Series 700 AV/T-S Power Conditioner Designed for Varian Novalis TM TX, Trilogy TM IX with ExacTRAC TM and OBI or Truebeam TM STX, Truebeam TM with ExacTRAC TM and OBI or any system with a high energy accelerator, Exactrac TM and OBI Input Voltages: 208 VAC, 240 VAC, 480 VAC or 600 VAC (60 Hz) Output Voltages: 208/120 VAC, 480/277 VAC and 120 VAC Integrated Input and Output Breakers One Input Breaker - Four Output Breakers Intelligent Voltage Regulation ( + 2.5% Output Typical) Output Voltage Monitoring Internal Bypass Switch Triple Shielded Isolation Transformer Internal TVSS Front Access “Zero Clearance” Cabinet Quadruple Output Power Conditioner with Voltage Regulation Submittal Package and Specifications

Upload: others

Post on 11-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

Series 700 AV/T-S Power ConditionerDesigned for Varian NovalisTM TX, TrilogyTM IX with ExacTRACTM and OBI or TruebeamTM STX, TruebeamTM with ExacTRACTM and OBI or any system with a high energy accelerator, ExactracTM and OBI

Input Voltages: 208 VAC, 240 VAC, 480 VAC or 600 VAC (60 Hz)Output Voltages: 208/120 VAC, 480/277 VAC and 120 VAC

▪ Integrated Input and Output Breakers

▪ One Input Breaker - Four Output Breakers

▪ Intelligent Voltage Regulation (+ 2.5% Output Typical)

▪ Output Voltage Monitoring

▪ Internal Bypass Switch

▪ Triple Shielded Isolation Transformer

▪ Internal TVSS

▪ Front Access “Zero Clearance” Cabinet

Quadruple Output Power Conditioner with Voltage RegulationSubmittal Package and Specifications

112283

Page 2: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

2Quadruple Output Power Conditioner l A&E Specs/Submittal Package

TRANSTECTOR SERIES 700 AV/T-S

Specifications for Quadruple Output 60 Hz Power Conditioner, designed for Varian NovalisTM TX, TrilogyTM IX with ExacTRACTM and OBI or TruebeamTM STX, TruebeamTM with ExacTRACTM and OBI or any system with a high energy accelerator, ExactracTM and OBI.

1.0 SCOPE

This specification covers the electrical characteristics of the Transtector Power Conditioner which provides clean regulated power for the entire NovalisTM TX, TrilogyTM IX with ExacTRACTM and OBI or TruebeamTM STX, TruebeamTM with ExacTRACTM and OBI or any system with a high energy accelerator, ExactracTM and OBI.

2.0 GENERAL

The Power Line Conditioner consists of a front access power cabinet incorporating an all copper, multiple tapped, triple shield isolation/regulation transformer. The ultra low output impedance of the transformer in conjunction with the electrostatic shields assures precision hospital grade performance with excellent noise and transient attenuation. Independently controlled inverse parallel electronic switches for each of the 7 taps per phase provide tight regulation over a wide input range. Linear devices are used for line synchronization to prevent phase shift errors normally associated with simple CT zero current crossing acquisition. The microprocessor control accurately selects the correct tap to maintain the output no greater than + 2.5% of nominal in a typical installation, correcting for voltage disturbances within one cycle. Digital processing technique provides fast and accurate regulation without output voltage over or undershoots.

2.01 MODEL NUMBERSMODEL INPUT VOLTAGE OUTPUT VOLTAGESStandard Models - other configurations are available including 50 Hz

Model Input Voltage Output Voltages8BLNX-160 K (i) -700AV/T-S 208 VAC nominal 208/120 VAC, 480/277 VAC and 120 VAC8CLNX-160 K (i) -700AV/T-S 240 VAC nominal 208/120 VAC, 480/277 VAC and 120 VAC8DLNX-160 K (i) -700AV/T-S 480 VAC nominal 208/120 VAC, 480/277 VAC and 120 VAC8ELNX-160 K (i) -700AV/T-S-C 600 VAC nominal 208/120 VAC, 480/277 VAC and 120 VAC

2.1 AGENCIES

2.1.1 STANDARDS

The systems shall be designed in accordance with:

▪ American National Standards Institute

▪ Institute of Electrical and Electronic Engineers

▪ National Electric Code (NEC)

▪ National Fire Protection Association (NFPA Article 70)

▪ Underwriters Laboratories (UL) 1449, 1012

▪ FCC Article 15, Section J, Class A

▪ ISO 9001

Page 3: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

3

1402-012

2.1.2 LISTINGS

▪ The system shall be listed to C-UL, UL standard UL1012.

▪ The system shall comply to: FCC Article 15, Section J, Class A and ANSI C62.14 (electromagneticcompatibility).

▪ The SPD shall be UL 1449 3rd Edition listed/recognized.

3.0 DYNAMIC ELECTRICAL CHARACTERISTICS

3.1 OPERATING VOLTAGE AND OUTPUTS

The input voltage shall be 208 VAC, 240 VAC or 480 VAC, Delta, three phase, 60Hz. Each unit will be pre-wired at the factory to accommodate the nominal input voltage selected. Units with other input voltages (i.e. 380, 400, 415, 600 VAC) and/or frequencies (50 Hz) are produced upon request.

3.2 LINE VOLTAGE REGULATION Usable Input Line Voltage +15%, -23%.Nominal Input Line Voltage +10% to -15%

3.2.1 OUTPUT VOLTAGE REGULATION + 2.0% to + 2.5% typical.

The design of the system shall indicate that with an input voltage of -10% of nominal, increasing the load to 1000% shall cause the output voltage to fall no lower than -6%.

3.3 OUTPUT VOLTAGES Two (2) separate WYE configured, 3 phase voltage outputs shall be provided as follows:Output #1 shall be 208/120 VAC.Output #2 shall be 208/120 VAC or 480/277 VAC.Output #3 shall be 480/277 VAC.A single (1) phase voltage output shall be provided as follows:Output #4 shall be 120 VAC.

3.4 OUTPUT CONNECTIONSOutput #1: 150 Amp, three (3) pole circuit breaker for Linear Accelerator or a 175 Amp three (3) pole circuit

breaker for TrueBeamTM.Output #2: 100 Amp, three (3) pole circuit breaker if 208 VAC, or a 70 Amp three (3) pole 480 VAC circuit breaker

for the BrainLAB ExacTRACTM generator.Output #3: 60 Amp, three (3) pole circuit breaker for OBI or a 90 Amp three (3) pole circuit breaker for

TrueBeamTM.Output #4: 30 Amp, single (1) pole circuit breaker for power peripherals.

** The TrueBeamTM STx will use either the 175 Amp 208 VAC output or the 90 Amp 480 VAC output.

Refer to chart on page #8 for output configurations.

Page 4: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

4Quadruple Output Power Conditioner l A&E Specs/Submittal Package

3.5 INPUT/OUTPUT WIRINGThe input/output wiring sizes are dependant upon the terminals provided by the circuit breakers.

Input wiring sizes: 208 VAC 300 Amp breaker #1 AWG to #600 KCMIL

240 VAC 250 Amp breaker #3/0 AWG to #350 KCMIL480 VAC 150 Amp breaker #14 AWG to #3/0 AWG600 VAC 100 Amp breaker #14 AWG to #3/0 AWG

Output wiring sizes:208/120 VAC 175 Amp breaker #4 AWG to #4/0 AWG

208/120 VAC 150 Amp breaker #14 AWG to #3/0 AWG208/120 VAC 100 Amp breaker #14 AWG to #3/0 AWG480/277 VAC 90 Amp breaker #14 AWG to #3/0 AWG480/277 VAC 70 Amp breaker #14 AWG to #3/0 AWG480/277 VAC 60 Amp breaker #14 AWG to #3/0 AWG120 VAC 30 Amp breaker #14 AWG to #2 AWG

Refer to chart on page #8 for output configurations.

The ILSCO TA-2/0 terminal allows wire sizes from #14 to 2/0 to be connected to the ground.

3.6 RESPONSE TIME Response time is less than 1/2 cycle.

3.7 CORRECTION TIME The output voltage is corrected within 1 cycle.

3.8 LOAD REGULATION The output is maintained to within 2% of nominal or less, from no load to full load.

3.9 IMPEDANCEOutput #1 and Output #2: impedance shall be less than 2.33% (Measured with the linear accelerator in beam on state @45 KVA with the OBI and ExacTracTM @ 5 KVA continuous) Output #3: impedance shall be less than 1.45% (45.9 mOhms) (Measured with the linear accelerator in ready state @ 20 KVA, ExacTracTM in standby state @ 5 KVA and OBI @ 45.7 KVA (55 A) momentary)

3.10 OPERATING FREQUENCY 60 Hertz + 3 Hertz.

3.11 HARMONIC DISTORTIONLess than 1% THD added to the output waveform under any dynamic linear loading conditions presented to the line regulator.

3.12 TURN-ON CHARACTERISTICS When energized the voltage overshoot is 5% or less of the nominal voltage for less than 1 cycle.

3.13 OVERLOAD RATING 200% for ten seconds. 1000% for one cycle.

Page 5: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

5

1402-012

3.14 NOISE ATTENUATION Common mode noise attenuation is typically 140 dB or greater.Transverse mode noise attenuation is 3 dB down at 1000 Hertz, 40 dB down per decade to below 50 dB with a resistive load.

3.15 AUDIBLE NOISE Not to exceed 55dB measured @1 meter.

3.16 EFFICIENCY Efficiency shall be > 96% typical at full load. Excitation losses shall be less than 1.5% of KVA rating.

3.17 BTU The Power Line Conditioner shall generate no more than 10,230 BTU/Hour in typical use.

3.18 POWER FACTORInput power factor shall be greater than .95 with a resistive load and reflect no triplen harmonics to the utility under non-linear loads.

3.19 LINE TO LINE BALANCE The Power Line Conditioner shall not produce more than a 2% phase to phase unbalance.

3.20 MTBFThe system shall exhibit a MTBF > 10,000Hr.

3.21 ENHANCED TRANSIENT OVER VOLTAGE SURGE SUPPRESSIONAn enhanced surge protection device (SPD) shall be installed parallel to the secondary output of the power line conditioner to provide all mode, bi-directional and bi-polar surge protection. The SPD is rated for 100 KA per phase, 50 KA per mode capacity. (L-L, L-N, L-G, N-G) The surge suppressor shall be listed to UL 1449 3rd edition ratings when subjected to ANSI/IEEE 62.41-1991 category 3 waveforms. Unit shall conform to UL 1283 requirements for electromagnetic interference. Unit shall provide attenuation against EMI/RFI noise up to 50 dB at 1MHz. The surge suppressor is installed on the load side of the transformer, connected in parallel.

4.0 MAIN TRANSFORMER

4.1 BASIC CONSTRUCTION The transformer windings are of all copper conductor construction with separate primary and secondary isolated windings.

4.2 MAGNETICGrain oriented, stress relieved silicon transformer steel is utilized to minimize losses and provide maximum efficiency. Flux density will not exceed 14k gauss.

4.3 INSULATION Class N (200° C) insulation is utilized throughout.

4.4 SHIELDING The transformer has multiple (three) copper shields to minimize inner winding capacitance, transient and noise coupling between primary and secondary windings. Inner winding capacitance is limited to .001 pf or less.

4.5 COOLING The transformer is designed for natural convection cooling. Fans are inside the unit.

Page 6: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

6Quadruple Output Power Conditioner l A&E Specs/Submittal Package

4.6 OPERATING TEMPERATURE The system operating range: 0 to 40 degrees Celsius, 32 to 104 degrees Fahrenheit.

4.7 OPERATING HUMIDITY0-95% relative humidity, non-condensing.

5.0 MAIN INPUT BREAKER

A main input molded case thermal magnetic circuit breaker, rated at 125 % of the full load input current, is furnished as an integral part of the unit. For example, a 300 Amp input breaker will be provided for 208 VAC input, a 250 Amp input breaker will be provided for a 240 VAC input, or a 150 Amp input breaker will be provided for 480 VAC input. A 100 Amp input breaker will be provided for 600 VAC input.

6.0 BY-PASS SWITCH

A manually operated rotary bypass switch provides bypassing of the SCR controlled voltage regulator portion of the Power Line Conditioner. The Power Line Conditioner can be operated in either the on-line or bypassed mode with one turn of the switch. The transformer and surge suppression circuitry remains in the circuit when in the bypass mode. All nominal output voltages, 208/120 and 480/277 VAC three (3) phase 4 wire, and 120 VAC single (1) phase will remain available in the bypass mode. The bypass switch is located on the front of the unit.

7.0 MONITORING

7.1 ALERT LIGHT An indicator light shall annunciate that the output has been disabled by one of the following conditions. (1) Transformer over-temperature.(2) SCR thermal over-temperature.

7.2 INPUT OUT OF RANGE LEDs

Input “Out of Range” red indicating LEDs shall be provided for each phase. If any phase of the input voltage source exceeds the specified +10%, -15% from nominal, the appropriate phase LED shall be illuminated.

7.3 INDICATING LAMPS Output “ON” indicating lamps shall provided for each phase.

7.4 OUTPUT VOLTAGE METERS

Two (2) digital output meters shall be provided. One shall monitor the 208/120 VAC output and the other shall monitor the 480/277 VAC output. Each meter shall monitor and display Line-to-Line and Line-to-Neutral voltages.

Minimum and maximum voltage values shall be recorded in EPROM. High and low output voltage limit alarms shall be programmable and a visual indication given if limits are exceeded.

8.0 CABINET

8.1 TERMINATIONTerminations is front access with input and output connections to the input and output circuit breakers. A ground terminal will also be provided, Ilsco TA-2/0.

The unit is constructed using an isolation transformer and is considered to be a “separately derived system”. It should be grounded in accordance with the NFPA 70 article 250.20 “Alternating-Current Circuits and Systems to be Grounded”, article 250.20 (D) “Separately Derived Systems” and article 250.30 “Grounding Separately Derived Alternating-Current Systems”.

The Output Neutral and Ground connection points are common to both WYE outputs (208/120 VAC & 480/277 VAC).Therefore, it is considered a single, separately derived, power source and should be wired accordingly.

Page 7: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

7

1402-012

8.2 VENTILATIONVentilation originates from the front of the cabinet, exiting through the top and/or sides.

8.3 MOBILITYThe Power Line Conditioner cabinets are equipped with angle iron supports that allow for transport by pallet jack or fork lift. These can be used for mounting unit to the floor in seismic zones.

8.4 ACCESSIBILITYThe Power Line Conditioner will have front access. Access to all wiring inputs, outputs, and breakers will be accessible through the front access door or panels. The output meters and bypass switch will be located on the front of the unit. The back of the unit may be set next to a wall without impeding access. It will also incorporate lift off side panels.

8.5 WEIGHTUnit weight: Approximately 2,194 lbs. (995.18 kg.).

8.6 DIMENSIONS76” Height x 34.5” Wide x 35.875” Deep (193.04 cm. x 87.63 cm. x 91.12 cm.).

9.0 CONTROLS

The control portion of the cabinet containing the circuit boards and connection to the semi-conductor devices is separate from the transformer section and apart from the input/output power connections.

10.0 WARRANTY

Units shall include a comprehensive warranty for the first year, covering all parts and workmanship, inclusive of on site labor and travel expenses in geographical areas covered. Consult factory for details. All units are provided with a standard two (2) year warranty covering parts and workmanship.

11.0 SERVICE

Transtector shall provide immediate phone support/consultation and if possible, same day parts shipment. (contact must be prior to 12:00 PM PST). If necessary, on site service shall be scheduled the same day for service to be conducted within 24 to 48 hours, based on customer requirements. Typical service hours are 8 AM to 5 PM Monday through Friday.

12.0 CONTACT

Rick Ribbeck

Phone: (O) 208-635-6400 Ext. 5867 | (M) 208-762-6112

E-mail: [email protected]

E-mail: [email protected]

Transtector Systems

10701 Airport Dr.

Hayden Lake ID 83835

Page 8: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

8Quadruple Output Power Conditioner l A&E Specs/Submittal Package

OUTPUT KVA INPUT BREAKER SIZE

160 K(I)

300 A @ 208 VAC 250 A @ 240 VAC

150 A @ 480 VAC

100 A @ 600 VAC

WEIGHTS, BTU AND DIMENSIONS UNIT SIZE IN

KVA (I)WEIGHT OPERATIONAL

BTU/HR TYPICALMAXIMUM

BTU/HRDIMENSIONS

160 K(I) 2,194 lbs. 995.18 kg.

5,115* 10,230 76” H x 34.5” W x 35.875” D(193.04 cm. x 87.63 cm. x 91.12 cm.2)

* Stated BTU’s / Hr is at 100% rated load, 100% duty cycle. Operational BTU’s / Hr is typically at 50% of ratedload. Input over current protection provided by others.

INPUT AND OUTPUT CONFIGURATION

INPUT

OUTPUT

NovalisTM TX - TrilogyTM with ExacTRACTM Output Breaker Configurations

TrueBeamTM STx - TrueBeamTM with ExacTRACTM Output Breaker Configurations

BRL #1 BRL #2 (Varian preferred breakers) BRL #6150 A 208 VAC CB for the Linear Accelerator 150 A 208 VAC CB for the Linear Accelerator 150 A 208 VAC for the Linear Accelerator

100 A 208 VAC CB for the BL Generator 70 A 480 VAC CB for the BL Generator 70 A 480 VAC CB for the BL Generator

60 A 480 VAC CB for the OBI 60 A 480 VAC CB for the OBI 60 A 480 VAC CB for the OBI

30 A 120 VAC CB for the BL Computers 30 A 120 VAC CB for the BL Computers 30 A 120 VAC CB for the BL Computers

30 A 120 VAC CB for the VMS Computers

BRL #4 BRL #5150 A 208 VAC CB (Not Used) 175 A 208 VAC CB for the TrueBeam

TM

70 A 480 VAC CB for the BL Generator 70 A 480 VAC CB for the BL Generator

90 A 480 VAC CB for the TrueBeamTM

60 A 480 VAC CB (Not Used)

30 A 120 VAC CB for the BL Computers 30 A 120 VAC CB for the BL Computers

Page 9: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

9

1402-012

SEISMIC CALCULATIONS

Coastal California, Zone 4 Z = 0.4

Equipment Anchorage I = 1.5

Uniform Building Code, Table 160 Cp = 0.75

Cabinet Weight 2100 lbs.

Center of Gravity Height 33 in.

Wp (max) 2,415.0 lbs.

Wp (min) 1,785.0 lbs.

Vertical Force Fp = 0.45 x 2415 = 1,086.8 lbs.

Moment Fp = 0.15 x 2415 = 1,086.8 lbs.

Mo = 33 x 1086 = 35,863 in. lbs.

Corners (a,b) 43.1 in

V1 = V4 = 21.6

Tension = Fp x Cg / V4 = 3,213.6 lbs.

Shear = Wp(max)Fp/4 lbs., each anchor = 603.8

EXAMPLE: <Rawl Power Bolt # 6913>

3/8” embedded 2.5”in minimum 2000 psi concrete

Tension rating of bolt: 5,200 lbs.

Shear rating of bolt: 7,270 lbs.

Interaction = (T/Tbolt) + (S/Sbolt)

Interaction = 0.70

Interaction = < 1 (OK)

Page 10: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

10Quadruple Output Power Conditioner l A&E Specs/Submittal Package

CABINET OUTLINE

8.00

0

T.SZW

AST

REVI

SED

OUTP

UT B

REAK

ER, A

DDED

KNO

CK-O

UTS I

NTO

P, R

ECRE

ATED

IN C

ADRA

FOR

POW

ERW

AY D

ATA

BASE

3/30

/090

84 3

ADDE

D OU

TPUT

MET

ERS A

ND M

OVED

125A

OUT

PUT

BREA

KER

4/28

/200

8D.

CAR

LSON

2UP

DATE

D PR

INT T

O RE

FLEC

T ONL

Y NE

UTRA

L AND

GROU

ND O

UTPU

T BUS

.2/

1/20

08D.

CAR

LSON

1CH

ANGE

D IN

PUT G

ROUN

D LU

G AN

D CH

ANGE

D (1

)20

8V O

UTPU

T CIR

CUIT

BRE

AKER

TO 15

0A.

D. C

ARLS

ON1/

21/2

008

REV.

DESC

RIPT

ION

DATE

BY

REVI

SIONS

OU

TPU

T M

ETER

S

(4)Ø

.875

KN

OC

K-O

UTS

4

BOTT

OM

VIE

W

R

36" C

LEA

RAN

CE

REQ

UIR

ED O

N F

RON

T SI

DE

FOR

SERV

ICE

OF

REG

ULA

TOR

SEC

TIO

NLI

FT O

FF A

CC

ESS

PAN

ELS

REQ

UIR

E C

OM

MO

N H

AN

D T

OO

L FO

R A

CC

ESS.

REA

R PA

NEL

S M

AYB

E RE

MO

VED

TO

DEC

REA

SE D

EPTH

. MU

ST B

E RE

INST

ALL

ED.

NO

TE:

.060

= 1

6GA

. .1

35 =

10G

A.

.048

= 1

8GA

. .1

05 =

12G

A.0

36 =

20G

A.

.075

= 1

4GA

.

Hay

den

Lake

, ID

838

35SU

PERI

OR

SURG

E SU

PPRE

SSIO

N

OU

TPU

T VO

LTA

GES

: 120

/208

VA

C &

277

/480

VA

CIN

PUT

VOLT

AG

ES: 2

08, 2

40, 4

80, O

R 60

0VA

C8(

BCD

E)LN

X-16

0K(i)

-700

AV/

T-S

SERI

ES 7

00A

FRO

NT

AC

CES

S C

ABI

NET

OU

TLIN

E

COPP

ER B

US

OU

TPU

T TE

RMIN

ATI

ON

S

WIR

E RA

NG

E: 2

/0-#

14 S

TR.

INPU

T G

ROU

ND

LU

G (1

) TA

2/0

TO T

OP)

(CU

STO

MER

'S IN

PUT

CO

NN

ECTS

INPU

T C

IRC

UIT

BRE

AKE

R

TERM

INA

TIO

N A

CC

ESS

ARE

ATO

P IN

PUT/

OU

TPU

T

OU

TPU

T FI

LTER

, OU

TPU

T TE

RMIN

ATI

ON

S A

ND

CU

RREN

T TR

AN

SFO

RMER

S.RE

MO

VABL

E TO

P PA

NEL

PRO

VID

ES A

CC

ESS

TO T

RAN

SFO

RMER

,

TERM

INA

TIO

N A

CC

ESS

ARE

AO

PTIO

NA

L SI

DE

OU

TPU

T

TERM

INA

TIO

N A

CC

ESS

ARE

ABO

TTO

M IN

PUT/

OU

TPU

T

DET

AIL

A

AIR

FLO

W

1/18

/200

8D

.CA

RLSO

N

76.0

00

3.00

0

BYPA

SS S

WIT

CH

AIR

INLE

T

14.0

006.

895

23.0

00

15.8

75

3.71

6

5.69

0

FRO

NT

VIEW

4.85

0

8.25

0

4.50

02.

810

35.5

00

RIG

HT

SID

E VI

EWLE

FT S

IDE

VIEW

6.89

5

30.2

71

2.73

9

30.7

50

Ø.5

62 x

1.0

00

2.46

9

3.87

5

34.5

00

REA

R VI

EW

AIR

FLO

W

OU

T O

F RA

NG

E &

MO

NIT

OR

ALE

RTLI

GH

TS

: 1:8

SCA

LETO

LERA

NCE

3O

F1

SHEE

T

PRO

HIB

ITED

.TR

AN

STEC

TOR

IS

WIT

HO

UT

THE

WRI

TTEN

PER

MIS

SIO

N O

FRE

PRO

DU

CTI

ON

IN P

ART

OR

AS

A W

HO

LETR

AN

STEC

TOR.

AN

Y D

RAW

ING

IS T

HE

SOLE

PRO

PERT

Y O

FTH

E IN

FORM

ATI

ON

CO

NTA

INED

IN T

HIS

.

.

MA

TERI

AL:

REV

DW

G.

NO

.

DA

TE

CHEC

KED

DRA

WN

PRO

PRIE

TARY

AN

D C

ON

FID

ENTI

AL

"

SEE D

ETAI

L A

SEE D

ETAI

L A INPU

T BR

EAKE

R SI

ZE30

0 A

@ 2

08 V

AC25

0 A

@ 2

40 V

AC15

0 A

@ 4

80 V

AC10

0 A

@ 6

00 V

AC

BR

L #1

BR

L #2

(Var

ian

pref

erre

d br

eake

rs)

BR

L #6

150

A 20

8 VA

C C

B fo

r the

Lin

ear A

ccel

erat

or15

0 A

208

VAC

CB

for t

he L

inea

r Acc

eler

ator

150

A 20

8 VA

C fo

r the

Lin

ear A

ccel

erat

or10

0 A

208

VAC

CB

for t

he B

L G

ener

ator

70

A 4

80 V

AC C

B fo

r the

BL

Gen

erat

or70

A 4

80 V

AC C

B fo

r the

BL

Gen

erat

or60

A 4

80 V

AC C

B fo

r the

OBI

60 A

480

VAC

CB

for t

he O

BI60

A 4

80 V

AC C

B fo

r the

OBI

30 A

120

VAC

CB

for t

he B

L C

ompu

ters

30 A

120

VAC

CB

for t

he B

L C

ompu

ters

30 A

120

VAC

CB

for t

he B

L C

ompu

ters

30 A

120

VAC

CB

for t

he V

MS

Com

pute

rs

BR

L #4

BR

L #5

150

A 20

8 VA

C C

B (N

ot U

sed)

175

A 20

8 VA

C C

B fo

r the

Tru

eBea

mTM

70 A

480

VAC

CB

for t

he B

L G

ener

ator

70 A

480

VAC

CB

for t

he B

L G

ener

ator

90 A

480

VAC

CB

for t

he T

rueB

eam

TM60

A 4

80 V

AC C

B (N

ot U

sed)

30 A

120

VAC

CB

for t

he B

L C

ompu

ters

30 A

120

VAC

CB

for t

he B

L C

ompu

ters

Nov

alis

TM T

X - T

rilog

yTM w

ith E

xacT

RAC

TM O

utpu

t Bre

aker

Con

figur

atio

ns

True

Bea

mTM

STx

- Tr

ueB

eam

TM w

ith E

xacT

RAC

TM O

utpu

t Bre

aker

Con

figur

atio

ns

Page 11: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

11

1402-012

CABINET OUTLINE

8.00

0

Ø.8

75 K

NO

CK-

OU

TS

4

R

TOP

INPU

T/O

UTP

UT

TERM

INA

TIO

N A

CC

ESS

ARE

A

REM

OVA

BLE

TOP

PAN

EL

OPT

ION

AL

SID

E O

UTP

UT

TERM

INA

TIO

N A

CC

ESS

ARE

A

BOTT

OM

INPU

T/O

UTP

UT

TERM

INA

TIO

N A

CC

ESS

ARE

A

SERI

ES 7

00A

FRO

NT

AC

CES

S C

ABI

NET

OU

TLIN

E8(

BCD

E)LN

X-16

0K(i)

-700

AV/

T-S

INPU

T VO

LTA

GES

: 208

, 240

, 480

, OR

600V

AC

OU

TPU

T VO

LTA

GES

: 120

/208

VA

C &

277

/480

VA

C

SUPE

RIO

R SU

RGE

SUPP

RESS

ION

Hay

den

Lake

, ID

838

35

.036

= 2

0GA

. .0

75 =

14G

A.

.048

= 1

8GA

. .1

05 =

12G

A.0

60 =

16G

A.

.135

= 1

0GA

.

AIR

FLO

WRI

GH

T SI

DE

VIEW

4.75

0

2.81

24.

500

8.25

0

FRO

NT

AIR

OU

TLET

35.5

15

15.8

75

3.71

6

5.75

034

.389

23.0

00

1.87

5

3.36

4

6.89

514

.000

FRO

NT

Ø.562 X 1.000

3.72

36.

895

2.46

9

30.2

501.

954

30.7

50

2.73

9

D.C

ARL

SON

1/18

/200

8

BOTT

OM

VIE

W

TOP

VIEW

: 1:4

SCA

LETO

LERA

NCE

THE

INFO

RMA

TIO

N C

ON

TAIN

ED IN

TH

ISD

RAW

ING

IS T

HE

SOLE

PRO

PERT

Y O

FTR

AN

STEC

TOR.

AN

Y RE

PRO

DU

CTI

ON

IN P

ART

OR

AS

A W

HO

LEW

ITH

OU

T TH

E W

RITT

EN P

ERM

ISSI

ON

OF

TRA

NST

ECTO

R IS

PRO

HIB

ITED

.

SHEE

TO

F

"

PRO

PRIE

TARY

AN

D C

ON

FID

ENTI

AL

DRA

WN

CHEC

KED

DA

TE

DW

G.

NO

.RE

V

MA

TERI

AL:

.

.

Page 12: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to:

12Quadruple Output Power Conditioner l A&E Specs/Submittal Package

CABINET OUTLINE

4

R

SERI

ES 7

00A

FRO

NT

ACC

ESS

CABI

NET

OU

TLIN

E8(

BCD

E)LN

X-16

0K(i)

-700

AV/

T-S

INPU

T VO

LTA

GES

: 208

, 240

, 480

, OR

600V

AC

OU

TPU

T VO

LTA

GES

: 120

/208

VA

C &

277

/480

VA

C

SUPE

RIO

R SU

RGE

SUPP

RESS

ION

Hay

den

Lake

, ID

838

35

.036

= 2

0GA

. .0

75 =

14G

A.

.048

= 1

8GA

. .1

05 =

12G

A.0

60 =

16G

A.

.135

= 1

0GA

.

TOP

VIEW

FRO

NT

D.C

ARL

SON

1/18

/200

8

: 1:4

SCA

LETO

LERA

NCE

THE

INFO

RMA

TIO

N C

ON

TAIN

ED IN

TH

ISD

RAW

ING

IS T

HE

SOLE

PRO

PERT

Y O

FTR

AN

STEC

TOR.

AN

Y RE

PRO

DU

CTIO

N IN

PA

RT O

R A

S A

WH

OLE

WIT

HO

UT

THE

WRI

TTEN

PER

MIS

SIO

N O

FTR

AN

STEC

TOR

IS P

ROH

IBIT

ED.

SHEE

TO

F

"

PRO

PRIE

TARY

AN

D C

ON

FID

ENTI

AL

DRA

WN

CHEC

KED

DAT

E

DW

G.

NO

.RE

V

MA

TERI

AL:

.

.

Page 13: New Series 700 AV/T-S Power Conditioner - Transtector · 2019. 3. 15. · 1402-012 2.1.2 LISTINGS The system shall be listed to C-UL, UL standard UL1012. The system shall comply to: