method for determination of 25-oh vitamin d2 and d3 in ... · interface temperature 300 °c dl...
TRANSCRIPT
Three different immunological methods at four different locations were used in our laboratory for determination of vitamin D. Comparison of results showed some discrepancies in methods.
Objective
Results and discussion References
Standards and controls were made in-house in 5% albumin solved in Krebs-Ringer buffer. A five point calibration curve was used and it covered 10 to 360 nmol\L for both 25-hydroxyvitamin D2 and 25-hydroxyvitamin D3. 25-hydroxyvitamin D2-d6 and 25-hydroxyvitamin D3-d6 was used as internal standards.
Department of Medical Biochemistry, Vestre Viken Hospital Trust, Drammen, Norway
Per Olov Nordstrøm, Sylwia Wolosowska, Vivi-Ann Tennfjord and Trine Lauritzen
Development of an automated high throughput HPLC-ESI-MS method for determination of 25-OH vitamin D2 and D3 in serum
The figures above shows product ion scan for 25-OH vitamin D2 and D3. The known vitamin D quantitation methods without derivatization, use different ions and collision energies for quantification [4,5,6]. The transitions found for each compound are presented in Table 3. The quantitation was performed using Shimadzu LCMS instrument which allows making a total ion chromatogram (TIC) for each compound. The aim of using TIC was to increase the sensitivity and improve the repeatability for the method. However, our study showed only a minor impact on these parameters.
The purpose of this project was to develop a LCMS procedure that could replace these methods. It should be fully automated and be able to handle up to 600 samples per day on one LCMS instrument.
Table 1. LC configuration and parameters
LC system Shimadzu Nexera HPLC with Shimadzu trippel quadrupole MS 8050
Column Raptor biphenyl (30 x 2.1 mm )
Column temp 40 °C
Flow rate 0.8 mL/min
Mobile phase 0.1 % Formic acid in water : Acetonitile (45:55); isocratic
Injection volume 15 µL
Run time 1 min
Sample preparation was performed on a pipetting robot. A precipitation reagent containing internal standard and magnesium sulphate was used for protein crash from serum. Hexane was added for vitamin D liquid–liquid extraction. The hexane layer was then transferred to a new sample plate, evaporated to dryness and the samples were reconstituted in a solvent compatible to the LCMS method. Similar method of sample purification has been used by others [2,3,4]. The details of sample preparation is given in Figure 1. The separation was performed on a Restek’s biphenyl column. Mass spectrometry was conducted using positive electrospray ionization. MS conditions are presented in Table 2. Run time:
One minute!
Table 4. Representative accuracy and precision for analytes
25-hydroxyvitamin D3 25-hydroxyvitamin D2
CV (%) Accuracy (%) CV (%) Accuracy (%)
10 nmol/L 6.4 129 13.6 116
50 nmol/L 9.1 100 7.0 106
260 nmol/L 3.5 97 3.1 100
Conclusions
We have developed a high throughput LCMS method for analyzing 25-OH vitamin D2 and D3 in serum that is well suited for large samples amount. The procedure includes an automated liquid-liquid extraction performed on a pipetting robot. The method was accurate and reproducible and was successfully applied to routine analysis.
Table 3. MS parameters (* qualifier)
Compound Precursor
ion Fragment
ion Q1 Pre Bias Q3 Pre Bias CE
25-OH vit D2 395,20 259.40 -14 -29 -14
119.00* -27 -12 -21
25-OH vit D3 383,00 229.30 -30 -11 -20
185.20* -30 -12 -21
297.50 -30 -29 -17
211.20 -30 -10 -23
128.10 -30 -24 -55
25-OH vit D2-d6
401,20 119.20* -28 -22 -23
269.20 -19 -30 -23
25-OH vit D3-d6
389,40 303.30 -11 -30 -18
128.10* -14 -28 -55
Measuring vitamin D has been increasingly popular over the last decade. This is probably caused by the large number of studies suggesting links between vitamin D status and body functions other than the well established bone related. There have been found links to the immune system, cardiovascular conditions and certain types of cancer etc [1].
25-hydroxyvitamin D2 and D3 calibrations curves were constructed based on biologically relevant concentrations of 10, 25, 60, 150 and 390 nmol/L (standard curve fit type: quadratic, weighting method 1/c˄2). This calibration model worked well with a coefficient of determination (r2) 0.994 and 0.993 for 25-OH vitamin D2 and D3 respectively. The limit of quantitation for both analytes were established at 10 nmol/L.
Experimental
Sample preparation
50,0 75,0 100,0 125,0 150,0 175,0 200,0 225,0 250,0 275,0 300,0 325,0 350,0 375,0 m/z
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1,0
1,1
1,2
1,3
1,4
1,5
Inten.(x100,000)
395
378
9177115
17611391
126
12877 11591
142
55
153 189
177217
115
91
129
16555 79
178
141
191
212
105
55
6791 128
141
165
208178
195223
11955
95
69
107
169
145
155197
212
251224
269
281
395119
37783
269
137
251
161
197 229211 335
107
171
57 69325
310283
50,0 75,0 100,0 125,0 150,0 175,0 200,0 225,0 250,0 275,0 300,0 325,0 350,0 375,0 m/z0,00
0,25
0,50
0,75
1,00
1,25
1,50
1,75
2,00
2,25
2,50
2,75
3,00
3,25
3,50
3,75
4,00
Inten.(x1,000,000)
126 17777 1551035742 14114112877 16515210355 65
11577
14115265 16655
102
91
128
141
55
65 153105 165
91
128
55141
105
67
153
167
79143
55 93 157119
183169
21169
237225
365
383
185229
211
171253121
135
29781 159 323
95
57271
283
383
365
Figure 3. Chromatogram of 25-OH vitamin D2 and D3
25-hydroxyvitamin D3
25-hydroxyvitamin D2
The sample preparation was performed using a robot with a maximum capacity of 384 samples per one setup. Preparation time for one sample set is approximately three hours. Maximum capacity required in our laboratory is about 600 samples per day. Further work is planned to increase the sample preparation capacity.
25-hydroxyvitamin D3
25-hydroxyvitamin D2
The Restek’s biphenyl column was used for the separation. 25-hydroxyvitamin D3 and 25-hydroxyvitamin D2 were almost fully separated and eluted at 0.55 and 0.65 minute in LC respectively (Figure 3). The total time per sample on the instrument was 1.5 minute. Equally separation were not achievable with the commonly used C18 columns.
The precision was determined by analyzing six replicates at each level. Results are presented in table 4. An external control from Chromsystems (MassCheck level II, c = 100 nmol/L) is analyzed on a daily basis. The reproducibility is 7.4% and 6.0% and the accuracy 110% and 103%, for 25-OH vitamin D2 and D3 respectively (results of the last seven months). We also participate in the DEQAS quality control program for vitamin D. The accuracy on the latest round was between 82% to 108% (average 95%, n=5) for 25-OH vitamin D3. The method has been in use in our laboratory for ten month with good results.
Acknowledgement
1. Teresa Kulie et al, Vitamin D: An Evidence-Based Review, JABFM, 2009, vol. 22 no 6 2. Naesgaard Patrycja A et al, Serum 25(OH)D Is a 2-Year Predictor All-Cause Mortality, Cardiac
Death and Sudden Cardiac Death in Chest Pain Patints from Northen Argentina, PLOS ONE 2012 Vol 7 Issue 9
3. Netzel Brian C et al, Increasing Liquid Chromatography-Tandem mass Spectrometry Throughput by mass Tagging: A Sample-Multiplexed High- Throughput Assay for 25-Hydroxyvitamin D2 and D3, Clinical Chemistry 57:3 (2011) p. 431-440
4. Maunsell Zoë et al, Routin Isotope-Dilution Liquid Chromatography-Tandem Mass Spectrometry Assay for Simultaneous Measurement of the 25-Hydroxy Metabolites of Vitamin D2 and D3, Clinical Chemistry 521:9 (2005) p. 1683-1690
5. Shah Iltaf et al, Misleading measures Vitamin D analysis: A novel LC-MS/MS assay to account for epimers and isobars, Nutrition Journal 2011, 10:46
6. Calton Lisa J. et al, The Analysis for 25-Hydroxyvitamin D in serum using UPLC/MS/MS, Waters application note # 720002748 (2008)
Thanks for support and good cooperation to Shimadzu Europa GMBH, Restek and Teknolab AS.
Table 2. MS/MS conditions Ion mode Positive Interface temperature 300 °C DL temperature 125 °C Nebulizer gas flow 3 L/min Heating gas flow 8 L/min Heat block temperature 250 °C Drying gas flow 8 L/min
Figure 1. Sample preparation
Figure 2. Product ion scan of 25-OH vitamin D2 and D3
Sample: 50 µL serum
Add 75µL of precipitation reagent with IS
Add 100µL of MgSO4
Add 600 µL of hexan
Shaking Centrifuge
Transfer 450 µL of hexane phase
Evaporation to dryness
Reconstitution in 200 µL of solvent
Analyze by LCMSMS
Mixing