m.i.b. - mini implant ball - anthogyr · ta6v titanium implant, medical grade 5 5.end the...

2
Removable prosthesis stabilisation simple - economical - pRactical M.I.B. - MInI IMplant Ball

Upload: vuongminh

Post on 28-Aug-2018

213 views

Category:

Documents


0 download

TRANSCRIPT

Removable prosthesis stabilisation

s i m p l e - e c o n o m i c a l - p R a c t i c a l

M . I . B . - M I n I I M p l a n t B a l l

ta6V titaniumImplant, medical grade 5

5.end the tightening of the implant with the click - and tightening - wrenches. the ideal in depth burying of the miB must let the implant’s head go over the gingival without any threading or apparent part of the implant’s neck. the implant can be buryied till its shows a primary stability good enough for immediate loading. if the tightening of the implant is blocked before its complete insertion, take out the implant by reversing the direction of rotation. introduce the drill an proceed to a little more drilling, before placing the implant again.

InDICatIOnSthis implant allows the stabilisation of a whole removable mandibular or maxillar prosthesis. its use will have to take into account the following precautionary measures : the estimation of the different forces, the precise evaluation of the bone volume and the rebasing of the removable prosthesis before the implant insertion.these implants maintain the prosthesis in place. in no case they are meant to support the occlusion charges. the rebasing will have to be done regularly in order not to loose the mucous supports.

2237 avenue andré lasquin- 74700 sallanches - Francephone +33 (0)4 50 58 02 37 - Fax +33 (0)4 50 93 78 60

www.anthogyr.com

pho

tos

cred

its: a

ntho

gyr

- a

ll ri

ghts

res

erve

d

©anthogyr - miB - 0710

3 µm

Ø 2,6 mm

Ø 3,5 mm

2,3 mm

6. implant and prosthesis in place.

Implants :

lg 10 mm: ref. miB V 26100 (per unit) ref. miB V 26100 4 (by 4)lg 12 mm: ref. miB V 26120 (per unit) ref. miB V 26120 4 (by 4)lg 14 mm: ref. miB V 26140 (per unit) ref. miB V 26140 4 (by 4)

ReFeRenCeS

Separating ring ref. in miB 002

analogref. in miB 003

O’Ring female attachment ref. Ho o4840 (supplied with a 50 shores o’Ring)

available separately by : - 5 o’Ring 50 shores ref. Hoo1415 50- 5 o’Ring 60 shores ref. Hoo1415 60

Kitref. in KtmiBmini

including :

> Click wrenchref. in cc

> Rack ref. in sQ miBaccessories of the rack:

- 1 round burref. in FB 20

- 1 drill Ø 2 mmref. in FHi 20 130

- 1 o’ring wrenchref. in coio

- 2 paralleling pins ref. in Jpa

- 1 gingival cutter ref. in miB 001

1

2

3

4

5

6

3,35

mm

2,8

mm

3,3

mm

10 ,

12 o

u 14

mm

Ø 5 mm

R e m o V a B l e p R o s t H e s i s s ta B i l i s at i o n

1.incision made with the gingival cutter. according to the visibility required and the bone volume, the opening of the flap is advised.in presence of a thin ridge, the use of a round bur may be necessary and shall be used in order to mark the bone before the next step (speed from 1000 à 1200 rpm).

2.Drilling with the Ø2 mm drill up to 2/3 of the implant length with a speed of 1000 to 1200 rpm with a «coma-like» drilling. Depending on the bone density, the depth can be adapted in order to optimize the primary retention of the implant.

3.assessment of the parallelism. the guide can be maintained in order to help the drilling tofollow.

4.pre-tightening of the miB implant with the implant holder. tighten the implant with a sight vertical pressure toward the apical end.this protocol starts the auto-tapping process. as soon as the tightening resistance is becoming too strong, proceed to the next step.

o’ring fitted out with a silicone joint designed to

fix the prosthesis

Ball for the removable prosthesis fixing

octagon for theimplant screwing

sanded and etched surface for a good osteo-integration

smooth machined neck in contact with the gingiva and the

junctional epithelium

anti-rotational and self-taping groove

SURGICal StepS :