One room - all applicationsPhilips EasyDiagnost Eleva Specification Digital radiography and fluoroscopy room with the DRF room solution
The EasyDiagnost Eleva is a R/F system designed to perform a wide range of
applications from gastro-intestinal, various iodine and vascular procedures to
standard radiography. By combining the EasyDiagnost Eleva with a choice of a
digital vertical stand and/or a digital detector in the table you can create your
own 2-in-1 DRF room solution. Set up a radiography and fluoroscopy room
that perfectly fits your needs – from a single detector version to a high-end
full digital room solution with all advantages of filmless and even cassetteless
workflow.
�
Content
04 Main components and options
06 Fluoroscopy unit EasyDiagnost Eleva
10 DoseWise
11 X-ray generation
13 Image detection
14 Image display
15 Extended Digital Imaging
19 Ceiling suspension CS
20 Over-table technique
22 Digital vertical stand VS
24 Vertical stand VS advanced
26 Eleva workspot
27 Height adjustable trolley TA-M
29 Image plate reader PCR Eleva S Plus
30 DRF room solution – Configuration proposals
32 Room planning
35 General
�
Main components and options*
EasyDiagnost Eleva ✓
Spot-film device (digital camera) ✓
Under-table X-ray tube assembly ✓
Tube lift ❍
Image detection, incl. image intensifier and TV camera ✓
Image display, incl. console Eleva Examination Control ✓
LCD monitor in examination room ✓
Monitor trolley for LCD monitor ✓
Ceiling suspension for one or two LCD monitors ❍
Extended digital imaging, incl. image processing and storage with ViewForum ✓
Angiography (with Bolus Chase) ❍
Footswitch for fluoroscopy and exposure ✓
Accessories ✓/❍
DoseWise ✓/❍
High voltage generator Velara RF ✓
DICOM Worklist Management ❍
DICOM MPPS ❍
DICOM image export ✓
DICOM print ✓
DICOM Media on CD-R ❍
mShield ✓
Bucky tray ❍
EasyLat ❍
Digital in-table detector ❍
� ✓ = standard ❍ = optional
Main components and options
* Please contact your local sales specialist for additional system extensions and accessories
�
Main components and options
Ceiling suspension with X-ray tube assembly with control grip and collimator (CS2/4)
❍
Digital vertical stand VSDigital flat detectorMotorized tilting – 20º to +90ºRemovable gridAutomatic exposure control (AEC)� AEC measuring fields
❍
✓
✓
✓
✓
✓
Vertical stand VS advanced tiltCassette trayMotorized tilting – 20º to +90ºRemovable gridAutomatic exposure control (AEC)� AEC measuring fields
❍
✓
✓
✓
✓
✓
Vertical stand VS advancedCassette trayRemovable gridAutomatic exposure control (AEC)� AEC measuring fields
❍
✓
✓
✓
✓
Height adjustable trolley TA-M ❍
Eleva workspot for displaying and processing of digital flat detector images ❍
UNIQUE image processing ✓
Reject analysis ❍
PCR Eleva S Plus ❍
Integration in existing PCR Eleva system ❍
iSite integration ❍
Philips CAD Chest solution xLNA ❍
✓ = standard❍ = optional
�
The EasyDiagnost Eleva stand
consists of:
• Table
• Cassette tray, digital detector unit or
EasyLat (option)
• Spot-film device (Digital Camera)
• Compressor
• Tube assembly
• Tube lift (option)
Table
DimensionsTable heightTable top
8�0 mm2,000 mm x �90 mm
(�2.7")(78.7" x 27.2")
✓
✓
Table tiltTilt angle
Speed
90° / –20°90° / –�0°90° / –��°90° / –8�°1°/s to �°/s variable
✓
❍
❍
❍
✓
Table top
MovementsLateralLongitudinal
–100 mm to +90 mm8�0 mm
(–�.9" to +�.�")(��")
✓
✓
Movement speedLateralLongitudinalwith bolus chase acquisition
�2 mm/s�0 mm/s0–1�0 mm/s
(1.7"/s)(2.�"/s)(0–�.9"/s)
✓
✓
✓
X-ray attenuation aquivalent0.7 mm Al typical(at 100 kV, 2.7 mm Al HVL)
✓
Maximum patient weightfor all movementsstatic, in horizontal or verticaltable position (incl. footrest)
180 kg2�0 kg
(�00 lbs)(��0 lbs)
✓
✓✓ = standard❍ = optional
Fluoroscopy unitEasyDiagnost Eleva
7
2115
(6'
11.
3")
2950
(9'
8.1
")
1558 (5' 1.3")
1775
(5'
9.9
")
1500 (4' 11.1")1075 (3' 6.3")
554
(21.
8")
2399
for
23 c
m I.
I. (7
' 10.
5" fo
r 9"
I.I.)
2390
for
38 c
m I.
I.
(7' 1
0.1"
for
15"
I.I.)
2950
(9'
8.1
")
All dimensions in mm (feet/inches)
750 (29.5")
340 (13.4") 520
(20.5")
1340 (4' 4.8")
620 (24.4")
film
38 cm I.I. (15")
975 (38.4") 1
2000 (6' 6.7")
835 (32.9")
835 (32.9")
690
(27.
2")
100
(3.9
")85
(3
.3")
554
(21.
8")
833
(32.
8")
max
. 210
4 (m
ax. 6
' 10.
8")
626
(24.
6")
155
(6.1
")49
0(1
9.3"
)
512
190
750
(29.
5")
140
1075 (3' 6.3") 885 mm (34.8") – with EasyLat
108.5 108.5 (4.3")
1000 (3' 3.4")
1
2115
(6'
11.
3")
2950
(9'
8.1
")
1558 (5' 1.3")
1775
(5'
9.9
")
1500 (4' 11.1")1075 (3' 6.3")
554
(21.
8")
2399
for
23 c
m I.
I. (7
' 10.
5" fo
r 9"
I.I.)
2390
for
38 c
m I.
I.
(7' 1
0.1"
for
15"
I.I.)
2950
(9'
8.1
")
2115
(6'
11.
3")
2950
(9'
8.1
")
1558 (5' 1.3")
1775
(5'
9.9
")
1500 (4' 11.1")1075 (3' 6.3")
554
(21.
8")
2399
for
23 c
m I.
I. (7
' 10.
5" fo
r 9"
I.I.)
2390
for
38 c
m I.
I.
(7' 1
0.1"
for
15"
I.I.)
2950
(9'
8.1
")
8
Spot-film device (Digital Camera)
MovementsLateralParkingLongitudinal
220 mm�20 mm7�0 mm
(8.7")(20.�")(29.�")
✓
✓
✓
Gridwithout tube lift option
with tube lift option
�0 l/cmr = 10:1, fo = 800 mm�0 l/cmr = 10:1, fo = 900 mm
✓
✓
Grid parking Automatic movement to parking position ✓
Compressor
Compression cone Integrated in spot-film device ✓
Compression movement �0� mm (12") ✓
Compression stop 10 steps, manually selectable ✓
Compression pressure servo supported ✓
Tube lift (option)
Speed �0 mm/s (2"/s) ✓
Focus – Table top distance �10 mm to ��0 mm (20" to 2�.�") ✓
✓ = standard❍ = optional
Fluoroscopy unitEasyDiagnost Eleva (contd.)
9
750 (29.5")
340 (13.4") 520
(20.5")
1340 (4' 4.8")
620 (24.4")
film
38 cm I.I. (15")
975 (38.4") 1
2000 (6' 6.7")
835 (32.9")
835 (32.9")
690
(27.
2")
100
(3.9
")85
(3
.3")
554
(21.
8")
833
(32.
8")
max
. 210
4 (m
ax. 6
' 10.
8")
626
(24.
6")
155
(6.1
")49
0(1
9.3"
)
512
190
750
(29.
5")
140
1075 (3' 6.3") 885 mm (34.8") – with EasyLat
108.5 108.5 (4.3")
1000 (3' 3.4")
1
All dimensions in mm (feet/inches)
750 (29.5")
340 (13.4") 520
(20.5")
1340 (4' 4.8")
620 (24.4")
film
38 cm I.I. (15")
975 (38.4") 1
2000 (6' 6.7")
835 (32.9")
835 (32.9")
690
(27.
2")
100
(3.9
")85
(3
.3")
554
(21.
8")
833
(32.
8")
max
. 210
4 (m
ax. 6
' 10.
8")
626
(24.
6")
155
(6.1
")49
0(1
9.3"
)
512
190
750
(29.
5")
140
1075 (3' 6.3") 885 mm (34.8") – with EasyLat
108.5 108.5 (4.3")
1000 (3' 3.4")
1
10
DoseWise
✓ = standard❍ = optional
Less radiation time
FluoroGrab ✓
Dynamic FluoroGrab ✓
Preset collimation ✓
Collimation on LIH ❍
Automatic parameter setting ✓
High resolution LCD monitor (flicker-free) ✓
More awareness
Dose calculation ✓
Dose measurement ❍
Dose display ✓
Dose reporting ✓
X-ray disable button ✓
X-ray shieldingfixed and flexible lead curtains againstscattered radiation
✓
Smart beam
Pulsed Fluoroscopy (PF), In-pulse controlled ❍
Grid Controlled Fluoroscopy (GCF), In-pulse controlled
❍
Intelligent Exposure (IQX), In-pulse controlled ✓
SpectraBeam RF ✓
Adaptive Measuring Field ✓
Tube lift ❍
Iris shutter and two pairs of parallel shutters ✓
Parkable Grid (in spot-film device) ✓
✓ = standard❍ = optional
11
X-ray generation
Generator Velara RF
Power pack �0 kW�� kW80 kW
✓
❍
❍
Tube support SRO 2��0SRM 22�0 GSSRO ��100 (for CS)
✓
❍
❍
Exposure techniques • Manual: kV-mAs or kV-mA-s• Automatic Exposure Control (AEC), based on
falling load and fixed current techniques• Intelligent Exposure (IQX), In-Pulse controlled• Automatic kV reduction techniques• Tomo Density Control (TDC)
✓
✓
✓
❍
❍
Fluoroscopy techniques • Automatic kV/mA control during fluorscopy• Pulsed Fluoroscopy (PF), In-pulse controlled• Grid Controlled Fluoroscopy (GCF),
In-pulse controlled
✓
❍
❍
Tube voltage exposureTube voltage fluoroscopy
�0 to 1�0 kV�0 to 110 kV
Tube current exposureTube current fluoroscopy
1 to 1100 mA0.2 to � mA
depending on power pack
mAs product 0.� to �00 mAs0.� to 8�0 mAs
with AECwithout AEC
Exposure times 1 ms to � s with AEC
Access times (tube dependent)Standby to fluoroscopyFluoroscopy to radiographyRadiography to fluoroscopy
< 0.� s< 0.8 s< 0.8 s
Dose managementDose calculationDose measurement
Entrance dose, entrance dose rate, dose-area productDose-area product
✓
❍
Safety Tube overload protectionAutomatic mains voltage compensation
✓
✓
✓ = standard❍ = optional
12
Tube SRO 2550 SRM 2250 GS SRO 33100 (for CS)
Focal spot 0.� / 1.0 0.� / 1.0 0.� / 1.0
Anode diameter 90 mm 100 mm 90 mm
Anode angle 1�º 1�º 1�º
Anode heat storage capacity �00 kHU (220 kJ) �80 kHU (280 kJ) �80 kHU (280 kJ)
Maximum voltage 1�0 kV 12� kV 1�0 kV
Max. continuous heat dissipation 2�0 W �00 W 2�0 W
Max. heat content of assembly 1700 kHU (12�0 kJ) 20�0 kHU (1�10 kJ) 1700 kHU (12�0 kJ)
Grid switch No Yes No
Collimator for Under-table technique Over-table technique
Collimator controlBuilt-in -controller and motor controllersPower-up self testAuto calibration at power-upError logging, reporting
✓ ✓
SpectraBeam RFProgrammable filter pre-setsMax. number of pre-filters
✓
�✓
�
Rectangular collimationMotor drivenTwo pairs of parallel shuttersMax. opening �1 cm (20")Positioning accuracy 0.�% of SID
✓ ✓
Circular collimationOctagonal ShapeMotor drivenPositioning accuracy 0.�% of SID
✓ –
Collimation on LIHFor regular shuttersOn last fluoroscopy imageOn last exposure image
✓
✓
✓
–––
X-ray generation (contd.)
1�
Image Intensifier (I.I.)
Mode2�0 mm, 170 mm, 1�0 mm �10 mm, 2�0 mm, 200 mm, 170 mm�80 mm, �10 mm, 2�0 mm, 170 mm
(9", 7", �.�") or(12", 10", 8", 7") or(1�", 12", 10", 7")
Input window Titanium
Detective Quantum Efficiency ��%
Modulation Transfer Factor ��% at 1 lp/mm in �8 cm (1�'') mode
Central spatial resolution �80 mm (1�'') mode:170 mm (�.7'') mode:
� lp/mm (1�.7 lp/ '')� lp/mm (2�.� lp/ '')
Conversion factor in �8 cm (1�'') mode,typical
�00 cd/m2/(mR/s); ��,8 cd/m2/(µGy/s)
Structure noise Negligible
Contrast ratio170 mm (�.7'') (10% disc): �80 mm (1�'') (10% disc):
��:122:1
Image detection
Camera
Type CCD
Matrix 102� x 102� and �12 x �12
Dynamic range Between �� and 70 dB
Digital output 12 bit
Horizontal / Vertical scan reverse
Yes
Dose control, automatic In-pulse controlled
Automatic Gain Control In-frame gain control
Measurement fields 2 fields configurable
Application range (dose) � to 10 000 nGy/image
Image display
Monitor
Type LCD (Liquid Crystal Display) ✓
Monochrome display ��0 mm (18'') ✓
Progressive display (flicker free) ✓
Native format 1280 x 102� SXGA ✓
Nominal brightness �00 Cd/m2 ✓
Monitor carriers
Trolley for one LCD monitorfor two LCD monitors
✓
❍
Ceiling suspension for one LCD monitorfor two LCD monitors
❍
❍
✓ = standard❍ = optional
1�
1�
Extended Digital Imaging
Fluoroscopy
Digital high resolution continuous fluoroscopy ✓
Grid Controlled Fluoroscopy: 0,�-�0 pulses/sec (�0/�0Hz) ❍
Pulsed fluoroscopy, 0,�-� pulses/sec (�0/�0Hz) ❍
FluoroGrab ✓
Dynamic FluoroGrab (up to �0 images/sec) ✓
Fluoro images freeze (last image hold) ✓
Trace subtract fluoroscopy ❍
Dynamic noise reduction ✓
Acquisition
First time right IQ with harmonization and auto windowing ✓
Matrix size 102�2 and �122 ✓
Acquisition speed: up to 8 images/sec (102�2 and �122) ✓
Acquisition speed: up to �0 images/sec (�122) ❍
Adjustable acquisition speed (during run) ✓
Live subtraction ❍
Standard acquisition RAM capacity (0.� GB) ✓
Max. acquisition RAM capacity (1 GB) ❍
External video display ✓
Bolus chase (non subtracted) ❍
✓ = standard❍ = optional
1�
Image processing
Harmonization ✓
Processing depth (bit) 1�
Video invert ✓
Zoom variable up to x �.7 ✓
Magnifying glass and panning ✓
Manual adjustable contrast/brightness/edge enhancement ✓
Electronic shutters (automatic & manual) ✓
Flip/rotate ✓
Annotation 'free text and advanced' ✓
Measurements: length, angle and area ✓
Vascular image processing
Subtract (single mask) ❍
Run subtract (multiple mask) ❍
Measurements extended: area, histogram, vessel analysis (automatic) ❍
Re-masking ❍
Pixel shift ❍
Split screen pixel shift ❍
Auto warp (automatic pixel shift) ❍
Land marking/road mapping: acquisition ❍
View trace (max. opacification) ❍
CO2 imaging (min. opacification) ❍
✓ = standard❍ = optional
Extended Digital Imaging (contd.)
17
Viewing
View switch time between patients stored in RAM: 0.1 sec ✓
View switch time between runs stored in RAM (within one patient): 0.1 sec ✓
Dynamic view & run cycle: upt to �0 fps ✓
Overview (user selectable number of images) ✓
Pictorial index ✓
Parallel working ❍
Multimodality viewing (CT/MR/…) in control- and exam room ❍
Direct Mouse Manipulation (DMM) ✓
Default display protocols ✓
User log-on ✓
Image storage
Hard disk ✓
CD-recordable (background storage and retrieval) ❍
Flagging for selective storing, printing and export of images ✓
(Automatic) export of image to network servers (e. g. PACS) ❍
Image reconstruction
ColonMap (complete abdominal overview) ❍
Bolus Chase Reconstruction (complete overview of peripheral vasculature) ❍
✓ = standard❍ = optional
18
Printing
Direct HCU printing (point to point in background) ❍
DICOM Print SCU ✓
Paper printing ✓
Manual compose print (user dependent layout) ✓
Protocol print (user selects layout from a list) ✓
Auto protocol print ✓
Print protocol editor ❍
Automatic annotation of hardcopies ✓
Networking
DICOM store SCP ✓
DICOM store SCU/SCP ❍
DICOM storage commit SCU ❍
DICOM query/retrieve SCU/ SCP ❍
DICOM worklist management SCU ❍
DICOM MPPS SCU ❍
DICOM presentation state ✓
DICOM grey scale display standard ✓
Multitasking (background communication) ✓
Serviceability❍
Remote service support ✓
Digital image quality measurement tool ✓
✓ = standard❍ = optional
Extended Digital Imaging (contd.)
19
The ceiling suspension CS2 / CS4
for over-table technique consists of:
• Column
• Rail system
• X-ray tube assembly with collimator
and control grip
Ceiling suspension CS2 / CS4 option
Column �-part Al telescopic column;adaptable to individual room heights
Weight-balanced vertical movement 1,�80 mm (�8.�") ✓
Format-sensing to control the automatic collimator
✓
Tube support SRO 2��0 (see page 1�)SRO ��100 (see page 1�)
✓
❍
Longitudinal travelwith extension rails
�,��0 mm�,0�0 mm
(1�9")(2��")
✓
❍
Transverse movementCS2CS�
1,�7� mm�,19� mm
(�0")(118")
✓
✓
Longitudinal/transverse centering up to 7 lock-in positions can be set
Tube assembly rotationaround horizontal axisaround vertical axis
12�º+1��º / –180º
✓
✓
Laser slit light ✓
Ceiling suspension CS
✓ = standard❍ = optional
brake rail
max
. cei
ling
heig
ht 3
110
(10'
2.4
") * min
. 125
0 (4
' 1.2
")m
in. 1
235
(4' 0
.6")
(w
ithou
t ß-
stop
)
mov
emen
t ra
nge
0...1
480
(4' 1
0.3"
)m
ovem
ent
rang
e 0.
..149
5 (4
' 10.
9")
(with
out
ß-st
op)
max
. 273
0 (8
' 11.
5")
380
(15"
)
* for max. maneuverability
All dimensions
in mm
(feet/inches)
20
Over-table technique
EasyLat
Type Fold-out cassette holder forhigh quality lateral exposures
Amplimat measuring chamber ✓
Cassette-size sensing ✓
Grid 70 l/cm, r = 1�:1,f0 = 110 cm, stationary
✓
Cassette formatsfold-in
fold-out
1� cm x 18 cm to �� cm x �� cm(�" x 7" to 1�" x 17")�� cm x �� cm to �� cm x �� cm(1�" x 1�" to 1�" x 17")
✓ = standard❍ = optional
900 (35.4")
min
. cei
ling
heig
ht 2
720
(8' 1
1.1"
) *
min
. 860
(33
.9")
mov
emen
t ra
nge
0...1
480
(4' 1
0.3"
)
mov
emen
t ra
nge
0...1
495
(4' 1
0.9"
) (w
ithou
t ß-
stop
)m
in. 8
45 (
33.3
")
(with
out
ß-st
op)
max
. 234
0 (7
' 8.1
")38
0(1
5")
1200 (3' 11.2")
510
(20.
1")
1500 (4' 11.1")2653 (8' 8.5")
3000 (9' 10.1")
2000 (6' 6.7")1322 (4' 4.1") 1678 (5' 6.1")
1875 (6' 1.8")1090 (3' 6.9")
588(23.1")
375(14.8")
72(2.8")
* for max. maneuverability
All dimensions in mm (feet/inches)
Bucky unit in table
Amplimat measuring chamber ✓
Automatic cassette size sensing ✓
Removable grid �� l/cmr = 12:1 or 8:1,f0 = 100 cm, oscillating(additional grid types are optional available)
✓
Cassette formats 1� cm x 18 cm to �� cm x �� cm(�'' x 7'' to 1�'' x 17'')
✓ = standard❍ = optional
Digital detector unit
Amplimat measuring chamber ✓
Preset collimation ✓
Exam-type dependent collimation presets ✓
Type Flat detector made of amorphous silicon
Detector size �� cm x �� cm (17" x 17")
Image matrix size �,000 pixels x �,000 pixels
Pixel size 1�� m
Image resolution up to �.� Lp/mm
Dimension �9� mm x �7� mm x 18� mm
(2�.�" x 22.�" x 7.2")
Removable grid �� lines/cm, r = 8, f0 = 1�00 mm (�'7.1")
(additional grid types are optional available)
✓ = standard❍ = optional
21
Digital vertical stand VS
Column
Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit
Vertical movement range �00 mm to 1,800 mm (center beam) (11.8" to �’11")
Installation Floor and wall attachment or floor only (option)
Grips
Two patient grips On left and right of the detector unit
Patient stretch grip (option)
• Arranged on left or right of the detector unit• Rotatable
Brakes
Brakes Electromagnetic, self-locking
22
Detector unit
Type Flat detector made of amorphous silicon
Detector size �� cm x �� cm (17" x 17")
Image matrix size �,000 pixels x �,000 pixels
Pixel size 1�� m
Image resolution up to �.� Lp/mm
Dimension �9� mm x �7� mm x 18� mm (2�.�" x 22.�" x 7.2")
Angle of tilt – 20º to +90º, motorized tilting
Weight approx. 2�.� kg (�8.� lbs)
Removable grid �� l/cm, r = 8, f0 = 1�00 mm (�'7.1")(additional grid types are optionally available)
Grid storage For up to 2 grids within the detector unit
2�
58 (2.3")
102 (4")
575
(22.
6")
406 (16")710 (28")
825
(32.
5")
1173
(3'
10.
2")
1819
(5'
11.
6")
596 (23.5") 227(8.9")
1990
(6'
6.4
")20
80 (
6' 9
.9")
≤150
0 (4
' 11.
1")
≥300
(11.
8")
≤210
5 (6
' 10.
9")
417 (16.4")
dete
ctor
pla
ne
Buck
y un
it fr
ont
side
596
54 (2.1")
902
(35.
5")
1286 (4' 2.6")
dete
ctor
pla
ne
Buck
y un
itfr
ont
side
596
(23.
5")
54 (2.1")
300(11.8")
902
(35.
5")
with spacer (option)
spacer
64 (2.5")107 (4.2")
1990
(6'
6.4
")20
80 (
6' 9
.9")
986 (38.8")
596
(23.
5")
902
(35.
5")
713 (28.07")
20°
≥300
(11.
8")
≤150
0 (4
' 11.
1")
≤210
5 (6
' 10.
9")
≥550
(21
.7")
dete
ctor
pla
ne
Buck
y un
itfr
ont
side
54 (2.1")
with tilting
with tilting
58 (2.3")
102 (4")
575
(22.
6")
406 (16")710 (28")
825
(32.
5")
1173
(3'
10.
2")
1819
(5'
11.
6")
596 (23.5") 227(8.9")
1990
(6'
6.4
")20
80 (
6' 9
.9")
≤150
0 (4
' 11.
1")
≥300
(11.
8")
≤210
5 (6
' 10.
9")
417 (16.4")
dete
ctor
pla
ne
Buck
y un
it fr
ont
side
596
54 (2.1")
902
(35.
5")
1286 (4' 2.6")
dete
ctor
pla
ne
Buck
y un
itfr
ont
side
596
(23.
5")
54 (2.1")
300(11.8")
902
(35.
5")
with spacer (option)
spacer
The digital vertical stand VS
consists of:
• Column
• Detector unit
• Grid
• 2 user interfaces (left & right)
• 2 patient grips (left & right)
All dimensions
in mm (feet/inches)
2�
Vertical stand VS advanced
Column
Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit
Vertical movement range �00 mm to 1,800 mm (center beam) (11.82" to �'10.87")
Installation Floor and wall attachment or floor only (option)
Grips
Two patient grips On left and right of the detector unit
Patient stretch grip(option)
• Arranged on left or right of the detector unit• Rotatable
Brakes
Brakes Electromagnetic, self-locking
Vertical stand VS advanced
with/without tilt
Detector unit (cassette tray)
Dimensions (w x h) �9� mm to �7� mm (2�.�" to 22.�")
Angle of tilt (option) – 20º to +90º, motorized tilting
Automatic exposure control (AEC)
� AEC measuring fields
Operating 2 control units (left & right) and wireless remote control
Removable grid �� l/cm, r = 8, f0 = 1�00 mm(additional grid types are optionally available)
Grid storage For up to 2 grids within the detector unit
Cassette formats 18 cm x 2� cm to �� cm x �� cm(8" x 10" to 1�" x 17")
2�
Column
Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit
Vertical movement range �00 mm to 1,800 mm (center beam) (11.82" to �'10.87")
Installation Floor and wall attachment or floor only (option)
Grips
Two patient grips On left and right of the detector unit
Patient stretch grip(option)
• Arranged on left or right of the detector unit• Rotatable
Brakes
Brakes Electromagnetic, self-locking
All dimensions in mm (feet/inches)
58 (2.3") 102 (4")
575
(22.
6")
406 (16")710 (28")
825
(32.
5")
1173
(3'
10.
2")
1819
(5'
11.
6")
596 (23.5") 227 (8.9")
160(6.3")
1990
(6'
6.3
")20
80 (
6' 9
.9")
≤150
0 (4
' 11.
1")
≥300
(11.
8")
≤210
5 (6
' 10.
9")
537 (21.1")
dete
ctor
pla
ne
Buck
y un
it fr
ont
side
596
(23.
5")
54 (2.1") 54 (2.1")
902
(35.
5")
1281 (4' 2.4")
dete
ctor
pla
ne
Buck
y un
itfr
ont
side
596
(23.
5")
300(11.8")
902
(35.
5")
with spacer (option)
spacer
58 (2.3")102 (4")
1990
(6'
6.3
")20
80 (
6' 9
.9")
981 (38.6")
596
(23.
5")
120
(4.7
")
902
(35.
5")
708 (27.9")
20°
≥300
≤150
0 (4
' 11.
1")
≤210
5 (6
' 10.
9")
≥549
(21.
6")
dete
ctor
pla
ne
Buck
y un
itfr
ont
side
54 (2.1")
with tilting (option)
with tilting (option)
The vertical stand VS advanced
consists of:
• Column
• Detector unit (cassette tray)
• Grid
• 2 user interfaces (left & right)
• 2 patient grips (left & right)
2�
Eleva workspot
Computer
Hard disk 1�0 GB (equivalent to more than �,000 images)
RAM storage capacity 2 GB
Operating frequency 2 GHz
Interfaces • Gigabit Ethernet / DICOM interface• Detector interface
Image data
Data volume Up to 18 MB/image
Matrix depth 1� bit/pixel
Console
Monitor • 17"-TFT display with resolution 1,280 x 1,02�• 19"-TFT display with resolution 1,280 x 1,02� (optional)• Touch screen operation
Keyboard with mouse and function buttons
For entering administrative patient data and for operating the screen menus
Dimensions A B C
Touch monitor 17" ��� mm ��0 mm 2�� mm
(17.1") (1�.9") (9.�")
Touch monitor 19" (optional)
��� mm ��7 mm 2�1 mm
(17.2") (18.0") (9.�")
PC 21� mm ��8 mm 27� mm
(8.�") (17.�") (10.8")
C
C
B
B
A
A
• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •
27
Height adjustable trolley TA-M
Trolley
Height of table adjustment �00 mm to 870 mm (2�.�" to ��.�")
Dimensions with accessories, max. 2,270 mm x 900 mm (7' �.�" x ��.�")
Total weight 1�0 kg (28� lbs)
Accessories All accessories for the table TH-S can be used with this trolley as well
Brakes At each side
Table top
Type Floating table top of sandwich design with carbon fiber overlay
Dimension table top (l x w)Standard versionBariatric version
2,200 mm x �20 mm2,200 mm x �70 mm
(7' 2.�" x 2�.�")(7' 2.�" x 2�.�")
X-ray transparent areaStandard versionBariatric version
1,7�0 mm x �20 mm1,7�0 mm x �70 mm
(�' 8.1" x 2�.�")(�' 8.1" x 2�.�")
Thickness of table top �� mm (2.1")
Range of horizontal movementstransverselongitudinal
100 mm (�.�")floating by fixing the vertical axes of the wheels
Attenuation equivalent 1.2 mm Al equivalent at 100 kV
Patient load 22� kg (�9� lbs)
28
2270 (7' 5.4")
620
(24.
4")1)
100
(3.9
")10
0(3
.9")
600
(23.
6")
870
(34.
3")
890
(35"
)
54 (
2.1"
)
1) Bariatric version 670 mm (26.4")
All dimensions in mm (feet/inches)
Height adjustable trolley TA-M (contd.)
Image plate reader PCR Eleva S Plus
590 (23.3")
810
(31.
9")
43 (
1.7"
)
380 (15.0")
Image plate reader PCR Eleva S Plus
Type The image plate reader PCR Eleva S Plus consists of• Cassette feeding unit for the cassette,• Reader unit.
Plate througput Up to 97 plates/h (18 cm x 2� cm), see table on page 1� (high-speed mode allowing up to 9� plates/h for �� cm x �� cm)
Cassette release �7 s to �� s, depending on cassette size and type
Cassette feeding unit• Removal of exposed image plate from the cassette• Reading of the image plate (cassette) barcode• Transport of the exposed image plate to and from the
reader unit• Loading the cassette with the erased plate• Release of newly loaded cassette
Reader unit:Scanning resolution
Read functionErase function
100 m or 200 m(200 m through either software re-sizing or high-speed mode)�0 m (Option. Not available in the US)12 bit analog/digital conversion of informationExposure-dependent erasure of residual information on the image plate using white light
All dimensions in mm (inches)
29
DRF room solution – Configuration proposals
Configuration 1
EasyDiagnost Eleva with Bucky tray or EasyLat• Ceiling suspended tube carrier (CS2/�)• Tiltable digital vertical stand VS
Configuration 2
EasyDiagnost Eleva with digital detector in table• Ceiling suspended tube carrier (CS2/�)
Configuration 3
EasyDiagnost Eleva with digital detector in table• Ceiling suspended tube carrier (CS2/�)• Tiltable digital vertical stand VS
�0
Optional available*:
• CR-integration
• Height adjustable trolley TA-M
• Vertical stand VS advanced
with/without tilt
* Please contact your local sales specialist for additional configurations and accessories
�1
�2
Room planning
12
1150
280 91
31 28
10502040
31005500
240
824
24
14
22
2321
32
29
600
1200
1000
30
5050
9
10
8
34
34
6
7
1
33
25
13
2726
16 3
2
18
20
11
17
19
54
35 EasyDiagnost Eleva
Key
1 DI cabinet
2 Image Receiver Module
� Eleva examination control (EEC)
� Keyboard
� Mouse
� Generator cabinet (Velara)
7 EasyDiagnost Eleva main cabinet
8 DI cabinet rear cover
9 Generator cabinet rear cover
10 Main cabinet rear cover
11 System controller (Sysca) PC
12 Indication box (IBO)
Key (contd.)
1� Acquisition work spot
1� Infra red receiver
1� Slave monitor (18")
1� Monitor Acquisition work spot (18")
17 ViewForum computer
18 Mouse
19 Keyboard
20 Viewing monitor (18") (maybe combined with �WM [shared made])
21 Footswitch
22 Monitor carriage
2� Tilting axis
Key (contd.)
2� Center of table top
2� Speed control
2� Handswitch colon map
27 Exposure handswitch
28 Area for service purposes
29 Space required for installing the X-ray module
�0 Space required for inserting the table top (alternatively: 2.0� m at the head end in 0° position)
�1 If necessary, order an insulating kit for ceiling rails locally
Key (contd.)
�2 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision
�� For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)
�� Min. space �–7 mm between 2 Cabinet Rear Cover
�� EasyDiagnost Eleva
Layout proposal for EasyDiagnost Eleva with under-table technique (single plane)
All dimensions in mm (feet/inches)
��
Room planning
�3220�6360
3140*
32
15
54SID max. 1800
835
16
985.5 � 7.5
29
1000 1000
28
35
6
835
31
1935
1500
*
5800 17
18
19
479
11
730
27
2
23
24
22
5
20
21
25
26
14
4
3
13
10
12
33
8
9
1
34
34
36 EasyDiagnost Eleva
Layout proposal for EasyDiagnost Eleva with over-table technique (second plane)
Key
1 DI cabinet
2 Image Receiver Module
� Eleva examination control (EEC)
� Keyboard
� Mouse
� Ceiling rails ceiling suspension
7 Ceiling rails MCS
8 Generator cabinet (Velara)
9 EasyDiagnost Eleva main cabinet
10 DI cabinet rear cover
11 Monitor ceiling suspension
Key (contd.)
12 Generator cabinet rear cover
1� Main cabinet rear cover
1� System controller (Sysco) PC
1� Indication box (IBO)
1� BuckyDiagnost VS advanced tilt
17 Infra red receiver
18 Examination monitor (18")
19 Reference monitor (18")
20 Slave monitor (17")
21 ViewForum computer
22 Mouse
2� Keyboard
Key (contd.)
2� Viewing monitor (18") (maybe combined with �WM [shared made])
2� Speed control
2� Handswitch colom map
27 Exposure handswitch
28 Footswitch
29 Ref. cross point
�0 If necessary, order an insulating kit for ceiling rails locally
�1 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision
Key (contd.)
�2 SID max. 1800 mm with table in upright position
�� For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)
�� Min. space �–7 mm between 2 Cabinet Rear Cover
�� Tube in position “free exposure”
�� EasyDiagnost Eleva
All dimensions in mm (feet/inches)
��
Key
1 DI cabinet
2 Image Receiver Module
� Eleva examination control (EEC)
� Keyboard
� Mouse
� Ceiling rails ceiling suspension
7 Ceiling rails MCS
8 CS
9 Generator cabinet (Velara)
10 EasyDiagnost Eleva main cabinet
11 DI cabinet rear cover
Room planning
Layout proposal for DRF room solution with digital vertical stand (minimum room layout without applicational restrictions)
Key (contd.)
12 Monitor ceiling suspension
1� Generator cabinet rear cover
1� Main cabinet rear cover
1� System controller (Sysco) PC
1� Indication box (IBO)
17 Acquisition work spot
18 BuckyDiagnost VS digital with motorized tilting
19 Infra red receiver
20 Examination monitor (18")
21 Reference monitor (18")
22 Monitor Acquisition work spot (18")
Key (contd.)
2� ViewForum computer
2� Mouse
2� Keyboard
2�Viewing monitor (18")(maybe combined with �WM) [shared model])
27 Footswitch
28 Speed control
29 Handswitch colon map
�0 Exposure handswitch
�1 If necessary, order an insulating kit for ceiling rails locally
Key (contd.)
�2 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision
�� External rails (optional)
�� Scopomat in work position
�� For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)
�� Min. space �–7 mm between 2 Cabinet Rear Cover
�7 EasyDiagnost Eleva
All dimensions in mm (feet/inches)
15
18
8
19 7
12
21
20
29
28
17 22 3
4 5
225
26
24
15
23
31
30
35
36
1 10 9
36
27
32
336
11 1314
� 6100
5800
37 EasyDiagnost Eleva
��
General
Power supply Digital verticalstand VS
Eleva workspot EasyDiagnost Eleva Eleva Examination Control
Generator 50 kW
Mains voltage 2�0 V ±10% 100 V to 2�0 V ±10% 2�0 V ±10% 2�0 V ±10%
�00 V ±10%; � phase
Mains frequency �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz
Current input max. 1.7 A max. � A max. 7.� A max. � A max. 1�� A
Fuse �A 10 A 10 A 10 A � x �� A
Heat dissipation (average)
100 W 12� W �00 W - max. 1�00 W �0 W �00 W
Safety Tube overload protection
Automatic mains voltage compensation
Ambient conditions
Working temperature +18˚C to +�0˚C
Storage temperature –10˚C to +��˚C
Rel. humidity (in operation) 20% to 7�% (without condensation)
Rel. humidity (storage) 10% to 9�%
Air pressure 700 hPa to 10�0 hPA
Conformity
The DRF system complies with the provisions of the Medical Device Direktive 9�/�2 EEC (9�) and with IEC, FDA, PAL and CSA standards.
© 2007 Koninklijke Philips Electronics N.V.All rights are reserved.
Philips Medical Systems DMC GmbH reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.
Printed in The Netherlands.��12 9�2 291�1/70� * NOV 2007
Philips Medical Systems is part of
Royal Philips Electronics
Interested?
Would you like to know more about our imaginative
products? Please do not hesitate to contact us.
We would be glad to hear from you.
On the web
www.medical.philips.com
Via email
By fax
+�1 �0 27 �� 887
By mail
Philips Medical Systems
Global Information Center
P.O. Box 128�
��02 BG Eindhoven
The Netherlands
By phone
Asia
Tel: +8�2 2821 �888
Europe, Middle East, Africa
Tel: +�9 70�1 ��� 22��
Latin America
Tel: +�� 11 212� 07��
North America
Tel: +1 800 229 ��17