Medical Imaging in ThickTissues Using Diffuse Optics

Laser Microbeam and Medical Program (LAMMP)Beckman Laser Institute

Department of Biomedical Engineering

University of California, Irvinewww.bli.uci.edu

Bruce J. Tromberg

BLI

Page 2: Medical Imaging in Thick Tissues Using Diffuse Optics

Australian “National” Team

International rugby sevens tournament: 1982

Page 3: Medical Imaging in Thick Tissues Using Diffuse Optics

Beckman Laser Institute and Medical Clinic

http://www.bli.uci.edu/Five Beckman Institutes in U.S.

Univ. of Illinois

UC Irvine

(1982)

Caltech

Stanford

City of Hope

Nat. Med. Ctr.

BLI: Co-founders Michael Berns and Arnold Beckman

BLI

Page 4: Medical Imaging in Thick Tissues Using Diffuse Optics

Optical Imaging in Thick Tissues

800 nm NIR light

Optical Imaging in Thick Tissues

• What is the biologic origin of contrast?

• Can contrast be quantified?

• Can light be localized?

Key Questions

Page 5: Medical Imaging in Thick Tissues Using Diffuse Optics

Optical Imaging in Thick Tissues

• What is the biologic origin of contrast?

• Can contrast be localized?

• Can contrast be quantified?

Key Questions

Intrinsic Optical Contrast0.1 µm

1.0 µm

10 µm

1 cm

5 cm

500 µm

5mm

1mm

{Sub-cellular Structures:

Size/Shape/Density

Scattering

(lsc ~20 �m)

Scattering

&

Absorption

(labs ~10 cm)

Scattering and absorption: across

spatial scales

{Cell Proliferation

Hypoxia

Fibrosis

Edema

Necrosis

Angiogenesis{

Page 6: Medical Imaging in Thick Tissues Using Diffuse Optics

Tissue Spectroscopy

Scattering

600-1000 nm

NIR Optical Spectrosocpy

600 650 700 750 800 850 900 950 10000.0

0.2

0.4

0.6

0.8

1.0

AB

SO

RP

TIO

N(m

m-1

mM

-1

)

WAVELENGTH (nm)

HHb

O2Hb

Page 7: Medical Imaging in Thick Tissues Using Diffuse Optics

NIR Optical Spectroscopy

600 650 700 750 800 850 900 950 10000.0

0.2

0.4

0.6

0.8

1.0

AB

SO

RP

TIO

N(m

m-1

mM

-1

)

WAVELENGTH (nm)

HHb

O2Hb

Lipid

H2O

Optical Imaging in Thick Tissues

• What is the biologic origin of contrast?

• Can contrast be localized?

• Can contrast be quantified?

Key Questions

Page 8: Medical Imaging in Thick Tissues Using Diffuse Optics

Multiple Light Scattering

Light Propagation in TissueOptical Imaging in Thick Tissues

• What is the biologic origin of contrast?

• Can contrast be localized?

• Can contrast be quantified?

Key Questions

Page 9: Medical Imaging in Thick Tissues Using Diffuse Optics

Technology for Tissue Spectroscopy

and Imaging

• Based on broadband modulation of semiconductor diode

lasers, “photon diffusion” models

• Quantitatively separates absorption from scattering

• Tomography of biochemical composition and structure

Time and Frequency-Domain Photon Migration

Tromberg et al. Neoplasia 2, (2000).

Diffuse Optical Imaging/Spectroscopy

Quantitative Measurements, Biochemical Composition

Page 10: Medical Imaging in Thick Tissues Using Diffuse Optics

Quantitative Measurements, Biochemical Composition

• Hemoglobin– Oxy-, Deoxy-, Met-, Total, Tissue oxygen saturation

• % Water– Protein bound, deep tissue temperature

• % Lipid

• Cytochrome oxidase– Oxidized, reduced forms

• Tissue Scatter Power– Density of cells, collagen, lipid

• Exogenous probes– ICG, Methylene Blue, Evans Blue…NIR fluorescent probes,

etc.

Diffuse Optical Imaging/Spectroscopy Diffuse Optics Technologies

Diffuse Optical Imaging (DOI)

Diffuse Optical Spectroscopy (DOS)

• Humans– Breast, Brain, Bone, Muscle

• Small animal models– Whole body

Page 11: Medical Imaging in Thick Tissues Using Diffuse Optics

• Cancer– Detection, image-guided therapy

• Functional Activation– Brain, muscle

• Wound Healing– Tissue viability, perfusion

• Therapeutic drug monitoring

Diffuse Optics Applications

Mammography: Poor Performance inDense Breast (~60% sensitivity)

Breast Cancer Motivation

Page 12: Medical Imaging in Thick Tissues Using Diffuse Optics

Consequences of Age Related

Changesx-ray mammograms (normal breasts)

DENSEMILDY

DENSE FATTY

AGE

http://homearts.com/depts/health/a8bhty51.htm

Page 13: Medical Imaging in Thick Tissues Using Diffuse Optics

1) Detection

• Pre- Peri-menopausal, high risk subjects

2) Guiding Therapies• Intraoperative (nodes, margins)

• Neoadjuvant Chemotherapy

locally advanced disease

Breast Cancer Role

(>600 subjects NTROI-wide)

Page 14: Medical Imaging in Thick Tissues Using Diffuse Optics

Broadband Frequency Domain Photon MigrationPham, Tromberg, et al., Rev. Sci. Instr., 71, 2500, (2000)

light scattering tissuesµ µa s’,

Frequency-DomainInstrument

I

time (ns)

sourcelight

detectedlight

�(�,�)

�(�,�) NonlinearLeast

Square Fits

Experimental Response

�(�,�)

�(�,�)

NIR TissueSpectroscopy

SpectroscopicAnalysis

Bulk TissueFunction &Structure

Theoretical Response

|�(�,µ ,µ )|a s’

�{ �,µ ,µa s’�( )}

::

White

light

Laser

diodes

Spectro-

graph

APDFDPM

SS

z

Measurement

time: ~15 sec

Bevilacqua, et al., Applied Optics, 39, 2000.

Combined FDPM and Steady State Spectroscopy

Broadband Diffuse Optical Spectroscopy

Page 15: Medical Imaging in Thick Tissues Using Diffuse Optics

The Laser Breast Scanner

-50-500 MHz (FDPM)

-Full NIR (600-1000nm) ~ 30 s

-Point-point scan measurement

The LBS handheld probe

Pham, TH., et al. Review of Scientific Instruments, 71 , 1 – 14, (2000).Bevilacqua, F., et al. Applied Optics, 39, 6498-6507, (2000).

The Laser Breast Scanner

-50-500 MHz (FDPM)

-Full NIR (600-1000nm) ~ 30 s

-Point-point scan measurement

The LBS handheld probe

Pham, TH., et al. Review of Scientific Instruments, 71 , 1 – 14, (2000).Bevilacqua, F., et al. Applied Optics, 39, 6498-6507, (2000).

Page 16: Medical Imaging in Thick Tissues Using Diffuse Optics

Hand-held Scanner

10 mm

+30

-30

-10

0

-20

+10

+20

SCAN

DIRECTION

-Y

+Y x y

Linescan Geometry

10 mm

+30

-30

-10

0

-20

+10

+20

600 650 700 750 800 850 900 950 1000 10500.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

AB

SO

RP

TIO

N (

mm

-1

)

WAVELENGTH (nm)

600 650 700 750 800 850 900 950 1000 10500.66

0.68

0.70

0.72

0.74

0.76

0.78

0.80

0.82

0.84

0.86

0.88

0.90

0.92

RE

DU

CE

D S

CA

TT

ER

ING

(m

m-1

)

WAVELENGTH (nm)

600 650 700 750 800 850 900 950 1000 10500.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

AB

SO

RP

TIO

N (

mm

-1

)

WAVELENGTH (nm)

600 650 700 750 800 850 900 950 1000 10500.66

0.68

0.70

0.72

0.74

0.76

0.78

0.80

0.82

0.84

0.86

0.88

0.90

0.92

RE

DU

CE

D S

CA

TT

ER

ING

(m

m-1

)

WAVELENGTH (nm)

600 650 700 750 800 850 900 950 1000 10500.66

0.68

0.70

0.72

0.74

0.76

0.78

0.80

0.82

0.84

0.86

0.88

0.90

0.92

RE

DU

CE

D S

CA

TT

ER

ING

(m

m-1

)

WAVELENGTH (nm)

600 650 700 750 800 850 900 950 1000 10500.000

0.002

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

AB

SO

RP

TIO

N (

mm

-1

)

WAVELENGTH (nm)

Y= +20 mm

Y= 0 mm

Y= -30 mm

SCAN

DIRECTION

-Y

+Y

Page 17: Medical Imaging in Thick Tissues Using Diffuse Optics

Linescan Geometry

Parameters:�Hb-R

�Hb-O2

�Lipid

�H2O

�SP

�PRE

Indices:�THC

�Hb-Sat

�…-5 -4 -3 -2 -1 0 1 2 3 4 5

0.4

0.5

0.6

0.7

0.8

0.9

1.0

PA

RA

ME

TE

R

POSITION

TMAX

TBASE

TPEAK

TAVG

– What Do Tumors “Look” Like?

– Can Optical Signatures Be Used for Diagnosis?

– Can Optics Monitor Therapy?

– Validate Optical Signatures by Conventional Imaging?

– Can Optics Predict Individual Therapeutic Response?

Major Questions

Page 18: Medical Imaging in Thick Tissues Using Diffuse Optics

– What Do Tumors “Look” Like?

– Can Optical Signatures Be Used for Diagnosis?

– Can Optics Monitor Therapy?

– Validate Optical Signatures by Conventional Imaging?

– Can Optics Predict Individual Therapeutic Response?

Major Questions Population Statistics

27±21Avg. Tumor Size (mm)

27.5±7.1BMI (m2/kg)

23.5Median Tumor Size (mm)

50.5± 13.8Age (years)

58Lesions (#)

57Subjects (#)

ValueItem

Invasive Ductal Carcinoma Study

Page 19: Medical Imaging in Thick Tissues Using Diffuse Optics

Averages (N=58)

650 700 750 800 850 900 950 1000

0.004

0.006

0.008

0.010

0.012

0.014

0.016

0.018

0.020

0.022

0.024

TBASE

AB

SO

RP

TIO

N(m

m-1

)

WAVELENGTH(nm)

TMAX

HHb & O2Hb

H2O & Lipid

Max Contrast (N=58)

0.026#37.5146PRE

0.0038*0.634±0.2780.830± 0.412SP

<0.0001*20.0±10.533.8±21.0WATER

<0.0001*63.2±12.349.7±18.0LIPID

<0.0001*14.6±7.521.5±11.3Hb-O2

<0.0001*5.93±2.429.98±5.02Hb-R

pNormalTumor

Invasive Ductal Carcinoma Study, <d>=2.7±2.1 cm

Page 20: Medical Imaging in Thick Tissues Using Diffuse Optics

Tumor Stratification by Age

650 700 750 800 850 900 950 10000.000

0.005

0.010

0.015

0.020

0.025

0.030

E

C

D

A

AB

SO

RP

TIO

N(m

m-1

)

WAVELENGTH ( nm)

B

AGE(< 30)

(30-39)

(40-49)

(50-59)

(>60)

Page 21: Medical Imaging in Thick Tissues Using Diffuse Optics

Summary: Tumor Detection

Encouraging Findings:

– Optical contrast of tumors: more than Hb

– Functional baselines are age dependent

– Evidence of success in women < 50

• Not prospective study

– What Do Tumors “Look” Like?

– Can Optical Signatures Be Used for Diagnosis?

– Can Optics Monitor Therapy?

– Validate Optical Signatures by Conventional Imaging?

– Can Optics Predict Individual Therapeutic Response?

Major Questions

Page 22: Medical Imaging in Thick Tissues Using Diffuse Optics

Monitoring of NAC

– Individualize treatment to optimize survivaland quality of life

– Complete pathological response increasedsurvival (NSABP trial, Fisher et al J Clin Onc, 1998)

– Need imaging to predict pathologicalresponse (pR)

Conventional Methods

• Recent 31 patient study correlationwith pathology:

– Palpation: 19%

– Mammography: 26%

– Ultrasound: 35%

– MRI: 71%

Yeh, E., et al., AJR Am J Roentgenol, 184, 868-77 (2005)

Page 23: Medical Imaging in Thick Tissues Using Diffuse Optics

Kinetic MRI and DOS

+33.0-17.1-43.7-69.0-39.7-36.4Difference

(%)

Lipid

avg, %

Water

avg, %

ctTHb

avg, �M

% tumor

volume

SER �1.30

Peak

Enhancement

Tumor

Volume

(cc)

Shah, N., et al. J. Biomed Opt, (2005)

Post 1 Post 4

Jakubowski, D. et al. J Biomed Opt, (2004)

Long-term tracking: ~12 weeks

Page 24: Medical Imaging in Thick Tissues Using Diffuse Optics

Why Optics and not MRI?

• Do methods probe different regions?

• Differential sensitivity?

– DCE-MRI: vessel resolution limit ~mm

– Optics: Sensitivity to <1 �M changes,

microvessels

Page 25: Medical Imaging in Thick Tissues Using Diffuse Optics

MRI-Optics Co-Registration Possible Mechanisms

• Early optical sensitivity to cell death

– Drop in Hb: reduced O2 consumption,

– Drop in H2O: loss of cellular water (MRI:increase ADC)

• Changes prior to MRI

– Similar to MRS choline signal

Page 26: Medical Imaging in Thick Tissues Using Diffuse Optics

– What Do Tumors “Look” Like?

– Can Optical Signatures Be Used for Diagnosis?

– Can Optics Monitor Therapy?

– Validate Optical Signatures by Conventional Imaging?

– Can Optics Predict Individual Therapeutic Response?

Major Questions Complete Responder

0 20 40 60 80 100 1200.04

0.05

0.06

0.07

0.08

0.09

0.10

0.11

0.12

TO

IA

VE

RA

GE

DAY

TUMOR AVERAGENORMAL AVERAGE

2306-13

Tumor approaches contra-lateral normal baseline

pathologic Response (pR)

Predict pR after 1 week?

Page 27: Medical Imaging in Thick Tissues Using Diffuse Optics

Pathology Response Predictions

757510050100SPECIFICITY (%)

4343715786SENSITIVITY (%)

SPLIPIDH2OO2HbHbITEM

100

100

Hb

& H2O

Based on Optical Measurements at day 6

11 patients, AC Therapy

Page 28: Medical Imaging in Thick Tissues Using Diffuse Optics
Page 29: Medical Imaging in Thick Tissues Using Diffuse Optics