module 3 use, calibration and maintenance of laboratory equipment 1

67
Module 3 Use, calibration and maintenance of laboratory equipment 1

Upload: alexa-gardiner

Post on 11-Dec-2015

216 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Module 3 Use, calibration and maintenance of laboratory equipment 1

Module 3 Use, calibration and maintenance

of laboratory equipment

1

Page 2: Module 3 Use, calibration and maintenance of laboratory equipment 1

Learning objectives

At the end of this module you will be able to: describe briefly the operating characteristics of

biological safety cabinets, centrifuges, autoclaves, inspissators, pH meters, incubators;

correctly use biological safety cabinets, centrifuges, autoclaves, inspissators, pH meters, incubators;

describe the main maintenance operations for biological safety cabinets, centrifuges, autoclaves, inspissators, pH meters, incubators.

2

Page 3: Module 3 Use, calibration and maintenance of laboratory equipment 1

Content outline

Principles, use and maintenance of:

• biological safety cabinet (BSC)• centrifuge• autoclave• inspissator (coagulator)• pH meter• incubator

3

Page 4: Module 3 Use, calibration and maintenance of laboratory equipment 1

Biological safety cabinet (BSC)

A ventilated contained area providingprotection for the operator and theenvironment against infectious aerosolsduring the handling of hazardous microorganisms

The most important equipment in all diagnostic mycobacteriology laboratories is the BSC.

4

Page 5: Module 3 Use, calibration and maintenance of laboratory equipment 1

HEPA filter

HEPA = high-efficiency particulate air

The HEPA filter traps and remove 99.97% of airborne particles of diameter 0.3 µm or more.

5

Page 6: Module 3 Use, calibration and maintenance of laboratory equipment 1

• BSCs should be ducted or vented to the outside.

• No recirculation into the room is allowed.

• The BSC should be connected to a UPS (Universal Power Supply).

BSC class I and II – main differences

BSC class I and II

BSC class Minimal facevelocity(m/s)

Air flow (%)

Re- Exhaustedcirculated

Exhaust system

I 0.38 0 100 Hard duct

IIA1 0.38 70 30 Thimble connection

IIA2 0.51 70 30 Thimble connection

IIB1 0.51 30 70 Hard duct

IIB2 0.51 0 100 Hard duct6

Page 7: Module 3 Use, calibration and maintenance of laboratory equipment 1

Class I BSC Protection of the user and the environment

Air is ducted to the outside of the building.

HEPA filter

7

Page 8: Module 3 Use, calibration and maintenance of laboratory equipment 1

Class II BSC

Protection of the user, the environment and the product

Class IIA BSC Class IIB BSC

8

Page 9: Module 3 Use, calibration and maintenance of laboratory equipment 1

Laminar flow hood for product (not user) protection

• Air blows towards workers.• Used only to protect product for sterile preparations

(media preparation).

Laminar vertical flowHorizontal airflow

9

Page 10: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC location

A B

CD

A , B – not recommendedD, C – recommended

10

Page 11: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – operations for correct use• Switch on the BSC airflow and the UV light (if

present) 5–15 minutes before use.• Check the airflow velocity on the display.• Complete the logbook, recording details of operation

of the BSC.• Switch off the UV lamp.

Never work with the UV lamp on..• Disinfect the work surface, interior walls  and interior

window surface with 70% ethanol or other suitable compounds.Do not use corrosive substances..

11

Page 12: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – rules for correct use• Organize all items needed for work in the BSC to avoid

frequent displacements.• Organize the work to minimize arm movements. • Do not overcrowd the working area – this disturbs the

airflow. • Conduct all manipulations within the BSC as far towards

the back of the unit as possible (at least 15cm from the front grille).

• In class II BSCs, never allow the front grille to be covered with anything.

• Do not use large open flames in the BSC.• Do not accumulate waste in the BSC; remove it when

activities are finished.12

Page 13: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – after use

• Autoclave the waste.

• Wipe down the inside of the BSC and work surface with 70% alcohol.

• Switch on the UV light for 1 hour (if the lamp is in good working order).

• Switch off the BSC fan.

13

Page 14: Module 3 Use, calibration and maintenance of laboratory equipment 1

Certification and maintenance

Regular certification and maintenance, with replacement of filters, are needed and represent a major challenge in low-income countries. If maintenance is not carried out regularly, the filters may become clogged and tubercle bacilli may be blown into the face of the operator. A BSC that is not well maintained is more hazardous than protective.

14

Page 15: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – evaluation of performancePerformance of the BSC should be evaluated by the manufacturer or a qualified professional, with specific, well-calibrated equipment.

15

Page 16: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – maintenance• Modern cabinets are equipped

with airflow indicators and warning devices.

• Maintenance, with replacement of filters, should be carried out by professionals when the airflow falls below the minimum recommended level.

The BSC should not be used until this maintenance has been carried out.

Out of order

16

Page 17: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – maintenanceAirflow  smoke pattern tests are performed to determine:

– whether the airflow along the entire perimeter of the work access opening is inward;

– whether airflow within the work area is downward with no dead spots or refluxing;

– whether ambient air passes onto or over the work surface;

– whether there is refluxing to the outside at the window wiper gasket and side seals.

17

Page 18: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – maintenance

A cabinet leak test is carried out:– to determine whether exterior surfaces of all

plenums, welds, gaskets, and plenum penetrations or seals are free of leaks;

– before initial installation or annually on fully installed cabinets

18

Page 19: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – fumigation• Decontamination of the BSC with formaldehyde gas.

• Hydrogen peroxide.

• Fumigation should be performed only by professionals.

System DECONTAKIT 19

Page 20: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – fumigation

Fumigation of the BSC should be carried out:

– before release of the BSC for use after a major biohazardous spill;

– before replacement of HEPA filters;– before repair work requiring access to the sealed plenum;– before service or replacement of the circulation fan or

components;– before maintenance work in contaminated areas;– before performance tests requiring entry into contaminated

areas;– before movement of the BSC to another laboratory or area of

use;– before changing work programmes, e.g. to non-TB work;– before release of the BSC for resale or salvage.

20

Page 21: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – logbook

21

Page 22: Module 3 Use, calibration and maintenance of laboratory equipment 1

BSC – maintenance logbook

22

Page 23: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – selection of proper equipment for TB culture

• The centrifuge should:

• Have aerosol-free (O-ring sealed) swing buckets.

• Have locking mechanism for protection to prevent opening before rotation has ceased.

• Be able to generate 3000g.

• Be refrigerated.

23

Page 24: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – rotors

24

Page 25: Module 3 Use, calibration and maintenance of laboratory equipment 1

Inadequate centrifuges and tubes

25

Page 26: Module 3 Use, calibration and maintenance of laboratory equipment 1

Prevent accidents by regular maintenance

26

Page 27: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – use• Inspect the centrifuge chamber for liquid or signs of

corrosion. • Check that there are shock-absorbing pads in the bottom

of the centrifuge buckets. • Balance the sample tubes and insert them in the buckets

in the appropriate position.–Never add water to a specimen to balance paired tubes

–Use an empty tube filled with water as a “balance” .

• Select the required time and speed.• Stop centrifuge immediately if any abnormal noise is

noticed.• Never operate a centrifuge with the lid open.• Open the sealed buckets in the BSC.

27

Page 28: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge tubes – adaptors

• Centrifuge tubes must tolerate g-forces of at least 3000g.

• Centrifuge tubes must be used with appropriate rubber or plastic cushions matched to the tube and bucket-holder.

• Use only adaptors recommended by the manufacturer of the centrifuge in use.

28

Page 29: Module 3 Use, calibration and maintenance of laboratory equipment 1

RCF, g, RPM

RCF (or g) = relative centrifugal force

rpm = revolutions per minute

RCF = 1.12 * 10−6 * r * v2

or

RCF (g) = 1118 x r x (v/1000)2

r = radius in mm v = speed in rpm

rpm = 1000 *√RCF/1.12 * r

Always work in RCF, not rpm: newcentrifuges can convert automatically.

r (mm)

29

Page 30: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – installation

• Ensure that the centrifuge is installed on a rigid, flat, level surface.

• Because centrifuges produce vibration during operation, they must not be installed next to balances or other sensitive equipment.

• Allow sufficient free space around the centrifuge for ventilation to prevent overheating.

• Initial calibration should be performed only by a qualified service technician .

30

Page 31: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – daily maintenance

• Wipe the inside bowl with disinfectant solution and rinse thoroughly.

• If a refrigerated centrifuge is turned off at night, open the top to allow the bowl to dry. During the day when the unit is under refrigeration, keep the top closed to avoid condensation and ice build-up.

31

Page 32: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – monthly maintenance

• Clean the centrifuge housing, rotor chamber, rotors and rotor accessories with a neutral cleaning agent.

• Clean plastic and non-metal parts with a fresh solution of 5% sodium hypochlorite (bleach).

32

Page 33: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – annual maintenance

Annual service should be carried out by a qualified service technician: centrifuge brushes, timer, speed and electrical leaks should all be checked.

33

Page 34: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – logbook

34

Page 35: Module 3 Use, calibration and maintenance of laboratory equipment 1

Centrifuge – maintenance logbook

ITEM IDENTIFICATION

Equipment CENTRI-FUGE

Brand name

Purchase date

Model/type

Location within lab.

Serial no.

Warranty expiration date

Manufacturer

Address

Tel. Contact person

Technical service representative

Tel.

PERIODICITY :

Date Maintenance operation Operator

FAILURE EVENTS

Date Event Corrective action taken

Operator

35

Page 36: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave

The autoclave using saturated steam under pressure is the most efficient means of:•sterilization in a diagnostic TB laboratory;•decontamination of biological material consisting of infectious waste.

36

Page 37: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave

For optimum function:

• All of the air in the chamber should be replaced by steam.

• The temperature must be 121 ºC.

• Materials to be sterilized must be packed loosely.

37

Page 38: Module 3 Use, calibration and maintenance of laboratory equipment 1

Tips for loading the autoclave

• Only appropriate containers should be used.• Discarded cultures should be in solid-bottomed

containers no more than 40 cm deep.• Leave a large air space around each container.• Never cover containers.• If autoclaving closed containers, add water to

generate steam.

Separate autoclaves should be used for sterilization of solutions or glassware (clean materials) and for decontamination of infectious materials.

38

Page 39: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclaves suitable for laboratory use

Pressure-cooker laboratory autoclave

Autoclave with air discharge by gravity displacement

39

Page 40: Module 3 Use, calibration and maintenance of laboratory equipment 1

Examples of autoclaves

40

Page 41: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave – use• Prepare material for autoclaving with thermosensitive paper.

• Do not load material to be sterilized with material to be decontaminated.

• Fill the bottom of the autoclave with distilled water.

• Open the air outlet valve.

• Turn on the heating.

• Close the outlet valve.

• Do not touch the drainage tap or the outlet or the safety valve while heating under pressure.

• When the required time is up, turn off the heating completely.

• When the temperature falls below 100 ºC, open the outlet valve slowly.

41

Page 42: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave – use• Never unscrew the lid clamps and open the lid

until the hissing sound has stopped.• Leave the sterilized/decontaminated material to

cool before removing it from the autoclave.• Check whether the autoclave tape has turned

black and the covering paper has turned brown. If they have not, the material is not decontaminated. Check the autoclave for malfunction.

• Biological indicators should be used after 40 hours of use.

42

Page 43: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave – suggested protocols

• Solid material for sterilization: 121 ºC (appropriate pressure 115 kPa), 20 min.

• Solid material for decontamination: 121 ºC (appropriate pressure 115 kPa), 30 min.

• Liquid material for sterilization: 15 min at 121 ºC if “liquid cycle” is available.

43

Page 44: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave – maintenance

• Check door gaskets for cracks and pitting.

• Check for proper operation of door clamps and door locks.

• Inspect valve discs and seats for signs of wear or cutting.

• Check pressure-release safety valves and thermometers.

44

Page 45: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave – logbook

45

Page 46: Module 3 Use, calibration and maintenance of laboratory equipment 1

Autoclave – maintenance and incidents logbook

46

Page 47: Module 3 Use, calibration and maintenance of laboratory equipment 1

Inspissator• Essential when egg media are “home-made” (made in the

laboratory).

47

Page 48: Module 3 Use, calibration and maintenance of laboratory equipment 1

Inspissator – use

• Carefully control the amount of heating.• Ensure the required temperature has been

reached before loading the tubes• Heat the tubes at constant temperature of

80–85 ºC for 45 minutes.• Tubes or bottles loaded at an angle of 5–

10º ( 10° ).• Always check the quality of media after

cooling.48

Page 49: Module 3 Use, calibration and maintenance of laboratory equipment 1

Inspissator – maintenance

• Check temperature at each use.• Clean after preparation of each batch of culture

media.

49

Page 50: Module 3 Use, calibration and maintenance of laboratory equipment 1

Inspissator – logbook

50

Page 51: Module 3 Use, calibration and maintenance of laboratory equipment 1

Inspissator – maintenance and incidents logbook

51

Page 52: Module 3 Use, calibration and maintenance of laboratory equipment 1

pH meter

52

Page 53: Module 3 Use, calibration and maintenance of laboratory equipment 1

pH meter – use

• Calibrate the pH meter before use at three pH values (pH 4.0, 7.0 and 10.0), using the calibration solutions provided by the manufacturer.

• Immerse the electrode in the test solution.

• Rinse the electrode after use.

• Cover the pH meter after use.

53

Page 54: Module 3 Use, calibration and maintenance of laboratory equipment 1

pH meter – calibration

• The pH meter must be calibrated according to the manufacturer’s instructions.

• Calibration must be performed once daily or, in case of infrequent use, at least on the day of pH testing. It must be performed before the first measurement of the day.

• Standardize with pH 4.0 and 7.0 buffers before each test or series of tests.

• Discard contaminated or cloudy standard buffers.

• Calibration results are acceptable if the pH values of the reference buffer solutions are within 0.1 pH units of the expected pH.

54

Page 55: Module 3 Use, calibration and maintenance of laboratory equipment 1

pH meter – Maintenance

• Keep the instrument clean. Cover it after each use.

• Rinse the electrode after use.

• Make sure that the electrode is always filled with electrolyte according to the manufacturer’s instructions.

• Do not touch the electrode membrane, which can be easily damaged.

• A glass electrode should be kept immersed in a standard salt solution for long-term storage.

• Calomel electrodes must be kept in a potassium chloride (KCl) buffer solution when not in use.

• Date buffer solutions and discard when unsatisfactory.55

Page 56: Module 3 Use, calibration and maintenance of laboratory equipment 1

Calibration solution Solution test

Date Temp.(ºC)

Electrode serial no.

pH 4 pH 7 pH8* Signature Solution/buffer

pH read Signature

pH-meter – calibration and record sheet

*pH 10 can be used if commercially purchased56

Page 57: Module 3 Use, calibration and maintenance of laboratory equipment 1

pH-meter – logbookITEM IDENTIFICATION

Equipment pHmeter Brand name

Purchase date

Model/type

Location within the lab

Serial no.

Warranty expiration dateManufacturer

Address

Tel. Contact person

Technical service representative

Tel.

FAILURE EVENTS

Date Event Corrective action taken

Operator

57

Page 58: Module 3 Use, calibration and maintenance of laboratory equipment 1

IncubatorsIncubators ensure optimum growth conditions for culture of M. tuberculosis.

Incubators are available in various sizes, from small (benchtop) models to large closets up to several hundred liters. Walk-in incubator rooms with circulating fans to obtain an homogeneous temperature are also frequently used.

58

Page 59: Module 3 Use, calibration and maintenance of laboratory equipment 1

Incubator – use

• Keep door(s) closed to prevent heat loss to the environment.

• Ensure that rack positions are clearly marked.

• Switch the incubator off when not in use.

• Do not overload.

59

Page 60: Module 3 Use, calibration and maintenance of laboratory equipment 1

Incubator – maintenance

• Record temperatures daily on appropriate forms.

• Check that the actual temperature corresponds with the thermostat setting.

• Clean every 14 days with 70% ethanol.

• Clean immediately after any infectious spills.

60

Page 61: Module 3 Use, calibration and maintenance of laboratory equipment 1

Incubator – calibration

Calibrate:

–before use;–after temperature changes have been detected and rectified;–following a power failure;–after cleaning of spillages.

61

Page 62: Module 3 Use, calibration and maintenance of laboratory equipment 1

Incubator – calibration• Ensure that the door is closed and that the incubator is

switched on.• Set the required temperature using the temperature

control and leave the incubator to run for 1 hour.• Place a thermometer in the centre of the incubator with

the probe away from the heating element.• Take the temperature reading after 30 minutes. If the

temperature is not 36 ± 1 ºC, adjust the control. • Repeat this process every 30 minutes until the required

temperature is reached.• Continue to take readings until two consecutive readings

(30 minutes apart) are 36 ± 1 ºC.

62

Page 63: Module 3 Use, calibration and maintenance of laboratory equipment 1

Incubator – maintenance form

63

Page 64: Module 3 Use, calibration and maintenance of laboratory equipment 1

Incubator logbook

Trimester : Year :

Month Temp. Visa Month Temp. Visa Month Temp. Visa

°C (Initials) °C (Initials) °C (Initials)

1 1 1

2 2 2

3 3 3

4 4 4

5 5 5

6 6 6

7 7 7

8 8 8

9 9 9

10 10 10

11 11 11

12 12 12

13 13 13

14 14 14

15 15 15

16 16 16

17 17 17

18 18 18

19 19 19

20 20 20

21 21 21

22 22 22

23 23 23

24 24 24

25 25 25

26 26 26

27 27 27

28 28 28

29 29 29

30 30 30

31 31 31

64

Page 65: Module 3 Use, calibration and maintenance of laboratory equipment 1

True and false exercise

1. The biological safety cabinet is the most important equipment for TB culture laboratories.

2. Only class II BSCs can be used for TB culture.

3. Centrifuge speed should always be expressed in rpm.

4. Centrifugation may produce hazardous aerosols.

5. Checking the autoclave regularly is unnecessary if it is used only for decontamination.

6. Equipment maintenance is crucial for optimal performance.

7. Maintenance can always be performed in-house.65

Page 66: Module 3 Use, calibration and maintenance of laboratory equipment 1

Module review: take-home messages

1. Equipment is expensive: all equipment must be treated with care and properly maintained.

2. Quality-control checks on equipment should be run regularly and results should be recorded on appropriate forms.

3. The biological safety cabinet is the most important item of equipment for TB culture under safe conditions.

4. Read carefully and always follow manufacturers’ instructions for routine work and maintenance.

5. Familiarize yourself with all working parts of all the instruments.

6. Call for help when troubleshooting any equipment malfunction.

66

Page 67: Module 3 Use, calibration and maintenance of laboratory equipment 1

Self-assessment

• Describe the differences between the two types of BSC.• What precautions must you take when working in a

BSC? • Describe the operations to be undertaken before starting

a centrifuge. • Explain the principles of sterilization using an autoclave.

67