10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory Compliance Expert Vaisala, Inc Page 2 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala Review Mapping Regulations. Discuss 5 Rules for Sensor Placement. Review Current Guidance on Sensor Placement. Goals

Upload: others

Post on 13-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

1

Modern Rules for an Old Practice:

Sensor Placement for GMP Temperature Mapping

Paul DanielSenior Regulatory Compliance ExpertVaisala, Inc

Page 2 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

�Review Mapping Regulations.

�Discuss 5 Rules for Sensor Placement.

�Review Current Guidance on Sensor Placement.

Goals

Page 2: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

2

Page 3 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Page 4 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� Health Canada

– GUI-0069 – Guidelines for Temperature Control of Drug Products during Storage and Transportation

� ISPE

– Good Practice Guide: Cold Chain Management

Recent History

Page 3: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

3

Page 5 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� PDA

– Technical Report 58 – Risk Management for Temperature Controlled Distribution

� CDSCO (India)

– Draft Guidelines on Good Distribution Practice for Biological Products

Recent History

Page 6 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� CFDA (China)

– Good Supply Practices for Pharmaceutical Products

� EMA (Europe)

– (2013/C 68/01) Good Distribution Practice of Medicinal Products for Human Use

� USP 36

– Chapter <1079> Good Storage and Distribution Practices for Drug Products

Recent History

Page 4: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

4

Page 7 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

�The regulations don’t tell us how to map, or even to map.

�Regulatory Summary

1. Determine if environmental parameters affect product quality.

2. Do stability testing to determine appropriate storage specifications.

3. Ensure that storage spaces are controlled to meet specifications.

�Mapping and monitoring is industry GMP response to #3.

�Therefore, regulations aren’t very helpful for sensor placement!

Regulations and Mapping

Page 8 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

�Tradition and Precedence

� What has been done at your facility before?

�Equipment

� What equipment do you already have?

�Science

� What makes sense based on scientific principles?

� Guidance

Resources for Sensor Placement

Page 5: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

5

Page 9 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

TOP

FRONT

SIDE

Page 10 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

5 Rules

Map the Extremes.

Map in 3 Dimensions.

For Large Spaces, Map Storage Only.

Identify and Address Variables

If it’s worth mapping, its worth monitoring.

Page 6: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

6

Page 11 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 1: Map the Extremes.

�Geometric

� Place sensors at each end of the storage area.

�Conditions

� Apply validation principle of worst cases.

Page 12 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Parts of a Cube

Corner – 3 planes

Edge – 2 planes

Side – 1 plane

Space – 0 planes

Page 7: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

7

Page 13 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Parts of a Cube

Corner – 3 planes

Space – 0 planes

Rule 1: Map the Extremes

Page 14 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Corollary 1A: If ≤ 2m3, use 9 + 1.

TOP

FRONT

SIDE

Page 8: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

8

Page 15 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

9 spots = worst case?

�Air Flow:

� Corners have least air circulation.

� Center has most air circulation.

�Heat Exchange:

� Corners have the most exposure (3 planes)

� Center has the least exposure (0 planes)

Page 16 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

How many? Where? < 20m3

TOP

FRONT SIDE

Page 9: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

9

Page 17 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

How many? Where? < 20m3

TOP

FRONT

SIDE

Page 18 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

How many? Where? < 20m3

TOP

FRONT

SIDE

Page 10: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

10

Page 19 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

How many? Where? < 20m3

TOP

FRONT

SIDE

Page 20 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

How many? Where? < 20m3TOP

FRONT

SIDE

Page 11: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

11

Page 21 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Corollary 1B: If ≤ 20m3, use 15 +1.

TOP

FRONT

SIDE

Page 22 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� ISPE Good Practice Guide: Cold Chain Management (2011)

Guidance describes 9 +1 and 15 +1

Page 12: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

12

Page 23 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 2: Map in 3 Dimensions.

3 planes.

Left to right.

Page 24 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 2: Map in 3 Dimensions.

3 planes.

Top to bottom.

Page 13: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

13

Page 25 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 2: Map in 3 Dimensions.

3 planes.

Front to back.

Page 26 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Corollary 2A: If ≥ 20m3, use Stacks of 3.

High

Middle

Low

Page 14: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

14

Page 27 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 28 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 15: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

15

Page 29 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 30 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 16: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

16

Page 31 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 32 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Corollary 2A: If ≥ 20m3, use Stacks of 3.

High

Middle

Low

Page 17: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

17

Page 33 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Corollary 2B: Remove sensors if possible.

�Remove “extra” sensors.

� 15 +1 diagram.

� With “Stacks of 3, in 3-D”, it would have 27 sensors.

Page 34 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 18: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

18

Page 35 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Stacks of 3, in 3 dimensions.

Page 36 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

“Completing a three-dimensional temperature profile shouldbe achieved by measuring points at not less than three

dimensional planes in each direction/axis—top-to-bottom, left-to-right, front-to-back…”

USP 36 Chapter <1079> Good Storage and Distribution

Practices for Drug Products (2013)

More published support....

Page 19: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

19

Page 37 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 3: If ≥ 20m3, map storage only.

�As the space gets larger, it is not necessary to map hallways and access areas.

�Map only the shelves and areas

where product is actually stored.

�Requires controls to prevent storage in wrong areas.

� SOPs

� Training

� Signs

Page 38 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

“Completing a three-dimensional temperature profile shouldbe achieved by measuring points at not less than three

dimensional planes in each direction/axis—top-to-bottom, left-to-right, front-to-back, where product will be present.”

USP 36 Chapter <1079> Good Storage and Distribution

Practices for Drug Products (2013)

More published support....

Page 20: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

20

Page 39 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

�Rule 1 – Map the Extremes

� Corollary 1A – If ≤ 2m3, use 9+1.

� Corollary 1B – If ≤ 20m3, use 15+1

Rule Summary (so far…)

Page 40 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

�Rule 1 – Map the Extremes

� Corollary 1A – If ≤ 2m3, use 9+1.

� Corollary 1B – If ≤ 20m3, use 15+1

�Rule 2 - Map in 3 Dimensions

� Corollary 2A – If ≤ 20m3, use Stacks of 3

� Corollary 2B – Remove Sensors if Possible

Rule Summary (so far…)

Page 21: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

21

Page 41 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

�Rule 1 – Map the Extremes

� Corollary 1A – If ≤ 2m3, use 9+1.

� Corollary 1B – If ≤ 20m3, use 15+1

�Rule 2 - Map in 3 Dimensions

� Corollary 2A – If ≤ 20m3, use Stacks of 3

� Corollary 2B – Remove Sensors if Possible

�Rule 3 - If ≥ 20m3, map storage only.

Rule Summary (so far…)

Page 42 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� “…additional points may be needed depending on airflow sources/characteristics, shelving (storage locations), external

temperature sources, and previous experience with similar units and their thermal behavior.”

– ISPE Good Practice Guide: Cold Chain Management (2011)

It’s just guidance. It’s a place to start.

Page 22: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

22

Page 43 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

�The following factors should be considered during temperature mapping of storage locations:

� Size of the space

� HVAC Locations

� Sun-facing walls

� Low ceilings or roofs

� Geography

� Airflow

� Outside temperature variability

� Weekly workflow variation

� Equipment movement

� Storage patterns of product

� Equipment capabilities

� SOPs.

More from USP 36 Chapter <1079>...

Page 44 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 4: Identify Variables

�Volume

�Temperature ∆

�Height

�Exterior Walls

�Construction Materials

�Doors and Windows

�Lighting

�Gradients

�HVAC Vents and Returns

�Air Circulation

�Control Sensors

�Energy Sources

�Racks and Shelving

�Traffic Patterns

�Human Factors

Page 23: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

23

Page 45 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Volume

Page 46 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Temperature ∆

Page 24: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

24

Page 47 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Height

Page 48 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Exterior Walls

Page 25: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

25

Page 49 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Construction Materials

Page 50 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Doors and Windows

Page 26: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

26

Page 51 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Lighting

Page 52 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Gradients

Page 27: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

27

Page 53 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

HVAC Vents and Returns

Page 54 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Air Circulation

Page 28: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

28

Page 55 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Control Sensors

Page 56 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Energy Sources

Page 29: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

29

Page 57 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Racks, Shelves, and Product

Page 58 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Traffic Patterns

Page 30: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

30

Page 59 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Human Factors

Page 60 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 4: Identify Variables

�Place sensors near representative variables!

�Adjust ideal “grid” to intersect

variables.

Page 31: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

31

Page 61 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

TOP

FRONT (South Wall)

SIDE

Page 62 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

TOP

FRONT (South Wall)

SIDE

Page 32: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

32

Page 63 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

TOP

FRONT (South Wall)

SIDE

Page 64 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

TOP

FRONT (South Wall)

SIDE

Page 33: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

33

Page 65 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

TOP

FRONT (South Wall)

SIDE

Page 66 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Tem

pera

ture

Distance (X)

Distance Between Points

Page 34: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

34

Page 67 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Tem

pera

ture

Distance (X)

Distance Between Points

Page 68 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Tem

pera

ture

Distance (1/2X)

Distance Between Points

Page 35: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

35

Page 69 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Tem

pera

ture

Distance (1/2X)

Distance Between Points

Page 70 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Tem

pera

ture

Distance (1/4X)

Distance Between Points

Page 36: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

36

Page 71 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Tem

pera

ture

Distance (1/8X)

Distance Between Points

Page 72 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Tem

pera

ture

Distance (1/16X)

Distance Between Points

Page 37: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

37

Page 73 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

TOP

FRONT (South Wall)

SIDE

Page 74 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

Rule 5: If it is worth mapping…It is worth monitoring.

�Find the hot and cold spots and monitor them.

�Select the right monitoring solution.

�Validate the monitoring system.

Page 38: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

38

Page 75 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

“Temperature monitoring… documents the temperatures during the qualification and operation. It allows conclusions on

whether the product is stored and shipped within the approved temperature range. Therefore, an appropriate temperature

monitoring strategy should be selected and qualified.”

PDA Technical Report No. 58: Risk Management for Temperature Controlled Distribution (2012)

Monitoring....

Page 76 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� Rule 1 – Map the Extremes

� Corollary 1A – If ≤ 2m3, use 9+1.

� Corollary 1B – If ≤ 20m3, use 15+1

� Rule 2 - Map in 3 Dimensions

� Corollary 2A – If ≤ 20m3, use Stacks of 3

� Corollary 2B – Remove Sensors if Possible

� Rule 3 - If ≥ 20m3, map storage only.

� Rule 4 – Identify Variables

� Rule 5 – If it is worth Mapping, it is worth Monitoring

Rule Summary

Page 39: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

39

Page 77 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� Its always best to measure air temperature.

� Worst case scenario provides best challenge.

“It is normal practice to use air temperature as the reference source for the stored [product]…” and is ‘the worst case representation of the stored product temperatures.”

-ISPE Good Practice Guide: Cold Chain Management (2011)

What about product temperature?

Page 78 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� The same concepts apply, but concerned with sources of moisture.

� Use the same number of sensors unless you have an expert who understands humidity theory.

“Where relative humidity is a critical factor, then humidity sensors could be located in the same locations – or fewer locations used with the impact of the temperature considered for other locations – as absolute humidity will be very similar throughout the space.”

-ISPE Good Practice Guide: Cold Chain Management (2011)

What about mapping humidity?

Page 40: Modern Rules for an Old Practice - Vaisala · 2017-08-23 · 10/15/2013 1 Modern Rules for an Old Practice: Sensor Placement for GMP Temperature Mapping Paul Daniel Senior Regulatory

10/15/2013

40

Page 79 / 2013 / Sensor Placement for Validation Mapping – GxP Webinar / ©Vaisala

� Reviewed Mapping Regulations.

� Explored 5 Rules for Sensor Placement for Mapping Storage Areas.

� Reviewed Mapping Guidance.

Summary