miz-28 operating guide (180 - 10020153 - 1 - d) - 1
TRANSCRIPT
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
1/169
Operating Guide
MIZ-28
Eddy Current Test Instrument
ITEM 10020153
24 Apri l, 2007
Rev. D
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
2/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
3/169
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . .1-1About This Operating Guide 1-1
Warranty & Quality Information 1-1
Electronic and Mechanical Equipment 1-1
Probes, Cables and Spare Parts 1-2
General Terms 1-2
O.E.M. Equipment 1-2
Customer-Paid Equipment Repairs 1-2
Return Policy 1-2
Product Description. . . . . . . . . . . . . . . . . . . .2-1Cleaning 2-2
Maintenance 2-2
Environment Information 2-2
General Operation. . . . . . . . . . . . . . . . . . . . .3-1
MIZ-28 Checklist 3-1Initial Setup of the MIZ-28 3-2
Front Panel Buttons 3-4
Navigating Through the MIZ-28 3-5
The MIZ-28 Main Menu 3-7
MIZ-28 Keyboard Operation 3-9
Selecting the Language 3-11
Quick Tutorials 3-13
Bobbin Coil Exam 3-13
MRPC Exam 3-15
RFT Setup for Ferromagnetic Tubing 3-17
HVAC 3-19
Configuration . . . . . . . . . . . . . . . . . . . . . . . .4-1Naming the Configuration 4-4
Selecting Configuration Options 4-5
Config Type 4-5
External Encoder 4-7
Adjusting the Sample Rate 4-8
Sample Rates in Multiplexed Mode 4-8
Sample Rates in Simultaneous Mode 4-8
Setting Test Frequency Values 4-9
Selecting a Drive Setting 4-10
Selecting a Channel 4-11
Storing a Configuration 4-12
Config List . . . . . . . . . . . . . . . . . . . . . . . . . .5-1
Viewing the List Options 5-3Scrolling Through the Configuration List 5-3
Saving a Configuration 5-3
Recalling a Configuration 5-4
Copying To and From Internal and Mass Storage 5-5
Removing a Configuration 5-7
Changing the Directory 5-8
Naming the Configuration 5-9
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
4/169
Table of Contents
Test / Review . . . . . . . . . . . . . . . . . . . . . . . .6-1Testing Mode 6-2
Data Control 6-4
Encoder 6-5
Manual Tube Info 6-6
Message Summary 6-7
User Message 6-8
Summary Form 6-9
Start / Stop Recording 6-10
Stop Acquisition 6-10
Tube End Detection 6-10
Adjust Tolerance 6-12
Adjust Data 6-13
Run Test Plan 6-13
Start / Stop Recording 6-13
Stop Acquisition 6-13
Review Mode 6-14
Data Cursor Control 6-16Adjust Tolerance 6-17
Adjust Data 6-17
Measure 6-18
Probe Speed Check 6-19
Tube End Detection 6-20
Filter Setup 6-21
Start Acquire 6-22
Analysis Setup. . . . . . . . . . . . . . . . . . . . . . . .7-1Curve Setup 7-3
Cursor Control 7-3
Measure 7-4
Adjust Curve 7-5
Curve Table 7-6Data Mixing 7-7
Setup File 7-9
Using the Toolbox Screen . . . . . . . . . . . . . . . .8-1Setting the Frequency Plot 8-3
Select Coil 8-3
Adjust Position 8-3
Adjust Range 8-3
Adjust Scale 8-3
Adjust Drive 8-4
Setting the Drive Plot 8-5
Select Coil 8-5
Adjust Frequency 8-5Viewing the Material List 8-6
Scrolling Through the Material List 8-6
Setting the Frequency 8-6
Setting the Wall Thickness 8-6
mm Ms/m Mils IACS 8-6
Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1Selecting the Language 9-3
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
5/169
Table of Contents
Setting the Speaker On/Off 9-4
Setting the Time/Date 9-5
Setting the IP Address 9-6
Selecting the Graphics Dump Destination 9-7
Setting the Remote Inputs 9-8
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . .10-1
Test Plans . . . . . . . . . . . . . . . . . . . . . . . . .11-1Viewing the Current Test Plan 11-3
Scroll 11-3
Go To Top 11-3
Find Next Untested 11-3
Mark Unmark 11-3
Selecting a Test Plan 11-4
Scroll 11-4
Open 11-4
Remove Plan 11-4
Importing a Test Plan 11-5
Changing the Source & Destination 11-5
Scroll 11-5
Import Plan 11-5
Mass Storage Setup . . . . . . . . . . . . . . . . . . .12-1Selecting the Recording Directory 12-3
Recording Options 12-4
Selecting the File Format 12-4
Selecting the Record To Destination 12-5
Selecting the Spool To Destination 12-6
Setting the Network IP Address 12-7
Setting the Mount Point 12-8Mounting the Network Drive 12-9
Setting the Time Zone 12-10
Mass Storage. . . . . . . . . . . . . . . . . . . . . . . .13-1Data Files 13-2
Scroll 13-2
Open 13-3
Delete 13-3
Selecting the List Directories 13-3
Scroll 13-3
Open a Directory 13-3
Deleting a Directory 13-4
Copying Directories to Mass Storage 13-4Copying Directories to the Network Drive 13-4
Detailed Tutorials . . . . . . . . . . . . . . . . . . . .14-1Bobbin 14-1
Bobbin Hardware Setup 14-1
Bobbin System Setup Steps 14-2
MRPC 14-9
MRPC Hardware Setup 14-9
MRPC System Setup Steps 14-10
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
6/169
Table of Contents
MRPC Rate Calculations 14-16
RFT 14-17
RFT Hardware Setup 14-17
Configuration Setup Single Frequency 14-18
HVAC 14-20Hardware Setup 14-20
HVAC System Setup Steps 14-21
Test Plan Setup 14-27
Specifications & Features . . . . . . . . . . . . . . .15-1Basic Specifications 15-1
Pin Functions 15-2
Instrument Control 15-4
Front Panel Control 15-4
VGA Monitor Output 15-4
Data Display 15-4
Configuration Display 15-4
Display Rotation 15-4Display Scaling 15-4
Display Positioning 15-4
Keyboard Control 15-4
MIZ-28 Features 15-5
Frequency Plot 15-5
Drive Plot 15-5
Memory 15-5
USB Devices 15-5
Digital Recording 15-5
Configuration Storage 15-5
Firmware Upgrades 15-6
Probe Drive Adjustment 15-6
Filters 15-6
Sample Rate 15-6Signal Measurement Capability 15-7
Appendix A. . . . . . . . . . . . . . . . . . . . . . . . . A-1
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I-1
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
7/169
List of Figures
Figure 2-1 MIZ-28 Front and Back Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1Figure 2-2 WEEE Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2Figure 3-1 On/Off Switch & Coil Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2Figure 3-2 Bobbin Coil Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3Figure 3-3 Front Panel Buttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-4Figure 3-4 Navigation Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5Figure 3-5 MIZ-28 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5Figure 3-6 Primary & Secondary Function Buttons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6Figure 3-7 MIZ-28 Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7Figure 3-8 Plugging in the Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9Figure 3-9 Keyboard Icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9Figure 3-10 Select Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11Figure 3-11 Select Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12Figure 3-12 Bobbin Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13Figure 3-13 Bobbin Probe Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14Figure 3-14 Bobbin Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Figure 3-15 MRPC Hardware Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15Figure 3-16 MRPC Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15Figure 3-17 MRPC Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16Figure 3-18 RFT Hardware Setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17Figure 3-19 RFT Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-17Figure 3-20 HVAC Hardware Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-19Figure 3-21 AC3 Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-19Figure 3-22 HVAC Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20Figure 4-1 Open Configuration Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1Figure 4-2 Configuration Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2Figure 4-3 Configuration Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3Figure 4-4 Config Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4Figure 4-5 Config Options. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Figure 4-6 Config Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6Figure 4-7 External Encoder. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7Figure 4-8 Sample Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8Figure 4-9 Frequency Select. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9Figure 4-10 Drive Select. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10Figure 4-11 Channel Select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11Figure 4-12 Config Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-12Figure 5-1 Open Config List Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-1Figure 5-2 Config List Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2Figure 5-3 Config List Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2Figure 5-4 List Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3Figure 5-5 Recall Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4Figure 5-6 Copy to Mass Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Figure 5-7 Change Directory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6Figure 5-8 Copy to Internal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6Figure 5-9 Remove Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-7Figure 5-10 Change Directory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8Figure 5-11 Config Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-9Figure 6-1 Open Test/Review Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1Figure 6-2 Test/Review Menu 1 StructureTest Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2Figure 6-3 Test/Review Menu 2 StructureTest Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3Figure 6-4 Testing Mode Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4Figure 6-5 Encoder Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
8/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
9/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
10/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
11/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
12/169
Product Description
Page 2-1
2.0 Product Description
The MIZ-28 instrument is used for eddy current inspections of heat exchanger tubing in balance-of-plant applica-
tions. It works seamlessly with all existing Zetec software and hardware products.
Zetecs MIZ-28 offers a color display, selectable data presentation, menu-driven user interface, high digital sampling
rates, bandpass and differential filters, and two independent balance controls. This versatile, fully digital tester
provides maximum stability, reliability, and sensitivity.Testing capabilities include heat exchanger and air condi-
tioner tube testing, weld inspection, and a broad variety of other eddy current applications. Packaged in a rugged
portable carrying case, the MIZ-28 is ideal for industrial environments.
An active matrix color display shows test information from almost every angle. The use of color provides quick chan-
nel identification without having to read the setup screen. The real-time display constantly indicates the coil
impedance point shift and closely emulates a storage oscilloscope display.
The MIZ-28 presents the phase and amplitude vectors of the eddy current signal as a two-dimensional Lissajous fig-
ure. Simultaneous display of up to four Lissajous figures and two strip charts enables you to efficiently monitor mul-
tiple data channels.
The MIZ-28 internal hard drive adds 40 GB of data storage capacity to the MIZ-28 for data acquisition and tube list
storage.
Figure 2-1 MIZ-28 Front and Back Panels
It is recommended that you are familiar with Eddy Current theory and application before using the
MIZ-28, and that you use it only for the functions for which it has been designed.
Zetec offers a wide variety of courses dealing with electromagnetic inspection theory and applica-
tions. This includes classes that are specific to either eddy current testing (ECT) or remote field
testing (RFT). Some formal courses can be used to help qualify inspectors to meet specific certifi-
cation criteria for either data acquisition or data analysis. Other informal courses can be designed
with variable content and duration to meet specific application needs. Contact the Training Depart-
ment at 425-974-2700 for more details.
CAUTI
ON
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
13/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 2-2
2.1 Cleaning
2.2 Maintenance
The MIZ-28 is designed to be maintenance free. However, if it does require repair, do not attempt to repair it your-
self. Contact Zetec for more information.
2.3 Environment InformationThis electronic equipment was manufactured according to high quality standards to ensure safe and reliable opera-
tion when used as stated in this manual. Due to its nature, this equipment may contain small quantities of sub-
stances known to be hazardous to the environment or to human health if released in the environment. For this
reason, Waste Electrical and Electronic Equipment (commonly known as WEEE) should never be disposed of in the
public waste stream. The Crossed-Out Waste Bin label affixed to this product is a reminder to dispose of this end
of life product in accordance with local WEEE regulations.
Figure 2-2 WEEE Label
The next section of this operating guide explains general operation of the MIZ-28.
When cleaning the display, care must be taken to avoid damage. For best results, use a lint free
cloth with minimal water or a mild detergent. Use only detergents or cleansers which are recom-
mended for use on plastics.
WA
RNIN
G
Do not open the MIZ-28 box. Any maintenance to be performed on internal components must be
carried out by trained maintenance personnel at Zetec only.
CAUTION
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
14/169
General Operation
MIZ-28 Checklist
Page 3-1
3.0 General Operation
This section provides an overview of MIZ-28 setup and basic operation. It includes a checklist to ensure that you
have received the complete MIZ-28 package, basic tutorials, and instructions on the following: initial setup of theMIZ-28, how to navigate through its menus and how to select the operating language. For detailed tutorials, see Sec-
tion 14.0.
3.1 MIZ-28 Checklist
When you unpack the MIZ-28, verify that the following items are present:
MIZ-28 Eddy Current Test Instrument
MIZ-28 Operating Guide
Bobbin probe adapter
USB 512 MB Flash drive
USB keyboard
Power cord Shipping case
If anything is missing, please contact Zetecs Customer Service department at 425-974-2700 immediately.
http://../TutorialDetailed.pdfhttp://../TutorialDetailed.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
15/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-2
3.2 Initial Setup of the MIZ-28
Because the most common use of the MIZ-28 will be for examinations of heat exchanger tubing with a pair of bobbin
coils, that is the type of setup described here. This ECT technique provides a rapid volumetric inspection of the
tube wall, assuring 100% coverage. Many other probe designs can be accommodated if special inspection criterianeed to be met. Up to eight (8) coil pairs can be used.
1. Turn the instrument on.
2. Connect the bobbin probes to the Coil ports on the back plane. Specific probe adaptors will be required (see
system drawings in Appendix A). NOTE: It is possible to connect and/or disconnect the probe adaptors without
turning the power off.
Figure 3-1 On/Off Switch & Coil Ports
Coil ports
On/off switch
http://../AppA.pdfhttp://../AppA.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
16/169
General Operation
Initial Setup of the MIZ-28
Page 3-3
Figure 3-2 Bobbin Coil Setup
3. Proceed to Section 3.8 for quick tutorials or Section 14.0 for detailed tutorials.
http://../TutorialDetailed.pdfhttp://../TutorialDetailed.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
17/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-4
3.3 Front Panel Buttons
All MIZ-28 controls are easy to access via front panel buttons.
If not already done, toggle the MIZ-28 power switch to the ON position. Below is a brief description of the MIZ-28
front panel control.
Figure 3-3 Front Panel Buttons
9
6
8
1
2
4
3
5
9
7
Control Knob. The rotating Control Knob enables you to
select menu items and adjust parameters. As you rotate the Con-
trol Knob, a highlight moves through the parameter fields to indi-
cate the current selection. The control button functions change
according to the selected parameter.
Coarse/Fine Adjustment. The Coarse/Fine Adjustment
button works in conjunction with the Control Knob. It switches
between coarse (fast) or fine (slow) adjustment of parameters.
Coarse/Fine LED Indicator. The red LED is lit when coarse
adjustment is selected, and is off when fine adjustment is
selected.
Select Button. When a parameter field is highlighted, press
the Select Button to choose it.
Help Button. Press the Help button to display information
about the current page selected. Press the Help button again to
exit the Help display.
Print Button. Press the Print button to send the current
screen image to the local USB printer or to a mass storage device
in .bmp file format. Print button operation is setup on the Utili-
ties page.
Menu Display/Hide Button. Any time during operation, you
can press the Menu Display button to display the Main Menu.
Press the Menu Display button again to hide the Main Menu. The
Menu organizes the MIZ-28s functions into related tasks for
instrument setup and testing.
Function Keys. Twelve function keys are available for oper-
ating the MIZ-28 features. Function key actions and labels change
according to the selected mode. The horizontal (primary) func-
tion keys select a feature for adjustment. The vertical (second-ary) function keys select a parameter value to change for the
selected feature.
Menu Button Labels. The display labels identify the func-
tion of each menu button and change according to the selected
mode.
1
2
3
4
5
6
7
8
9
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
18/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
19/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
20/169
General Operation
The MIZ-28 Main Menu
Page 3-7
3.5 The MIZ-28 Main Menu
The MIZ-28 Main Menu functions are described briefly below, with hyperlinked references to the Sections which
describe them in detail.
Figure 3-7 MIZ-28 Main Menu
Configuration Screen. A configuration is a set of MIZ-28 test parameters that are used for NDT inspection. These
parameters include sample rate, test frequencies, drive and channel. Use the Configuration screen to create and
maintain these configurations. See Section 4.0 for details.
Config List Screen. The Configuration List screen provides the ability to manage test configurations. It also lists all
saved configurations and allows you to select a configuration to edit. Configurations can be stored internally or on
an external storage device. The number of stored configurations is limited only by the size of the storage medium.
See Section 5.0 for details.
Test/Review Screen. Use the Test/Review screen to acquire and review data. See Section 6.0 for details.
Analysis Setup Screen. Use the Analysis Setup screen to set up curve parameters and mix data channels, and to
adjust the volt scale. See Section 7.0 for details.
Toolbox Screen. The Toolbox screen allows you to plot graphs of amplitude versus frequency (freq plot) and ampli-
tude versus drive (drive plot), as well as to view a list of materials. There are also prime frequencies for a given wall
thickness and standard depth of penetration at a given frequency. See Section 8.0 for details.
Utilities Screen. The Utilities screen provides access to Language Selection (English, German, Italian, Spanish, and
French), Speaker Control, Time/Date Input, IP Address Setting, Graphics Dump, and Remote Inputs. The MIZ-28 has
built-in USB and TCP/IP support. See Section 9.0 for details.
Diagnostics Screen. The Diagnostics screen contains diagnostic tools that can be activated to assure that the instru-
ment is within calibration specifications, as well as to troubleshoot inputs and outputs from other devices. The fea-
http://../Configurations.pdfhttp://../ConfigList.pdfhttp://../TestReview.pdfhttp://../AnalysisSetup.pdfhttp://../UsingToolbox.pdfhttp://../Utilities.pdfhttp://../AnalysisSetup.pdfhttp://../Utilities.pdfhttp://../UsingToolbox.pdfhttp://../TestReview.pdfhttp://../ConfigList.pdfhttp://../Configurations.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
21/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-8
tures in the Diagnostics menu are intended for use by Zetec calibration and repair or other personnel who are
trained in MIZ-28 maintenance, and are not intended for typical field use applications.See Section 10.0 for details.
Test Plans Screen. Use the Test Plans screen to import and utilize test plans. See Section 11.0 for details.
Mass Storage Setup Screen. The Mass Storage Setup screen allows you to specify how to record the data. Use it toset the recording directory, file format, recording destination, network IP address, mount point, and time zone.
Data can only be recorded when the Acquire function is active. See Section 12.0 for details.
Mass Storage Screen. The Mass Storage screen enables you to open, delete and move data files between internal
and mass storage directories. See Section 13.0 for details.
http://../Diagnostics.pdfhttp://../TestPlans.pdfhttp://../MassStorageSetup.pdfhttp://../MassStorage.pdfhttp://../MassStorage.pdfhttp://../MassStorageSetup.pdfhttp://../TestPlans.pdfhttp://../Diagnostics.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
22/169
General Operation
MIZ-28 Keyboard Operation
Page 3-9
3.6 MIZ-28 Keyboard Operation
To use a keyboard, simply plug it into one of the USB ports in the back of the MIZ-28.
Figure 3-8 Plugging in the Keyboard
When a keyboard is plugged in, an icon displays in the upper right corner of the screen.
Figure 3-9 Keyboard Icon
When the main MIZ-28 menu is displayed, use the Enter key to select the current highlighted selection.
When using the data cursor, the Up and Down arrow keys scroll the data, and the left and right arrow keys zoom the
charts in and out, respectively.
USB Ports
Keyboard icon
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
23/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-10
The Home key on the keyboard toggles between coarse and fine adjustment.
When adjusting the configuration, press the Enter key to perform the selected function (i.e., highlight a drive,
press the Enter key to make an adjustment and press Enter again to select the next one).
In Test/Review, when the SELECTQUADRANT/CHANNEL(F8) button is selected, use the left and right arrow keys tochange the quadrant, and use the up and down arrow keys to change the channel.
When vertical or horizontal position is selected, the up and down keys will adjust the vertical position, and the right
and left arrow keys will adjust the horizontal position.
In testing mode (acquiring data) with a Model III Controller connected, the following keys can be used.
Up arrow key runs the pusher forward at the fwd run speed.
Down arrow key runs the pusher reverse at the reverse speed.
Right arrow key runs the pusher forward at fwd jog speed.
Left arrow key runs the pusher reverse at the rev jog speed.
Enterkey starts recording and starts the pusher running reverse.
Enterkey again stops recording and stops the pusher.
Tube ID fields in testing mode (acquiring data).
+ key increments the current tube encode field.
key decrements the current tube encode field.
* key increments the field to adjust.
/ key decrements the field to adjust.
When using a keyboard, the following CNTRL functions will take you directly to a screen.
Ctrl C Configuration screen
Ctrl L Configuration list screen
Ctrl T Test / review screen
Ctrl A Analysis Setup
Ctrl B Toolbox screen
Ctrl U Utilities screen
Ctrl D Diagnostics screen
Ctrl P Test plans screen
Ctrl S Mass storage setup
Ctrl R Mass storage dir listing
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
24/169
General Operation
Selecting the Language
Page 3-11
3.7 Selecting the Language
The MIZ-28 has built-in multilingual support for English, German, Italian, Spanish, and French. It powers up in the
language selected during its last work session.
To change t he language select i on:
1. Press the Menu Display button.
2. Use the Control Knob to select (highlight) UTILITIES.
Figure 3-10 Select Utilities
3. Press the Select button to open the Utilities screen.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
25/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-12
4. Press the SELECT LANGUAGE (F1) button.
Figure 3-11 Select Language
5. Press the corresponding language (F8 to F12) button to select one of the following languages:
English (F8)
Deutsch (F9)
Italiano (F10)
Espaol (F11)
Franais (F12)
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
26/169
General Operation
Quick Tutorials
Page 3-13
3.8 Quick Tutorials
This Section contains tutorials on the various inspection types, and is intended as an overview for users who are
already familiar with Zetecs instruments. For detailed tutorials, see Section 14.0.
3.8.1 Bobbin Coil Exam
1. Connect bobbin hardware per the bobbin system drawing (see Drawing Number 10020312 in Appendix A for a
larger illustration that includes the parts list.)
Figure 3-12 Bobbin Hardware Setup
2. Verify the bobbin probe frequency, drive and coils.
In some testers a mode of operation (differential, absolute, driver/pick-up, etc.) might be
selected in the test configuration process. In the MIZ-28 this step is not required. The specific probe
adaptor selected will define whether the displayed information is differential or absolute once
the coil functionality is selected (On/Off). For the bobbin probe application, those channels
activated under Coil 1 will be operating in the self-comparison differential mode. Channels
activated under Coil 2 will be operating in the external-reference differential mode.
NOTE
http://../TutorialDetailed.pdfhttp://../AppA.pdfhttp://../AppA.pdfhttp://../TutorialDetailed.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
27/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-14
Figure 3-13 Bobbin Probe Configuration
3. Verify system sensitivity and signal rotation.
4. Select data storage mode or device: USB, internal hard drive, or network.
5. Store all system parameters.
6. Record summary, calibration runs, and data.
Figure 3-14 Bobbin Data
Note: For a step by step bobbin application process refer to Section 14.1.
http://../TutorialDetailed.pdfhttp://../TutorialDetailed.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
28/169
General Operation
Quick Tutorials
Page 3-15
3.8.2 MRPC Exam
1. Connect MRPC hardware per the MRPC system drawing (see Drawing Number 10020311 in Appendix A for a
larger illustration that includes the parts list.)
Figure 3-15 MRPC Hardware Setup
2. Verify MRPC 3C probe frequency, drive and coils.
Figure 3-16 MRPC Configuration
http://../AppA.pdfhttp://../AppA.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
29/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-16
3. Verify system sensitivity and signal rotation.
4. Select data storage mode or device: USB, internal hard drive, or network.
5. Store all system parameters.
6. Record summary, calibration runs, and data.
Figure 3-17 MRPC Data
Note: For a step by step MRPC application process refer to Section 14.2.
http://../TutorialDetailed.pdfhttp://../TutorialDetailed.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
30/169
General Operation
Quick Tutorials
Page 3-17
3.8.3 RFT Setup for Ferromagnetic Tubing
1. Connect RFT hardware per the RFT system drawing (see Drawing Number 10020313 in Appendix A for a larger
illustration that includes the parts list.)
Figure 3-18 RFT Hardware Setup
2. Verify RFT probe frequency, drive and coils.
Figure 3-19 RFT Configuration
http://../AppA.pdfhttp://../AppA.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
31/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 3-18
3. Verify system sensitivity and signal rotation.
4. Select data storage mode or device: USB, internal hard drive, or network.
5. Store all system parameters.
6. Record summary and calibration runs for the settings used in the calibration standard.
7. Based on responses from tubing in the heat exchanger it will be necessary to modify the set-up parameters
(predominantly frequency) to achieve maximum signal/noise ratio and flaw sensitivity in the tubes to be
inspected. This process may have to be performed more than once in a given heat exchanger.
Note: For a step-by-step RFT application process refer to Section 14.3
http://../TutorialDetailed.pdfhttp://../TutorialDetailed.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
32/169
General Operation
Quick Tutorials
Page 3-19
3.8.4 HVAC
1. Connect HVAC hardware per the HVAC system drawing (see Drawing Number 10020340 in Appendix A for a larger
illustration that includes the parts list.)
Figure 3-20 HVAC Hardware Setup
2. Verify AC3 probe frequency, drive and coils.
Figure 3-21 AC3 Configuration
http://../AppA.pdfhttp://../AppA.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
33/169
3. Verify system sensitivity and signal rotation.
4. Select data storage mode or device: USB, internal hard drive, or network.
5. Store all system parameters.
6. Record summary, calibration runs, and data.
Figure 3-22 HVAC Data
Note: For a step by step AC3 application process refer to Section 14.4
The next section of this operating guide describes the Configuration screen.
http://../TutorialDetailed.pdfhttp://../TutorialDetailed.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
34/169
Configuration
Page 4-1
4.0 Configuration
A configuration is a set of MIZ-28 test parameters, including sample rate, test frequencies, drive and channel, used
for NDT inspection. Use the Configuration screen to create and maintain these configurations.
In addition to any custom configurations you might create, the MIZ-28 also includes four canned configurations: Bob-
bin Coil, MRPC, RFT and HVAC. These are accessed via the Config List screen (Section 5.0).
To open the Configuration screen, follow these steps:
1. Press the Menu Display button.
2. Use the Control Knob to select (highlight) CONFIGURATION.
3. Press the Select button.
Figure 4-1 Open Configuration Screen
The standard configurations are not write-protected. Use care when working with them.
WARNIN
G
http://../ConfigList.pdfhttp://../ConfigList.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
35/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 4-2
Figure 4-2 Configurat ion Screen
The default Configuration screen has the following primary function buttons across the bottom of the screen:
Config Name (F1)
Config Options (F2)
Sample Rate (F3)
Frequency Select (F4)
Drive Select (F5)
Channel Select (F6)
Config Storage (F7)
When you press these buttons, the F8 F12 secondary function buttons change, as illustrated in Figure 4-3.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
36/169
Configuration
Page 4-3
Figure 4-3 Configuration Menu Structure
The rest of this section describes each menu function in the same order as shown above.
CONFIG NAME (F1)
CONFIG OPTIONS (F2)
SAMPLE RATE (F3)
FREQUENCY SELECT (F4)
DRIVE SELECT (F5)
CHANNEL SELECT (F6)
CONFIG STORAGE (F7)
Right arrow (F8)
Left arrow (F9)
SPACE (F10)
BACK SPACE (F11)
CLEAR (F12)
RETURN (F7)
SELECT (F8)
ADJUST (F9)
SAMPLE RATE (F8)
SELECT FILTER (F9)
FREQUENCY NUMBER (F8)
ADJUST (F9)
DRIVE (F8)
ADJUST (F9)
SET ALL (F10)
COIL NUMBER (F8)
FIELD (F9)
TOGGLE ON/OFF (F10)
SET ALL (F11)
SAVE CONFIG (F8)
CONFIG LIST (F9)
Secondary Function ButtonsPrimary Function Buttons
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
37/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 4-4
4.1 Naming the Configuration
You can use up to 30 characters for the configuration name (letters A through Z and numbers 0 through 9).
If you are using a USB keyboard, simply type in the desired configuration name. Otherwise, follow the steps below.
1. Press the CONFIGNAME (F1) button to open the character table.
2. To entirely erase the current name (displayed in the NAME field), press the CLEAR(F12) button.
3. To edit the current name, use the RIGHT ARROW (F8) and LEFT ARROW (F9) buttons to move the cursor withinthe name field, the SPACE (F10) button to add spaces, and the BACK SPACE (F11)button to delete spaces andcharacters.
4. To enter a new configuration name, rotate the Control Knob to highlight a character in the table.
5. Press the Select button to input the character.
6. Repeat Steps 4 and 5 until the new name is spelled out.
7. Press the RETURN (F7) button to exit the character table.
Figure 4-4 Conf ig Name
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
38/169
Configuration
Selecting Configuration Options
Page 4-5
4.2 Selecting Configuration Options
Press the CONFIG OPTIONS (F2) button to set the variables described below.
Figure 4-5 Conf ig Options
4.2.1 Config Type
The MIZ-28 supports both multiplexed and simultaneous injection testing.
Mul t ip lexed
It is recommended you use the multiplexed mode for the following probes:
Bobbin (differential and absolute with external reference)
MRPC
Array probes (that do not require external multiplexing hardware)
AC3 probes
The MIZ-28 supports four test frequencies in multiplexed mode.
Simul t aneous Inject ion
In simultaneous injection mode, the number of available frequencies depends on the number of coils being used andthe number of boards present in the unit. With the two analog board MIZ-28, you can run a two coil test at four fre-
quencies, a four coil test at two frequencies, or an eight coil test at one frequency.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
39/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 4-6
Figure 4-6 Conf ig Type
To select a config type:
1. Press the CONFIG OPTIONS (F2) button.
2. Use the Control Knob to scroll to CONFIG TYPE.
3. Press the ADJUST (F9) button and use the Control Knob to select MULTIPLEXED or SIMULTANEOUS.
4. Press the SELECT (F8) button.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
40/169
Configuration
Selecting Configuration Options
Page 4-7
4.2.2 External Encoder
Selecting External Encoder allows you to set the number of encoders. If you enable the encoder (up to two), you
must also configure the encoder.
Figure 4-7 External Encoder
To configure an encoder:
1. Use the Control Knob to scroll through the encoder configuration options.
2. Press the ADJUST (F9) button to activate the highlighted option.
3. Use the Control Knob to scroll through the selected option configurations.
4. Press the SELECT (F8) button to confirm the new option configuration.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
41/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 4-8
4.3 Adjusting the Sample Rate
You can adjust the sample rate from 1 up to a maximum of 16,000 samples per second. The maximum sample rate
allowed depends on the number of test frequencies in use and the value of the test frequencies.
To adjust the sample rate:
1. Press the SAMPLE RATE (F3) button to enable the function.
2. Use the Control Knob to select the samples per second.
3. In Simultaneous mode only, press the SELECT FILTER(F9) button to set the filter rate.
Figure 4-8 Sample Rate
4.3.1 Sample Rates in Multiplexed Mode
Four different filters 4K, 8K, 12K and 16K are available for use in the signal processing circuitry when operating
in multiplexed mode. The 4K filter produces the best signal-to-noise, though at the cost of sample rate. The 16K fil-
ter allows faster sampling rates at the cost of signal-to-noise. Switching from the 4K filter to the 8K filter, you can
expect to lose 3dB of signal-to-noise. Switching from 8K to 12K or 12K to 16K filter, you can expect to lose 1.5dB of
signal-to-noise. The instrument automatically selects the best possible filter for the configured parameters and dis-
plays it next to the maximum sample readout. As you increase or decrease the sample rate, the filter setting will
change.
4.3.2 Sample Rates in Simultaneous Mode
Four filters are available in simultaneous mode: 240Hz pass band, 400Hz pass band, 800Hz pass band and 1600Hz
pass band. The trade-off when using these filters is signal-to-noise versus pull speed. Given the general rule of 10Hz
of pass band per inch pulling speed, the 240Hz filter has about 24ips max pull speed and the 400Hz has about 40ips
max pull speed. As you start decreasing test frequencies, the max pull speed value displayed to the right of the
sample rate decreases. You should expect the 240Hz filter to have a 2.5dB gain in signal to noise.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
42/169
Configuration
Setting Test Frequency Values
Page 4-9
4.4 Setting Test Frequency Values
The MIZ-28 supports up to four test frequencies from 5 Hz to 10 MHz. To eliminate cross talk between frequencies in
simultaneous mode, a minimum frequency separation rule must be followed. If there is not enough separation
between test frequencies, the frequencies in violation will turn red and the message, Not enough frequency sepa-ration will be displayed at the bottom of the screen.
A special filter design is implemented in the MIZ-28 to improve performance at low test frequencies. This is to
improve RTF testing performance and signal noise. This filter will automatically be selected if the configuration
allows for it. When the filter is applied, you will see an increase in the max pull speed (ips) value on the configura-
tion screen and an LF (low frequency filter) will appear. This filter is only applicable in the SI mode.
You can select up to four instrument test frequencies, depending on the test configuration.
To select a test frequency:
1. Press the FREQUENCY SELECT(F4) button.
2. Use the Control Knob to select the frequency you want to change.
3. Press the ADJUST (F9) button and use the Control Knob to adjust the frequency value.
Figure 4-9 Frequency Select
For help on frequency selection, see Section 8.3, Viewing the Material List .
NOTE
http://../UsingToolbox.pdfhttp://../UsingToolbox.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
43/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 4-10
4.5 Selecting a Drive Setting
In multiplexed mode, each drive can be set from 1.0 to 20.0 volts. The drive is adjustable in 0.1 volt increments.
In simultaneous mode, all four drives must add up to a maximum of 20 volts. If the sum of all drives is greater than
20 volts, the drives will turn red and the message Drive greater than 20Vpp will appear at the bottom of the
screen.
To select a drive setting:
1. Press the DRIVE SELECT (F5) button to enable the function.
2. Press the DRIVE (F8) button, and then use the Control Knob to select a drive.
3. Press the ADJUST (F9) button to increase or decrease the drive value.
4. Press the SET ALL (F10) button to set all the drives to the same value.
Figure 4-10 Drive Select
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
44/169
Configuration
Selecting a Channel
Page 4-11
4.6 Selecting a Channel
Each channel is described by a frequency and coil combination. You can turn a channel ON or OFF. Once a coil has
been activated, a balance number is associated with it. The balance assignment is by coil. There are two separate
balances available in the MIZ-28.
In multiplexed mode, you can turn on two coils per timeslot per board for a total of four channels per timeslot.
In simultaneous mode, the number of available coils depends on the number of test frequencies used. The two ana-
log board MIZ-28 is capable of running a two coil test at four frequencies, a four coil test at two frequencies, or an
eight coil test at one frequency.
To select a channel:
1. Press the CHANNEL SELECT (F6) button.
2. Press the COIL NUMBER(F8) button, and then use the Control Knob to select a coil number.
3. Press the FIELD (F9) button to set the field.
4. When the field is in one of the frequency lines, press the TOGGLE ON OFF (F10) button to toggle the channels
ON or OFF. In simultaneous mode all channels for that coil will be turned ON or OFF.
5. When the cursor is scrolled down to the balance number line, the F10 button label changes to BALANCE NUM-
BER. Press F10 and use the Control Knob to assign a balance number to the channels configured to that coil.
Figure 4-11 Channel Select
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
45/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
46/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
47/169
Config List
Page 5-1
5.0 Config List
The Configuration List screen enables you to manage test configurations. It also lists all canned and created config-
urations and allows you to select a configuration to edit. Configurations can be stored internally or on an external
storage device. The number of stored configurations is limited only by the size of the storage medium.
To open the Configuration List screen:
1. Press the Menu Display button.
2. Use the Control Knob to select (highlight) CONFIG LIST.
3. Press the Select button to open the Configuration List screen. (To select the Configuration List from the Config-
uration screen, press CONFIG STORAGE (F7) > CONFIG LIST (F9).)
Figure 5-1 Open Config List Screen
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
48/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
49/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
50/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
51/169
Config List
Viewing the List Options
Page 5-5
5.1.4 Copying To and From Internal and Mass Storage
Configurations can be copied back and forth between internal storage and mass storage. If a configuration of the
same name exists in the location that you are copying the configuration to, you will be prompted as to whether or
not you want to overwrite the existing configuration.
Copyi ng from Inter nal t o Mass St ora ge
If a USB mass storage device is attached to the instrument, the COPY TO MASS STORAGE(F11) button will be active.
(If there is no USB mass storage device attached, this button label will be blank.)
1. Press the COPY TO MASS STORAGE (F11) button.
2. Highlight the configuration you want to copy.
Figure 5-6 Copy to Mass Storage
3. Press the Select button.
4. The message Configuration copied to mass storage will be displayed at the bottom of the screen.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
52/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
53/169
Config List
Viewing the List Options
Page 5-7
5. Highlight the configuration you want to copy.
6. Press the Select button.
7. The message Configuration copied to internal storage will be displayed at the bottom of the screen.
5.1.5 Removing a Configuration
To remove a configuration:
1. Press the REMOVE CONFIG (F12) button to enable configuration selection.
2. Rotate the Control Knob to select (highlight in green) the Configuration in the list.
Figure 5-9 Remove Configurat ion
3. Press the Select button to remove the configuration.
The configuration is deleted immediately, so be certain you want to remove it before pressing the
Select button.
CAUTI
ON
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
54/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 5-8
5.2 Changing the Directory
If you have a USB mass storage device attached to the instrument, the CHANGE DIRECTORY (F2) button appears.
This button enables you to switch between configuration lists stored internally or on the mass storage device.
1. Press the CHANGE DIRECTORY (F2) button.
2. Select the INTERNAL (F8) or MASS STORAGE (F9) button.
Figure 5-10 Change Directory
Change Directory button
When the current directory is Mass Storage, the F11 key of the List Options menu changes to Copy
to Internal.
NOTE
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
55/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
56/169
Test / Review
Page 6-1
6.0 Test / Review
Use the Test/Review screen to acquire and review data.
To open the Test/Review screen, follow these steps:
1. Press the Menu Display button.
2. Use the Control Knob to select (highlight) TEST/REVIEW.
3. Press the Select button.
The menu functions change based on whether you are in Testing Mode (Section 6.1) or Review Mode (Section 6.2).
Figure 6-1 Open Test/Review Screen
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
57/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-2
6.1 Testing Mode
The Testing Mode screen contains all the functionality to collect eddy current test data. Semi-automatic data col-
lection can be achieved using the test plans and a Model III Controller. When you press the START ACQUIRE (F7) but-
ton, you will be in testing mode. To toggle between menus 1 and 2, push the MORE button (F1).
Figure 6-2 Test/Review Menu 1 StructureTest Mode
DATA CONTROL (F2)
MANUAL TUBE INFO (F4)
MESSAGE SUMMARY (F5)
When you press this button, the function
keys to the right appear and the overall
menu changes. Use the keys to the right
to work with the message window.
SELECT QUADRANT CHANNEL (F8)
CHART RATE (F9)
CLEAR (F10)
PERSIST TIME OFF (F11)
BALANCE (F12)
SELECT ID FIELD (F8)
ADJUST (F9)
TOGGLE FIELD(F10)
ENABLE STD RUN (F11)
DISPLAY NULL (F12)
SCROLL (F8)
WRITE MESSAGE (F9)
CHANGE MESSAGE (F10)
STOP ACQUIRE (F7)
When you press this button, Acquisition stops and
the menu changes to that shown in Figure 6-12
START RECORDING (F6)
When you press this button, recording starts
and the button label changes to Stop Recording.
.
ENCODER (F3)
If you have only one encoder, this button dis-
plays ZERO ENCODER, and no secondary func-
tion buttons are displayed. Depending on
how many encoders you have, there could be
less secondary functions.
ZERO ENCODER 1 (F8)
ZERO ENCODER 2 (F9)
Secondary Function ButtonsPrimary Function Buttons
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
58/169
Test / Review
Testing Mode
Page 6-3
Figure 6-3 Test/Review Menu 2 StructureTest Mode
ADJUST DATA(F4)
RUN TEST PLAN (F5)
SELECT QUADRANT (F8)
SPAN (F9)
CHANNEL ROTATION (F10)
VERTICAL ADJUST (F11)
HORIZ ADJUST (F12)
SCROLL (F8)
GO TO TOP (F9)
FIND NEXT UNTESTED (F10)
MARK UNMARK (F11)
STOP ACQUIRE (F7)
When you press this button, Acquisition stops and
the menu changes to that shown in Figure 6-12
START RECORDING (F6)
When you press this button, recording startsand the button label changes to Stop Recording.
.
TUBE END DETECTION (F2)This button will only be present if a Model III
Controller is connected to the RS-232 con-
nector on the back panel of the MIZ-28. If a
Model III is connected, a pusher icon will dis-
play in the top right corner of the display.
SET THRESHOLD (F8)Tube End Detection ENABLED (F9)
THRESHOLD DELAY (F10)
Auto Retract ENABLED (F11)
Secondary Function ButtonsPrimary Function Buttons
ADJUST TOLLERANCE(F3) SELECT QUADRANT (F8)
TOGGLE (F9)
VERTICAL ADJUST (F10)
HORIZ ADJUST (F11)
HORZ (F12)
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
59/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-4
6.1.1 Data Control
Press the DATA CONTROL (F2) button to change the channel and quadrant, chart rate, and persist time. From this
submenu, you can also manually clear and balance the data display.
Figure 6-4 Testing Mode Screen
Select Quad ra nt / Channel
1. Press the SELECT QUADRANT CHANNEL (F8) button. When the word QUADRANT is capitalized, you can use theControl Knob to scroll through the four X-Y displays and strip charts.
2. Press F8 again and CHANNEL is capitalized. Now the Control Knob is enabled to select a channel for the high-lighted X-Y or strip chart display.
Chart Rat e
Press the CHART RATE (F9) button, and then use the Control Knob to adjust the rate at which the strip charts scroll.
The number that displays on the Chart Rate button indicates how many slices of data will be plotted to the chart
per second.
Clear
Press the CLEAR(F10) button to clear the data off the display.
Persist Tim e
Press the PERSIST TIME (F11) button to adjust the amount of time that will elapse between automatic screen clears.
Adjustment ranges from OFF to 10 seconds in 100 millisecond internals.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
60/169
Test / Review
Testing Mode
Page 6-5
Balance
Press the BALANCE (F12) button to electronically balance the eddy current signals. This function will take the cur-
rent data values and shift them to a value of zero. Two balance functions are available if more than one balance is
selected in the configuration.
6.1.2 Encoder
The ENCODER(F2) button will only be present if two encoders have been configured.
Figure 6-5 Encoder Screen
Zeroing an Encoder
1. Press the ENCODER(F3) button.
2. Press the ZERO ENCODER(F8 or F9) for the encoder you would like to zero.
If there is no encoder is configured, the ENCODER button will not display. If only one encoder is
configured, the ENCODER button will display ZERO ENCODER.
NOTE
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
61/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-6
6.1.3 Manual Tube Info
If test plans are not being used, it is necessary to manually encode the data files. Press the MANUAL TUBE INFO (F3)
button to access the tube info control buttons.
Figure 6-6 Manual Tube Info Screen
Select ing an ID Fiel d
Press the SELECT ID FIELD (F8) button to enable the Control Knob to scroll through the tube identification fields.
The selected field is highlighted in blue.
Adjust ing an ID Field
Press the ADJUST (F9) button to adjust the selected field. The selected field is highlighted in green when the
adjustment is enabled.
Toggling an ID Fiel d
Press the TOGGLE FIELD (F10) button to toggle the ID field between alpha numeric and numeric.
Enable/ Disable St andar d Run
Press the ENABLE STD RUN (F11) button to change the row and column values to the standard run values (typically
999). The values are operator selectable using the SELECT and ADJUST buttons above. When standard run is
Pressing the Selectbutton in this mode will toggle the selection between select and adjust. After
making an adjustment, pressing the Adjust(F9) button again will change the row or column encode
to increment or decrement.NOTE
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
62/169
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
63/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-8
6.1.5 User Message
In the Message Summary screen (Section 6.1.4), press the USER MESSAGE (F4) button to enable a blank text screen.
This screen is used for message input to the data directory for later use when analyzing data.
Figure 6-8 User Message Screen
Scrol l Message
Press the SCROLL (F8) button to enable the Control Knob to scroll the cursor through the message.
Wri t e Message
Press the WRITE MESSAGE (F9) button to write the message to the data directory.
Clea r Message
Press the CLEAR MESSAGE (F10) button to clear the current message.
Sav e Message
Press the SAVE THIS MESSAGE (F11) button to save the current message form.
Recal l Message
Press the RECALL MESSAGE (F12) button to recall the saved message form.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
64/169
Test / Review
Testing Mode
Page 6-9
6.1.6 Summary Form
In the Message Summary screen (Section 6.1.4), press the SUMMARY FORM (F5) button to access the test summary
form. This default form includes parameters such as owner, plant, unit, and a configuration summary of the current
parameters. To fill out the summary form, you must attach a USB keyboard.
Press the Tab button to move the cursor to the next valid entry location.
Press the Enter button to move the cursor to the beginning of the next line.
Figure 6-9 Summary Form
Scroll
Press the SCROLL (F8) button to enable the Control Knob to scroll the cursor through the Summary Form.
Wri t e Summary
Press the WRITE SUMMARY (F9) button to write the current form.
Defaul t Summary
To restore the summary form default parameters, press the DEFAULT SUMMARY (F10) button.
Save Summar y For m
Press the SAVE THIS FORM (F11) button to save any changes made to the summary form.
When the file format is set to Eddynet in the mass storage setup, writing a summary form will
automatically create a new data directory (cal group) for recording.
NOTE
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
65/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-10
Recall Summary Form
Press the RECALL FORM (F12) button to recall the summary form that was last saved. This function can be useful for
clearing the specific operator data out of the form.
6.1.7 Start / Stop Recording
Press the START RECORDING (F6) button to start recording a data file to the current data directory. The current
recording path and file name is shown under the tube encode.
When the START RECORDING (F6) buttonis pressed, the F6 button label changes to STOP RECORDING and a list of
the last 5 files recorded displays in the lower right corner of the screen.
6.1.8 Stop Acquisition
Press the STOP ACQUISITION (F8) button to stop the data acquisition and switch to data review mode (Section 6.2).
6.1.9 Tube End Detection
The TUBE END DETECTION (F2) button on the second page of buttons, will only be present if a Model III Controller
is connected to the RS-232 connector on the back panel of the MIZ-28. When a Model III is connected, a pusher icon
is displayed in the top right corner of the display.
The tube end detection submenu allows you to enable or disable tube end detection, set a tube end threshold delay
(which allows for entry into the tube without detecting the near side tube end), enable or disable the auto retract
function and, when in review mode, set the tube end threshold.
Figure 6-10 Tube End Detection Screen
Set t ing t he Tube End Thr eshold
1. Acquire the data from a tube end and press the STOP ACQUIRE button.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
66/169
Test / Review
Testing Mode
Page 6-11
2. Scroll the tube end signal to the cursor and press the SET THRESHOLD button. This will take the peak measure-ment of the tube end and save it to NVRAM. A message at the bottom of the page will display saying, Thresh-
old has been set.
Tube End Detect ion
Press the TUBE END DETECTION (F2) button to enable or disable the function.
When tube end detection is enabled, every time the pusher starts running the data will be monitored for a signal
that is greater than 80% of the set threshold limit (tube end signal). When a signal of great enough amplitude is
detected, the instrument will turn off the pusher. The exception is when Standard Run is enabled. In this case, the
instrument will not monitor for the threshold signal.
Thr eshold Delay
Press the THRESHOLD DELAY (F10) button, and then use the Control Knob to change the delay time in milliseconds
(mS).
The Threshold Delay is the amount of time that the instrument will wait to begin processing data after a pusher
move has been initiated. This delay allows the probe to enter the tube without triggering a threshold event.
Auto Retr act
Press the AUTO RETRACT (F11) button to enable or disable the function. When auto retract is enabled and a thresh-
old event is encountered in the forward direction, recording will be started and the pusher will automatically start
running in the reverse direction. The next threshold event will stop the pusher and stop recording.
In addition, when auto retract is enabled, the pusher will start running in the reverse direction whenever the START
RECORDING button is pressed. Any time the pusher is running and the STOP RECORDING button is pressed, the
pusher will be stopped.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
67/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-12
6.1.10 Adjust Tolerance
Press the ADJUST TOLERANCE (F3) button on the second page of buttons to change configured channels, toggle each
alarm on or off, adjust the vertical and horizontal size of the alarm, and adjust the vertical to horizontal scale of
the displayed data and alarms
Figure 6-11 Adjust Tolerance Screen
Select Quadr ant Channel
Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob
to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now you can
use the Control Knob to select a channel for the highlighted X-Y or strip chart display.
Toggle the alar m on or of f
Press the TOGGLE button (F9) to turn the alarm on or off.
Adjust t he height of t he alar m area
1. Press the VERTICAL ADJUST (F10) button to activate the height of the alarm area.
2. Use the control knob to adjust the height of the alarm area.
Adjus t t he wid th of t he alarm area
1. Press the HORIZ ADJUST (F11) button to activate the width of the alarm area.
2. User the control knob to adjust the width of the alarm area.
Adjusting the scaling (HORIZ/#) does not effect the analysis calculations.
NOTE
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
68/169
Test / Review
Testing Mode
Page 6-13
Adjust t he scale of t he data
1. Press the HORIZ (F12) to scale the flaw data.
2. Use the control knob to adjust the scale of the flaw data.
6.1.11 Adjust DataRefer to Adjust Data on page 6-17.
6.1.12 Run Test Plan
The RUN TEST PLAN (F4) button is only present if a test plan has been imported and selected. See Section 11.0 for
more information.
6.1.13 Start / Stop Recording
Press the START RECORDING (F6) button to start recording a data file to the current data directory. The current
recording path and file name is shown under the tube encode.
When the START RECORDING (F6) buttonis pressed, the F6 button label changes to STOP RECORDING and a list ofthe last 5 files recorded displays in the lower right corner of the screen.
6.1.14 Stop Acquisition
Press the STOP ACQUISITION (F8) button to stop the data acquisition and switch to data review mode (Section 6.2).
http://../TestPlans.pdfhttp://../TestPlans.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
69/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-14
6.2 Review Mode
The Review mode provides all the functionality to review, adjust and filter data, as well as to check pusher speed
and set the tube end threshold. To toggle between menus 1 and 2, push the MORE button (F1).
Figure 6-12 Test/Review Menu 1 StructureReview Mode
CURSOR CONTROL (F2)
ADJUST DATA (F4)
MEASURE (F5)
START ACQUIRE (F7)
When you press this button, Acquisition
begins, and the menu structure changes.
See Figure 6-2.
SELECT QUADRANT CHANNEL (F8)
CHART ZOOM (F9)
DATA POSITION (F10)
CURSOR SIZE (F11)
DISPLAY NULL(F12)
SELECT QUADRANT CHANNEL (F8)
SPAN (F9)
CHANNEL ROTATION (F10)
VERTICAL ADJUST(F11)
HORIZ ADJUST (F12)
SELECT QUADRANT CHANNEL (F8)
CHANNEL ROTATION (F9)
PEAK TO PEAK (F10)
VERTICAL MAX (F11)
MAX RATE (F12)
Secondary Function ButtonsPrimary Function Buttons
ADJUST TOLERANCE (F3) SELECT QUADRANT CHANNEL (F8)
TOGGEL (F9)VERTICAL ADJUST (F10)
HORIZ ADJUST (F11)
HORIZ (F12)
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
70/169
Test / Review
Review Mode
Page 6-15
Figure 6-13 Test/Review Menu 2 StructureReview Mode
SPEED CHECK(F2)
FILTER SETUP(F4)
START ACQUIRE (F7)
When you press this button, Acquisition
begins, and the menu structure changes.
See Figure 6-2.
SELECT QUADRANT CHANNEL (F8)CHART ZOOM (F9)
DATA POSITION (F10)
CURSOR SIZE (F11)
ADJUST DISTANCE (F12)
SELECT QUADRANT CHANNEL (F8)
FILTER TYPE OFF (F9)
HIGH FREQ CUTOFF (F10)
DIFF DELTA (F11)
SET ALL (F12)
Secondary Function ButtonsPrimary Function Buttons
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
71/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-16
6.2.1 Data Cursor Control
Press the CURSOR CONTROL (F2) button to access the cursor control options.
Figure 6-14 Data Cursor Screen
Select Quad ra nt / Channel
Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knobto scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now you can
use the Control Knob to select a channel for the highlighted X-Y or strip chart display.
Chart Zoom
Press the CHART ZOOM (F2) button to expand the data in the strip chart. The chart will zoom from the cursor loca-
tion. At a chart zoom of 100%, there is one data point per strip chart line. If you cannot zoom in as far as you would
like, reduce your cursor size.
Dat a Posit ion
Press the DATA POSITION (F10) button to scroll the data through the cursor.
Cursor Siz e
Press the CURSOR SIZE (F11) button to change the amount of data that is displayed in the four X-Y displays.
Display Null
Press the DISPLAY NULL (F12) button to center the data back in the X-Y displays.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
72/169
Test / Review
Review Mode
Page 6-17
6.2.2 Adjust Tolerance
See Adjust Tolerance on page 6-12.
6.2.3 Adjust DataPress the ADJUST DATA (F4) button to access the data adjustment control menu. Here you have the ability to select
displayed channels, adjust the size of the displayed signal, rotate the channels data, and adjust the datas vertical
and horizontal position. For quick adjustment of data, zoom the strip chart and only cursor the data of interest.
Figure 6-15 Adjust Data Screen
Select Quad ra nt / Channel
Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob
to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now the
Control Knob is enabled to select a channel for the highlighted X-Y or strip chart display.
Span
Press the SPAN (F9) button and rotate the Control Knob to adjust the size of the displayed signal for the selected X-
Y display or strip chart.
Channel Rot at ion
Press the CHANNEL ROTATION (F10) button, and then use the Control Knob to rotate the selected channels rotation
from 0 to 360 degrees.
Vert ical Adj ust ment
Pressthe VERTICAL ADJUST (F11) button, and then rotate the Control Knob to adjust the vertical data position of
the current X-Y display. The vertical position has no effect when a strip chart is highlighted.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
73/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-18
Horizontal Adj ust ment
Pressthe HORIZONTAL ADJUST (F11) button, and then rotate the Control Knob to adjust the horizontal data posi-
tion of the current X-Y display. The horizontal position has no effect when a strip chart is highlighted.
6.2.4 Measure
Press the MEASURE (F5) button to access the measure controls. There are three measurement types in the MIZ-28:
peak to peak, vertical max, and max rate. For quick adjustment and measurement of data, zoom the strip chart and
only cursor the data of interest.
Figure 6-16 Measure Screen
Select Quad ra nt / Channel
Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob
to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now the
Control Knob is enabled to select a channel for the highlighted X-Y or strip chart display.
Channel Rot at ion
Press the CHANNEL ROTATION (F9) button to rotate the current channels data to achieve the desired angle of the
flaw response.
Peak t o Peak
Press the PEAK TO PEAK(F10) button to show all the channels displayed peak to peak measurement. This measure-
ment determines the maximum signal amplitude in any phase. Measurement displays in A to D counts and angle.
Vert ical Maximum
Press the VERTICAL MAX (F11) to display all the channels vertical maximum measurement. Vertical maximum mea-
surement determines the maximum signal amplitude in the vertical direction only, and displays in A to D counts.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
74/169
Test / Review
Review Mode
Page 6-19
Maximum Rat e
Press the MAX RATE (F12) button to display the maximum rate measurement. The maximum rate measurement
determines the greatest rate of change in the data signal. This measurement displays the peak to peak A to D mea-
surement and the maximum rate angle.
6.2.5 Probe Speed Check
The probe speed check controls are the same as the cursor controls with the addition of a button labelledADJUST
SPEED. To check the speed of the probe when the data was collected, follow the steps below.
1. Press the SPEED CHECK(F2) button on the second button page.
2. Position a known landmark (typically a tube support) in the center of the data cursor in the strip chart.
3. Adjust the cursor size to place one of the outer cursor lines on another known landmark. The distance and
speed are displayed at the bottom of the screen.
4. Press the ADJUST DISTANCE (F12) button, and then use the Control Knob to adjust the distance value to the
known distance between the two landmarks of the cursor location.
The speed value will update to reflect the rate of travel at the time of acquisition.
Figure 6-17 Speed Check Screen
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
75/169
6.2.6 Tube End Detection
The TUBE END DETECTION (F3) button on the second page of buttons will be present only if a Model III Controller is
connected to the RS-232 connector on the back panel of the MIZ-28. When a Model III is connected, a pusher icon is
displayed in the top right corner of the screen. This menu is the same as previously discussed in the test mode sec-
tion with one exception in review mode a new button, SET THRESHOLD, is available.
Figure 6-18 Tube End Detection
Set t ing t he Tube End Thr eshold
Acquire the data from a tube end, and then press the STOP ACQUIRE button. Scroll the tube end signal to the cursor
and press the SET THRESHOLD (F8) button. This will take the peak measurement of the tube end and save it to
NVRAM. A message at the bottom of the page will appear saying, Threshold has been set.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
76/169
Test / Review
Review Mode
Page 6-21
6.2.7 Filter Setup
Filter adjustments do not affect the analog probe signal. They are digital filters that optimize the presentation and
aid in signal analysis. This allows you to experimentally change the filters in review mode without the need to re-
acquire the data. There are two types of filters available with the MIZ-28: a first order differential filter and a low
pass filter. The differential filter is adjustable from 2 to 40 points; the low pass filter is adjustable from 0 Hz to
about half of the current sample rate. When a channel has a filter associated with it, you will see the words DIFF, LP,
or LP+DIFF on the end of the X-Y channel information.
Figure 6-19 Filter Setup Screen
Select Quad ra nt / Channel
Press the SELECT QUADRANT channel (F8) button. When the word QUADRANT is capitalized, use the Control Knob
to scroll through the four X-Y displays and strip charts. Press F8 again and CHANNEL will be capitalized. Now the
Control Knob is enabled to select a channel for the highlighted X-Y or strip chart display.
Fi l te r Type
Press the FILTER TYPE (F9) button and rotate the Control Knob to change the filter type associated with the
selected channel. The options are off, low pass, differential, or a combination of low pass and differential. The
low pass filter is a 41 coefficient digital filter with sharp cutoff characteristics. Its useful in suppressing a variety ofunwanted signals including electrical interference. The first order differential filter displays the difference between
the present data point and a previous data point. The closer the data points are together in time, the more high-
pass action the filter creates. The process is sometimes known as continuous balance, because the displayed signal
has been compared to some recent value, rather than a fixed point. This filter is useful when the probe is scanned
across dissimilar materials, as it tends to keep the signal on screen. The differential filter is sensitive to the direc-
tion of data change. Rising edges will show as positive signals, and falling edges will show as negative signals. For
this reason, a single unfiltered pulse, which deflects upwards and then returns to null, will appear filtered as a pos-
itive and then a negative signal.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
77/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 6-22
High Frequency Cut off
Press the HIGH FREQ CUTOFF (F10) button and rotate the Control Knob to adjust the cutoff frequency of the low
pass filter for the selected channel. The range is 0 Hz to about half of the current sample rate.
Di f ferent ia l Del ta
Press the DIFF DELTA (F11) button to adjust the number of data points that are applied to the differential filter.
Set All
Press the SET ALL (F12) button to set all channels configured to the current channels filter settings.
6.2.8 Start Acquire
When you press the START ACQUIRE (F7) button, Acquisition begins and the screen changes to Test Mode (Section
6.1).
The next section of this operating guide describes the Analysis Setup screen.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
78/169
Analysis Setup
Page 7-1
7.0 Analysis Setup
To open the Analysis Setup screen, follow these steps:
1. Press the Menu Display button.
2. Use the Control Knob to select (highlight) ANALYSIS SETUP.
3. Press the Select button.
Figure 7-1 Open Analysis Setup Screen
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
79/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 7-2
Figure 7-2 Analysis Setup Menu Structure
CURVE SETUP (F1)
MEASURE (F3)
ADJUST CURVE (F4)
CURVE TABLE (F5)
SELECT CHANNEL (F8)
CURVE TYPE (F9)
NUM ID POINTS (F10)
NUM OD POINTS (F11)
SELECT CHANNEL (F8)CHANNEL ROTATION (F9)
PEAK TO PEAK (F10)
VOLTAGE MULT (F11)
MAX RATE (F12)
SELECT POINT (F8)
ADJUST PERCENT (F9)
DATA POSITION (F10)
SET THIS CHANNEL (F11)
SET ALL FREQS (F12)
SELECT CHANNEL (F8)
SETUP FILE (F7)
DATA MIXING (F6) SELECT NUM MIXES (F8)
MIX 1 CHAN 1 (F9)
MIX 1 CHAN 2 (F10)
PERFORM MIX (F12)
CURSOR CONTROL (F2)
Secondary Function ButtonsPrimary Function Buttons
SAVE SETUP (F8)
RECALL SETUP (F9)
SELECT QUADRANT CHANNEL (F8)
CHART ZOOM (F9)
DATA POSITION (F10)
CURSOR SIZE (F11)
DISPLAY NULL(F12)
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
80/169
Analysis Setup
Curve Setup
Page 7-3
7.1 Curve Setup
Set the curve parameters by selecting the CURVE SETUP (F1) button.
Figure 7-3 Curve Setup Screen
Select i ng t he Curve Type
1. Press the SELECT CHANNEL (F8) button.
2. Rotate the Control Knob to select a channel.
3. Toggle the CURVE TYPE (F9) button to select ANGLE, VOLT, or NO CURVE for the selected channel.
4. If you selected ANGLE or VOLT, press NUM ID POINTS (F10) to set the number of ID points, then press NUM ODPOINTS (F11) to set the number of OD points. When you select an angle curve, you can set from 1 to 4 ID pointsand 3 to 5 OD points. When you select a volt curve, you can set from 3 to 7 points to describe the curve.
7.2 Cursor Control
Use the CURSOR CONTROL (F2) button to scale and adjust the signal of interest in the cursor window. These cursor
control options work the same as those in Data Review. See Section 6.2.1 for details.
http://../TestReview.pdfhttp://../TestReview.pdf -
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
81/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 7-4
7.3 Measure
Press the MEASURE (F3) button to select the measurement type of peak to peak or max rate. The measurement
results will be displayed at the top of the curve display.
Figure 7-4 Select Measurement Type
Volt age Mult
Use the VOLTAGE MULT (F11) button to normalize all channels.
Channel Rot at ion
Use the CHANNEL ROTATION (F9) button to rotate the signals.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
82/169
Analysis Setup
Adjust Curve
Page 7-5
7.4 Adjust Curve
Use the ADJUST CURVE (F4) button to scroll through all the curve points and adjust the actual flaw percentages per
the standard drawing.
Figure 7-5 Adjus t Curve
1. Press the ADJUST CURVE (F4) button.
2. Press the SELECT POINT (F8) button and select the first point to setup.
3. Press the DATA POSITION (F10) button and position the data to the appropriate flaw signal.
4. To set only the current channel selected, press SET THIS CHANNEL (F11).
5. To set all the channels for this coil, press SET ALL FREQS (F12).
6. The selected point will automatically increment to the next point.
7. Adjust the data position to the next flaw on the standard and repeat for all the curve points.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
83/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 7-6
7.5 Curve Table
Press the CURVE TABLE (F5) button to display a table of angles versus percent through wall.
Figure 7-6 Curve Table Screen
Select i ng a Diff erent Channel for t he Curve Table
1. Press the SELECT CHANNEL (F8) button.
2. Rotate the Control Knob to scroll through the channels.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
84/169
Analysis Setup
Data Mixing
Page 7-7
7.6 Data Mixing
To perform a data mix, follow the steps below.
Figure 7-7 Data Mixing Screen
1. Collect a tube support signal in test mode.2. In the Analysis Setup screen, press the DATA MIXING (F6) button.
3. Adjust SELECT NUM MIXES (F8) to 1.
4. Press the MIX CHAN 1 (F9) button and select the first channel of the mix.
5. Press the MIX CHAN 2 (F10) button and select the second channel of the mix.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
85/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 7-8
6. Press the PERFORMMIX (F12) button to calculate and apply the mix coefficients.
Figure 7-8 Per fo rm Mix
7. To add another mix channel, press SELECT NUM MIXES (F8) and use the Control Knob to select 2. Repeat theprevious channel selection and perform mix process as above.
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
86/169
Analysis Setup
Setup File
Page 7-9
7.7 Setup File
To save or recall an analysis setup file, press the SETUP FILE (F7) button.
Figure 7-9 Setup Fi le
Sav e Set up
Press the SAVE SETUP (F8) button to save the current analysis setup to the current recording directory. A directorywill be created in the current recording directory (cal group) named MIZ28_SETUP, and within that directory the
setup file will be saved as analysis.stp.
Recall Set up
Press the RECALL SETUP (F9) button to recall the setup file from the current directory.
The next section of this operating guide describes the Toolbox screen
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
87/169
Using the Toolbox Screen
Page 81
8.0 Using the Toolbox Screen
The Toolbox screen enables you to plot graphs of amplitude versus frequency (freq plot) and amplitude versus drive(drive plot), as well as view a list of materials that includes prime frequencies for a given wall thickness and stan-
dard depth of penetration at a given frequency.
To open the Toolbox screen, follow these steps:
1. Press the Menu Display button.
2. Use the Control Knob to select (highlight) TOOLBOX.
3. Press the Select button.
Figure 8-1 Open Toolbox Screen
-
7/27/2019 MIZ-28 OPERATING GUIDE (180 - 10020153 - 1 - D) - 1
88/169
MIZ-28 Eddy Current Test Instrument
Operating Guide
Page 8-2
The Toolbox screen initially displays the following system information: the instruments serial number, the firmware
revision number and the revision date.
Figure 8-2 Toolbox Screen
Figure 8-3 Toolbox Menu Structure
Serial number
Firmware revisionnumber
Revision date
FREQUENCY PLOT (F1)
DRIVE PLOT (F2)
MATERIAL LIST (F4)
SELECT COIL (F8)
ADJUST POSIT