OCL (Oslo Cyclotron Lab)
Scanditronix MC 35 @ OCL
![Page 2: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/2.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 2
OCL Statistics & Achievements
In addition we spend ~200hrs Beam Time on a 11C PET Target installation which led to the succesful production of 11C
in December 2013.
OCL Beam Time Statistic 2013
Beam Proton Deuteron 3He 4He Overall
Hours ~600 ~200 ~280 ~320 ~1400
![Page 3: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/3.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 3
Troubles...
● RF Short-Circuit inside Cyclotron
● Anode Power Supply
● Accidently moving of Deflector
● Moisture
![Page 4: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/4.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 4
RF Short-Circuit inside Cyclotron
Fault Indication:
● RF would not start
● "Cathode delay" interlock on RF-2 turn off with irregular intervals
● "Air Grid" interlock present, with airflow OK
● Problem tuning RF-1, no reading on U-dee meter on control desk even when attempting pulse start
![Page 5: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/5.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 5
RF Short-Circuit inside Cyclotron
What to look for...
Second harmonic problem on RF-1: Seems independent of grid tuning capacitor
RF Detector for RF1 & 2 Anode & Dee Pickoffs RF 1 & 2
![Page 6: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/6.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 6
RF Short-Circuit inside Cyclotron
Reason: Grounding Mesh came loose and short-circuited the RF1-Dee
![Page 7: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/7.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 7
Anode Power Supply
Fault indication:
● Power supply would not start
● Error indicators
Reason:
● Airflow monitoring circuit broken
![Page 8: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/8.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 8
Accidently moving of Deflector
What happened:
● After a Blackout the Main Sond would not move
● Locating the responsible Servo Control Unit
● Accidently activated the Deflector Positioning Servos
● Position readback not plausible
![Page 9: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/9.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 9
Accidently moving of Deflector
Reason & Consequnces:
● The Timing Belt connecting the position encoding potmeter with the dc servo motor has disintegrated over time, therefor the Deflector Position was unknown
● Manual repositioning the Deflector with factory settings
● Makeshift solution for replacing the Timing Belt
● New Servo Control Unit (Ongoing Project)
![Page 10: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/10.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 10
Moisture
Fault Indication:
● Moisture on all cooled Devices in Summer
Reason:
● Very high humidity and therefor a dew point above the cooling water temperature → Condensation
![Page 11: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/11.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 11
Servo Control Unit
● After the Blackout it was decided that the old Servo Amplifier Moduls (SAMs) should be replaced
● 3x 19 inch Rack Units with 11 DC SAMs
● 2x 19 inch Rack Unit for stepper motor control
● Prove of concept for 3x DC Servo and 1x Stepper Motor Controller (Deflector and Sonds) based on NI cRIO
● 2x more Units planed for the remaining 8 DC SAMs
![Page 12: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/12.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 12
Servo Control Unit
![Page 13: OCL Oslo Cyclotron Lab) - Rigshospitalet SCX User Meeting 2014.pdf · 2014-09-10 · J. C. Müller OCL for the Cyclotron Workshop 2014 11 Servo Control Unit After the Blackout it](https://reader034.vdocument.in/reader034/viewer/2022042107/5e86a8b958f7f502e224f7b5/html5/thumbnails/13.jpg)
J. C. Müller OCL for the Cyclotron Workshop 2014 13
OCL...
Thank you for your attention!