mbaa-rocky mountain district technical summit 25 june 2010 measuring dissolved oxygen with optical...
TRANSCRIPT
MBAA-Rocky Mountain District Technical Summit
25 June 2010Measuring Dissolved Oxygen with Optical Technology
Brian VaillancourtMettler-Toledo Ingold
Bedford MA 2010
2
Agenda
Introduction
Current Technology
Challenges with DO measurements
Oxygen Measurement in Breweries
New Optical Technology
Theory of Operation
Benefits
Summary
4
Current DO Measurement
Dissolved oxygen measurement in Breweries is predominantly amperometric
- Proven technology
- Technology offered by a multiple manufacturers
- Extensive portfolio for a wide application coverage
- Wide temperature range
- CIP & Sterilizable
- Accurate at low oxygen levels
There are challenges with amperometric technology:
- Process conditions can damage the membrane
- Speed of response from saturation values is slow
- Flow dependences
- The high Impedance measurement makes it susceptible to moisture problems
Optical DO measurement offers a solutions for these challenges with amperometric technology
Today
But
5
Why Measure Oxygen – Key to Quality
Requirements to oxygen measurement equipment:- Ability to measure in beer
- Low limit of detection
- Stable measurement signal
- No flow dependence
- Fast response
- Low maintenance
Reduction of oxygen level in beer is directly linked to product quality and shelf life cost savings
Oxygen is considered one of the top beer spoilers
Oxygen in beer reduces the shelf life
The lower the DO when the product is packaged, the longer it will remain “Fresh, Crisp & Clean tasting”
DO in the beer before filling, contributes to nearly 1/3 of the total packaged oxygen “TPO”
Key requirements for successful oxygen control:- Avoid any ingress of oxygen at all process stages
- Increasing demand for lower oxygen value in water and CO2
6
Oxygen Measurements in the Brewery
water preparation mash tun lauter tun wort copper whirlpool
wort cooler
yeast propagationstorage tank fermentation tank
CIP stations
separator
Kieselguhr filter
PVPP filter
water deaeration
filling lines
waste water treatment
bright beer tank
DO
DO
DO
O2
DO DO
DO
DO
DO DO
Fermentation/Storage
Brew house
Filtration/Filling
DO DO
7
Challenges for new Technology
Reliable Measurement
-Robustness-Ease of use
Maintenanceplanning
-Diagnostics
The measurement works
- Accuracy- Reliability
8
Keep your focus
Optical DO systems allow you to concentrate on your process
9 Internal usage only
Every Day Challenges
85%AirIs this value
correct
How long will it be correct?
What happens if the instrument has a
problem?
When do I have to perform recalibration /
maintenance?
Do I have all Information?
Will I be informed about the
status or problems? Is the instrument
o.k.?
10 % Air!
ISM offers a comprehensive approach for reliable measurement andmaintenance planning
5 Internal usage only
Challenges for new Technology
Reliable Measurement
-Robustness-Ease of use
I can planmaintenance
-DLI-ACT
-Diagnostics
The measurement works
- Accuracy
18 Internal usage only
Simplified SOP through Optical DO
1. Detach cap sleeve
2. Detach membrane body
3. Dispose electrolyte
4. Clean or replace membrane body
5. Clean electrode
6. Fill in electrolyte
7. Bubble free installation of membrane body
8. Clean outside
9. Install cap sleeve
10.Polarize sensor (6h)11.Calibration
1. Detach cap sleeve
2. Detach OptoCap
3. Install new OptoCap
4. Install cap sleeve
5. Calibration
Optical sensors offer higher operational availability
Total: more than 6 hours Total: few minutes
Today's standard Optical Systems
9
What is Fluorescence Quenching?
Fluorescence quenching is the basic principle of the optical oxygen measurement
Fluorescence is a phenomenon where a material absorbs light (energy) of a specific wavelength (color) and after a short time emits light with a different wavelength (color)
Fluorescence quenching describes a reduction of the fluorescence intensity and a time shift caused by another substance (quencher, e.g. oxygen).
The quenching depends on the amount of oxygen present in the process solution. The oxygen is quantified by measuring the time shift.
O2
O2
O2
LED
DetectorEmitted fluorescence light
Opto-Layer
Sensor tip
10
Partial Pressure and Dissolved Oxygen
Henry’s Law states “ The partial pressure of a gas in a liquid is
equal to the partial pressure of the gas in the vapor above the
liquid.”
The sensors deliver information which is proportional to the oxygen partial pressure in the liquid. This information is translated by the transmitter into % saturation, mg/l or ppm
Transmitter100%
Partial PressureO2 in Air
Partial PressureO2 in Liquid
Equilibrium
11
PAir =
760 mm Hg
PAir =
1580 mm Hg
System Pressure =
760 mm Hg
System Pressure =
1580 mm Hg
The Dissolved Oxygen concentration in solution changes with change in partial pressure.
The user must compensate for changes in pressure to ensure an accurate measurement
Partial Pressure
12
Tank Pressure
Tank hydrostatic pressure has virtually no influence on DO measurement up to 100 meters depth. (<1.0%)
10M
PAir =
760 mm Hg
PAir =
760 mm Hg
System Pressure =
760 mm Hg
System Pressure =
1580 mm Hg
13
Partial Pressure
Partial Pressure is the pressure that a single gas exerts in a mixture of gases
- Oxygen is 160 mm or 212.2 mBar at saturation
Humid Air displaces the Partial Pressure of Oxygen
- Example At 20oC 0% Humidity the Partial Pressure of Oxygen is 212.2 mbar
23.3 mbar Humidity the Partial Pressure of Oxygen is 207.4 mbar
4.8 mbar or 2.26% Difference between the Partial Pressure of Oxygen in Dry Air vs Humid Air
14
Amperometric Sensor
Teflon
Silicone
S.S. Mesh
ElectrolyteLayer
Teflon
Cross Section of the Electrode Tip
(not to scale)
15
Theory of Operation of Amperometric Sensors
O2 diffuses through the gas- permeable membrane (the higher the partial pressure in the liquid, the more O2 diffuses)
O2 is dissolved in the electrolyte
O2 is reduced at the cathode
The oxidation-reduction reaction generates a current
The current is measured by the transmitter and converted
1
2
3
1
23
20
0
0.2
0.4
0.6
0.8
1
0% 20% 40% 60% 80% 100% 120% 140%
Oxygen – Fluorescence Relation
Due to the nonlinearity of the sensor signal, accurate calibration is essential for high accuracy
The decay time and therewith the delay time (phase-shift) of the fluorescence light is directly related to the concentration of oxygen (quencher). But the shape of the function is not linear like amperometric and follows the so call Stern-Volmer equation.
O2 (Air concentration)
De
lay
tim
e
Optical
21
AmperometricOptical
Calculated non linear signal Sensor-specific calibration Calibration necessary because ageing
of the sensor influences the whole
calibration curve Two-point calibration (Air & Zero)
Linearity between nA and Oxygen value Direct information from the raw signal Offset or slope correction possible
because ageing prevailing influences
the slope
0
0.2
0.4
0.6
0.8
1
0% 20% 40% 60% 80% 100% 120% 140%
De
lay
tim
e
O2 (Air concentration)
0
0.2
0.4
0.6
0.8
1
0% 20% 40% 60% 80% 100% 120% 140%
Se
ns
or
Cu
rre
nt
(nA
)
O2 (Air concentration)
Optical vs. Amperometric Technology
22
Technology Polarographic OpticalDetection limit 1-3ppb 1-3ppbAccuracy ± (1ppb +1% of the
reading)± (2ppb +1% of the reading)
Response Time
T98 (Air – N2)
< 90 s < 20 s
Temperature during measurement
-5 to 80°C -5 to 60°C
Pressure during measurement
9 bar 12 bar
Material wetted parts 316L stainless steel
Membrane: Silicone /Teflon
316L stainless steel:
Sensor: Silicone
Optical vs. Amperometric Technology
23
Key Enhancements / Improvements: SOP
1. Detach cap sleeve
2. Detach membrane body
3. Dispose electrolyte
4. Clean or replace membrane body
5. Clean electrode
6. Fill in electrolyte
7. Bubble free installation of membrane body
8. Clean outside
9. Install cap sleeve
10.Polarize sensor (6h)
11.Calibration
1. Detach cap sleeve
2. Detach OptoCap
3. Install new OptoCap
4. Install cap sleeve
5. Calibration
Optical sensors offer higher operational availability and improves handling safety
Total: more than 6 hours Total: few minutes
Today's standard Optical Systems
24
Key Enhancements / Improvements
Time consuming sensor verification is replaced by enhanced self testing of the whole measuring system
Performance Check-Time consuming controlling and
documentation Response time Air and zero current Slope Drift
Automated Self Test-Communication
-Electronic component
-Optical component
-OptoCap quality
Total: about 30 minutes Sensor status directly available without additional testing
Today's standard Optical Systems
25
Sensor Performance: Response Time
Optical
Amperometric
60 Seconds0
5
10
15
20O
2 / p
pb
The response time in liquid phase of the optical is 50% faster than amperometric systems leading to higher efficiency
DeaeratedWater
Beer
26
Sensor Performance: Response Time
Optical
Amperometric
< 1 Minute
O2
/ ppb
The response time after a CIP cycle using non-degassed water is significantly shorter for optical sensors
30 Minutes
400
200
0
50
2000
Water Beer
27
Sensor Performance: During No Flow
Optical
Amperometric
Time / h
0
2
4
6
8
10O
2 / p
pb
The optical sensor shows no significant stop-flow effect leading to reduced alarm frequency
Flow Stop
1 2 3 4 5 6
28
Sensor Performance: During No Flow
Optical Sensor
Flow Stopped
Process conditions that affect the operation of amperometric sensors are not affected with the optical sensor
Optical sensors will show actual DO in process which is difficult to accept
Which results in blaming the instrumentation and not dealing with actual oxygen ingress
29
Sensor Performance: Extensively Tested Multiple optical system manufacturers were tested
Test included amperometric technology
The test period lasted 14 months
30
Sensor Performance: Other Benefits
Not susceptible to Hydraulic Shocks (measurement Stable)
No Damage from Hydraulic Spikes (Press-Vac)
Does not see CO2 bubbles as O2- Only responds to the presents of O2
Process Orientation of sensor is not important- Does not contain an electrolyte
Opto Cap life expectancy is 12+ Months- Easily replaced onsite and recalibrated
Does not require frequent “calibration”, but only “validation” - Verification has been necessary to become comfortable with this
new technology
31
Sensor Performance: Not without issues
Optical spot can not be pulsed during CIP process or at high temperatures- Results in a shift in the calibration values
- Most manufacturers deal with this issue by turning off the LED by a temperature shut-off or remote signal to the transmitter
More frequent pulse rate will deplete (bleach out) the optical spot at a faster rate- The pulse rate can be programmed
Multiple/Frequent (weekly) process calibrations will eventually require a two-point calibration be done- Drift rate is less than 1 ppb per month
32
Summary
Optical oxygen measurement systems
Provide
- Signal stability
- Faster Response time
- Extensively less maintenance then amperometric systems
- Ease of maintenance
- Process improvement
- Improved product quality