metric optimized gating for fetal cardiac mri · the assessment of the fetal heart. the subsequent...
TRANSCRIPT
Metric Optimized Gating for Fetal Cardiac MRI
by
Michael Jansz
A thesis submitted in conformity with the requirements for the degree of Master of Science Department of Medical Biophysics
University of Toronto
© Copyright by Michael Shelton Jansz 2010
ii
Metric Optimized Gating for Fetal Cardiac MRI
Michael Jansz
Master of Science
Department of Medical Biophysics University of Toronto
2010
Abstract
Phase-contrast magnetic resonance imaging (PC-MRI) can provide a complement to
echocardiography for the evaluation of the fetal heart. Cardiac imaging typically requires gating
with peripheral hardware; however, a gating signal is not readily available in utero. In this
thesis, I present a technique for reconstructing time-resolved fetal phase-contrast MRI in spite of
this limitation. Metric Optimized Gating (MOG) involves acquiring data without gating and
retrospectively determining the proper reconstruction by optimizing an image metric, and the
research in this thesis describes the theory, implementation, and evaluation of this technique. In
particular, results from an experiment with a pulsatile flow phantom, an adult volunteer study, in
vivo application in the fetal population, and numerical simulations are presented for validation.
MOG enables imaging with conventional PC-MRI sequences in the absence of a gating signal,
permitting flow measurements in the great vessels in utero.
iii
Acknowledgments
This work was supported by a Canadian Graduate Scholarship from the Natural Sciences and
Engineering Research Council of Canada.
I would like to begin by thanking my committee members, Mark Henkelman and Graham
Wright, for their flexibility, guidance, and confidence.
Thank you to Lars Grosse-Wortmann, Mike Seed, Derek Wong, and Shi-Joon Yoo for their
enthusiastic support of my work and their help acquiring clinical data. Mike, in particular, has
been the driving force behind this project and I am deeply indebted to him for his contributions
to this thesis, and his invaluable feedback that has influenced not only this project but also my
future career path.
Sina Fazelpour, Peter Leimbigler, and Chris Wernik for a great work environment, and Joshua
van Amerom, who is the cornerstone of our lab. While he truly has a hand in everything that
happens in our lab, he was particularly instrumental in this project.
My supervisor, Chris Macgowan, for his unwavering support and for teaching me more than he
will ever realize. I will always count myself lucky to have spent the last two years in his lab.
Finally, my family for supporting me in everything I do, and my fiancée, Nicole, for her love,
encouragement, patience—as well as all the little things.
iv
Table of Contents
Acknowledgments.......................................................................................................................... iii
Table of Contents........................................................................................................................... iv
List of Tables ................................................................................................................................ vii
List of Figures .............................................................................................................................. viii
Prologue ...........................................................................................................................................1
Chapter 1 Background .....................................................................................................................2
1.1 Motivation............................................................................................................................2
1.2 Relevant Biology .................................................................................................................3
1.2.1 Fetal Circulation.......................................................................................................3
1.2.2 Congenital Heart Disease.........................................................................................5
1.3 Fetal Cardiac Imaging..........................................................................................................7
1.3.1 Cardiac Evaluation...................................................................................................7
1.3.2 Imaging Modalities ..................................................................................................8
1.4 Cardiac Gating ...................................................................................................................10
1.4.1 The Role of Gating.................................................................................................10
1.4.2 Alternative Solutions .............................................................................................12
1.5 Fetal Heart Rate Variation .................................................................................................14
1.5.1 General Structure ...................................................................................................14
1.5.2 Indices ....................................................................................................................15
1.5.3 Modelling...............................................................................................................15
1.6 Image Metrics and Autocorrection ....................................................................................16
1.7 Thesis Statement ................................................................................................................17
v
Chapter 2 Metric Optimized Gating ..............................................................................................18
2.1 Introduction........................................................................................................................18
2.2 Theory ................................................................................................................................18
2.3 Oversampling.....................................................................................................................18
2.3.1 Metric Optimized Gating .......................................................................................19
2.3.2 HR Modeling .........................................................................................................20
2.3.3 Misgating Artifacts ................................................................................................20
2.3.4 Image Metrics ........................................................................................................22
2.4 Methods..............................................................................................................................24
2.4.1 Heart Rate Models .................................................................................................24
2.4.2 MR Data.................................................................................................................24
2.4.3 Simulation ..............................................................................................................25
2.4.4 Phantom .................................................................................................................26
2.4.5 Volunteer Experiment............................................................................................26
2.5 Results................................................................................................................................27
2.5.1 Heart Rate Models .................................................................................................27
2.5.2 Phantom Experiment..............................................................................................28
2.5.3 Volunteer Measurements .......................................................................................29
2.5.4 Fetal Measurements ...............................................................................................31
2.5.5 Simulation ..............................................................................................................33
2.6 Discussion ..........................................................................................................................35
2.6.1 Validation...............................................................................................................35
2.6.2 Fetal Application....................................................................................................35
2.6.3 Limitations .............................................................................................................35
2.7 Conclusion .........................................................................................................................36
vi
Chapter 3 Future Work ..................................................................................................................37
3.1 Introduction........................................................................................................................37
3.2 Improvements ....................................................................................................................37
3.3 Further Investigation..........................................................................................................38
3.4 Extensions ..........................................................................................................................40
3.5 Final Remarks ....................................................................................................................40
References......................................................................................................................................41
Copyright Acknowledgements.......................................................................................................47
vii
List of Tables
Table 1: Results of 1000 simulated reconstructions testing the ability of the three heart-rate
models to account for variability in the simulated heart rate traces. The simulated traces were
generated using Eq. 2, with σ = 7, Δ = 0.1, and a baseline heart rate of 150 bpm. The same 1000
simulated measurements were reconstructed using each heart rate model. The residual is the
mean reduced χ2 residual between the measured and known flow patterns. ................................ 28
Table 2: Measured mean flow values and fractional distributions corresponding to the 37 week
fetal case shown in Figure 5, as well as reference values derived from the literature (5,7,9,25-28).
Literature values are based on measurements with Doppler ultrasound and experiments involving
the injection of radionuclide-labelled microspheres. .................................................................... 33
viii
List of Figures
Figure 1: Diagram of the fetal circulation reproduced from the third edition of “Congenital
Diseases of the Heart: Clinical-Physiological Considerations” written by Abraham M. Rudolph
and published by Wiley-Blackwell (5).1 DV denotes the ductus venosus; DA denotes the ductus
arteriosus; and forked arrow in the right atrium shows the foramen ovale, which is not explicitly
labeled. ............................................................................................................................................ 5
Figure 2: Illustration of the temporal averaging that reduces the pulsatility of the flow. The
diagram on the left shows properly gated flow and the diagram on the right shows flow with
linearly accumulating CPE. On the top axes the line shows the true flow and the dots the
measured values. The power in k-space is shown on the far right and determines the relative
weighting in the averaging kernel. Each dashed line shows a frame in the series of images and
the solid line shows the frame of interest. The slope of the dashed lines is inversely related to the
difference between the true and reconstructed heart rates. ........................................................... 21
Figure 3: Results from an experiment with a pulsatile flow phantom. (a) Images of the tube
reconstructed at a range of hypothetical pump frequencies using the one-parameter heart rate
model. The frequency at which the images were reconstructed is given on the axis below, where
the axis values denote the difference between the supposed pump frequency used in the
reconstruction and the true frequency. (b) The time-entropy values corresponding to the images
in (a). (c) Flow patterns extracted from the image series indicated with the corresponding letters,
as well as the flow pattern extracted from the ECG gated images. .............................................. 29
Figure 4: Results from an experiment using the carotid arteries in adult volunteers. (a)
Comparison of the optimized heart rate model and the true heart rate trace, as measured by ECG.
The two-parameter model was used, and the two model parameters converged to nearly identical
values. (b) and (c) show a single frame from the ECG gated and MOG image series, respectively.
The magnitude images are shown on the left and the masked phase images are shown on the
right. (d) Right carotid artery flow patterns extracted from the images shown in (b) and (c). The
left carotid artery flow patterns were indistinguishable and were not included for clarity. (e)
ix
Metric values corresponding to four one-parameter heart rate searches in the left carotid artery
(LC), right carotid artery (RC), left jugular vein (LJ), and right jugular vein (RJ). ..................... 30
Figure 5: Results from a 37 week fetal case with normal cardiac anatomy and function. (a) The
metric value as a function of the model parameters for one representative measurement. (b) The
magnitude and phase images corresponding to the optimum in (a), with the fetal pulmonary
artery indicated by the arrow. (c) Three repeated measurements of the flow in the pulmonary
artery. The inset shows the expected shape of the flow pattern in the pulmonary artery in a late-
gestation fetus as measured by Doppler ultrasound...................................................................... 32
Figure 6: Results of a simulation with σ = 7, Δ = 0.1, and a baseline heart rate of 150 bpm. (a)
The simulated heart rate trace generated by Eq. 2 using the aforementioned parameters, as well
as the optimized piecewise-constant heart rate model. (b) Comparison of the reference and
calculated flow patterns. ............................................................................................................... 34
Figure 7: Results of a Monte Carlo simulation testing the effects of HRV on the quality of MOG
reconstruction. Each data point represents the mean and standard deviation of 1000 simulations
with random HRV, determined by Δ = 0.1 and σ as shown. The mean and pulsatility are
normalized against the correct values. The two-sided arrow denotes the typical range of fetal
HRV based on reports in the literature (55,76)............................................................................. 34
1
Prologue
Quantitative fetal blood flow measurements provide an important tool for the assessment of the
healthy and diseased circulation. Specifically, flow measurements in utero can improve our
understanding of fetal physiology and pathophysiology, and aid in the assessment of congenital
heart disease and pregnancy management. This thesis is concerned with the measurement of
volumetric blood flow rates (volume of blood flow through a vessel per unit time) in fetal vessels
with phase-contrast MRI (PC-MRI). It describes a new technique that allows quantitative, time-
resolved flow measurements by addressing the issue of cardiac gating, which has prevented the
application of MR to fetal cardiac assessment until now.
Chapter 1 describes the motivation, applications, and relevant background; chapter 2 addresses
the theory, implementation, and validation of the technique; and chapter 3 discusses of the
potential for further validation and improvements.
2
Chapter 1 Background
This chapter provides an overview of the background information on which the second chapter is
built. It begins by outlining the motivation for fetal flow measurements, including a discussion
of fetal circulation and its associated pathology, as well as an overview of the role of imaging in
the assessment of the fetal heart. The subsequent sections summarize a number of relevant
technical fields including the use of gating in MRI, the principles and limitations of phase-
contrast MRI, the quantitative assessment of fetal heart rate variability (HRV), and metric-based
correction of MR images. Finally, this chapter concludes by introducing metric optimized
gating, which will be discussed in detail in the second chapter.
1.1 Motivation
Congenital heart disease (CHD) is the most common congenital abnormality in North America
with an incidence of nine per thousand live births (1). It is responsible for more deaths than any
other congenital defect, and 2.3 per thousand live births will require invasive treatment or result
in death in the first year of life (1). Even cases that are successfully treated are often associated
with significant morbidity and a lifetime of medical care.
Fetal cardiac imaging is a useful tool for the assessment of CHD. In cases such as transposition
of the great arteries, pulmonary atresia, and coarctation of the aorta, it is crucial that CHD be
treated immediately after birth. Imaging enables prenatal diagnosis of CHD, allowing for
informed pregnancy counseling and improved management. Furthermore, advances in
quantitative imaging could facilitate further study of the healthy and diseased circulation during
development. This would improve our understanding of normal fetal physiology and
development, and lead to better characterization of the natural history of CHD. A reliable
measure of disease progression may also aid prognosis.
Ultrasound is the standard modality for fetal cardiac imaging. Its high spatial and temporal
resolutions, lack of ionizing radiation, and relative availability make it well suited to this
application. It is not without limitations, however, and several investigators have suggested that
MRI may provide a useful adjunct (2,3). Of particular interest is the ability to make quantitative
3
measurements of blood flow with phase-contrast MRI (PC-MRI). Unfortunately, fetal cardiac
MRI is currently limited by the inability to synchronize acquisitions to the beating of the fetal
heart (4). Without accurate synchronization, the motion of the heart and the pulsatility of the
blood in the great vessels introduce artifacts, preventing time-resolved measurements and
corrupting mean flow measurements.
The inability to properly gate acquisitions is the primary limitation of PC-MRI flow
measurements in utero, and it is this problem that is addressed in my research. My ultimate goal
is to enable time-resolved flow measurements in utero, to investigate fetal cardiac physiology
and eventually improve patient management. Here I present the theory and experimental
validation for a new technique to reconstruct PC-MRI data in the absence of a gating signal.
1.2 Relevant Biology
As the principal application of the research in this thesis is assessment of the fetal circulation and
its pathology, it is necessary to begin with a discussion of the relevant biology in order to
appreciate the motivation and potential applications of this research. It this section I will present
an overview of the fetal circulation and its pathology.
1.2.1 Fetal Circulation
As shown in Figure 1, the fetal circulation differs significantly from that of an adult. The primary
difference is that the placenta—as opposed to the lungs—is the site of gas exchange and the
source of oxygen. There are a number of adaptations that compensate for this difference. First,
umbilical circulation exists to exchange blood between the fetus and the placenta. The umbilical
arteries draw blood from the internal iliac arteries and direct it to the placenta, and the umbilical
vein and the ductus venosus stream highly oxygenated blood directly from the placenta to the
right atrium. Second, the foramen ovale connects the two atria and directs blood from the right
heart to the left, reducing the flow to the pulmonary circulation and directing the most highly
oxygenated to the brain via the left heart. Third, the ductus arteriosus connects the main
pulmonary artery to the aortic arch, bypassing the lungs. This decreases the afterload on the
right ventricle and allows distribution of blood to the body. Together, these three shunts act to
reduce the blood flow to the lungs and preferentially direct oxygenated blood to the brain. These
modifications results in a greater relative flow through the right heart, reduced flow to the lungs,
4
and considerably more mixing between the pulmonary and systemic systems, as compared to the
adult circulation.
At birth, the process of gas exchange is transferred to the lungs and several changes occur in the
circulation. First, the umbilical cord is tied off and severed leading to the closure of the ductus
venosus, umbilical arteries and umbilical vein. Second, the ventilation and oxygenation of the
lungs decreases the pulmonary vascular resistance, which increases the blood flow to the
pulmonary circulation. This increases the pressure in the left side of the heart, leading to the
closure of the foramen ovale. Finally, a decrease in prostaglandin levels leads to the closure of
the ductus arteriosus.
Although the general course of the human fetal circulation is known, many elements have not
been well characterized. Much of the current knowledge of the fetal circulation is based on
experiments in fetal sheep and postnatal measurements due to the difficulties involved in
studying the human fetus (5). The experiments in sheep determined the volumetric flow rates,
fractional distributions, and oxygen saturations using injections of radionuclide-labeled
microspheres and invasive pressure and oxygen saturation measurements (6). Subsequent
experiments with radionuclide-labeled microspheres in pre-viable human fetuses indicate that the
human fetal circulation is similar to that in the lamb, in general, but with significant differences
in flow rates, fractional distributions, and oxygen saturations (7). The difficulties associated with
non-invasive measurements of flow in the human fetus have prevented proper characterization of
these differences in the healthy human fetus (5).
5
Figure 1: Diagram of the fetal circulation reproduced from the third edition of “Congenital Diseases of the Heart: Clinical-Physiological Considerations” written by Abraham M. Rudolph and published by Wiley-Blackwell (5).1 DV denotes the ductus venosus; DA denotes the ductus arteriosus; and forked arrow in the right atrium shows the foramen ovale, which is not explicitly labeled.
1.2.2 Congenital Heart Disease
Overview
CHD refers to any cardiac defect acquired during development and can be present in utero or
acquired during the transition from prenatal to postnatal life. These abnormalities include
aberrant connections, patent shunts, hypoplasia, valvular defects, septal defects, and stenoses—or
any combination of these. Although there is a wide range of potential conditions, all types of
CHD are associated with anomalous blood flow patterns.
Some instances of CHD are simple lesions, such as minor septal defects, that cause mild
symptoms and occasionally resolve spontaneously. Some, however, are very complex and
severe, requiring major surgery and a lifetime of follow-up care. Complex lesions are usually
6
fatal without treatment (8), and it is often the closing of shunts at birth that triggers the
development of severe symptoms. Although they are not strictly malformations of the heart,
intra-uterine growth restriction and twin-twin transfusion syndrome are also developmental
abnormalities associated with altered blood flow. Intra-uterine growth restriction refers to
underdevelopment and reduced fetal size, and is often associated with insufficient placental
blood flow (9), and twin-twin transfusion syndrome is defined as the transfusion of blood
between twins sharing a single placenta.
Altered blood flow in cases of CHD may cause changes in both hemodynamics and oxygen
delivery. Before birth, these changes can affect development and may lead to fetal hydrops in
some cases. Hemodynamic perturbations are known to induce morphologic changes, and
restrictive inflow and outflow tracts can lead to hypo- and hyperplasia of the cardiac chambers,
respectively (10). Similarly, increased flow distribution to the lungs can lead to increased
pulmonary vascular resistance, and decreased pulmonary blood flow can lead to hypoplasia of
the pulmonary vasculature (11,12). Furthermore, development may also be affected by changes
in oxygen delivery. While the inter-connected nature of the fetal circulation may provide some
protection against damage to end organs resulting from obstructive lesions, these malformations
can disrupt the streaming of oxygenated blood to the brain resulting in biochemical and
morphological changes and delay in brain maturation (13). In addition to developmental
changes, CHD can cause fetal hydrops due to raised systemic venous pressures associated with
obstructive lesions and cardiac dysfunction (5). Hydrops is the collection of fluid in fetal
compartments—including the peritoneum, pleural cavity, pericardium, and skin—and is
commonly associated with fetal demise. After birth, the primary consequence of CHD is
hypoxemia, or insufficient oxygen delivery to tissues. The degree of hypoxemia varies with the
lesion type and severity, with clinical presentations ranging from circulatory collapse with or
without cyanosis in the neonatal period through to incidental detection later in childhood or
adulthood.
Treatment
The treatments of CHD vary considerably, depending on the type and severity. Mild simple
lesions are often untreated and may resolve spontaneously. Complex and severe lesions, on the
7
other hand, require prompt treatment to avoid poor outcomes. Antenatal diagnosis allows proper
planning of perinatal management, improves outcomes, and reduces the cost of neonatal care.
Complex lesions are usually palliated with open-heart surgery and less severe lesions can be
corrected with cardiac surgery or percutaneous interventions. Prior to surgery, ductal patency
may be maintained with prostaglandins to provide adequate systemic or pulmonary blood flow to
prevent hypoxia. Conversely, non-steroidal prostaglandin inhibitors may be successful in closing
a patent ductus arteriosus associated with premature delivery. Other medications such as
inotropic agents, ACE inhibitors, beta blockers, and diuretics can be used on occasion to improve
symptoms associated with cardiac failure, and anti-arrhythmic drugs may also be required in
patients with CHD. In addition to conventional therapies, in utero interventions are now under
investigation at a number of institutions. These procedures include laser ablation for the
treatment of twin-twin transfusion syndrome (14,15) as well as atrial septoplasties and aortic
valvuloplasties in cases of hypoplastic left heart syndrome with a highly restrictive atrial septum
(16,17).
1.3 Fetal Cardiac Imaging
Having discussed the fetal circulation and CHD, we now move on to investigate the role of
imaging in the assessment of the healthy and diseased fetal circulation. This section outlines the
typical methodology of assessing the fetal heart, as well as contributions of imaging to this field.
It ends with a discussion of the role PC-MRI may play in this assessment, touching on the
principles, strengths, and limitations of this technology.
1.3.1 Cardiac Evaluation
Standard pregnancy monitoring involves two ultrasound scans at around 7 and 18 weeks. The
early scan is used to confirm cardiac pulsation and measure fetal size to determine gestational
age. The later scan screens for malformations and includes a brief examination of the fetal heart.
In addition to these two scans, a thorough echocardiographic workup may be completed in high-
risk pregnancies or those suspected of having CHD.
Fetal echocardiography involves the use of several ultrasound modes to assess both the structure
and function of the heart. 2D greyscale (B-mode) imaging is used to establish the sequential
8
segmental anatomy of the heart and assess cardiac function. Rapid 1D (M-mode) imaging is
used to evaluate the motion of the chamber walls as well as the rhythm of the heart. Doppler
ultrasound can measure blood velocities through the cardiac chambers, valves, great vessels and
ductus arteriosus, providing information about the spatial and temporal patterns of the flow.
Disease is indicated by structural abnormalities, reduced or altered wall movement, or changes in
the blood flow patterns in the heart and great vessels. Changes in velocity waveforms may
indicate reduced ventricular function; the presence or absence of shunts; or abnormalities in
vascular resistance, blood pressure, vessel diameter, or valve function.
Time-resolved blood flow measurements provide a valuable tool for assessing congenital heart
disease after birth (18,19); however, flow volumes are not typically evaluated in utero due to
technical limitations. If these technical limitations can be overcome, reliable measurement of
volume blood flow may assist in the assessment of CHD by providing information regarding
ventricular function, flow redistribution, and perfusion of blood to vital tissue beds (20).
Additionally, it could provide a measure of the severity and progression of disease, aiding in
prognosis and indicating which cases may benefit from in utero intervention.
In addition to clinical management, imaging also plays an important role in the basic scientific
understanding of the fetal circulation and the pathogenesis of CHD. Flow measurements, in
particular, may be used to validate baseline flow parameters in the healthy human fetus,
characterize altered blood flow in fetuses with lesions such as hypoplastic left heart syndrome,
and evaluate the hemodynamic changes in twin-twin transfusion syndrome and intrauterine
growth restriction.
1.3.2 Imaging Modalities
Ultrasound
Ultrasound is the primary modality for fetal imaging. It is relatively cheap and accessible,
capable of providing high spatial resolution in real-time, and does not involve ionizing radiation.
Furthermore, it is also capable of hemodynamic assessment with the use of power and colour
Doppler. Ultrasound does, however, have two main limitations.
9
First, the nature of the measurements requires an acoustic window between the transducer and
the structure to be imaged, resulting in acoustic shadowing where acoustically opaque structures,
such as bone, obscure the tissue beneath them. This can prevent proper assessment in cases with
low amniotic fluid levels, adverse fetal position, or maternal obesity (21).
Second, the assessment of volume flow rates requires the measurement of both blood velocity
and the cross-sectional area of the vessel. While Doppler ultrasound can effectively assess blood
velocities, the velocity measurements are confined to the component parallel to the ultrasound
beam, making simultaneous measurement of vessel area and blood velocity impractical.
Ultrasound flow measurements typically rely on assumptions regarding the shape of the vessel,
the insonation angle, and the velocity profile (22). These assumptions, along with the
challenging nature of vessel diameter measurement, introduce large uncertainties in the final
flow rates (23,24). While there have been many reports of flow volume measurements in fetal
vessels with Doppler ultrasound in the literature (9,25-28), there is significant disagreement
between studies. Furthermore, validation experiments have also demonstrated large inter- and
intra-observer variability and poor repeatability (23,29). As a result, ultrasound has been used
primarily to assess blood velocities.
MRI
MR may serve two roles in fetal cardiac imaging. First, it can offer an alternative in cases where
ultrasound is not feasible due to acoustic shadowing. Second, PC-MRI provides a well-
established technique for measuring blood flow (30), which may provide additional
hemodynamic information beyond that which is provided by Doppler assessment of blood
velocity. The research in my thesis focuses on the technical challenges associated with the
application of PC-MRI for time-resolved flow measurement in the fetus.
Phase-contrast MRI
MR data has both a magnitude and a phase; however, conventional MR imaging considers only
the signal magnitude. Phase-contrast imaging, on the other hand, uses the signal phase in each
voxel to encode the component of the local tissue velocity in an arbitrary, user-specified
direction. In practice, eddy currents, field inhomogeneities, and gradient imperfections also
produce phase shifts; however, these can be removed by taking the difference between two
10
interleaved acquisitions with positive and negative velocity encodings. If the imaging plane is
prescribed orthogonal to the vessel and through-plane velocity encoding is used, a reliable
measure of the volume flow rate can be determined by integrating the velocity over the vessel
lumen (30). If a series of images are acquired at different points in the cardiac cycle, the time-
evolution of the flow rate can also be quantified.
There are several limitations of PC-MRI, especially when applied in the fetal population. First,
low spatial resolution can lead to partial volume effects where the phases in voxels at the vessel
boundary do not reflect the true mean velocity of the protons in those voxels. This problem will
always be present to some degree, but it can be adequately reduced with sufficiently high spatial
resolution (31,32). Second, the need for interleaved positive and negative flow encodes and
short echo times makes accelerating phase-contrast scans difficult, resulting in long scan times.
This can be overcome by acquiring the data in segments, where several lines of k-space are
acquired in each R-R interval in an interleaved fashion. This decrease the scan time by an
integer factor given by the number of lines-per-segment, although it also reduces the temporal
resolution by the same factor. Third, there is an unavoidable trade-off between spatial resolution
and scan time. In the case of fetal imaging high spatial resolution is important to reduce partial
volume effects; however, long scan times increase the likelihood of fetal movement during the
scan. Fourth, accurate flow measurements require orthogonal slice prescriptions which are more
difficult in fetal subjects due to the possibility of bulk motion between the acquisition of
localizers and the prescription of the phase-contrast imaging plane. Finally, there is the issue of
cardiac gating, which will be discussed below.
1.4 Cardiac Gating
The absence of cardiac gating is the primary limitation of the assessment of the fetal heart with
PC-MRI. This section describes the role of cardiac gating in cine PC-MRI, and further explains
the consequences of not having a gating signal. It concludes with an overview of the literature,
discussing potential solutions to this problem that have already been proposed.
1.4.1 The Role of Gating
In PC-MRI, the time required to acquire a single image is comparable to the length of a cardiac
cycle. This poses a problem, as both the position of the heart and the velocity of the blood in the
11
great vessels will vary considerably during the time each image is acquired, and k-space will not
reflect a single point in the cardiac cycle. This results in image artifacts such as blurring and
ghosting of moving structures and temporal smoothing of the flow waveform.
To overcome this problem, the data acquisition must be divided over several heartbeats with each
segment acquired in a different cardiac cycle. The acquisition is then gated such that every echo
is acquired within an appropriately small window centred on the desired cardiac phase. In this
case, data can also be acquired at other cardiac phases at no cost to scan time, to enable
reconstruction of a time-series of images reflecting a number of different points in the cardiac
cycle. This is referred to as cine imaging.
There are two strategies for cardiac gating. The first is prospective gating where the scanner
waits for the beginning of each cardiac cycle and then acquires a predetermined number of
measurements with specific delays before stopping to wait for the beginning of the next cycle.
These measurements are then inserted into an equivalent number of k-space matrices that are
reconstructed to produce a series of images corresponding to precise points in the cardiac cycle,
with the disadvantage that imaging at late diastole is difficult. The second is retrospective gating
where the data are acquired continuously and the segment number is incremented with each
cardiac cycle. The ECG signal is recorded simultaneously and the cardiac phase of each
measurement is determined retrospectively, relative to the start and end of the cardiac cycle in
which it was acquired. The data are then interpolated to a set of desired cardiac phases. This
produces a series of images with slightly reduced temporal resolution due to the interpolation,
but with full coverage of the cardiac cycle. A modified retrospective gating approach is used in
this work.
Needless to say, cardiac gating requires the measurement of a signal that indicates the beginning
of each cardiac cycle. In adults, this is accomplished with either ECG electrodes attached to the
patient’s chest or a peripheral pulse monitor attached to an extremity; however, these signals are
not readily available in utero—precluding the use of gating for fetal cardiac PC-MRI. This
results in an appreciable loss of image quality, loss of the dynamic information in flow
measurements, and corruption of mean flow values. The next section discusses potential
solutions to this problem.
12
1.4.2 Alternative Solutions
In the absence of a traditional ECG or peripheral pulse monitor signal, there is a range of
potential solutions for cardiac gating for MRI. These include the detection of an alternative
gating signal with additional hardware, and the use of real-time imaging, self-gating, or
nontriggered acquisitions.
Hardware
There are a number of hardware solutions capable of measuring the fetal heart rate, although
their applicability to human fetal cardiac MR gating has yet to be determined. These include
fetal electrocardiography (fECG), magnetocardiography (MCG), and both imaging and non-
imaging ultrasound.
fECG involves the acquisition of a composite fetal-maternal ECG signal with electrodes attached
to the mother’s abdomen, and the extraction and amplification of the fetal component with
advanced signal processing. Unfortunately, the fetal signal is typically an order of magnitude
weaker than the maternal signal, and the extraction is only possible with a sufficient SNR (33).
In an MR magnet, the excess noise due to gradient switching and magnetohemodynamic effects
is likely to make fECG extraction unfeasible.
MCG involves the use of a superconducting quantum interference device (SQUID) in a
magnetically shielded environment to measure minute magnetic fields associated with the
beating of the fetal heart. Needless to say, this technology is also not applicable for MR gating.
The last, and most promising, hardware for measuring the fetal heart rate is ultrasound.
Ultrasound is an established technique for measuring the fetal heart rate, and MR-compatible
ultrasound devices have been reported in the literature (34). The most common technology for
measuring the fetal heart rate is cardiotocography (CTG), which operates by identifying the
dominant frequency in a signal acquired with a non-imaging ultrasound transducer directed at the
fetal heart. Typically, CTG returns a time-averaged measure of the fetal heart rate at regular
intervals, which is insufficient for MR gating; however, it may be possible to extract a more
useful gating signal with improved signal processing. Alternatively, feature tracking with
conventional B-mode ultrasound has been applied to respiratory gating in adults and it may be
13
possible to extend this approach to fetal cardiac gating. These techniques both show potential for
fetal cardiac gating; however, the extension is non-trivial, and there have not been any reports of
fetal cardiac gating with ultrasound to date.
Real-time MRI
Real-time imaging does not require a gating signal because the acquisition of each image is
sufficiently fast to eliminate motion artifact. Significant advances have been made in real-time
PC-MRI; however, these sequences necessarily compromise spatial and temporal resolution as
compared to a regular ECG-gated acquisition. In fetal cardiac imaging the spatial and temporal
resolution are both at a premium, as the vessels of interest are on the order of 5 mm or less, the
cardiac cycle length (R-R interval) is shorter than 500 ms, and the field of view is necessarily
large to encompass the mother’s abdomen and avoid the image artifacts known as “wrap”.
Current state-of-the-art “real-time” methods provide phase-contrast data with 20x20 cm2
coverage, 1.5 mm in-plane resolution, and 150 ms temporal resolution, which are still
insufficient for this application (35).
Self-gating
Self-gating or “wireless” strategies were first proposed by Spraggins (36) and Hinks (37). These
techniques sample k-space such that a periodic gating signal can be extracted from the MR data
themselves, which can then be used to sort the data retrospectively. Modern implementations
exist for both radial and Cartesian trajectories and use a variety of signal detection algorithms
including signal intensity modulation, centre-of-mass tracking, image correlation, and low-
resolution flow measurement (38-40). Previous publications have suggested fetal imaging as a
potential application for their techniques and Nieman (41) and Holmes (42) have even
implemented cardiac self-gating for the mouse and chick fetus, respectively. These techniques
may be applied to human fetal cardiac PC-MRI, but no such application has been reported in the
literature to date. Additionally, these techniques often require specialized pulse sequences in
order to extract a suitable self-gating signal.
Nontriggered imaging
A nontriggered acquisition provides a simple alternative to complex gating schemes by
circumventing the need for gating altogether. In this case, an image is acquired with complete
14
disregard for the cardiac phase, generally with several averages. In theory, the acquisition will
sample evenly across a number of cardiac cycles producing a flow value in each pixel that
reflects the mean flow during the scan. This process is complicated by factors such as oscillation
of the vessel size and position, and varying magnitude in the vessel due to inflow effects. It has
been shown that the pulsatility of the flow as well as the choice of scan parameters can greatly
affect the degree of inflow enhancement and consequently the measured mean flow (43,44).
Sequences that minimize inflow effects may reduce this problem; however, problems with vessel
motion and dilation will remain. Additionally, these methods disregard the dynamic information,
producing only a mean flow measurement.
1.5 Fetal Heart Rate Variation
In cine PC-MRI imaging, the objective is to produce a series of images where the dynamic
information in the cardiac cycle is resolved in a series of time-frames. It is therefore important
that each image corresponds to data from a single phase of the cardiac cycle, which is typically
accomplished with gating.
In the absence of any heart rate variability, gating would be unnecessary as the period of the
acquisition could simply be matched to the R-R interval length. Unfortunately, this is not usually
the case and a more sophisticated acquisition and reconstruction is required. While the case with
no gating or heart rate variability is clearly extreme, it highlights the fact that the level of heart
rate variability, the complexity of the reconstruction method, and the quality of the
reconstruction are all intimately related. In light of this, it is important to understand the nature
of fetal rate heart variability as it will constrain and motivate possible solutions, and ultimately
determine the quality of the reconstruction.
1.5.1 General Structure
Fetal HRV is used as a clinical measure of fetal well-being both ante- and intrapartum (45),
which has motivated an appreciable amount of research in the area. Heart rate traces are visually
inspected by the obstetrician, and guidelines for this evaluation are well-established (45). A
healthy human fetus usually displays moderate variation, asymmetrically distributed about a
baseline value which is between 110-160 bpm (45). Healthy variation includes plateau-like
accelerations, random variation, and some oscillatory components related to fetal proto-breathing
15
(46-48). In this work, random motion refers to the heart rate variation that is not obviously
structured, although it is typically fractal in nature (49). Heart rate variation also depends on
fetal behavioural state with more pronounced variation in the active awake and active sleep states
(50).
HRV can be subdivided into short- and long-term variation, where short-term variation refers to
beat-to-beat changes and high frequency changes, and long-term variation refers to the power in
the low frequency components and the breadth of the distribution of achieved heart rates.
Compared with adults, fetal HRV shows a smaller dynamic range and less beat-to-beat variation
(45,51).
1.5.2 Indices
Due to the subjective nature of fetal HRV interpretation, many researchers have attempted to
quantify and automate the process through the use of computer-calculated indices. More than 20
different indices have been proposed over the last 40 years, with varying degrees of utility (52).
Very few of these have entered standard clinical practice; however, these indices allow the
quantification and characterization of fetal HRV, permitting validation of simulations against the
literature. An exhaustive description of the definitions and normal values of these indices is
beyond the scope of this thesis.
1.5.3 Modelling
There have been a number of attempts to simulate fetal heart rate traces and their corresponding
signals as measured by different devices. These simulations were created using a wide range of
methodologies and varying degrees of realism.
Trace Generation
Several authors have attempted to generate simulated fetal heart rate traces based on existing
knowledge of the characteristics and structure of fetal HRV, generally with poor results.
Cesarelli et al. (53) generated synthetic traces for signal processing validation by combining an
existing model for simulating adult ECG signals with knowledge about the spectral properties of
fetal HRV from the literature (46). van Meurs et al. (54) produced a simulation that mimics fetal
distress by generating realistic decelerations, which were used for educational purposes.
16
Magalhães et al. (55) adapted this technique and added white noise, which they then applied to
signal-processing validation. In all of these cases the resulting traces were qualitatively and
quantitatively unrepresentative of real fetal HRV. While these simulations may be appropriate
for validation of signal-processing algorithms, their unrealistic nature makes them inappropriate
for validation in this work.
Parameter Extraction
In addition to the generation of simulated fetal heart rate traces based on existing parameters,
several groups have attempted to analyze real heart rate traces within the confines of different
models.
Jarisch and Detwiler (56) attempted to overcome the inherently random nature of fetal HRV
through the use of a Kalman filter to fit a parameterized stochastic model to real datasets. Their
model included white noise for beat-to-beat variation, asymmetric variation around a slowly
varying baseline, and a “jitter” component that had no physiologic motivation but was
determined empirically to be necessary. They reported good fit quality in general, but
unfortunately, their work did not include parameter value ranges derived from their training data,
precluding the use of this model for the generation of simulated traces in this work.
Gough (57) modeled the fetal heart rate as Brownian motion and determined the fractal
dimension through repeated measurements of the curve length with varying ruler sizes. This
model assumes nothing about the underlying structure of the fetal HRV, other than the existence
of fractal properties, which had already been demonstrated (49). This model has since been
adapted by a number of researchers (58-60), and motivated the simulation used in this work.
1.6 Image Metrics and Autocorrection
The final section of this chapter discusses an existing image correction technique of particular
relevance to the work. As will become apparent, my research draws upon the existing theory and
methodology of metric-based autocorrection. Metric-based autocorrection has been applied in a
number of fields, including astronomy, photography, and most recently medical imaging.
Although the undesired artifact, cost function, and correction strategy vary between applications,
17
at the most basic level all implementations use an image metric to quantify image quality, and
then iteratively adjust the image to optimize the metric value.
Metric-based autocorrection was first applied to MR imaging by Noll et al. in 1992 (61), where a
maximum-intensity focus criterion was used to reduce blurring caused by field inhomogeneities.
It was subsequently applied to ghost artifact suppression by Xiang and Henkleman (62) with an
image gradient power criterion. The most relevant implementation of metric-based
autocorrection was first introduced by Atkinson et al. in 1997 (63) and further developed by the
Ehman group (64). This implementation determines a “motion record” that describes the
position and orientation of the patient during the scan, and applies phase shifts and rotations in k-
space to reduce artifacts resulting from in-plane motion during the scan.
These authors used a number of different focus criteria including image entropy (65) and the
entropy of the gradient image (66,67), among others (68). The technique has been applied to a
number of different image types including brain (65) and shoulder imaging (65,66) as well as
angiography (67). In all cases the metric was evaluated on the anatomical magnitude images.
1.7 Thesis Statement
The objective of this work is to develop a retrospective data-processing technique that permits
time-resolved flow measurements in utero without the need for gating. I propose a new gating
strategy called metric optimized gating (MOG), that is based on previous work in metric-based
autocorrection that is motivated by knowledge of fetal heart rate variation. In the next chapter I
discuss the theoretical basis and practical implementation of this technique, and provide
experimental validation. This technique enables time-resolved PC-MRI in utero, providing a
useful tool for the assessment of the fetal heart. The hope is that this will lead to improved
clinical management for cases of complex CHD and a better understanding of the fetal
circulation.
18
Chapter 2 Metric Optimized Gating*
2.1 Introduction
The previous chapter outlined the potential utility of cine PC-MRI for evaluating the fetal heart,
as well as the issue of gating that has precluded these types of measurements. Without the ability
to gate acquisitions, cardiac motion produces appreciable artifacts that corrupt both time-
resolved and time-average flow measurements.
This chapter draws on several concepts presented in the last chapter including retrospective
cardiac gating, fetal heart rate variability, and metric-based image correction. These concepts
motivate an alternative strategy for reconstructing fetal PC-MRI data acquired without explicit
gating, where the gating is retrospectively determined by adjusting a heart rate model to optimize
an image metric. In this chapter I present the theory and implementation of this technique, as
well as the results from a number of validation experiments. These include the construction of a
numerical simulation, an experiment with a pulsatile flow phantom, a study in adult volunteers,
and finally preliminary application in the target fetal population.
2.2 Theory
The technique proposed in this study involves the use of image metrics to detect misgating
artifact. Specifically, the image metrics are evaluated on the time-series of PC-MRI images of
vessels with pulsatile flow. For this reason, it is important to understand how the data are sorted
according to cardiac phase, what artifacts result from incorrect sorting, and how these artifacts
can be detected.
2.3 Oversampling
As described in the previous chapter, retrospective cardiac gating involves the continuous
acquisition of data with each block of consecutive k-space lines (referred to as a segment) being
* This chapter is based on a manuscript that has been accepted for publication in the journal Magnetic Resonance in
Medicine (69).2
19
acquired for an entire R-R interval. In addition to recording the times of the R-waves during the
scan for later use, the heart rate monitor triggers the transition from the acquisition of one
segment to the next. In this sense, the technique is not strictly retrospective; however, the term
“retrospective” is used to reflect the fact that the data are subsequently interpolated to a
predetermined set of cardiac phases. Larson et al. (39) implemented a more literal interpretation
of retrospective gating where each segment of k-space was acquired repeatedly for a
predetermined period that was longer than the longest anticipated R-R interval. These data were
then temporally interpolated in the traditional way. This technique involves a slight loss in scan
efficiency but improves the flexibility in accommodating heart rate variability (HRV) and
eliminates the requirement for any real-time feedback from a heart rate monitor.
To allow retrospective reconstruction, each segment must be sampled at every cardiac phase.
For fetal imaging, this requirement is complicated by the fact that the heart rate is unknown and a
heart monitor is unavailable to trigger the transition from one segment to the next. The technique
implemented by Larson et al. can be used to circumvent this problem if the period of acquisition
is chosen to be sufficiently long to accommodate the full range of heart rates found in the fetal
population. This can reduce the scan efficiency by up to 30 % in some cases, but it also
guarantees that a complete reconstruction is possible. Additionally, the excess data can be used
to improve the SNR in the final images. As the fetal heart rate usually ranges between 110-
180 bpm throughout pregnancy, oversampling of the cardiac cycle can be accomplished for the
entire fetal population with a sampling period greater than or equal to 545 ms (the reciprocal of
110 bpm) (45).
2.3.1 Metric Optimized Gating
The oversampling process described above ensures that every segment of k-space is acquired at
every cardiac phase. This guarantees the existence of perfect reconstruction, provided the data
can be properly sorted according to cardiac phase retrospectively. The sorting by cardiac phase
is accomplished by distributing hypothetical triggers throughout the scan according to a
parameterized model of the fetal heart rate. The data are grouped according to cardiac phase and
temporally interpolated to produce a series of images and an image metric is evaluated on the
reconstructed images to determine the level of misgating artifact. This process is repeated and
the heart rate model parameters are iteratively adjusted to optimize the image metric.
20
2.3.2 HR Modeling
The first chapter discussed the fact that fetal HRV is fractal and random in nature over short
periods of time (51,57), and displays a much smaller dynamic range (45) and less short-term
variation (51) as compared to adult HRV. These characteristics suggest that it may be possible to
approximate the fetal heart rate well with a fairly simple model, as the variation is relatively
small and unstructured. In addition, the fact that the velocity-induced phase shifts in PC-MRI are
primarily encoded in the central portion of k-space (70) suggests that the heart rate model need
only be accurate over the short period of time during which the centre of k-space is acquired.
For linear phase-encode orderings, as considered in this study, this is the middle of the scan.
An ideal heart rate model provides sufficient flexibility to account for fetal HRV, reasonable
search times, and adequate SNR in the image metric to eliminate over-fitting. On one extreme, a
one-parameter model assumes the heart rate is constant and provides the least flexibility with the
fastest search times. On the other extreme, a many-parameter model, where each trigger time is
specified by an independent parameter, can completely account for the HRV, at the cost of a long
search time. Between these extremes, there exists a wide range of potentially applicable models.
2.3.3 Misgating Artifacts
Misgating refers to the incorrect detection of one or more triggers that mark the beginning of
each cardiac cycle during the scan. This results in a cardiac phase error (CPE) between the true
cardiac phase at which an echo was acquired and the presumed cardiac phase based on the
incorrect gating. In metric optimized gating (MOG), the timing of each trigger is determined by
“dead reckoning” according to the parameterized model of the fetal heart rate during the scan.
This makes CPE cumulative from one R-R interval to the next, as it is given by the integral of
the difference between the true and modelled heart rates. In general, misgating can be either
systematic or random. Systematic misgating refers to a structural discrepancy between the
modelled and true heart rates, causing the modelled and true trigger times to drift apart and the
CPE to grow from one interval to the next. Random misgating refers to short-term variability in
the heart rate that is not properly accounted for in the model, causing the CPE to wander
randomly about zero throughout the scan.
21
Systematic misgating produces two principal effects in vascular PC-MRI images: ghosting in the
phase-encoding direction and a loss of pulsatility in the vessels, where pulsatility is defined as
, and Q is the volume flow rate. The ghosting in the phase-encoding
direction is caused by the distribution of the non-zero temporal frequency components in the
vessel. This is a well-known artifact that was clearly explained by Xiang (71) using a k-t-space
formalism. The locations, powers, and phases of these ghosts can be determined analytically in
very simple cases and numerically in more realistic cases. The loss of pulsatility in the extracted
flow patterns is due to mixing of data from different parts of the cardiac cycle into each image.
As shown in Figure 2, reconstruction of data at an incorrect heart rate produces an oblique
trajectory in k-t-space due to an accumulation of CPE with time, or equivalently ky. In this case,
each cardiac phase represents a weighted average of the entire cardiac cycle, where the weighting
is determined by the power distribution in k-space.
The artifacts resulting from random misgating are more difficult to characterize. This type of
CPE produces dispersion in the phase-encoding direction that is smeared rather than discrete,
with pulsatility less affected.
Figure 2: Illustration of the temporal averaging that reduces the pulsatility of the flow. The diagram on the left shows properly gated flow and the diagram on the right shows flow with linearly accumulating CPE. On the top axes the line shows the true flow and the dots the measured values. The power in k-space is shown on the far right and determines the relative weighting in the averaging kernel. Each dashed line shows a frame in the series of images and the solid line shows the frame of interest. The slope of the dashed lines is inversely related to the difference between the true and reconstructed heart rates.
22
2.3.4 Image Metrics
In MOG, the “correctly gated” series of images is defined as the one that optimizes the image
metric value, making the selection of an appropriate metric essential. There are several criteria
that define an effective image metric: the metric should be able to discriminate between good and
bad reconstructions (sensitivity); the definition of a “good” reconstruction should agree with
reality (accuracy); the difference in the metric value between correct and incorrect
reconstructions should be large compared to random error in the metric value (high contrast-to-
noise); the metric should operate well on images with low SNR and be insensitive to user-
controlled parameters such as ROI placement (robustness); and it should have a single, well-
defined optimum (well-behaved). Additionally, the metric should be straightforward to compute
in a short period of time.
The characteristic effects of misgating provide several means by which misgated reconstructions
might be identified. Spatially, the smearing of the vessels in the phase-encoding direction makes
the power distribution more diffuse and the vessel boundaries less abrupt. Temporally, the loss
of pulsatility results in a more even distribution of flow across the cardiac cycle and a reduction
in the power of the derivative of the flow waveform. Previous work in metric-based auto-
correction (63,68,71) provided several potential metrics that have been shown to be sensitive to
similar effects in other applications. The metrics investigated in this study included the image
gradient power (with a variety of derivative kernels), the image entropy, the entropy of the
gradient, and the normalized gradient squared. These metrics were implemented spatially,
temporally, and spatio-temporally. In addition to these metrics, the pulsatility index and mean
value of the flow waveform extracted from the vessel lumen were tested for their ability to detect
misgating.
Of the proposed metrics, time-entropy was best able to satisfy the above criteria and hence was
selected for use in this work. Time-entropy refers to the entropy evaluated in time on a voxel-by-
voxel basis and summed spatially, which provides a surrogate measure of pulsatility. This is
because pulsatility is equivalent to non-uniform flow distribution in time, which has a lower
likelihood of random occurrence. It relies on the assumption that the most correct reconstruction
produces the most pulsatile waveform. Apart from small effects where noise can add
constructively to slightly increase the pulsatility near the correct reconstruction, it is true that the
23
pulsatility only decreases with misgating. This provides a type of template-matching, where the
reconstruction is guided by an assumption regarding the measured flow; however, it requires less
a priori knowledge about the anatomy and physiology of the vessel than a true template-
matching approach. Additionally, the fact that time-entropy considers the data as a set instead of
a sequence makes it less susceptible to noise and less biased by the specific shape of the flow
curve than derivative-based metrics. Finally, time-entropy is insensitive to ROI placement and it
does not require vessel boundary delineation prior to the optimization, which is important
because severe misgating can obscure and distort images to the point where even locating vessels
can be challenging.
The time-entropy was evaluated on the phase images of the PC-MRI series, as they contain most
of the dynamic information. It was necessary to mask out regions with large phase error (i.e. low
signal) as they impair the performance of the metric. This was accomplished by multiplying the
phase images by the corresponding magnitude images. Although the magnitude images contain
strictly non-negative values, the phase images can contain any value between , making it
necessary to rectify the masked images before evaluating the metric. Evaluation of the metric on
the entire field of view introduces background signal and noise, which decreases the sensitivity
and contrast of the metric. This was solved by evaluating the metric on a small neighborhood
surrounding the vessel of interest. An 11×11 voxel region was chosen as a compromise between
robustness and sensitivity. With the resolution used in this study and the typical size of fetal
vessels, 11 voxels corresponded to approximately 2-3 vessel diameters, which was sufficiently
large to allow reliable placement around a vessel and its accompanying artifact. To compensate
for the fact that this neighborhood contains a large proportion of extra-vascular voxels, the
spatial sum of the entropy values was weighted by the total signal in each voxel. This increases
the sensitivity of the metric to changes in pulsatility in the vessel by enhancing the contribution
from the voxels in the vessel to the spatial sum.
The time-entropy was given by the expression:
€
E =Bi
Bii∑i
∑ Si,tBit
∑ logSi,tBi
⎛
⎝ ⎜
⎞
⎠ ⎟ (1)
24
where i indexes over space, t over time, S is the product of the phase and magnitude images,
and
€
Bi = Si,t2
t∑[ ]
1/ 2. The normalization factor used in this work was adapted from the work of
Atkinson et al. (63) and was found to provide less contrast but more accuracy than the usual
entropy normalization of
€
Bi = Si,t2
t∑ .
2.4 Methods
2.4.1 Heart Rate Models
In this study, a matter of minutes was considered an acceptable search time, which restricted the
heart rate models to one- and two-parameter searches based on available computer hardware.
The one-parameter model assumed a constant heart rate throughout the scan with the single
parameter specifying this constant, unknown heart rate. In addition, two two-parameter models
were considered—each motivated by different assumptions and a priori information. First, the
knowledge that the fetal heart rate typically contains less short-term variation than that of an
adult motivated a linear model where the heart rate is specified by a constant baseline and a
“drift” term (i.e. slope). Second, the knowledge that the majority of the dynamic information is
contained in the centre of k-space motivated a model where the two degrees of freedom were
chosen to maximize the flexibility during the middle of the scan when these data are acquired.
This model is piecewise-constant with a discontinuity in the middle of the scan, and the heart rate
in each half of the scan is specified by an independent, constant value.
2.4.2 MR Data
All clinical measurements were acquired with informed written consent and approval from the
hospital research ethics board, and were consistent with the guidelines for fetal MR imaging at
our institution. Patients were directed to MR if the prior fetal echocardiography was difficult or
inconclusive, as in patients with oligohydramnios or severe obesity. Patients who were referred
for fetal MR for extracardiac reasons were also studied with consent and hospital IRB approval.
The data were acquired on a 1.5T Avanto Syngo system (SIEMENS, Erlangen, Germany), using
an abdominal multi-channel surface coil. A conventional gradient echo phase-contrast sequence
was used with flip angle = 30°, TE/TR = 2.9/6.6 ms, and VENC = 150 cm/s. The positive and
25
negative velocity encodes were acquired alternately, doubling the time for each measurement to
13.15 ms. A voxel size of 1.25×1.25×5 mm3 was required due to the small size of fetal
structures, and a relatively large field of view of 32×48 cm2 was used to prevent excessive
wrapping of the maternal abdomen. To simplify off-line reconstruction, parallel imaging was not
used. Instead, every 4 lines of k-space were segmented to accelerate the scan, giving an effective
temporal resolution of 52.6 ms. Ten repetitions were acquired regardless of the fetal heart rate,
resulting in the continuous acquisition of each segment for 526 ms. Although this is slightly
shorter than the upper bound provided in the literature (545 ms), estimates of fetal heart rates
with cardiotocography prior to each MR examination suggest that it was sufficiently long in all
cases. In addition, the number of repetitions can be adjusted to accommodate lower heart rates,
if necessary. The scan time was 34 s, requiring that the acquisition be free-breathing. Data were
retrospectively reconstructed offline (MATLAB, The Mathworks Inc, Natick, MA, USA) and
reconstructed images were analyzed with clinical flow quantification software (QFlow®, Medis,
Leiden, Netherlands). Multicoil data were combined using quadrature magnitude weighting (72)
and phase images were corrected for background phase offsets (73). Linear interpolation was
used to generate intermediate cardiac phases through view-sharing (74,75), resulting in 30
cardiac phases.
2.4.3 Simulation
A phase-contrast acquisition of pulsatile flow through a vessel was numerically simulated to
study the effects of HRV on the reconstruction algorithm. The simulated matrix size, spatial and
temporal resolutions, and view ordering were all equal to their respective clinical parameters and
the vessel diameter, signal-to-noise ratio (SNR), and parabolic flow profile were based on direct
fetal measurements. An arterial flow pattern was simulated as
where , , and ,
and T is the cardiac period. The final calculated flow pattern was compared to the known input
flow to determine the quality of the reconstruction. The start of the cardiac cycle cannot be
determined by this technique, so the temporal offset between the two curves was fit as a free
parameter and agreement was quantified by the reduced χ2 residual.
A fetal heart rate simulation was constructed to test the effects of heart rate variability on the
reconstruction process. The observation that fetal heart rate traces are well described by
26
Brownian motion (57) motivated the simulation used in this work. The heart rate was simulated
as a bounded random walk superimposed on a constant baseline:
€
RRn+1 = RRn +N Δ⋅ RR0 − RRn( ) , σ2[ ] (2)
where N is a normally distributed random variable, RR0 is the baseline heart rate, σ is the
standard deviation of the beat-to-beat step size in milliseconds, and
€
Δ ∈ 0,1[ ] is the strength of
the bias bounding the walk around the baseline value. Values of Δ = 0.1 and σ = 7 ms were found
to give HRV that was quantitatively representative of typical fetal HRV, as measured by a
number of HRV indices (60,76). The reconstruction process was tested over a range of HRV
values, with 1000 random simulated heart rate traces at each value.
2.4.4 Phantom
A phantom experiment was completed to test the effects of misgating and to validate the heart
rate search algorithm and image metric. The phantom consisted of a computer-controlled
servomotor (FlexDriveII, Baldor Motors and Drives, Fort Smith, AR, USA) and gear pump
(Shertech AK Series AMBV1A, Hypro Corporation, New Brighton, MN, USA) connected to a
tube with 10 mm inner-diameter. The programmable nature of the pump allowed for the
combination of physiologically realistic flow patterns with the convenience of a relatively
constant, known pump frequency. The setup also permitted electrical triggering of the MR
scanner to the pump cycle to provide gated PC-MRI measurements for reference. The pump was
cycled at the maximum achievable frequency of 80 bpm. The diameter was larger than that of
typical fetal vessels; however, the spatial resolution was reduced to 2x2 mm to compensate. This
resulted in an equivalent number of voxels per vessel, albeit with a higher SNR than would be
expected in a fetal measurement.
2.4.5 Volunteer Experiment
Due to the difficulty associated with validating fetal PC-MRI flow measurements and the
absence of heart rate variability in the in vitro model, a validation experiment was carried out
using the carotid arteries in adult volunteers. This provided a good model of the fetal great
arteries with a more straightforward validation. The carotid arteries were chosen because they
provide similar vessel diameters and flow patterns to what would be expected in the great vessels
27
in a fetus. Exercising on an MR compatible cycling apparatus achieved heart rates comparable
to that of a fetus (120-150 bpm), as well as an appreciable degree of HRV. The carotid arteries
were imaged at the same resolution as the fetal vessels, providing approximately the same
number of voxels per diameter and a similar SNR.
There are several factors that make adult carotid arteries well suited for the validation of flow
curves. First, orthogonal slice prescriptions through the neck are easy to achieve, eliminating
effects due to poor slice prescriptions. Second, there are two carotid arteries in each image, as
well as two vertebral arteries and two jugular veins. This provides a wealth of data for validation
as well as the possibility of internal comparisons in the data. Third, adult volunteers are less
prone to random movement than fetal subjects and the carotid arteries do not move appreciably
with respiration, eliminating the confounding effects of motion. Finally, and most importantly,
measurements in adults made it possible to acquire an ECG signal in synchrony with the PC-
MRI measurements. Each data set could then be reconstructed with ECG gating and MOG,
allowing the comparison of the measured and modeled heart rate traces as well as the ECG gated
and MOG flow measurements.
2.5 Results
2.5.1 Heart Rate Models
Search times varied depending on the number of coils activated during the scan, the number of
function evaluations required by the search function, the number of parameters used in the
model, and the computer hardware; however, in general, a one-parameter search could be
completed in approximately 1 minute, and a two-parameter search could be completed in 4-5
minutes on a conventional desktop computer with non-optimized coding. The reconstruction
quality, as measured by the flow-fit residual, for the different heart rate models averaged over
1000 simulation runs are shown in Table 1. It is apparent that the addition of the second
parameter provides a significant improvement over the constant heart rate model and that the
piecewise-constant model provides better results than the linear model. For this reason, the
piecewise-constant model was used for optimization of the volunteer and fetal data. Due to the
relative stability of the pump frequency in the phantom, the one-parameter model was sufficient
to provide an accurate reconstruction.
28
Table 1: Results of 1000 simulated reconstructions testing the ability of the three heart-rate models to account for variability in the simulated heart rate traces. The simulated traces were generated using Eq. 2, with σ = 7, Δ = 0.1, and a baseline heart rate of 150 bpm. The same 1000 simulated measurements were reconstructed using each heart rate model. The residual is the mean reduced χ2 residual between the measured and known flow patterns.
Model Residual (×103) Constant 12.00 ± 0.04 Linear 7.24 ± 0.02
Piecewise-constant 6.65 ± 0.02
2.5.2 Phantom Experiment
Results from the flow phantom experiment are shown in Figure 3. These results demonstrate the
characteristic effects of misgating as well as the sensitivity of the metric to these effects. Figure
3a and Figure 3c depict images and flow patterns corresponding to reconstructions at a range of
assumed pump frequencies. The effects of misgating are apparent in both the images and the
flow patterns. The misgated images contain severe smearing in the phase-encode direction and
the flow patterns have reduced pulsatility, even when the reconstruction frequency was incorrect
by only a few cycles per minute. On the other hand, the optimized images are sharp and artifact
free, and the flow extracted from the optimized images is almost indistinguishable from the gated
reference flow. The time-entropy values plotted in Figure 3b show excellent discrimination
between correct and incorrect reconstructions, with a pronounced minimum at the correct
frequency.
29
0102030405060708090100
0 0.2 0.4 0.6 0.8 1
Flow
(ml/
s)
Cardiac Phase
GatedABCDE
A B C D E
2
2.5
3
3.5
4
-10 -5 0 5 10
Tim
e-en
trop
y
Frequency Offset (bpm)
a
b
c
Figure 3: Results from an experiment with a pulsatile flow phantom. (a) Images of the tube reconstructed at a range of hypothetical pump frequencies using the one-parameter heart rate model. The frequency at which the images were reconstructed is given on the axis below, where the axis values denote the difference between the supposed pump frequency used in the reconstruction and the true frequency. (b) The time-entropy values corresponding to the images in (a). (c) Flow patterns extracted from the image series indicated with the corresponding letters, as well as the flow pattern extracted from the ECG gated images.
2.5.3 Volunteer Measurements
Results from the volunteer carotid artery experiment are depicted in Figure 4. The heart rate
trace in Figure 4a shows a large degree of HRV during the scan, both in terms of random
fluctuations and a cyclic oscillation due to respiration. The model reflects the “mean” or
consensus of this trace; however, it fails to account for much of this variation. Regardless, there
is good agreement between the ECG gated and MOG images shown in Figure 4b and Figure 4c,
respectively, as well as the ECG gated and MOG flow patterns shown in Figure 4d.
30
Finally, Figure 4e presents the time-entropy values as a function of reconstruction heart rate
(using the one-parameter model for display purposes), evaluated on four different neighborhoods
of the images. The 11×11 voxel window was placed on the left and right carotid arteries and the
left and right jugular veins. It is apparent that despite differences in waveform shapes and vessel
sizes, all four regions are able to identify an optimal reconstruction heart rate. Furthermore, the
metric minima all occur at the same heart rate, which is consistent with the known mean heart
rate measured with ECG.
-5
0
5
10
15
20
0 0.2 0.4 0.6 0.8 1
Flow
(ml/
s)
Cardiac Phase
RC Gated
RC MOG
a
b
d
120
125
130
135
140
0 10 20 30
Hea
rt R
ate
(bpm
)
Time (s)
ECGHR Model
c
2.4
2.6
2.8
3
3.2
3.4
3.6
110 120 130 140 150
Tim
e-en
trop
y
Heart Rate (bpm)
LCRCLJRJ
e
Figure 4: Results from an experiment using the carotid arteries in adult volunteers. (a) Comparison of the optimized heart rate model and the true heart rate trace, as measured by ECG. The two-parameter model was used, and the two model parameters converged to nearly identical values. (b) and (c) show a single frame from the ECG gated and MOG image series, respectively. The magnitude images are shown on the left and the masked phase images are shown on the right. (d) Right carotid artery flow patterns extracted from the images shown in (b) and (c). The left carotid artery flow patterns were indistinguishable and were not included for clarity. (e) Metric values corresponding to four one-parameter heart rate searches in the left carotid artery (LC), right carotid artery (RC), left jugular vein (LJ), and right jugular vein (RJ).
31
2.5.4 Fetal Measurements
The results from a single fetal case with normal cardiac function are shown in Figure 5 and Table
2. The MR exam involved several measurements, each of which was optimized independently.
The full set of optimized heart rates indicates that the fetal heart rate varied between 138-175
bpm over the course of the exam. The MOG optimization and images for one representative
measurement are shown in Figure 5a and Figure 5b. The two-parameter metric optimization
shows a pronounced minimum and the images corresponding to this minimum depict a clear
view of the fetal pulmonary artery with little image artifact. Figure 5c shows three repeated
measurements of the pulmonary artery made at different times during the scan. They show good
agreement with each other and are consistent with the expected waveform shape shown in the
inset, which was taken from a normal fetal pulmonary artery Doppler trace from another subject
at the same gestational age. Additionally, the mean flow values of 287, 275, and 281 ml/min/kg,
and pulsatility indices of 3.87, 3.76, and 3.82 suggest that the measurements are reproducible.
Table 2 presents the mean flow rates and fractional distributions compiled from the full set of
measurements, as well as estimates of these values derived from the literature. Flow volumes
were divided by fetal weight, following the convention in the literature. The fractional
distribution of flow shows good agreement with previous measurements in most vessels, and the
mean flow values are within the expected range. Differences between individual values may
reflect natural variation, random error, or inaccuracies in the literature values.
The data can also be tested for internal consistency based on the conservation of blood volumes.
The flow in the main pulmonary artery is distributed between the lungs and the ductus arteriosus,
and indeed, two times the flow in the right pulmonary artery plus the flow in the ductus
arteriosus gives 295 ml/min/kg, which is in good agreement with the flow in the main pulmonary
artery (281 ml/min/kg). Also, the flow in the descending aorta is given by the sum of the flows
in the ascending aorta and the ductus arteriosus, minus the flow to the head and arms that returns
to the heart via the superior vena cava. This combination gives a value of 224 ml/min/kg, which
agrees with the measured flow in the descending aorta (217 ml/min/kg).
32
a
b
c -10
0
10
20
30
40
50
60
70
0 0.2 0.4 0.6 0.8 1
Flow
(ml/
s)
Cardiac Phase
Heart Rate 1 (bpm)
Hea
rt R
ate
2 (b
pm)
120 130 140 150 160 170120
130
140
150
160
170
2.4
2.6
2.8
3
3.2
Figure 5: Results from a 37 week fetal case with normal cardiac anatomy and function. (a) The metric value as a function of the model parameters for one representative measurement. (b) The magnitude and phase images corresponding to the optimum in (a), with the fetal pulmonary artery indicated by the arrow. (c) Three repeated measurements of the flow in the pulmonary artery. The inset shows the expected shape of the flow pattern in the pulmonary artery in a late-gestation fetus as measured by Doppler ultrasound.
33
Table 2: Measured mean flow values and fractional distributions corresponding to the 37 week fetal case shown in Figure 5, as well as reference values derived from the literature (5,7,9,25-28). Literature values are based on measurements with Doppler ultrasound and experiments involving the injection of radionuclide-labelled microspheres.
Flow (ml/min/kg) Distribution (% Combined Ventricular Output)
Vessel Measured Reference Measured Reference Combined Ventricular Output 481 400 – 553 100 100
Pulmonary Artery 281 248 – 302 58 55 – 60 Ascending Aorta 200 174 – 202 42 40 – 45 Descending Aorta 217 171 45 38 Ductus Arteriosus 165 135 – 196 34 30 – 46
Right Pulmonary Artery 65 23 – 113 13 5.5 – 12.5 Superior Vena Cava 141 83 – 165 29 23 – 37
Umbilical Vein 153 112 – 128 32 26 – 32
2.5.5 Simulation
The results of a randomly selected simulation are shown in Figure 6. The heart rate trace in
Figure 6a shows a significant amount of HRV, most of which is not accounted for in the heart
rate model. Despite this, the extracted and reference flow patterns shown in Figure 6b show good
agreement. Both the mean flow and pulsatility index are preserved to within 3% or their true
values.
Figure 7 shows the effects of increasing HRV on the flow pattern accuracy through Monte Carlo
simulation. Each bar reflects the mean and standard deviation of 1000 simulation runs with
simulated heart rate traces. For each execution, the data were optimized with the piecewise-
constant heart rate model and the mean flow and pulsatility index were extracted from the
measured flow. In general, as HRV increases, the measured flow pattern becomes less accurate;
however, over the range of expected fetal HRV the mean flow and pulsatility index are preserved
to within approximately 5 and 10 percent of their true values, respectively.
34
a
b
140
145
150
155
160
165
0 10 20 30H
eart
Rat
e (b
pm)
Time (s)
Simulated TraceOptimized Model
-5
0
5
10
15
20
0 0.2 0.4 0.6 0.8 1
Flow
(ml/
s)
Cardiac Phase
True FlowReconstructed Flow
Figure 6: Results of a simulation with σ = 7, Δ = 0.1, and a baseline heart rate of 150 bpm. (a) The simulated heart rate trace generated by Eq. 2 using the aforementioned parameters, as well as the optimized piecewise-constant heart rate model. (b) Comparison of the reference and calculated flow patterns.
0.6
0.7
0.8
0.9
1
1.1
1.2
1.3
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Flow
Par
amet
er
Heart Rate Variability ( )
PulsatilityMean Flow
Figure 7: Results of a Monte Carlo simulation testing the effects of HRV on the quality of MOG reconstruction. Each data point represents the mean and standard deviation of 1000 simulations with random HRV, determined by Δ = 0.1 and σ as shown. The mean and pulsatility are normalized against the correct values. The two-sided arrow denotes the typical range of fetal HRV based on reports in the literature (55,76).
35
2.6 Discussion
2.6.1 Validation
Collectively, the phantom, volunteer, and simulation experiments presented in this study provide
a strong validation of the MOG technique. The phantom experiment demonstrates the
effectiveness of the proposed metric in detecting the expected effects of misgating, and its ability
to accurately identify correctly gated images. The volunteer data provided in vivo application in
a representative model, and the simultaneous acquisition of an ECG signal permitted a direct
validation of both the optimized heart rate model and the measured flow pattern against accepted
techniques (30). Finally, numerical simulations showed that MOG can accurately optimize data
in the presence of realistic fetal HRV.
2.6.2 Fetal Application
The fetal measurements provided proof-of-principle application in the target population. The
fetal data reflect preliminary experience with a small number of cases, but the results are
encouraging. Validation of fetal data is difficult because a gold-standard technique is
unavailable for flow volume measurements in utero; however, these early results were
reproducible, in agreement with previous estimates, and internally consistent. It should be noted
that the preliminary fetal data set presented in this work is but one of several. MOG has been
attempted in 12 cases to date and the piecewise-constant heart rate model has provided an
acceptable reconstruction for all image series with accurate slice prescriptions.
2.6.3 Limitations
MOG
The primary limitation of MOG lies in its inability to fully correct data acquired with wildly
varying heart rates. This is evident in the Monte Carlo simulation where an obvious bias is
introduced in both the mean flow and pulsatility in the presence of high HRV. The gradual loss
in pulsatility reflects the temporal blurring that results from the inevitable scrambling of data in
the presence of high HRV. The overestimation of the mean value of a pulsatile waveform is a
well-characterized effect (43,77) that results from weighted temporal averaging due to inflow
effects in the signal magnitude. While these effects are undesirable, the results of the simulation
36
suggest that they are small in practice, and future research may further reduce this problem, as
will be discussed in the next chapter.
The second limitation of MOG is the inability to confirm the validity of the optimized
measurements. Although many of the experiments in this study included a means of validation,
no good alternative is available for fetal measurements. While the results of this study suggest
that MOG is capable of optimizing fetal data with typical levels of HRV, it is conceivable that a
particular measurement may contain sufficiently large HRV that even the optimized
reconstruction is not necessarily an accurate representation of reality. As MOG continues to be
validated in the fetal population, repeated measurements can be used to test reproducibility.
General fetal cardiac MRI limitations
Although MOG may be effective in mitigating problems associated with a lack of cardiac gating
in the fetus, there exist a wide variety of challenges associated with fetal cardiac MRI in general.
These include partial-volume effects in PC-MRI flow measurements, inaccurate slice
prescriptions, bulk fetal movement, and maternal breathing. There are a number of
improvements that may be made to address these issues in the future, which will be discussed in
the next chapter.
2.7 Conclusion
In summary, this study has demonstrated the feasibility of MOG through successful application
in a flow phantom, adult volunteers, and in utero. In addition, simulation results confirmed that
MOG is expected to perform well in the presence of HRV typical of the fetal population.
Although many other challenges remain in fetal cardiac MRI, this technique allows for
reconstruction of nontriggered phase-contrast images acquired with conventional pulse
sequences, facilitating fetal cardiac flow measurements with existing protocols and hardware.
The next chapter will discuss potential solutions to some of the issues identified in both the
methodology and results in this work. Additionally, it will include suggestions for further
avenues of research and potential extensions of the MOG technique.
37
Chapter 3 Future Work
3.1 Introduction
The previous chapter presented the theory, implementation, and validation of a new technique
for reconstructing PC-MRI data acquired without a gating signal. This chapter outlines potential
improvements to the reconstruction process, further investigation that would improve the
soundness of the conclusions, and future directions for research into the extension, adaptation,
and application of this technique.
3.2 Improvements
Although a large effort was made to optimize the scan protocol and MOG process, there is
certainly room for improvement. The previous chapter identified several limitations of MOG
and fetal MR in general, including the long scan time, the relatively low spatial resolution, the
difficulty associated with accurately prescribing image slices, and the bias introduced to flow
measurement in the presence of large HRV. These limitations may be addressed with the
improvements to the acquisition and reconstruction outlined below.
The long scan time precluded breath-holding, which introduced motion artifact due to maternal
respiration. Additionally, the long scan time increased the likelihood of bulk fetal motion,
physiologic changes, and HRV during the scan. The temporal resolution was near the minimum
allowable value, precluding the use of an increased number of views-per-segment to accelerate
the scan, although the data were not SNR limited so acceleration by some other means may be
possible. Spatial (78,79), temporal (80), or spatio-temporal undersampling (81) are all promising
options for accelerating the scan. In addition to reducing motion artifact, shorter scan times
would also decrease the amount of HRV that occurs during a scan, allowing the heart rate to be
better described by a simple model.
Second, increased spatial resolution may be beneficial for fetal flow measurements. It has been
shown that spatial resolution of at least 3 voxels per diameter is necessary for reliable flow
measurements (31,32), which is typically achievable in the fetal great vessels with the voxel size
38
used in this study; however, this is the lower bound on the acceptable resolution. While spatial
resolution cannot be increased without sacrificing scan time and SNR, automated lumen
segmentation (31) or model based approaches (82,83) may be used to increase the accuracy of
PC-MRI flow measurements in small vessels.
The presence of fetal movement between the acquisition of localizers and the PC-MRI series
often results in incorrect slice prescriptions. While, it will always depend on the experience of
the radiologist to prescribe slices accurately and identify slices that are erroneous, improvements
to the imaging protocol can also help. Minimizing scan times, prematurely terminating multi-
slice scans after the relevant anatomy has been identified, and continuously reacquiring
localizers may all help reduce this problem by ensuring that the localizer images reflect the
current position of the fetus.
The final issue that will need to be addressed is the bias in the mean flow and pulsatility that is
due to the inability of the heart rate model to fully account for the fetal HRV. This can be
overcome by adding complexity to the heart rate model used in the MOG reconstruction to better
accommodate fetal HRV. This will necessarily increase search times, but improved hardware for
computation and better code optimization may accelerate the reconstruction to the point where
higher order models are feasible.
3.3 Further Investigation
In addition to the experiments presented in the previous chapter, MOG may be further validated
with additional experiments and improvements on existing experiments. These include further
simulations with an improved simulation of the fetal heart rate, an experiment combining a gold
standard validation technique with physiological fetal HRV, a comparison against other
potentially applicable techniques, and evaluation of the technique in a larger population. These
suggestions for further validation are discussed below.
The heart rate simulation used in this work was chosen because fetal HRV is fractal in nature and
was adjusted to be quantitatively similar to true fetal HRV. It is true, however, that not all
fractals are alike, and that fetal HRV is not perfectly described by Brownian motion. A
generalized form of Brownian motion called fractional Brownian motion may provide a more
39
representative model by allowing the fractal dimension to be exactly matched to the literature.
The simulations could then be rerun with this new heart rate simulation, although the effects are
likely to be minimal.
The experiment with an adult volunteer provided excellent validation through the simultaneous
measurement of an ECG signal; however, it was limited by the fact that there are important
differences between adult and fetal HRV. An ideal experiment would involve acquiring images
in a fetal animal model, with a direct, invasive measurement of the cardiac cycle to combine a
gold standard validation with more representative fetal HRV. The gating would be acquired in
parallel with the regular oversampled acquisition, just as it was in the adult volunteer experiment,
permitting comparison between the optimized heart rate model and the measured heart rate, as
well as the MOG and traditionally gated images. Recently, Yamamura et al. (84) published
preliminary work involving PC-MRI in fetal lambs where cardiac gating was accomplished with
catheter measurements in the fetal carotid artery, suggesting that such an experiment is possible.
In addition to the validation of MOG against gold standard techniques in research settings, it
must also be directly compared against alternative techniques in the target population.
Specifically, MOG and self-gating should be applied back-to-back in several fetal cases and
compared in terms of their reconstruction qualities. In fact, some implementations of self-gating
involve sampling strategies that are consistent with MOG, allowing a single dataset to be
reconstructed with both techniques for a direct comparison.
Finally, the most important addition is the validation of MOG in a broader population. Due to
patient availability, only a limited number of fetal scans have been completed to date. Large
patient numbers could provide tests of reproducibility and reliability, further strengthening the
validation. Additionally, application in known cases of congenital heart disease could provide
measurements demonstrating the ability of PC-MRI to differentiate between healthy and
diseased states, demonstrating the utility of time-resolved PC-MRI measurements for the
assessment of the fetal heart.
40
3.4 Extensions
The most obvious extension of this technique would be its application to anatomical image
series, expanding the role of fetal MR to the assessment of cardiac anatomy and wall motion.
Although the artifacts resulting from misgated anatomical images are less severe than those in
PC-MR images, they do exist and have been well-characterized (85). The extension of this
technique to different types of images would simply involve an investigation of alternative image
metrics to identify one that is appropriate for that particular type of data. Furthermore, time-
entropy has been shown to be sufficient for MOG of PC-MRI images, but additional research in
image metrics may yield more effective metrics for this application as well.
3.5 Final Remarks
In conclusion, I have demonstrated that MOG provides an effective and reliable method of
reconstructing fetal PC-MRI data acquired without a gating signal. In all validation experiments,
MOG was able to produce measurements that were consistent with reference values, suggesting
that MOG is both reliable and accurate. While there are still several limitations that must be
addressed, the implementation of the recommendations in this chapter may resolve, or at least
reduce, many of these. More widespread application in the fetal population is still necessary;
however, MOG appears to provide an effective tool, enabling time-resolved PC-MRI flow
measurements in utero. It is my hope that the simplicity and efficacy of this technique will
promote its adoption in the lab and the clinic, contributing to a better understanding of human
fetal circulation and improved patient management in the years to come.
41
References 1. Lloyd-‐Jones D, Adams R, Carnethon M, De Simone G, Ferguson TB, Flegal K, Ford E, Furie K, Go A,
Greenlund K, Haase N, Hailpern S, Ho M, Howard V, Kissela B, Kittner S, Lackland D, Lisabeth L, Marelli A, McDermott M, Meigs J, Mozaffarian D, Nichol G, O'Donnell C, Roger V, Rosamond W, Sacco R, Sorlie P, Stafford R, Steinberger J, Thom T, Wasserthiel-‐Smoller S, Wong N, Wylie-‐Rosett J, Hong Y, Subcommittee AHASCaSS. Heart disease and stroke statistics-‐-‐2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 2009;119(3):e1-‐e161.
2. Manganaro L, Savelli S, Di Maurizio M, Perrone A, Francioso A, La Barbera L, Totaro P, Fierro F, Tomei A, Coratella F, Giancotti A, Ballesio L, Ventriglia F. Assessment of congenital heart disease (CHD): is there a role for fetal magnetic resonance imaging (MRI)? Eur J Radiol 2009;72(1):172-‐180.
3. Sklansky M. Advances in fetal cardiac imaging. Pediatr Cardiol 2004;25(3):307-‐321.
4. Van Mieghem T, DeKoninck P, Steenhaut P, Deprest J. Methods for prenatal assessment of fetal cardiac function. Prenat Diagn 2009;29(13):1193-‐1203.
5. Rudolph AM. Congenital Diseases of the Heart: Clinical-‐Physiological Considerations. Chichester: Wiley-‐Blackwell; 2009.
6. Rudolph AM, Heymann MA. The circulation of the fetus in utero. Methods for studying distribution of blood flow, cardiac output and organ blood flow. Circ Res 1967;21(2):163-‐184.
7. Rudolph A, Heymann M, Teramo K, Barrett C, Raiha N. Studies on the circulation of the previable human fetus. Pediatr Res 1971;5(9):452-‐465.
8. Boneva RS, Botto LD, Moore CA, Yang Q, Correa A, Erickson JD. Mortality associated with congenital heart defects in the United States: trends and racial disparities, 1979-‐1997. Circulation 2001;103(19):2376-‐2381.
9. Kiserud T, Ebbing C, Kessler J. Fetal cardiac output, distribution to the placenta and impact of placental compromise. Ultrasound Obstet Gynecol 2006;28(2):126-‐136.
10. Hove JR, Köster RW, Forouhar AS, Acevedo-‐Bolton G, Fraser SE, Gharib M. Intracardiac fluid forces are an essential epigenetic factor for embryonic cardiogenesis. Nature 2003;421(6919):172-‐177.
11. Rabinovitch M, Haworth SG, Castaneda AR, Nadas AS, Reid LM. Lung biopsy in congenital heart disease: a morphometric approach to pulmonary vascular disease. Circulation 1978;58(6):1107-‐1122.
12. Haworth SG, Reid L. Quantitative structural study of pulmonary circulation in the newborn with pulmonary atresia. Thorax 1977;32(2):129-‐133.
42
13. Limperopoulos C, Tworetzky W, McElhinney DB, Newburger JW, Brown DW, Robertson RL, Guizard N, McGrath E, Geva J, Annese D, Dunbar-‐Masterson C, Trainor B, Laussen PC, du Plessis AJ. Brain volume and metabolism in fetuses with congenital heart disease: evaluation with quantitative magnetic resonance imaging and spectroscopy. Circulation 2010;121(1):26-‐33.
14. De Lia JE, Cruikshank DP, Keye WR. Fetoscopic neodymium:YAG laser occlusion of placental vessels in severe twin-‐twin transfusion syndrome. Obstet Gynecol 1990;75(6):1046-‐1053.
15. Rossi AC, D'Addario V. Laser therapy and serial amnioreduction as treatment for twin-‐twin transfusion syndrome: a metaanalysis and review of literature. Am J Obstet Gynecol 2008;198(2):147-‐152.
16. Marshall AC, van der Velde ME, Tworetzky W, Gomez CA, Wilkins-‐Haug L, Benson CB, Jennings RW, Lock JE. Creation of an atrial septal defect in utero for fetuses with hypoplastic left heart syndrome and intact or highly restrictive atrial septum. Circulation 2004;110(3):253-‐258.
17. Tworetzky W, Wilkins-‐Haug L, Jennings RW, van der Velde ME, Marshall AC, Marx GR, Colan SD, Benson CB, Lock JE, Perry SB. Balloon dilation of severe aortic stenosis in the fetus: potential for prevention of hypoplastic left heart syndrome: candidate selection, technique, and results of successful intervention. Circulation 2004;110(15):2125-‐2131.
18. Chung T. Assessment of cardiovascular anatomy in patients with congenital heart disease by magnetic resonance imaging. Pediatr Cardiol 2000;21(1):18-‐26.
19. Powell A, Geva T. Blood flow measurement by magnetic resonance imaging in congenital heart disease. Pediatr Cardiol 2000;21(1):47-‐58.
20. Vimpeli T, Huhtala H, Wilsgaard T, Acharya G. Fetal cardiac output and its distribution to the placenta at 11-‐20 weeks of gestation. Ultrasound Obstet Gynecol 2009;33(3):265-‐271.
21. Cannie M, Jani J, Dymarkowski S, Deprest J. Fetal magnetic resonance imaging: luxury or necessity? Ultrasound Obstet Gynecol 2006;27(5):471-‐476.
22. Gill RW. Measurement of blood flow by ultrasound: accuracy and sources of error. Ultrasound Med Biol 1985;11(4):625-‐641.
23. Beeby AR, Dunlop W, Heads A, Hunter S. Reproducibility of ultrasonic measurement of fetal cardiac haemodynamics. Br J Obstet Gynaecol 1991;98(8):807-‐814.
24. Kiserud T, Rasmussen S. How repeat measurements affect the mean diameter of the umbilical vein and the ductus venosus. Ultrasound Obstet Gynecol 1998;11(6):419-‐425.
25. De Smedt M, Visser G. Fetal cardiac output estimated by Doppler echocardiography during mid-‐ and late gestation. Am J Cardiol 1987;60(4):338-‐342.
26. Kenny JF, Plappert T, Doubilet P, Saltzman DH, Cartier M, Zollars L, Leatherman GF, St John Sutton MG. Changes in intracardiac blood flow velocities and right and left ventricular stroke volumes with gestational age in the normal human fetus: a prospective Doppler echocardiographic study. Circulation 1986;74(6):1208-‐1216.
43
27. Mielke G, Benda N. Cardiac output and central distribution of blood flow in the human fetus. Circulation 2001;103(12):1662-‐1668.
28. Rasanen J, Wood DC, Weiner S, Ludomirski A, Huhta JC. Role of the pulmonary circulation in the distribution of human fetal cardiac output during the second half of pregnancy. Circulation 1996;94(5):1068-‐1073.
29. Simpson J, Cook A. Repeatability of echocardiographic measurements in the human fetus. Ultrasound Obstet Gynecol 2002;20(4):332-‐339.
30. Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-‐contrast MR imaging: basic facts and implementation. Radiographics 2002;22(3):651-‐671.
31. Hofman M, Visser F, Van Rossum A, Vink G, Sprenger M, Westerhof N. In vivo validation of magnetic resonance blood volume flow measurements with limited spatial resolution in small vessels. Magn Reson Med 1995;33(6):778-‐784.
32. Tang C, Blatter D, Parker D. Accuracy of phase-‐contrast flow measurements in the presence of partial-‐volume effects. J Magn Reson Imaging 1993;3(2):377-‐385.
33. Peters M, Crowe J, Piéri JF, Quartero H, Hayes-‐Gill B, James D, Stinstra J, Shakespeare S. Monitoring the fetal heart non-‐invasively: a review of methods. J Perinat Med 2001;29(5):408-‐416.
34. Feinberg DA, Giese D, Bongers DA, Ramanna S, Zaitsev M, Markl M, Günther M. Hybrid ultrasound MRI for improved cardiac imaging and real-‐time respiration control. Magn Reson Med 2010;63(2):290-‐296.
35. Liu C, Varadarajan P, Pohost G, Nayak K. Real-‐time color-‐flow MRI at 3 T using variable-‐density spiral phase contrast. Magn Reson Imaging 2008;26(5):661-‐666.
36. Spraggins TA. Wireless retrospective gating: application to cine cardiac imaging. Magn Reson Imaging 1990;8(6):675-‐681.
37. Hinks RS; Picker International, Inc., assignee. Monitored echo gating for the reduction of motion artifacts. United States patent 4,761,613. 1988 August 12, 1987.
38. Crowe M, Larson A, Zhang Q, Carr J, White R, Li D, Simonetti O. Automated rectilinear self-‐gated cardiac cine imaging. Magn Reson Med 2004;52(4):782-‐788.
39. Larson A, White R, Laub G, McVeigh E, Li D, Simonetti O. Self-‐gated cardiac cine MRI. Magn Reson Med 2004;51(1):93-‐102.
40. Thompson RB, Mcveigh ER. Flow-‐gated phase-‐contrast MRI using radial acquisitions. Magn Reson Med 2004;52(3):598-‐604.
41. Nieman BJ, Szulc KU, Turnbull DH. Three-‐dimensional, in vivo MRI with self-‐gating and image coregistration in the mouse. Magn Reson Med 2009;61(5):1148-‐1157.
44
42. Holmes W, Mccabe C, Mullin J, Condon B, Bain M. Noninvasive self-‐gated magnetic resonance cardiac imaging of developing chick embryos in ovo. Circulation 2008;117(21):e346-‐e347.
43. Bakker CJ, Kouwenhoven M, Hartkamp MJ, Hoogeveen RM, Mali WP. Accuracy and precision of time-‐averaged flow as measured by nontriggered 2D phase-‐contrast MR angiography, a phantom evaluation. Magn Reson Imaging 1995;13(7):959-‐965.
44. Hofman M, Kouwenhoven M, Sprenger M, Van Rossum A, Valk J, Westerhof N. Nontriggered magnetic resonance velocity measurement of the time-‐average of pulsatile velocity. Magn Reson Med 1993;29:648-‐648.
45. Unknown. Electronic fetal heart rate monitoring: research guidelines for interpretation. National Institute of Child Health and Human Development Research Planning Workshop. Am J Obstet Gynecol 1997;177(6):1385-‐1390.
46. Oppenheimer LW, Lewinsky RM. Power spectral analysis of fetal heart rate. Bailliere Clin Ob Gy 1994;8(3):643-‐661.
47. Sontag L, Richards T. Studies in fetal behavior: I. Fetal heart rate as a behavioral indicator. Monogr Soc Res Child Dev 1938;3(4):i-‐72.
48. Wheeler T, Murrills A. Patterns of fetal heart rate during normal pregnancy. Br J Obstet Gynaecol 1978;85(1):18-‐27.
49. Gough NA. Fractals, chaos, and fetal heart rate. Lancet 1992;339(8786):182-‐183.
50. Lange S, Van Leeuwen P, Schneider U, Frank B, Hoyer D, Geue D, Grönemeyer D. Heart rate features in fetal behavioural states. Early Hum Dev 2009;85(2):131-‐135.
51. Ortiz MR, Aguilar SD, Alvarez-‐Ramirez J, Martínez A, Vargas-‐Garcia C, González-‐Camarena R, Echeverría JC. Prenatal RR fluctuations dynamics: detecting fetal short-‐range fractal correlations. Prenat Diagn 2006;26(13):1241-‐1247.
52. Parer WJ, Parer JT, Holbrook RH, Block BS. Validity of mathematical methods of quantitating fetal heart rate variability. Am J Obstet Gynecol 1985;153(4):402-‐409.
53. Cesarelli M, Romano M, Bifulco P, Fedele F, Bracale M. An algorithm for the recovery of fetal heart rate series from CTG data. Comput Biol Med 2007;37(5):663-‐669.
54. van Meurs WL, Couto PMS, Couto CDS, Bernardes JF, Ayres-‐de-‐Campos D. Development of foetal and neonatal simulators at the University of Porto. Med Educ 2003;37 Suppl 1:29-‐33.
55. Magalhaes F, Marques de Sa JP, Bernardes J, Ayres-‐de-‐Campos D. Characterization of fetal heart rate irregularity using approximate entropy and wavelet filtering. Comput Cardiol 2006;33:933-‐936.
56. Jarisch W, Detwiler J. Stochastic modeling of fetal heart rate variability with applications to clinical risk-‐assessment and detection of fetal respiratory movements-‐preliminary results. Cybernet Sys 1980;11(3):215-‐243.
45
57. Gough NA. Fractal analysis of foetal heart rate variability. Physiol Meas 1993;14(3):309-‐315.
58. Echeverría JC, Hayes-‐Gill BR, Crowe JA, Woolfson MS, Croaker GDH. Detrended fluctuation analysis: a suitable method for studying fetal heart rate variability? Physiol Meas 2004;25(3):763-‐774.
59. Felgueiras CS, de Sá JP, Bernardes J, Gama S. Classification of foetal heart rate sequences based on fractal features. Med Bio Eng Comput 1998;36(2):197-‐201.
60. Signorini M, Magenes G, Cerutti S, Arduini D, di Bioingegneria D. Linear and nonlinear parameters for the analysis of fetal heart rate signal from cardiotocographic recordings. IEEE Trans Biomed Eng 2003;50(3):365-‐374.
61. Noll DC, Pauly JM, Meyer CH, Nishimura DG, Macovski A. Deblurring for non-‐2D Fourier transform magnetic resonance imaging. Magn Reson Med 1992;25(2):319-‐333.
62. Xiang QS, Bronskill MJ, Henkelman RM. Two-‐point interference method for suppression of ghost artifacts due to motion. J Magn Reson Imaging 1993;3(6):900-‐906.
63. Atkinson D, Hill D, Stoyle P, Summers P, Keevil S. Automatic correction of motion artifacts in magnetic resonance images using an entropy focus criterion. IEEE Trans Med Imaging 1997;16(6):903-‐910.
64. McGee KP, Felmlee JP, Manduca A, Riederer SJ, Ehman RL. Rapid autocorrection using prescan navigator echoes. Magn Reson Med 2000;43(4):583-‐588.
65. Atkinson D, Hill DL, Stoyle PN, Summers PE, Clare S, Bowtell R, Keevil SF. Automatic compensation of motion artifacts in MRI. Magn Reson Med 1999;41(1):163-‐170.
66. Manduca A, McGee KP, Welch EB, Felmlee JP, Grimm RC, Ehman RL. Autocorrection in MR imaging: adaptive motion correction without navigator echoes. Radiology 2000;215(3):904-‐909.
67. McGee KP, Felmlee JP, Jack CR, Manduca A, Riederer SJ, Ehman RL. Autocorrection of three-‐dimensional time-‐of-‐flight MR angiography of the Circle of Willis. Am J Roentgenol 2001;176(2):513-‐518.
68. McGee K, Manduca A, Felmlee J, Riederer S, Ehman R. Image metric-‐based correction (autocorrection) of motion effects: analysis of image metrics. J Magn Reson Imaging 2000;11(2):174-‐181.
69. Jansz MS, Seed M, van Amerom JFP, Wong D, Grosse-‐Wortmann L, Yoo SJ, Macgowan CK. Metric Optimized Gating for Fetal Cardiac MRI. Mag Reson Med n/a. doi: 10.1002/mrm.22542.
70. Markl M, Hennig J. Phase contrast MRI with improved temporal resolution by view sharing: k-‐space related velocity mapping properties. Magn Reson Imaging 2001;19(5):669-‐676.
71. Xiang QS, Henkelman RM. K-‐space description for MR imaging of dynamic objects. Magn Reson Med 1993;29(3):422-‐428.
46
72. Bernstein M, Grgic M, Brosnan T, Pelc N. Reconstructions of phase contrast, phased array multicoil data. Magn Reson Med 1994;32(3):330-‐334.
73. Walker PG, Cranney GB, Scheidegger MB, Waseleski G, Pohost GM, Yoganathan AP. Semiautomated method for noise reduction and background phase error correction in MR phase velocity data. J Magn Reson Imaging 1993;3(3):521-‐530.
74. Foo T, Bernstein M, Aisen A, Hernandez R, Collick B, Bernstein T. Improved ejection fraction and flow velocity estimates with use of view sharing and uniform repetition time excitation with fast cardiac techniques. Radiology 1995;195(2):471-‐478.
75. Polzin J, Frayne R, Grist T, Mistretta C. Frequency response of multi-‐phase segmented k-‐space phase-‐contrast. Magn Reson Med 1996;35(5):755-‐762.
76. Ortiz R, González R, Peña M, Carrasco S, Gaitán M, Vargas C. Differences in foetal heart rate variability from phonocardiography and abdominal electrocardiography. J Med Eng Technol 2002;26(1):39-‐45.
77. Hangiandreou N, Rossman P, Riederer S. Analysis of MR phase-‐contrast measurements of pulsatile velocity waveforms. J Magn Reson Imaging 1993;3(2):387-‐394.
78. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med 1999;42(5):952-‐962.
79. Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med 1997;38(4):591-‐603.
80. van Vaals JJ, Brummer ME, Dixon WT, Tuithof HH, Engels H, Nelson RC, Gerety BM, Chezmar JL, den Boer JA. "Keyhole" method for accelerating imaging of contrast agent uptake. J Magn Reson Imaging 1993;3(4):671-‐675.
81. Tsao J, Boesiger P, Pruessmann KP. k-‐t BLAST and k-‐t SENSE: dynamic MRI with high frame rate exploiting spatiotemporal correlations. Magn Reson Med 2003;50(5):1031-‐1042.
82. Hoogeveen R, Bakker C, Viergever M. MR phase-‐contrast flow measurement with limited spatial resolution in small vessels: value of model-‐based image analysis. Magn Reson Med 1999;41(3):520-‐528.
83. van der Weide R, Bakker C, Hoogeveen R, Viergever M. On-‐line flow quantification by low-‐resolution phase-‐contrast MR imaging and model-‐based postprocessing. J Magn Reson Imaging 2000;12(4):623-‐631.
84. Yamamura J, Schnackenburg B, Kooijmann H, Frisch M, Hecher K, Adam G, Wedegärtner U. High resolution MR imaging of the fetal heart with cardiac triggering: a feasibility study in the sheep fetus. Eur Radiol 2009;19(10):2383-‐2390.
85. Wood ML, Henkelman RM. MR image artifacts from periodic motion. Medical physics 1985;12(2):143-‐151.
47
Copyright Acknowledgements 1 Congenital Diseases of the Heart: Clinical-Physiological Considerations, 3rd Ed. Copyright © 2009 Abraham Rudolph; Reprinted with permission of John Wiley & Sons, Inc.
2 Metric Optimized Gating for Fetal Cardiac MRI, doi: 10.1002/mrm.22542. Copyright © 2010 Wiley-Liss, Inc.; Reprinted with permission of John Wiley & Sons, Inc.