mod.4. notes v.f1

Upload: sefer-avdiaj

Post on 02-Jun-2018

230 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/11/2019 Mod.4. Notes v.f1

    1/71

    INTERNATIONAL UNION FOR VACUUM SCIENCE,TECHNIQUE AND APPLICATIONS

    UNION INTERNATIONALE POUR LA SCIENCE,

    LA TECHNIQUE ET LES APPLICATIONS DU VIDE

    INTERNATIONALE UNION FRVAKUUM-FORSCHUNG, -TECHNIK UND -ANWENDUNG

    VISUAL AIDSFOR INSTRUCTIONS IN

    VACUUM TECHNOLOGY

    & APPLICATIONSSeries Editor

    John L Robins

    Module 4:

    PARTIAL PRESSURE ANALYZERS AND

    ANALYSISSecond Edition

    Leonard Beavis

    Presenters Notes

  • 8/11/2019 Mod.4. Notes v.f1

    2/71

    INTERNATIONAL UNION FOR VACUUM

    SCIENCE, TECHNIQUE AND

    APPLICATIONS

    VISUAL AIDS

    FOR INSTRUCTIONS IN

    VACUUM TECHNOLOGY

    & APPLICATIONS

    Module 4

    PARTIAL PRESSURE ANALYZERS AND

    ANALYSIS

    Second Edition

    Presenters Notes

    by

    Leonard Beavis

    Series Editor

    John L Robins

  • 8/11/2019 Mod.4. Notes v.f1

    3/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    3

    INTRODUCTION TO THE SECOND EDITION

    This revision of the Partial Pressure Analyzers and Analysis module of the IUVSTA

    Visual Aids Project replaces the 1stEdition which had been produced in 1972. Many

    changes in partial pressure analysis have occurred in the 30 intervening years. It isintended that this revision accurately reflect these changes. As this collection ofslides is intended to be an instructional tool in the technology and not an historical

    documentation of partial pressure analysis, all of the original material is beingreplaced with both new content and format.

    It is assumed that the students have a familiarity with elementary chemistry and

    physics and basic vacuum technology. It is further assumed that the instructor has

    experience with partial pressure analyzers and analysis as well as an advanced level of

    understanding of vacuum science. Accordingly, in most cases the material on the

    slide will not be explained in detail, i.e. this material has not been prepared as a stand-

    alone course for either students or instructors but is intended as material to enhancethe presentation of a course by a competent instructor.

    This Instructors Guide is arranged in the following fashion: For each slide, from the

    compendium of slides, there is accompanying text which includes a statement of the

    purposeof the slide, some points of emphasisregarding the material on the slide andsometimes some advanced topicsthe students may wish to discuss (possible subjects

    for an advanced course on partial pressure analysis).

    Leonard BeavisAlbuquerque, USA.

    June 2004

  • 8/11/2019 Mod.4. Notes v.f1

    4/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    4

    LIST OF SLIDES

    Section 1. INTRODUCTORY MATERIAL AND TERMINOLOGY

    Slide 1.01 Comparative Need for Total Press Gauges and Partial Press Analyzers

    Slide 1.02 Relationship of a PPA to an Ion Gauge and a Mass SpectrometerSlide 1.03 Terminology ReviewSlide 1.04 Terminology General 1

    Slide 1.05 Terminology General 2Slide 1.06 Determination of Resolving Power/Resolution

    Slide 1.07 Examples of Alternative Determinations of Resolution

    Slide 1.08 Terminology Ion Source

    Slide 1.09 Terminology Ion Source and Analyzer

    Slide 1.10 Terminology Analyzer

    Slide 1.11 Terminology Detectors

    Section 2. INFLUENCE OF THE COMPONENTS OF A PPA ON DATA

    Slide 2.01 Block Diagram of a PPA System

    Slide 2.02 Overview: Potential Impact of the Ion Source on Data

    Slide 2.03 Overview: Potential Impact of the Analyzer Section on DataSlide 2.04 Overview: Potential Impact of the Ion Detector on Data

    Slide 2.05 Overview: Data Collection and Display

    Section 3. ION SOURCES

    Slide 3.01 Methods for Producing Ions

    Slide 3.02 Electron Bombardment Ion Source

    Slide 3.03 Ionization Efficiency Curves

    Slide 3.04 An Open Ion SourceSlide 3.05 Nier Type Closed Ion Source

    Slide 3.06 Table of Parent Ions Expected in a Leak Tight Vacuum System

    Slide 3.07 Table of Parent Ions with Major Isotopes Included

    Slide 3.08 Metastable Ions Generated in the Ion Source

    Slide 3.09 Typical Hydrogen/Deuterium Spectrum

    Slide 3.10 Triatomic Deuterium FormationSlide 3.11 Triatomic Hydrogen Ion Formation

    Slide 3.12 Commonly Seen Multicharged Ions

    Slide 3.13 Species Produced by the Filament

    Slide 3.14 Fragmentation Species Produced in an Ion SourceSlide 3.15 Effect of Bombardment Voltage on Fragmentation

    Slide 3.16 Thermal Fragmentation by Hot Filaments

    Slide 3.17 Occasionally Seen Important Contaminants

  • 8/11/2019 Mod.4. Notes v.f1

    5/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    5

    Section 4. SAMPLES OF SPECTRA; ION SOURCE EFFECTS

    Slide 4.01 A Typical Spectrum of a Relatively Clean System

    Slide 4.02 A Spectrum with Air and Organic Contamination

    Slide 4.03 Scans during Pump Down at 70eV and 105eV Electron Energy

    Section 5. ION DETECTORS

    Slide 5.01 Faraday Cup DetectorSlide 5.02 Faraday Collector Features

    Slide 5.03 Secondary Electron DetectorsSlide 5.04 Discrete Dynode Electron Multiplier

    Slide 5.05 Continuous Dynode Electron Multiplier

    Slide 5.06 Secondary Electron Multiplier for Extreme Low Pressure

    Slide 5.07 Diagram of an Ion Counting System

    Section 6. MASS ANALYZERS

    Slide 6.01 Mass Analyzers Readily Available

    Slide 6.02 Magnetic Sector AnalyzerSlide 6.03 Operation of the Magnetic Sector Spectrometer

    Slide 6.04 Special Issues with the Magnetic Sector

    Slide 6.05 Time of Flight Mass Analyzer

    Slide 6.06 Operation of the Time of Flight Spectrometer

    Slide 6.07 Quadrupole Mass Filter

    Slide 6.08 Quadrupole Operation

    Slide 6.09 Stability Diagram for the Quadrupole Mass Filter

    Slide 6.10 Location in Time of Mass Peaks for Various Sweep ParametersSlide 6.11 Quadrupole Issues

    Slide 6.12 Symptoms of Improper TuningSlide 6.13 Mislabelled Spectrum for Hydrogen Selenide

    Section 7. DATA STORAGE AND DISPLAY

    Slide 7.01 Data Display and Storage

    Slide 7.02 Logarithmic Display of m/q = 1 - 72Slide 7.03 Linear Display of Data in Slide 7.02, m/q = 1 69Slide 7.04 Block Diagram of Data Acquisition with Feedback Control

    Section 8. SOME APPLICATIONS

    Slide 8.01 Air Leak Spectrum

    Slide 8.02 Leaky N2Backfill Valve

    Slide 8.03 Vacuum System Bakeout

  • 8/11/2019 Mod.4. Notes v.f1

    6/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    6

    Section 9. QUANTITATIVE AND QUALITATIVE DATA

    Slide 9.01 Qualitative vs. Quantitative Data

    Slide 9.02 Diagram of an Orifice Flow Calibration System

    Section 10. BIBLIOGRAPHY/REFERENCES

    Slide 10.01 Bibliography I (Props of Elements, Isotopes & Compounds)Slide 10.02 Bibliography II (Vacuum)

    Slide 10.03 Bibliography III (PPA Spectra Interpretation)Slide 10.04 Bibliography IV (Instrumentation)

  • 8/11/2019 Mod.4. Notes v.f1

    7/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    7

    PRESENTERS GUIDE

    Section 1. INTRODUCTORY MATERIAL AND TERMINOLOGY

    Slide 1.01 Comparing the Need for a Total Pressure Gauge with that for aPartial Pressure Analyzer

    Purpose: To introduce the idea that not every vacuum system requires a PPA but thatmuch of modern vacuum technology depends upon a qualitative and possibly

    quantitative knowledge of the constituents of the gases in the vacuum.

    Points of emphasis: Total pressure gauges are used when the reactants and/or

    products of the vacuum process are known or controlled independently of the process.

    For example: freeze drying, or chemical vapor deposition (CVD). Total pressure

    gauges are also used when the results of particle to particle interaction are practically

    independent of the species involved. For example: in high energy particleaccelerators. As total pressure gauges give a single output and are much less

    expensive than a partial pressure analyzer, they are the gauges of choice in the above

    instances.

    Partial pressure analyzers (PPA) are used in applications where the species to bedetected or controlled is unknown or known to be a small fraction of the total gas

    present. Examples include leak detecting, multi-layer semiconductor processing, gas

    surface studies (catalysis) and contamination control processes. The requirement for

    qualitative and/or quantitative gas species information is paramount. In these

    instances it is necessary to discern a particularly active gas species or one that is at alow level compared to the total gas present. Anything but a PPA is inappropriate.

    Advanced topics:

    Relationship between advance of vacuum technology and the advent of

    partial pressure analysis.

    History of mass spectrometry and partial pressure analysis.

    Development of the mass spectrometer leak detector (the first PPA) and

    improved leak detection.

  • 8/11/2019 Mod.4. Notes v.f1

    8/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    8

    Slide 1.02 Relationship of a PPA to an Ion Gauge and a Mass Spectrometer

    Purpose: To illustrate the similarities and differences between these three very

    important instruments.

    Points of emphasis: Point out the increase in complexity and cost in going from therather simple ionization gauge to the analytical mass spectrometer.

    Also discuss the similarities of the PPA to the mass spectrometer. For example theyeach measure mass to charge ratio. They each have a similar trade-off of sensitivity

    and resolution.

    The PPA is generally an instrument that can determine one mass number from an

    integer adjacent mass number, whereas an analytical mass spectrometer can discern

    masses that may be one thousandth of a mass unit or less apart.

    Advanced topics: What must be sacrificed in order to increase resolution of a PPA?

    Inlet systems for a PPA.

  • 8/11/2019 Mod.4. Notes v.f1

    9/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    9

    Slide 1.03 Terminology - Review

    Purpose: To review some of the terminology that will be used throughout the

    remainder of the illustrations

    Points of emphasis: Each element has unique chemical properties and has associated with it an

    atomic number. The atomic number is the number of protons or electrons

    associated with that element. For example: hydrogen is atomic number 1 with

    one proton and one bound electron. Neon is atomic number 10 with ten

    protons and ten bound electrons.

    Most elements have more than one naturally occurring isotope associated with

    them. An elemental isotope has the identical number of protons and bound

    electrons and differs from other isotopes of a particular element in the number

    of neutrons in the nucleus. Hydrogen has three naturally occurring isotopes.

    These are uniquely named protium, deuterium, and tritium and have zero, one

    and two neutrons in addition to a single proton respectively in their nuclei.

    Neon also has three naturally occurring isotopes. These are designated neon-

    20, neon-21 and neon-22 and contain in addition to ten nuclear protons 10, 11

    and 12 neutrons respectively.

    An atom has associated with it a mass, which for the purposes of partial

    pressure analysis is an integer value. This integer, atomic mass, is the sum of

    the number protons and neutrons in the nucleus. Thus hydrogen with threeisotopes has masses one, two and three. An atom is the single smallest

    chemical entity of any element. Atoms of an element behave chemically

    nearly identically no matter what the mass of the particular isotope

    Associated with every element is the naturally occurring isotopic abundance ofthe element. These abundances are tabulated in various handbooks. In the

    case of hydrogen most of it is protium, but 15 parts per hundred thousand aredeuterium and about 1 part in 1017 is tritium. The accurately determined and

    known isotopic abundances are a tool used in partial pressure analysis.

    Chemical combinations of atoms make molecules. Most elements that are

    insulating and many semiconducting elements appear in the vapor/gas phase

    as multi-atomic molecules. Examples include nitrogen as N2 and arsenic asAs3and As4. These have masses associated with the number of atoms present.

    Thus N2has masses of 28, 29 and 30. Arsenic vapor has masses 225 and 300

    associated with As3 and As4respectively.

    Molecules may be mono-elemental, as above or multi-elemental as in water ormethane. Multi-elemental molecules are called compounds. Associated with

    a compound is also a mass. It is the sum of the masses of the elements (byisotope) that make the compound. For example most water has a mass of 18,

    and most methane has a mass of 16.

    Advanced topics:

    Determination of isotopic abundances of elements.

    Determination of relative abundances of molecules.

    Determination of isotopic masses to 1ppm.

  • 8/11/2019 Mod.4. Notes v.f1

    10/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    10

    Slide 1.04 Terminology General 1

    Purpose: To become familiar with terminology encountered in partial pressure

    analysis.

    Points of emphasis: Although at one time PPA, vacuum analyzer and RGA had somewhat different

    meanings, for the past several years these names have been used

    interchangeably. Occasionally one will also hear mass spectrometer mixed in

    as a synonym also. It is probably not prudent to do so as most, if not all, mass

    spectrometers have vastly different capabilities than PPAs.

    In partial pressure analysis the mass number is always expressed in atomic

    mass units (AMUs) rather than in metric or international mass units. 1 AMU

    = 1.66x10-27 kg. Additionally, and in contrast to analytical mass spectrometry,

    masses of atoms, molecules and molecular fragments are always expressed in

    integer AMUs.

    PPAs, as well as mass spectrometers always measure mass to charge ratio

    (m/q) and not mass independent of charge. This is true of all possible schemes

    for mass separation (analysis). This fact has large implications when making

    quantitative interpretation of mass spectral output from a PPA. Multicharged

    ions appear at mass numbers in the spectra different from singly charged ions.

    Doubly charged carbon dioxide ions appear at mass 22 for example.

    Gas pressure indicates a force per unit area. As pressure drops below about

    10-8

    atmospheres it becomes impossible to measure pressure directly. Gas

    density (the number of molecules per unit volume) is the parameter that is

    measured by ion gauges and PPAs. Gas density, !, is related to gas pressure,

    p, through the ideal gas law: P=!RT. The measurement of gas density by aPPA becomes an issue when reporting pressures quantitatively measured at

    elevated temperatures.

    Whilst the SI units for pressure are the Pascal (Pa) (defined as one Newton per

    square meter) and the millibar (mbar), an earlier unit, the Torr (or torr), is still

    often used, particularly in the USA. 1 Torr = 1.33 mbar = 133 Pa. It is aconvenient approximation, particularly where orders of magnitude only are

    being considered, to treat the units mbar and Torr as equivalent.

    Advanced topics:

    Peaks at non-integer mass numbers.

    Elimination of multi-charged ions.

  • 8/11/2019 Mod.4. Notes v.f1

    11/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    11

    Slide 1.05 Terminology General 2

    Purpose: To define these four terms and show the relationship of the first two and the

    relationship of the third and fourth.

    Points of emphasis: The ability to detect one ion species in the presence of all other ion species is

    called the dynamic range of the instrument. For some PPAs this may be as

    small as one part per thousand. Modern quadrupoles have a dynamic range of

    one part in 106-107.

    The minimum detectable partial pressure is an expression of the possibility of

    detecting the presence of a few molecules/unit volume of a particular

    molecule. For example most modern quadrupoles can detect 10-14

    mbar of

    nitrogen under ideal conditions. Obviously these conditions must include a

    total pressure of 10-8

    mbar or less if the dynamic range of the instrument is

    106. The dynamic range and minimum detectable partial pressure should be

    specified in some relational form to assure there is no misunderstanding about

    the capabilities of the PPA.

    Sensitivity for PPAs is normally expressed in amp/torr (or amp/mbar) for a

    particular gas, e.g. nitrogen, rather than in torr-1 (or mbar-1) as is done for

    ionization gauges. To assure there is no confusion about such a specification

    the amp/torr figure should include the electron bombarding current and the

    electron accelerating potential. For example 1ma at 70eV electron energy.Resolution is expressed as the inverse of the peak width of a mass spectral peak in

    AMU-1

    . Resolving power is the ratio of the mass of the spectral peak being measured

    to its width. Clearly resolution and resolving power are related. As mass peaks do

    not have vertical sides to them the point on the side of the peak where width is to bemeasured must be specified. Slides 1.06 and 1.07 give examples of resolving power

    and resolution measurements. Sensitivity and resolution are inversely related in all

    PPAs. An increase in one of these factors comes at the expense of a corresponding

    decrease of the other. Sensitivity " resolution-1

    . For this reason specifications

    involving resolution or sensitivity must be related i.e. specifying resolution without a

    corresponding sensitivity and vise versa is meaningless.

  • 8/11/2019 Mod.4. Notes v.f1

    12/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    12

    Slide 1.06 Determination of Resolving Power/Resolution

    Purpose: To give an example of making a resolving power estimate.

    Points of emphasis:

    Resolving power is the mass number of a peak divided by its peak width inAMU.

    In the example, the peak width for m/q=86 is found by taking the distance D

    between peaks 84 and 86, in this case estimated to be 23 units, and dividing it

    into the m/q difference of 2 AMU, and multiplying this quotient by the peakwidth of m/q=86 at the 10% amplitude point, which is estimated to be 11 units

    (i.e. peak width = 11x(2/23) AMU).

    Thus the resolving power is equal to 86/(22/23) #90.

    In this example the 10% peak amplitude was taken as the place to measure the

    peak width. This follows from the AVS recommended practice on PPA

    calibration. Obviously the resolving power would be greater the higher up thepeak at which its width is measured.

    Advanced topics

    Is resolving power constant across the mass range?

    How does resolving power vary with instrument type?

  • 8/11/2019 Mod.4. Notes v.f1

    13/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    13

    Slide 1.07 Examples of Alternative Determinations of Resolution

    Purpose: To demonstrate various concepts for measuring resolution.

    Points of emphasis:

    In the first example the peak width $m is measured at the 50% level. Clearlythe resolution will be greater the further up the peak that the width is measured

    In the second example resolution is defined by the ratio Y/xY divided by themass number difference between the interfering peaks. As the peaks narrow

    the resolution increases as expected.

    In the third example resolution is defined by the ratio of the sum of the two

    peak heights divided by the height of the valley between them divided by the

    mass difference between the measured peaks. It is likely that as the peaks

    narrow the valley height will decrease thus increasing the resolution.

    In the fourth example the resolution is treated as in the third, above, except

    that rather than dividing the valley into the sum of the neighboring peaks, thevalley height is divided into either of the equal amplitude peak heights.

    Advanced topics:

    Is resolution constant across the mass range?

    Is it instrument type dependent?

    Are there expressions that relate the various definitions of resolution

    displayed above?

  • 8/11/2019 Mod.4. Notes v.f1

    14/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    14

    Slide 1.08 Terminology Ion Source

    Purpose: To discuss a number of terms relevant to ion source operation. Ion sources

    are discussed in section 3, slides 3.01 to 3.17

    Points of emphasis: Fragmentation and cracking are used interchangeably. These terms refer to a

    molecule or molecular ion coming apart, i.e. fragmented, because of an

    introduction of excess energy from one of a variety of sources. The term

    cracking comes from the petrochemical industry; a major user of mass

    spectrometers.

    Impact fragmentation is caused by the collision of an energetic electron with a

    molecule or molecular ion.

    Thermal fragmentation occurs when a molecule or molecular ion collides with

    a high temperature surface. The high temperature causes the particle tofragment.

    Ionization efficiency is the number of electron/ion pairs produced per electron

    colliding with an atom or molecule. Ionization probability is the probability

    that an electron will ionize an atom or molecule. In each of these cases the

    efficiency or probability is zero until the bombarding electrons energy

    exceeds the ionization potential. At energies above the ionization potential theprobability and efficiency are greater than zero and a function of electron

    energy.

    When an ion is formed if it loses an electron the ion will take on one positive

    charge. This is a singly charged ion. If the ion that is formed loses more than

    one electron during the ionization process it becomes a multi-charged ion.

    In an ion source, electrons collide with various electrode surfaces in thesource. It is possible if atoms/molecules are not tightly bound to the surface

    that the electron can cause the atom/molecule to desorb into the vacuum. This

    is called electron induced desorption EID.

  • 8/11/2019 Mod.4. Notes v.f1

    15/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    15

    Slide 1.09 Terminology Ion Source and Analyzer

    Purpose: To define terms that apply to the ion source and the analyzer sections of a

    PPA.

    Points of emphasis: Charge exchange occurs when an ion collides with a neutral molecule, and the

    molecule loses an electron to the ion (possibly neutralizing the ion). The

    molecule becomes an ion. Charge exchange becomes an issue only when the

    gas density is high enough so that particle-particle collisions are likely, i.e. the

    mean free path becomes smaller than the dimensions of the ion source or the

    analyzer section.

    Space charge occurs when the ion density becomes large enough to cause the

    ions to be deflected, or spread. This generally causes the loss of ions.

    Ion induced desorption (IID) is similar to EID. In this case the particle

    causing desorption is an ion rather than an electron.

    Depending upon the particular type of PPA, ions that are formed in the ion

    source may be injected into the analyzer at an angle and from a position such

    that they traverse the analyzer as if they were ions at the m/q in tune, thus

    giving a false signal to the detector. These non-tuned ions are called off axis

    ions.

  • 8/11/2019 Mod.4. Notes v.f1

    16/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    16

    Slide 1.10 Terminology Analyzer

    Purpose: To define terms that apply to the analyzer section of a PPA . Mass

    analyzers are discussed in section 6, slides 6.01 to 6.13.

    Points of emphasis The Mass Range of a PPA is defined by the operating parameter of the

    analyzer section. It is typically from AMU=1 to 100, but may be 1 to 300 or

    greater in some instruments. The Mass Range of magnetic sector instruments

    may consist of multiple spans e.g. 1- 50 and 12 300. This typically is

    accomplished by an adjustment to the magnet.

    Scan Rate is expressed either as AMU/second or full range scans e.g. AMU 1-

    100 per unit of time. In a quadrupole, mass peaks can be scanned at a

    maximum rate of a few hundred AMU per second. A time of flight analyzer

    can scan at rates 105 AMU /second depending upon the mass range

    interrogated or thousands of scans /second.

    Scan Parameter is the operating parameter of the instrument (analyzer) that is

    adjusted, typically as a function of time, to move from one mass peak to

    another. In a magnetic sector instrument, the scan parameter is typically the

    ion injection energy into the magnet field region in less expensive instrument,

    but may also be the magnet field intensity in some sector instruments. In the

    time of flight instrument the scan parameter is a combination of injection

    pulse length and ion energy. Quadrupoles scan the D.C. voltage applied to thequadrupole rods. The scan parameter relation to mass spectra presentation

    will become more apparent in the Mass Analyzer section of this module, slides

    6.01 to 6.13.

    Transmission is usually expressed as amperes/pressure unit. It is anexpression of the efficiency with which the instrument (analyzer) can extract

    from the ion source, transmit through the analyzer and detect ions of a specificspecies e.g. nitrogen.

    Ghost Peaks are those apparent mass peaks that are not representative of any

    actual species in the vacuum. For example: a peak at mass 19.5 is sometimes

    seen in a magnetic sector due to ion induced desorption/charge change that

    occurs from non-resonant ions striking the analyzer tube wall.

  • 8/11/2019 Mod.4. Notes v.f1

    17/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    17

    Slide 1.11 Terminology Detectors

    Purpose: To define terms and describe briefly the types of ion detection used in

    Partial Pressure Analyzers. Ion detectors are discussed in Section 5, slides 5.01 to

    5.07.

    Points of emphasis:

    The Faraday detector is a detector that measures ions by the flow of electrons

    to neutralize the ions as they collide with the metal surface of the ion

    collector. The collector is usually at ground (earth) potential. Suppression of

    secondary electron emission from ion bombardment or the photoelectron

    effect is assumed.

    The Multiplier detector is used to amplify ion currents in the vacuum system

    through secondary electron emission from ion and electron collisions. It

    depends upon high ion to electron and electron to electron conversion

    efficiencies. These are undesirable attributes in a Faraday detector.

    Multiplier detectors are used for detecting lower partial pressures than

    Faraday detectors can detect.

    Ion counting is a technique using a multiplier detector to detect very low

    partial pressures. It is capable of measuring at pressure levels lower than

    those using a multiplier detector in its analog mode. This detection capability

    comes at the expense of more complex electronics.

    ======================= End Section 1 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    18/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    18

    Section 2. OVERVIEW OF INSTRUMENTATION

    Slide 2.01 Block Diagram of a PPA System

    Purpose: To define by diagram a complete partial pressure analysis system.

    Points of emphasis:

    In addition to the basic partial pressure analyzer instrument i.e. ion source,

    analyzer and detector, the data collection, display, storage and analysis

    equipment is an important part of any partial pressure analysis system.

    The impact and interaction of each of these subsystems will be illustrated.

  • 8/11/2019 Mod.4. Notes v.f1

    19/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    19

    Slide 2.02 Overview: Potential Impact of the Ion Source on Data

    Purpose: To illustrate the importance of understanding the ion source as a generator

    of data.

    Points of emphasis: Much of the information in a PPA spectrum is not representative of the

    contents of the vacuum system the PPA is interrogating, but is rather a

    convolution of interactions of the vacuum species and the ion source materials

    and potentials.

    The superfluous information is often useful in interpreting the low-resolution

    mass spectra of a PPA.

    Nearly all of the information/misinformation in a PPA spectrum is generated

    in the ion source. This fact must always be kept in mind when describing the

    remainder of the vacuum based upon PPA data.

    Atomic and radical ions are very reactive and capable of generating species

    not present in the remainder of the vacuum system.

    Neutral atoms and radicals are also very reactive and capable of generating the

    same sorts of misinformation.

    The sources of artifacts are alluded to on the slide.

    Advanced topics:

    How would misinformation be reduced?

    Is there some useful misinformation?

  • 8/11/2019 Mod.4. Notes v.f1

    20/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    20

    Slide 2.03 Overview: Potential Impact of the Analyzer Section on Data

    Purpose: To give an overview of how the analyzer section influences PPA data.

    Points of emphasis:

    The analyzer section converts a heterogeneous mixture of different ions intoindividual atomic or molecular ionic constituents.

    It is important that resonant (tuned to the chosen mass number) ions be

    transmitted with maximum efficiency to the ion detector.

    It is also important that non-resonant species not be transmitted to the detector.

    They lead to detector signal noise. Analyzer bypass is such a source of noise.

    Collisions with the walls or analyzer elements can lead to spurious signal

    generation or analyzer detuning through insulating layer build-up.

  • 8/11/2019 Mod.4. Notes v.f1

    21/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    21

    Slide 2.04 Overview: Potential Impact of the Ion Detector on Data

    Purpose: To give an overview of the influence of the ion detector on data.

    Points of emphasis:

    Contrast the energies of ions collected on Faraday and electron multiplierdetectors.

    Ion neutralization requires one or more electrons/ion to flow in the detector

    circuitry.

    Detectors in optical sight of the ion source may produce photoelectrons.

    Describe the possible consequences.

    Ions striking the ion collector surface can cause secondary electron emission.

    Describe the possible consequences.

    Advanced topics:

    How can electron currents in the detector be modified?

  • 8/11/2019 Mod.4. Notes v.f1

    22/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    22

    Slide 2.05 Overview: Data Collection and Display

    Purpose: To describe some possible impacts of the data collection scheme on the

    storage display and analysis of the data.

    Points of emphasis: Data output from most PPAs is analog. Describe what is an adequate

    conversion to digital information.

    Describe how computers handle data.

    Describe non-linearities in data management e.g. amplifier saturation,

    logarithmic and linear data conversions.

    Describe sources of noise in the data and its impact.

    Advanced topics:

    When is PPA data already digital?

    How can non-linearities be minimized? How can noise be minimized?

    ======================= End Section 2 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    23/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    23

    Section 3. ION SOURCES

    Slide 3.01 Methods for Producing Ions

    Purpose: To illustrate some of the ways that energy input to a molecule can bereflected in changes to the molecule.

    Points of emphasis:

    The purpose of the ion source is to convert a neutral atom or molecule into an

    ion that is an accurate reflection of the original neutral species, e.g. correct

    mass, correct structure.

    Not all energy inputs to a neutral atom or molecule accomplish this goal.

    Advanced topics:

    What are some ways of maximizing the ion sources purpose?

    Can an ion source make a perfect conversion from neutral to ion?

  • 8/11/2019 Mod.4. Notes v.f1

    24/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    24

    Slide 3.02 Electron Bombardment Ion Source

    Purpose: To show what a typical ion source looks like and how ions are generated.

    Points of emphasis:

    Describe the purpose of each electrode and the corresponding potential appliedto the electrode.

    Point out the flow of ions and electrons.

    Describe what happens during electron collisions with neutrals or ions.

    Describe the energetics of reactions in the ion source.

    Advanced topics:

    What reactions occur when neutrals or ions contact the filament?

    What interactions occur when atoms or molecules contact the other

    electrodes?

    What interactions occur when ions contact the other electrodes?

  • 8/11/2019 Mod.4. Notes v.f1

    25/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    25

    Slide 3.03 Ionization Efficiency Curves

    Purpose: To show how the electron bombardment energy relates to the ability to

    produce ions of various gasses.

    Points of emphasis: Note that the peak in the ionization efficiency does not occur at the same

    energy for all gasses.

    Note that the peak efficiency of ionization is not the same for all gasses.

    Describe what the impact of these two facts is.

    Advanced topics:

    Why do most PPAs operate with 70 eV electron energy?

    What electron energy should be used for leak detecting with helium? With

    neon?

  • 8/11/2019 Mod.4. Notes v.f1

    26/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    26

    Slide 3.04 An Open Ion Source

    Purpose: To show a schematic diagram of an open ion source used with most

    commercially available PPAs.

    Points of emphasis: Describe the advantages and disadvantages of this type of ion source.

    Describe the materials usually used for each of the electrodes.

    Describe the advantages and disadvantages of these selections.

    Advanced topics:

    How would you optimize the electrode materials? Why?

    What are the optimum potentials on the electrodes? Why?

  • 8/11/2019 Mod.4. Notes v.f1

    27/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    27

    Slide 3.05 Nier Type Closed Ion Source

    Purpose: To show a diagram of a Nier type closed ion source.

    Points of emphasis:

    Describe where gas is injected into this type of ion source. Describe differential pumping and how it applies to this type of ion source.

    Describe the advantages and disadvantages of the Nier type of ion source.

  • 8/11/2019 Mod.4. Notes v.f1

    28/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    28

    Slide 3.06 Table of Parent Ions Expected in a Leak Tight Vacuum System

    Purpose: To show a table of mass spectral peaks of an idealized hypothetical

    spectrum of residuals in a non-leaking vacuum system.

    Points of emphasis: This spectrum is not real. It is hypothetical only.

    It assumes that only the major isotope of each element exists.

    It also assumes no multiple ionization, no fragmentation, and no meta-stable

    ions exist.

  • 8/11/2019 Mod.4. Notes v.f1

    29/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    29

    Slide 3.07 Table of Parent Ions with Major Isotopes Included

    Purpose: To show a table of a hypothetical spectrum similar to that in slide 3.06, but

    with isotopic peaks added. The importance of isotopes in identifying elements

    becomes apparent.

    Points of emphasis:

    This spectrum is not real.

    Isotopic peaks to the 0.001 level of the major peak have been added.

    It is through isotopic peaks that low-resolution mass spectra species are often

    unequivocally identified.

  • 8/11/2019 Mod.4. Notes v.f1

    30/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    30

    Slide 3.08 Metastable Ions Generated in the Ion Source

    Purpose: To draw attention to two major meta-stable ions, that often appear, in PPA

    spectra.

    Points of emphasis: These meta-stable ion species are generated in the ion source.

    The amount of each of these species that is generated is dependent upon: the

    concentration of the materials that go to make them up in a non-linear fashion;

    the electron current; the ion source geometry; and the electron energy in theion source.

    H3+ and H3O

    + are only present in the PPA and not in the vacuum system.

    Dempster discovered the meta-stable tri-atomic hydrogen ion in the early 20th

    century.

    Advanced topics:

    What happens to these ions if they collide with a surface?

    What happens to these ions if they drift freely in a vacuum?

  • 8/11/2019 Mod.4. Notes v.f1

    31/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    31

    Slide 3.09 Typical Hydrogen/Deuterium Spectrum

    Purpose: To show a spectrum of tri-atomic hydrogen (deuterium).

    Points of emphasis:

    Deuterium was used to accurately discern the amount of tri-atomic hydrogenions produced.

    Identify the potential makeup of each mass peak from AMU=1-6.

    Masses 1, 5 and 6 have single constituents.

    Masses 2, 3 and 4 consist of multiple constituents.

  • 8/11/2019 Mod.4. Notes v.f1

    32/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    32

    Slide 3.10 Triatomic Deuterium Formation as a Function of Ion Source

    Operating Parameters

    Purpose: To show the dependence of tri-atomic hydrogen ion formation upon various

    operating parameters in an ion source.

    Points of emphasis:

    Point out dependence of tri-atomic ion formation on the electron current.

    Point out how tri-atomic hydrogen ion formation depends upon electron

    bombardment energy.

    Advanced topics:

    Why does the tri-atomic ion formation depend upon electron current?

    Why does tri-atomic ion formation depend upon electron energy?

  • 8/11/2019 Mod.4. Notes v.f1

    33/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    33

    Slide 3.11 Triatomic Hydrogen Ion Formation as a Function of Molecular

    Hydrogen Concentration

    Purpose: To show the dependence of tri-atomic hydrogen ion formation on the di-

    atomic hydrogen concentration.

    Points of emphasis:

    The left side plot is for deuterium, but would apply to hydrogen (protium) as

    well.

    The data was taken from two different quadrupoles with differing geometriesin their ion sources.

    The concentration dependence of the tri-atomic ion on di-atomic hydrogen is

    of the form H3+ "H2

    n . Where n = 1.3 and 1.5 respectively for left and right

    plots.

    Advanced topics: What would be the expected maximum value for n?

    What conditions would maximize n?

    Why does n differ for the two ion sources, and from nmax?

  • 8/11/2019 Mod.4. Notes v.f1

    34/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    34

    Slide 3.12 Commonly Seen Multicharged Ions

    Purpose: To give examples of the common multi-charged species and where they will

    appear in a spectrum.

    Points of emphasis: Multi-charged species are present in the PPA and not in the remainder of the

    vacuum system.

    They are, however, usually due to parent molecules that are extant in the

    vacuum system.

    They are stable until they collide with a surface or a gas molecule.

    Multi-charged ions of other atoms and molecules are possible, but usually in

    lower relative concentrations.

    Advanced topics:

    How is the concentration of multi-charged ions distorted in a Faradaycollector?

    How is the concentration of multi-charged ions distorted in an electron

    multiplier detector?

  • 8/11/2019 Mod.4. Notes v.f1

    35/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    35

    Slide 3.13 Species Produced by the Filament

    Purpose: To give examples of species that are generated from an ion source filament.

    Points of emphasis:

    Carbon oxides produced by a filament outgassing and reaction may beindependent of gasses in the vacuum system.

    A common source of oxygen for production of carbon oxides by the filament

    is water vapor from the vacuum system.

    Oxygen ions (O2+) are produced by ion sputtering or evaporation from thoria

    impregnated filaments.

    Any of the filament-produced species may escape to the vacuum system,

    although escape is unlikely.

    Advanced topics:

    How can carbon oxides be produced on the filament independent of carbon oroxygen in the vacuum?

    Why doesnt a cold filament produce carbon oxides?

    Is the amount of filament-produced species representative of the concentration

    of the elements of these species in the ion source? Why?

    Why is escape of filament-produced species to the vacuum unlikely?

  • 8/11/2019 Mod.4. Notes v.f1

    36/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    36

    Slide 3.14 Fragmentation Species Produced in an Ion Source

    Purpose: To tabulate expected fragment ions from commonly seen parent molecules.

    Points of emphasis:

    The tabulated species are nearly always present in PPA spectra. Fragment ion species peak amplitudes depend upon the operating parameters

    of the PPA ion source i.e. filament material, operating temperature, electron

    energy etc.

    Fragment ions and neutrals are well confined to the PPA. They are generally

    not out in the vacuum system.

    Advanced topics:

    What are the consequences of fragment species in the ion source?

    Are fragment ions useful in identifying parent molecules? How?

    What can be done to control fragmentation? Why do fragment species remain primarily in the PPA?

  • 8/11/2019 Mod.4. Notes v.f1

    37/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    37

    Slide 3.15 Effect of Bombardment Voltage on Fragmentation

    Purpose: To show the impact of electron bombardment energy on fragmentation.

    Points of emphasis:

    The spectra were taken in a stable, i.e. pressure constant, vacuum system. The only change was the electron bombardment energy being changed from

    50 to 70 eV.

    The mass 17 fragment ion of water is 0.31 of the parent mass 18 peak when

    the electron energy is 50 eV and 0.39 when the electron energy is 70 eV.

    Note that all of the peaks have smaller amplitude with the lower electronbombardment energy.

    Advanced topics:

    Are the mass peak amplitudes expected to increase as electron energy is

    increased? Why? What is the optimum electron energy? Explain?

  • 8/11/2019 Mod.4. Notes v.f1

    38/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    38

    Slide 3.16 Thermal Fragmentation by Hot Filaments

    Purpose: To show the impact of filament operating temperature on fragment ion

    production.

    Points of emphasis: The PPA had filaments of W (ThO2) and LaB6 symmetrically mounted with

    respect to the ion source and analyzer.

    Ion source voltages and emission currents were held constant. Only the

    operating temperature and material of the filament were varying. Thehexaboride filament operated at approximately 1000 K. The thoria filament

    operated at approximately 1650 K.

    The PPA was a R.F. linear type instrument.

    Note the spectrum complexity increases with temperature.

    The thoria filament shows the characteristic oxygen peak at m/q= 32.

  • 8/11/2019 Mod.4. Notes v.f1

    39/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    39

    Slide 3.17 Occasionally Seen Important Contaminants

    Purpose: To tabulate contaminant species and fragments from primarily acid

    treatment residue.

    Points of emphasis: Hydrochloric acid appears at masses 36 and 38 in a 3:1 ratio. Chlorine at

    masses 35 and 37 will be seen in the same ratio as fragments.

    Nitric acid residue often appears as nitrogen (m/q=28) and nitric oxide

    (m/q=30). The parent (m/q=63) may or may not appear. Other sources ofN2and NO are also possible.

    Sulfuric acid and other sulfur residues usually appear at m/q=48 (SO+) and

    m/q=64 (SO2+). The appearance of only one of the peaks indicates that the

    species is probably not sulfur related.

    ======================= End Section 3 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    40/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    40

    Section 4. SAMPLES OF SPECTRA; ION SOURCE EFFECTS

    Slide 4.01 A Typical Spectrum of a Relatively Clean System

    Purpose: To show a PPA spectrum of a relatively clean vacuum system and identifythe peaks.

    Points of emphasis:

    Point out and identify the parent, isotopic, meta-stable, multi-charged and

    fragment ion mass peaks on this spectrum.

    Note the logarithmic amplitude scale and its implications.

    The amplitudes of the mass peaks are in amperes not pressure units.

    Advanced topics:

    Is there a possibility that the identification of any of the peaks is wrong?

    Which one(s)? Why?

    What would one need to do to relate peak amplitudes as pressure rather than

    ion currents?

  • 8/11/2019 Mod.4. Notes v.f1

    41/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    41

    Slide 4.02 A Spectrum with Air and Organic Contamination of the Vacuum

    System

    Purpose: To show and explain a PPA spectrum of a vacuum system with an air leak

    and organic contamination.

    Points of emphasis:

    Contamination in the system is primarily organic molecules.

    Because of the nature of fragmentation and complexity of organic

    compounds, the spectrum becomes much more complex.

    Major peaks of nearly all organic compounds are fragments. In the case of

    inorganic gasses the major peak is nearly always the parent peak.

    Odd numbered mass peak fragments of organics are larger than the adjacent

    even numbered mass peak fragments.

    Peak amplitudes are in amperes.

    Advanced topics:

    Identify the species contributing to each mass peak.

    Why are odd numbered mass peak fragments larger than adjacent even

    numbered mass peak fragments?

    What causes the series of prominent peaks separated by fourteen mass units?

    Are there ambiguities in the identification of the mass peaks? Why?

  • 8/11/2019 Mod.4. Notes v.f1

    42/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    42

    Slide 4.03 Scans during Pump Down at 70ev and 105ev Electron Energy

    Purpose: To show the impact of increasing the electron energy from 70 eV (left

    spectrum) to 105 eV (right spectrum) on the mass peak amplitudes.

    Points of emphasis: The two spectra were taken in sequence during a pump-down of a vacuum

    system.

    Most of the spectral peaks are seen to decrease as the system is evacuated.

    The helium sensitivity increases by a factor of greater than two with the

    increase of electron energy in the ion source in this particular PPA. See mass

    4.

    When leak detecting with helium the electron energy should be temporarily

    increased in order to give greater sensitivity.

    Note the appearance of O++(m/q = 8) when the electron energy is increased.

    Advanced Topics:

    Why does the helium sensitivity increase with electron energy?

    What are the disadvantages of operating the ion source with greater electron

    bombardment energy?

    ======================= End Section 4 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    43/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    43

    Section 5. ION DETECTORS

    Slide 5.01 Faraday Cup Detector

    Purpose: To show the electrode configuration of a typical Faraday Cup ion collector.

    Points of emphasis:

    The collector shape is important as well as the presence of the associated

    electrodes.

    Secondary electron suppression is important.

    Advanced topics:

    Why is the Faraday collector shape important?

    Why is secondary electron suppression important?

    Slide 5.02 Faraday Collector Features

    Purpose: To give the main features of a Faraday collector.

    Points of emphasis:

    For high vacuum and ultra-high vacuum (10-5 to 10-9 mbar total pressure)

    partial pressure measurements, a Faraday collector is adequate.

    A Faraday collector is the simplest to use and is usually drift free.

    Multi-charged ions generate larger signals due to their degree of charge.

    Therefore multi-charged ion peaks appear larger than they are in a PPAspectrum.

    The range of a Faraday collector detector can be extended (2-3 orders) tolower currents (pressure/gas density) through the use of a vibrating reed

    electrometer amplifier in the detector circuit.

  • 8/11/2019 Mod.4. Notes v.f1

    44/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    44

    Slide 5.03 Secondary Electron Detectors

    Purpose: To describe features of two secondary electron multiplier detectors.

    Points of emphasis:

    All secondary electron multiplier detectors (SEMDs) are sensitive to theirenvironment. This causes the multiplier gain to drift downward with exposure

    to the vacuum.

    SEMDs come in two types. The older but still used discrete dynode electron

    multiplier (DDEM) and the newer continuous dynode electron multiplier(CDEM).

    Multi-charged ions, because they have greater velocity at a given energy lead

    to potentially larger signals.

    Determine the overall gain of an electron multiplier.

  • 8/11/2019 Mod.4. Notes v.f1

    45/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    45

    Slide 5.04 Discrete Dynode Electron Multiplier

    Purpose: To show a diagram of a discrete dynode electron multiplier and describe

    major features of its operation.

    Points of emphasis: Dynodes need high ion to electron and electron to electron conversion

    efficiencies.

    Common dynode materials include silver-magnesium and copper- beryllium

    alloys.

    Stability of dynodes is an issue. Some recent improvements are claimed. See

    the final section of the bibliography (Slide 10.04).

    Advanced Topics:

    Calculate the average gain per electron required for 10 dynode and 14 dynode

    multipliers to give a total gain of the multiplier of 106.

    Ion to electron conversion is Gaussian. Assume the conversion of ions to

    electrons is 3 with a %of 1. How many ions/100 are not counted?

  • 8/11/2019 Mod.4. Notes v.f1

    46/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    46

    Slide 5.05 Continuous Dynode Electron Multiplier

    Purpose: To show and describe important features of the continuous dynode electron

    multiplier detector.

    Points of emphasis: Explain the bend or twist in the CDEM type of detector.

    Describe the coating and its functionality.

    Describe degradation and end of life of the CDEM and DDEM multiplier

    detectors.

  • 8/11/2019 Mod.4. Notes v.f1

    47/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    47

    Slide 5.06 Secondary Electron Multiplier for Extreme Low Pressure

    Purpose: Describe how DDEM or CDEM detectors are used for detecting extremely

    low pressures or gas densities.

    Points of emphasis: The multipliers are used in a pulse counting mode. Each ion is counted. But

    spurious electronic (noise) signals are not counted.

    The useful pressure range for this type of ion detection is about 10-22to about

    10-15

    mbar. The lower limit is determined by dark current of the entiremultiplier. This can be as low as 0.01 counts /sec with a cooled detector. The

    upper limit is determined by counting electronics of about 100 Mhz to 1 Ghz.

    and/or the multiplier output dispersion of 1-10 ns.

    Ion counting requires more complicated electronics.

  • 8/11/2019 Mod.4. Notes v.f1

    48/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    48

    Slide 5.07 Diagram of an Ion Counting System

    Purpose: To show a block diagram of the type of electronics used for ion pulse

    counting.

    Points of emphasis: Cooling the electron multiplier decreases the number of non-ion produced

    pulses.

    Most ion counting systems are used in conjunction with short pulse time of

    flight spectrometers.

    Small gain changes in the multiplier are generally not critical to the ion count.

    Large gain changes in the multiplier are compensated for by adjusting the

    discrimination level.

    ======================= End Section 5 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    49/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    49

    Section 6. MASS ANALYZERS

    Slide 6.01 Mass Analyzers Readily Available

    Purpose: To describe briefly the history of the mass analyzers used in partial pressureanalysis and the dominance of the quadrupole analyzer for the past several years.

    Points of emphasis:

    There were a large variety of PPAs available and in use during the early use of

    partial pressure analysis.

    Presently the quadrupole analyzer is used for nearly all partial pressure

    analysis.

    Niche applications may still use the Magnetic Sector or Time of Flight type of

    analyzers.

    Most analytical mass spectrometry is still done with magnetic sector type of

    instruments.

  • 8/11/2019 Mod.4. Notes v.f1

    50/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    50

    Slide 6.02 Magnetic Sector Analyzer

    Purpose: To show a diagram of a simple magnetic sector mass analyzer.

    Points of emphasis:

    The particular analyzer shown in the diagram is a 60 sector. Other possibledeflection angles of resonant ions are possible e.g. 45, 90, and 180 sectors.

    Note that masses lower than the resonant ion mass strike the inner wall of the

    analyzer tube. While those of mass greater than the resonant ion mass strike

    the outer wall of the analyzer tube.

  • 8/11/2019 Mod.4. Notes v.f1

    51/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    51

    Slide 6.03 Operation of the Magnetic Sector Spectrometer

    Purpose: To describe the operation of the magnetic sector analyzer.

    Points of emphasis:

    Most inexpensive magnetic sector instruments use a permanent magnet andscan the mass range by varying the energy of the ions injected into the

    magnetic field.

    Most magnetic sector PPAs have fixed slit widths, and thus have a fixed

    sensitivity and resolution that is mass dependent.

    Because of the small size of magnetic PPAs, the instruments have two mass

    ranges. The upper mass range is obtained by placing a magnetic shunt on the

    magnet.

    The lower mass range is typically from mass 2 to mass 15. The upper range is

    typically from mass 12 to 100.

    Advanced Topics:

    What are the advantages of ion energy mass scanning?

    What are the advantages of magnetic field mass scanning?

    What other parameter could be used to scan the mass range? Is such scanning

    practical?

  • 8/11/2019 Mod.4. Notes v.f1

    52/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    52

    Slide 6.04 Special Issues with the Magnetic Sector

    Purpose: To describe some of the issues and concerns with using a magnetic sector

    analyzer.

    Points of emphasis: Mechanical slit adjustment becomes troublesome in high and ultra-high

    vacuum applications.

    Stray fields from the magnet can have an impact on either or both the ion

    source and the detector. This is a particular issue when B field massscanning is being used.

    Off axis ion injection contributes to spurious (ghost) mass peaks.

  • 8/11/2019 Mod.4. Notes v.f1

    53/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    53

    Slide 5.05 Time of Flight Mass Analyzer

    Purpose: To show a schematic diagram of the Time of Flight (TOF) analyzer.

    Points of emphasis:

    The drift tube of length L is about 1 meter for a TOF partial pressureanalyzer

    Clearly to get ion separation in 1 meter or less at any reasonable ion injection

    energy, the ion pulse must be short, e.g. 1s or less.

    The injection circuitry and detector electronics must operate in the sub-

    microsecond region.

  • 8/11/2019 Mod.4. Notes v.f1

    54/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    54

    Slide 6.06 Operation of the Time of Flight Mass Spectrometer

    Purpose: To describe the operation of the Time of Flight analyzer, and some of the

    advantages and disadvantages of this analyzer.

    Points of emphasis: The TOF analyzer is the only PPA that measures all of the masses present

    during the time of the injection pulse.

    The TOF is particularly useful in measuring rapidly occurring reactions. For

    example: intermediates and by products from explosions, materials ablatedfrom a surface by a laser pulse, progress and products of combustion, etc.

    A compromise between the injection energy and the electronic injection and

    detection electronics must always be considered. Low injection energy leadsto the possibility of space charge driving ions, particularly those of larger

    mass, out of the detector. Whereas greater injection energy means the drifttimes are shortened and the response of the detector must be faster.

  • 8/11/2019 Mod.4. Notes v.f1

    55/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    55

    Slide 6.07 Quadrupole Mass Filter

    Purpose: To show a diagram of the quadrupole type of analyzer.

    Points of emphasis:

    Cross sections of the quadrupole rods that solve the Mathieu equations arehyperbolic. Nearly all commercially available quadrupole analyzers use

    circular cross section rods.

    The value of r0 is modified to compensate for the rod cross section shape

    change.

    Rod length of commercially available quadrupoles varies from a few

    centimeters to about !meter.

    Paul and his students built and successfully operated a quadrupole analyzer

    with 10 meter long rods. The resolution of such an instrument was between

    10-5and 10-6.

  • 8/11/2019 Mod.4. Notes v.f1

    56/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    56

    Slide 6.08 Quadrupole Operation

    Purpose: To describe the basic operation of the quadrupole analyzer.

    Points of emphasis:

    Ions traversing the quadrupole fields follow paths that are solutions to theMathieu Differential equations. These are the same equations that are used to

    focus ions in high-energy accelerators such as an alternating gradient

    synchrotron.

    Stable R.F. and D.C. power supplies are required for successful quadrupoleoperation.

    The radio frequency used with most instruments is of the order of 10 Mhz.

    The spectra produced by quadrupoles that use voltage scan linear with time are

    m/q linear with time scans.

    The injection energy of ions has little influence on the spectra produced.

    Injection energies of a few to a few tens of electron volts are common.

  • 8/11/2019 Mod.4. Notes v.f1

    57/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    57

    Slide 6.09 Stability Diagram for the Quadrupole Mass Filter

    Purpose: To show a diagram of a stable solution to the Mathieu differential equation,

    and point out that resonant ions in a quadrupole traverse this region in going from the

    ion source to the detector.

    Points of emphasis:

    Ions follow paths that oscillate in both the X and Y direction.

    Resonant ions follow paths where the magnitude of the oscillations are

    confined to the region within the rod structure, and thus pass from the ionsource to the detector.

    Ions with m/q less than resonance have confined oscillations in Y direction,

    but oscillations in the X direction increase in magnitude until the ions strikethe X rods and the ions are collected (neutralized).

    Ions with m/q greater than resonance have confined oscillations in the X

    direction, but oscillate in the Y direction with increasing amplitude until they

    strike the Y rods and are lost.

  • 8/11/2019 Mod.4. Notes v.f1

    58/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    58

    Slide 6.10 Location in Time of Mass Peaks for Various Sweep Parameters

    Purpose: To show a scan parameter plot and describe how it has led to the

    ascendancy of the quadrupole for partial pressure analysis.

    Points of emphasis: K is the sweep or mass scan parameter e.g. magnetic field intensity or voltage.

    The top line shows the mass scan as the constant voltage ratio D.C./R.F. is

    increased linearly with time in a quadrupole analyzer.

    The middle line shows the mass scan for a linear-with-time B field scan in a

    magnetic sector analyzer. This scan spacing is also how ions arrive in time at

    the detector in a Time of Flight analyzer.

    The bottom line gives the mass scan for a linear-in-time ion injection energy

    (voltage) scan in a magnetic sector analyzer.

    The easier to read and interpret top line generated by a quadrupole in

    conjunction with the ability to electrically adjust the resolution-sensitivitytrade-off have led to the dominance of the quadrupole analyzer for partial

    pressure analysis.

    Non-linear-with-time scans of the scan parameter are possible and have been

    used to give linear-with-time mass spectra from magnetic sector analyzers.Such scans do not resolve the resolution-sensitivity adjustment issues in the

    magnetic sector analyzer.

  • 8/11/2019 Mod.4. Notes v.f1

    59/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    59

    Slide 6.11 Quadrupole Issues

    Purpose: To show some important issues encountered when using commercially

    available quadrupoles.

    Points of emphasis: Although quadrupole analyzers are the dominant instruments used for partial

    pressure analysis, they do have some shortcomings.

    At pressures above about the mid 10-5 mbar range most quadrupoles suffer

    from serious non-linear output. Small quadrupoles exist that extend thepressure range an order higher, but these typically have high background noise

    so they cannot measure below about 10-8

    mbar.

    Every electronic drive package must be specifically tuned to a particular

    quadrupole head. This is required because of the small differences in

    capacitance from head to head and cable to cable. These small electronicdifferences lead to distorted radio frequency voltages that are not tolerated by

    the analyzer.

    The total pressure read-out assumes a specific mixture of gasses in the vacuum

    for accuracy. Vacuum systems seldom have the usual residual mixture. The

    total pressure measure is probably no better than plus or minus a half order of

    magnitude in pressure.

    Advanced topics:

    How might a quadrupole be modified so that it would be not only a mass filter

    but also an energy filter?

    What contributes to detector noise from the quadrupole analyzer?

  • 8/11/2019 Mod.4. Notes v.f1

    60/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    60

    Slide 6.12 Symptoms of Improper Tuning

    Purpose: To show some of the characteristics of an improperly tuned quadrupole, and

    describe how improper tuning comes about.

    Points of emphasis: In this particular spectrum the higher mass peaks (m/q >44) are lost due to

    improper matching of quadrupole head and the electronic drive unit.

    The head and drive unit are from the same manufacturer.

    The drive unit was tuned to the head, after this spectrum was taken, and

    spectra showed typical higher mass organic fragments.

    Advanced topics:

    What are other symptoms of improper tuning?

    Is it possible to lose the low mass number peaks rather than the high masses

    due to improper tuning?

  • 8/11/2019 Mod.4. Notes v.f1

    61/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    61

    Slide 6.13 Mislabelled Spectrum for Hydrogen Selenide

    Purpose: To show a spectrum misidentified because of an improper mass scale and

    describe how to this has occurred and how to correct the problem.

    Points of emphasis: This particular spectrum is displayed in bar mode. Thus the peaks do not

    display their characteristic shape.

    The mass spectrum numbering is one number low.

    Advanced topics:

    How could the numbering of the mass spectrum be displaced?

    What information is available to assure that the spectrum is displaced?

    How much of each peak is due to each H2Se species?

    What are the advantages and disadvantages of a bar mode display?

    ======================= End Section 6 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    62/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    62

    Section 7. DATA STORAGE AND DISPLAY

    Slide 7.01 Data Display and Storage

    Purpose: To describe two methods of displaying and storing the same spectrum.

    Points of emphasis:

    Linear amplitude displays of mass spectra are particularly useful when only

    the major constituents of the vacuum are of interest.

    With a linear amplitude display it is important to assure that the largest peak is

    fully displayed i.e. it does not saturate the amplifier with too large a signal.

    Logarithmic amplitude displays of mass spectra are useful when minor

    constituents may play a significant role in the vacuum.

    Logarithmic data take up half the computer space that the same range in

    amplitude of linear data occupy.

    The full dynamic range of an analyzer (typically 5 to 7 orders) can be

    displayed on a single logarithmic mass spectrum.

  • 8/11/2019 Mod.4. Notes v.f1

    63/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    63

    Slide 7.02 Logarithmic Display of m/q 1 - 72

    Purpose: To display a spectrum using a multi-decade logarithmic amplifier.

    Points of emphasis:

    Each major division e.g. 60 to 70 is a decade on the logarithmic amplifier. The dynamic range of data on this plot is at least 6 decades. The amplifier

    used to obtain this data was a 7-decade amplifier.

    Assign a mass number to each mass peak. Be careful with this.

    This spectrum and the next one (7.03) were taken on the same pressure-stable

    vacuum system. They were taken within 3 minutes of each other. Theyshould be viewed together.

    Advanced topics:

    What are the constituents of each mass peak

    There is at least one non-integer mass peak present on this spectrum. Can youidentify its location and composition?

  • 8/11/2019 Mod.4. Notes v.f1

    64/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    64

    Slide 7.03 Linear Display of Data in Slide 7.02, m/q 1 - 69

    Purpose: To show a linearly displayed spectrum of the same data as displayed in

    7.02.

    Points of emphasis: Two mass peaks appear on this spectrum that do not appear on 7.02. Which

    are they? Why dont they appear on 7.02?

    A number of peaks that appeared on 7.02 do not appear on this spectrum.

    Which are they? Why dont they?

    Point out where this spectrum would fall on the logarithmic spectrum (7.02).

    Identify by mass number and composition the peaks that appear on this

    spectrum.

  • 8/11/2019 Mod.4. Notes v.f1

    65/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    65

    Slide 7.04 Block Diagram of Data Acquisition with Feedback Control

    Purpose: To give an illustration of how PPA data may be used as feedback control

    information in process control.

    Points of emphasis: The dotted line is the connection between the partial pressure analyzer i.e.

    sensor and the control functions in the process chamber.

    Establishing the proper feedback and control algorithms is a non-trivial

    undertaking for any process control system.

    Making the dotted line operate properly is usually the most expensive and

    difficult part of an automatically controlled process.

    Such systems do exist to control oxidation in semiconductor processing.

    In experimental and non-repetitive processes a human often acts as the dotted

    line and the data analysis part of the diagram.

    ======================= End Section 7 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    66/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    66

    Section 8. SOME APPLICATIONS

    Slide 8.01 Air Leak Spectrum

    Purpose: To show a PPA spectrum taken from a vacuum system with a large air leak.

    Points of emphasis:

    Point out the peaks from the major constituents of air, i.e. nitrogen, oxygen,

    water, argon and carbon dioxide.

    Point out the fragments and multi-charged species associated with the air

    peaks.

    Advanced topics:

    What mass numbers are due to contaminant peaks?

    Identify and justify your identification of the contaminants.

  • 8/11/2019 Mod.4. Notes v.f1

    67/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    67

    Slide 8.02 Leaky N2Backfill Valve

    Purpose: To show a spectrum of a vacuum system with another type of leak.

    Points of emphasis:

    This spectrum clearly does not indicate the presence of an air leak as can benoted by the absence of many of the constituents of air present in the previous

    spectrum.

    Many high and ultra-high vacuum systems are backfilled with dry nitrogen or

    argon in order to make the return to the vacuum condition easier.

    After many operations of backfill valves they often develop small leaks that

    will eventually need to be repaired as in this case.

    Identify all of the mass peaks due to nitrogen in this spectrum. There are more

    than two.

    Identify the remaining peaks in the spectrum.

  • 8/11/2019 Mod.4. Notes v.f1

    68/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    68

    Slide 8.03 Vacuum System Bakeout

    Purpose: To show a PPA spectrum of a vacuum system during bake-out with some

    peculiar species present.

    Points of emphasis: Identify the mass peaks in the spectrum.

    Note the hydrocarbon fragments.

    Note the large peaks due to zinc at masses 64, 66 and 68.

    Advanced topics:

    What is the evidence that the peaks at m/q=64, 66 and 68 are due to zinc?

    Do other possibilities exist? What are they?

    Do multiple possibilities exist for the other mass peaks in the spectrum? What

    are they?

    ======================= End Section 8 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    69/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    69

    Section 9. QUANTITATIVE AND QUALITATIVE DATA

    Slide 9.01 Qualitative vs. Quantitative Data

    Purpose: To describe how a PPA is used to acquire quantitative partial pressure data.

    Points of emphasis:

    A source of high purity gas is required for calibrating the sensitivity of a PPA

    to that gas.

    It is possible to use certified high purity gas mixtures to calibrate for several

    species.

    Calibrations should be made before and after making measurements of

    unknowns until reliable stable performance of the instrument can be assured.

    Calibration procedures as developed and refereed by reputable, knowledgeable

    organizations should be used to calibrate partial pressure analysis systems.

    See as an example the AVS Recommended Practice listed in the bibliography

    (Slide 10.02).

    Advanced topics:

    In calibrating for methane one notices that the m/q=17 peak amplitude is 2.5%

    of the m/q=16 peak amplitude at a presumed high purity methane pressure of

    1X10-6mbar. Does this have an impact on the calibration for methane?

    If it does have an impact, what is its magnitude?

    What would the expected impact be when calibrating at a presumed pressure

    of 1X10-7mbar?

  • 8/11/2019 Mod.4. Notes v.f1

    70/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    70

    Slide 9.02 Diagram of an Orifice Flow Calibration System

    Purpose: To show a block diagram of a recommended calibration system for

    acquiring quantitative partial pressure information.

    Points of emphasis: The diagram is taken from the AVS Recommended Practice for PPA

    calibration (Slide 10.02).

    The orifice flow procedure is one of several methods for calibrating a PPA.

    In all procedures some sort of standard certified by a standards organization is

    required. The certified standard may be a gauge calibrated for pressure of a

    particular gas or a gas flow certified standard, as is used in the orifice flow

    calibration system.

    Using the method shown in the diagram, in addition to having gas flow

    standards for calibrating the gases of interest, the orifice must be carefully

    fabricated and measured in order to have an accurate calibration.

    The value of C is gas and temperature dependent.

    ======================= End Section 9 ====================

  • 8/11/2019 Mod.4. Notes v.f1

    71/71

    Module 4: Partial Pressure Analyzers and Analysis IUVSTA Visual Aids Program

    Section 10. BIBLIOGRAPHY/REFERENCES

    Slide 10.01 Bibliography I (Props of Elements, Isotopes & Compounds)

    Slide 10.02 Bibliography II (Vacuum)

    Slide 10.03 Bibliography III (PPA Spectra Interpretation)

    Slide 10.04 Bibliography IV (Instrumentation)

    Purpose: These slides display a bibliography of information pertaining to all Slides

    from 1.01 through 9.02.

    Points of emphasis:

    The first reference material covers general data on physical and chemical

    properties of elements and compounds. Information contained in thesereferences should be useful in determining parent peak masses as well as

    isotopic peak masses.

    The second set of references concerns information of a general vacuum nature.Basic operating principles of some PPAs and vacuum gauges are given.

    Collisions of electrons, ions and neutrals with gasses and surfaces are

    described in some detail.

    The third set of references is about recommended calibration procedures for

    gauges and PPAs. Some additional information on operating conditions and

    principles for PPAs and gauges is given, particularly in Lecks book.

    The fourth set of references deals with interpreting mass spectra. Referencespectra are given. Caution should be used when using reference spectra.

    Although the location of m/q peaks for various species is useful theamplitudes quoted will most certainly be quite different for various PPAs.

    The last set of information gives references with more detailed information

    about the various currently used PPAs as well as those from the past. Other

    instrumentation detailed information is also included.

    The intent is not that the bibliography included here be all inclusive, but rather

    that it be a start for further investigation and incentive to gain greater

    understanding of the power and usefulness of partial pressure analysis.

    ======================= End Section 10 ====================

    ================== End Module 4================

    !