OHMEffect of low-to-moderate degrees
of refractive astigmatism onContrast sensitivity and Reading
speed in pseudophakic eyes
Shetal M. Raj, DO, MS, Abhay R. Vasavada, MS, FRCS, Vaishali Vasavada, MS, Viraj A. Vasavada, MS, M.R. Praveen, DO
The authors have received travel expense reimbursement from Alcon Laboratories, USA
Iladevi Cataract and IOL Research Center, Ahmedabad,
India
![Page 2: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/2.jpg)
To evaluate effect of low to moderate degrees of refractive astigmatism (0.5 to 0.75 diopters (D)) on low and high contrast sensitivity and reading speed in pseudophakic eyes.
Purpose
![Page 3: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/3.jpg)
• Prospective, controlled, observational study
Materials and Methods
Study Population
• 40 eyes of 40 pseudophakic patients with single-piece monofocal hydrophobic acrylic IOL
• Monofocal IOL• BCVA of 20/20 • Refractive astigmatism of 0.5 and 0.75D
Inclusion criteria
Study Design-
![Page 4: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/4.jpg)
1. ETDRS acuity using CSV- 3000 chart 100%, 25% , 10% contrast.
2. Contrast sensitivity using the Vector Vision chartsphotopic, mesopic and glare conditions
3. Reading acuity on Radner reading chartsLogRAD scoreMean reading speedMaximum reading speed
Observations
All observations were done monocular with - spherical equivalent (uncorrected cylinder) and - with sphere and cylinder correction (corrected cylinder)
![Page 5: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/5.jpg)
1. Distance contrast acuity
Results
0.5 cyl (n=21) 0.75 cyl (n=19)
S.E. Corr. P value
S.E. Corr. P value
100% -0.14 + 0.1 -0.08 + 0.11 0.0 -0.15 + 0.09 -0.07 + 0.1 0.0
25% -0.06 + 0.13 0.00 + 0.13 0.0 -0.06 + 0.13 0.05 + 0.12 0.0
10% 0.05 + 0.12 0.11 + 0.10 0.1 0.06 + 0.10 0.17+ 0.14 0.0
S.E. = Spherical equivalent (uncorrected cylinder)Corr. = Corrected sphere and cylinder (Corrected cylinder)
P = Paired sample T test
![Page 6: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/6.jpg)
2. Contrast sensitivity: photopic
Results
3 cpd 6 cpd 12cpd 18 cpd
S.E. Corr S.E. Corr. S.E. Corr. S.E. Corr.
0.5 cyl (n=21)
1.54 + 0.14
1.6 + 0.15
1.79 + 0.18
1.91 + 0.16
1.46 + 0.27
1.58 + 0.20
0.96 + 0.22
1.15 + 0.17
P value 0.1 0.01 0.03 0.00.75 cyl (n=19)
1.55 + 0.14
1.62 + 0.12
1.79 + 0.23
1.93 + 0.21
1.31 + 0.3
1.52 + 0.26
0.78 + 0.37
1.02 + 0.32
P value 0.01 0.01 0 0.01
S.E. = Spherical equivalent (uncorrected cylinder)Corr. = Corrected sphere and cylinder (Corrected cylinder)
P = Wilcoxon signed ranks paired test
![Page 7: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/7.jpg)
Contrast sensitivity: mesopic
Results
3 cpd 6 cpd 12cpd 18 cpd
S.E. Corr S.E. Corr S.E. Corr S.E. Corr
0.5 cyl (n=21)
1.39 + 0.17
1.45 + 0.19
1.41 + 0.3
1.55 + 0.27
0.87 + 0.31
1.01 + 0.33
0.45 + 0.29
0.57 + 0.31
P value 0.13 0.0 0.06 0.15
0.75 cyl (n=19)
1.4 + 0.25
1.48 + 0.21
1.41 + 0.23
1.52 + 0.29
0.85 + 0.34
0.84 + 0.31
0.42 + 0.32
0.47 + 0.34
P value 0.05 0.03 0.89 0.29
S.E. = Spherical equivalent (uncorrected cylinder)Corr. = Corrected sphere and cylinder (Corrected cylinder)
P = Wilcoxon signed ranks paired test
![Page 8: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/8.jpg)
Contrast sensitivity: glare conditions
Results
3 cpd 6 cpd 12cpd 18 cpd
S.E. Corr S.E. Corr S.E. Corr S.E. Corr
0.5 cyl (n=21)
1.3 + 0.22
1.45 + 0.19
1.34 + 0.33
1.43 + 0.31
0.88 + 0.24
0.99 + 0.33
0.33 + 0.29
0.40 + 0.27
P value 0.0 0.02 0.17 0.17
0.75 cyl (n=19)
1.31 + 0.2
1.41 + 0.24
1.21 + 0.31
1.4 + 0.32 0.74 + 0.28
0.81 + 0.26
0.22 + 0.14
0.39 + 0.29
P value 0.02 0.01 0.09 0.01
S.E. = Spherical equivalent (uncorrected cylinder)Corr. = Corrected sphere and cylinder (Corrected cylinder)
P = Wilcoxon signed ranks paired test
![Page 9: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/9.jpg)
3. Reading acuity on Radner reading charts
Results
LogRAD Score Mean Reading Speed
Max Reading Speed
S.E. Corr. S.E. Corr. S.E. Corr.
0.5 cyl (n=19)
-0.28 + 0.34
-0.29 + 0.39
117.08 + 29.42
120.42 + 25.53
142.08 + 31.63
155.14 + 34.67
P value 0.33 0.05 0.3
0.75 cyl (n=21)
-0.2 + 0.01 -0.2 + 0.0 115.32 + 25.44
126.32 + 24.53
145.16 + 33.29
161.22 + 35.11
P value 0.33 0.01 0.0
S.E. = Spherical equivalent (uncorrected cylinder)Corr. = Corrected sphere and cylinder (Corrected cylinder)
P = paired sample t test
![Page 10: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/10.jpg)
• The performance of eyes with 0.5 diopter cylindrical error fully corrected resulted in significantly better: – Distance contrast acuity at 100% and 25%– Photopic contrast at 6, 12, 18 cpd– Mesopic contrast at 6 cpd– Glare contrast at 3, 6 cpd– Mean reading speed
Summary
• The performance of eyes with 0.75 diopter cylindrical error fully corrected resulted in significantly better: – Distance contrast acuity at 100% , 25% and 10%– Photopic contrast at 3, 6, 12, 18 cpd– Mesopic contrast at 3, 6 cpd– Glare contrast at 3, 6, 18 cpd– Mean and maximum reading speed
![Page 11: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/11.jpg)
• Correction of astigmatism as low as 0.50 and 0.75 diopters improves the visual performance
• Surgeons must strive to adopt techniques and IOLs that eliminate the pre-existing
astigmatism, such as microincision phacoemulsification approach and using Toric IOLs
Implications
![Page 12: OHM Effect of low-to-moderate degrees of refractive astigmatism on Contrast sensitivity and Reading speed in pseudophakic eyes Shetal M. Raj, DO, MS, Abhay](https://reader035.vdocument.in/reader035/viewer/2022062519/5697bff11a28abf838cbb678/html5/thumbnails/12.jpg)
High and Low contrast acuity, contrast sensitivity, Reading acuity was significantly better when correcting 0.5 and 0.75 dioptres of refractive astigmatism as compared to correcting with spherical equivalent and not giving cylindrical correction in the spectacles.
Conclusion