objective testing of flicker and stroboscopic effects resulting from

22
Pierre Beeckman Philips Innovation Services May 23, 2014 Objective testing of flicker and stroboscopic effects resulting from mains voltage fluctuations

Upload: lyliem

Post on 14-Feb-2017

219 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Objective testing of flicker and stroboscopic effects resulting from

Pierre Beeckman Philips Innovation Services

May 23, 2014

Objective testing of flicker and stroboscopic effects resulting from mains voltage fluctuations

Page 2: Objective testing of flicker and stroboscopic effects resulting from

2 May 23, 2014 Philips Innovation Services

• Modeling & Simulations • Design Methods & Tools • Standardization

• Design reviews • Risk analysis • Trainings & audits

• (Pre-)compliance EN, IEC/CISPR, FCC, ETSI • ISO/IEC 17025 accreditation

Technology development

Consultancy Testing

Competences:

EMC

PQ

ESD

SI / PI

EMF

Wireless

EMC Center Philips Innovation Services

Page 3: Objective testing of flicker and stroboscopic effects resulting from

3 May 23, 2014 Philips Innovation Services

Introduction

Page 4: Objective testing of flicker and stroboscopic effects resulting from

Why ‘flicker’ matters

Human impact

• Headaches

• Neurological problems: photosensitive epilepsy

• Autistic sensitivity

• Performance reduction

Safety

• Distraction

• Possible hazard from stroboscopic effect: apparent

stopping or slowing of motion of machinery

4 May 23, 2014 Philips Innovation Services

Page 5: Objective testing of flicker and stroboscopic effects resulting from

5 May 23, 2014 Philips Innovation Services

Different types of temporal light artefacts

Flicker

• The perception of temporal changes in the intensity

(luminance flicker) or color (chromatic flicker) of the light

Stroboscopic effect

• The perception that objects illuminated by fluctuated

• light move discretely rather than continuously

Ghost effect (phantom array)

• The perception of a spatially extended series of light spots

when making a saccade across a light spot that fluctuates

over time

Page 6: Objective testing of flicker and stroboscopic effects resulting from

6 May 23, 2014 Philips Innovation Services

Root Cause: light output modulation Light source technology • Incandescent, fluorescent, solid-state: inherently they have different flicker/strobo properties

Drivers • Electronics convert AC mains to ‘DC’ LED current and voltage • Always some ripple in the LED current will remain which translates toripple in the light output • Suppressing the ripple requires buffering which costs money, physical space, and impacts lifetime

Dimming • Dimming SSL sources is challenging particularly with phase-cut dimmers • Wide variation & no standard test procedure

Camera interactions • Artifacts also result from camera's interaction; especially sensitive for high-speed cameras

Mains voltage fluctuations • Arising from the grid or from other connected equipment or systems

Dimmer voltage

‘variations’ supply

Driver

Light source technology

EMC/PQ

Page 7: Objective testing of flicker and stroboscopic effects resulting from

7 May 23, 2014 Philips Innovation Services

Examples light ripple due to driver & dimming

single stage SMPS

dual stage SMPS

linear (tapped-linear driver)

linear (quasi-DC)

some examples for major driver categories

examples of SSL lamp + phase-cut dimmer

Page 8: Objective testing of flicker and stroboscopic effects resulting from

Visibility: when can people see the stroboscopic effect and flicker?

Flicker influence quantities • frequency • mean luminance • retinal position of the stimulus • size of the stimulus • shape of the waveform (e.g. sinusoidal, squares) • adaptation state of the observer • perceptual load of the observer

Metrics (using light output):

• Flicker Index 𝐹𝐼 = 𝐴𝑟𝑒𝑎 1

𝐴𝑟𝑒𝑎 2+𝐴𝑟𝑒𝑎 1

• Modulation depth (flicker percent)

MD =𝐿𝑚𝑎𝑥 − 𝐿𝑚𝑖𝑛

(𝐿𝑚𝑎𝑥+ 𝐿𝑚𝑖𝑛) ∗ 100%

• IEC short-term flicker metric (Pst)

• Philips Stroboscopic Visibility measure (SVM)

light output

8 May 23, 2014 Philips Innovation Services

Frequency dependent ‘summation’ of spectrum of light output

Page 9: Objective testing of flicker and stroboscopic effects resulting from

9 May 23, 2014 Philips Innovation Services

Standardization activities

Page 10: Objective testing of flicker and stroboscopic effects resulting from

Future: diversity of lighting technologies & diversity of voltage-fluctuation immunity performance

requires flicker immunity test

Voltage fluctuations: low-frequency EMC

∆U voltage ‘variations’

on the mains

Appliances (threat)

Lighting equipment (victim)

Emission test: IEC 61000-3-3 voltage changes , -fluctuations & flicker (regular & fast)

EMC

Until now: no flicker immunity test for lighting equipment; incandescent

lamp is considered the most sensitive apparatus connected to a

LV-mains

11 May 23, 2014 Philips Innovation Services

Page 11: Objective testing of flicker and stroboscopic effects resulting from

10 May 23, 2014 Philips Innovation Services

IEC flickermeter: emulates average perception of flicker from a 60 W incandescent lamp IEC

Mains voltage + variations

∆U

Un

Various flicker metrics and limits for various product categories: Pst

Plt

Pinst

d dmax

Limits & test conditions in

IEC 61000-3-3

flickermeter (IEC61000-4-15) emulating 60 W lamp + eye/brain response

Page 12: Objective testing of flicker and stroboscopic effects resulting from

12 May 23, 2014 Philips Innovation Services

Flicker and stroboscopic effect testing preparatory work IEC TC34-MT1 on future objective immunity test method lighting equipment IEC TR 61547-1

Page 13: Objective testing of flicker and stroboscopic effects resulting from

IEC voltage flickermeter & the light flickermeter

Mains voltage

13

light flickermeter

Model 60 W incandescent lamp

Light output

IEC 61000-3-3 flickermeter

May 23, 2014 Philips Innovation Services

Flicker metrics

Page 14: Objective testing of flicker and stroboscopic effects resulting from

Block diagram PInS test setup Flicker & strobo metrics:

General: MD: Modulation depth/flicker percent FI: Flicker index

Philips: SVM: Strobo effect visibility measure FVM: Flicker visibility measure

IEC: Pst: short-term flicker

Synthesized a.c. mains +

(optional) voltage fluctuations

NEMA SSL 7A phase cut dimmer

(optional)

Light flicker- and

stroboscopic effect meter

EUT Light sensor

Measurement test voltage properties

Metrics

General: rms, max, min, mean

IEC: d: relative voltage fluctuation Pst: short-term flicker

14 May 23, 2014 Philips Innovation Services

Page 15: Objective testing of flicker and stroboscopic effects resulting from

Photograph setup Ulbricht sphere including EUT &

CIE compliant light sensor

NI USB-6251 data acquisition

16 bits; 16 inputs, 2 outputs

Test voltage signal generation

&

Calculation of flicker & strobo metrics

using Matlab

Chroma 61601 amplifier

Inverto Light sensor amplifier

Tektronix DPO 4104

Oscilloscope (monitoring)

15 May 23, 2014 Philips Innovation Services

Page 16: Objective testing of flicker and stroboscopic effects resulting from

All test levels (Pst=1) for immunity testing (draft of IEC TR 61547-1)

16

Rectangular amplitude modulations with duty cycle of 50 % 1) 3) 4)

Voltage changes per minute Frequency

Relative voltage fluctuation

d =∆U/U

(cpm) (Hz) (%)

39 0,3250 0,894

110 0,9167 0,722

1056 8,8 0,2752)

1620 13,5 0,407

4000 33,3 2,343

NOTE 1 – See Table 5 of IEC 61000-4-15

NOTE 2 – See Table 2b of IEC 61000-4-15 for P inst =1; the value of d = 0,196 % is increased to 0,275 % to give Pst =1

NOTE 3 –Duration of the voltage fluctuation recommended to be minimally 60 seconds

NOTE 4 – Recommended absolute tolerance for the duty cycle is ±2 %, for the modulation frequency the recommended tolerance is ±1 % and for the relative voltage fluctuation the recommended tolerance is ±2 %

1

May 23, 2014 Philips Innovation Services

Page 17: Objective testing of flicker and stroboscopic effects resulting from

Example test voltage fluctuation (1) EUT: 9 W self-ballasted CFL lamp

Test voltage: d = 0,407 % @ 13,5 Hz (Pst = 1)

Relative illuminance

0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.5-400

-300

-200

-100

0

100

200

300

400

Time(sec)

Main

s v

oltage (

V)

EUT1

17

0 0.05 0.1 0.15 0.2 0.25 0.3 0.35 0.4 0.45 0.50.6

0.7

0.8

0.9

1

1.1

1.2

1.3

Time (sec)

Rela

tive illu

min

ance

EUT2

May 23, 2014 Philips Innovation Services

Page 18: Objective testing of flicker and stroboscopic effects resulting from

100

101

0

0.2

0.4

0.6

0.8

1

Frequency(Hz)

Pst

(pu)

Results voltage fluctuation immunity tests

Pst=1

EUT1

EUT2

EUT3

Example test voltage fluctuation (2) Results for 3 EUTs @ all test points

18

EUT1: 60 W incandescent lamp EUT2: 9 W self-ballasted CFL lamp EUT3: 7 W self-ballasted LED lamp

May 23, 2014 Philips Innovation Services

Page 19: Objective testing of flicker and stroboscopic effects resulting from

Summary

19

• Flicker and stroboscopic effect are important light quality aspects for lighting equipment

• There is an increasing diversity of flicker sensitivity of lighting equipment

• New lighting technologies drive – EMC and performance standards to include requirements for flicker and/or

stroboscopic effects – Need for an immunity test method, metrics and limits

• An objective way of flicker and strobo testing is developed, capable of – Applying a variety of test voltages – Analyzing illuminance waveforms using a variety of metrics

May 23, 2014 Philips Innovation Services

Page 20: Objective testing of flicker and stroboscopic effects resulting from

20 May 23, 2014 Philips Innovation Services

Further reading

• M. Perz et al, Modeling the visibility of the stroboscopic effect occurring in temporally modulated light systems, Lighting Research and Technology, May 13, 2014 : http://lrt.sagepub.com/content/early/2014/05/12/1477153514534945.full.pdf?ijkey=GcQ3UW7Qz2UwqtM&keytype=ref

• US DoE – Solid-state lighting technology fact sheet:

http://apps1.eere.energy.gov/buildings/publications/pdfs/ssl/flicker_fact-sheet.pdf

Page 21: Objective testing of flicker and stroboscopic effects resulting from

21 May 23, 2014 Philips Innovation Services

Acknowledgement

This work is supported by the Eniac Joint Undertaking project Enlight: Energy efficient and intelligent lighting systems

Page 22: Objective testing of flicker and stroboscopic effects resulting from