medical image analysis

81
Medical Image Analysis Medical Image Analysis Medical Imaging Modalities: Magnetic Resonance Imaging Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Upload: candie

Post on 20-Mar-2016

65 views

Category:

Documents


0 download

DESCRIPTION

Medical Image Analysis. Medical Imaging Modalities: Magnetic Resonance Imaging. Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011. Magnetic Resonance Imaging. Nuclear magnetic resonance The selected nuclei of the matter of the object - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Medical Image Analysis

Medical Image AnalysisMedical Image AnalysisMedical Imaging Modalities: Magnetic Resonance Imaging

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 2: Medical Image Analysis

Magnetic Resonance Magnetic Resonance ImagingImagingNuclear magnetic resonance

◦The selected nuclei of the matter of the object

◦Blood flow and oxygenation◦Different parameters: weighted,

weighted, Spin-density◦Advance: MR Spectroscopy and

Functional MRI◦Fast signal acquisition of the order of

a fraction of a secondFigures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

1T 2T

Page 3: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure comes from the Wikipedia, www.wikipedia.org.

Page 4: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure 4.12. MR images of a selected cross-section that are obtained simultaneously using a specific imaging technique. The images show (from left to right), respectively, the T1-weighted, T-2 weighted and the Spin-Density property of the hydrogen protons present in the brain.

Page 5: Medical Image Analysis

Magnetic Resonance Magnetic Resonance ImagingImaging1H: high sensitivity and vast

occurrence in organic compounds13C: the key component of all organic15N: a key component of proteins and

DNA19F: high relative sensitivity31P: frequent occurrence in organic

compounds and moderate relative sensitivity

Adapted from the Wikipedia, www.wikipedia.org.

Page 6: Medical Image Analysis

MR SpectroscopyMR Spectroscopy

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure comes from the Wikipedia, www.wikipedia.org.

Page 7: Medical Image Analysis

MR SpectroscopyMR Spectroscopy

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure comes from the Wikipedia, www.wikipedia.org.

Page 8: Medical Image Analysis

Functional MRIFunctional MRI

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure comes from the Wikipedia, www.wikipedia.org.

Page 9: Medical Image Analysis

MRI PrinciplesMRI Principles : spin-lattice relaxation time : spin-spin relaxation time : the spin density

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

1T

2T

Page 10: Medical Image Analysis

MRI PrinciplesMRI Principles1. Great web sites

1. Simulations from BIGS - Lernhilfe für Physik und Technik

2. http://www.cis.rit.edu/class/schp730/bmri/bmri.htm

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 11: Medical Image Analysis

MRI PrinciplesMRI PrinciplesSpin

◦A fundamental property of nuclei with odd atomic weight and/or odd atomic numbers is the possession of angular moment

Magnetic moment◦The charged protons create a

magnetic field around them and thus act like tiny magnets

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 12: Medical Image Analysis

MRI PrinciplesMRI Principles : the spin angular moment : the magnetic moment : a gyromagnetic ratio, MHz/T

A hydrogen atom◦ :42.58 MHz/T

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

J

J

Page 13: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

NN

SS

JJ

JJ

Figure 4.13. Left: A tiny magnet representation of a charged proton with angular moment, J. Right: A symbolic representation of a charged proton with angular moment, J and a magnetic moment, μ.

Page 14: Medical Image Analysis

MRI PrinciplesMRI PrinciplesPrecession of a spinning proton

◦The interaction between the magnetic moment of nuclei with the external magnetic field

◦Spin quantum number of a spinning proton: ½

◦The energy level of nuclei aligning themselves along the external magnetic field is lower than the energy level of nuclei aligned against the external magnetic field

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 15: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure 4.14 (a) A symbolic representation of a proton with precession that is experienced by the spinning proton when it is subjected to an external magnetic field. (b) The random orientation of protons in matter with the net zero vector in both longitudinal and transverse directions.

Page 16: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

MRI PrinciplesMRI PrinciplesEquation of motion for isolated

spin

Solution:

J

kHHdtJd

00

kHdtd

0

00 H

Page 17: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Longitudinal Vector OX at the transverse position X

Net LongitudinalVector: Zero

Net TransverseVector: Zero

Net LongitudinalVector: Zero

Net TransverseVector: Zero

Page 18: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Lower EnergyLevel

Higher EnergyLevel

S

N

H

Lower EnergyLevel

Higher EnergyLevel

S

N

H0

Figure 4.15 (a). Nuclei aligned under thermal equilibrium in the presence of an external magnetic field. (b). A non-zero net longitudinal vector and a zero transverse vector provided by the nuclei precessing in the presence of an external magnetic field.

Page 19: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Non-zero Net Longitudinal Vector

x

y

z

x

y

zH0

Net Zero Transverse Vector

Page 20: Medical Image Analysis

MRI PrinciplesMRI PrinciplesThe precession frequency

◦Depends on the type of nuclei with a specific gyromagnetic ratio and the intensity of the external magnetic field

◦This is the frequency on which the nuclei can receive the Radio Frequency (RF) energy to change their states for exhibiting nuclear magnetic resonance

◦The excited nuclei return to the thermal equilibrium through a process of relaxation emitting energy at the same precession frequency

Page 21: Medical Image Analysis

MRI PrinciplesMRI Principles90-degree pulse

◦Upon receiving the energy at the Larmor frequency, the transverse vector also changes as nuclei start to precess in phase

◦Form a net non-zero transverse vector that rotates in the x-y plane perpendicular to the direction of the external magnetic field

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 22: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

S

N

S

N x

y

z

Figure 4.16. The 90-degree pulse causing nuclei to precess in phase with the longitudinal vector shifted clockwise by 90-degrees as a result of the absorption of RF energy at the Larmor frequency.

Page 23: Medical Image Analysis

MRI PrinciplesMRI Principles180-degree pulse

◦If enough energy is supplied, the longitudinal vector can be completely flipped over with a 180-degree clockwise shidf in the direction against the external magnetic field

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 24: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

S

N

S

N

x

y

z

Figure 4.17. The 180-degree pulse causing nuclei to precess in phase with the longitudinal vector shifted clockwise by 180-degrees as a result of the absorption of RF energy at the Larmor frequency.

Page 25: Medical Image Analysis

MRI PrinciplesMRI PrinciplesRelaxation

◦The energy emitted during the relaxation process induces an electrical signal in a RF coil tuned at the Larmor frequency

◦The free induction decay of the electromagnetic signal in the PF coil is the basic signal that is used to create MR images

◦The nuclear excitation forces the net longitudinal and transverse magnetization vectors to move

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 26: Medical Image Analysis

MRI PrinciplesMRI PrinciplesA stationary magnetization

vector

The total response of the spin system

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

N

nnM

1

1

0

2

)( T

kMMT

jMiMHM

dtMd zzyx

Page 27: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Dephasing

RF Pulse

Random Phase (Zero Transverse Vector)

In Phase Spin

Relaxation

Figure 4.18. The transverse relaxation process of spinning nuclei.

Page 28: Medical Image Analysis

MRI PrinciplesMRI PrinciplesThe longitudinal and transverse

magnetization vectors with respect to the relaxation times

where

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

tiTtyxyx eeMtM 02/

,, )0()(

11 //0 )0()1()( Ttz

Ttzz eMeMtM

piyxyx eMM 0)0()0( ',',

Page 29: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

t

t

Mx,y (t)

Mz (t)

Figure 4.19. (a) Transverse and (b) longitudinal magnetization relaxation after the RF pulse.

Page 30: Medical Image Analysis

MRI PrinciplesMRI PrinciplesThe RF pulse causes nuclear

excitation changing the longitudinal and transverse magnetization vectors

After the RF pulse is turned off, the excited nuclei go through the relaxation phase emitting the absorbed energy at the same Larmor frequency that can be detected as an electrical signal, called the Free Induction Decay (FID)

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 31: Medical Image Analysis

MRI PrinciplesMRI PrinciplesThe NMR spin-echo signal (FID

signal)

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

dxdydzezyxMS zyxizyx

zyx )(0 ),,(),,(

zyxzyxi

zyx dddeSMzyx zyx )(0 ),,(),,(

Page 32: Medical Image Analysis

MR InstrumentationMR InstrumentationThe stationary external magnetic

field◦Provided by a large superconducting

magnet with a typical strength of 0.5 T to 1.5 T

◦Housing of gradient coils◦Good field homogeneity, typically on

the order of 10-50 parts per million◦A set of shim coils to compensate for

the field inhomogeneity Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 33: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Gradient Coils

Magnet

Gradient Coils

RFCoils

PatientPlatform

Monitor Data-AcquisitionSystem

Figure 4.20. A general schematic diagram of a MR imaging system.

Page 34: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure comes from the Wikipedia, www.wikipedia.org.

Page 35: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure comes from the Wikipedia, www.wikipedia.org.

Page 36: Medical Image Analysis

MR InstrumentationMR InstrumentationAn RF coil

◦To transmit time-varying RF pulses◦To receive the radio frequency

emissions during the nuclear relaxation phase

◦Free Induction Decay (FID) in the RF coil

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 37: Medical Image Analysis

MR Pulse SequencesMR Pulse SequencesNMR signal

◦The frequency and the phaseSpatial encoding in MR imaging

◦Frequency encoding and phase encoding

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 38: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

x

z

y

y

xx

y

z

z

Sagital

Coronal

Axial

Figure 4.21 (a). Three-dimensional object coordinate system with axial, sagittal and coronal image views. (b): From top left to bottom right: Axial, coronal and sagittal MR images of a human brain.

Page 39: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 40: Medical Image Analysis

MR Pulse SequencesMR Pulse Sequences

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Z Gradient

90 RF Pulse(Slice Selection)

X Gradient

Phase-Encoding(x-scan selection)

Z Gradient

180 RF Pulse(Slice Echo Formation)

Y Gradient

Frequency Encoding(Read-Out Pulse)

Figure 4.22. (a): Three-dimensional spatial encoding for spin-echo MR pulse sequence. (b): A linear gradient field for frequency encoding. (c). A step function based gradient field for phase encoding.

Page 41: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Varying Spatially DependentLarmor Frequency

S

N

S

N

Linear Gradient

Precessing Nuclei

External Magnet

Positive PhaseChange

Negative PhaseChange

Phase -EncodingGradientStep

Page 42: Medical Image Analysis

MR Pulse SequencesMR Pulse SequencesFrequency encoding

◦A linear gradient is applied throughout the imaging space a long a selected direction

◦The effective Larmor frequency of spinning nuclei is also spatially encloded along the direction of the gradient

◦Slice selection for axial imaging

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 43: Medical Image Analysis

MR Pulse SequencesMR Pulse SequencesThe phase-encoding gradient

◦Applied in steps with repeated cycles◦If 256 steps are to be applied in the

phase-encoding gradient, the readout cycle is repeated 256 times, each time with a specific amount of phase-encoding gradient

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 44: Medical Image Analysis

Spin Echo ImagingSpin Echo Imaging :

◦Between the application of the 90 degree pulse and the formation of echo (rephasing of nuclei

:◦Between the 90 degree pulse and

180 degree pulse

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

ET

2/ET

Page 45: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

RF Energy: 90 Deg Pulse

Zero Net Vector:Random Phase

RelaxationDephasing

RF Energy: 180 Deg Pulse

Echo -Formation

RF Energy: 90 Deg Pulse

Zero Net Vector:Random Phase

In Phase

Rephasing

Echo -Formation

In Phase

Figure 4.23. The transverse relaxation and echo formation of the spin echo MR pulse sequence.

Page 46: Medical Image Analysis

Spin Echo ImagingSpin Echo ImagingK-space

◦The placement of raw frequency data collected through the pulse sequences in a multi-dimensional space

◦By taking the inverse Fourier transform of the k-space data, an image about the object can be reconstructed in the spatial domain

◦The NMR signals collected as frequency-encoded echoes can be placed as horizontal lines in the corresponding 2-D k-space

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 47: Medical Image Analysis

Spin Echo ImagingSpin Echo ImagingK-space

◦As multiple frequency encoded echoes are collected with different phase-encoding gradients, they are placed as horizontal lines in the corresponding k-space with the vertical direction representing the phase-encoding gradient values

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 48: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure comes from the Wikipedia, www.wikipedia.org.

Page 49: Medical Image Analysis

Spin Echo ImagingSpin Echo Imaging : the cycle repetition time weighted

◦A long and a long weighted

◦A short and a shortSpin-density

◦A long and a short

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

RT2T

RT ET1T

RT ET

RT ET

Page 50: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

90 deg RF pulse

180 deg RF pulse

Gz: Slice Selection Frequency EncodingGradient

Gx: Phase EncodingGradient

Gy: Readout Frequency EncodingGradient

TE /2

TE /2

TE

RF pulseTransmitter

NMRRF FIDSignal

Figure 4.24. A spin echo pulse sequence for MR imaging.

Page 51: Medical Image Analysis

Spin Echo ImagingSpin Echo Imaging

The effective transverse relaxation time from the field inhomogeneities

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

12 1),,(),,( 0TT

TT RE

eezyxzyx

2 11

2*

2

HTT

Page 52: Medical Image Analysis

Spin Echo ImagingSpin Echo ImagingThe effective transverse

relaxation time from a spatial encoding gradient

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

2 11

*2

**2

GdTT

Page 53: Medical Image Analysis

Inversion Recovery Inversion Recovery ImagingImagingIR imaging

◦IR imaging pulse sequence allows relaxation of some or all of before spins are rephased through 90-degree pulse and therefore emphasizes the effect of longitudinal magnetization

◦ 180-degree pulse is first applied along with the slice selection frequency encoding gradientFigures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

1T

Page 54: Medical Image Analysis

Echo Planar ImagingEcho Planar ImagingA single-shot fast-scanning methodSpiral Echo Planar Imaging (SEPI)

◦where

)(1)( tdtdtG xx

)(1)( tdtdtG yy

tttx cos )(

ttty sin )(

Page 55: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

90 deg RF pulse

90 deg RF pulse

Gz: Slice Selection Frequency EncodingGradient

Gx: OscillatingGradient

Gy: Readout Gradient

RF pulseTransmitter

NMRRF FIDSignal

Figure 4.25. A single shot EPI pulse sequence.

Page 56: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

gy

gx

x

y

Figure 4.26. The k-space representation of the EPI scan trajectory.

Page 57: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

x

y

SEPI Trajectory

Data SamplingPoints

Figure 4.27. The spiral scan trajectory of SEPI pulse sequence in the k-space.

Page 58: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

90 deg RF pulse

180 deg RF pulse

Gz: Slice Selection Frequency EncodingGradient

Gx Gradient

Gy Gradient

TE /2

TE /2

TE

RF pulseTransmitter

NMRRF FIDSignal

TD

Figure 4.28. The SEPI pulse sequence

Page 59: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure 4.29. MR images of a human brain acquired through SEPI pulse sequence.

Page 60: Medical Image Analysis

Gradient Echo ImagingGradient Echo ImagingFast low angle shot (FLASH) imaging

◦Utilize low-flip angle RF pulses to create multiple echoes in repeated cycles to collect the data required for image reconstruction

◦A low-flip angle (as low as 20 degrees)◦The readout gradient is inverted to re-

phase nuclei leading to the gradient echo during the data acquisition

◦The entire pulse sequence time is much shorter than the spin echo pulse sequence

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 61: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Low Flip Angle RF pulse

Gz: Slice Selection Frequency EncodingGradient

RF pulseTransmitter

Gx: Phase EncodingGradient

Gy: Readout Frequency EncodingGradient TE

NMRRF FIDSignal

Figure 4.30. The FLASH pulse sequence for fast MR imaging.

Page 62: Medical Image Analysis

Flow ImagingFlow ImagingTracking flow

◦Diffusion (incoherent flow) and perfusion (partially coherent flow)

◦The FID signal generated in the RF receiver coil by the moving nuclei and velocity-dependent factors

MR angiography

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Page 63: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

90 deg RF pulse

180 deg (selective)RF pulse

Gz: Slice Selection Frequency EncodingGradient

Gx: Phase EncodingGradient

Gy: Readout Frequency EncodingGradient

TE /2

TE

RF pulseTransmitter

NMRRF FIDSignal

Figure 4.31. A flow imaging pulse sequence with spin echo.

Page 64: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure 4.32: Left: A proton density image of a human brain. Right: The corresponding perfusion image.

Page 65: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

90 degree RF pulse

Gz: Slice Selection Frequency EncodingGradient

RF pulseTransmitter

Gx: Phase EncodingGradient

Gy: Readout Frequency EncodingGradient

TE NMRRF FIDSignal

Next 90 degree RF pulse

TR

Figure 4.33. Gradient echo based MR pulse sequence for 3-D MR volume angiography.

Page 66: Medical Image Analysis

Figures come from the textbook: Medical Image Analysis, Second Edition, by Atam P. Dhawan, IEEE Press, 2011.

Figure 4.34. An MR angiography image.

Page 67: Medical Image Analysis

Flow ImagingFlow Imaging

Figure comes from the Wikipedia, www.wikipedia.org.

angiography image

Page 68: Medical Image Analysis

FMRIFMRIfMRI imaging

◦Measure blood oxygen level during sensory stimulation or any task that causes a specific neural activity

◦Visual or auditory stimulation, finger movement, or a cognitive task

◦Blood oxygenated level dependence (BOLD)

◦Oxygenated hemoglobin ( ) is diamagnetic, while deoxygenated hemoglobin ( ) is paramagnetic

2HbO

Hb

Page 69: Medical Image Analysis

Figure comes from the Wikipedia, www.wikipedia.org.

Page 70: Medical Image Analysis

FMRIFMRIfMRI imaging

◦A reduction of the relative deoxy-hemoglobin concentration due to an increase of blood flow and hence increased supply of fresh oxy-hemoglobin during neural activity is measured as an increase in or weighted MR signals

2T2T

Page 71: Medical Image Analysis

Diffusion ImagingDiffusion ImagingDiffusion process

◦Water molecules spread out over time that is represented by Brownian motion

◦An anisotropic Gaussian distribution along a given spatial axis such that the spread of the position of molecules after a time along a spatial axis can be represented with a variance of

◦where is diffusion coefficient in the tissue

Tx

D

DTx 22

Page 72: Medical Image Analysis

Figure comes from the Wikipedia, www.wikipedia.org.

DTI color image

Page 73: Medical Image Analysis

Diffusion ImagingDiffusion ImagingDiffusion process

◦Anisotropic diffusion in the white matter◦Isotropic diffusion in the gray matter◦Motion probing gradients (MPG) to

examine the motion of water molecules in the diffusion process in a specific direction

◦The MR FID signal is decreased for healthy tissue, and increased with trapped-in water molecules

Page 74: Medical Image Analysis

Diffusion ImagingDiffusion ImagingDiffusion process

◦where is the gyromagnetic ratio, is diffusion coefficient, and is the strength of two MPG gradients each with duration separated by applied in spatial directions

DNGeSS )/(0

222

DG

N

Page 75: Medical Image Analysis

Diffusion ImagingDiffusion ImagingDiffusion process

z

y

x

zzzyzx

yzyyyx

xzxyxx

zyx

uuu

DDDDDDDDD

uuuD ),,(

321][trace zzyyxx DDDD

3

2

1

Page 76: Medical Image Analysis

Diffusion ImagingDiffusion ImagingDiffusion process

◦Fractional anisotropy (FA)

◦Multiple sclerosis, strokes, tumors, Parkinson’s and Alzheimer’s disease

◦Attention deficit hyperactivity disorder (ADHS)

23

22

21

213

232

221 )()()(

21

FA

Page 77: Medical Image Analysis

Contrast, Spatial Resolution, Contrast, Spatial Resolution, and SNRand SNRSpin-echo imaging pulse

sequence

Inversion recovery (180-90-180) imaging pulse sequence

211 TT

TT ER

eekS

21121 TT

TT

TT ERI

eeekS

Page 78: Medical Image Analysis

Contrast, Spatial Resolution, Contrast, Spatial Resolution, and SNRand SNRGradient echo imaging pulse

sequence

1

*21

cos1

sin1

TT

TT

TT

R

ER

e

ee

kS

Page 79: Medical Image Analysis

Contrast, Spatial Resolution, Contrast, Spatial Resolution, and SNRand SNRParamagnetic contrast agent

◦gadolinium (Gd) to change the susceptibility of the net magnetization vector

◦Reduces relaxation time and increases the signal intensity of -weighted images

Noise and field inhomogeneities◦RF noise, field inhomogeneities,

motion, chemical shift

1T1T

Page 80: Medical Image Analysis

Contrast, Spatial Resolution, Contrast, Spatial Resolution, and SNRand SNRChemical shift

◦The deviation of its effective resonance frequency in the presence of other nuclei from a standard reference without any other nuclei with their local magnetic fields present

◦ppmref

ref

610)(

Page 81: Medical Image Analysis

Contrast, Spatial Resolution, Contrast, Spatial Resolution, and SNRand SNR

◦Induced magnetic field in alkenes

◦Induced magnetic field in alkynes

Figure comes from the Wikipedia, www.wikipedia.org.