mechanical infusion pumps › accuflo™ › accuflux™ · pdf...

12
Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump with a soft or hard case. It is used for the continuous application of drugs. www.pfmmedical.com

Upload: phamdung

Post on 28-Feb-2018

223 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps› AccuFlo™ › AccuFlux™

The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump with a soft or hard case. It is used for the continuous application of drugs.

www.pfmmedical.com

Page 2: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

02

The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump with a soft or hard case. It is used for the continuous application of drugs.

AccuFlo™/AccuFlux™ elastomeric infusion pumps offer clinicians and nurses the possibility to give patients medication in a safe and accurate manner. The pumps are designed to cater for continuous infusion with predetermined dose. As there are no battery or mains supply needed, the patient can be ambulatory, and also free from risks associated with device failure during infusion. The elastomeric in-fusion pumps are available in two versions —AccuFlo™ with a soft case and AccuFlux™ with a hard case.

Which version used is dependent on the patient’s needs, the infusion time, the per-sonal preference and the regulations for disposables.

Both AccuFlo™/AccuFlux™ come complete with integrated extension line, clamp, air venting and particle filter and instructions for use. In some models, there is also a light protective pouch provided. Optional acces-sories available are carrying bags.

Application RangeAccuFlo™/AccuFlux™ is designed for pa-tients requiring intravenous, subcutaneous or epidural administration of drugs. Drugs such as cytostatics, antibiotics, antimy-cotics, virustatics and analgesics can be infused. In addition it can also be used for rehydration and parenteral nutrition.

Benefits

▸ Innovative piston design▸ Low initial flow rate peak▸ Constant flow rate▸ AccuFlo™: Soft case▸ AccuFlux™: Hard case▸ No battery or power supply required

Page 3: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

03

▸ Silicone balloon▸ AccuFlo™: Lightweight▸ Easy filling▸ Accuracy: ± 10%▸ AccuFlux™: High mechanical rigidity▸ Integrated kink-proof tube▸ Air and particle filter▸ DEHP-free▸ Latex-free▸ Light-protection bag

Details

AccuFlo™/AccuFlux™ has a specially de-signed balloon-like silicone reservoir that exerts mechanical pressure when filled. Medication flow is regulated by a microbore tube or glass orifice. The combination of the pressure and flow restriction deter-mines the desired flow rate. The fluid path materials used in AccuFlo™/AccuFlux™ are silicone, ABS, DEHP free PVC and Polycar-bonate.

AccuFlo™/AccuFlux™ has a unique design in its extendable telescopic core. When medication is filled, this core extends lengthwise while the balloon reservoir expands outwards. This feature reduces the initial spike in flow rate which is common in elastomeric pumps. Together with the

design of the one-way valve, located at the filling port, this feature helps to lower the force needed to fill the pumps. The valve is anti-siphon, so removal of contents after the pump is filled is not possible.

The silicone reservoir allows trapped air to dissipate even though the pump has an integrated air venting filter.

The clamp allows priming, infusion initia-tion as well as interruption.

Functions & Processes

Page 4: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

04

View

Filling portOuter soft caseElastomeric membraneClampTubingAir and particle filterFlow restrictorPatient-side plugExpandable pistonFilling port cap (individually packed)Patient-side plug cap (individually packed)

AccuFlo™ AccuFlux™Filling port with capOuter hard caseElastomeric membraneClampTubingAir and particle filterFlow restrictorPatient-side plugExpandable pistonPatient-side plug cap (individually packed)

❶❷❸❹❺❻❼❽❾❿

❶❷❸❹❺❻❼❽❾❿

❷❸

❻❼❽

❾ ❶

❻❼❽

Page 5: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

05

Filling instructions

Manual fillingUse a luer lock syringe, preferably with 50 ml/60 ml maximum volume for manual filling. A smaller syringe will reduce the strength required to overcome resistance. In order to reduce the number of aspira-tions, connections and disconnections, a 3 way stopcock or a transfer set may be used. Follow the manufacturer’s recom-mendations for drug reconstitution and instructions for use.

DirectionsUse aseptic techniques at all times dur-ing filling. Use appropriate solution for dilution to the desired concentration and

volume. Aspirate the infusion solution into the syringe with male luer lock tip. It is recommended to purge the tubing with saline (NaCl) before adding medication.

Priming technique for drugs prone to precipitation, e.g. 5-FU▸ Fill the AccuFlo™/AccuFlux™ with 10 ml

of diluent first.▸ Using the below priming method, prime

the tubing.▸ Fill the remaining volume with diluent

and medication. At completion, the dilu-ent will fill the tubing, protecting it from precipitation, while the pump reservoir will contain the medication.

Step ❶ Aspirate the infusion solution into the syringe with male luer lock tip.

Step ❷ Flip open the top cap of AccuFlo™ to expose the filling port. Unwind the cap of AccuFlux™.

Step ❸ Attach the filled syringe. Be careful not to overtighten the luer fitting as this could damage the medication filling port. On no account is a needle to be used during the filling process as this will puncture the valve in the filling port.

Step ❹ Close the tubing clamp. Inject the contents of the syringe. Remove the syringe from the filling port. Repeat this process until the pump is filled to the predeter-mined volume.

Continued on next page ‹

❶ ❷ ❸ ❹

Page 6: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

100%

Min Vol (ml)

Nom

inal

Flo

w R

ate

(ml/H

r)

Max Vol (ml)

Flow Profile

Mechanical Infusion Pumps

06

Technical Data

▸ AccuFlo™: Sizes: ∅ 120 × 80 mm (up to 120 ml)/∅ 120 × 100 mm (from 120 ml), Weight: 61 g (up to 120 ml)/71 g (from 120 ml)

▸ AccuFlux™: Size: ∅ 70 × 138 mm, Weight: 110 g

▸ Length of infusion tube: 1,450 mm▸ Max. filling pressure: < 1 bar▸ Particle filter: 1,2 µm▸ Air filter: 0,02 µm▸ Delivery accuracy: ± 10 %

Use

▸ Chemotherapy▸ Analgesia/pain treatment▸ Antibiotic treatment▸ Desferal therapy▸ Parenteral nutrition

Step ❺ When the AccuFlo™ has been filled, remove the filling port cap from its sterile unit pack and use it to close the filling port by turning clockwise. Flip the top cap to close it. AccuFlux™: Put the cap back on the filling port. The medication label (if provided) can be folded as a flag around the tubing or pasted over the outer cover of the pump.

Step ❻ To prime, attach the patient end cap to the patient connector. The end cap is red/orange and a sterile unit packed. Open the clamp to allow the tubing to be filled. Slightly loosen the patient end cap and any air trapped in the line will be purged. By loosening and tightening the patient end cap the air bubbles will dislodge themselves to the end of the tubing. When all air is expelled, close the clamp and tighten the patient end cap.

❺ ❻

Filling instructions

Page 7: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

2827 3029 3231 33

Flow Rate ml/Hr

Nominal Flow Rate

10% belowNominal Flow Rate

–10%

+10%

Flo

w R

ate

Nominal Fill Volume

Flow Duration

More than Nominal Fill Volume

Less than Nominal Fill Volume

0.9% NaCl D5W

Flo

w R

ate

Flow Duration1 2 3 4 5 6 7 8 9 10

Mechanical Infusion Pumps

07

Knowledge

Factors affecting flow rates Effects of temperatureAccuFlo™/AccuFlux™ are calibrated for 31 deg C / 88 deg F. In order to maintain stated flow rates during infusion, the glass flow restrictor should be in close contact with the patient’s skin. For every 1 deg C above or below this temperature, the flow rate will increase or decrease by approximately 2.5 %. An increase in temperature results in an increase in flow rate and vice versa.

Effects of fill volumeFilling the pump LESS than nominal volume generally results in SLOWER flow rates. The flow duration will however be shorter as a result of less volume being delivered. Filling the pump MORE than nominal volume gen-erally results in FASTER flow rates. The flow duration will however be longer in view of the higher volume being delivered.

Effects of diluentAccuFlo™/AccuFlux™ flow rates are calculat-ed using 0.9 % normal saline (NaCl) as dilu-ent. Using dextrose 5 % (D5W) as a diluent or adding any drug of greater viscosity than normal saline will increase delivery time by 10 %. When filled to nominal volume, the flow accuracy is ± 15 % of the labelled flow rate for pumps with 50 ml/hrs or higher flow rate. For pumps with slower than 50 ml/hrs flow rate, the flow accuracy is ± 10 % of the labelled flow rate against a back pressure of 40 cm of water.

Page 8: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

08

Model number CT-0010-200H

CT-0010-168H

CT-0010-160

CT-0010-120

CT-0010-100

CT-0010-050

CT-0015-240

CT-0015-180H

CT-0020-240

CT-0020-200H

CT-0020-150

CT-0020-100H

CT-0020-100

CT-0020-050

CT-0020-048H

CT-0030-150

CT-0040-200H

CT-0040-200

CT-0040-100H

CT-0040-100

CT-0050-270

CT-0050-240

CT-0050-120

CT-0050-060

CT-0080-200H

CT-0100-240

Nominal flow rate (ml/hr) 1 1 1 1 1 1 1,5 1,5 2 2 2 2 2 2 2 3 4 4 4 4 5 5 5 5 8 10Nominal fill volume (ml) 200 168 160 120 100 50 240 180 240 200 150 100 100 50 48 150 200 200 100 100 270 240 120 60 200 240Maximum fill volume (ml) 200 180 160 120 100 60 240 200 270 200 150 110 120 50 60 200 200 200 110 100 270 240 120 65 200 240Residual volume (ml) < 3,5 < 3,5 < 3,5 < 2,5 < 2,5 < 2,5 < 3,5 < 3,5 < 3,5 < 3,5 < 2,5 < 2,5 < 2,5 < 2,5 < 2,5 < 3,5 < 3,5 < 2,5 < 2,5 < 3,5 < 3,5 < 2,5 < 2,5 < 2,5 < 3,5 < 3,5

Refrigerator to room temp (hr) 8 8 8 8 6 6 8 8 8 8 8 8 6 6 6 8 8 8 6 6 8 8 8 6 8 8

Freezer to room temp (hr) 10 10 10 10 8 8 10 10 10 10 10 10 8 8 8 10 10 10 8 8 10 10 10 8 10 10

Approx. infusion time Fill volume to obtain desired infusion time Fill volume to obtain desired infusion time12 hrs 6018 hrs 36 72 90 1801 day 50 48 100 100 120 200 240

1 1/2 days 144 1802 days 50 100 96 150 200 200 270 240

2 1/2 days 60 120 120 1803 days 1504 days 100 2005 days 120 180 240

5 1/2 days 200 2656 days 144 2167 days 168 160 240

7 1/2 days 1808 days 200

Model number CT-0100-080H

CT-0500-120

CT-0500-050

CT-0800-240

CT-1000-400

CT-1000-200

CT-1000-100

CT-1000-050

CT-1200-240

CT-2000-200

CT-2000-100

CT-2400-240

CT-2400-120

Nominal flow rate (ml/hr) 10 50 50 80 100 100 100 100 120 200 200 240 240Nominal fill volume (ml) 80 100 50 240 400 200 100 50 240 200 100 240 120Maximum fill volume (ml) 85 120 60 260 400 250 120 55 250 250 120 260 150Residual volume (ml) < 2,5 < 2,5 < 2,5 < 2,5 < 5,0 < 3,5 < 2,5 < 2,5 < 3,5 < 3,5 < 2,5 < 3,5 < 2,5

Refrigerator to room temp (hr) 6 6 6 6 10 8 6 6 8 8 8 8 8

Freezer to room temp (hr) 8 8 8 8 12 10 8 8 10 10 10 10 10

Approx. infusion time Fill volume to obtain desired infusion time30 min 50 100 12036 min 55 120 14060 min 50 100 200 24062 min 210 25065 min 55 110 220 26070 min 60 120 230

2 hrs / 120 min 100 200 2402 hrs 5 min / 125 min 210 250

2 hrs 10 min / 130 min 110 2202 hrs 20 min / 140 min 120

3 hrs / 180 min 240

3 hrs 15 min / 195 min 260

4 hrs / 240 min 4008 hrs / 480 min 80

Reference list: Infusion time – Fill volume

Page 9: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

09

Volume

Volume

Model number CT-0010-200H

CT-0010-168H

CT-0010-160

CT-0010-120

CT-0010-100

CT-0010-050

CT-0015-240

CT-0015-180H

CT-0020-240

CT-0020-200H

CT-0020-150

CT-0020-100H

CT-0020-100

CT-0020-050

CT-0020-048H

CT-0030-150

CT-0040-200H

CT-0040-200

CT-0040-100H

CT-0040-100

CT-0050-270

CT-0050-240

CT-0050-120

CT-0050-060

CT-0080-200H

CT-0100-240

Nominal flow rate (ml/hr) 1 1 1 1 1 1 1,5 1,5 2 2 2 2 2 2 2 3 4 4 4 4 5 5 5 5 8 10Nominal fill volume (ml) 200 168 160 120 100 50 240 180 240 200 150 100 100 50 48 150 200 200 100 100 270 240 120 60 200 240Maximum fill volume (ml) 200 180 160 120 100 60 240 200 270 200 150 110 120 50 60 200 200 200 110 100 270 240 120 65 200 240Residual volume (ml) < 3,5 < 3,5 < 3,5 < 2,5 < 2,5 < 2,5 < 3,5 < 3,5 < 3,5 < 3,5 < 2,5 < 2,5 < 2,5 < 2,5 < 2,5 < 3,5 < 3,5 < 2,5 < 2,5 < 3,5 < 3,5 < 2,5 < 2,5 < 2,5 < 3,5 < 3,5

Refrigerator to room temp (hr) 8 8 8 8 6 6 8 8 8 8 8 8 6 6 6 8 8 8 6 6 8 8 8 6 8 8

Freezer to room temp (hr) 10 10 10 10 8 8 10 10 10 10 10 10 8 8 8 10 10 10 8 8 10 10 10 8 10 10

Approx. infusion time Fill volume to obtain desired infusion time Fill volume to obtain desired infusion time12 hrs 6018 hrs 36 72 90 1801 day 50 48 100 100 120 200 240

1 1/2 days 144 1802 days 50 100 96 150 200 200 270 240

2 1/2 days 60 120 120 1803 days 1504 days 100 2005 days 120 180 240

5 1/2 days 200 2656 days 144 2167 days 168 160 240

7 1/2 days 1808 days 200

Model number CT-0100-080H

CT-0500-120

CT-0500-050

CT-0800-240

CT-1000-400

CT-1000-200

CT-1000-100

CT-1000-050

CT-1200-240

CT-2000-200

CT-2000-100

CT-2400-240

CT-2400-120

Nominal flow rate (ml/hr) 10 50 50 80 100 100 100 100 120 200 200 240 240Nominal fill volume (ml) 80 100 50 240 400 200 100 50 240 200 100 240 120Maximum fill volume (ml) 85 120 60 260 400 250 120 55 250 250 120 260 150Residual volume (ml) < 2,5 < 2,5 < 2,5 < 2,5 < 5,0 < 3,5 < 2,5 < 2,5 < 3,5 < 3,5 < 2,5 < 3,5 < 2,5

Refrigerator to room temp (hr) 6 6 6 6 10 8 6 6 8 8 8 8 8

Freezer to room temp (hr) 8 8 8 8 12 10 8 8 10 10 10 10 10

Approx. infusion time Fill volume to obtain desired infusion time30 min 50 100 12036 min 55 120 14060 min 50 100 200 24062 min 210 25065 min 55 110 220 26070 min 60 120 230

2 hrs / 120 min 100 200 2402 hrs 5 min / 125 min 210 250

2 hrs 10 min / 130 min 110 2202 hrs 20 min / 140 min 120

3 hrs / 180 min 240

3 hrs 15 min / 195 min 260

4 hrs / 240 min 4008 hrs / 480 min 80

To find the appropriate model for desired infusion time, find the time in the left column, then slide across to find volume, and to the top for the appropriate model number.

To find the partial or overfill volume, look for desired delivery time in the far left column, then slide to the column with model number and find the corresponding fill volume.

❶ ❷

❸ Model number

Model number❷

Page 10: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

10

Ordering Information

AccuFlo™REF Flow rate Volume Infusion time PU

CT-0010-050 1 ml/h 50 ml 2 days 10

CT-0010-100 1 ml/h 100 ml 4 days 10

CT-0010-120 1 ml/h 120 ml 5 days 10

CT-0010-160 1 ml/h 160 ml 7 days 10

CT-0015-240 1,5 ml/h 240 ml 7 days 10

CT-0020-050 2 ml/h 50 ml 1 day 10

CT-0020-100 2 ml/h 100 ml 2 days 10

CT-0020-150 2 ml/h 150 ml 3 days 10

CT-0020-240 2 ml/h 240 ml 5 days 10

CT-0030-150 3 ml/h 150 ml 2 days 10

CT-0040-100 4 ml/h 100 ml 1 day 10

CT-0040-200 4 ml/h 200 ml 2 days 10

CT-0050-060 5 ml/h 60 ml 12 hrs 10

CT-0050-120 5 ml/h 120 ml 1 day 10

CT-0050-240 5 ml/h 240 ml 2 days 10

CT-0050-270 5 ml/h 270 ml 2 days 10

CT-0070-270 7 ml/h 270 ml 1,5 days 10

CT-0100-240 10 ml/h 240 ml 1 day 10

CT-0500-050 50 ml/h 50 ml 1 hr 10

CT-0500-100 50 ml/h 100 ml 2 hrs 10

CT-0800-240 80 ml/h 240 ml 3 hrs 10

CT-1000-050 100 ml/h 50 ml 30 min 10

CT-1000-100 100 ml/h 100 ml 1 hr 10

CT-1000-200 100 ml/h 200 ml 2 hrs 10

CT-1000-400 100 ml/h 400 ml 4 hrs 10

CT-1200-240 120 ml/h 240 ml 2 hrs 10

CT-2000-100 200 ml/h 100 ml 30 min 10

CT-2000-200 200 ml/h 200 ml 1 hr 10

CT-2400-120 240 ml/h 120 ml 30 min 10

CT-2400-240 240 ml/h 240 ml 1 hr 10

Page 11: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

11

Ordering Information

AccuFlux™REF Flow rate Volume Infusion time PU

CT-0010-168H 1 ml/h 168 ml 7 days 10

CT-0010-200H 1 ml/h 200 ml 8 days 10

CT-0015-180H 1,5 ml/h 180 ml 5 days 10

CT-0020-048H 2 ml/h 48 ml 1 day 10

CT-0020-100H 2 ml/h 100 ml 2 days 10

CT-0020-200H 2 ml/h 200 ml 4 days 10

CT-0040-100H 4 ml/h 100 ml 1 day 10

CT-0040-200H 4 ml/h 200 ml 2 days 10

CT-0080-200H 8 ml/h 200 ml 1 day 10

CT-0100-080H 10 ml/h 80 ml 8 hrs 10

Carry bagREF 53100

Ordering Information Accessories

Bag › with AccuFlo™ logoREF 06-B2-01

Bag › with AccuFlux™ logoREF 07-B2-01

▸ Adjustable carrying belt▸ Zip with tube exit

▸ Comfortable to wear▸ Waterproof inner lining▸ Toggle for safe closure of the bag▸ Adjustable cord length

Page 12: Mechanical Infusion Pumps › AccuFlo™ › AccuFlux™ · PDF fileMechanical Infusion Pumps 02 The AccuFlo™/AccuFlux™ elastomeric infusion pump is a mechanical disposable pump

Mechanical Infusion Pumps

PB32

10EN

/11.

2010

Should you have any questions our Customer Service & Sales Support team is happy to advise you.

Contact

www.pfmmedical.com – Your information source for the products of pfm medical ag.

Distributorpfm medical UK LtdSuite 2, Armcon Business Park,London Road South, PoyntonCheshire, SK12 1LQ, England

ManufacturerMedpro Corporation PTE LTD, SingaporeProducts are subject to alterations

For [email protected] +44 (0)7789 8383-64F +44 (0)1753 6448-69

Outside [email protected] +49 (0)2236 9641-70F +49 (0)2236 9641-744