Microwave DigestionGamal A. Hamid

Page 2: Microwave digestion

2

Thanks

To everyone who has helped us with support,

new books, hard/soft ware

And over the internet. Special thanks for Milestone

Page 3: Microwave digestion

3

Basic theory

Microwave Types

Hardware

Safety

Software

Samples Preparations

Basic Rules

Applications

Contents

Page 4: Microwave digestion

Basic Theory

Page 5: Microwave digestion

5

Microwaves Rays

Microwaves are a form of electromagnetic radiation with wavelengths ranging from one

meter to one millimeter; with frequencies between 300 MHz (100 cm) and 300 GHz (0.1 cm).

Microwaves are electromagnetic radiations with relatively low energy.

Microwaves promote the rotation of specific molecules in a reaction mixture. this rotation

results in increased molecular collisions and generation of heat.

The Polar reaction mixtures have the physical capability to interact with microwaves and

generate heat.

Page 6: Microwave digestion

6

Microwave Rays

c

H

λ= Wavelength

ε = wave height

C = Velocity

Page 7: Microwave digestion

7

Conductive Heating

• Heat passes through the vessel walls prior reaching

the reactants

• Slow and inefficient method dependent from

vessel material thermal conductivity

• Vessel temperature in excess of reaction mixture

temperature

Page 8: Microwave digestion

8

Microwave Heating

Liquids rapidly absorb microwave energy and Microwaves couple

with reaction mixture molecules

• Rapid temperature increase

• Independent from vessel material thermal conductivity

• Instantaneous localized superheating by ionic conduction

and dipole rotation

• No inertia (instant on-off)

• Dependant from ionic conduction and dipolar polarization

Page 9: Microwave digestion

9

Microwave Advantages

• Short digestion times. Minutes, not hours!

• No loss of volatile elements. Complete recovery of Hg,

As, Cd etc.

• No acid fumes. Improved laboratory working

conditions.

• No sample contamination from the environment.

• No cross contamination.

• Low blanks, as minimal quantities of acids are used.

• Unattended operation.

• Easily reproducible and fully documented methods.

Page 10: Microwave digestion
Page 11: Microwave digestion

11

Microwaves Types

• Microwave Acid Digestion

• Microwave Extraction

• Microwave Synthesis

• Microwave Ashing

Page 12: Microwave digestion

12

Microwave Acid Digestion

Microwave digestion is widely recognized as

the most robust sample prep technique for AA,

ICP-OES, or ICP-MS analyses.

Microwave digestion involves combining the

sample matrix and acids in a pressurized

container and elevating the solution past the

boiling point of the acid which significantly

accelerates the digestion.

Page 13: Microwave digestion

13

Microwave Extraction

A solvent or combination is chosen to absorb

energy and heat quickly.

When performed in a closed-vessel, the power

is controlled to heat the solvents above their

boiling points and allow the extractions to be

performed much faster than older methods.

From a simple set-up of multiple vessels inside a

microwave cavity, the speed and productivity

enhancement has enabled a critical

advancement in sample preparation for the

organics field.

Page 14: Microwave digestion

14

Microwave Synthesis

• Chemical synthesis is the purposeful execution of

chemical reactions among opportune starting

materials (reagents) aimed to get a product or a

mixture of products (reaction mixture).

• Microwave instruments for synthesis have the

requirements of a very homogeneous microwave

field, a stirring device (normally magnetic) and

reliable sensors to measure operative variables

such as temperature, pressure and irradiated

power.

Page 15: Microwave digestion

15

Microwave Ashing

• The PYRO reduces ashing times, from hours to

minutes.

• allows the ashing of large quantities (50 grams and

up) of multiple samples.

• microwave ashing system can accommodate sample

crucibles made of different materials, such as

metals, quartz, porcelain etc.

• Just 30 seconds are enough for the Ultra FAST

crucible to go from 1000 degrees to room

temperature.

Page 16: Microwave digestion
Page 17: Microwave digestion

17

Microwave Hardware

1. Magnetron

2. Rotating diffuser

3. Constraction

4. Sensors

5. Rotors MDR

6. Vessel

Page 18: Microwave digestion

18

1. Dual Magnetron

• Highest microwave power provides fast vessel heating

• The microwave is equipped with two magnetrons.

• Installed power is 1.800 watts allowing rapid heating of

even high throughput.

• Delivered microwave power: 1000W, controlled via

microprocessor in 10-Watt increments.

Page 19: Microwave digestion

19

Microwave Output Power Test

• A 1.0 litter container of water. Use a low, wide microwave

transparent container

• Place it in the rear of the cavity.

• The thermocouple should be removed and the jumper plug

installed.

• Stir the water and measure the initial temperature with an

accurate thermometer.

• Heat the water at maximum power for 60 seconds.

• Immediately after, stir the water and measure the maximum

final Temperature of the water.

Output power = (Tf - Ti) x 70

Page 20: Microwave digestion

20

2. Rotating Diffuser

Homogeneous microwave distribution

A pyramid-shaped rotating diffuser evenly

distributes the microwaves in the cavity.

This unique design ensures optimal microwave

distribution, preventing localized hot and

cold spots.

Page 21: Microwave digestion

21

3. Constraction

• Stainless steel door. Or impact-resistant glass window.

• Large illuminated microwave cavity

• Microwave cavity stainless steel housing with multilayer PTFE

plasma coating applied at over 350ºC.

• All hardware protected against acid/organic solvents with

polymer coating both inside and outside surfaces.

• Total of 7 micro switches of which 4 safety interlocks to prevent

microwave emission with open door.

• Exhaust located above the cavity separate from electronics to

prevent any corrosion.

Page 22: Microwave digestion

22

Pressure-Responsive Door

• The microwave door is based on the “moving wall”

principle used in high pressure autoclaves.

• The door is mounted on spring-loaded high-impact

steel bars.

• If there is a sudden pressure increase the door moves

out a small distance to release the excess pressure,

then reseals.

• Integrated safety switches cut off microwave power in

the event of pressure release.

Page 23: Microwave digestion

23

Enhanced Air Cooling

• Milestone has available a “Enhanced Air Cooling

Kit”.

• This kit contains adapters for the Ethos lab station

that direct the flow of air through the top of the

Ethos lab station and around vessels in segmented

rotors.

• Using this kit can provide more effective air cooling

after the completion of the microwave heating

process, but it is not as effective as water cooling.

Page 24: Microwave digestion

24

Exhaust

• Warning: In laboratories where acids and acid

vapors are not present (common to solvent

extraction or most organic synthesis applications),

there should be little or no consequence to

installing the Labstation in a dedicated (solvent

free), active fume hood.

• As with all electronics, one should take care to

never operate a microwave lab station in an

explosive atmosphere.

Page 25: Microwave digestion

25

4. Reaction Sensors

The Microwave is equipped with reactions sensors for

temperature and pressure control.

Temperature is measured either with

1. Direct sensor in a reference vessel, or via

2. A contact-less high sensitivity infrared sensor,

which effectively measures and controls the

temperature in all vessels.

Page 26: Microwave digestion

26

The ATC Automatic Temperature Control

system allows for direct continuous

monitoring and control of a reference

vessel up to 300°C.

• Easy and rapid insertion and removal of the

sensor

• No sensor bending

• Triple layer of PTFE/ceramic/PTFE

protection

A- Direct Temperature Control

Page 27: Microwave digestion

27

B- Contact-less Temperature Control

• Contact-less temperature control in all vessels

• The Microwave features a focused, high

sensitivity

• Infrared sensor for contact-less temperature

monitor and control of all vessels, with

individual temperature profiles for each vessel.

Page 28: Microwave digestion

28

C- Direct Pressure Control

• The APC Automatic Pressure Control system allows for

direct monitoring and control of a reference vessel up

to 100 bar.

• Pressure control up to 55bar

• Pressure monitoring and control are ideal for highly

reactive organic samples or ‘unknowns’, and for

method development, to maintain pressure limits

within the vessel’s specifications.

• Solid state sensor with built-in electronics fully

protected from acid vapors

Page 29: Microwave digestion

29

D- Automatic Gas Detector NOX

• NOX are released if vessels vent

• Monitoring the NOX presence in the microwave cavity allow

safer working conditions and consistant results

• This sensor continuously monitors the concentration of acid

vapors in the microwave cavity.

• Onceover a preset level, the sensor automatically adjusts the

microwave power, to effectively limit the pressure in each

vessel and prevent overpressure and venting.

• All vessels in the cavity are monitored simultaneously.

Page 30: Microwave digestion

30

5. Microwave Digestion Rotors MDR

a) High Pressure Rotor

b) Medium Pressure Rotor

c) Medium Pressure High Throughput Rotors

d) High Temperature Rotor

e) High Throughput Rotor

Page 31: Microwave digestion

31

TFM Features

• Trade name for chemically modified PTFE poly tetra

fluoro ethylene

• Dense structure

• Chemically inert

• Thermal insulator

• Microwave transparent

• Melting range 320-340°C

• Maximum working temperature for extended use

260°C, for short time 300°C

Page 32: Microwave digestion

32

a- High Pressure Rotor (SK10)

• High pressure rotor suitable for all samples,

including difficult organic and inorganic refractory

materials

• Ensures complete digestion of larger amounts of

highly reactive organics, such as food/feed, oils,

plastics, pharmaceuticals.

Page 33: Microwave digestion

33

b- Medium Pressure Rotor (SK 12)

• General purpose rotor suitable for most samples.

• The high-strength rotor body holds multiple digestion

vessels made of high purity TFM.

• Allows for the complete digestion of water, sludge,

municipal waste, biological materials, metals and

alloys.

Page 34: Microwave digestion

34

C- Medium PressureHigh Throughput Rotors (PRO 16/24)

• High throughput rotors suitable for easy organic

samples or environmental applications,

according to the U.S. EPA method 3051A.

• Easy handling and rapid cooling capabilities.

Page 35: Microwave digestion

35

d- High Temperature Rotor (Nova 10)

• High temperature rotor able to withstand extremely

high temperatures for long lasting complete

digestion of the most challenging samples.

• Suitable for very difficult inorganic materials such as

silicon carbide, metal oxides, geological, and

refractory materials.

Page 36: Microwave digestion

36

e- High Throughput Rotor (Multiprep)

• High throughput carousel, able to perform easy

digestions or leaching-type reactions on a variety

of samples, such as soils, sediments, wastewater

etc.

Page 37: Microwave digestion

37

Rotors VS Application fields

Page 38: Microwave digestion

38

Rotors Specifications

Page 39: Microwave digestion

39

6. The Vessel

Milestone ‘vent-and-reseal’ vessels are the

foundation of it’s microwave digestion

technology.

Milestone offers a wide variety of vessels for all

applications from digestion, synthesis, extraction,

evaporation, fusion, and protein hydrolysis.

Page 40: Microwave digestion

40

Advanced Vessel Technology

• ‘Vent-and-reseal’ vessels are the foundation

of Milestone’s microwave digestion

technology.

• Continuously enhanced, this patented*

technology provides analysts with

unsurpassed performance capabilities.

• Highest temperature and pressure

• Highest safety standards

• Easy of use

• Fast air cooling

Page 41: Microwave digestion

41

The Segment components

A vessel that contains the sample,

A vessel cover that seals the vessel and acts as a

overpressure release valve,

A safety shield that surrounds the vessel and provides

structure and safety for operation at elevated

pressures,

A pressure adapter plate, that fits on top of the vessel

cover,

A spring rated to the pressure performance

specifications of the particular vessel/rotor

combination, and

A vent indicator ring.

Page 42: Microwave digestion

42

The Vessel Material

• Microwave transparency

• Chemical resistance

• Extrusion capability (thermoplasticity)

• Machining capability

• High working temperature

• High density (low porosity)

• Thermal insulation

Page 43: Microwave digestion

43

Cleaning Vessels and Covers

• The simplest method for cleaning vessels and covers is to process

a blank (acids or solvents) under the same process conditions

(reaction parameter profile) that is used for preparing the

analytical samples.

• Another traditional method for cleaning vessels and covers is to

soak them in an acid or solvent bath, at elevated temperature,

for an extended period of time before they are rinsed, dried and

stored.

• Typical conditions include 1:1 nitric acid and water in a Teflon

container maintained at 80 - 90 degrees for an hour or more.

Page 44: Microwave digestion

44

Reference Vessel(s): purpose

• The reference vessel is a specialized vessel that facilitates

reaction parameter monitoring (temperature and / or

pressure) and allows feedback process control.

• To allow feedback process control, a reference vessel must

be prepared to contain chemistry identical to any other

sample vessel being processed during the same run.

Page 45: Microwave digestion

45

Cleaning The Reference Vessel

• The only way to clean the reference vessel and assembled

cover is to process a blank (using the same process acids or

solvents) under the same process conditions (reaction

parameter profile) that are used for the analytical samples.

• Although the reference vessel can be cleaned using the

soaking method or by using Milestone’s trace CLEAN

system, these methods cannot be used to clean the

reference vessel cover.

Page 46: Microwave digestion

46

Loading Samples

• Sample materials, normally, are quantitatively introduced into a

vessel using an analytical balance.

• Acid or solvent mixtures, normally, are quantitatively introduced

into a vessel using an analytical balance or calibrated pipettes.

• Note: Sample materials should be placed into the bottom of the

vessel, not on the sides, and should be completely submerged by

the acid or solvent mixture.

• Any sample materials that stick to the sides of the vessel should

be washed down into the pool to avoid potential vessel damage.

Page 47: Microwave digestion

47

Minimum Volume

• The volume of liquid in the reference vessel must be sufficient to

submerge the end of the thermowell to a depth of at least one

centimeter.

• In the standard 100 mL vessels, the minimum recommended volume

is 8 mL at room temperature.

• More importantly, the thermowell must be submerged to a depth of

at least one centimeter at the maximum temperature of the heating

process.

• This fact should be taken into consideration when using liquids with a

high vapor pressure that will result in a substantial fraction of the

liquid mass in the vapor phase at the maximum temperature of the

heating process.

Page 48: Microwave digestion

48

Contamination Test

• Run the following program:

o 1 minute 1000 W (for 10 Vessels)

o 1 minute 600 W (for 6 or less Vessels)

• As soon as the run is finished, check the temperature of

all Vessels (must be at room temperature).

• With temperature > 50°C please make the cleaning

procedure

• The yellow TFM vessel has been contaminated by NOx

developed during strong decomposition.

Page 49: Microwave digestion

49

Water Cooling

• Milestone’s Segmented and monobloc rotors can be

effectively water cooled at the completion of the microwave

heating process.

• Water cooling can be accomplished by using the standard

cooling station (available from Milestone) or by simply placing

the monobloc rotor or individual segments into any

convenient container that can hold water.

• A lab sink filled with water, so that the vessels are immersed

up to the bottom of the indicator ring (a little more than three-

quarters up from the bottom of the safety shield) is

appropriate for the purpose.

Page 50: Microwave digestion
Page 51: Microwave digestion

51

Safety 1

• External fire protection should be installed according to local regulations for equipment

operating unattended.

• If materials inside the equipment should ignite, keep the cavity closed, switch off the

equipment and disconnect the power cord or turn off the power at the fuse or circuit

breaker.

• The equipment shall not be used for heating and ventilation of laboratories.

• The equipment shall not be used as a sterilizing equipment.

• The equipment shall not be used as a medical electrical equipment.

• The equipment shall not be used for heating food and beverages.

• The equipment shall not be used for mass production.

• The equipment shall not be used in an explosive atmosphere.

Page 52: Microwave digestion

52

Safety 2

• The equipment shall only be operated indoors, where the altitude does not exceed

2,000 meters (1,242 feet), at temperatures between +5°C and +40°C +(41°F and +104°F).

• The maximum relative humidity is 80% for temperatures up to 31°C (87.8°F) decreasing

linearly to 50% relative humidity at 40°C (104°C).

• The equipment shall be operated in places where fluctuations of the main supply

voltage are not exceeding values of 1.10 times the nominal voltage and are not below

0.90 times the nominal voltage.

• The equipment shall only be operated in places where transient overvoltage on the

main supply are not exceeding values of overvoltage category II specified in IEC 60364-

4-443.

Page 53: Microwave digestion

53

Safety 3

The power supply must be protected by a residual current circuit breaker with other

current protection, interrupt capability equal or more than 4500A, nominal current 16A (a

230V) and 32A (a 115V), characteristics C-curve, residual current 30mA.

Power line must have strong earth connection.

The equipment shall only be operated in places where normally only non-conductive

pollution occurs and occasionally, however, a temporary conductivity caused by

condensation may occur.

The equipment shall not be stored or operated in places where special conditions prevail,

such as the presence of a corrosive or explosive atmosphere (dust, vapor or gas).

Page 54: Microwave digestion
Page 55: Microwave digestion

55

Heating Process

• When an exothermic reaction drives digestion

temperature above the programmed ramp rate,

EasyCONTROL instantly shuts off microwave

power, applying it again when the temperature

falls back to the ramp rate.

• Determining the microwave power settings

appropriate for a given run, as it automatically

adjusts the power to match the number of

samples in the rotor.

Page 56: Microwave digestion

56

Easycontrol Software

• The terminal runs Milestone’s unique Easycontrol software, to

provide simple, user-friendly control of the microwave sample

preparation process with automatic, real-time monitoring and

feedback-based control of multiple parameters.

• Simply recall a factory stored method or create a new one;

• press ‘START’ and the system will automatically follow the user

defined temperature or pressure profile, utilizing a

sophisticated PID algorithm to regulate the microwave power.

Page 57: Microwave digestion

57

The Benefit of Software

On-line control of all parameters.

• Direct temperature control 20 times per second.

• Temperature driven microwave power output:

o Reaction is controlled at any temperature

o Better control of exothermic reactions.

• Same digestion quality for different sample sizes.

• One method only needed for any number of samples.

Page 58: Microwave digestion

58

Login

• Switch on the main switch of the

START, located on the front panel.

• The Easycontrol display will show:

• Type your identification and confirm

you with o.k.

Page 59: Microwave digestion

59

Program

The program page contain four minus and six icons

1. Method 2. Parameters 3. Wave 4. Run

Page 60: Microwave digestion

60

1. Method

The theoretical expectation of :

1. Temperature curve

2. Pressure curve

Of the reactions according the methods.

Page 61: Microwave digestion

61

2. Parameter

• Select the QP on “50” medium sensitivity by scrolling up or down.

• Select the “QP Check” (activated icon). Press in the window of Ventilation.

• The display will show a keyboard: Digit the time “10”, confirm by pressing OK.

• Control for permits the selection of different controls such as T1, T2,P, if only the

Thermocouple ATC-400 is installed to use T1.

• The Twist must be always ON (activated icon) when temperature sensor T1 and/or

pressure sensor P are installed.

• The Rotor control ON (activated icon) when the IRTC sensor T2 (External temperature) is

installed.

• The Start parameters always ON (activated icon) when the temperature sensor T1

(Internal temperature) is installed.

Page 62: Microwave digestion

62

3. Wave

• Introduce the desired Time parameters:

• Press in the window “00.00.00” of t.

• The display will show a keyboard.

• Digit in the window “00.10.00” confirm by

pressing OK.

• Introduce the desired Energy parameters:

• Press in the window “0” of E.

• The display will show a keyboard:

• Digit into the window “1000”, after confirm

pressing OK

Page 63: Microwave digestion

63

4. Run

Complete “on-line” control and graphic

representation of all reaction parameters

1. Time

2. Microwave power

3. Internal temperature

4. External temperature

5. Pressure

6. Acid vapors level in the microwave cavity

Page 64: Microwave digestion
Page 65: Microwave digestion

65

Laboratory Requirements

• The lab bench can easly handle the combined weight of cmplete station “ 100 kg”.

• The exhaust hose must be connected to some source of active ventilation:

• The labstation draws a maximum of 16 amps on a 230 VAC single phase line.

• The labstation must be connected to a grounded, single phase line, with 220–230 vac output.

• Milestone recommends that all addition of acid mixtures or hazardous solvents to vessels

should take place inside an appropriately vented fume hood.

• Milestone recommends that vessels be opened inside an appropriately vented fume

hood to contain and convey reaction vapors.

• The instrument requires sufficient clearance (approximately 15 centimeters space) on all sides

of the chassis to allow adequate movement of air.

• Microwave labstation should not be permanently installed inside an active fume hood due to

the potential for acid vapors to attack and accelerate the corrosion of internal electronic.

Page 66: Microwave digestion

66

Sampling Method Structure

1. Summary

2. Microwave equipment

3. Sample amount

4. Reagents

5. Procedure

6. Microwave program

7. Temperature profile

8. Notes

Page 67: Microwave digestion

67

Example

1- SUMMARY

This method provides for the acid digestion of the Sludge sample in a closed

vessel device using temperature control microwave heating for the metal

determination by spectroscopic methods.

2- MICROWAVE EQUIPMENT

Milestone ETHOS Labstation with easy WAVE or Easycontrol software

HPR1000/10S high pressure segmented rotor.

3- SAMPLE AMOUNT 1.0 g

4- REAGENTS

8 ml of HNO3 65%, 2 ml of HCl 37%

Page 68: Microwave digestion

68

5- PROCEDURE

a) Place a TFM vessel on the balance plate, tare it and weigh of the sample.

b) Introduce the TFM vessel into the HTC safety shield.

c) Add the acids; if part of the sample stays on the inner wall of the TFM vessel, wet it by

adding acids drop by drop, then gently

d) swirl the solution to homogenize the sample with the acids.

e) Close the vessel and introduce it into the rotor segment, then tighten by using the torque

wrench.

f) Insert the segment into the microwave cavity and connect the temperature sensor

g) Run the microwave program to completion.

h) Cool the rotor by air or by water until the solution reaches room temperature.

i) Open the vessel and transfer the solution to a marked flask.

Page 69: Microwave digestion

69

6- MICROWAVE PROGRAM

7- TEMPERATURE PROFILE

8- NOTES

Use up to 500 Watt for operations with 3 or less vessels simultaneously.

This procedure is only a guideline and it may need to be modified or changed to obtain the

required results on your sample.

Always use hand, eye and body protection when operating with the microwave system.

Page 70: Microwave digestion

70

Acid Classification in Digestion Reactions

NONOXIDIZING

• Hydrochloric acid

• Hydrofluoric acid

• Phosphoric acid

• Diluted sulfuric acid

• Diluted perchloric acid

OXIDIZING

• Nitric acid

• Hot concentrated perchloric acid

• Concentrated sulfuric acid

• Hydrogen peroxide

Page 71: Microwave digestion

71

Nitric Acid

• It dissolves most metals forming soluble nitrates, exceptions are

Au and Pt (not oxidated) and Al, B, Cr, Ti and Zr (passivated)

• These metals require acid mixtures or diluted nitric acid

• Often mixed with H2O2, HCl and H2SO4

• Available in high purity for trace analysis

Page 72: Microwave digestion

72

Nitric Acid

• Boiling point is 120°C at 65% concentration

• Poor oxidizing strength at concentrations less than 2M; oxidizing

strength increases with concentration and reaction T&P

• Most common acid for oxidation of organic matrices

(CH2)X + HNO3 = CO2(g) + NOX(g) + H2O

Page 73: Microwave digestion

73

Hydrofluoric Acid

DIGESTION

• Boiling point is 108°C at 40% concentration

• Nonoxidizing, strong complexing nature

• Used in digestion of minerals, ores, soils, rocks and even botanical samples

• Major use is the decomposition of silicates

• SiO2 + 6HF = H2SiF6 + 2H2O

• Often used in combination with HNO3 or HClO4

Page 74: Microwave digestion

74

Hydrofluoric Acid

• EVAPORATION/CONCENTRATION

• Following dissolution, many analyses require removal of HF to

prevent equipment damage or to resolubilize insoluble fluorides

H2SiF6 = SiF4 + 2HF

• Many analytes such as As, B, Se, Sb, Hg, Cr may volatilize

Page 75: Microwave digestion

75

Hydrochloric Acid

• Boiling point of azeotropic mixture with H2O with 20,4% HCl is 110°C

• Available with 38% concentration

• Nonoxidizing

• It dissolves salts of weak acids (carbonates, phosphates) and most

metals are soluble with the exception of AgCl, HgCl and TiCl

• Excess of HCl improves the solubility of AgCl, converted into AgCl2-

Page 76: Microwave digestion

76

Hydrochloric Acid

• Strong complexing nature

• Widely used for iron-based alloys because of its ability to hold

large amounts of chloro-complex in solution

• Other complexes formed are Ag (I), Au (II), Hg (II), Ga (III), Tl

(III), Sn (IV), Fe (II) and Fe (III)

• It does not dissolve oxides of Al, Be, Cr, Ti, Zr, Sn and Sb;

sulphates of Ba and Pb, group II fluorides, SiO2, TiO2 and ZrO2

Page 77: Microwave digestion

77

Sulfuric Acid

• Boiling point is 340°C at 98% concentration, exceeding max

working temperature of Teflon vessels

• Careful reaction temperature monitoring is required to prevent

vessel damages

• It destroys organics by dehydrating action

• Many sulfates are insoluble (Ba, Sr, Pb)

Page 78: Microwave digestion

78

Perchloric Acid

• Boiling point is 203°C at 72% concentration

• Powerful oxidizing acid when used warm

• Hot and concentrated decomposes violently organic matter

• Nearly all perchlorates are soluble

• HClO4 decomposes at 245°C in microwave closed vessel with

dangerous amounts of by-products and tremendous excess

pressure

Page 79: Microwave digestion

79

General Rules for Usage of Perchloric Acid

• Rule #1: do not use it conc.

• Use only very diluted perchloric acid

• Mix it with other acids (but never with sulfuric acid)

• Never exceed 200°C

• Use it only to perform a two-step digestion

• Perchloric acid is normally not required for the closed vessel

microwave digestion of organic samples

Page 80: Microwave digestion

80

Hydrogen Peroxide

• Oxidizing agent

• 2H2O2 = 2H2O + O2

• Added to HNO3 it reduces nitrous vapors and it

accelerates the digestion of organic samples by raising

the temperature

• Typical mixture ratio is HNO3:H2O2=4:1

Page 81: Microwave digestion
Page 82: Microwave digestion

82

Rules 1

• All vessels must contain same type and amount of sample/reaction mixture.

• The minimum and maximum volumes are set in accordance with the rotor

specification.

• The maximum volume must be approximately 50/60% of the total volume of the

vessel.

• Never use a blank in the reference vessel.

• Never heat water in the reference and/or in one of the standard vessels.

• Before adding reaction mixture rinse the vessels with distilled water in order to

remove any electrostatic field.

• To prevent hot spot place the sample on the bottom of the vessels, if part of the

sample stays on the inner wall of the vessel, wet it by adding acids drop by drop.

Page 83: Microwave digestion

83

Rules 2

• If during the addition of the reagents at your sample an exothermic reaction is observed,

leave the vessel under fume hood for five minutes, before closing.

• Before running any test check that the method (temperature and pressure) is suitable for

your type of reactor.

• For strong exothermic reactions add 2 ml of distilled water.

• Periodically check the status of vessels, covers, shields, adapters, springs.

• After starting the microwave program always check if the temperature and/or pressure

sensor are free to rotate.

• Never open the reaction vessels at high temperature or above the boiling temperature of

the reagents

• Do not use ice to cool down the rotor after microwave heating as the thermal shock could

damage the safety shields.

Page 84: Microwave digestion

84

Basic Method Optimization

POOR DIGESTION QUALITY

• Increase T1 of the two last program Steps in 10°C increments.

• Extend Time of the last step.

• Reduce weight and / or increase quantity of chemicals /concentration of acids.

• Optimize acid mixtures particularly with inorganic samples.

• Use stirrer.

• With good quality of digestion, the last program step can be reduced

• For method development use a small sample amount (e.g. 100 mg), low temperature,

low power and short time

Page 85: Microwave digestion
Page 86: Microwave digestion

86

Page 87: Microwave digestion

87

Approved Methods

• ASTM (American Society of Testing and Materials) Methods

• USCPSC (US Consumer Product Safety Commission) Methods

• US Pharmacopeia Methods

• U.S. EPA Methods

Page 88: Microwave digestion

88

Applications

• Food/Feed

• Environment

• US EPA methods

• Clinical/Pharmaceutical

• Petrochemical

• Plastic/Polymer

• Metal/Alloy

• Geochemistry/Mining

• Ceramic/Refractory

• Various

Page 89: Microwave digestion

89

Application Table

Applications NO. Applications NO.

Food/Feed 72 Plastic/Polymer 15

Environment 45 Metal/Alloy 50

High temp. rotor 12 Geochemistry/Mining 71

Clinical/Pharmaceutical 25 Ceramic/Refractory 24

Petrochemical 21 Chemicals 18

Agriculture 9 Catalyst-pigment 12