Understand the components & the operating principles
of the anaesthesia machine, namely, gas supply, O2
safety devices, and flowmeters.
Know how to deal with problems arising from these
components.
![Page 5: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/5.jpg)
An anaesthesia system consists of the various
components that communicate with each other
during the administration of inhalation anaesthesia.
A thorough understanding of these parts is essential
to the safe practice of anaesthesia.
![Page 6: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/6.jpg)
![Page 7: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/7.jpg)
1. O2 & N2O Sources: consisting of a pipeline supply
source and a cylinder supply source. The cylinder
source is regulated from 2,200 to approximately 45
Psig, and the N2O cylinder source is regulated from
745 to approximately 45 Psig.
2. Fail-safe valve: is a safety device located downstream
from the N2O supply source. This valve shuts off or
proportionally decreases the supply of N2O if the O2
supply pressure decreases. Recent machines have an
alarm device to monitor the O2 supply pressure.
![Page 8: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/8.jpg)
3. Second-stage O2 regulator: in most Ohmeda machines
have located downstream from the O2 supply source. It
is adjusted to a precise pressure level, such as 14
Psig.
4. Flow control valves: are an important anatomic
landmark because they divide the anaesthesia machine
into two parts.
5. Machine outlet check valve: in many Ohmeda
machines have located between the vaporizers and the
common gas outlet.
![Page 9: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/9.jpg)
1. The high–pressure circuit: is the part of the machine
that is confined to the cylinders and their primary
pressure regulators.
2. The intermediate–pressure circuit: is the part of the
machine that begins at the regulated cylinder supply
sources downstream the flow control valve.
3. The low–pressure circuit: is the part of the machine
that extends from the flow control valves to the
common gas outlet.
1. The high–pressure circuit: is the part of the machine
that is confined to the cylinders and their primary
pressure regulators.
2. The intermediate–pressure circuit: is the part of the
machine that begins at the regulated cylinder supply
sources downstream the flow control valve.
3. The low–pressure circuit: is the part of the machine
that extends from the flow control valves to the
common gas outlet.
![Page 10: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/10.jpg)
1. The high-pressure circuit: consists of those parts
which receive gas at cylinder pressure:
Hanger yoke (including filter and unidirectional valve).
Yoke block.
Cylinder pressure gauge.
Cylinder pressure regulators.
![Page 11: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/11.jpg)
2. The intermediate – pressure circuit: receives gases
at low, relatively constant pressures (37–55 psig,
which is pipeline pressure).
Ventilator power inlet.
Oxygen pressure–failure device (fail-safe) and alarm.
Flowmeter valves.
Oxygen second-stage regulator.
Oxygen flush valve.
![Page 12: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/12.jpg)
3. The low-pressure circuit: includes components
distal to the flowmeter needle:
Valves.
Flowmeter tubes.
Vaporizers.
Check valves (if present).
Common gas outlet.
![Page 13: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/13.jpg)
![Page 14: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/14.jpg)
Most hospitals today have a central piping system to
deliver medical gases such as O2, N2O, & air to the
operating room.
In a survey of approximately 200 hospitals in 1976,
31%% reported difficulties with pipeline systems.
Inadequate oxygen pressure.
Excessive pipeline pressures.
Accidental crossing of oxygen and nitrous oxide
pipelines.
![Page 15: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/15.jpg)
The pipeline inlet fittings are gas-specific Diameter
Index Safety System (D.I.S.S.).
![Page 16: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/16.jpg)
A pipeline pressure gauge
must be located on the
pipeline side rather than on
the machine side of the
check valve. A value of 50
psig, however, does not
guarantee that the pipeline
is supplying the machine.
![Page 17: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/17.jpg)
Anaesthesia machines have reserve E cylinders if a
pipeline supply source is not available or if the
pipeline fails.
![Page 18: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/18.jpg)
Color–coded cylinders are attached to the
anaesthesia machine through the hanger yoke
assembly, which orients and supports the cylinder,
provides a gas – tight seal, and ensures a
unidirectional flow of gases into the machine.
![Page 19: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/19.jpg)
Each hanger yoke is equipped with the Pin Index
Safety System.
A check valve is located downstream from each
cylinder if a double-yoke assembly is used.
![Page 20: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/20.jpg)
The check valve has several functions:
1. It minimizes gas transfer from a cylinder at high
pressure to one with lower pressure.
2. It allows an empty cylinder to be exchanged for a
full one while gas flow continues from the other
cylinder into the machine with minimal loss of
gas.
3. It minimizes leakage from an open cylinder to the
atmosphere if one cylinder is absent.
![Page 21: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/21.jpg)
Each cylinder supply source has a pressure-reducing valve
known as the cylinder pressure regulator.
The O2 cylinder pressure regulator reduces the O2 cylinder
pressure from a high of 2,200 psig to approximately 45 psig.
The N2O cylinder pressure regulator receives pressure of up to
745 psig and reduces it to approximately 45 psig.
![Page 22: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/22.jpg)
The cylinders should be turned off except:
1. During the preoperative machine-checking period .
2. When a pipeline source is unavailable.
![Page 23: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/23.jpg)
O2 & N2O supply sources existed as independent
entities in older models of anaesthesia machines, and
they were not pneumatically or mechanically
interfaced.
Therefore, abrupt or insidious oxygen pressure failure
had the potential to lead to the delivery of a hypoxic
mixture.
![Page 24: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/24.jpg)
Contemporary anaesthesia machines have a number
of safety devices that act together in a cascade
manner to minimize the risk of hypoxia as oxygen
pressure decreases.
![Page 25: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/25.jpg)
Many older anaesthesia machines have a pneumatic
alarm device that sounds a warning when the O2
supply pressure decreases to a predetermined
threshold value, such as 30 psig.
![Page 26: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/26.jpg)
Electronic alarm devices have both audible and visual
alarms. The O2 pressure threshold value for the
Ohmeda Modulus II Plus & the Ohmeda CD is 27 psig,
and for the N.A.Dräger Narkomed 2B, 3, & 4, the value
is 30 ± 3 psig.
![Page 27: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/27.jpg)
It is present in the gas line supplying each of the
flowmeters except that for oxygen. These valves,
controlled by oxygen pressure, shut off or
proportionally decrease the supply pressure of all
other gases (N2O, air, CO2, helium, and N2) as the O2
supply pressure decreases.
![Page 28: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/28.jpg)
The misnomer "fail-safe" has led to the misconception
that the device prevents administration of a hypoxic
mixture.
This is not the case.
![Page 29: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/29.jpg)
They are equipped with the
pressure-sensor shutoff valve.
It is either open or closed with
a threshold pressure of 25 psig.
![Page 30: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/30.jpg)
An O2 pressure > the threshold
value is exerted upon the
mobile diaphragm.
This pressure moves piston,
pin, and valve off the valve
seat.
N2O flow passes freely to the
N2O flow control valve.
![Page 31: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/31.jpg)
They uses the Oxygen Failure Protection Device
(OFPD), which interfaces the O2 pressure with that of
other gases, such as N2O, air, CO2, helium, and N2.
The OFPD is based on a proportioning principle rather
than a threshold principle.
The pressure of all gases controlled by the OFPD will
decrease proportionally with the oxygen pressure.
![Page 32: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/32.jpg)
The OFPD consists of a seat-
nozzle assembly connected
to a spring-loaded piston.
Nozzle
Valve seat
Spring
![Page 33: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/33.jpg)
When the O2 supply
pressure is 50 psig,it
pushes the piston upward,
forcing the nozzle away
from the valve seat.
N2O advances toward the
flow control valve at 50
psig. 50 Psig
50 Psig
N2O
O2
50 Psig
![Page 34: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/34.jpg)
When the O2 pressure is
zero psig, the spring is
expanded and forces the
nozzle against the seat,
preventing N2O flow
through the device.
0 Psig
50 Psig
N2O
O2
0 Psig
![Page 35: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/35.jpg)
When the O2 pressure is
intermediate at 25 psig.
The force of the spring
partially closes the valve.
The N2O pressure
delivered to the flow
control valve is 25 Psig.
25 Psig
50 Psig
N2O
O2
25 Psig
![Page 36: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/36.jpg)
There is a continuum of
intermed.configurations
between the extremes (0
– 50 Psig) of O2 supply
pressure.
25 Psig
50 Psig
N2O
O2
25 Psig
![Page 37: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/37.jpg)
These intermediate valve
configurations are
responsible for the
proportional nature of
the OFPD.
25 Psig
50 Psig
N2O
O2
25 Psig
![Page 38: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/38.jpg)
Most contemporary Ohmeda machines have a second-
stage oxygen pressure regulator set at a specific value
ranging from 12 to 19 Psig.
O2 flowmeter output is constant when the O2 supply
pressure exceeds the set value.
![Page 39: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/39.jpg)
Ohmeda pressure-sensor shutoff valves are set at a
higher threshold value (20 to 30 Psig).
This ensures that O2 is the last gas flow to decrease if
O2 pressure fails, and act together in a cascade manner
to minimize the risk of hypoxia as oxygen pressure
decreases.
![Page 40: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/40.jpg)
Opening the flow control valve allows
gas to travel through the space between
the float and the flowtube [the annular
space].
The indicator float hovers freely in an
equilibrium position, where the upward
force resulting from gas flow equals the
downward force on the float resulting
from gravity at a given flow rate.
![Page 41: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/41.jpg)
These flowmeters are commonly referred
to as constant – pressure flowmeters
because the pressure decrease across
the float remains constant for all
positions in the tube.
![Page 42: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/42.jpg)
It is composed of a flow control
knob, a needle valve, a valve
seat, and a pair of valve stops.
It can receive its pneumatic input
either directly from the pipeline
source (50 Psig) or from a
second-stage pressure regulator.
![Page 43: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/43.jpg)
The location of the needle valve in
the valve seat changes to establish
different orifices when the flow
control valve is adjusted.
Gas flow increases when the flow
control valve is turned counter-
clockwise, and vice versa.
![Page 44: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/44.jpg)
Extreme clockwise rotation results
in damage to the needle valve and
valve seat.
![Page 45: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/45.jpg)
Therefore, flow control valves
are equipped with valve stops to
prevent this occurrence. The
stops come into contact with
each other at zero flow on most
flow control valves.
![Page 46: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/46.jpg)
The Ohmeda Modulus I, Modulus
II, Modulus II Plus, and Ohmeda
8000 are set at an O2 flow rate of
approximately 200 ml.min-1.
![Page 47: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/47.jpg)
On recent Ohmeda machines,
minimum O2 flow results from
incomplete closure of the O2 flow
control valve.
N.A.Dräger machines, the O2 flow
control valve does close
completely.
![Page 48: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/48.jpg)
1. The oxygen flow control knob is
physically distinguishable [fluted,
projects beyond the control knobs
of the other gases, and is larger in
diameter] from other gas knobs.
![Page 49: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/49.jpg)
2. All knobs are color-coded for the
appropriate gas & the name of the
gas is permanently marked on
each.
![Page 50: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/50.jpg)
3. The knobs are recessed or
protected with a shield or barrier
to minimize inadvertent change
from a preset position.
![Page 51: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/51.jpg)
The flowmeter subassembly consists of:
1. The flow tube.
2. The indicator scale.
3. The indicator float with float
stops.
![Page 52: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/52.jpg)
They are made of glass, have a
single taper in which the inner
diameter of the flow tube
increases uniformly from bottom
to top.
![Page 53: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/53.jpg)
Manufacturers provide double flow
tubes for O2 & N2O to provide
better visual discrimination at low
flow rates.
![Page 54: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/54.jpg)
A fine flow tube indicates flow
approximately 1-200 ml.min-1.
A coarse flow tube indicates flow
approximately 1-10 L.min-1.
![Page 55: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/55.jpg)
The two tubes are connected in
series & supplied by a single flow
control valve.
The total gas flow is that shown on
the higher flowmeter.
![Page 56: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/56.jpg)
Some older machines have two
flow tubes for a single gas
arranged in parallel.
Each of the tubes has a flow
control valve.
The total flow is the sum of the
individual flows.
![Page 57: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/57.jpg)
There are different types of bobbins
or floats, including plumb-bob floats,
rotating skirted floats & ball floats.
![Page 58: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/58.jpg)
Flow is read at the top of plumb-bob
and skirted floats and at the center of
the ball on the ball-type floats.
Flow tubes are equipped with float
stops at the top & bottom of the tube.
![Page 59: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/59.jpg)
The upper stop prevents the float
from ascending to the top of the
tube, plugging the outlet & ensures
that the float will be visible at
maximum flows instead of being
hidden in the manifold.
![Page 60: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/60.jpg)
The bottom float stop provides a
central foundation for the indicator
when the flow control valve is
turned off.
![Page 61: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/61.jpg)
The flowmeter scale
can be marked directly
on the flow tube or can
be located to the right
of the tube.
![Page 62: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/62.jpg)
Gradations are closer together at
the top of the scale because the
annular space increases more
rapidly than does the internal
diameter from the bottom to the
top of the tube.
![Page 63: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/63.jpg)
Rib guides are used in N.A.Dräger
flow tubes with ball-type
indicators to minimize the
compression effect.
They are tapered glass ridges that
run the length of the tube.
![Page 64: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/64.jpg)
There are usually 3 rib guides,
which are equally spaced around
the inner circumference of the
tube.
![Page 65: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/65.jpg)
The annular space from the
bottom to the top of the tube
increases almost proportionally
with the internal diameter,
resulting in a nearly linear scale.
![Page 66: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/66.jpg)
1. The flowmeter subassembly for each gas on the Ohmeda Modulus
I, Modulus II, Modulus II Plus, and CD is housed in an independent,
color-coded, pin-specific module.
2. The flow scale and the chemical formula or name of the gas is
permanently etched on the backing to the right of the flow tube.
3. Flowmeter scales are individually hand-calibrated by use of the
specific float to provide a high degree of accuracy.
4. The tube, float, and scale make an inseparable unit. The entire set
must be replaced if any component is damaged.
![Page 67: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/67.jpg)
5. N.A.Dräger does not use a modular system for the flowmeter
subassembly. The flow scale, the chemical symbol, and the gas-
specific color codes are etched directly onto the flow tube. The
scale in use is obvious when two flow tubes for the same gas are
used.
![Page 68: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/68.jpg)
1. Leaks:
It is a substantial hazard because the flowmeters are
located downstream from all machine safety devices
except the O2 analyzer.
It can occur at the O rings between the glass flow tube
& the metal manifold and even in the glass flow tubes.
Gross damage to glass flow tubes is usually apparent,
but subtle cracks & chips may be overlooked, resulting
in errors of delivered flows.
![Page 69: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/69.jpg)
1. Leaks:
Eger et al (1963) demonstrated that in the presence of a
flowmeter leak, a hypoxic mixture is less likely to occur
if the oxygen flowmeter is located downstream from all
other flowmeters.
O2 escapes through the leak & N2O flows toward the
common outlet, particularly at high N2O/ O2 flow ratios.
![Page 70: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/70.jpg)
Ohmeda Mod II Anaesthesia Machine
![Page 71: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/71.jpg)
2. Inaccuracy:
Flow error can occur even when flowmeters are
assembled properly with appropriate components.
1. Dirt or static electricity can cause a float to stick, and
the actual flow may be higher or lower than that
indicated.
2. Sticking is more common in the low flow range
because the annular space is smaller.
![Page 72: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/72.jpg)
2. Inaccuracy:
3. A damaged float can cause inaccurate readings
because the precise relationship between the float
and the flow tube is altered.
4. Back-pressure from the breathing circuit can cause a
float to drop so that it reads less than the actual flow.
![Page 73: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/73.jpg)
2. Inaccuracy:
5. Finally, if flowmeters are not aligned properly in the
vertical position, readings can be inaccurate because
tilting distorts the annular space.
![Page 74: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/74.jpg)
3. Ambiguous Scale:
Before the standardization of flowmeter scales and
the widespread use of oxygen analyzers, at least two
deaths resulted from confusion created by
ambiguous scales.
The operator read the float position beside an
adjacent but erroneous scale in both cases.
![Page 75: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/75.jpg)
3. Ambiguous Scale:
Today this error is less likely to occur because
contemporary flowmeter scales are marked either
directly onto or to the right of the appropriate flow
tube, therefore confusion is minimized.
![Page 76: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/76.jpg)
3. Proportioning System:
Manufacturers have equipped newer machines with
proportioning systems in an attempt to prevent
delivery of a hypoxic mixture.
N2O & O2 are interfaced either mechanically or
pneumatically so that the minimum O2 concentration
at the common outlet is 25%%.
![Page 77: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/77.jpg)
3. Proportioning System:
4. Ohmeda Link-25 Proportion Limiting Control System.
It allows independent
adjustment of either
valve, yet automatically
intercedes to maintain a
min.25%% O2 concentration
with a maximum N2O/O2
flow ratio of 3:1.
The N2O & O2 flow control
valves are identical.
![Page 78: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/78.jpg)
A 14-tooth sprocket is
attached to the N2O flow
control valve, and a 28-
tooth sprocket is attached
to the O2 flow control
valve.
3. Proportioning System:
4. Ohmeda Link-25 Proportion Limiting Control System.
A chain physically links
the sprockets.
![Page 79: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/79.jpg)
When N2O flow control
valve is turned through
two revolutions, O2 flow
control valve will revolve
once because of the 2:1
gear ratio.
3. Proportioning System:
4. Ohmeda Link-25 Proportion Limiting Control System.
![Page 80: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/80.jpg)
It is used on the
N.A.Dräger Narkomed 2A,
2B, 3, and 4.
it is a pneumatic N2O/O2
interlock system designed
to maintain a FG O2 conc.
of at least 25 ±± 3%%.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
![Page 81: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/81.jpg)
It limits N2O flow to prevent delivery of a hypoxic
mixture.
This is unlike the Ohmeda Link-25, which actively
increases O2 flow.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
![Page 82: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/82.jpg)
It is composed of an O2
chamber, a N2O chamber
& a N2O slave control
valve; all are inter-
connected by a mobile
horizontal shaft.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
O2 chamberN2O chamber
![Page 83: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/83.jpg)
The pneumatic input into
the device is from the O2
& N2O flowmeters.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
![Page 84: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/84.jpg)
These flowmeters are
unique because they have
specific resistors, located
downstream from the flow
control valves, which create
back-pressures directed to
the O2 & N2O chambers.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
Resistor
![Page 85: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/85.jpg)
The value of the O2 flow
tube’s resistor is 3–4 times
that of the N2O flow tube’s
resistor& the relative value
of these resistors
determines the value of
the controlled FG O2 conc.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
Resistor
The backpressure in the
O2 and N2O chambers
pushes against rubber
diaphragms attached to
the mobile horizontal
shaft.
Diaphragms
Movement of the shaft
regulates the N2O slave
control valve, which
feeds the N2O flow
control valve.N2O slave control valve
![Page 86: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/86.jpg)
If the O2 pressure is proportionally > the N2O
pressure, the N2O slave control valve opens more
widely, allowing more N2O to flow.
As the N2O flow is increased manually, the N2O
pressure forces the shaft toward the O2 chamber.
The valve opening becomes more restrictive and
limits the N2O flow to the flowmeter.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
![Page 87: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/87.jpg)
The N2O slave control valve is closed because of
inadequate O2 back-pressure.
3. Proportioning System:
4. N.A. Dräger Oxygen Ratio Monitor Controller (ORMC).
![Page 88: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/88.jpg)
1.1. Wrong Supply Gas: Wrong Supply Gas:
In the Link-25 system, the N2O and O2 flow control
valves will continue to be mechanically linked, and
a hypoxic mixture will proceed to the common
outlet.
1.1. Wrong Supply Gas: Wrong Supply Gas:
The oxygen rubber diaphragm of the ORMC will
recognize adequate " O2 " pressure, and flow of
both the wrong gas plus N2O will result.
1.1. Wrong Supply Gas: Wrong Supply Gas:
The oxygen analyzer is the only machine monitor
that will detect this condition in both systems.
![Page 89: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/89.jpg)
2.2. Defective Pneumatics Or Mechanics: Defective Pneumatics Or Mechanics:
Pneumatic integrity in the Ohmeda system depends
on properly functioning second–stage pressure
regulators. A N2O/O2 ratio other than 3:1 will result if
the regulators are not precise.
The chain connecting the two sprockets must be
intact; a 97% N2O concentration can occur if it is cut
or broken.
![Page 90: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/90.jpg)
2.2. Defective Pneumatics Or Mechanics: Defective Pneumatics Or Mechanics:
In the N.A.Dräger system, a functional OFPD is
necessary to supply appropriate pressure to the
ORMC.
The mechanical aspects of the ORMC [e.g. rubber
diaphragms, flow tube resistors & N2O slave
control valve] must likewise be intact.
![Page 91: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/91.jpg)
3.3. Leaks Downstream: Leaks Downstream:
The oxygen analyzer is the only machine safety
device that can detect the problem.
N.A.Dräger system recommends a preoperative
+ve pressure leak test to detect such a leak.
Ohmeda recommends a preoperative –ve pressure
leak test because of the check valve located at the
common outlet.
![Page 92: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/92.jpg)
3.3. Leaks Downstream: Leaks Downstream:
![Page 93: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/93.jpg)
![Page 94: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/94.jpg)
4.4. Inert Gas Administration: Inert Gas Administration:
Administration of a third inert gas, such as helium,
N2, or CO2, can cause a hypoxic mixture because
contemporary proportioning systems link only N2O &
O2.
The oxygen analyzer is the only machine safety
device that can detect the problem.
![Page 95: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/95.jpg)
![Page 96: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/96.jpg)
Understanding of the components & the operating
principles of the gas supply system, safety devices,
and flowmeters, is a part from the corner stone in
anaesthesia practice.
![Page 97: mohamedrefaateltahan@yahoo.com Understand the components & the operating principles of the anaesthesia machine, namely, gas supply, O 2 safety](https://reader038.vdocument.in/reader038/viewer/2022110322/56649d205503460f949f4bd1/html5/thumbnails/97.jpg)
1. Andrews JJ: Inhaled Anesthetic Delivery Systems. In:
Miller RD, ed., Anesthesia, 6th ed., Philadelphia, Churchill
Livingston, pp. 273 – 317, 2005.
2. Andrews JJ: Delivery system for inhaled anesthetics. In:
Barash PG, Cullen BF, Stoelting RK, eds., Clinical
Anesthesia, 4th ed., Philadelphia, Pennsylvania, Lippincott
Williams and Wilkins, pp 567 – 594, 2001.