D-Sense
Safe and effective operation
Large LED display
Compact and Easy to Use
Battery Operation
Simulate five type of waveforms
Ideal for All AEDs
Meditech Equipment Co., Ltd (Meditech Group) Address: Qingdao, Shandong Province, P. R. China
Tel: +86-532-85832673 81705331
Fax: 86-532-81705332
Email: [email protected]
www.meditech.com.cn
![Page 2: Meditech D-Sense automated external defibrillator analyzer](https://reader038.vdocument.in/reader038/viewer/2022100803/5a6841bf7f8b9a45728b5095/html5/thumbnails/2.jpg)
OUTPUT POWER MEASUREMENT:
Load Resistance: 50ohms±1%. 100 watt
Range : 2-250J Resolution : 0.1 Joule Max. Voltage : 2400 Volts Max. Current : 48 Amps Measurement Trip Level : 70±5 Volts Pulse Width : 0.5-60 ms Accuracy : ±2%
ECG WAVEFORMS: QRS complex Rates : 60,90 BPM Rate
Accuracy : ±1%
Amplitude : Fixed at 1mV (Apex-Sternum)
Accuracy : ±5%
PERFORMANCE WAVEFORMS: Sine Wave: 10Hz
Amplitude: Fixed at 1mV (Apex-Sternum)
Accuracy: ±5%
ARRHYTHMIAS: Vfib,Vtach
Rate Accuracy: 1% Amplitude: Fixed at 1mV
(Apex-Sternum) PEAK/OVERSHOOT: Voltage Accuracy:±5Volts POWER: CR123A-4x2N LiMnO2 12-Volt WEIGHT: 1Kg DIMENSIONS: 256x220x(L:65 H:80)mm
ENVIRONMENTAL: Storage Temperature: -20 to 55℃
Operating Temperature : 0 to 40℃