Rev # Date/Quote Description
Equipment References DrawingIndex #
Note that not all drawings will be required for each mount / drawings package
A, B, C
00a
_1
Room Layout - Contains overhead view of Skytron equipment
SkyVision Routing - Contains overhead view of Skytron SkyVision conduit routing
Elevation/Mounting Details - Contains elevation drawings of equipment with mounting details
Carrier Details - Contains front, left and right side views of equipment carriers detailing accessories and outlet placement
Medical Gas Details - Contains medical gas details required for the riser plateinstallation
Electrical Wiring Detail - Contains detailed wiring diagram and circuit requirements for equipment booms and lights
Communication Details - Contains a breakdown of required communications cabling and connections
Light Fixture Details - Contains details of light wiring for fixture mounts and back box details
SKYTRON products found within this package are specifically built for your facility. Ensure that all of your custom specifications have been included.Deviating from the specific equipment placement could cause conflicts within the room and should be approved by a Skytron Representative.
I have read the appropriate requirements from the Skytron Pre-Installation Guide for Ergon Skybooms and consulted with the individualtrades. I understand there are specific ceiling height, medical gas, electrical, video/communications and structural requirements that
must be supplied by the customer's representatives to support this project. Please be aware modification to the site specific documentcould result in Change Order/Drawing Change fees. These fees take place after the second Submital Revision or the first Fabrication
Revision. Please speak with your Skytron Representative if you have any questions with regards to these fees.
PLEASE READ THE FOLLOWING CAREFULLY
EACH DRAWING MUST BE INITIALED AND DATED
Accepted by:(signature)
00b
_2
_3
_4
_5
_6
Accepted by:(print)
Requested Delivery Date
Date:
P.O. #:
NORTHWEST HOSPITAL CENTER
New Submittal Package
Drawing Package Revision Summary
0
1
2
3
4
5
6
7
6/20/201337178 - 5
Accepted - with changes RejectedAccepted - as is
Updated Room Layout6/21/201337178 - 5
01 9
A7.01
5
6
7
8
OR #5E32
RESTRICTEDCORRIDOR
E17
CLEANCOREE20
?
?
SCRUBE17H
5'
4'-014"
4'
5'
DESC
RIPT
ION:
ROO
M LA
YOUT
00aSHEETINITIAL:
DATE:
NORT
HWES
T HO
SPIT
ALCE
NTER
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
ROOM
TYP
E: G
EN O
RRE
V. #:
16/2
1/201
3
EQUIPMENT SCHEDULE
ID DESCRIPTION
A ANESTHESIA / MONITOR BOOM
B EQUIPMENT / MONITOR BOOM WITH LIGHTS
C UTILITY / MONITOR BOOM
IN THE EVENT THAT FIXTURE LOCATIONS NEED TOCHANGE PLEASE CHECK WITH SKYTRON ON THE
REVISED LAYOUT LOCATION*SITE SPECIFIC LAYOUT*
* IF SKYVISION IS APPLICABLESEE END OF BOOM PACKAGE*
A7.01
3
2
A7.0110OR #04
E31
ALCOVEE20C
?
?
??
SCRUBE17G
4
1
4'
5'
5'
4'
DESC
RIPT
ION:
ROO
M LA
YOUT
01aSHEETINITIAL:
DATE:
NORT
HWES
T HO
SPIT
ALCE
NTER
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
ROOM
TYP
E: G
EN O
RRE
V. #:
16/2
1/201
3
EQUIPMENT SCHEDULE
ID DESCRIPTION
A ANESTHESIA / MONITOR BOOM
B EQUIPMENT / MONITOR BOOM WITH LIGHTS
C UTILITY / MONITOR BOOM
IN THE EVENT THAT FIXTURE LOCATIONS NEED TOCHANGE PLEASE CHECK WITH SKYTRON ON THE
REVISED LAYOUT LOCATION*SITE SPECIFIC LAYOUT*
* IF SKYVISION IS APPLICABLESEE END OF BOOM PACKAGE*
OR #4E31
OR #3E30
A7.024
1
2
3
5'
4'
5'
4'
DESC
RIPT
ION:
ROO
M LA
YOUT
02aSHEETINITIAL:
DATE:
NORT
HWES
T HO
SPIT
ALCE
NTER
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
ROOM
TYP
E: G
EN O
RRE
V. #:
16/2
1/201
3
EQUIPMENT SCHEDULE
ID DESCRIPTION
A ANESTHESIA / MONITOR BOOM
B EQUIPMENT / MONITOR BOOM WITH LIGHTS
C UTILITY / MONITOR BOOM
IN THE EVENT THAT FIXTURE LOCATIONS NEED TOCHANGE PLEASE CHECK WITH SKYTRON ON THE
REVISED LAYOUT LOCATION*SITE SPECIFIC LAYOUT*
* IF SKYVISION IS APPLICABLESEE END OF BOOM PACKAGE*
A1SHEET
STRUCTURAL REQUIREMENTS - Architect and Structural Engineer
Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.
Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces place 90° apart and positioned at a 45° and 60° angle is recommended.
Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications that thesway bracing be fastened to the structural ceiling.
Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.
Expansion AnchorsTest 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors mustmeet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.
Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate is ASTM 500 Grade B, 6" diameter tube. Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.
Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.
Mounting Structure DesignSeismic structural applications may differ. Please contact your local SKYTRON distributor for specific calculations. The mountingstructure must be designed and fabricated to position the bottom of the SKYTRON mounting plate 10" (-0", +1") above thefinished ceiling. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. Themounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation at the mountingplate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior to installation ofthe fixture. Please contact your SKYTRON representative to schedule testing.
Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.
Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.
SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.
Additional Skytron Supplied ItemsIn addition to the pre-installation kit. SKYTRON provides the following items:
(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)INITIAL:DATE:
STRUCTURALCEILING
STRUCTURALCEILING
STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)
SWAY BRACING3" x 3" x 1/4" (4)
GUSSET 6" DIA.SUPPORT
TUBE
FINISHED CEILING
20"DIAMETER OPENING
REQUIRED(CENTERED UNDER MOUNT)
12" MIN. / 60" MAX.
*10" (-0"/+1") TYPICALREFER TO JOB SPECIFIC ELEVATION ABOVE
MOUNTING PLATE(SKYTRON SUPPLIED)
45°- 60°
FIXTURE HUBTYP. 1/4"
PJP
ACCESS DOOR24" x 24"
(REQUIRED)
TYP. 1/4"TOP & BOTTOM
STRUCTURALCEILING PLATE
17" DIA. x 1"
EXPANSION ANCHORS:HILTI KB-II (ICBO 4627)
CARBON STEEL
NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult
specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld, weld slag or etc.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each
application.• *Critical Dimension
11" HIGH CLEAR SPACEABOVE MOUNTING PLATE
REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL
COMPONENT INSTALLATION
FIXTURE WEIGHT:MOMENT LOAD:
809 LBS.4584 FT. LBS.
OPA NUMBER: 2510-07
DESC
RIPT
ION:
ELE
VATI
ON D
ETAI
LS
120" CEILING HEIGHT
0" FLOOR
*SITE SPECIFIC ELEVATION DETAILS* 6/20/2
013
72"
61"
31 1/2"
11 1/16"
3" 10"
51 3/16"
35 7/16"
12"
18"
70"
34"
25"
NORT
HWES
T HO
SPIT
ALPR
OJEC
T #:
13-0
72-K
PSU
BMIT
TAL
PLOT
DAT
E:
MDL:
ET2F
VBM3
6/2AF
C2QT
Y.: 3
REV.
#: 0
9"
77"
GAS OUTLETS:
NORT
HWES
T HO
SPIT
AL
DIMENSIONS:INITIAL:
PLO
T DA
TE: 6
/20/
2013
SUBM
ITTA
LPR
OJE
CT #
: 13-
072-
KP
DESC
RIPT
ION:
CAR
RIER
DET
AILS
SHEETA2
QTY
: 3
R
EV #
: 0
DATE: ELECTRICAL:
MO
DEL
#: E
T2FV
BM36
/2AF
C2
CARRIER
*SITE SPECIFIC CARRIER DETAILS*
WHITE - REQUIRES"J" BOX
RED
VACUUMVACUUM
OXYGENOXYGEN
NITROUS OXIDEWAGDMEDICAL AIR
VERTICAL NITROGENSYSTEM (DISS)
(2) 125V, 20A DUPLEX - WHITECHEMETRON37"H x 11 1/2"W x 8"D(4) 125V, 20A DUPLEX - RED
ACCESSORY LIST(2) BOLT ON VACUUM SLIDES
A3SHEET
MEDICAL GAS REQUIREMENTS - Medical Gas / Piping Engineer
Notes:• Ohio Medical manufactured gas outlets supplied unless otherwise specified.Pricing and lead time could vary if another MFGR is chosen.• Medical gas riser plates comply with NFPA 99-2005 Edition Cleanliness of gas outlets must be maintained through installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal
instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **
Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawingssupplied by SKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hosesconnect the fixture to the riser plate connectors. The customer is responsible to deliver the appropriate medical gas from thefacility supply to the riser plate connectors and for the design of a medical gas system with adequate flow capacity capable ofcompensating for the accumulative flow restrictions associated with conventional construction methods i.e. flex gas hoses.SKYTRON provides medical gas riser plate(s) and the appropriate connectors for attachment to the Mounting Plate.The riser plate attaches to the mounting plate and will accommodate up to 12 gas connectors. The connectors are D.I.S.S.type medical gas connectors with removable single check valves and provide a 1/2" copper tube for attachment to facility supplylines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.
All connection and testing of medical gas piping to be performed in accordance with NFPA 9-2005 Edition-5.1.6 requirements.
Recommended Gas Riser PlacementWhen mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.
INITIAL:DATE:
GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*
9"
DISS CONNECTOR(w/ REMOVABLE PRIMARY
CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT
HAVE PRIMARY CHECK.)*SUPPLIED w/MOUNTING KIT*
D I S SCHEMETRON
* VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED *
CGA ColorStandard
Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.
Note #2 - For testing and certification purposes, individual station inlets shall be capable of a flow rate of: 3 SCFM, while maintaining a system pressure of not less than 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.
Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.
Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.
Additional references: Health Care Facilities Handbook 2002, Section 5.1.11, page 211 and section 5.1.12.13.10.1 thru 5.1.12.3.10.5, page 224. NFPA 99, 2002 guideline figure A.5.1.6.
• TYPE K COPPERTUBING
• 1/2" O.D. UNLESSOTHERWISE STATED
Medical Air(Yellow)
Test GasCanada Color
StandardMedical Air
(Black)
AbbreviatedName
MedAir
StandardPressure
50 - 55 psig
MaximumPressure
55 psig
AllowablePressure Drop
5 psig
Minimum Flow Rates
3.5 SCFM per outlet (100NL/min)N2NF see note #1Carbon Dioxide
(Gray)Carbon Dioxide
(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox
(Brown)Heliox
(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)
Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig
5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4
Nitrous Oxide(Blue)
Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF
Oxygen(Green)
Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1
Vacuum(White)
Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2
Waste AnestheticGas Disposal
(Purple) WAGDVaries withsystem type
At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)
through the interface
Waste AnestheticGas Disposal
(Purple) see note #3
PURITANOHMEDA
NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult
specific structural criteria defined by a structural engineer.• Mounting bolts and nuts shipped with fixture.
INLET TUBE
RISER PLATE
JUNCTIONBOX
BRACKET
GENERIC RISER PLATE INSTALLATION
DESC
RIPT
ION:
MED
ICAL
GAS
DET
AILS
*SITE SPECIFIC GAS DETAILS*
6/20/2
013
NORT
HWES
T HO
SPIT
ALPR
OJEC
T #:
13-0
72-K
PSU
BMIT
TAL
PLOT
DAT
E:
MDL:
ET2F
VBM3
6/2AF
C2QT
Y.: 3
REV.
#: 0
CHECK VALVEREMOVAL TOOL
OXYG
EN
OXYG
EN
NITR
OUS
OXID
E
NITR
OGEN
MED
ICAL
AIR
WAG
D
VACU
UM
VACU
UM
DESC
RIPT
ION:
ELE
CTRI
CAL W
IRIN
G DE
TAILS
A4SHEET
ELECTRICAL REQUIREMENTS - Electrical Engineer
Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixtureto junction box or pump enclosure (if applicable). SKYTRON supplies either an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate fieldwiring for up to six circuits that is mounted on the mounting plate in the correct position OR if applicable, a hydraulic pump enclosure/junction box(18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped withthe installation kit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz power supply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal,state and local codes. It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patient care area.
Specific conductor colors and/or wiring for isolated applications are available upon request.
INITIAL:DATE:
3 INPUT LEADS
CARRIER UTILITIES PANELFIXTURE JUNCTION BOX
AT MOUNTING PLATE
120 VAC, 20 AMP
120 VAC, 20 AMP CIRCUIT 1
CIRCUIT 2
(4) 125V, 20A, DUPLEX OUTLETS - RED(2) 125V, 20A, DUPLEX OUTLETS - WHITE
NON-ISOLATED POWERSYSTEMS
INTERNAL FIXTURE WIRINGTYPICAL WIRE TYPE:
12AWG, 600V, XHHW-2, 90°CWHITE, BLACK, GREEN
120 VAC, 20 AMP CIRCUIT 3
ARM COVERARM COVER
ARM COVERARM COVER
GROUND(FOR ELECTRICAL)
PROTECTED EARTH(FOR CHASSIS)
NORMALDUPLEX
CRITICALDUPLEX
CRITICALDUPLEX
GENERIC BOOM ELECTRICAL WIRINGDIAGRAM FOR FIXED / SPRING ARMS
LGN
GENERIC WIRING DIAGRAM
ULTRAFLEXTYPE USL CONDUIT
ULTRAFLEXTYPE USL CONDUIT
GREENBLACK (L)WHITE (N)
GREENBLACK (L)WHITE (N)
CRITICAL NORMAL
*SITE SPECIFIC WIRING DETAILS* 6/20/2
013
NORT
HWES
T HO
SPIT
ALPR
OJEC
T #:
13-0
72-K
PSU
BMIT
TAL
PLOT
DAT
E:
MDL:
ET2F
VBM3
6/2AF
C2QT
Y.: 3
REV.
#: 0
RED
RED
WHITE
B1SHEET
STRUCTURAL REQUIREMENTS - Architect and Structural Engineer
Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.
Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces place 90° apart and positioned at a 45° and 60° angle is recommended.
Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications that thesway bracing be fastened to the structural ceiling.
Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.
Expansion AnchorsTest 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors mustmeet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.
Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate is ASTM 500 Grade B, 6" diameter tube. Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.
Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.
Mounting Structure DesignSeismic structural applications may differ. Please contact your local SKYTRON distributor for specific calculations. The mountingstructure must be designed and fabricated to position the bottom of the SKYTRON mounting plate 10" (-0", +1") above thefinished ceiling. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. Themounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation at the mountingplate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior to installation ofthe fixture. Please contact your SKYTRON representative to schedule testing.
Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.
Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.
SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.
Additional Skytron Supplied ItemsIn addition to the pre-installation kit. SKYTRON provides the following items:
(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)INITIAL:DATE:
STRUCTURALCEILING
STRUCTURALCEILING
STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)
SWAY BRACING3" x 3" x 1/4" (4)
GUSSET 6" DIA.SUPPORT
TUBE
FINISHED CEILING
20"DIAMETER OPENING
REQUIRED(CENTERED UNDER MOUNT)
12" MIN. / 60" MAX.
*10" (-0"/+1") TYPICALREFER TO JOB SPECIFIC ELEVATION ABOVE
MOUNTING PLATE(SKYTRON SUPPLIED)
45°- 60°
FIXTURE HUBTYP. 1/4"
PJP
ACCESS DOOR24" x 24"
(REQUIRED)
TYP. 1/4"TOP & BOTTOM
STRUCTURALCEILING PLATE
17" DIA. x 1"
EXPANSION ANCHORS:HILTI KB-II (ICBO 4627)
CARBON STEEL
NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult
specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld, weld slag or etc.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each
application.• *Critical Dimension
11" HIGH CLEAR SPACEABOVE MOUNTING PLATE
REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL
COMPONENT INSTALLATION
FIXTURE WEIGHT:MOMENT LOAD:
1028 LBS.5607 FT. LBS.
OPA NUMBER: 2510-07
DESC
RIPT
ION:
ELE
VATI
ON D
ETAI
LS
120" CEILING HEIGHT
0" FLOOR
*SITE SPECIFIC ELEVATION DETAILS* 6/20/2
013
MODE
L #: E
CT2F
KM48
/2AFC
2/AUT
75QT
Y.: 3
REV.
#: 0
72"
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
61"
31 1/2"
11 1/16"
3" 10"
51 3/16"
35 7/16"
12"
18"9"
77"
23"
70"
16 1/2"
19"
42 15/16"
80"
35 1/4"
80"
REFER TO AUT75 DATA SHEETFOR SURGICAL LIGHT
SPECIFICATIONS
NORT
HWES
T HO
SPIT
AL
CHEMETRON(4) 125V, 20A DUPLEX - RED
(4) 125V, 20A DUPLEX - WHITE54"H x 22"W x 30"D
ACCESSORY LIST(2) BOLT ON VACUUM SLIDES(3) KMSH
NORT
HWES
T HO
SPIT
AL
GAS OUTLETS:DIMENSIONS:INITIAL:
PLO
T DA
TE: 6
/20/
2013
SUBM
ITTA
LPR
OJE
CT #
: 13-
072-
KP
DESC
RIPT
ION:
CAR
RIER
DET
AILS
SHEETB2
QTY
: 3
R
EV #
: 0
DATE: ELECTRICAL:
MO
DEL
#: E
CT2F
KM48
/2AF
C2/A
UT75
CARRIER
*SITE SPECIFIC CARRIER DETAILS*
VACUUMVACUUM
NITROGEN CONTROLSYSTEM - DISS
CARBON DIOXIDE - DISSCARBON DIOXIDE - DISS
RED
WHITEREQUIRES"J" BOX
B3SHEET
MEDICAL GAS REQUIREMENTS - Medical Gas / Piping Engineer
Notes:• Ohio Medical manufactured gas outlets supplied unless otherwise specified.Pricing and lead time could vary if another MFGR is chosen.• Medical gas riser plates comply with NFPA 99-2005 Edition Cleanliness of gas outlets must be maintained through installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal
instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **
Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawingssupplied by SKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hosesconnect the fixture to the riser plate connectors. The customer is responsible to deliver the appropriate medical gas from thefacility supply to the riser plate connectors and for the design of a medical gas system with adequate flow capacity capable ofcompensating for the accumulative flow restrictions associated with conventional construction methods i.e. flex gas hoses.SKYTRON provides medical gas riser plate(s) and the appropriate connectors for attachment to the Mounting Plate.The riser plate attaches to the mounting plate and will accommodate up to 12 gas connectors. The connectors are D.I.S.S.type medical gas connectors with removable single check valves and provide a 1/2" copper tube for attachment to facility supplylines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.
All connection and testing of medical gas piping to be performed in accordance with NFPA 9-2005 Edition-5.1.6 requirements.
Recommended Gas Riser PlacementWhen mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.
INITIAL:DATE:
GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*
9"
DISS CONNECTOR(w/ REMOVABLE PRIMARY
CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT
HAVE PRIMARY CHECK.)*SUPPLIED w/MOUNTING KIT*
D I S SCHEMETRON
* VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED *
CGA ColorStandard
Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.
Note #2 - For testing and certification purposes, individual station inlets shall be capable of a flow rate of: 3 SCFM, while maintaining a system pressure of not less than 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.
Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.
Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.
Additional references: Health Care Facilities Handbook 2002, Section 5.1.11, page 211 and section 5.1.12.13.10.1 thru 5.1.12.3.10.5, page 224. NFPA 99, 2002 guideline figure A.5.1.6.
• TYPE K COPPERTUBING
• 1/2" O.D. UNLESSOTHERWISE STATED
Medical Air(Yellow)
Test GasCanada Color
StandardMedical Air
(Black)
AbbreviatedName
MedAir
StandardPressure
50 - 55 psig
MaximumPressure
55 psig
AllowablePressure Drop
5 psig
Minimum Flow Rates
3.5 SCFM per outlet (100NL/min)N2NF see note #1Carbon Dioxide
(Gray)Carbon Dioxide
(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox
(Brown)Heliox
(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)
Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig
5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4
Nitrous Oxide(Blue)
Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF
Oxygen(Green)
Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1
Vacuum(White)
Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2
Waste AnestheticGas Disposal
(Purple) WAGDVaries withsystem type
At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)
through the interface
Waste AnestheticGas Disposal
(Purple) see note #3
PURITANOHMEDA
NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult
specific structural criteria defined by a structural engineer.• Mounting bolts and nuts shipped with fixture.
INLET TUBE
RISER PLATE
JUNCTIONBOX
BRACKET
GENERIC RISER PLATE INSTALLATION
DESC
RIPT
ION:
MED
ICAL
GAS
DET
AILS
*SITE SPECIFIC GAS DETAILS*
6/20/2
013
CHECK VALVEREMOVAL TOOL
MODE
L #: E
CT2F
KM48
/2AFC
2/AUT
75QT
Y.: 3
REV.
#: 0
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
NORT
HWES
T HO
SPIT
AL
NITR
OGEN
CARB
ON D
IOXI
DE
CARB
ON D
IOXI
DE
VACU
UM
VACU
UM
DESC
RIPT
ION:
ELE
CTRI
CAL W
IRIN
G DE
TAILS
B4SHEET
ELECTRICAL REQUIREMENTS - Electrical Engineer
Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixtureto junction box or pump enclosure (if applicable). SKYTRON supplies either an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate fieldwiring for up to six circuits that is mounted on the mounting plate in the correct position OR if applicable, a hydraulic pump enclosure/junction box(18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped withthe installation kit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz power supply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal,state and local codes. It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patient care area.
Specific conductor colors and/or wiring for isolated applications are available upon request.
INITIAL:DATE:
5 INPUT LEADS
CARRIER UTILITIES PANEL
120 VAC, 15 AMPAURORA LIGHT
LIGHT CIRCUITTO WALL CONTROL
SEE PAGE 6
FIXTURE JUNCTION BOXAT MOUNTING PLATE
CIRCUIT 5
120 VAC, 20 AMP
120 VAC, 20 AMP
120 VAC, 20 AMP
120 VAC, 20 AMP
CIRCUIT 1
CIRCUIT 2
CIRCUIT 3
CIRCUIT 4
(4) 125V, 20A, DUPLEX OUTLETS - RED
NON-ISOLATED POWERSYSTEMS
INTERNAL FIXTURE WIRINGTYPICAL WIRE TYPE:
12AWG, 600V, XHHW-2, 90°CWHITE, BLACK, GREEN
ARM COVERARM COVER
ARM COVERARM COVER
GROUND(FOR ELECTRICAL)
PROTECTED EARTH(FOR CHASSIS)
NORMALDUPLEX
CRITICALDUPLEX
CRITICALDUPLEX
GENERIC BOOM ELECTRICAL WIRINGDIAGRAM FOR FIXED / SPRING ARMS
LGN
GENERIC WIRING DIAGRAM
ULTRAFLEXTYPE USL CONDUIT
ULTRAFLEXTYPE USL CONDUIT
GREENBLACK (L)WHITE (N)
GREENBLACK (L)WHITE (N)
CRITICAL NORMAL
*SITE SPECIFIC WIRING DETAILS* 6/20/2
013
MODE
L #: E
CT2F
KM48
/2AFC
2/AUT
75QT
Y.: 3
REV.
#: 0
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
NORT
HWES
T HO
SPIT
AL
RED
RED
WHITE
WHITE
(4) 125V, 20A, DUPLEX OUTLETS - WHITE
DESC
RIPT
ION:
LIGH
T FI
XTUR
E DE
TAILS
B6SHEETINITIAL:
DATE:
JUNCTION BOX BY OTHERSWITHIN 12" OF MOUNT
TYPE USL CONDUITBY SKYTRON
3/4" METAL CONDUIT 12AWG,3 WIRES PER LIGHTHEADRED, GREEN, WHITE30VDC OUTPUT
3/4" METAL CONDUIT120VAC, 15AINPUT DEDICATED
WALL CONTROL(SKYTRON SUPPLIED)
TO BE MOUNTED NEARENTRANCE DOOR OR
AT N.D.C.
THIS DIAGRAM IS FOR INFORMATIONAL PURPOSES ONLY.THIS WILL NOT MATCH YOUR EXACT MODEL.
FLOOR
SPECIAL GROUNDING REQUIREMENTS - Electrical Engineer
Proper performance and safety of this fixture can only be achieved by an adequate grounding system. Fixture ground must be a dedicated ground point ultimately bonded to the facilities grounding system to prevent the migration of electrical interference generated by other devices.
Notes:• 2 Dedicated conduit runs required at wall control to separate 120VAC input lines from 30VDC output lines to light fixture to prevent migration of
electrical magnetic interference which will disrupt the operation of the light.• **No shared ground.** Each light head must have separate individual ground.
GENERIC AURORA WALL CONTROLMOUNTING DETAILS
RECESSED MOUNT
A B
C
ANCHOR (4)
RECESS MOUNTFLANGE
SCREW (4)
RECESS DEPTH
FRONTPANEL
SURFACE MOUNT
ANCHOR (4)
E
D
SCREW (4)
FRONTPANEL
OPTIONAL BACK-BOX INSTALLATION
ANCHOR (4)
RECESS MOUNTFLANGE
SCREW (4)
GROMMETS (2)
CONTROL BOX WITH RECESSMOUNT FLANGE INSTALLED
"BACK-BOX" DIMENSIONSSINGLE / DUAL
8-1/2"W x 11-5/8"H x 4"DTRIPLE
10-1/2"W x 14-1/2"H x 4"D
FINISHED WALL120VAC
30VDC
CONDUIT CONNECTOR
3/4" CONDUIT
DIMENSION
SINGLE / DUAL
8"
4"
10"
6 - 7/8"
7 - 5/8"
5 - 7/8"
11 - 3/4"W x 14"H
TRIPLE
10"
4"
13 - 1/2"
8 - 5/8"
11"
6 - 3/8"
13 - 3/4"W x 17 - 1/4"HRECESS MOUNT FLANGE
A
B
C
D
E
F
F3"
C
3"
A
1"
JUNCTION BOX
OPA #: OPA-1607-07
*GENERIC LIGHT FIXTURE DETAILS* 6/20/2
013
MODE
L #: E
CT2F
KM48
/2AFC
2/AUT
75QT
Y.: 3
REV.
#: 0
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
NORT
HWES
T HO
SPIT
AL
C1SHEET
STRUCTURAL REQUIREMENTS - Architect and Structural Engineer
Mounting Structure ComponentsThe fabrication of each mounting structure may be slightly different but they each require the same basic components to ensurestability.
Sway Bracing (by others)Sway bracing is designed to rigidly affix the mounting plate to the structural ceiling. The primary purpose of sway bracing is toeliminate sway, or lateral twisting and flexing of the mounting structure as it "reacts" to dynamic load changes caused by moving thefixture radial arms. The sway bracing should be welded to the mounting plate and extend away from the center of the mount. Aminimum of four sway braces place 90° apart and positioned at a 45° and 60° angle is recommended.
Minimum recommended material for sway bracing is 3" x 3" x 1/4" angle iron. It is recommended that in all applications that thesway bracing be fastened to the structural ceiling.
Structural Ceiling Plate (by others)The structural ceiling plate rigidly attaches the mount to the structural ceiling using structural anchors appropriate for the ceilingconstruction. The structural ceiling plate should be a minimum of, 1" ASTM A36 steel plate, 17" diameter with (6) 5/8" diameterholes for structural anchors and is fabricated by others.
Expansion AnchorsTest 50% of the anchors at 2,000 pounds (907 kg) tension, or 50 ft. lb. (68 N●m) torque per CBC 1925A.3.5. Installed anchors mustmeet the following criteria:1. Hydraulic Ram Method: The anchor should have no observable movement at the applicable test load. For wedge and sleevetype anchors, a practical way to determine observable movement is that the washer under the nut becomes loose.2. Torque Wrench Method (Wedge or Sleeve Type): The applicable test torque must be reached within one-half (1/2) turn of the nut.Testing should occur no sooner than 24 hours after installation of anchors. If any anchor fails testing, test all anchors until 20 consecutive anchors pass, then resume the initial testing frequency. Test equipment is to be calibrated by an approved testinglaboratory in accordance with standard recognized procedures.
Support Tube (by others)The support tube required to attach the mounting plate to the structural ceiling plate is ASTM 500 Grade B, 6" diameter tube. Support tube is to be welded to structural ceiling plate and mounting plate. A minimum of 6 gussets placed 60° apart should bewelded to support tube at the structural ceiling plate and the mounting plate.
Mounting Plate (SKYTRON supplied)The 17.5" x 17.5" x 1" ASTM A36 steel mounting plate is a SKYTRON supplied item. The Support tube and sway bracing are welded to the mounting plate. The mounting plate contains the corresponding bolt pattern for attaching the fixture and provides themounting areas for the junction box and gas riser plates.
Mounting Structure DesignSeismic structural applications may differ. Please contact your local SKYTRON distributor for specific calculations. The mountingstructure must be designed and fabricated to position the bottom of the SKYTRON mounting plate 10" (-0", +1") above thefinished ceiling. This is a critical dimension in order to accommodate proper clearance required for ceiling cover function. Themounting plate must be perfectly level (+/- 0.1°) and allow no more than two-tenths of a degree (0.2°) of rotation at the mountingplate when the specified load is applied. The mounting structure must be tested for strength and stiffness prior to installation ofthe fixture. Please contact your SKYTRON representative to schedule testing.
Please consult your SKYTRON representative during early stages of construction to facilitate this process. The testingprocess is a required, documented function prior to closing of the finished ceiling.
Ceiling RequirementsA 24" x 24" access door must be mounted adjacent to the mounting structure for entry by service personnel for serviceaccess.
SKYTRON provides a 24" ceiling cover designed to cover 20" diameter ceiling hole cutout.
Additional Skytron Supplied ItemsIn addition to the pre-installation kit. SKYTRON provides the following items:
(6) 1-1/4" x 10" threaded rods. (24) 1-1/4" hex nuts, pump enclosure (if applicable)INITIAL:DATE:
STRUCTURALCEILING
STRUCTURALCEILING
STRUCTURAL CONCRETE SLABNORMAL WEIGHTfc=3,000 PSI (20,684 kPa)
SWAY BRACING3" x 3" x 1/4" (4)
GUSSET 6" DIA.SUPPORT
TUBE
FINISHED CEILING
20"DIAMETER OPENING
REQUIRED(CENTERED UNDER MOUNT)
12" MIN. / 60" MAX.
*10" (-0"/+1") TYPICALREFER TO JOB SPECIFIC ELEVATION ABOVE
MOUNTING PLATE(SKYTRON SUPPLIED)
45°- 60°
FIXTURE HUBTYP. 1/4"
PJP
ACCESS DOOR24" x 24"
(REQUIRED)
TYP. 1/4"TOP & BOTTOM
STRUCTURALCEILING PLATE
17" DIA. x 1"
EXPANSION ANCHORS:HILTI KB-II (ICBO 4627)
CARBON STEEL
NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult
specific structural criteria defined by a structural engineer.• Do not cover or block any holes with sway bracing, gussets, weld, weld slag or etc.• Typical dimensions shown. Refer to specific structural drawings and/or Seismic drawings for each
application.• *Critical Dimension
11" HIGH CLEAR SPACEABOVE MOUNTING PLATE
REQUIRED ON 3 SIDES FORGAS AND ELECTRICAL
COMPONENT INSTALLATION
FIXTURE WEIGHT:MOMENT LOAD:
809 LBS.4584 FT. LBS.
OPA NUMBER: 2510-07
DESC
RIPT
ION:
ELE
VATI
ON D
ETAI
LS
120" CEILING HEIGHT
0" FLOOR
*SITE SPECIFIC ELEVATION DETAILS* 6/20/2
013
MODE
L #: E
T2FV
BM36
/2AFC
2QT
Y.: 3
REV.
#: 0
PROJ
ECT
#: 13
-072
-KP
SUBM
ITTA
LPL
OT D
ATE:
72"
61"
31 1/2"
11 1/16"
3" 10"
51 3/16"
35 7/16"
12"
18"
70"
34"
25"
NORT
HWES
T HO
SPIT
AL77"
9"
WHITE - REQUIRES"J" BOX
RED
MEDICAL AIR
VACUUMVACUUM
VERTICAL NITROGENSYSTEM - DISS
NITROUS OXIDE
WAGDOXYGEN
OXYGEN
CHEMETRON
ACCESSORY LIST(2) BOLT ON VACUUM SLIDE
37"H x 11 1/2"W x 8"D(4) 125V, 20A DUPLEX - RED
(2) 125V, 20A DUPLEX - WHITEDIMENSIONS:
NORT
HWES
T HO
SPIT
AL
INITIAL:C2
PLO
T DA
TE: 6
/20/
2013
ELECTRICAL:GAS OUTLETS:
PRO
JECT
#: 1
3-07
2-KP
DESC
RIPT
ION:
CAR
RIER
DET
AILS
SHEET
SUBM
ITTA
LQ
TY: 3
REV
#: 0
DATE:
MO
DEL
#: E
T2FV
BM36
/2AF
C2
CARRIER
*SITE SPECIFIC CARRIER DETAILS*
C3SHEET
MEDICAL GAS REQUIREMENTS - Medical Gas / Piping Engineer
Notes:• Ohio Medical manufactured gas outlets supplied unless otherwise specified.Pricing and lead time could vary if another MFGR is chosen.• Medical gas riser plates comply with NFPA 99-2005 Edition Cleanliness of gas outlets must be maintained through installation.• Purge gas lines and test at least 24 hrs. prior to installation.• Test results must be made available to SKYTRON for verification & comparison.• Check valve must be removed during installation to prevent damage to o-rings, refer to Ergon pre-installation guide for removal
instructions.** Caution - Check valves have 2 different sizes, please keep Instrument Air and Oxygen separate from the others **
Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The configuration drawingssupplied by SKYTRON will indicate the type and quantity of gas supply lines required. D.I.S.S. connection medical grade hosesconnect the fixture to the riser plate connectors. The customer is responsible to deliver the appropriate medical gas from thefacility supply to the riser plate connectors and for the design of a medical gas system with adequate flow capacity capable ofcompensating for the accumulative flow restrictions associated with conventional construction methods i.e. flex gas hoses.SKYTRON provides medical gas riser plate(s) and the appropriate connectors for attachment to the Mounting Plate.The riser plate attaches to the mounting plate and will accommodate up to 12 gas connectors. The connectors are D.I.S.S.type medical gas connectors with removable single check valves and provide a 1/2" copper tube for attachment to facility supplylines. Caps are provided with Vacuum & WAGD gas connectors to facilitate testing procedures.
All connection and testing of medical gas piping to be performed in accordance with NFPA 9-2005 Edition-5.1.6 requirements.
Recommended Gas Riser PlacementWhen mount is in center of room, riser plate should face head end of room, when mount is near walls, riser plate should be adjacentto wall. If more than one riser plate is required they should be placed next to each other.
INITIAL:DATE:
GAS CONNECTOR ASSEMBLY (w/ SECONDARY CHECK)*SUPPLIED w/FIXTURE*
9"
DISS CONNECTOR(w/ REMOVABLE PRIMARY
CHECK. PASSIVE EVAC,VACUUM & WAGD DO NOT
HAVE PRIMARY CHECK.)*SUPPLIED w/MOUNTING KIT*
D I S SCHEMETRON
* VERIFY AND INITIAL GAS FACEPLATE STYLE REQUESTED *
CGA ColorStandard
Note #1 - Any room (Critical Care Area) designed for a permanently located respiratory ventilator or anesthesia machine shall have an outlet capable of a transient flow rate of:170 LPM (6 SCFM) for 3 seconds at the station outlet.
Note #2 - For testing and certification purposes, individual station inlets shall be capable of a flow rate of: 3 SCFM, while maintaining a system pressure of not less than 12" (300mm) at the nearest adjacent vacuum inlet. Facility supply must be 115 LPM MINIMUM. (Vacuum D.I.S.S. connectors omit primary check valves for optimal flow). 12in/HG.
Note #3 - WAGD (Waste Anesthetic Gas Disposal) systems employing a design where the WAGD lines are "tied in" to MedVac lines must produce the same flow rates as theMedVac inlets.
Note #4 - Nitrogen system requires nitrogen supplied directly from facility supply line rated at 185psi MIN to 200psi MAX. Avoid designs which feature multiple-in-line Nitrogencontrol systems in order to avoid loss of flow capability.
Additional references: Health Care Facilities Handbook 2002, Section 5.1.11, page 211 and section 5.1.12.13.10.1 thru 5.1.12.3.10.5, page 224. NFPA 99, 2002 guideline figure A.5.1.6.
• TYPE K COPPERTUBING
• 1/2" O.D. UNLESSOTHERWISE STATED
Medical Air(Yellow)
Test GasCanada Color
StandardMedical Air
(Black)
AbbreviatedName
MedAir
StandardPressure
50 - 55 psig
MaximumPressure
55 psig
AllowablePressure Drop
5 psig
Minimum Flow Rates
3.5 SCFM per outlet (100NL/min)N2NF see note #1Carbon Dioxide
(Gray)Carbon Dioxide
(Gray) CO2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFHeliox
(Brown)Heliox
(Brown) Heliox 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NFNitrogen(Black)
Nitrogen(Black) N2 or HPN2 160 - 185 psig 200 psig 5 psig
5 SCFM per outlet (140NL/min)free air per outletN2NF see note #4
Nitrous Oxide(Blue)
Nitrous Oxide(Blue) N2O 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF
Oxygen(Green)
Oxygen(White) O2 50 - 55 psig 55 psig 5 psig 3.5 SCFM per outlet (100NL/min)N2NF see note #1
Vacuum(White)
Vacuum(Yellow) MedVac 12in/Hg (300mm) N/A 3 SCFM per outlet (85NL/min) see note #2
Waste AnestheticGas Disposal
(Purple) WAGDVaries withsystem type
At a minimum, each inlet must be able todraw a continuous 50 lpm (1.8 SCFM)
through the interface
Waste AnestheticGas Disposal
(Purple) see note #3
PURITANOHMEDA
NOTES:• This illustration depicts a generic mounting structure design and its components. Always consult
specific structural criteria defined by a structural engineer.• Mounting bolts and nuts shipped with fixture.
INLET TUBE
RISER PLATE
JUNCTIONBOX
BRACKET
GENERIC RISER PLATE INSTALLATION
DESC
RIPT
ION:
MED
ICAL
GAS
DET
AILS
*SITE SPECIFIC GAS DETAILS*
6/20/2
013
NORT
HWES
T HO
SPIT
ALPR
OJEC
T #:
13-0
72-K
PSU
BMIT
TAL
PLOT
DAT
E:
MDL:
ET2F
VBM3
6/2AF
C2QT
Y.: 3
REV.
#: 0
CHECK VALVEREMOVAL TOOL
OXYG
EN
OXYG
EN
NITR
OUS
OXID
E
NITR
OGEN
MED
ICAL
AIR
WAG
D
VACU
UM
VACU
UM
DESC
RIPT
ION:
ELE
CTRI
CAL W
IRIN
G DE
TAILS
C4SHEET
ELECTRICAL REQUIREMENTS - Electrical Engineer
Each Skyboom fixture is fabricated in accordance to the specifications required by the customer. The Configuration drawings supplied by SKYTRON will indicate the type and quantity of circuits required. SKYTRON provides all wiring and electrical materials for connection from fixtureto junction box or pump enclosure (if applicable). SKYTRON supplies either an electrical junction box (8-5/8" x 4-5/8" x 1-3/4") to facilitate fieldwiring for up to six circuits that is mounted on the mounting plate in the correct position OR if applicable, a hydraulic pump enclosure/junction box(18"L x 6"W x 12"H) that is to be remote mounted within 24" of the mounting structure (by contractor). The pump enclosure can be shipped withthe installation kit upon request. Typical wire type is 12AWG, 600V, XHHW-2. Each circuit requires a separate, properly circuit protected, 120VAC, 60Hz power supply line enclosed in rigid metal conduit. All electrical materials for connection to SKYTRON supplied junction box or pump enclosure and installation labor for such materials to be provided by customer. All wiring and materials to be in accordance with federal,state and local codes. It is the customer's responsibility to meet conformity to NFPA and NEC standards with respect to the number of receptacles provided in a patient care area.
Specific conductor colors and/or wiring for isolated applications are available upon request.
INITIAL:DATE:
3 INPUT LEADS
CARRIER UTILITIES PANELFIXTURE JUNCTION BOX
AT MOUNTING PLATE
120 VAC, 20 AMP
120 VAC, 20 AMP CIRCUIT 1
CIRCUIT 2
(4) 125V, 20A, DUPLEX OUTLETS - RED(2) 125V, 20A, DUPLEX OUTLETS - WHITE
NON-ISOLATED POWERSYSTEMS
INTERNAL FIXTURE WIRINGTYPICAL WIRE TYPE:
12AWG, 600V, XHHW-2, 90°CWHITE, BLACK, GREEN
120 VAC, 20 AMP CIRCUIT 3
ARM COVERARM COVER
ARM COVERARM COVER
GROUND(FOR ELECTRICAL)
PROTECTED EARTH(FOR CHASSIS)
NORMALDUPLEX
CRITICALDUPLEX
CRITICALDUPLEX
GENERIC BOOM ELECTRICAL WIRINGDIAGRAM FOR FIXED / SPRING ARMS
LGN
GENERIC WIRING DIAGRAM
ULTRAFLEXTYPE USL CONDUIT
ULTRAFLEXTYPE USL CONDUIT
GREENBLACK (L)WHITE (N)
GREENBLACK (L)WHITE (N)
CRITICAL NORMAL
*SITE SPECIFIC WIRING DETAILS* 6/20/2
013
NORT
HWES
T HO
SPIT
ALPR
OJEC
T #:
13-0
72-K
PSU
BMIT
TAL
PLOT
DAT
E:
MDL:
ET2F
VBM3
6/2AF
C2QT
Y.: 3
REV.
#: 0
RED
RED
WHITE
INITIAL:DATE:
MOUNTING STRUCTURE TEST JIG
WEIGHT SUPPORT TUBE
1-1/4" NUTS
1-1/4" THREADED RODS (6)
MOUNTING HUBASSEMBLY
12mm NUT (6)
12mm x 30mm BOLT (4)
CHAIN
CHAIN FALL
LOWER RADIAL ARM
UPPER RADIAL ARM
5/8" NUT (2)
12mm NUT (2)
WEIGHTS (6)
TESTING - MOUNTING STRUCTURE
Each mounting structure must be tested by a Skytron representative to verify that there is no more than two-tenths of a degree of rotational movement at the mounting plate prior to installing a fixture. A mounting structure test jig is available from a Skytron representative to facilitate thisprocess. The mounting structure test jig is a fixture which simulates the weights and moment loads created by a Skytron fixture. It consists of ahub assembly, an upper and lower radial arm, a chain fall, weight support tube and six 100 lb. weights (refer to illustration).
The drawing package and the test jig instructions are required to perform an accurate test.
The test jig instructions include the load simulator weight chart, the test jig report form and complete instructions for performing the test.
upper armlength (major)
lower armlength (major)
upper armlength (minor)
lower armlength (minor)
† equipmentweight capacity
minimumceiling height fixture weight moment load * test weight
200 lbs400 lbs400 lbs400 lbs600 lbs600 lbs600 lbs600 lbs600 lbs600 lbs
600 lbs600 lbs600 lbs
500 lbs
600 lbs600 lbs200 lbs300 lbs
400 lbs400 lbs
N/A
200 lbs300 lbs
100 lbs200 lbs
1410 ft lbs2843 ft lbs3074 ft lbs3149 ft lbs4584 ft lbs4272 ft lbs6102 ft lbs5318 ft lbs4158 ft lbs4526 ft lbs
4362 ft lbs5239 ft lbs5607 ft lbs
3994 ft lbs
4927 ft lbs5288 ft lbs1690 ft lbs2058 ft lbs
2888 ft lbs6256 ft lbs
N/AN/A
1204 ft lbs1810 ft lbs
549 lbs622 lbs611 lbs646 lbs809 lbs830 lbs
1100 lbs1108 lbs743 lbs832 lbs
858 lbs939 lbs
1028 lbs
769 lbs
960 lbs1049 lbs397 lbs486 lbs
597 lbs686 lbs
N/AN/A
478 lbs434 lbs
145 lbs
N/AN/A
N/AN/A
N/AN/A
N/AN/A
N/AN/A
N/AN/A
650 ft lbs
7'-9"8'-8"8'-2"8'-5"
9'-11"9'-11"8'-9"8'-9"8'-9"9'-5"8'-9"9'-5"
9'-6" ***9'-11"
9'-6" ***9'-11"9'-6"9'-6"
9'-11"9'-11"9'-0"7'-7"8'-5"
275 lbs275 lbs
N/A275 lbs
205/66 lbs210/66 lbs
205/175 lbs225/175 lbs
215 lbs215 lbs
255 lbs215/66 lbs215/66 lbs
255 lbs
255/66 lbs255/66 lbs
66 lbs66 lbs
66/66 lbs66/66 lbs
235 lbs184 lbs
N/A
N/AN/AN/AN/A
N/AN/AN/AN/A
N/A
N/AN/AN/AN/A
N/AN/A
N/AN/A N/A
N/AN/A
N/AN/AN/AN/AN/A
N/AN/AN/A
N/AN/A
35.5"35.5"
35.5"35.5"35.5"35.5"
35.5"35.5"35.5"
35.5"
35.5"35.5"
31.5"31.5"31.5"
31.5"
31.5"31.5"31.5"31.5"31.5"31.5"31.5"31.5"31.5"31.5"31.5"31.5"31.5"
39.5"
39.5"39.5"39.5"
39.5"39.5" 31.5"
44.5"
44.5"
61"
61"
61"61"
61"61"
51"
51"51"
51"51"51"51"
51"51"
51"51"
51"
59"59"
41.5"48"
41.5"41.5"
E1_E2_
E2VBM36E2H_
ET2_/2AF_ET2-C_/2AF_
ETM2_/2_ETM2-C_/2-C_EC2_/LIGHT(1)
EC2_/LIGHTS(2)EC2-C_/LIGHT(1)
EC2-C_/LIGHTS(2)ECT2_/2AF_/LIGHT(1)
ECT2_/2AF_/LIGHTS(2)ECT2-C_/2AF_/LIGHT(1)
ECT2-C_/2AF_/LIGHTS(2)LC2AF_/LIGHT(1)
LC2ALF_/LIGHTS(2)LCT2AF_/2AF_/LIGHT(1)
LCT2AF_/2AF_/LIGHTS(2)COLUMN (ALL)
Q1_Q2_
(ALL) LFSLFS(ALL) EXTENDED LIGHT ARMS
BOOM
* SKYTRON recommends testing all mounts to 600 lbs **** Center of flatscreen will be 70" at level† Refer to Operator's Manual for exact weight capacity
LC2ALF_/LIGHT(1)LC2AF_/LIGHTS(2)
300 lbs1800 ft lbs400 lbs9'-6"66 lbsN/AN/A35.5"59"300 lbs2150 ft lbs500 lbs9'-6"66 lbsN/AN/A35.5"59"