non-invasive estimation of pressure gradients in pulsatile...

1
Non-invasive Estimation of Pressure Gradients in Pulsatile Flow using Ultrasound Jacob Bjerring Olesen, Carlos Armando Villagómez Hoyos Marie Sand Traberg and Jørgen Arendt Jensen Center for Fast Ultrasound Imaging, Dept. of Elec. Eng., Bldg. 349, Technical University of Denmark, DK-2800 Kgs. Lyngby, Denmark Background Carotid artery suffering from atherosclerosis. Objective: Show that pressure gradients can be measured non-invasively. Background: Pressure gradients in hemodynamics provides important information for diagnosing various cardiovascular diseases. The gradients are used to study how flow changes, caused by for instance plaque formation, affect the risk of embolism. Current measures: Today’s measure of pressure gradients is an invasive procedure done by catheters under the guidance of X-ray. Long-term aim: The idea of estimating pressure gradients non-invasively, is to replace catheterization thereby minimizing the risks that are associated with invasive procedure. Governing Equation and Method The Navier-Stokes equation for an incompressible fluid ρ v t + v · v = -p + ρ g + μ 2 v. (1) ρ : Density. v t : Temporal acceleration. v · v: Spatial acceleration. p: Pressure gradient. g: Gravity. μ 2 v: Viscous force. The grav- itational forces are negligible as the patient is placed in a supine position, thus; p = -ρ v t + v · v + μ 2 v. (2) Equipment used in the experimental setup Flow phantom mimicking a carotid artery with 70 % atherosclerosis Magnetic resonance imaging was used for scanning the flow phantom. Image slides were collected and the fluid domain was extracted using segmentation software. An MRI scanner and an ultrasound scanner were used in the set-up Visual comparison between the simulated model and the estimated results (a) (b) Fig. A shows the simulated velocity field through the center of the constricted vessel, while Fig. B presents the estimated field obtained from ultrasound flow data. (a) (b) Fig. A shows the simulated pressure gradient field through the center of the constricted vessel and Fig. B presents the estimated pressure field. The estimated pressure field is estimated from the velocities shown in the top Fig. B. using (2). Standard deviation and bias of estimator The standard deviation and bias of the estimator is calculated to 13 % and 10 % relative to the peak estimated gradient, respectively. Single point comparison over the pulse 0 0.222 0.84 -0.2 -0.1 1.5 1.6 Axial gradient component Time [s] Gradient magnitude [kPa/m] Measured Simulated 0 0.222 0.84 -0.2 0 3.1 4.2 Lateral gradient component Time [s] Gradient magnitude [kPa/m] Measured Simulated The graphs show the magnitude of the axial and lateral gradient component. The plotted data is taken from the peak pressure value in the upstreams region, also marked by the dark square seen on the pressure plots. In-vivo measurement at peak systole Vector velocity flow and pressure gradients measured on a common carotid artery at peak systole. Conclusion Pressure gradients can be derived non-invasively us- ing vector velocity ultrasound data. Poster presented at the DTU Compute/KU Comp.Sci. Summer school on deep learning for image analysis, August 18-22, 2014 Preprints available from: [email protected] Web: www.cfu.elektro.dtu.dk

Upload: others

Post on 23-Apr-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Non-invasive Estimation of Pressure Gradients in Pulsatile ...deep-learning.compute.dtu.dk/wp-content/uploads/... · Jacob Bjerring Olesen, Carlos Armando Villagómez Hoyos Marie

Non-invasive Estimation of Pressure Gradients in Pulsatile Flow usingUltrasound

Jacob Bjerring Olesen, Carlos Armando Villagómez Hoyos Marie Sand Traberg and Jørgen Arendt Jensen

Center for Fast Ultrasound Imaging, Dept. of Elec. Eng., Bldg. 349, Technical University of Denmark, DK-2800 Kgs. Lyngby, Denmark

Background

Carotid artery suffering from atherosclerosis.

Objective:Show that pressure gradients can be measured non-invasively.

Background:Pressure gradients in hemodynamics provides important informationfor diagnosing various cardiovascular diseases. The gradients areused to study how flow changes, caused by for instance plaqueformation, affect the risk of embolism.

Current measures:Today’s measure of pressure gradients is an invasive procedure doneby catheters under the guidance of X-ray.

Long-term aim:The idea of estimating pressure gradients non-invasively, is to replacecatheterization thereby minimizing the risks that are associated withinvasive procedure.

Governing Equation and MethodThe Navier-Stokes equation for an incompressible fluid

ρ

[∂v∂ t

+v ·∇v]=−∇p + ρg + µ∇

2v. (1)

ρ: Density.∂v∂ t

: Temporal acceleration. v ·∇v: Spatial acceleration.

∇p: Pressure gradient. g: Gravity. µ∇2v: Viscous force. The grav-itational forces are negligible as the patient is placed in a supineposition, thus;

∇p =−ρ

[∂v∂ t

+v ·∇v]+ µ∇

2v. (2)

Equipment used in the experimental setup

Flow phantom mimicking a carotid artery with 70 % atherosclerosis

Magnetic resonance imaging was used for scanning the flow phantom. Image slides were collected

and the fluid domain was extracted using segmentation software.

An MRI scanner and an ultrasound scanner were used in the set-up

Visual comparison between the simulatedmodel and the estimated results

(a) (b)Fig. A shows the simulated velocity field through the center of the constricted vessel, while Fig. B

presents the estimated field obtained from ultrasound flow data.

(a) (b)Fig. A shows the simulated pressure gradient field through the center of the constricted vessel and

Fig. B presents the estimated pressure field. The estimated pressure field is estimated from the

velocities shown in the top Fig. B. using (2).

Standard deviation and bias of estimatorThe standard deviation and bias of the estimator is calculated to 13 %and 10 % relative to the peak estimated gradient, respectively.

Single point comparison over the pulse

0 0.222 0.84−0.2−0.1

1.51.6

Axial gradient component

Time [s]

Gradientmagnitude[kPa/m]

Measured

Simulated

0 0.222 0.84−0.2

0

3.1

4.2

Lateral gradient component

Time [s]

Gradientmagnitude[kPa/m]

Measured

Simulated

The graphs show the magnitude of the axial and lateral gradient component. The plotted data is

taken from the peak pressure value in the upstreams region, also marked by the dark square seen

on the pressure plots.

In-vivo measurement at peak systole

Vector velocity flow and pressure gradients measured on a common carotid artery at peak systole.

Conclusion

Pressure gradients can be derived non-invasively us-ing vector velocity ultrasound data.

Poster presented at the DTU Compute/KU Comp.Sci. Summer school on deep learning for image analysis, August 18-22, 2014 Preprints available from: [email protected] Web: www.cfu.elektro.dtu.dk