1

Nikon A1si Training Guide

Page 2: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

2

Table of Contents

Hardware Review…………………………................Pgs 3-19 Fluorescence and Brightfield…….........................Pgs 18-19 Software Review…………………..........................Pgs 20-31 Basic Confocal.……………………….......................Pg 32-37 Spectral Confocal…………………………………….Pg 38-41 Post Acquisition/Image Saving……………………..Pg 42-48

Page 3: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

3

Hardware Guide

Page 4: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

4

Turn On Sequence (off-reverse order)

1. Laser Launch (key switch on front, clockwise) 2. Fluorescence Light Source-switch on front 3. Transmitted Light Supply-switch on front 4. PC 5. Microscope-switch on back right 6. Motorized Stage Controller-switch on back 7. A1 Confocal Controller (push button on top left side)

Page 5: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

5

Main System Components

Page 6: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

6

Lasers and Controller

Scan Head

Fluorescence

light source

Stage controller

Transmitted light

Confocal keypad

Stage Joystick

Page 7: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

7

Stage Joystick *Only button 5 is being used on this system

Page 8: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

Fluorescence Light Source

8

*Shutter on top; ND filter positions 1-32 (1=full light passes, 32=1/32 of light passes)

Page 9: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

9

Confocal Keypad

*Provides external adjustment of laser settings

*Channel Select will allow you to switch between wavelength for adjustment

-software will show a small arrow next to laser being adjusted, i.e. ->488

Page 10: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

10

Confocal Lasers and Controller 405-457-476-488-514-561-640

Page 11: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

11

Microscope: Nikon TiE

Page 12: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

12

Front

Display:

Brightness Adjustment and Display Info

Resets Current Z position to Zero

Perfect Focus Adjustments

(N/A on this system)

Magnification Changer: 1x or 1.5x

*Please default to 1x

Light Path:

Eye-Eyepiece Viewing

L100-Confocal

R100-NOT USED

L80-NOT USED

Page 13: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

13

Left Side Z focus adjustment

Objective Adjustment-will rotate to next objective lens

Turn On/Off Transmitted Light (brightfield)

Transmitted Light Intensity adjustment knob

Focus Knob

Page 14: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

14

Right Side Z focus adjustment

Focus Knob

Fluorescence Shutter

Filter position

Escape-lowers objective

Refocus-raises objective back up

*When “escaped” focus is disabled

Page 15: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

15

Top Sliders

*Push all sliders in the light path as

default

Depending on configuration, usually you will have three or four sliders: 1. ND=Neutral Density (cuts down light) 2. GIF=Green Interference Filter (changes white light to green

for better contrast) 3. D=Diffuser (spreads light out evenly)

Page 16: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

16

Transmitted Light Adjustments Field Aperture for adjusting Kohler Alignment

Condenser Height-set at green arrows

Condenser Prisms-controlled by software

(may differ from this selection)

A-open

DICN2-DIC

PH1-(10x/20x) Phase

Polarizer-push in for DIC and rotate for shading

Centering Screws for adjusting Kohler Alignment

Page 17: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

17

Quick Review on Kohler Alignment

1. Focus on a sample

2. Close down your field aperture

3. Focus your condenser (adjust condenser height)

4. Center your condenser

5. Open field aperture

*Use your 10x objective and put the condenser in the A position

Page 18: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

18

To use fluorescence: 1. Make sure to turn on your

fluorescence light source

2. Place your sample on the stage and choose an objective

3. Select “EYE” from front panel

4. On the right panel, rotate “FL Block” to desired fluorescent filter and open shutter with “Epi Shutter” button

Page 19: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

19

To use brightfield: 1. Place your sample on the stage and

choose an objective

2. Select “EYE” from front panel

3. On the right panel, rotate “FL Block” to “DIA” (diascopic/brightfield)

4. Turn on the diascopic light from the left side panel and adjust the light as needed

*Note: to use a contrast method, you must match

your condenser prism to the appropriate objective; each objective will have writing on the side to denote the condenser prism to be used i.e. “10x, PH1”

Page 20: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

20

Software Guide

Page 21: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

21

General Software Statements As a general reference, the devices connected to the software can be reviewed in the

“Device Manager” under “Devices”. This is a good place to confirm all portions of the system are connected and working properly.

*On this system, you should see the Nikon Ti and the Proscan III as being connected. If there is a break in the chain, make sure the device is physically turned on, then highlight the device and select “Connect” to reconnect.

Page 22: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

22

Good to know…. • Most dialogs can be found under the View

tab (Acquisition/Analysis/Visualization Controls) in Elements

-As a shortcut these can also be found by right clicking in the open space of Elements

• Elements obeys most Windows-based functions, shift/control and Alt are used often; right-clicks are used to display options/properties

• Elements also uses “drag and drop” for overlaying windows/datasets

Page 23: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

23

System Information

*The system information will tell you how long the software has been running for that session and will also tell you what direction you are focusing.

Page 24: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

24

Microscope Control • The Ti Pad can control almost anything

on the microscope from within the software.

• Each button is clickable (i.e., click on 20x to move to the 20x objective).

• The Light Path is dependent on the type of imaging you want to do:

o E100: 100% to the eyepieces

o L100: confocal

o R100: usually fluorescence/brightfield camera or empty

*Remember to select the appropriate Zoom level (1x or 1.5x both on the microscope and software)

Page 25: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

25

LUTS and Histogram The LUTs menu can be used for brightness/contrast adjustment (using triangular sliders) as well as for showing a saturation indicator which can be used to show what pixels are saturated. Use the drop down arrow to select “complementary color”

The traditional Histogram gives similar information as the LUTs but is very useful in determining a confocal “offset” adjustment (see page 42)

Page 26: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

26

ND Experiments *Experiments can be run on time, stage positions, Z-stacks, wavelength switching and/or image stitching.

-The next few slides will explain each dimension.

Page 27: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

27

Time Lapse

• To add a timelapse, check off the checkbox for “Time”.

• Add a time phase by checking one box; you can change the interval (how often to image) and the duration (total time). Alternatively you can input a number of loops (# of images required).

• You can add additional time phases to be run in sequence.

• The “X” will reset the time phases.

Page 28: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

28

XY-Stage Positions

• To add XY positions, check off the checkbox for “XY Pos”.

• Anytime you would like a position to be remembered, check off a box. This will add the XY coordinate as well as Z (if checkbox for Include Z is checked) or the PFS offset (if PFS is set).

*There is also a checkbox for “move stage to selected point”, which will allow you to freely move between points by clicking on them in the XY Pos box.

Page 29: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

29

Z-Stacking/Optical Sectioning There are three methods for optical sectioning….select one

1. Choose a middle plane and click “reset”, go to a live view and focus towards the bottom (click bottom) then focus to top (click top)

*Set your step size (distance between each plane) and/or the # of steps you want (image total)

2. Choose a middle plane and click “home” then input a distance range (equal from top to bottom)

3. Choose a plane somewhere in the sample and click “home” then input a staggered range above and below

Page 30: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

30

Lambda-wavelength • You can add automated wavelength switching to an experiment by checking off the checkbox for “Lambda”.

*This is most used for widefield imaging as this is already automated for confocal as default

• You can add an optical configuration (DAPI/FITC/CY3/DSRED) by clicking on the down arrows

*Make sure to adjust your camera settings before adding

Page 31: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

31

Image Stitching

• You can add image stitching to an experiment by checking off the checkbox for “Large Image”.

*You will need to make sure to run a calibration before starting

1. Place a high contrast slide on the stage (like an H&E)

2. Locate CalibrationRecalibrate “x” objective

3. Select “Auto” for calibration

Page 32: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

32

Confocal Imaging: A1si

Page 33: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

33

Layout

Page 34: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

34

Using Optical Configurations

*You have a series of “shortcut” keys on the top toolbar for confocal imaging and control of the microscope with a single button. These are referred to as “optical configurations”. They may show as different than the ones discussed here.

Click on the configuration you want to use……

1. You have a few shortcuts to view your fluorescence quickly to the eyepieces (DAPI-EYES_FITC-EYES_CY3-EYES_DSRED-EYES)

2. A1-Home will setup the microscope for confocal mode

3. RETURN TO CONFOCAL MODE can be used to use the confocal after viewing your sample in the eyepieces; it will show the last used laser settings

Page 35: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

35

Confocal Acquisition: Quick Start

*The A1 Simple GUI is a wizard that will walk you through acquiring a confocal image

1. Make sure you have previewed your slide through the eyepieces and focused on an area of interest.

2. Click on the “A1-Home” optical configuration on the top toolbar

3. Click on “Remove Interlock” in the GUI (will change from red to grey).

4. Choose Galvano for standard imaging, Resonant for fast imaging

5. Select what laser lines you want to acquire (check boxes for DAPI/FITC/TXRED/CY5/TD (transmission)

• HV-Gain, start at 100

• Offset-background adjustment, start at 0

• Laser Power, start at 5%

6. Start with a scan size of 512x512 and scan speed of 1.9usec.

7. Select “Normal” and “Ch Series”

8. Select 1.2 AU for Pinhole

9. Start “SCAN”

10. Adjust if needed

11. Click “CAPTURE”

*To save, click on FileSave As and select either a .tif or .jp2 for raw data

*To convert into a viewable image, click on ImageConvertConvert to 8bit RGB and save an additional copy

Page 36: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

36

1) Stops/Starts Live Image Acquisition (as a default, please do not use any averaging or integrating in this mode, see #10)

2) Captures an image (using user defined settings below)

3) “Find Mode”: Implements a fast preview scan a) Defined by Find Mode Settings in

4) Eye Port-changes to last used non-confocal setting

5) AG: Auto Gain feature; sets the HV (gain) for optimum based on user settings (for more info see Help, 4.2.5)

6) Skip: skips scan lines, increases speed

7) Unidirectional/Bidirectional: scan direction (Uni-one direction, traditional scanning; Bi-directional, used for live imaging for speed increase—does need to be aligned during scan). See Help, 8.1.3 for more info.

8) Pixel Dwell/Frame per Sec: scan speed; pixel dwell in micro seconds

9) Size: Scan Size in XY

10) Averaging and Integrating : average signal by line or integrate signal by line (see Help, 8.1.4). Normal mode does not use either. Typically averaging is used only for image capture, not for live preview.

11) Ch Series: Channel Series; setup will allow for each laser to be run sequentially rather than simultaneously *important for closely excited fluorophores! See Help, 4.1.4

12) Pinhole: adjusts size of pinhole, 1.2 AU can be set via button as standard, see Help 4.2.3 for more information.

13) Optical Path Settings: to set up lasers, dichroics, filters; for experienced users only!

14) Laser Selections: check off laser lines to be used. HV is your amplification gain (100 is a good starting point), offset will black values (0 is a good starting point) and the wavelength shown last is your laser power in % (5 is a good starting point).

Note: Offset Adjustment

Offset values should be close to a number value of 0 (zero) but a histogram should be previewed to adjust this value accurately. The left side of each wavelength’s peak should be close to 0,0 (bottom left of histogram). A more positive # will move the peak to the right, a more negative # will move it to the left. Please ask for assistance if you need help—this step is VERY important!

The A1 GUI-further explained (top to bottom)

Page 37: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

37

Scan Area/Zoom/ Resolution Improvements

• Locate the “A1 Scan Area” dialog.

• You can click and drag the green borders of the image to zoom into an area of interest. While zooming in normally decreases resolution, it can increase resolution when using a confocal. The pizel size will decrease and keep the image size the same.

*The “Nyquist XY” button will select the appropriate settings to set the selected area at Nyquist resolution (based on preferences set with drop down arrow).

Page 38: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

38

Spectral Acquisition

• To acquire a spectral image, choose the from the A1 Simple Gui.

• This will allow you to adjust the resolution and range for spectral detection.

*Spectral detection can be great for increasing the # of fluorophores you can use in a single experiment; it can also be used to subtract autofluorescence.

Page 39: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

39

Spectral Settings

Page 40: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

40

Spectral Unmixing Spectral Images can be unmixed a few different ways. It is best to decide which method based on the following questions: 1. Do you have all of your controls, i.e. separate slides per

wavelength as well as a slide will all wavelengths? Yes-we will need to acquire spectrums from each of your control

slides in order to use them when unmixing your combined wavelength’s slide

No-see next question

2. Do you only have a slide will all wavelengths? Yes-ok, instead of preparing controls, we will use either a “blind”

unmixing algorithm, pick from regions in your sample, or choose from stored known spectra.

Page 41: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

41

Blind Unmixing:

Unmixing from ROI/Library:

Under Image you can find all of the spectral unmixing adjustments

Page 42: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

42

Post Acquisition Image Adjustments

Page 43: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

43

Scale Bars

A Scale Bar can be added at any time from the right side image window’s toolbar. Once positioned where you choose, you can right click on it to change its properties or to “burn it” into the image

*it is recommended to save an original image before burning a scalebar into an image, save the second as a copy

Page 44: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

44

How you can visualize your data…….

2D Sliced Tiled

3D Projection EDF

*Can also make really nice 3D rotational movies as .avi’s!

Page 45: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

45

Movie Maker Click the film icon to open movie maker

Click film strip (black to red) to go into editing mode

Make rotations/movements/adjustments and click where you want them to show up in the filmstrip. Once you play it back and it looks good, click on the film icon to create an .avi.

No compression

About 150-200ms is usually good

Page 46: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

46

“X” Key Snapshots

*Use the “x” key on your keyboard to take snapshots for publication/presentation

-can be used on any open image window, in any orientation

“Split Window”

Page 47: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

47

Magnifying Glass

• The magnifying glass can be used to show zoomed portions of an image

• Under ViewMagnifying glass options, you can change the zoom and shape/border

*Holding down “shift” after choosing an area can allow it to be moved to another spot on the image; click on “X” to take a snapshot (as shown here)

Page 48: Nikon A1si Training Guide - Department of Biologybiology4.wustl.edu/imaging-facility/A1 Training.pdf · Nikon A1si Training Guide . 2 Table of Contents ... Nikon Ti. and the . Proscan

48

Image Saving Helper….

Yay, I just captured an Image!!

Do I want to use my image for analysis? Intensity Measurements

Area/Length Measurements Cell Counts

ETC.

Do I want to use my image for publication? Or open in later in PhotoShop?

I don’t have time to mess with many adjustments, I just want to send the image to a colleague or open it

on my own computer without hassle

I need to save my image as either a .TIF or .JP2

I need to convert my captured image to 8bits/channel, I can do this in Elements by:

ImageConvertConvert to 8bit RGB

Now I can save my image as a .tif and it

will open up in color for publication or adjustment in photoshop

If I want to, I can adjust my brightness and contrast within Elements:

ViewVisualization ControlsLUTS (drag bounding bars to make darker/brighter)

Click on the “X” key on your keyboard to make a compressed,

RGB copy of the image to send to my colleague

Save new copy as a .tif