  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    1/64

    FT-NIR Spectroscopy for Food and

    Subtitle

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    2/64

    If youre going to need a year to train a horseto locate contriband pick a musical one

    We all know that exceptionally well-

    designed, manufactured, delivered and

    (necessarily) make great Analytical

    Methods

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    3/64

    If youre going to need a year to train a horseto locate contriband pick a musical one

    We all know that exceptionally well-

    designed, manufactured, delivered and

    (necessarily) make great Analytical

    Methods

    Clever p e o p l e make great AnalyticalMethods

    A clever person can make sound and

    successful methods with the simplest of

    tools or lar el incom lete tools for the

    job, by learning to apply them to the

    greatest advantage

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    4/64

    If youre going to need a year to train a dogto locate contriband pick a musical one

    However, given the chance, clever

    p e o p l e will pick exceptionally well, ,

    installed i n st r u m e n t s because it will

    make their job a whole lot easier!

    Clever people together with greatinstruments can make exceptional

    na y ca e o s

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    5/64

    If youre going to need a year to train a dogto locate contriband pick one with a greatgreat nose

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    6/64

    If youre going to need a year to train a dogto locate contriband pick one with a greatgreat nose - an an inc ination to use it!

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    7/64

    The MPA Multi-Purpose Analyzer

    multiple measurement accessories for

    different sampling techniques: er p c ro es

    Integrating Sphere

    automated Sample Wheel

    Sam le Rotator

    software controlled measurement

    accessories selection full line of sampling accessories and

    options

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    8/64

    Unique Features of Bruker MPA

    Custom-made Instrument:

    Each customer buys exactly the instrument that matches his application.There is no need to purchase anything that is not wanted or needed.

    MPA fully equippedMPA basic configuration

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    9/64

    Maximum Flexibility:

    Fi er OpticProbes

    Samp eCompartment

    IntegratingExternalp ereransm ss on

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    10/64

    The MATRIX-F

    compact Process Spectrometer

    with RocksolidTM

    Interferometer mo u ar es gn

    robust, rugged and waterproof

    high long-run Stability

    high measuring Accuracy

    unrestricted Method Transfer possible

    all Components in 19 Geometry

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    11/64

    The MATRIX-I

    based on the optical module of MATRIX-F

    robust, rugged and waterproof fixed, gold-coated Integration Sphere

    vibration and temperature insensitive

    at-line and in-line Installation

    used for heterogeneous and coarse-

    gra ne ma er a s

    much better light efficiency than probes

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    12/64

    Calibration Transfer:

    e except ona per ormance o t e nstrument ensures a smootcalibration transfer not only to other MPA instruments but also to the

    MATRIXprocess spectrometers.

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    13/64

    Calibration Transfer:

    e except ona per ormance o t e nstrument ensures a smootcalibration transfer not only to other MPA instruments but also to the

    MATRIXprocess spectrometers.

    One Calibration

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    14/64

    Spectrometer Calibration at H20 Vapor Bands

    . 025

    a

    resolution = 25 cm -1.

    020

    eams

    pectr

    resolution = 2 cm -1. 015S

    ing

    le

    . 010

    target: 7.306,74 cm-1

    7800 7600 7400 7200 7000 6800 6600Wavenumber / cm-1

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    15/64

    OPUS spectroscopy software:

    ex e so tware, opt m ze orthe special demands of the FT-

    NIR technology.

    Various software acka es enable a ualitative as well as a uantitativeanalysis of the samples.

    Modular design, additional software packages can seamlessly beintegrated in the existing OPUS suite.

    The customer is able to use all software functions without any restriction.

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    16/64

    OPUS/IDENT:

    Identification software with factor

    analysis as well as with the Euclidian

    distance calculations

    Hierarchical libraries offer the

    identification of samples in multiple

    lar e libraries ossible >2 000

    samples)

    g egree o separa on, even

    with very similar products!

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    17/64

    OPUS/QUANT:

    Quantification software with

    automated calibration function:

    user defined optimization regions

    functions

    even inex erienced users to

    generate calibrations at a push of a

    button

    the process of time consuming

    method generation and optimization

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    18/64

    OPUS/LAB:

    Easy-to-use interface for the

    even inex erienced users can

    use OPUS/LAB

    pre-defined order of events

    no mistakes

    freely configurable to the needs

    of your users

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    19/64

    Full GMP and 21 CFR part 11 compliance:

    OPUS in combination with the MPAcomplies with regulations issued

    by ISO, ASTM, FDA and USP.

    full GMP capabilities like multiplesecurity levels, non-editable data files

    .

    electronic records and signatures.

    Bruker Optics is a certified supplier and

    companies.

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    20/64

    Training:

    free training after the installationof the instrument

    erman ra n ng courses w ce a year

    English training course once a year

    personal in-house training at thecustomers facility on demand

    every instrument is supported by anexpert who is available for you when

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    21/64

    Areas of Application

    ChemistryPetro & Polymer

    Food & FeedPharma

    Cosmetics

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    22/64

    Sample Forms

    Powders Tablets Seeds Paper Liquids

    Granulates Phials Foils

    Coatings

    Emulsions

    Pastes Resins

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    23/64

    Limits of NIR Technology

    Detection of C-H-, N-H-, O-H- and C=O-vibrations

    (mainly examination of organic compounds).

    The NIR-technique is currently not suitable for trace analysis

    Usually in the percentage range, sometimes down to 0.1%

    Examination of complex mixtures (> 5-7 components)usually demands a higher calibration effort.

    NIR is often not suitable for the identification of unknown samples.

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    24/64

    -

    How does a quantitative analysis work?

    Analysis

    rr

    r

    rrr

    rr

    r

    r

    rr

    Factors

    1

    2

    Spectra

    0.6

    0.

    7

    0.8

    0.9

    Units

    Reference

    rrrr

    rr

    r

    r

    r

    r3

    4

    5500060007000800090001000011000Wavenumber cm-1

    0.2

    0.3

    0.4

    0.5

    Absorbance

    Building up acalibration curve

    Calculation of thespectral parameters

    Measurement of spectra

    with known concentration

    - uantif in unknown Sam les

    1.5 % Comp. A

    34 % Comp. B

    62% Comp. C

    e erm na on oconcentration(s)

    na ys s ounknown spectrum

    Measurement ofunknown samples

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    25/64

    Pharmaceutical A lications

    Subtitle

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    26/64

    NIR Applications of Pharmaceuticals

    Substance Identification in the Receipt of Goods

    Quantification of active Materials (APIs)

    auxiliary Materials

    Fi ers

    in Powders, Capsules, Tablets or Vials

    Drying Processes

    Blending Solvent Recovery

    ...

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    27/64

    Calibration of active Ingredients

    Calibration of

    Ibuprofene Paracetamol

    in Tablets

    measued with

    External Transmission Unit

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    28/64

    Ibuprofen in Tablets

    73

    69

    e/%

    65

    67

    NIRva

    lu

    61

    63

    59

    59 61 63 65 67 69 71 73

    reference value / %

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    29/64

    Paracetamol in Tablets

    94

    90

    e/%

    86

    88

    NIRva

    lu

    82

    84

    80

    80 82 84 86 88 90 92 94

    reference value / %

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    30/64

    Calibration of active Ingredients

    Results:

    R2

    Range RMSECV

    Ibuprofene 99.41 60 - 74% 0.3%

    Paracetamol 99.46 81 - 94% 0.3%

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    31/64

    Ingredients of Vitamin Solutions

    Calibration of

    Methanol Water

    Sugar

    in Vitamin B2 Solution

    measured with

    1 mm Transmission Probe

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    32/64

    Methanol in Vitamin B2 Solution

    59

    53

    55

    e/%

    49

    51

    NIRvalu

    45

    47

    43

    43 45 47 49 51 53 55 57 59

    reference value / %

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    33/64

    Water in Vitamin B2 Solution

    50

    45

    e/%

    40

    NIRvalu

    35

    30

    30 35 40 45 50

    reference value / %

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    34/64

    Sugar in Vitamin B2 Solution

    10

    8e/%

    7

    NIRvalu

    6

    5

    5 6 7 8 9 10

    reference value / %

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    35/64

    Calibration of Ingredients

    Results:

    R2

    Range RMSEP

    Methanol 99.87 44 - 59% 0.18%

    Water 99.92 31 - 49% 0.15%

    Sugar 99.33 5.5 - 9.6% 0.11%

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    36/64

    Cosmetics A lications

    Subtitle

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    37/64

    Calibration of Shampoo

    Calibration of

    Wash Active Substancesin Shampoo

    measured at 30

    oC

    in the Sample Compartment

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    38/64

    Wash Active Substances in Shampoo

    265

    kg

    225

    245

    /mmol

    185

    205

    IRvalu

    145

    165

    145 165 185 205 225 245 265 285

    reference value / mmol/kg

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    39/64

    Calibration of Shampoo

    Results:

    R

    2

    Range RMSECV

    WAS 99.43 150-282 2.5 [mmol/kg]

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    40/64

    Surfactant in aqueous Solution

    80

    lue/%

    20NIR

    v

    0

    0 20 40 60 80

    reference value / %

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    41/64

    Food & Feed A lications

    Subtitle

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    42/64

    M P A with Integrating Sphere and Sample Compartment

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    43/64

    M P A Int.Sphere/Sample Comp./Transmission Unit

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    44/64

    MPA with Petri Dish Rotator

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    45/64

    MPA with Rotator and Transmission Unit

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    46/64

    Starter Calibrationsfor Food Aapplications

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    47/64

    Food Starter Calibrations for turn-keysolutions

    Multiple Packages

    - -

    B-MEA-M Meat

    - -

    Single Package

    PriceList

    B-DAI-S Dairy

    B-MEA-S Meat

    B-OIL-S Edible Oils

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    48/64

    NIR Starter Calibration Packages: Dairy

    Liquid Milk and Milk Products in Transmission (1mm flow cell)

    aw an w o e m

    Whey

    Whey protein concentrate

    Lactose Concentrate

    Brine (Mozzarella)

    Cleaning and disinfection solvents

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    49/64

    Starter Calibration for Milk Powder

    Calibration of

    Protein

    Ash

    Lactose

    in Milk Powder

    Measured on the sphere with IN311/C

    + IN311-S

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    50/64

    Measurements of melamine in milk powder

    .8

    0.9

    1.2

    1.4

    Pure milk powder (red)

    0.7

    nits

    0.6

    0.8

    1.0

    0.5

    0.6

    Absorb

    ance

    400050006000700080009000100001100012000

    0.2

    0.4

    0

    .3

    0.4

    Pure melamine (blue)

    400050006000700080009000100001100012000

    Wavenumber cm-1

    50

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    51/64

    Measurements of melamine in milk powder

    1.4

    0

    . .

    .0

    1.2

    s

    0.3

    5

    0.

    -1

    0.8

    1.

    Absorbance

    Unit

    0.3

    0

    0.4

    0.6 60006200640066006800700072000.2

    5

    400050006000700080009000100001100012000

    Wavenumber cm-1

    0.2

    51

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    52/64

    Measurements of melamine in milk powder

    NIR-Calibration melamine in milk powder3.2

    2.2

    2.7

    ion/%

    = .

    RMSECV= 0.057

    1.2

    1.7

    R-Predict

    0.7

    NI

    .

    0.2 0.7 1.2 1.7 2.2 2.7 3.2

    True Value / %

    52

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    53/64

    Measurements of melamine in liquid milk

    0

    3.5

    Liquid milk of 1.5% and 3.5% of fat with admixture of melamine from 0.05% - 0.3%

    .0

    2.5

    3.

    1.0

    1.5

    2

    Absorbanc

    e

    Units

    0.0

    0.5

    400050006000700080009000100001100012000

    Wavenumber cm-1

    -0.5

    53

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    54/64

    Measurements of melamine in liquid milk

    NIR-Calibration melamine in liquid milk

    0.31

    0.26

    ion/%

    = .

    RMSECV= 0.0042

    0.16

    .

    R-Predict

    0.11NI

    .

    0.06 0.11 0.16 0.21 0.26 0.31

    True Value / %

    54

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    55/64

    Measurement: Oils in Sample Compartment

    Instrumentation:

    - u urpose na yzer

    - MPA-C: Sample Compartment

    -

    - IN601C08: Insert for Vials

    -

    Resolution: 8 cm-1

    No. of Scans: 32 Scans Temperature: 50

    oC

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    56/64

    External Heating Block

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    57/64

    Universal Calibrations

    Property Data Set Performance

    Name Unit n Min Max Rank R2

    SEPIodine Value IV 607 0.1 133 9 99.9 0.56

    Trans Fatty Acids % 532 0.0 60.4 13 99.9 0.61

    TFA low range % 243 0.0 2.6 11 93.2 0.17

    These calibrations should normally work without further adaptation

    with common customer samples.

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    58/64

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    59/64

    Measurement: Trans content in Fats/Oils

    trans content wasexpressed as %FA

    Universal trans contentcalibration covers transcontent from 0 to 56 FA

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    60/64

    Starter Calibrations

    Property Data Set Performance

    Name Unit n Min Max Rank R2

    SEPAnisidine Value AnV 270 0 195 14 95.2 8.64

    Free Fatty Acids % 790 0.1 6.0 9 99.8 0.09

    C16:0 (Palmitic Acid) % 608 4 16 10 95.2 0.50

    C18:0 (Stearic Acid) % 660 1.8 92 15 98.2 1.41

    C18:0 (Stearic Acid) % 482 1.6 6 14 94.3 0.25

    C18:1 (Oleic Acid) % 800 0 85 17 99.9 0.65

    C18:2 (Linoleic Acid) % 597 2.5 67 12 99.9 0.20

    C18:3 (Linolenic Acid) % 412 0 9.5 10 99.3 0.20

    Customer samples should be added for optimum performance.

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    61/64

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    62/64

    Mix of ingredients for animal feed

    Energy Source

    wheat, barley, corn, oats, fat, molassesetc.

    Vegetable Proteinsoya, rapesee s, sun ower see s etc.

    Crude Fibers , , .

    Minerals

    , , ,etc.

  • 8/10/2019 NIR Applications Phk [Compatibility Mode]

    63/64

    Task of a Nutritionist: the best feed inconstant quality for the lowest price

    For the optimization of the mixed feed the Nutritionist needs to know

    the property values of the raw materials for:protein, moisture (dry matter), fiber, oil (fat) and ash.

    For this task classical wet chemistry is slow and expensive!

    This is a NIR task!