower Seal Disc Purge Gas: (Upper 2 ports on the manifold)

5 cc/min of (typically) same gas as carrier

Total switching time: 150 ms (total time from initiation of command signal to completion of change in valve state)

Body size: 57.2 mm (21/4") diameter by 76.2 mm (3") heightMounting: Clamp type

7.5 Model 11 Low Dead Volume Valve (LDV) Specifications

Internal loop sample size: 2 and 5 μl standard. Other sizes on special request.External loop sample size: 15 μl and larger. 170 μl is standard.Sample filter: 5 μm recommended (not furnished with valve).Cap screw (33) Torque Setting: 3.95±0.11 Nm (35±1 inch-pounds)Base Screw (12) Torque Setting: 0.73±0.11 Nm (6.5±1 inch-pounds)Maximum operating temperature: 122°C (250°F) standard

66°C (150°F) 2000627-014Maximum sample pressure: Up to 2068 kPa (300 psig). Up to 414 kPa (60 psig) 2000627-014

Valve Specifications7.5 Model 11 Low Dead Volume Valve (LDV) Specifications

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 27

Minimum sample pressure before it is neces‐sary to apply vacuum assistance to lower chamber:

34.5 kPa (5 psig)

Maximum leak rate between ports: Less than 1 μl per minuteConstruction material in contact with sample: 316 stainless steel and Teflon, standard. Monel and Teflon, or Hastelloy and

Teflon available on request.Connection size and type: 1.6 mm (1/16") O. D. tubing, with Swagelok® fittingsActuating air: (Lower port on manifold) Max. 345 kPa (50 psig) instrument air standard. Helium or nitrogen may be

substituted for air. 20 ml (at standard temperature and pressure) of actuating air required for each valve operation.

Lower Seal Disc Purge Gas: (Upper 2 ports on the manifold)

5 cc/min of carrier (normally helium)

Total switching time: 150 ms (total time from initiation of command signal to completion of change in valve state)

Body size: 57.2 mm (2.25") diameter by 76.2 mm (3") heightMounting: Clamp type

7.6 Liquid Injection Valve

Table 7-4 Siemens Liquid Injection Valve

Vaporization temperature 60 to 350°C (140 to 662°F) depending on application and temperature classInjection volume 0.1 to 9.5 μlSample temperature -20 to +150°C (-4 to 302°F)Material in Contact with Sample Stainless steel, mat. no. 1.4571, Hastelloy, Monel, Rulon, Teflon or special materialsControl pressure 400 to 600 kPa (58 to 87 psig)Sample pressure Max. 6 000 kPa (870 psig), recommended 50 to 100 kPa (7.25 to 14.5 psig)Connections for pipe 3.14 mm (1/8") outer diameter

Table 7-5 Liquid sample inlet conditions

Sample Flow 5-20 cc/min (application dependent)Sample Filtration 0.3-5 micron (sample valve dependent)Minimum Sample Pressure 35 kPa (5 psig), lower pressure optionalMaximum Sample Pressure 2070 kPa (300 psig) standard; higher pressure optionalMaximum Sample Temperature 121°C (250°F) standard; higher temperature optional

Valve Specifications7.6 Liquid Injection Valve

Maxum II Official Product Specifications28 Brochure, 12/2016, A5E03012338001 Rev03

Air Treater Specifications 8Dimensions Approximately 210 x 210 x 300 mm (81/4" x 81/4" x 113/4") (w x d x h)

A minimum of 400 mm (153/4") total clearance is required in depth for removal of the bottom coverWeight 10.5 kg (23 pounds)Power requirements 115 or 230 VAC, 50 wattsInstallation Panel, rack, or wall mounted near the analyzerConstruction 300 Series Stainless Steel in contact with the treated air. Mounted in an explosion proof aluminum

enclosureInput air quality ● Meet or exceed ISO8573-4 (Class 3) requirements.

● Maximum total hydrocarbon content less than 1 ppm and ● Dew point -20°C (-4°F)

Input air flow ● 250 to 1500 ml/min (depending on requirements)Input pressure 34 to 1034 kPa (5 to 150 psig), depending on flow rate desiredWarm-up time Approximately 90 minutesTreated air purity Up to 95% of the trace hydrocarbons typically found in instrument quality air are removed by the

air treaterCatalyst life Catalyst is not consumed as part of the air treatment so the life is indefinite. However the catalyst

can be poisoned when exposed to sulfur compounds.Safety ● CSA certified for use in Class I, Division 1, Groups B, C, D Hazardous Locations for US and

Canada● ATEX certified for use in Zone 1 or 2, gas group IIB + H2 explosive atmosphere (II 2 G) (Ex d

IIB+H2 T5 Gb)● Temperature Class Rating: T5

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 29

Air Treater Specifications

Maxum II Official Product Specifications30 Brochure, 12/2016, A5E03012338001 Rev03

Network Access Unit 9Table 9-1 Dimensions

Mounting Options ● Rack-mount case● Wall-mount case

Weight 15 kg (33 lbs.)

Figure 9-1 Rack Mount Cabinet

Table 9-2 Rack Mount Cabinet Dimensions

Label Description Metric mm

U.S. inches

A Rack Height 177 6.97B Rack Total Width 483 19 C Rack Mounting Holes Spacing 468 18.43D Rack Depth Case Only 413 16.26E Rack Handle 174 6.83F Rack Depth with Handles 448 17.63 G Case Width 440 17.32

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 31

Figure 9-2 Wall Mount Cabinet

Table 9-3 Wall Mount Cabinet Dimensions

Label Description Metric mm

U.S. inches

A Wall Width 444 17.48B Wall Mount Holes Horizontal Spacing 396 15.59C Wall Mount Holes Vertical Spacing 450 17.72D Wall Height Including Cable Housing 611 24.04E Wall Height w/o Cable Housing 415 16.32F Wall Mount Total Depth 199 7.81G Wall Mount Depth Case Only 176 6.91

Table 9-4 Enclosure Protection

Explosion protection ● Standard configurations:● Suitable for use in non-hazardous areas and with non-dangerous conditions● Certified according to CSA C/US for use in Class 1, Div. 2, Groups A, B, C, D.

EMI/RFI design ● CE-compatible; in conformity with 2014/30/EU (EMC directive) ● CE-compatible; in conformity with 2014/35/EU (low-voltage directive)● Tested according to EN 61010-1 / IEC 61010-1

Table 9-5 Electrical characteristics

Power supply ● Single-phase AC, 100 ... 130 V or 195 ... 260 V (selectable), 47 … 63 Hz● 300 W● G fuse element per IEC 127-2, 4A rating, slowblow● Power wiring must conform to local installation codes and requirements. If

conformance to CE certified installation requirements is required, power cable must be shielded or installed in conduit.

Network Access Unit

Maxum II Official Product Specifications32 Brochure, 12/2016, A5E03012338001 Rev03

Table 9-6 Electrical inputs and outputs

Standard input and output ● 2 analog outputs● 4 digital outputs (1 for output of system faults, 3 are user-configurable)● 4 digital inputs;

Card slots for optional inputs and outputs via internal I2C bus

7

Input and output cards ● I2C_AIO 8: 8 analog outputs, 8 analog inputs, 2 digital inputs● I2C_DIO: 6 digital inputs and 8 digital outputs ● I2C_ADIO: 4 digital inputs and 4 digital outputs, 4 analog inputs and 4 analog

outputsDigital inputs ● Optically coupled with a common for all inputs.

● Self powered floating-contact input, or configurable for sinking or sourcing current. ● Sourcing current mode: 24V internal isolated supply, with positive terminal of supply

at common. ● Sinking current mode: 5V internal isolated supply, with negative terminal of supply

at common.Digital outputs Floating double-throw contacts, maximum contact load rating 1A at 30V (AC or DC).

External diode shunt suppression should be used for inductive DC loads, preferably at the load

Analog inputs -20 to +20mA into 50Ω or-10 to +10V Rin = 0.1MΩ, alternate insulation up to 10V

Analog outputs 0/4 to 20 mA into max. 750 Ω, common negative pole, electrically isolated from ground; freelyconnectable to ground

Termination ● SYSCON-based I/O: Terminal strip for braided or solid cable with maximum section of 16AWG (1.5mm2)

● Expansion board-based I/O: Spring contacts for shielded or solid cable with a maximum area of 18AWG or 0.82 mm2

Table 9-7 Communication

Serial ports RS 232, RS 485, e.g. ModbusEthernet ● Standard 10/100 BaseT Ethernet with 4 RJ45 connectors e.g. Modbus TCP IP or

OPC● Optional ESBF board Fiber-optic 100Base FX multimode with ST connection 3 x

RJ45 and 1 x optical

Network Access Unit

Maxum II Official Product SpecificationsBrochure, 12/2016, A5E03012338001 Rev03 33