modified 9-lead diagnostic ecg using the physio-control lp10 “old dog tricks” gary denton
TRANSCRIPT
![Page 1: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/1.jpg)
Modified 9-lead Diagnostic ECG
Using the Physio-Control LP10“Old dog tricks”
Gary Denton
![Page 2: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/2.jpg)
Test
Confirm option is activated
Press RECORD for 1.0 - 2.0 seconds DIAG should appear on screen.
![Page 3: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/3.jpg)
If DIAG does not appear...
YOU MUST LOAD THE OPTION
![Page 4: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/4.jpg)
To activate diagnostic quality option...
• Locate LP 10 programming key
![Page 5: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/5.jpg)
• Place battery into slot one.• Turn power off• Insert LP 10 programming key into AUX port
![Page 6: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/6.jpg)
Turn Power Switch to BATT 1 & Remove key
![Page 7: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/7.jpg)
Verify o01 appears on status display
Press ECG SIZE to change option status to o01
![Page 8: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/8.jpg)
Verify 1 appears in upper right screen
Press QRS VOL until # 1 appears in upper right screen
![Page 9: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/9.jpg)
Set Option
Press CAL, BEEP will sound. Option is now loaded
![Page 10: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/10.jpg)
Confirm option is activated
Confirm option is activated
Press RECORD for 1.0 - 2.0 seconds DIAG should appear on screen.
![Page 11: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/11.jpg)
Obtaining the Diagnostic ECG
Go directly to these steps if DIAG option is already activated on your
LP10
![Page 12: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/12.jpg)
Place Electrodes
![Page 13: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/13.jpg)
Place Lead Wires
![Page 14: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/14.jpg)
Select Lead 1
![Page 15: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/15.jpg)
Select DIAG optionThis activates the recorder
Press RECORD for 1.0 - 2.0 seconds DIAG should appear on screen.
![Page 16: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/16.jpg)
Scroll through leads 1,2,3 (recorder will be on)
![Page 17: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/17.jpg)
Watch for baseline to stabilize
Obtain one full cardiac cycle for each lead
![Page 18: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/18.jpg)
Leave switch in lead 3 positionManually obtain V-leads
![Page 19: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/19.jpg)
Move from V-1 through V-6
![Page 20: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/20.jpg)
Watch for baseline to stabilize
Obtain one full cardiac cycle for each lead
![Page 21: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/21.jpg)
Stop recorder when complete
![Page 22: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/22.jpg)
Mark leads with pen
![Page 23: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/23.jpg)
Cut and Paste ECG
![Page 24: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/24.jpg)
Diagnostic 9 Lead
Advantages
• Can be done with existing equipment
• Provides adequate diagnostic picture for AMI
• Requires no special equipment
![Page 25: Modified 9-lead Diagnostic ECG Using the Physio-Control LP10 “Old dog tricks” Gary Denton](https://reader030.vdocument.in/reader030/viewer/2022032523/56649d8c5503460f94a731ae/html5/thumbnails/25.jpg)
Diagnostic 9 Lead
Disadvantages
• Does not produce a “standard format” ECG
• Take more time than conventional 12-lead
• Harder to mount and label
• Often rejected by ED staff
• Lacks leads aVR, aVL, and aVF
• No computer interpretation