Page 1: Meditech D-Sense automated external defibrillator analyzer

D-Sense

Safe and effective operation

Large LED display

Compact and Easy to Use

Battery Operation

Simulate five type of waveforms

Ideal for All AEDs

Meditech Equipment Co., Ltd (Meditech Group) Address: Qingdao, Shandong Province, P. R. China

Tel: +86-532-85832673 81705331

Fax: 86-532-81705332

Email: [email protected]

www.meditech.com.cn

Page 2: Meditech D-Sense automated external defibrillator analyzer

OUTPUT POWER MEASUREMENT:

Load Resistance: 50ohms±1%. 100 watt

Range : 2-250J Resolution : 0.1 Joule Max. Voltage : 2400 Volts Max. Current : 48 Amps Measurement Trip Level : 70±5 Volts Pulse Width : 0.5-60 ms Accuracy : ±2%

ECG WAVEFORMS: QRS complex Rates : 60,90 BPM Rate

Accuracy : ±1%

Amplitude : Fixed at 1mV (Apex-Sternum)

Accuracy : ±5%

PERFORMANCE WAVEFORMS: Sine Wave: 10Hz

Amplitude: Fixed at 1mV (Apex-Sternum)

Accuracy: ±5%

ARRHYTHMIAS: Vfib,Vtach

Rate Accuracy: 1% Amplitude: Fixed at 1mV

(Apex-Sternum) PEAK/OVERSHOOT: Voltage Accuracy:±5Volts POWER: CR123A-4x2N LiMnO2 12-Volt WEIGHT: 1Kg DIMENSIONS: 256x220x(L:65 H:80)mm

ENVIRONMENTAL: Storage Temperature: -20 to 55℃

Operating Temperature : 0 to 40℃

Maximum Humidity: 90% Relative Humid