MR Safety. Dr William Holmes. Effects of MR. Direct biological effects to tissues of the body by, 1. Static magnetic field (ie large superconducting magnet) 2. Magnetic gradients (only when scans are running) 3. RF fields (only when scans are running) Indirect effects on metal - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/1.jpg)
MR Safety
Dr William Holmes
![Page 2: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/2.jpg)
Effects of MR
• Direct biological effects to tissues of the body by, 1. Static magnetic field (ie large superconducting magnet)
2. Magnetic gradients (only when scans are running)
3. RF fields (only when scans are running)
• Indirect effects on metal
1. Strong attractive force on metal object/implants.
2. Heating effects on implants
![Page 3: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/3.jpg)
1. Direct effects of static field
• Many studies have been performed on fruit flies to primates.
• No long term effects shown at clinical field levels (up to 4T).
• At fields >2T rapidly turning one’s head can result in ‘magnetophosphenes’ due to torque on retinal cone cells, also metallic taste and dizziness
Important, do not rapidly move your head in proximity to magnet bore. The concern is you may get dizzy, fall over and then bang your head.
![Page 4: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/4.jpg)
Courtesy Prof. A. Geim
![Page 5: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/5.jpg)
2. Direct effects of magnetic gradients
• These can induce electrical currents in the peripheral nervous system, causing twitching and pain.
• Fast sequences such as EPI, turbos and diffusion weighted gradients can be the worst.
Important, avoid being in the magnet room when scans are being run.
Your can generally tell scans are running by the loud clicking sound.
If you do need to be in the magnet room during scanning, then wear ear protection!!.
![Page 6: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/6.jpg)
• People vary markedly in terms of their sensitivity to such gradient effects.
Important
Seek advice from staff before entering magnet room when scans are running. Especially the first time!
2. Direct effects of magnetic gradients
![Page 7: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/7.jpg)
3. Direct effects of Radiofrequency (RF)
RF radiation can cause an heating effect on biological tissue.
Metal implants can concentrate RF energy and cause local burns
Non-MR compatible pulse oxymeter used on baby on ward
Transponder implant.
Important
1. Check list for implants
2. Avoid being in Magnet room when scans are run. Especially close to bore!
3. Ask staff before entering Magnet room during scanning.
![Page 8: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/8.jpg)
Aneurysm clips
Implants
Implant RF ID chips
Pacemakers
Aneuysm clip
Transponders
Metal plates or pins
Vascular clamp, coil or stent
Infusion pump
Electronic implant (Neurostimulator)
Metal from accident or military service (bullet)
Artificial joints, limbs
Metal fragments in the eye
Important,
On check list identify ANY implant you may have. Ask staff for advice.
![Page 9: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/9.jpg)
You will be asked to fill out and sign an,
“MRI safety checklist form”
It is your responsibility to read and respond accurately.
If you answer “yes” to any question, discuss with staff member.
![Page 10: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/10.jpg)
“Missile” effect of magnetic field
IMPORTANT
7.0 Tesla is a VERY VERY strong magnetic field!!!
http://www.youtube.com/watch?v=vIQaGt_fkkw
Watch the youtube video showing a pair of scissors being pulled into a 1.5Tesla magnet.
Imagine a 7tesla magnet!!!
![Page 11: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/11.jpg)
“Missile” effect of magnetic field
IMPORTANT
• Never take metal object into Magnet room.
• If you need to take an object into Magnet room, ask staff to check the object and authorise it to enter the Magnet room.
• The red/white line marks safe distance limit
![Page 12: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/12.jpg)
![Page 13: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/13.jpg)
Missile injuries in MR units caused by:
• Hairpins
• Earrings
• Pens
• Paperclips
• Cigarette lighters
• Pocket knives
• Surgical scissors/clamps
• Wrenches/screwdrivers
• ID Badges
• Radiation badges
• Needles
• Oxygen tanks
• IV poles/pumps
• Step stools
• Floor buffers
• Wheelchairs
• etc. etc. etc.
![Page 14: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/14.jpg)
MRI: Reported Adverse Incidents
• 17 pacemaker/defibrillator deaths• 1 death due to implant migration/rotation• 1 death due to ferromagnetic missile impact• 1 blinding• >200 cases of burns• >100 impact injuries? Are these reports the tip of the iceberg?
17/11/05
![Page 15: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/15.jpg)
Metal in eye
• One case of a blinding due to fragment rotation/migration
• If metal in eye they must be screened first with a plain film x-ray (detects metal down to 0.1x0.1x0.1.mm)
![Page 16: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/16.jpg)
Safety measures in GEMRIC
• Warning signs posted on all approaches to high field area
• ‘Controlled area’ accessed only through coded self-locking doors
• Extensive staff safety checklist• Emergency policies eg
firesuper-conducting magnet quenchheart attack etc.
![Page 17: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/17.jpg)
Emergencies
• Cardiac arrest: patient must be removed from Controlled area before crash team can have access to them
• Fire: Firemen must be taken through checklist before entering magnet room.
• Quench of super-conducting magnet
![Page 18: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/18.jpg)
Super-conducting quenches• If the quench pipe is blocked there can be a
blowback of helium into the magnet room. Helium is not toxic but can displace oxygen causing asphyxiation
• In the event of a quench the magnet suite must be evacuated immediately!.
![Page 19: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/19.jpg)
![Page 20: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/20.jpg)
![Page 21: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/21.jpg)
![Page 22: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/22.jpg)
![Page 23: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/23.jpg)
![Page 24: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/24.jpg)
![Page 25: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/25.jpg)
![Page 26: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/26.jpg)
![Page 27: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/27.jpg)
![Page 28: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/28.jpg)
![Page 29: MR Safety](https://reader035.vdocument.in/reader035/viewer/2022081603/56814347550346895dafbe62/html5/thumbnails/29.jpg)