METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL PRESSURE AND COMPOSITIONAL ANALYSIS

Robert E. Ellefson REVac Consulting

Dayton, OH 45459 USA [email protected]

ERMP IND12 – QMS Workshop Bled, Slovenia

April 10-13, 2012

Page 2: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

What do Industrial & R/D Users

Need and Expect?

• Partial Pressures or Composition of Process Gas

• Expect Stable Operation to Trust the Data

– How do they Know it is Stable? Accurate?

• Process Control practices typically call for a Verification of Performance

– Test Stand Calibration of RGA may Qualify the MS

– In Situ Measurement of Reference Mixture is needed to provide Verification that the RGA is “In Control”

Page 3: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Define a MS Measurement and Calibration Method

for a Process or Vacuum System

Logic

Plan:

Application

Sampling

Design

Select and

Qualify MS

Calibration

Design

Quality

Assurance

CVD, Etch, ALD, Degas

0.01 mb < PProcess< 100 mb

Vis., Trans. or Molecular

Process Gas Species(Any Condensibles?)

Viscous, Transition or

Molecular Flow [PProcess ]

Orifice / Small Tube toMS

(Heated Inlet and QMS)

Closed Ion Source (CIS)

Low eV: e.g. 40 eV

Low Emission: e.g. 0.2 mA

Mass Range for Species

Detector: FC & EM(?)

Demo: Linearity, S vs t, A-B

Std Mix: Process Species

Reservoir: PV for 1 year

Cal Flow Std: 1x10-4

mb-l/s

Flow Mix into CIS

Use Std Mix for QA Data

Log Data & Plot PPi vs Time

Plot Xi =PPi / Sum PPi vs Time

Establish Action Limits for Sens or Gain

Atmospheric Processes

100 mb < PProcess < 2 bar

Viscous Flow

Process Gas Species(Any Condensibles?)

Viscous Flow

Capillary Sampling with

Sample Pump and

Molecular Leak to MS

(Heated Inlet and QMS)

Std Mix: Process Species

Volume: 5-10 L@ 1 Atm

Capillary Sampling of

Standard Mixture

Vacuum: PVD- UHV -XHV

PProcess < 0.01 mb

Molecular Flow

Process Gas Species(Any Condensibles?)

Molecular Flow

Direct Insertion of QMS

(Isolation Valve)

Bakeout Provision

Open Ion Source (CIS)

High eV: 70 - 100 eV

High Emission: e.g. 2 mA

Mass Range for Species

Detector: FC & EM(?)

Demo: Linearity, S vs t, A-B

Page 4: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

RGA Conditions for Accurate Partial Pressure and Compositional Analysis begins in the Ion Source

• Vacuum System Base Pressure PP Measurements Requires Sensitivity

– OIS: 70 – 110 eV e-; 1-2 mA Emission

• Process Gas Composition allows Lower Sensitivity operation which minimizes Fragmentation

– CIS: 35-40 eV e-; 200 μA Emission

• Ion Extraction, Focus and Ion Energy Potentials affect Space Charge, Ion Residence Times and Linearity

Page 5: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Evidence of Gas Scattering Loss at Higher Pressure is seen as Ions Traverse the Mass Analyzer to the Detector.

XPR Correction for Gas Scattering is I = Imeas eKP ; Gives PPi’s

Page 6: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

The Potential Well formed by the Ionizing Electron Beam Lengthens Ion Residence Time (Longer Path) in Ion Source

CIS equipotentials

1V well

(SIMION)

10V

70V

75V

79V

79.9V

E-Beam Well Depth For the Geometry of this CIS:

Vwell = - 3800 ie / (Ve)1/2

Vwell(40eV/200uA) = - 0.12 V

Vwell(70eV/2000uA) = - 0.90V

Note for a given ie the well gets deeper

for lower Ve (slower electrons; higher ρe)

R Ellefson and M Vollero, AVS-57, 2010

Page 7: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

The Same Open Ion Source can be Linear or Non-Linear depending on Operating Potentials

Page 8: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

The Ar and O2 sensitivity decreases in OIS with time due to O2 Oxidation of Ion Source Grids changing surface potentials.

However Ratio O2 /[O2 +Ar] remains Nearly Constant

Data provided by Wm Sproul

0.0

1.0

2.0

3.0

4.0

5.0

0.0 50.0 100.0 150.0 200.0 250.0

Time, minutes

O2 a

nd

Ar

Pa

rtia

l P

res

su

res

an

d

(O2/O

2+

Ar)

*10

0 R

ati

o

Ratio = 0.0348 + 0.4%

O2+ (E-10 A)

Ar+ (E-9 A)

• Conditioning • RGA exposed to pure Ar

for many hours Reduces the Surface Oxide:

H(Diff) + OH(S) H2O

• Exposure to Ar +3% O2

changes the sensitivity for both Ar and O2 by surface re-oxidation with a

time constant of ~ 2hrs.

• Practical Solution for PVD Process Control:

PP (O2) = R(O2)*P(CDG)

Where R (O2) = O2 /[O2 +Ar]

Recommended Cal Reference: 3% O2 in Ar

Page 9: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Examples of In Situ RGA Calibration Methods

• In Situ Calibration for an Open Ion Source (OIS) RGA

• In Situ Calibration for a Closed Ion Source (CIS) RGA

– System Calibration using a Fixed Reference Supply with Sampling equivalent to the Process Sampling

– Portable Calibration Reference Source introduced directly into the CIS

Page 10: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

In Situ Calibration: A Reference Pressure & Composition is established at the IG and RGA when

the Valve to the Gas Mixture is Open

• Pumping is provided by the Vacuum System

• A Calibrated Fixed-Flow Rate produces a Reproducible Pressure and Composition at IG and RGA Ionizers

• The Pressure is

Pcal = Qcal / Ccal

where CCal can be Calculated

from Geometry.

• Mixture Composition chosen for Application

• Vacuum System must tolerate the Qcal Flow Rate

Page 11: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

For UHV and XHV Systems, a Lower Calibration Pressure can be created with a Q vs PFill Calibration

The Plot shows Flow Rate, Q; The Pressure generated in the example is ~ Q/10. PFill can be adjusted without altering Composition using the Gas Pipettes. Suggested UHV Composition: 90% H2; 9% CO; 1% CO2

Q = A Pfill2

Page 12: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

A Two-Stage Pressure Reduction with multiple probe types enables sampling at a representative

point over a wide range of process pressures

• Process Gas Species in CIS reflect Process: Molecular Flow into CIS; Molecular Flow out • Calibration Reference Mix Species are Altered by Mass:

Viscous Flow into CIS; Molecular Flow out (Addressed Later)

Page 13: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Sensitivity Monitoring Data shows Shifts

in Sensitivity Related to Process Changes

MASS(40) Sensitivity vs Daysfor CIS2 [W Filament; 40eV/200uA]

Reference Gas: VTI Flow Std with INFICON Mix

2.0E-04

2.2E-04

2.4E-04

2.6E-04

2.8E-04

3.0E-04

3.2E-04

3.4E-04

0.00 5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00

Time (Days)

Ar-

40

Se

ns

itiv

ity

(A

/To

rr)

Event: Accidental Air Venting of

CIS2 Vacuum due to Power Failure

+ 4.8% 2-Sigma

+ 5.5% 2-Sigma

Page 14: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Tritium Fume Hood

Process Capillaries

Tritium Glove Box

Tritium Glove Box

Standards Capillaries

D/TPump

V

T

T

P

D/TPump

D/T/He - 1

D/T/He - 4

D/T/He - 2

D/T/He - 3

30 – 60 m

CDG

3 L

Magnetic SectorMass

Spectrometer

Process Room Analytical Lab

<< 1 mbSample

Pressure

Remote Process Monitoring: D/T/He Gas Mixture Compositional Measurements

Sampling Capillaries 30 – 60 m long Capillaries to Standard Mixtures (Similar) Capillary Sample Purged 3 – 5 Capillary Vol’s ~ 0.1 mb sample P in 3 L volume

Species deplete through Molecular Leak Record time for peak for t=0 Composition Analysis Time: ~ 20 minutes

Qualify MS with Std Mix each AM + Run Time Enables < 0.5% relative accuracy for D/T/He-3

Page 15: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

The D/T/He-3 mixtures were used to Assess Accuracy and do Daily Qualification of the MS’s

• A daily routine for operators was to run an analysis of a known mixture to check performance

• Call Customers “We are ‘up’ today” so they could run a Test, etc.

• If a MS was “Out of Control”, a re-run was done to note permanent shifts in calibration (due to power failures, etc)

Page 16: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

RGA Calibration Source

•Fixed Flow Rate Gas Source •Pumping by Process through a Fixed Conductance (~ 10 L/s)

Process Monitor

Calibration Gas Source

•Fixed Flow Rate Gas Source •Pumping by Monitor Vacuum System through the Closed Ion Source (~ 1 L/s)

Can we define a Multi-Application,

Portable InSitu Calibration Gas Source?

2x10 -5 mbar

Vacuum Process

2x10 -7 mbar

Q=2x10 -4 mbar-l/s

Process

Q=2x10-4 mbar-l/s

2x10-4 mbar

1x10-5 mbar

Page 17: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Portable Calibration Sources have

been used for many years

• Crimped-Capillary Leak Standards (Boeckmann, et al.)

• VTI Positive Shut off Flow Standards (Traceable to NIST) • Sandia Flow Standards [Chamberlin, et al. JVST A7 (1989)]

• High accuracy primary flow references (Traceable to NIST)

INFICON has used their compact Reference Gas Source

for InSitu Calibration of OIS and CIS products for over

10 years

• SS Frit Flow Element For Stable Flow

• Low dead volume valve to minimize turn-on bursts

• Measured Flow Rate stated on Label (+ 10% of Rate)

• Predictable Flow Rate (Depletion < 10%/Yr normal use) • Gas mixture composition is constant (viscous flow)

• Gas fill Pressure < 2.8 bar (non-compressed gas for simple shipping)

• Flow Rate can be independently Certified at a Standards Lab

Page 18: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

INFICON Calibration Reference Source

• 4-VCR Male Connection • Air Operated Valve • 38mm- (+VCR) x 270mm • 132 cc Volume

Page 19: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Flow Through Element goes as Pfill

2

RGA Calibration Pressure vs Fill Pressure of Reference

0.0E+00

2.0E-06

4.0E-06

6.0E-06

8.0E-06

1.0E-05

1.2E-05

1.4E-05

1.6E-05

0 0.5 1 1.5 2 2.5 3

Fill Pressure (Ar) [bar]

RG

A C

alib

rati

on

Pre

ss

ure

: Q

(P

) / C

[m

bar]

PP(Ar) = Q(P) / C

Q(P) / C = A P2

Vacuum Process

Q=2x10 -4 mbar-l/s

Page 20: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

RGA Calibration Pressure Depletion is

Predictable with Time of Flow

0.0E+00

5.0E-06

1.0E-05

1.5E-05

2.0E-05

2.5E-05

0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0

Elapsed Flow Time (Days)

PC

al[A

r-4

0]

(mb

ar)

Model for Depletion of

Calibration Reference

Pressure:

PCal(t) = Pcal(0) / [1+Qo t/PoV]

Qo = 2.8x10-4

T-L/s

Fill Parameters:

Po = 2053 Torr [2.7 bar]

V = 0.135 L

PCal[Ar-40] (t)

Expected 10% Depletion in 1 year of automated use

(5 minutes of flow/day )

Page 21: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Composition Ratio of 50/50 He/Ar Mix does not change with

30% Depletion of Fill Gas

Ratio 40/4 = -0.0041 t + 1.1627

0.0

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

10.0

0.00 2.00 4.00 6.00 8.00 10.00 12.00

Time (Days)

Ion

Cu

rre

nt

(A -

Co

rre

cte

d t

o R

ef

mix

) &

Ra

tio

40

/4

Flow Rate Depletion: Ar-40

(Data Sensitivity Corrected)

Flow Rate Depletion: He-4

(Data Sensitivity Corrected)

Page 22: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Pcal produced from Cal Reference Source changes

with Ambient Temperature by + 0.15 %/oC

4.06E-09

4.08E-09

4.10E-09

4.12E-09

4.14E-09

4.16E-09

4.18E-09

4.20E-09

4.22E-09

4.24E-09

4.26E-09

20 25 30 35 40 45 50

Average Temperature [Valve & Tank] (C)

Ion

Cu

rre

nt [A

r-40

] (A

) fr

om

PC

al

Temperature Coefficient of Flow Rate

0.15% / oC

Page 23: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

INFICON Gas Mixture options allow

choice of gas pumped by the process

Component Argon (99.995%) Ar with 5% Impurities PVD Mix

Argon 99.995 % 95 % Balance

H2 - 1.0 % 200 ppm

He - 1.0 % 1000 ppm

N2 - 1.0 % 50 ppm

CO2 20 ppm

Kr - 1.0% 1000 ppm

Xe - 1.0 % 1000 ppm

Page 24: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Composition of the Gas Mixture in the Ionizer is

altered by the RGA’s molecular flow pumping

• The Mixture is in viscous flow from the Calibration Reference Source. The composition entering the vacuum chamber is the stated Mixture. The partial flow rate qi of a species is qi (in) = Xi (Ref) Qo (mbar-L/s) • The partial flow rate out of the chamber depends on the mass of the species, Mi and the pumping system conductance, CN2 at the ion source: qi (out) = PPi Ci = PPi CN2 [28 / Mi]

1/2

But qi (in) = qi (out)

So at the ionizer: PPi (Ion Source) = [Mi / 28]1/2 Xi (Ref) Qo / CN2

From which a Sensitivity Factor can be calculated:

SFi (mb/A) = PPi (Ion Source) / [ Ii – Interference Contributions ]

Page 25: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Composition of Mixture at the MS ionizer shows

depleted light gases and enhance heavy gases

CEM 1:1:1 Ar-He-N2 Mix

Component Tank Mix Xi -Ion Source

Ar 33.33 46.4492392

He 33.33 14.6885391

N2 33.33 38.8622217

Page 26: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Mass Spectrum of a PVD Mixture

Calibration Reference Source

1.0E-14

1.0E-13

1.0E-12

1.0E-11

1.0E-10

1.0E-09

1.0E-08

0 10 20 30 40 50 60 70 80 90 100

Mass

Ion

Cu

rre

nt

(A)

INFICON PVD Mixture

Calibration Reference Source Flow RATE 1x10-4 mbar-l/s @ 23.5 oC

Transpector CIS: 70eV/2000uA/EM

200 ppm

H 2 +

1000 ppm

He

Ar +

Xe ++

Kr +

40 Ar

++

Page 27: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Mass Scale Calibration with

Calibration Reference Source

H2 & He 36Ar & 38Ar Kr

Page 28: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Electron Multiplier Gain Adjustment

• Adjust RGA EM (HV) to give Ion Current Level:

• e.g.: PCal = 2x10-5 mbar

SFC = 2x10-4 A/mbar

Gain = 200

Species Abundance Ion current (EM)

Ar-36 3370 ppm 8.0x10-7 A

Ar-38 630 ppm 1.5x10-7 A

Ar-40 99.6% Off Scale

1% N2 1% 2.4x10-7 A

• Measure Gain of the EM at a value of High Voltage

Gain = I(Ar-38: EM)

I(Ar-38: FC)

Page 29: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Sensitivity and Gain can be quickly measured with sequential

FC/EM scans with Selected Ion Monitoring of Ar species.

1.0E-13

1.0E-12

1.0E-11

1.0E-10

1.0E-09

1.0E-08

1.0E-07

1.0E-06

1.0E-05

00:00.0 00:08.6 00:17.3 00:25.9 00:34.6 00:43.2 00:51.8 01:00.5 01:09.1 01:17.8 01:26.4

Time (m:s)

Ion

Cu

rre

nt

(A)

1.0E-10

1.0E-09

1.0E-08

1.0E-07

1.0E-06

1.0E-05

1.0E-04

1.0E-03

CIS

To

tal

Pre

ss

ure

(m

b)

Ar-40 EM (Off Scale)

FC FCEM

Pressure: Calibration Reference ON

Gain = 445 + 8

Gain = 443 + 38

S(Ar-40) = 1.1E-5 A/mb

Ar-36 FC

Ar-38 FC

Ar-40 FC

Ar-38 EM

Ar-36 EM

Page 30: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Sensitivity Monitoring Data shows Shifts

in Sensitivity Related to the Sensor

MASS(40) Sensitivity Vs TimeFor CIS2 (W filament; 40eV/200uA)

Reference Gas: Inficon/VTI Calibration Standard

2.20E-04

2.25E-04

2.30E-04

2.35E-04

2.40E-04

2.45E-04

2.50E-04

2.55E-04

2.60E-04

0 2 4 6 8 10 12 14 16

Time (Days)

Ar-

40

Se

ns

itiv

ity

(A

mp

/To

rr)

2-Sigma Deviation 5% of Average over 15 days

5% Shift in Average Sensitivity on Day 7

2-Sigma: 2% over 7 days

2-Sigma: 3% over 8 days

Page 31: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Composition Calculations: Accurate Compositional Analysis is the Goal for Process Monitoring

• QMS measures the Ion Current vs Time for the Gas being sampled

• Partial Pressures of Species Present are Calculated Ion Currents with Corrections for Interferences, and Sensitivities

• Mol-% Composition is defined as

Xi(mol-%) = 100 * PPi / ∑ PPj

For Quality Assurance, Accuracy is measured with known mixtures:

ΔXi = Xi - Xi (Std) [for Major Components]

TMPTMP

DragDrag

8 CapillariesPprocess = 0.5 – 2 Atm

i.d. = 0.25 mmL = 1.5 m

Diaphragm Pump

35μ Leak CIS 100 amu QMS w FC

Page 32: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Ion Currents vs Time from 2 Gas Streams

How do we relate Ion Current to Partial Pressure?

Calibration

Page 33: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Molecule e-Cross Sections at http://physics.nist.gov/PhysRefData/Ionization/molTable.html Atom e-Cross Sections inferred from Wutz Handbuch Vacuumtechnik Ed 9 - 2006, Bild 12.42

Ionization Cross Sections(Å2) differ substantially with e- Energy

• Ion Gauges use 150 eV

• INFICON OIS uses 105 eV

• Most OIS and CIS’s use

70 eV for high Sensitivity

• Some OIS and CIS use 40 eV for reducing Fragmentation and Multi-Charge peaks e.g. [Ar++]

• IG Sensitivity ratios to N2 are at best a guess for RGAs

• Sensitivity Measurements are Required for Species of interest for a RGA

Page 34: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Partial Pressures are Calculated from Ion Current in Real Time using a Previous Calibration

Partial Pressure = Sens Factor * Trans(M) * [Ion Current – Interferences]

PPi = SFi * MZ * [ Ii - Σk≠i Aik * Ik ]

Page 35: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

The Result of the Partial Pressure Calculation is Real Time Display of the Partial Pressures and Clearing Times when Gas Comp Changes

4m-% H2 in Ar 4m-% H2 in Ar Air

Page 36: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

More informative is the Composition of the Gas Species showing Stable Composition after Conditioning the Inlet System

5 Min Elapsed Time Composition Table at the Time with the Yellow Cursor

4m-% H2 in Ar 4m-% H2 in Ar Air

Page 37: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

“Safe” Hydrogen [4 mol-% H2 in Ar] is a Low-Cost Reference Mixture

• Initial Calibration for H2 was done with 2.0% H2 in Synthetic Air (N2 & O2) in August 2011 • Data for 4% H2 in Ar (at right) began a month later. Data shown are values from a single scan sampled at random during the sampling of the 4 mol-% H2 in Ar from the Trend Data over 12 days • The Mean Value for H2 over 12 days is 3.88 + 0.34 Mol-%

Page 38: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Air Components Altered by Seasons and Breathing

Sampling Room Air is a Convenient Reference Gas

• Capillary Sampling and CIS Analysis of Room Air over 160 Days

• Calibration for this data done in August 2011; No Sensitivity Drift Trend

• Mean Value of N2, O2, Ar is within 1 σ

• Data displayed is a randomly selected, single measurement value of Xi

• O2 and Ar have σ ~ 2 %-RSD

• H2O and CO2 are Biased due to Seasonal ΔRH and Breathing in Lab

Page 39: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Now to Summarize…

Page 40: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

In Situ Calibration by Two Methods Provides Accurate

Partial Pressures and/or Composition to a RGA / IG

• Calibrated Flow Reference Gas Mixtures Deliver Known PPi’s and Xi’s to the Ionization Region

• Method of Introduction is Designed for a Particular Application

• Regular Measurements for PPi’s and Xi’s Plotted over Time Reveals Accuracy, Drift Trends, Time for Recalibration and Data for Sensitivity or Gain Changes

• Flow Standards should be designed to last > 1 year or Flow Rate Re-Calculated Based on Time Measured Use

XHV UHV Low Pressure Processes Atmospheric Processes

Tritium Fume Hood

Process Capillaries

Tritium Glove Box

Tritium Glove Box

Standards Capillaries

D/TPump

V

T

T

P

D/TPump

D/T/He - 1

D/T/He - 4

D/T/He - 2

D/T/He - 3

30 – 60 m

CDG

3 L

Magnetic SectorMass

Spectrometer

Process Room Analytical Lab

<< 1 mbSample

Pressure

• A Tank of Reference Gas is Sampled by the Same Means as Process Gas

• Certified Gas Mixture should be Pertinent to the Process

• Do QA Measures and Plot Difference Results

• Reference Gas Supply should last > 1 Year in planned useage

Page 41: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Summary: Operation of RGAs for Analytical Measurements

• Choose Ion Source potentials that produce linear stable operation – Electron Energy, Emission Current, Extraction/Focus Voltages and Ion Energy

• Choose a Calibration Reference Gas for Calibration Verification (QA) – For UHV/XHV: Consider a Viscous Leak with Mix Pressure defining Flow Rate.

Needs Flow Rate vs Pmix Calibration + Conductance (Calc/Measure); or Total P @ RGA

– For UHV: Fixed Flow Rate Calibration Reference Mix (INFICON or VTI) + Conductance @ OIS

– 0.01 mb<Pprocess<100 mbar: Calibration Reference Mix into CIS; Conductance of CIS

– In each application, calculate Composition in Ion Source given Viscous in – Molecular Out

– For Atm Processes: Tank with Gas Mix representing the Process; Sample with Capillary [In this case, the composition in Ion Source is the same as Tank due to Mol. In - Mol. Out]

• Calculate Sensitivities from Partial Pressures at Ion Source from Cal Gases – Initial Calibration requires known gases in the list of Species to be analyzed.

– Determine Fragmentation Factors during these initial calibrations

– If a gas species not available, estimates of the Sens can come from Gauge Constants or σ(E)

– Choosing a model for Ion Transmission, e.g. T(M) = Mz where 0 < Z < 1 to make SF ≈ σ(E)

• For High Ion Source Pressure, Xi -mol-% more stable than PPi for species

• QA: Plot differences , Xi –Xi(Std) vs Time to monitor when to Re-Calibrate

Page 42: METHODS FOR IN SITU QMS CALIBRATION FOR PARTIAL … · ERMP IND12 – QMS Workshop Bled, Slovenia ... H ig h e V : 7 0 - 1 0 0 e V H ig h E m is s io n : e .g . 2 m A M a s s R a

Acknowledgements

• INFICON RGA Engineers (Developed the Calibration Reference Source)