measuring absorption and scattering properties for spider ... · the cosmic microwave background...
TRANSCRIPT
1
Measuring Absorption and Scattering
Properties for Spider Filters
Kristen McKee
Project Advisor: Professor John Ruhl
Physics Department
Case Western Reserve University
May 2, 2014
2
Table of Contents
Executive Summary…………………………….…….3
Background……………………………………….….3
Introduction…………………………………………..4
Objectives…………………………………………….7
Review of Previous Work……………………………..7
Design and Experimental Set-up……..........................11
Methods……………………………………………... 16
Procedure and Data Analysis…………………………19
Results…………………………………………..…….29
Conclusion…………………………………………....31
Sources……………………………………………….32
3
Executive Summary
The Cosmic Microwave Background (CMB) is a blackbody of thermal radiation leftover
from the Big Bang, which today is shifted into the range of microwave photons. The Spider
experiment is designed to detect a very small polarization signal in the CMB, which would
signify the existence of gravity waves in the early universe. Since the experiment is only trying
to detect in certain bands of microwave photons (centered at 90 GHz and 150 GHz), it contains
many filters which screen out photons outside of these ranges.
Filters ideally transmit all photons within the desired microwave frequency bands to the
detector; however, this is not always what happens. Some photons can be absorbed, reflected, or
scattered by the filter rather than being transmitted to the detector. If the fraction of scattering,
absorption, and reflection of these photons is too high, it can cause problems in the experiment.
In this project, I built and debugged a novel test setup to enable measurements of absorption,
reflection, and scattering properties of Spider filters. I then tested two different Spider low pass
filters in order to measure the fraction of incident photons which were scattered, absorbed, and
reflected by each of the filters.
Background
In the very early universe, baryonic matter formed a hot, ionized plasma. However, as the
universe cooled and matter became neutral, photons were scattered from electrons less and less
frequently. After their last scattering, photons became free to move throughout the universe
4
independently from baryonic matter. The photons from the time of last scattering which we
observe today (red-shifted to microwaves) compose the Cosmic Microwave Background (CMB).
The CMB was first measured by Penzias and Wilson in 1965 [1]. Since this discovery,
many experiments have engaged in measuring various properties of the CMB. In 1992, the
COBE satellite first measured small anisotropies in the intensity of the CMB and showed that the
power spectrum of the CMB was very close to that of a uniform blackbody [2]. This
measurement supported the theory of inflation in the early universe. The experiments WMAP [3]
and Boomerang [4] both measured the temperature anisotropies in the CMB, providing precise
measurements of the angular power spectrum. Measurements of these anisotropies are significant
because they provide information about important cosmological parameters. The goal of the
SPIDER experiment is to detect a small polarization signal in the CMB which would signify the
existence of gravity waves in the early universe.
Introduction
Optical filters are essential components of CMB experiments such as Spider. A filter is a
device which blocks out photons in specific frequency ranges. Since the Spider experiment is
measuring microwave photons, it has detectors which are sensitive to bands centered at 90 GHz
and 150 GHz, with 20% bandwidth. The goal of the filters is thus to screen out all photons which
do not lie in the frequency bands around 90 and 150 GHz and to transmit to the detector all
photons which are in these bands.
5
One challenge involved in using filters is that they can scatter, reflect, and absorb some
of the incident 90 and 150 GHz band photons rather than transmitting them straight to the
detector (see diagram below). Reflected photons return to the direction of incidence after
interacting with the filter. Scattered photons pass through the filter but in doing so the filter
scatters them in a direction away from the detector. Absorbed photons are absorbed by the filter
and thus do not get transmitted through it at all. The absorption, scattering, and reflection are
problems because the polarization signal due to gravity waves which Spider is attempting to
detect is approximately 100 nK, while the CMB is approximately 2.73 K. Thus, since this signal
is extremely small compared to the CMB, it is very undesirable to lose any of this tiny signal. In
addition, Spider consists of six telescopes, each of which contains three filters. Thus, with a total
of 21 filters, it is very important that the percentage of photons lost through interaction with each
one is very small.
Figure 1: Ideally, all incident photons within the desired frequency bands are transmitted
to the detector. However, in reality, some of the incident photons are reflected back in the
6
direction of the source, some photons are absorbed by the filter, and some photons are scattered
away from the detector when they pass through the filter.
Methods have already been developed for determining the transmission and reflectance
of these optical materials. One can test transmission by shining microwave photons at the
material and aligning a detector at the opposite side of the material. Similarly, reflectance can be
determined by shining microwave photons onto the material and aligning a detector on the same
side of the material in order to measure the percentage of incident photons which are reflected
back towards the source.
Although the methods for determining transmission and reflectance properties are fairly
straightforward, determining the absorption and scattering properties of these devices has always
posed more of a problem. While one can use measurements of transmission and reflectance
properties to infer the amount of scattering plus absorption together, it is complicated to
differentiate between scattered photons and absorbed photons. This is because absorbed photons
stay in the material and thus cannot be directly measured. The scattered photons, on the other
hand, consist of all photons which make it through the window/ filter material but not to the
detector. Consequently, they are projected onto such a wide area of space that it is very difficult
to make precise measurements of them [5]. For this project, I am building, debugging, and using
a novel test setup to enable measurements of absorption and scattering properties for two Spider
low pass filters.
7
Objectives
The first objective was to use SolidWorks to design a test setup for measuring the
absorption, scattering, and reflection of millimeter waves from various optical devices. This set-
up consisted of a scatterometer built in the lab as well as a thermal blackbody load for a source
and a test dewar containing the detector. The next goal was to assemble and debug the test setup
and use it to make measurements of absorption, reflection, and scattering properties for Spider
filters. After measuring the desired optical properties, the next objective was to write Matlab
code to analyze the data and determine the amount of absorption, scattering, and reflection
caused by each of the optical materials that are tested. Data was taken for several optical devices,
but the data analysis was eventually focused on two different Spider low pass filters.
Review of Previous Work
Many people in the Spider collaboration were concerned that there was too much
uncertainty in the amount of absorption and scattering from the Spider windows. The entire
Spider experiment resides inside of a cryostat under vacuum. The purpose of the windows is to
provide a way for photons to enter the cryostat. Thus, they must be composed of materials which
are strong enough to hold vacuum but are also microwave-transparent. For Spider the windows
are composed of ultra-high-molecular-weight polyethylene (UHMWPE). Spider collaboration
members at Caltech came up with the idea to test the scattering and reflection properties of the
Spider windows and other optical devices using a test baffle, or hollow metal cylinder, as a
scatterometer. In this setup, the test baffle is attached to the window of a test dewar. Underneath
the baffle is a thermal cold load. Photons sourced by the cold load travel through the baffle into
8
the dewar and are focused into a detector using a mirror. A diagram of this set-up is shown
below.
9
Figure 2: A diagram of the set-up that the Caltech Spider collaborators used to test the
absorption, reflection, and scattering caused by the Spider ultra-high-molecular-weight
polyethylene (UHMWPE) windows. This set-up consists of a thermal blackbody load which
serves as the source, a “baffle,” or hollow cylinder, for photons to travel through into the dewar,
and a test dewar which contains the detector.
Leaving the inside of the baffle shiny will cause scattered photons to be reflected off the
sides of the baffle and thus remain in the baffle. However, blackening the inside of the baffle will
cause scattered photons to be absorbed by the side of the baffle, and hence they will not make it
to the dewar and will not be detected. By performing this test with the baffle both shiny and
blackened, one can then separate the scattered photons from the absorbed photons, thus
achieving the goal of determining these properties.
Spider Collaboration members at Caltech used this principle to determine the scattering
and absorption for a few of the optical devices used in Spider. They first tested a shader, which is
a type of filter. They found a value of absorption minus reflection of around 0.3 %, and minimal
scattering from the shader. They also tested a Spider hot pressed filter and found that absorption
minus reflection was around 0.8%, which indicated some amount of absorption and reflection.
However, scattering was still minimal. They used this same procedure for testing an ultra-high-
molecular-weight polyethylene (UHMWPE) Spider window and found that absorption minus
reflection was around 0.7%, but there was a negligible amount of scattering. Similar tests of
foam windows provided inconclusive results, as there were large variations among each of the
measurements [6].
10
These results from the Caltech Spider collaborators proved that using this test set-up was
a possible way to find out information about absorption and scattering properties for various
optical devices. However, their set-up was thoroughly debugged, as it was only constructed
quickly to see if the concept would work. In addition, measurements were not repeated
extensively. They also had some problems including condensation on the thermal load, which
caused some of the signal to be absorbed.
The goal of my project was to create and refine a similar test set-up which would enable
systematic testing of the absorption, reflection, and scattering properties for various optical
devices. This refined set-up was conducive to systematic repetition of measurements to allow for
more accurate results. In addition, I also took extra measurements using a chopper wheel in order
to take into account a possible spillover of the main beam.
In my setup, I was also able to separate the percentage of absorption, reflection and
scattering caused by the filter of interest. This is important because each of these types of
photons can create different, undesirable effects. For example, reflected photons can be reflected
back into the focal plane of the detector and cause a false signal. In addition, since Spider
observes very close to sources of polarization contamination, such as the galaxy, it is necessary
to limit the amount of light that is observed at these wide angles on the sky. Scattered photons
are a problem because they contribute to this signal when they are scattered to wide angles.
11
Design and Experimental Set-Up
The set-up for this experiment consisted of a scatterometer built in the lab as well as a
thermal blackbody load for a source and a test dewar used for photon detection. A diagram of the
full set-up is shown below.
12
Figure 3: A diagram of the experimental set-up. It consists of a thermal blackbody load,
constructed out of HR-10, a microwave absorbing material. The scatterometer consists of a
“baffle” (hollow cylinder) which is covered in either shiny (aluminum) or black (HR-10)
material, and a shelf to put the filter on. The large blue device is a cross section model of the test
dewar. The entire dewar was kept under vacuum. The cylinder at the bottom of the dewar is the
window, which holds vacuum but allows microwave photons to pass through it. The test dewar
also contains an elliptical mirror (left) for focusing photons into the detector (right).
In this experimental set-up, the thermal blackbody load serves as a source of incident
photons. Measurements were made with this load functioning as a 77 K blackbody and again as a
300 K blackbody. In order to create a 300 K blackbody, a large square piece (approximately 1’x
1’) of HR-10 (a microwave-absorbing material) was placed under the baffle at room temperature.
In order to create a 77K blackbody, a small rectangular bucket lined with HR-10 was filled with
liquid nitrogen and placed directly under the baffle.
The scatterometer consisted of a hollow cylinder (called a “baffle”) and a shelf on which
to rest the filter so that it remained directly above the baffle opening during measurements.
Measurements were taken with the baffle being both shiny and then black. “Shiny” means that
the inside of the baffle is aluminum and thus all photons which scatter to the side of the baffle
will be reflected and continue to propagate within the baffle interior. On the other hand, “black”
means that the inside of the baffle was covered in HR-10, and thus all microwave photons which
were scattered to the side of the baffle were absorbed by this material and thus ceased to
propagate within the baffle. Two separate baffles (one shiny and one black) were constructed for
this purpose, and thus it was necessary to detach and change the baffle when switching between
13
the shiny and black scenarios. This entire scatterometer apparatus was then directly connected to
the window of the test dewar.
The next part of the set-up is the dewar, which contains the detector and an elliptical
mirror for focusing photons into the detector. The detector used for these measurements only
functions at very low temperatures. It is cooled on a cold stage by a closed cycle He4/He3
refrigerator to around 280 mK. In order for this to be possible, it is necessary to keep the entire
dewar cold. As a consequence, it is also necessary to keep the entire dewar under vacuum. This
is because keeping the dewar under vacuum means that there will be no air to conduct heat
between the different temperature stages in the dewar, and thus it is easier to maintain the
temperature of the cold stage that the detector rests on. Another reason why it is important to
keep the dewar under vacuum is to prevent the gases which compose air, such as N2, CO2, and
O2, from condensing inside the cryostat and possibly damaging equipment such as wires or the
detector.
The detector used in these measurements is called a bolometer. It has a resistance which
is dependent upon the temperature of the detected photons. A pre-amp followed by a lock-in
amplifier is used to measure the resistance of the bolometer. The bolometer is also hooked up in
series with two large bias resistors which are then hooked up to the output voltage. As the
bolometer changes temperature, this will affect the output voltage of the lock-in amplifier, which
is the quantity that is being measured in this experiment [7].
15
In order to choose what bias voltage to use for the measurements, a load curve was
created by systematically changing the bias voltage and recording the resulting output voltage
changes. The bias voltages used were 0.1, 0.2, 0.4, 0.6, 0.8, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5,
and 5 Volts. The output voltage was recorded at each of these bias voltages from 0.1 V to 5 V,
and then they were repeated on the way down from 5 V to 0.1 V. Load curves at 77 K and 300 K
were graphed using Matlab, and the bias voltage of 0.6 V was chosen for the measurements
because this voltage was close to maximizing the output voltage signal for all load curves. An
example for such a curve is shown below in figure.
16
Figure 5: Output voltage as a function of bias voltage. Each vertical line marks a change in bias
voltage. The values used for bias voltage increased through 0.1, 0.2, 0.4, 0.6, 0.8, 1.0, 1.5, 2.0,
2.5, 3.0, 3.5, 4.0, 4.5, and 5 Volts, and then decreased back down from 5 V to 0.1 V in these
same step increments.
Methods
Combinations
In order to determine the amount of absorption, scattering, and reflection, it is necessary
to take several different sets of measurements with different combinations of baffle interior,
blackbody load temperature, and the filter in the shelf as well as outside of the shelf. The eight
different combinations of these set-up possibilities are as follows:
Configuration Baffle Interior Load Temperature Filter
A Shiny 77K In Shelf
B Shiny 77K Out of Shelf
C Black 77K In Shelf
D Black 77K Out of Shelf
E Shiny 300K In Shelf
F Shiny 300K Out of Shelf
G Black 300K In Shelf
H Black 300K Out of Shelf
17
Table 1: A list of each of the combinations used in the set-up. These consist of all the possible
combinations of using either using a black or shiny baffle, a 77K or 300K load, and the filter
being either in or out of the shelf.
For each of these combinations, the final temperature measurement from the detector
depends on the amount of absorption, scattering, reflection, and the fraction of the main beam
that is scattered to the baffle. These properties can be modeled for each combination by adding
the temperature contributions from each of these terms. For example, when the filter is out of the
shelf, the total temperature will have no contribution from the terms involving absorption,
reflection, or scattering, since there is no filter in place to cause these to occur. Thus, the
temperature when the filter is out of the shelf will only depend on how much of the main beam is
spilled over to the baffle. However, when the filter is in the shelf, there will be a term for the
scattering, absorption, and reflection in addition to the spillover to the baffle.
Each configuration shown in the above table will have a different equation relating the
detected temperature to each of the terms for absorption, reflection, scattering, and spillover of
the main beam. This is because the photons interact differently with the set-up depending on the
temperature of the load, the material of the baffle, and whether the filter is in the shelf or not. In
reality, the thermal blackbody load serves as the source for these measurements, and photons are
detected inside of the dewar by the detector. However, it simplifies the problem to work in the
time-reverse case, where the beam is originating from the detector and propagating through the
filter and baffle to the load.
In this time-reversed outlook, when the baffle is shiny, the scattered photons will have the
same temperature as the load because when they hit the side of the baffle, they will be reflected
18
into the load. However, when the baffle is black, the scattered photons will always have a
temperature of 300 K since they will be absorbed in the baffle wall. The absorbed photons will
always have a temperature of 300 K since they are absorbed by the filter, which is at room
temperature (300 K). The reflected photons will always have a temperature that is the same as
the temperature of the dewar, since they are reflected from the filter back into the dewar. Below
is a table of all of the terms that contribute to the final temperature reading for each
configuration. The measured temperature of each configuration is equivalent to the terms in each
row being summed up.
Configuration Main Beam Main Beam
Spillover to Baffle
Absorption Reflection Scattering to
Baffle
A 77*m*(1-a-r-s) 77*(1-m)*(1-a-r-s) 300*a r*Tdewar 77*s
B 77*m 77*(1-m) 0 0 0
C 77*m*(1-a-r-s) 77*(1-m)*(1-a-r-s) 300*a r*Tdewar 300*s
D 77*m 77*(1-m) 0 0 0
E 300*m*(1-a-r-s) 300*(1-m)*(1-a-r-s) 300*a r*Tdewar 300*s
F 300*m 300*(1-m) 0 0 0
G 300*m*(1-a-r-s) 300*(1-m)*(1-a-r-s) 300*a r*Tdewar 300*s
H 300*m 300*(1-m) 0 0 0
Table 2: The temperature contribution of each term to the total (detected) temperature for each
configuration. The configuration labels refer to the combination that they were defined as in
19
Table 1. The terms which contribute to the final temperature are the main beam fraction, the
main beam spillover to the baffle, the absorption, the reflection, and the scattering terms. m, a, r,
and s are the fraction of the incident photons which are in the main beam, absorbed, reflected, or
scattered, respectively. Tdewar is the temperature of the dewar.
Equations Found
In order to get a meaningful relationship between the variables m, a, r, and s, I found the
difference between several of the different combinations (A through H in Tables 1 and 2). The
equations I found relating each of the variables are shown below
r = (F-E) / (300 - Tdewar) [Eq. 1]
a = 1 – r – (E-A) / (F-B) [Eq. 2]
m = 1 – (D-B) / 223 [Eq. 3]
s = (-1 / m) * (((D-C) - (B-A)) / 223 - (1-m)(a+r)) [Eq. 4]
Procedure & Data Analysis
Calibration
Each of the expressions shown in Table 2 has units of temperature. However, the signal
that was read out of the lock-in amplifier had units of voltage. In order to equate the actual
voltage measurements to these equations for the temperature, it was necessary to find a
relationship between the voltage of the lock-in amplifier and the corresponding temperature. In
order to do this, I took several load curves. A load curve is a measurement in which the baffle
20
was shiny, there was no filter in the shelf, the load was a constant temperature, but the bias
voltage was systematically changed and the resulting output voltages were recorded. The bias
voltages used were 0.1, 0.2, 0.4, 0.6, 0.8, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 4.5, and 5 Volts. The
output voltage was recorded at each of these bias voltages from 0.1 V to 5 V, and then they were
repeated on the way down from 5 V to 0.1 V.
I took a load curve with a 77 K load and another one with a 300 K load (see Figure 5).
Then, since I had chosen to bias all of my measurements at 0.6 V, I took the average of the
voltage during the 0.6 V bias sections for each of the load curves. Then I found the calibration
factor between temperature and voltage by assuming a linear dependence of voltage on
temperature and calculating the ratio C = ΔV/ΔT. ΔT is the difference in temperature between
the two load curves. In this case ΔT = 300 K – 77 K = 223 K. ΔV is the difference in the voltage
that was measured with 0.6 V bias for each of the load curves. In other words, ΔV = V(at 300 K,
bias 0.6 V) – V(at 77 K, bias 0.6 V). Once I found this conversion factor, it was possible to
directly translate the voltage measurements I took into temperatures which I could then equate to
the expressions shown in Table 2.
Temperature of the Dewar
As apparent from the expressions in Table 2, it is necessary to know the temperature of
the dewar in order to find m, a, r, and s. In order to do this, I put an aluminum plate in the shelf.
Since aluminum is completely reflective, all photons detected should have the same temperature
as the dewar. I took four load curves: The first and third load curve was with a 300 K load, shiny
tube, and no device in the shelf. The second and fourth load curves were with the aluminum plate
21
in the shelf. I found the voltage difference at the 0.6 bias voltage between the 300K load and the
aluminum plate. Then, using the known calibration factor of C = ΔV/ΔT, which was described in
the previous section under “Calibration,” I was able to find the temperature of the dewar using
the relation:
C = (V(300 K) – V(Tdewar)) / (300 - Tdewar)
where V(300 K) is the voltage measured at the 0.6 bias voltage for the 300K load curve, and
V(Tdewar) is the voltage measured at the 0.6 bias voltage for the aluminum plate load curve.
After manipulating this equation, one can find:
Tdewar = (V(Tdewar) - V(300 K)) / C + 300
This calculation was done twice, since there were two trials of both the aluminum plate
load curve and the 300 K load curve. These two values were then averaged, and this value was
reported as the temperature of the dewar. The value found was Tdewar = 27 ± 2 K.
Data Analysis for Equations 1 and 4
In order to find the quantity (A-B), the shiny baffle, 77 K load was set up. Then the
output voltage was measured as the filter was placed in the shelf and removed five times
consecutively. After taking these measurements, the results were plotted in Matlab, as shown
below. The graph below shows the voltage output as a function of time during this measurement,
for the Spider 15 ICM Low Pass filter.
22
Figure 6: This figure shows the lock-in voltage as a function of time as the Spider 15
ICM Low Pass filter is placed inside of the shelf and then taken out five times. The lower voltage
readings correspond to when the filter was out of the shelf, while the higher voltage readings
correspond to when the filter was inside of the shelf.
After plotting figure 6 in Matlab, I selected a uniform time interval of 10 seconds. Then I
took a 10 second sample from each of the 5 samplings for the filter inside of the shelf. Similarly,
I took the same size sample for each of the sections where the filter is out of the shelf. For each
of these samples, I took the average within the sample and plotted all of the ten averages in
Matlab, as shown in the first graph of Figure 7.
23
Then, as one can see from both Figures 6 and 7, it appears that there is a linear downward
drift in the signal over time. Thus, I performed a linear fit to all of the data in the first graph of
Figure 7, and subtracted this linear fit from the data points. The results are shown in the second
graph of figure 7.
Then, in order to find the desired quantity of (A-B), I found the difference between
consecutive trials of the filter being in the shelf and the filter being out of the shelf. The results of
this calculation are found in the third graph of Figure 7. In order to find one final value for the
quantity (A-B), I found the average of all five of the points. As an error estimate I calculated the
error on the mean among these five points.
24
Figure 7: The first graph shows the average voltage reading for each trial. The red line is
a linear fit to these points. In the second graph, the linear fit is subtracted from the original data
in the first graph. In the third graph, consecutive measurements with the filter being in and the
filter being out are subtracted, giving a difference which corresponds to the desired quantity (A-
B).
25
The same data analysis process was repeated in order to find (D-C), except with a black
baffle and a 77 K load. Similarly, (F-E) was also found using this same process except with a
shiny baffle and a 300 K load. All of these processes and data analysis were repeated using the
Spider 9 ICM Low Pass Ade filter in order to find the analogous quantities of (A-B), (D-C), and
(F-E) for this filter. After finding these values in volts, I used the calibration factor to convert
them into units of temperature.
Data Analysis for Equations 2 and 3
For the combinations shown in Table 1, data was taken by setting up a specific
combination of load temperature and baffle type, but alternating between having the filter in the
shelf and out of the shelf. Thus, data was taken simultaneously for the pairs of combinations (A,
B), (C, D), (E, F), and (G, H). This is why, for Eq. 1 and Eq. 4 it was possible to directly use this
data in order to find (F-E), (D-C), and (A-B). However, for Eq. 2 and Eq. 3, I could not use this
data because measurements were not taken simultaneously for the pairs (A, E), (F, B), or (D, B).
Thus, subtracting these combinations from one another would likely produce inaccurate results
because the data for each configuration was taken so far apart in time that the signal could have
drifted and there would be no way of predicting in which way. In addition, the set-up would most
likely have moved around or changed slightly while the baffle or load was being changed, which
would affect the signal and make it inaccurate to compare measurements taken before and after
the changes.
In order to find the quantity (D-B) for Eq. 3 in a more accurate way, I took four load
curves consecutively: the first and third load curves were taken with a black baffle and a 77 K
load, while the second and fourth load curves were taken with a shiny baffle and a 77 K load. I
26
found the average value of the output voltage at the 0.6 bias voltage points for each load curve.
Then I subtracted the values found for the first two load curves and found an amplitude
corresponding to (D-B). Similarly, I subtracted the values found for the second two load curves
to find another measurement of (D-B). I then averaged these two values and reported that
number as the final result for (D-B).
Chopper Wheel
In order to find the quantity (E-A) / (F-B) without having to subtract non-consecutive
measurements, a chopper wheel was added to the set-up, placed on top of a 77 K nitrogen
blackbody load but below the bottom of the baffle (see Figure 4). The chopper wheel was
divided into four sections. Two of the sections were covered in HR-10 and thus functioned as a
300 K thermal blackbody. The remaining two sections had the inside cut out, and thus the signal
from these sections would simply be that of the 77 K nitrogen blackbody which was stationed
directly below the wheel. Thus, by turning on the chopper wheel, which rotated with a frequency
of approximately 4.35 Hz, the signal output was a sine wave which corresponded to the
amplitude of the change in signal as the wheel chopped between 300 K blackbody and the 77K
nitrogen blackbody load.
27
Figure 4: A diagram of the set-up of the chopper wheel. The wheel was inserted directly
between the 77 K load and the baffle. The chopper wheel was divided into four sections,
alternating between black sections and carved out/empty sections. Thus, as it rotated, the signal
was chopping between that from a 77 K load and that of a 300 K load.
28
Figure 5: A drawing of the chopper wheel. The black sections correspond to sections
covered in HR-10, which function as 300 K blackbodies. The white sections correspond to cut
out sections. Thus, when placed over the 77 K nitrogen load, these sections function as 77 K
blackbodies. The black outline of the wheel is the metal rim which holds the wheel together.
With the chopper wheel continuously running, the filter was put inside of the shelf and
taken out of the shelf five times consecutively. While the filter was inside of the shelf, the
amplitude of the sine wave signal corresponded to a measurement of (E-A). Similarly, when the
filter was not inside of the shelf, the amplitude of the sine wave signal corresponded to a
measurement of (F-B). By taking the ratio of each consecutive in/out measurement, it was
29
possible to obtain a value for (E-A) / (F-B), which was the necessary quantity to measure for
subbing into Eq. 2.
Results
After doing the data analysis described in the previous section, I found the following
results for the Spider 9 ICM Low Pass filter:
Quantity Value Error
F-E 3.24 0.04
(E-A) / (F-B) 1.02 0.03
D-B -0.05 0.02
(D-C) - (B-A) -0.8 0.1
Table 3: Results found for the unknown quantities in equations 1-4 for the Spider 9 ICM Low
Pass filter. All units are in Kelvin.
For the Spider 15 ICM Low Pass filter I found the following results:
Quantity Value Error
F-E 14.5 0.2
(E-A) / (F-B) 1.01 0.05
D-B -0.05 -0.02
30
(D-C) - (B-A) -3.46 0.06
Table 4: Results found for the unknown quantities in equations 1-4 for the Spider 15 ICM Low
Pass filter.
The values found for each of the quantities in Table 3 can be directly substituted into
Equations 1, 2, 3, and 4 in order to find the fraction of incident photons which were absorbed (a),
reflected (r), scattered to the baffle (s), and remained in the main beam (m). Final results
indicated the following values for the Spider 9 ICM Low Pass filter:
Quantity Value Error
r 0.0118 0.0007
a -0.03 0.03
m 1.00 7*10^-5
s 0.0037 0.0006
Table 5: The results for r, a, m, and s for the Spider 9 ICM Low Pass filter. The quantities r, a,
m, and s represent the fraction of incident 150 GHz photons which were reflected, absorbed,
stayed in the main beam, and were scattered to the baffle, respectively.
The final results for reflection, absorption, main beam fraction, and scattering to the
baffle for the 15 ICM Low Pass filter were found by the same method of directly substituting the
values from table 4 into equations 1 through 4 for a, r, s, and m. The results are as follows:
31
Quantity Value Error
r 0.053 0.003
a -0.07 0.05
m 1.00021 7.83*10^-5
s 0.0155 0.0003
Table 6: The results for r, a, m, and s for the Spider 15 ICM Low Pass filter. The quantities r, a,
m, and s represent the fraction of incident 150 GHz photons which were reflected, absorbed,
stayed in the main beam, and were scattered to the baffle, respectively.
Conclusion
This project proved that it is possible to measure absorption, scattering, and reflection
properties for optical devices using a scatterometer and a thermal blackbody load. In addition, it
set limits on the amount of absorption, scattering, and reflection that are caused by the two
Spider low pass filters that were tested during this project. One area of concern in the results of
this experiment is the negative values found for the absorption fraction. Although these values
were mostly within the error estimates of zero, it is still a point of concern that the error was so
high. If this experiment is to be repeated in the future, it would be useful to improve the
absorption measurements.
32
Sources
[1] A.A. Penzias and R.W. Wilson. A Measurement of Excess Antenna Temperature at 4080
Mc/s. Atrophys. J., 142:419-421, July 1965.
[2] Smoot, G.F. et al. “Structure in the COBE DMR First Year Maps”. 1992, ApJ, 396, L1.
[3] C. L. Bennett, et. al. “First-Year Wilkinson Microwave Anisotropy Probe (WMAP)
Observations: Preliminary Maps and Basic Results”. 148:1-27, Setpember 2003. astro-
ph/0302207.
[4] W. C. Jones, et. al. “A measurement of the angular power spectrum of the CMB temperature
anisotropy from the 2003 flight of Boomerang”. The Astrophysics Journal , 647:823-832, 1006.
[5] J. Nagy, Personal Communication. 29 October 2013.
[6] J. Filippini, L. Moncelsi, B. Tucker, Personal Communication. 11 November 2013.
[7] M. Runyan. “A Search for Galaxy Clusters Using the Sunyaev-Zel’dovich Effect.” 20
December 2002.